Transcript of UV - VIS SW
![Page 1: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/1.jpg)
UV/VIS SOFTWAREGamal A. Hamid
![Page 2: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/2.jpg)
To everyone who has helped us with support, new books, hard / soft ware And over the internet Special thanks
for THERMO
Thanks
![Page 3: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/3.jpg)
3 Gamal A. Hamid
Contents Introduction
Software
Menus
Toolbar
Data Store
Status Bar
Reporting
Applications
![Page 4: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/4.jpg)
Introduction
![Page 5: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/5.jpg)
5 Gamal A. Hamid
UV/VIS An analytical for measure the concentration of
samples or samples components according to
Beer’s lambert Law depending on the ratio
between incident and transmitted light
through the cuvette.
Absorbance is expressed as
A = εbc where ε is a constant called absorptivity
Absorbance is directly proportional to the path
length, b, and the concentration, c, of the
absorbing molecule.
![Page 6: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/6.jpg)
6 Gamal A. Hamid
UV/VIS Region The UV is the electromagnetic
spectrum
shorter than those of ordinary, visible
violet light but longer than those of x-
rays.
The human eye is only sensitive
between approximately 380 and 780
nm and within this area we perceive
the colors of the rainbow from violet
through to red.
![Page 7: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/7.jpg)
7 Gamal A. Hamid
Spectrometer
A spectrophotometer is an instrument for
measuring the transmittance or
absorbance of a sample as a function of
the wavelength of electromagnetic
radiation.
Single beam instrument.
Double beam instrument.
![Page 8: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/8.jpg)
8 Gamal A. Hamid
Instrument Components
1. Light Source
2. Sample Chamber
3. Optical System
4. Detector
![Page 9: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/9.jpg)
Software
![Page 10: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/10.jpg)
10 Gamal A. Hamid
Main window
![Page 11: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/11.jpg)
Menus
11
![Page 12: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/12.jpg)
12 Gamal A. Hamid
Menus
File
View
Tools
Application
Command
Quant
Data Store
Table
Window
The Main Menu Bar gives you access to all of the
operations needed to obtain, process, store and
retrieve data.
Each pull-down menu gives you access to one or
more sub-menus.
The most frequently-used operations may also be
accessed directly via icons on the toolbar.
![Page 13: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/13.jpg)
13 Gamal A. Hamid
File menu
The File Menu is used for all operations connected with creating, opening, saving and printing files.
The File Menu is used for all operations
connected with creating, opening, saving
and printing files.
![Page 14: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/14.jpg)
14 Gamal A. Hamid
View menu
Customize
This opens a tabbed dialog which
gives you control over which
toolbar icons are displayed and
how they are grouped.
You also have some choices about
the appearance of the toolbar
buttons.
![Page 15: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/15.jpg)
15 Gamal A. Hamid
Tools menu
The Tools Menu enables you to configure
the software to suit your preferences.
Options
This item opens a tabbed dialog through
which you can control the following:
File Load/Save Behavior
System Settings
Batch Information Printout
Environment Information
![Page 16: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/16.jpg)
16 Gamal A. Hamid
Application menu
The Applications Toolbar enables you to
move between one application and
another with a single mouse click,
instead of accessing the applications via
the Application Menu.
![Page 17: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/17.jpg)
17 Gamal A. Hamid
Command menu
The available options in the Command
Menu will depend on which instrument
you are using and on the current context.
![Page 18: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/18.jpg)
18 Gamal A. Hamid
Baseline
100%T Baseline
Causes the instrument to do a Baseline
Scan over the Wavelength Range and Data
Interval specified in the current method.
This is used to compensate for introduced
differences between the sample and
reference beams other than the presence
of the sample (e.g., differences in
absorption between the Reference position
or sample and the Sample position or
sample).
![Page 19: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/19.jpg)
19 Gamal A. Hamid
Math menu
The Math Menu contains all the commands
needed to process your data post-run.
If you are going to repeat the manipulation
on a number of batches of results, the
window can be left open and other batches
dragged to it.
The Math Manipulation window will
remain on top on the work area all the
time it is open.
![Page 20: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/20.jpg)
20 Gamal A. Hamid
Quant menu
No curve fit
Select all curve fits
Select Linear fit
Select Linear through zero fit
Select Quadratic fit
Select Quadratic fit through zero
![Page 21: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/21.jpg)
21 Gamal A. Hamid
Data Store menu
The Data Store menu gives you access to
commands that operate on data in the
Data Store.
Not all commands will be applicable in all
circumstances.
Commands not available in the current
context will be grayed out.
![Page 22: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/22.jpg)
22 Gamal A. Hamid
Table- Graph- Report menus
When the Results Table has the
focus Table is added to the Main
Menu Bar.
When the Graph window has the
focus Graph is added to the Main
Menu Bar.
When a Report window exists and
has the focus Report is added to
the Main Menu Bar.
![Page 23: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/23.jpg)
23 Gamal A. Hamid
Window menu
The options in the Window Menu give
you control over the way in which
windows are displayed on the program
work area and enable you to transfer
the focus from one window to another.
Clicking on your selected window will
transfer the focus to that window and
re-open it if that window has previously
been closed or minimized.
![Page 24: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/24.jpg)
Toolbars
24
![Page 25: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/25.jpg)
25 Gamal A. Hamid
The toolbars
The toolbars contain icons for the most
commonly used operations, enabling them
to be accessed with a single mouse click
rather than via the pull-down menus on
the Main Menu Bar.
![Page 26: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/26.jpg)
26 Gamal A. Hamid
The toolbars groups
File operation group
Application group
Analysis group
Window management group
Window group
Graph group
![Page 27: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/27.jpg)
27 Gamal A. Hamid
File operation group
Create a new method in the current application
Load a method for the current application from disk
Save the Sample Method to disk using its current name
Load a result file for the current application from disk
Save the file currently selected in the Data Store to disk using its current name
![Page 28: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/28.jpg)
28 Gamal A. Hamid
Application group
Starts the Scan Application (shows the appearance of the button for the active application)
Starts the Fixed Application
Starts the Quant Application
Starts the Rate Application (available only if VISIONlife is installed)
Starts the MCA (Multicomponent Analysis) Application
Starts the DNA Melting Application (available only if VISIONlife and suitable accessories are
installed)
![Page 29: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/29.jpg)
29 Gamal A. Hamid
Analysis group
Starts a run in the current application using the current Application and Accessory Method
parameters.
Toggle to pause start/stop data collection.
Start a Zero measurement or a baseline scan as applicable using the current Application
Method parameters.
Stop the current run
Access this Help system
![Page 30: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/30.jpg)
30 Gamal A. Hamid
Window management group
Remember the Current Window Layout
Restore the remembered window layout
Arrange the windows as non-overlapping tiles
Cascade the Windows
Load a Window arrangement from disk. (Window arrangements can be saved via the
Save/Window Layout command in the File Menu)
![Page 31: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/31.jpg)
31 Gamal A. Hamid
Window group
Restores the Method window
Restores the Graph window
Restores the Results window
Restores the Calibration window
Restores the Kinetics window
Restores the Data Store
Restores the Report Composer window
![Page 32: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/32.jpg)
32 Gamal A. Hamid
Graph group
Cancels Tracking
Sets tracking
Clears the graph, without clearing the axes (Use the Graph Menu to clear axes)
Steps through grid displays (dots, lines, no grid)
Sets up Axes
Toggles display of the legend
Toggles display of selected labels
Toggles Graph background between white and gray
Toggles Graph Line Type between Solid and Points
![Page 33: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/33.jpg)
Data Store
33
![Page 34: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/34.jpg)
34 Gamal A. Hamid
Data Store
The Data Store lists all of the batches of
results that are currently held in memory.
When a batch of results has been run it will
appear in the Data Store.
When a results file is loaded from disk it
will be placed in the Data Store.
A batch is represented by
A sample is represented by
A cycle is represented by
![Page 35: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/35.jpg)
Status Bar
35
![Page 36: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/36.jpg)
36 Gamal A. Hamid
Status Bar
The following will be displayed on the status bar when appropriate:
The effect of a left mouse-click at the current cursor position (e.g., reminders of the action of toolbar
icons)
Icons for currently active accessories (e.g., sipper, Cell Changer, temperature control, stirrer)
The status of active accessories (e.g., current cell, cell temperature, equilibration status)
The presence of any Performance Verification accessory
Lamp status
The current absorbance reading
The current wavelength
The status of the instrument
Zero or user Baseline status
![Page 37: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/37.jpg)
Reporting
37
![Page 38: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/38.jpg)
38 Gamal A. Hamid
Report The Report Menu only appears on the Main Menu Bar when the Report Composer
window has the focus.
Click on the item you require to select it, Move the cursor into the Report window.
![Page 39: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/39.jpg)
Applications
39
![Page 40: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/40.jpg)
40 Gamal A. Hamid
Applications
Scan
Fixed
Quant
Rate
MCA
DNA
![Page 41: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/41.jpg)
41
1.Scan
Spectrum recording.
The spectrophotometer measures the amount
of light that passes through the sample over a
continuous range of wavelengths and plots a
graph of the resulting spectrum.
The starting and finishing wavelengths of the
scan are both set .
Select the appropriate Baseline Correction you
wish to perform.
Maximum wavelength range is determined by
the working range of the spectrophotometer.
![Page 42: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/42.jpg)
42
Scan Parameters
![Page 43: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/43.jpg)
43
Definitions No. Item Definition
1 Data Mode The Data Mode enables you to select what will be used as the vertical axis when the spectrum is first plotted.
2 Bandwidth The Effective Spectral Slit width of the monochromator.
3 Scan speed A measure of how fast the monochromator changes the wavelength of the light falling on the sample.
4 Data interval The Data Interval is the difference in wavelength between successive measurement points.
5 Lamp changeControl an instrument with a deuterium lamp installed to cover the UV wavelengths and a tungsten lamp to cover the visible range.It will not be available if the instrument has a xenon lamp installed,
6 Cycle Single measurement on a sample
7 Baseline correction The standard User Baseline Correction selection will be the 100%T .
![Page 44: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/44.jpg)
44 Gamal A. Hamid
Display Parameters
Off No results will appear in the Results Table
Manual The Results Table will contain the values of the Y-axis variable at up to 100 specified wavelengths.
Peak PickThe Results Table will contain the wavelengths and Y-axis variable values of up to 100 peaks, valleys or zero-crossings.
Peak Area
The Results Table will report the area of a selected peak.
Peak Height
The Results Table will report the height of a selected peak.
( Scan and MCA) methods – display Parameters – results table, enables you to choose from a number of ways of reporting and at run-time
![Page 45: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/45.jpg)
45
2. Fixed
Recording spectrometer readings at
different dedicated wavelengths and
performing simple calculations
In the Fixed application the instrument
measures the amount of light passing
through the sample at one or more
wavelengths.
Set the wavelengths at which the
instrument will make measurements.
![Page 46: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/46.jpg)
46
Fixed Parameters
![Page 47: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/47.jpg)
47
3. Quantitative
Quantifying sample concentrations via
calibration, based on Lambert- Beers law
The Quantification application is used to obtain
quantitative concentration data on a single
component from absorbance measurements.
A valid calibration must be present before
samples can be measured.
Enter the number of standards that will be used.
Once a calibration has been performed it can be
saved as part of the method and will be available
for use next time the method is loaded.
![Page 48: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/48.jpg)
48
Quant. Parameters
![Page 49: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/49.jpg)
49
4. Rate
The Rate application is used to measure the change
in concentration of a reactant or product with time.
The absorbance of the sample is measured
repeatedly at a fixed wavelength over a period of
time appropriate to the reaction being studied. The
software then offers a variety of methods to
calculate the rate of reaction from the variation of
absorbance with time.
Recording time dependent readings at a fixed
wavelength and is dedicated for measuring rate
curves.
![Page 50: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/50.jpg)
50
Rate Parameters
![Page 51: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/51.jpg)
51
5. MCA
It provides the ability to perform
simultaneous quantitative analysis on
samples for up to 20 components
Standards and components are scanned,
and each data point of the scan
contributes to the calculation.
A series of Standards, in which the
concentrations of the components are
known, is prepared and scanned.
There must be at least as many standards
as there are components to be analyzed.
![Page 52: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/52.jpg)
52
MCA Parameters
![Page 53: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/53.jpg)
53
6. DNA
During a DNA experiment, the sample is
heated and Absorbance v Temperature
data are collected at a suitable
wavelength.
Tm, the "melting point" of the DNA, may
be calculated at run-time, or post-run via
the Math menu.
![Page 54: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/54.jpg)
54
DNA Parameters
![Page 55: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/55.jpg)
Accessories Software
55
![Page 56: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/56.jpg)
56
Rapid Mixing Accessory
The Rapid Mixing accessory provides a very low
dead time of 8 ms, giving you access to reactions
that occur up to 1000 times faster than those
visible to manual mixing experiments.
The Rapid Mixing accessory provides a very low
dead time of 8 ms, giving you access to reactions
that occur up to 1000 times faster than those
visible to manual mixing experiments.
The Anaerobic accessory consists of a manifold,
which is mounted over the base of the drive
syringes and is purged with a steady stream of
inert gas.
![Page 57: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/57.jpg)
57
Stirrer Parameters
Off The stirrer is off
Continuous The cell is stirred
continuously
Off Before Measurement The stirrer is normally
on but is switch off prior to and during
measurement.
A programmable settling time elapses before the
measurement is made.
![Page 58: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/58.jpg)
58
Air-Cooled Peltier
The Air-cooled Peltier Accessory allows the cell to be
thermostatted during measurements. It comprises a
dedicated cell holder and an electronic control device, which
is connected to the PC.
The Air-cooled Peltier accessory delivers reliable temperature
control .
Wider temperature range from 0 to 110 °C
Multiple cell efficiency for higher sample throughput.
Designed for biologically relevant assays that require
temperature control at 25°, 37°, 40° and 50 °C, the Air-cooled
Peltier accessory is ideal for the life science laboratory.
![Page 59: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/59.jpg)
59
Temp. Control Parameters
Method off not selected
Tolerance Target Temperature ±
Tolerance
Start on using block or probe
The Equilibration Time permitted
range and start a measurement.
The Equilibrate Using sample or
sample and reference.
![Page 60: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/60.jpg)
60
Cell Changer
Using a cell changer, you load the system
with a batch of samples before starting
the measurement.
During the measurement the cell changer
moves the required sample automatically
into the sample beam of the
spectrophotometer.
This type of measurement procedure is
advisable for time consuming analyses,
e.g. rate measurements.
![Page 61: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/61.jpg)
61
Cell Changer Parameters
Method off not selected
Cell changer type. When an instrument
with a cell changer installed is connected,
this field will automatically show the
correct type.
Each cell can contain a Sample (green) or a
Reference (red), or not be used.
![Page 62: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/62.jpg)
62
Sipper
In sipper operation the sipper pump
transports the sample to the flow cell
located in the sample beam. Then the
measurement starts automatically.
Therefore, there is no need to change and
clean the cells.
This operation mode is recommended for
high sample throughput in routine
analyses.
![Page 63: UV - VIS SW](https://reader036.fdocuments.us/reader036/viewer/2022081514/589cb0c11a28abbe4a8b5b41/html5/thumbnails/63.jpg)
63
Sipper Parameters
Method off not selected
Sipper Mode. Choose between Sip Only, Sip and
Run, Continuous.
Sample Direction. When Waste is selected the
sample is pumped to waste after it has been
measured. Return sample is returned to the
sample container .
Sample Delay.
The Sample time between the end of pumping
the air gap and the start of sample measurement.
The purpose of the Sample Delay is to allow time
for the sample to settle down in the flowcell.