USP Workshop on Particle Size
Transcript of USP Workshop on Particle Size
-
Welcome to the
USP Workshop on Particle Size:Particle Detection and Measurement
Day 2
December 8-10, 2010
USP Headquarters, Rockville, MD
-
USP Workshop on Particle Size:
Particle Detection and Measurement
December 8-10, 2010; USP Headquarters
Desmond Hunt, Ph.D.
USP Senior Scientific Liaison
Welcome and Opening Remarks
-
USP Workshop on Particle Size:
Particle Detection and Measurement
December 8-10, 2010; USP Headquarters
Session III: USP General Chapters
Track 1: Parenterals
Spalding Auditorium
Track 2: Aerosols
Briggs/Parker/
Marshall/Wiley
-
USP Workshop on Particle Size:
Particle Detection and Measurement
December 8-10, 2010; USP Headquarters
Chair: Mr. Scott Aldrich
Ultramikro
Session III, Track 1: Parenterals - USP
General Chapters and
-
Quality Standards for Medicines, Supplements, and Food Ingredients throughout the World
8.1 Parenterals USP General Chapters /
Historical Perspective and Current Chapters ReviewScott Aldrich
USP Workshop on Particle Size:
Particle Detection & Measurement
-
Outline
Introduction
Background
History and Limits Evolution
History of liquid particle counting
Periods of Investigation
Harmonization of Aug 1 2007
Current chapters
Limits according to container volume and method used
informational chapter official USP 34, 2011
follows discussion
Chapter Crossover
What is Particulate Matter?
Test Methods
Light Obscuration Counting and Operations
Membrane Microscopic Counting and Operations
Comparison of LO and MM methods
-
Introduction
Parenteral injections and ophthalmic solutions are sterile
liquids, with expectations for physical appearance and
stability.
All released product must comply with two limit tests for
particle content:-
USP General Chapter Injections
See PF Stimuli Article Vol. 35(5) [Sept.-Oct. 2009], Visible
Particulates in Injections - A History and a Proposal to Revise USP
General Chapter Injections
Must contain low amounts of sub-visible particles
USP Injections
USP Ophthalmic solutions
-
History and Limits Evolution
History of liquid particle counting
Periods of Investigation1
Cognitive
Latent
Definitive
Evaluative
Harmonization
The Future
1Floyd Benjamin, IES conference 1990
-
History and Limits Evolution
Cognitive Period 1905-1948
1905 First USP reference to parenteral solution as a compendial drug. No references to particulate matter content or clarity.
1916 NF IV includes 6 monographs for parenterals; however, no mention of appearance.
aqueous solutions are to be clear, i.e., when observed over a bright light, they shall be substantially free from precipitate, cloudiness or turbidity. Specks or flecks; fibers or cotton hairs, or any undissolved
In 1942, USP XII and NF VII both attempt to further define the
published definition.
-
History and Limits Evolution
Latent Period 1949-1965 No changes in compendial requirements, although many reports in literature regarding medical events from particulate matter injection.
1949 USP revision committee eliminates the clarity test
should be exercised in the preparation of injections to
1965 - First Limit Standards (Canberra, Australia)In 1965, the National Biological Standards Laboratory, Canberra
-stimulated compendia and regulatory activities in the US, leading to the Definitive Period.
-
Definitive Period I 1966-19731966 FDA symposium on the safety of large volume parenterals1
1972 Australian comparison of Coulter and HIAC counts for parenteral products2
Australian and English standards for particulate matter in
Australia - ten containers must be less than 2x the following limits:
NMT 100 particles/mL 2m
NMT 5 particles/mL 20m
British Std: mean of 5 containers and individual
1973 Turco and Davis review the clinical consequences of injected particulate matter3
History and Limits Evolution
1. FDA - National symposium on safety of large volume parenteral solutions. U.S. Dept. of Health, Education and Welfare,
Washington DC, 1966.
2. Kendall and Tilley, Comparison of Coulter counter and HIAC counts for parenteral products
3. Turco S. Davis NM. Hosp Pharm. 1973;8:137-40.
-
History and Limits Evolution
Definitive Period II 19741974 Revision of Foreign Particulate Standards
Great BritainNMT 1000 particles/mL 2m
NMT 100 particles/mL 5m
AustraliaNMT 1000 particles/mL 2m
NMT 250 particles/mL 3.5m
NMT 100 particles/mL 5m
NMT 25 particles/mL 10m
NMT 2 particles/mL 20m
PMA, USP, FDA meet to discuss further refinement of the standard for
In December 1974, USP proposed:NMT 50 particles/mL 10m
NMT 5 particles/mL 25m
studies and collaboration with AMA, ASHP, PMA:Absence of particles 50m in a 10 sample mean
NMT 1 particle/mL 5m in a 10 sample mean
-
History and Limits Evolution
Definitive Period III 1975-19861975 USP XIX chapter used a membrane test to evaluate the particle
50 particles/ mL 10 m
5 particles/mL 25 m
electrical zone sensing (Coulter) or light extinction (HIAC) instruments.
10,000 particles/container 10m
1000 particles/container 25m
Throughout 1985, many meetings between PDA, PMA, FDA, industry and USP reviewed the limits and methods for all volumes
A new method, light extinction (obscuration), was added in 1985, in USP XXI chapter for small volume parenterals (SVP, 100mL and lower volume).
Thus, LVP standards were established in the US as well, and initiated the
Problems meeting particle standards for terminally sterilized dextrose-containing solutions were due to 5-hydroxymethylfurfural
USP and FDA to insert a disclaimer into the definition of valid particulate matter. do not attempt to size or enumerate amorphous, semi-liquid, or otherwise morphologically indistinct materials, that is, material that has the appearance of a stain or discoloration on the membrane surface.
January 1, 1986, the SVP standard and test method become official.
-
History and Limits Evolution
Evaluative Period 1986 1995The test methods for counting particles were continually improved through collaborative investigation and method revision.
1990 USP XXII defined new limits by membrane
for large-volume injections (>100mL):
for small-volume injections ( 100mL):
1995 USP 23 provided the most dramatic change. The light obscuration method became the preferred or Method 1 approach.
Ease of method control, objectivity, and efficiency
History of product experience and regulatory filing
for large-volume injections (>100mL):
for small-volume injections ( 100mL):
-
History and Limits Evolution
Evaluative Period 1995 2006In 2004, a new chapter for ophthalmic solutions was published in USP 27 for particulate matter in three size categories:
LO and MM50 particles/mL 10 m
5 particles/mL 25 m
MM testing by membrane methodologyThe same 10 m and 25 m limits and 2 particles/mL 50 m
Two important non-U.S. compendial organizations, European Pharmacopoeia (Ph. Eur.) and Japan Pharmacopoeia (JP) included the USP change to include the SVI products in particle limits.
Ph. Eur. changed to include SVI products in 2005
JP updated to include them in 2006.
-
History and Limits Evolution
Harmonization Period 2007 2010
2007 all three organizations, through the Pharmacopeial Discussion
Group, have harmonized the methods, definitions and limits.
This chapter is seen in the respective pharmacopeia as USP ,
EP 5.5 and JP XIV, XV, and is the current standard.
These standards monitor injectable solutions for the content of
particulate matter undetected by visual inspection. However, the
harmonized USP chapter provides much less method direction than in
previous USP versions, and is addressed by the planned
publication of chapter in 2011.
-
Harmonization - Current USP Guideline
USP 33 --InjectionsStipulates the expectations and tests for all injectable dose forms
Foreign and Particulate Matter
Test methods concern quantification of particulate matter in:
Pharmaceutical injectable products
Ophthalmic solutions
USP 33 --Particulate Matter in Injections2 methods: LO and MM
Limits depend upon product volume
USP 33 --Particulate Matter in Ophthalmic Solutions2 methods: LO and MM
Limits are per mL
Exempt from limits:
Radiopharmaceuticals
Parenteral products for which labeling specified use of a final filter (provided
available scientific data justify the exemption)
Irrigating solutions
-
Current Limits
The use of two methods is a "two -LO results are suspicious or fail limits, the microscope method is run
Method 1 LO Method 2 - Microscope
Parenteral Volume
10 m 25 m 10 m 25 m
SVI 100 mL and lower
6000 per container
600 per container
3000 per container
300 per container
LVI above 100 mL
25 per mL 3 per mL 12 per mL 2 per mL
-
Current Limits
Since 2004 Chapter --Particulate Matter in Ophthalmic
SolutionsOfficial for the sub-visible particle limits of Ophthalmic products
All limits on a per mL basis.
Methods are essentially and Limits are tight
Method 1 - LO Method 2 - Microscope
10 m 25 m 10 m 25 m 50 m
50 per mL 5 per mL 50 per mL 5 per mL 2 per mL
-
Comparison of Injectable & Ophthalmic Solution
Product Particulate Load Limits As 5mL Fill Volumes. Particle Size
by Membrane
Assay
USP Limits
5mL Injectable
Volume
USP Limits
5mL Ophthalmic Volume
10 m 3000 part./container 250 particles in a 5mL
container or 50 particles/mL
25 m 300 part./container 25 particles in a 5mL
container or 5 particles/mL
50 m No specification 10 particles in a 5mL
container or 2 particles/mL
Particle Count by Membrane Microscopy
Limits for Liquid Products
-
History and Limits Evolution
How will we characterize the next period of work?
-
USP Chapter Crossover
Two Particle Count Tests in USP Guideline Chapter
Light obscuration
Microscopic
Represent the baseline and standardized approach for particle
measurement, with which many companies have compiled deep
historical data.
Provide a snapshot of particle load in final product form
Are not necessarily optimal for every formula or dose form
Visual Inspection detects particles there is crossover
Note that
Analytical methods and limits evolve with technical improvements
Changes to methods and limits are influenced by regulatory,
commercial and harmonization efforts
-
Visible Sub-visible Sub-micrometer
1m25m 10m150m
Increasing Probability of Detection
Visible
Gray zone
Compendial Threshold
Size Domains
Where do the visible and sub-visible domains crossover?
-
Visual Inspection
Product development and release must include both visual
inspection and particle counting
Commercial products must comply with Chapters and
or
Size Crossover: Material may be visible down to ~50m
Particles are visible in the upper end of the detectable size
range by the primary sub-visible count method, light
obscuration
The microscopic method retains solids (some semi-solids) at
into the visible zone
-
Size domain intersection for Chapter visual
inspection and Chapter / sub-visible
methods
LO results are truly sub-visible
Membrane results span a wide size range, into the
visible
Product must meet all requirements from time of
release to end of shelf life
-
What is Particulate Matter?
Particulate matter in injections and parenteral infusions is
defined as extraneous, mobile, undissolved particles, other
than gas bubbles, unintentionally present in the solution.
Practically, any semi-solid to solid material, soft to hard,
transparent to opaque may be counted as a particle and may
be considered objectionable dependant upon the identity.Such as air, liquid, gel, singular solid, aggregate, agglomerate, drug
solid, salt, polymorph, lubricant, plasticizer
-
Particulate Matter Details
What forms do we see?Insoluble, mobile solids/semi-solids
Single entity alone or in aggregates
one specie or multiple species
chemical interactions yielding solids
Formulation change yielding solids
OriginsForeign to the Process: Presence due to Extrinsic Source
Part of the Process/Product: Formation due to Intrinsic Source
Process function failure
Formulation/Package origin
Inherent to the formulation Presence in biologic products
Studying the Particle Load aids Refinement of the Final
Product
-
Test Methods
Light Obscuration
Membrane Microscopy
-
Current Test Methods
Two procedures are specified in both Chapters and
has particle limits according to test article volume:Method 1 (Light Obscuration Particle Count Test)
1a. > 100mL per mL
1b. = 100mL per container
1b. < 100mL per container
Method 2 (Microscopic Particle Count Test) 2a. > 100mL per mL
2b. = 100mL per container
2b. < 100mL per container
has limits per mL regardless of product volume.
Method 1 is preferred when examining injections, parenteral infusions and ophthalmic solutions
It may be necessary to test some preparations by the light obscuration particle count test followed by the microscopic particle count test to reach a conclusion on conformance to the requirements.
It may be necessary to utilize membrane microscopic alone, based upon the nature of the formulation or package.
-
LO Particle Count Test
Equipment: an apparatus that sizes particles and counts
by size using light obscuration (extinction) principle
Not Included:Instrument Standardization TestsCalibration Options
Typical CalibrationMono-disperse spherical solids between 10m and 25m
System Suitability verified with USP Particle Count
Reference Standard
Blank ControlParticle-free water
Particle-free water is water that has been passed through a
0.22m filter
5 x 5ml aliquots, total load 25 particles
-
Product Samples
Volume 25mL
Pool 10 or more containers to allow sufficient aliquot
Necessary for LO/Unnecessary for MM
Volume 25-100mL
One container, minimally
Volume >100mL
One container, minimally
What is your statistical sampling consideration?
-
Light Obscuration Method
The product particulate matter content
is determined by counting particles in
pre-determined size bins, from small
portions of product liquid
Pools of sample are required for
products
-
Light Obscuration Method
Voltage response is recorded
for a sphere of equivalent
circular diameter
The resultant particle size is
generated from a size-
voltage response calibration
curve constructed using
spherical mono-disperse
certified size standards
33
Per HachUltra
-
Light Obscuration Method
Enumerates sub-visible
solids, but also liquid and gas
Samples test fluid via a
metered system
Vendor-specific analysis
routines for calibration and
daily sampling
LO method has few operator-
induced effects
Some articles cannot be
tested meaningfully by light
obscuration
34
-
Light Obscuration Method
Blank Control
Degas the open container, by
Sonication at 80 to 120 watts for
approximately 30 seconds, or
allowing the sample to stand to
dissipate gas bubbles, or
evacuate the sample to outgas
(new in USP
-
LO Instrument Standardization
Instrument qualification is essential to test performance
Light Obscuration sections emphasize criteria rather
than specific determination method:Calibrate with spheres 10-25m
System Suitability can be verified with USP PCRS
User is responsible for
Ensuring that the counter is used and operated according to the
manufacturer's instructions
Ensuring proper standardization methods are applied to instrument
see Chapter Instrument Standardization Tests
Calibration
Interim checks
Daily
Weekly
3 month
These principles must be followed ensure that instruments operate
accurately within defined ranges36
-
LO Test Particle Limits
Reporting: average of 3 aliquots not to exceed particle limit
General Chapter
Test 1.A. Large-Volume Injections
(more than 100mL)
25 per mL 3 per mL
Test 1.B. Small-Volume Injections (less
than and equal to 100mL)
6000 per
container
600 per container
Ophthalmic Solutions 50 per mL 5 per mL
1.A. > 100mL 1.B. 100mL 1.B.
-
Membrane Assay Method
Harmonized Compendial Methods USP/EP/JP
100 x 2% with graticule
Two illumination paths
USP Graticule verified at installation
Not included:Calibration day-of-use
Blank control considerations
QualificationTraining and Inter-Lab Certification
Mono-disperse spherical solids from 10m and upUSP Particle Count RS
-
Microscopic Particle Count Test
Blank Control for Particle-free water
USP Particle-0.22-
Considerations for ?
Sampling LVP: may use single containers
SVP: may use single containers
Unless volume 25mL use 10 fills, pooled
Dilution and Recon with particle-free water or suitable solvent
20x inversion to mix
Entire volume is sampled
Use wetted, full particle-free water rinses
-
Membrane Microscopy Method
Captures solid matter from liquid of one or more product containers
onto a filter membrane
Enumerates sub-visible to visible, solid to soft or semi-solid
particulate matter on the membrane surface
Retained particles are counted by scanning the retention surface at
100 with a compound binocular light microscope
Method is highly operator-dependent
Operator makes decisions regarding
-the-
40
-
Membrane Microscopy Lab
Minimize traffic
Separate HEPA areas for wet
filtration and dry counting
steps
Dedicated glassware
Dilution and pooling systems
developed for product
volumes, product types
Establish lab controls
Operator training essential
Operator performance
evaluation recommended
-
Microscopic Particle Count Test
Full
Count particles in entire effective filtration area
Wet zone of liquid passage
Inner diameter of funnel
undefined
defined
If 1000 or less particles are present, the membrane should be fully counted
If the count difference between a GFOV at center and one at the
Count 20 GFOV for 25mm (16mm EFA) membranes
Count 100 GFOV for 47mm (37mm EFA) membranes
-
General Chapter
Test 2.A. Large-Volume
Injections (more than 100 mL)
12 per mL 2 per mL Not applicable
Test 2.B. Small-Volume
Injections (less than and equal
to 100 mL)
3000 per
container
300 per
container
Not applicable
Ophthalmic Solutions 50 per mL 5 per mL 2 per mL
MM Test Particle Limits
Reporting: average count from units tested not to exceed
particle limit
2.A. > 100mL 2.B. 100mL 2.B.
-
Informational Chapter
Provides LO calibration and control guidance removed from
after harmonizationInstrument Standardization Tests
Flow Rate
Volume Accuracy
Calibration
Sensor Resolution
Particle Counting Accuracy
Provides microscopy setup and counting guidance
Provides discussion regarding pharmaceutical development
practices
Previewed in Pharm Forum 35(6) Nov.-Dec. 2009
Official USP34, 2011
-
LO and MM: Compare and Contrast
Light Obscuration
Provides a robust, vendor-supported, precise and accurate
electronic particle count methodology
It counts suspended particles that are solid and others that may
be semi-solid, liquid or gas
Membrane Microscopy
Provides a count of particles by a 100 microscopic scan of a
membrane surface
Provides an orthogonal means of counting
As a second-pass, is it a Referee method ?
The two methods yield different perspectives of the particle
population
45
-
LO and MM: Compare and Contrast
Light Obscuration
Effectively counts all particles as equivalent circular diameters,
over a wide range of concentration
Skews resultant sizes relative to shape when deviating from the
ideal spherical, or 1:1:1, axial ratio
Counts air bubbles and immiscible oil droplets
46
-
LO and MM: Compare and Contrast
Membrane Microscopy
Directly compares isolates to
calibrated 10m and 25m circles and
includes a linear scale
equivalent 10m and 25m circles
while counting the membrane isolates
Many users attempt to correlate the
membrane count to that obtained by
LO
47
-
Ongoing Revision Process
Current methods and reliability are continually reviewed and
improved by USP staffContribution from industry
Assistance by industry specialists
Contributions from industry and by industry specialists
are overseen by the USP-DF (Parenteral Products: Industry)
Expert Committee
USP publishes Stimuli Articles in Pharmacopeial Forum to
allow perspectives, dissension, and initiatives from
participants
Certain formulations cannot be tested directly by either
method:Examples: sterile suspensions, nano/micro-suspensions, some
emulsions, low-volume and special device products
-
USP Workshop ParticlesDecember 8 - 10, 2010
USP Chapters and Methods
- VS -
Alternative Methods for Investigation and Compliance
49
-
Biotech products
Vacines
Novel treatments for Cancer
Nanoparticles (overcoming insolubility)
Controlled release microspheres
Polymers
Crystalline nanoparticles
Liposomal formulations
50USP Workshop on Particles
-
Extrinsic
Glass/stopper fragments
Metals
Silicon oil
Filter/process particles
Skin flakes
Ordinary dirt particles
Insect parts
Fibers: clothing/hair/etc.
51
Intrinsic
Ingredient degradation
API / Excipient changes
Active ingredients
Protein aggregation
Aggregation/foreign matter
Silicon interaction
Process related
Ingredient anomalies
Immiscible droplets
USP Workshop on Particles
-
Leachables
Extractables
Foreign particles
Protein aggregation
Glass delamination
Increased use of plastics
Pre- filled syringes
New stopper materials/coatings
52USP Workshop on Particles
-
ComplianceDetect manufacturing problems
USP Particulate Matter, Lot release
Formulation degradation over time/stress/ stabil .
Research Particle Size Analysis (especially sub - micron)
Evaluation of processing methods
Evaluation of ingredients
New drug products are more complex and often require more characterization.
Investigational (methods, understanding)
53USP Workshop on Particles
-
Provides 100% inspection of the batch.
Provides a pass/fail from brief observation.
Over 12 variables affect results.
Particles in rejects may not be identified.
Plays vital role with complex formulations
54USP Workshop on Particles
-
What is the smallest particle size our inspectors see reliably ?
What factors in sub - visible results, show correlation with V.I. results?
What is the variability between our inspectors at this plant, and those in our other plants ? Same with results ?
How much does this variability affect our product reject and acceptance levels ? (cost ) ?
Do we understand the relations between Vis. Inspector Training, our actual inspection process, our rejection levels and USP with each of our products ?
What are the critical issues for our products when comparing human and automated inspection with Sub - Visible (USP) results ?
Attend the annual PDA Visual Inspection Forum in Fall/2011 .
55USP Workshop on Particles
-
METHOD I
Light Obscuration (LO) Laser (usually) light extinction (blockage).
METHOD II
Microscopic Membrane (MM) Sample is
membrane and examined by optical microscopy.
56USP Workshop on Particles
-
particle size.Different methods or instruments used to measure particle size (or count), produce different results.The current Compendial methods are LO and MM
Current (laser base) instruments for particle countingInvestigational (orthogonal) methodOther (analytical) complimentary principles
57USP Workshop on Particles
-
58USP Workshop on Particles
-
59
Formulation
Immiscib
droplets
Gels
Fibers &
discrete
particles
USP Workshop on Particles
-
USP Particulate Matter Methods I and II
Single Particle Optical Sensing Test Results (Summation Mode)
Particles per mL size (in microns)
Microscopic Membrane
Method
Sample
>0.5m > 0.7m > 0.8m > 1m > 2m > 5m > 10m > 25m > 10m > 25m
A
737995 562576 523208 460241 153786 36004 4441 0 4.7 2.6
B
1020601 768623 707751 603883 171300 41064 7145 203 2.3 0.7
C
1732076 1449868 1376814 1242247 497642 148685 22078 128 2.7 1.3
60USP Workshop on Particles
-
61USP Workshop on Particles
-
62USP Workshop on Particles
-
SAMPLER
Ensures consistent flow rate Low flow rates are advantageous for investigations/etc.
Volume precision
Commonly syringe based (1, 5, 10, 25 ml)Flow rates are usually limited to syringe size
(may not achieve 30ml/min with 1ml syringe)
Accepts large or small containers
63USP Workshop on Particles
-
USP Workshop on Particles 64
-
USP Workshop on Particles 65
-
SENSOR Single detector for USP
Minimum size sensitivity: 2 microns (or less)
Flow rate range from 10ml/min
Large cell opening (400 m) preferable
Know coincidence values for 2 to 25 microns
Remote sensor mounting may be helpful for Use in a small Biosafety cabinet or clean hood
Accommodates very large containers
Allow sampling directly from experiment
Easier to accommodate stirring boxes
66USP Workshop on Particles
-
Software (must be part 11)
Includes recipe structure
Includes particle size distribution capability
Export data to Excel
Includes custom alarms for size
Built - in PQ Check routine with trend monitor
Converts output report to per/ml or other
Graphic sample overlays for comparisons
67USP Workshop on Particles
-
68USP Workshop on Particles
-
69USP Workshop on Particles
-
LO instruments with < 15 channels of size resolution.
Provide essential reporting
Easier to calibrate since fewer sphere sizes used
Data generally reported in a table
Reports should provide easy to see
Individual runs, displayed to observe run - run informity
Display all sizes reported in Cummulative mode
Provide additional reports to graphically compare sample counts (per ml or per container) vs size
70USP Workshop on Particles
-
Verify volume accuracy
Verify flow rate
Verify size accuracy
Verify count accuracy
Measure system drift
Perform routinely
Verification tests can overlap
Use Polystyrene latex particles (PSL)
Use standards certified for Size AND Count
71USP Workshop on Particles
-
72USP Workshop on Particles
-
73
Particle Std Size (m)
Previous Method New Method Size
Total Count >
10mRatio
10/15 Mean Size Std Dev
Counts Error
Date Initials
Within in PQ Check
Dil Fac Peak percent Lot #
3/2/2009 AP 15.02 4092 1.54 15.12 0.42 3285 0.6658 JB16A
3/3/2009 AP 15.02 4041 1.48 15.10 0.42 3147 0.5326 JB16A
3/4/2009 AP 15.02 4049 1.51 15.13 0.42 3361 0.7324 JB16A
3/5/2009 AP 15.02 4028 1.68 15.07 0.41 3318 0.3329 JB16A
3/6/2009 AP 15.02 4169 1.73 15.04 0.42 3215 0.1332 JB16A
3/9/2009 AP 15.02 5083 1.33 15.22 0.42 2993 1.3316 JB16A
3/10/2009 AP 15.02 3925 15.28 0.52 3.68 4166 1.7310 JB16A
3/11/2009 AP 15.02 5359 1.33 15.38 0.55 3.67 4106 2.3968 JB16A
3/12/2009 AP 15.02 4383 1.43 15.34 0.50 3.69 3924 2.1305 JB16A
3/13/2009 AP 15.02 4028 1.63 15.19 0.55 3.95 4299 1.1318 JB16A
3/16/2009 AP 15.02 3885 1.65 15.20 0.52 4.01 4202 1.1984 JB16B
3/17/2009 AP 15.02 3865 1.64 15.19 0.50 3.68 2923 1.1318 JB16B
3/18/2009 AP 15.02 3867 1.81 15.10 0.51 3.94 4076 0.5326 JB16B
3/19/2009 AP 15.02 3921 1.79 15.10 0.53 3.94 4360 0.5326 JB16B
3/20/2009 AP 15.02 4011 1.79 15.11 0.53 3.81 3884 0.5992 JB16B
3/23/2009 AP 15.02 4199 1.58 15.36 0.52 3.57 4083 2.2636 JB16B
3/27/2009 AP 15.02 4356 1.36 15.35 0.53 3.68 4331 2.1971 JB16B
3/30/2009 AP 15.02 3928 1.15 15.35 0.52 3.73 4562 2.1971 JB16B
-
Electrozone Sensing (EZS)
Optical Microscopy (human observation)
Image Analysis (camera w/software analysis)Static
Dynamic
Single Particle Optical Sensing (SPOS)
Nano Particle Tracking (NTA)
Fluorescence NTA (FNTA)
Archimedes, mass of individual particles
Spectrex
74USP Workshop on Particles
-
75
Syringes1 ml, 10 ml, 25 ml
Flow rate settings10 100 ml
Sensors (MC05 is added)
Sampling Probes(added short - small bore probe)
USP Workshop on Particles
-
Interchangeable sampling probes, syringes, and sensors
Ensure you have one instrument to manage all applicationsNew MC05 sub - micron sensorChange configuration with no impact to instrument validation
Customized reporting
reportsAdd company logo, user defined descriptors
Customized test recipesProcedure Builder enables the development of unique test recipes for your applicationCopy a recipe and make the changes!
76USP Workshop on Particles
-
77
www.hachultra.comUSP Workshop on Particles
-
78USP Workshop on Particles
USP LO
-
79
-
80
-
81
-
82
-
83USP Workshop on Particles
-
84
-
HACH HRLD150 2.0 to 400mPMS APSS 200 2.0 to 120mHACH HRLD150 1.0 to 150mAccusizer LE400 - 1 1.0 to 400mAccusizer Fx 0.7 to 400m *Accusizer 780 0.5 to 400mHACH MicroCount 05 0.5 to 400mPMS S05 0.5 to 20mPMS S03 0.3 to 20mPMS S02 0.2 to 2mAccusizer FxNano 0.15 to 10m *Nanosight LM20 0.03 to 1m *Coulter Counter 0.4 to 1,000mDynamic Image Analyzers (next section )*Count determined via algorithm Contact Mfr for details
85USP Workshop on Particles
-
USP Workshop on Particles 86
-
USP Workshop on Particles 87
-
88USP Workshop on Particles
-
89
-
90USP Workshop on Particles
-
91USP Workshop on Particles
-
92USP Workshop on Particles
-
93USP Workshop on Particles
-
94
http://www.malvern.com/common/downloads/MRK652.pdf
USP Workshop on Particles
-
95USP Workshop on Particles
-
96USP Workshop on Particles
-
97USP Workshop on Particles
-
www.nanosight.com 98USP Workshop on Particles