USP Workshop on Particle Size

download USP Workshop on Particle Size

If you can't read please download the document

Transcript of USP Workshop on Particle Size

  • Welcome to the

    USP Workshop on Particle Size:Particle Detection and Measurement

    Day 2

    December 8-10, 2010

    USP Headquarters, Rockville, MD

  • USP Workshop on Particle Size:

    Particle Detection and Measurement

    December 8-10, 2010; USP Headquarters

    Desmond Hunt, Ph.D.

    USP Senior Scientific Liaison

    Welcome and Opening Remarks

  • USP Workshop on Particle Size:

    Particle Detection and Measurement

    December 8-10, 2010; USP Headquarters

    Session III: USP General Chapters

    Track 1: Parenterals

    Spalding Auditorium

    Track 2: Aerosols

    Briggs/Parker/

    Marshall/Wiley

  • USP Workshop on Particle Size:

    Particle Detection and Measurement

    December 8-10, 2010; USP Headquarters

    Chair: Mr. Scott Aldrich

    Ultramikro

    Session III, Track 1: Parenterals - USP

    General Chapters and

  • Quality Standards for Medicines, Supplements, and Food Ingredients throughout the World

    8.1 Parenterals USP General Chapters /

    Historical Perspective and Current Chapters ReviewScott Aldrich

    USP Workshop on Particle Size:

    Particle Detection & Measurement

  • Outline

    Introduction

    Background

    History and Limits Evolution

    History of liquid particle counting

    Periods of Investigation

    Harmonization of Aug 1 2007

    Current chapters

    Limits according to container volume and method used

    informational chapter official USP 34, 2011

    follows discussion

    Chapter Crossover

    What is Particulate Matter?

    Test Methods

    Light Obscuration Counting and Operations

    Membrane Microscopic Counting and Operations

    Comparison of LO and MM methods

  • Introduction

    Parenteral injections and ophthalmic solutions are sterile

    liquids, with expectations for physical appearance and

    stability.

    All released product must comply with two limit tests for

    particle content:-

    USP General Chapter Injections

    See PF Stimuli Article Vol. 35(5) [Sept.-Oct. 2009], Visible

    Particulates in Injections - A History and a Proposal to Revise USP

    General Chapter Injections

    Must contain low amounts of sub-visible particles

    USP Injections

    USP Ophthalmic solutions

  • History and Limits Evolution

    History of liquid particle counting

    Periods of Investigation1

    Cognitive

    Latent

    Definitive

    Evaluative

    Harmonization

    The Future

    1Floyd Benjamin, IES conference 1990

  • History and Limits Evolution

    Cognitive Period 1905-1948

    1905 First USP reference to parenteral solution as a compendial drug. No references to particulate matter content or clarity.

    1916 NF IV includes 6 monographs for parenterals; however, no mention of appearance.

    aqueous solutions are to be clear, i.e., when observed over a bright light, they shall be substantially free from precipitate, cloudiness or turbidity. Specks or flecks; fibers or cotton hairs, or any undissolved

    In 1942, USP XII and NF VII both attempt to further define the

    published definition.

  • History and Limits Evolution

    Latent Period 1949-1965 No changes in compendial requirements, although many reports in literature regarding medical events from particulate matter injection.

    1949 USP revision committee eliminates the clarity test

    should be exercised in the preparation of injections to

    1965 - First Limit Standards (Canberra, Australia)In 1965, the National Biological Standards Laboratory, Canberra

    -stimulated compendia and regulatory activities in the US, leading to the Definitive Period.

  • Definitive Period I 1966-19731966 FDA symposium on the safety of large volume parenterals1

    1972 Australian comparison of Coulter and HIAC counts for parenteral products2

    Australian and English standards for particulate matter in

    Australia - ten containers must be less than 2x the following limits:

    NMT 100 particles/mL 2m

    NMT 5 particles/mL 20m

    British Std: mean of 5 containers and individual

    1973 Turco and Davis review the clinical consequences of injected particulate matter3

    History and Limits Evolution

    1. FDA - National symposium on safety of large volume parenteral solutions. U.S. Dept. of Health, Education and Welfare,

    Washington DC, 1966.

    2. Kendall and Tilley, Comparison of Coulter counter and HIAC counts for parenteral products

    3. Turco S. Davis NM. Hosp Pharm. 1973;8:137-40.

  • History and Limits Evolution

    Definitive Period II 19741974 Revision of Foreign Particulate Standards

    Great BritainNMT 1000 particles/mL 2m

    NMT 100 particles/mL 5m

    AustraliaNMT 1000 particles/mL 2m

    NMT 250 particles/mL 3.5m

    NMT 100 particles/mL 5m

    NMT 25 particles/mL 10m

    NMT 2 particles/mL 20m

    PMA, USP, FDA meet to discuss further refinement of the standard for

    In December 1974, USP proposed:NMT 50 particles/mL 10m

    NMT 5 particles/mL 25m

    studies and collaboration with AMA, ASHP, PMA:Absence of particles 50m in a 10 sample mean

    NMT 1 particle/mL 5m in a 10 sample mean

  • History and Limits Evolution

    Definitive Period III 1975-19861975 USP XIX chapter used a membrane test to evaluate the particle

    50 particles/ mL 10 m

    5 particles/mL 25 m

    electrical zone sensing (Coulter) or light extinction (HIAC) instruments.

    10,000 particles/container 10m

    1000 particles/container 25m

    Throughout 1985, many meetings between PDA, PMA, FDA, industry and USP reviewed the limits and methods for all volumes

    A new method, light extinction (obscuration), was added in 1985, in USP XXI chapter for small volume parenterals (SVP, 100mL and lower volume).

    Thus, LVP standards were established in the US as well, and initiated the

    Problems meeting particle standards for terminally sterilized dextrose-containing solutions were due to 5-hydroxymethylfurfural

    USP and FDA to insert a disclaimer into the definition of valid particulate matter. do not attempt to size or enumerate amorphous, semi-liquid, or otherwise morphologically indistinct materials, that is, material that has the appearance of a stain or discoloration on the membrane surface.

    January 1, 1986, the SVP standard and test method become official.

  • History and Limits Evolution

    Evaluative Period 1986 1995The test methods for counting particles were continually improved through collaborative investigation and method revision.

    1990 USP XXII defined new limits by membrane

    for large-volume injections (>100mL):

    for small-volume injections ( 100mL):

    1995 USP 23 provided the most dramatic change. The light obscuration method became the preferred or Method 1 approach.

    Ease of method control, objectivity, and efficiency

    History of product experience and regulatory filing

    for large-volume injections (>100mL):

    for small-volume injections ( 100mL):

  • History and Limits Evolution

    Evaluative Period 1995 2006In 2004, a new chapter for ophthalmic solutions was published in USP 27 for particulate matter in three size categories:

    LO and MM50 particles/mL 10 m

    5 particles/mL 25 m

    MM testing by membrane methodologyThe same 10 m and 25 m limits and 2 particles/mL 50 m

    Two important non-U.S. compendial organizations, European Pharmacopoeia (Ph. Eur.) and Japan Pharmacopoeia (JP) included the USP change to include the SVI products in particle limits.

    Ph. Eur. changed to include SVI products in 2005

    JP updated to include them in 2006.

  • History and Limits Evolution

    Harmonization Period 2007 2010

    2007 all three organizations, through the Pharmacopeial Discussion

    Group, have harmonized the methods, definitions and limits.

    This chapter is seen in the respective pharmacopeia as USP ,

    EP 5.5 and JP XIV, XV, and is the current standard.

    These standards monitor injectable solutions for the content of

    particulate matter undetected by visual inspection. However, the

    harmonized USP chapter provides much less method direction than in

    previous USP versions, and is addressed by the planned

    publication of chapter in 2011.

  • Harmonization - Current USP Guideline

    USP 33 --InjectionsStipulates the expectations and tests for all injectable dose forms

    Foreign and Particulate Matter

    Test methods concern quantification of particulate matter in:

    Pharmaceutical injectable products

    Ophthalmic solutions

    USP 33 --Particulate Matter in Injections2 methods: LO and MM

    Limits depend upon product volume

    USP 33 --Particulate Matter in Ophthalmic Solutions2 methods: LO and MM

    Limits are per mL

    Exempt from limits:

    Radiopharmaceuticals

    Parenteral products for which labeling specified use of a final filter (provided

    available scientific data justify the exemption)

    Irrigating solutions

  • Current Limits

    The use of two methods is a "two -LO results are suspicious or fail limits, the microscope method is run

    Method 1 LO Method 2 - Microscope

    Parenteral Volume

    10 m 25 m 10 m 25 m

    SVI 100 mL and lower

    6000 per container

    600 per container

    3000 per container

    300 per container

    LVI above 100 mL

    25 per mL 3 per mL 12 per mL 2 per mL

  • Current Limits

    Since 2004 Chapter --Particulate Matter in Ophthalmic

    SolutionsOfficial for the sub-visible particle limits of Ophthalmic products

    All limits on a per mL basis.

    Methods are essentially and Limits are tight

    Method 1 - LO Method 2 - Microscope

    10 m 25 m 10 m 25 m 50 m

    50 per mL 5 per mL 50 per mL 5 per mL 2 per mL

  • Comparison of Injectable & Ophthalmic Solution

    Product Particulate Load Limits As 5mL Fill Volumes. Particle Size

    by Membrane

    Assay

    USP Limits

    5mL Injectable

    Volume

    USP Limits

    5mL Ophthalmic Volume

    10 m 3000 part./container 250 particles in a 5mL

    container or 50 particles/mL

    25 m 300 part./container 25 particles in a 5mL

    container or 5 particles/mL

    50 m No specification 10 particles in a 5mL

    container or 2 particles/mL

    Particle Count by Membrane Microscopy

    Limits for Liquid Products

  • History and Limits Evolution

    How will we characterize the next period of work?

  • USP Chapter Crossover

    Two Particle Count Tests in USP Guideline Chapter

    Light obscuration

    Microscopic

    Represent the baseline and standardized approach for particle

    measurement, with which many companies have compiled deep

    historical data.

    Provide a snapshot of particle load in final product form

    Are not necessarily optimal for every formula or dose form

    Visual Inspection detects particles there is crossover

    Note that

    Analytical methods and limits evolve with technical improvements

    Changes to methods and limits are influenced by regulatory,

    commercial and harmonization efforts

  • Visible Sub-visible Sub-micrometer

    1m25m 10m150m

    Increasing Probability of Detection

    Visible

    Gray zone

    Compendial Threshold

    Size Domains

    Where do the visible and sub-visible domains crossover?

  • Visual Inspection

    Product development and release must include both visual

    inspection and particle counting

    Commercial products must comply with Chapters and

    or

    Size Crossover: Material may be visible down to ~50m

    Particles are visible in the upper end of the detectable size

    range by the primary sub-visible count method, light

    obscuration

    The microscopic method retains solids (some semi-solids) at

    into the visible zone

  • Size domain intersection for Chapter visual

    inspection and Chapter / sub-visible

    methods

    LO results are truly sub-visible

    Membrane results span a wide size range, into the

    visible

    Product must meet all requirements from time of

    release to end of shelf life

  • What is Particulate Matter?

    Particulate matter in injections and parenteral infusions is

    defined as extraneous, mobile, undissolved particles, other

    than gas bubbles, unintentionally present in the solution.

    Practically, any semi-solid to solid material, soft to hard,

    transparent to opaque may be counted as a particle and may

    be considered objectionable dependant upon the identity.Such as air, liquid, gel, singular solid, aggregate, agglomerate, drug

    solid, salt, polymorph, lubricant, plasticizer

  • Particulate Matter Details

    What forms do we see?Insoluble, mobile solids/semi-solids

    Single entity alone or in aggregates

    one specie or multiple species

    chemical interactions yielding solids

    Formulation change yielding solids

    OriginsForeign to the Process: Presence due to Extrinsic Source

    Part of the Process/Product: Formation due to Intrinsic Source

    Process function failure

    Formulation/Package origin

    Inherent to the formulation Presence in biologic products

    Studying the Particle Load aids Refinement of the Final

    Product

  • Test Methods

    Light Obscuration

    Membrane Microscopy

  • Current Test Methods

    Two procedures are specified in both Chapters and

    has particle limits according to test article volume:Method 1 (Light Obscuration Particle Count Test)

    1a. > 100mL per mL

    1b. = 100mL per container

    1b. < 100mL per container

    Method 2 (Microscopic Particle Count Test) 2a. > 100mL per mL

    2b. = 100mL per container

    2b. < 100mL per container

    has limits per mL regardless of product volume.

    Method 1 is preferred when examining injections, parenteral infusions and ophthalmic solutions

    It may be necessary to test some preparations by the light obscuration particle count test followed by the microscopic particle count test to reach a conclusion on conformance to the requirements.

    It may be necessary to utilize membrane microscopic alone, based upon the nature of the formulation or package.

  • LO Particle Count Test

    Equipment: an apparatus that sizes particles and counts

    by size using light obscuration (extinction) principle

    Not Included:Instrument Standardization TestsCalibration Options

    Typical CalibrationMono-disperse spherical solids between 10m and 25m

    System Suitability verified with USP Particle Count

    Reference Standard

    Blank ControlParticle-free water

    Particle-free water is water that has been passed through a

    0.22m filter

    5 x 5ml aliquots, total load 25 particles

  • Product Samples

    Volume 25mL

    Pool 10 or more containers to allow sufficient aliquot

    Necessary for LO/Unnecessary for MM

    Volume 25-100mL

    One container, minimally

    Volume >100mL

    One container, minimally

    What is your statistical sampling consideration?

  • Light Obscuration Method

    The product particulate matter content

    is determined by counting particles in

    pre-determined size bins, from small

    portions of product liquid

    Pools of sample are required for

    products

  • Light Obscuration Method

    Voltage response is recorded

    for a sphere of equivalent

    circular diameter

    The resultant particle size is

    generated from a size-

    voltage response calibration

    curve constructed using

    spherical mono-disperse

    certified size standards

    33

    Per HachUltra

  • Light Obscuration Method

    Enumerates sub-visible

    solids, but also liquid and gas

    Samples test fluid via a

    metered system

    Vendor-specific analysis

    routines for calibration and

    daily sampling

    LO method has few operator-

    induced effects

    Some articles cannot be

    tested meaningfully by light

    obscuration

    34

  • Light Obscuration Method

    Blank Control

    Degas the open container, by

    Sonication at 80 to 120 watts for

    approximately 30 seconds, or

    allowing the sample to stand to

    dissipate gas bubbles, or

    evacuate the sample to outgas

    (new in USP

  • LO Instrument Standardization

    Instrument qualification is essential to test performance

    Light Obscuration sections emphasize criteria rather

    than specific determination method:Calibrate with spheres 10-25m

    System Suitability can be verified with USP PCRS

    User is responsible for

    Ensuring that the counter is used and operated according to the

    manufacturer's instructions

    Ensuring proper standardization methods are applied to instrument

    see Chapter Instrument Standardization Tests

    Calibration

    Interim checks

    Daily

    Weekly

    3 month

    These principles must be followed ensure that instruments operate

    accurately within defined ranges36

  • LO Test Particle Limits

    Reporting: average of 3 aliquots not to exceed particle limit

    General Chapter

    Test 1.A. Large-Volume Injections

    (more than 100mL)

    25 per mL 3 per mL

    Test 1.B. Small-Volume Injections (less

    than and equal to 100mL)

    6000 per

    container

    600 per container

    Ophthalmic Solutions 50 per mL 5 per mL

    1.A. > 100mL 1.B. 100mL 1.B.

  • Membrane Assay Method

    Harmonized Compendial Methods USP/EP/JP

    100 x 2% with graticule

    Two illumination paths

    USP Graticule verified at installation

    Not included:Calibration day-of-use

    Blank control considerations

    QualificationTraining and Inter-Lab Certification

    Mono-disperse spherical solids from 10m and upUSP Particle Count RS

  • Microscopic Particle Count Test

    Blank Control for Particle-free water

    USP Particle-0.22-

    Considerations for ?

    Sampling LVP: may use single containers

    SVP: may use single containers

    Unless volume 25mL use 10 fills, pooled

    Dilution and Recon with particle-free water or suitable solvent

    20x inversion to mix

    Entire volume is sampled

    Use wetted, full particle-free water rinses

  • Membrane Microscopy Method

    Captures solid matter from liquid of one or more product containers

    onto a filter membrane

    Enumerates sub-visible to visible, solid to soft or semi-solid

    particulate matter on the membrane surface

    Retained particles are counted by scanning the retention surface at

    100 with a compound binocular light microscope

    Method is highly operator-dependent

    Operator makes decisions regarding

    -the-

    40

  • Membrane Microscopy Lab

    Minimize traffic

    Separate HEPA areas for wet

    filtration and dry counting

    steps

    Dedicated glassware

    Dilution and pooling systems

    developed for product

    volumes, product types

    Establish lab controls

    Operator training essential

    Operator performance

    evaluation recommended

  • Microscopic Particle Count Test

    Full

    Count particles in entire effective filtration area

    Wet zone of liquid passage

    Inner diameter of funnel

    undefined

    defined

    If 1000 or less particles are present, the membrane should be fully counted

    If the count difference between a GFOV at center and one at the

    Count 20 GFOV for 25mm (16mm EFA) membranes

    Count 100 GFOV for 47mm (37mm EFA) membranes

  • General Chapter

    Test 2.A. Large-Volume

    Injections (more than 100 mL)

    12 per mL 2 per mL Not applicable

    Test 2.B. Small-Volume

    Injections (less than and equal

    to 100 mL)

    3000 per

    container

    300 per

    container

    Not applicable

    Ophthalmic Solutions 50 per mL 5 per mL 2 per mL

    MM Test Particle Limits

    Reporting: average count from units tested not to exceed

    particle limit

    2.A. > 100mL 2.B. 100mL 2.B.

  • Informational Chapter

    Provides LO calibration and control guidance removed from

    after harmonizationInstrument Standardization Tests

    Flow Rate

    Volume Accuracy

    Calibration

    Sensor Resolution

    Particle Counting Accuracy

    Provides microscopy setup and counting guidance

    Provides discussion regarding pharmaceutical development

    practices

    Previewed in Pharm Forum 35(6) Nov.-Dec. 2009

    Official USP34, 2011

  • LO and MM: Compare and Contrast

    Light Obscuration

    Provides a robust, vendor-supported, precise and accurate

    electronic particle count methodology

    It counts suspended particles that are solid and others that may

    be semi-solid, liquid or gas

    Membrane Microscopy

    Provides a count of particles by a 100 microscopic scan of a

    membrane surface

    Provides an orthogonal means of counting

    As a second-pass, is it a Referee method ?

    The two methods yield different perspectives of the particle

    population

    45

  • LO and MM: Compare and Contrast

    Light Obscuration

    Effectively counts all particles as equivalent circular diameters,

    over a wide range of concentration

    Skews resultant sizes relative to shape when deviating from the

    ideal spherical, or 1:1:1, axial ratio

    Counts air bubbles and immiscible oil droplets

    46

  • LO and MM: Compare and Contrast

    Membrane Microscopy

    Directly compares isolates to

    calibrated 10m and 25m circles and

    includes a linear scale

    equivalent 10m and 25m circles

    while counting the membrane isolates

    Many users attempt to correlate the

    membrane count to that obtained by

    LO

    47

  • Ongoing Revision Process

    Current methods and reliability are continually reviewed and

    improved by USP staffContribution from industry

    Assistance by industry specialists

    Contributions from industry and by industry specialists

    are overseen by the USP-DF (Parenteral Products: Industry)

    Expert Committee

    USP publishes Stimuli Articles in Pharmacopeial Forum to

    allow perspectives, dissension, and initiatives from

    participants

    Certain formulations cannot be tested directly by either

    method:Examples: sterile suspensions, nano/micro-suspensions, some

    emulsions, low-volume and special device products

  • USP Workshop ParticlesDecember 8 - 10, 2010

    USP Chapters and Methods

    - VS -

    Alternative Methods for Investigation and Compliance

    49

  • Biotech products

    Vacines

    Novel treatments for Cancer

    Nanoparticles (overcoming insolubility)

    Controlled release microspheres

    Polymers

    Crystalline nanoparticles

    Liposomal formulations

    50USP Workshop on Particles

  • Extrinsic

    Glass/stopper fragments

    Metals

    Silicon oil

    Filter/process particles

    Skin flakes

    Ordinary dirt particles

    Insect parts

    Fibers: clothing/hair/etc.

    51

    Intrinsic

    Ingredient degradation

    API / Excipient changes

    Active ingredients

    Protein aggregation

    Aggregation/foreign matter

    Silicon interaction

    Process related

    Ingredient anomalies

    Immiscible droplets

    USP Workshop on Particles

  • Leachables

    Extractables

    Foreign particles

    Protein aggregation

    Glass delamination

    Increased use of plastics

    Pre- filled syringes

    New stopper materials/coatings

    52USP Workshop on Particles

  • ComplianceDetect manufacturing problems

    USP Particulate Matter, Lot release

    Formulation degradation over time/stress/ stabil .

    Research Particle Size Analysis (especially sub - micron)

    Evaluation of processing methods

    Evaluation of ingredients

    New drug products are more complex and often require more characterization.

    Investigational (methods, understanding)

    53USP Workshop on Particles

  • Provides 100% inspection of the batch.

    Provides a pass/fail from brief observation.

    Over 12 variables affect results.

    Particles in rejects may not be identified.

    Plays vital role with complex formulations

    54USP Workshop on Particles

  • What is the smallest particle size our inspectors see reliably ?

    What factors in sub - visible results, show correlation with V.I. results?

    What is the variability between our inspectors at this plant, and those in our other plants ? Same with results ?

    How much does this variability affect our product reject and acceptance levels ? (cost ) ?

    Do we understand the relations between Vis. Inspector Training, our actual inspection process, our rejection levels and USP with each of our products ?

    What are the critical issues for our products when comparing human and automated inspection with Sub - Visible (USP) results ?

    Attend the annual PDA Visual Inspection Forum in Fall/2011 .

    55USP Workshop on Particles

  • METHOD I

    Light Obscuration (LO) Laser (usually) light extinction (blockage).

    METHOD II

    Microscopic Membrane (MM) Sample is

    membrane and examined by optical microscopy.

    56USP Workshop on Particles

  • particle size.Different methods or instruments used to measure particle size (or count), produce different results.The current Compendial methods are LO and MM

    Current (laser base) instruments for particle countingInvestigational (orthogonal) methodOther (analytical) complimentary principles

    57USP Workshop on Particles

  • 58USP Workshop on Particles

  • 59

    Formulation

    Immiscib

    droplets

    Gels

    Fibers &

    discrete

    particles

    USP Workshop on Particles

  • USP Particulate Matter Methods I and II

    Single Particle Optical Sensing Test Results (Summation Mode)

    Particles per mL size (in microns)

    Microscopic Membrane

    Method

    Sample

    >0.5m > 0.7m > 0.8m > 1m > 2m > 5m > 10m > 25m > 10m > 25m

    A

    737995 562576 523208 460241 153786 36004 4441 0 4.7 2.6

    B

    1020601 768623 707751 603883 171300 41064 7145 203 2.3 0.7

    C

    1732076 1449868 1376814 1242247 497642 148685 22078 128 2.7 1.3

    60USP Workshop on Particles

  • 61USP Workshop on Particles

  • 62USP Workshop on Particles

  • SAMPLER

    Ensures consistent flow rate Low flow rates are advantageous for investigations/etc.

    Volume precision

    Commonly syringe based (1, 5, 10, 25 ml)Flow rates are usually limited to syringe size

    (may not achieve 30ml/min with 1ml syringe)

    Accepts large or small containers

    63USP Workshop on Particles

  • USP Workshop on Particles 64

  • USP Workshop on Particles 65

  • SENSOR Single detector for USP

    Minimum size sensitivity: 2 microns (or less)

    Flow rate range from 10ml/min

    Large cell opening (400 m) preferable

    Know coincidence values for 2 to 25 microns

    Remote sensor mounting may be helpful for Use in a small Biosafety cabinet or clean hood

    Accommodates very large containers

    Allow sampling directly from experiment

    Easier to accommodate stirring boxes

    66USP Workshop on Particles

  • Software (must be part 11)

    Includes recipe structure

    Includes particle size distribution capability

    Export data to Excel

    Includes custom alarms for size

    Built - in PQ Check routine with trend monitor

    Converts output report to per/ml or other

    Graphic sample overlays for comparisons

    67USP Workshop on Particles

  • 68USP Workshop on Particles

  • 69USP Workshop on Particles

  • LO instruments with < 15 channels of size resolution.

    Provide essential reporting

    Easier to calibrate since fewer sphere sizes used

    Data generally reported in a table

    Reports should provide easy to see

    Individual runs, displayed to observe run - run informity

    Display all sizes reported in Cummulative mode

    Provide additional reports to graphically compare sample counts (per ml or per container) vs size

    70USP Workshop on Particles

  • Verify volume accuracy

    Verify flow rate

    Verify size accuracy

    Verify count accuracy

    Measure system drift

    Perform routinely

    Verification tests can overlap

    Use Polystyrene latex particles (PSL)

    Use standards certified for Size AND Count

    71USP Workshop on Particles

  • 72USP Workshop on Particles

  • 73

    Particle Std Size (m)

    Previous Method New Method Size

    Total Count >

    10mRatio

    10/15 Mean Size Std Dev

    Counts Error

    Date Initials

    Within in PQ Check

    Dil Fac Peak percent Lot #

    3/2/2009 AP 15.02 4092 1.54 15.12 0.42 3285 0.6658 JB16A

    3/3/2009 AP 15.02 4041 1.48 15.10 0.42 3147 0.5326 JB16A

    3/4/2009 AP 15.02 4049 1.51 15.13 0.42 3361 0.7324 JB16A

    3/5/2009 AP 15.02 4028 1.68 15.07 0.41 3318 0.3329 JB16A

    3/6/2009 AP 15.02 4169 1.73 15.04 0.42 3215 0.1332 JB16A

    3/9/2009 AP 15.02 5083 1.33 15.22 0.42 2993 1.3316 JB16A

    3/10/2009 AP 15.02 3925 15.28 0.52 3.68 4166 1.7310 JB16A

    3/11/2009 AP 15.02 5359 1.33 15.38 0.55 3.67 4106 2.3968 JB16A

    3/12/2009 AP 15.02 4383 1.43 15.34 0.50 3.69 3924 2.1305 JB16A

    3/13/2009 AP 15.02 4028 1.63 15.19 0.55 3.95 4299 1.1318 JB16A

    3/16/2009 AP 15.02 3885 1.65 15.20 0.52 4.01 4202 1.1984 JB16B

    3/17/2009 AP 15.02 3865 1.64 15.19 0.50 3.68 2923 1.1318 JB16B

    3/18/2009 AP 15.02 3867 1.81 15.10 0.51 3.94 4076 0.5326 JB16B

    3/19/2009 AP 15.02 3921 1.79 15.10 0.53 3.94 4360 0.5326 JB16B

    3/20/2009 AP 15.02 4011 1.79 15.11 0.53 3.81 3884 0.5992 JB16B

    3/23/2009 AP 15.02 4199 1.58 15.36 0.52 3.57 4083 2.2636 JB16B

    3/27/2009 AP 15.02 4356 1.36 15.35 0.53 3.68 4331 2.1971 JB16B

    3/30/2009 AP 15.02 3928 1.15 15.35 0.52 3.73 4562 2.1971 JB16B

  • Electrozone Sensing (EZS)

    Optical Microscopy (human observation)

    Image Analysis (camera w/software analysis)Static

    Dynamic

    Single Particle Optical Sensing (SPOS)

    Nano Particle Tracking (NTA)

    Fluorescence NTA (FNTA)

    Archimedes, mass of individual particles

    Spectrex

    74USP Workshop on Particles

  • 75

    Syringes1 ml, 10 ml, 25 ml

    Flow rate settings10 100 ml

    Sensors (MC05 is added)

    Sampling Probes(added short - small bore probe)

    USP Workshop on Particles

  • Interchangeable sampling probes, syringes, and sensors

    Ensure you have one instrument to manage all applicationsNew MC05 sub - micron sensorChange configuration with no impact to instrument validation

    Customized reporting

    reportsAdd company logo, user defined descriptors

    Customized test recipesProcedure Builder enables the development of unique test recipes for your applicationCopy a recipe and make the changes!

    76USP Workshop on Particles

  • 77

    www.hachultra.comUSP Workshop on Particles

  • 78USP Workshop on Particles

    USP LO

  • 79

  • 80

  • 81

  • 82

  • 83USP Workshop on Particles

  • 84

  • HACH HRLD150 2.0 to 400mPMS APSS 200 2.0 to 120mHACH HRLD150 1.0 to 150mAccusizer LE400 - 1 1.0 to 400mAccusizer Fx 0.7 to 400m *Accusizer 780 0.5 to 400mHACH MicroCount 05 0.5 to 400mPMS S05 0.5 to 20mPMS S03 0.3 to 20mPMS S02 0.2 to 2mAccusizer FxNano 0.15 to 10m *Nanosight LM20 0.03 to 1m *Coulter Counter 0.4 to 1,000mDynamic Image Analyzers (next section )*Count determined via algorithm Contact Mfr for details

    85USP Workshop on Particles

  • USP Workshop on Particles 86

  • USP Workshop on Particles 87

  • 88USP Workshop on Particles

  • 89

  • 90USP Workshop on Particles

  • 91USP Workshop on Particles

  • 92USP Workshop on Particles

  • 93USP Workshop on Particles

  • 94

    http://www.malvern.com/common/downloads/MRK652.pdf

    USP Workshop on Particles

  • 95USP Workshop on Particles

  • 96USP Workshop on Particles

  • 97USP Workshop on Particles

  • www.nanosight.com 98USP Workshop on Particles