User Guide KI 734x Series

Two way Loss & ORL Tester

www.kingfisher.com.au

Page 2: User Guide KI 734x Series Two way Loss & ORL Tester .

Two-way Loss Tester

The fastest and easiest two-way loss tester

Real time display of bi-directional average multi- link loss on both instruments 3 year warranty & calibration cycle No warm up required

360 hours battery life & interchangeable connectors 1200 storage space for dual test results

Page 3: User Guide KI 734x Series Two way Loss & ORL Tester .

Fast & Easy For Bi-directional Testing

Step 1: Set references

Step 2: Press the “Auto test” button

Real time bi-directional average multi- link loss display on both instruments

Page 4: User Guide KI 734x Series Two way Loss & ORL Tester .

Step 1 - Set Reference At Both Ends

Two-way port

Meter port

Test lead

End A

End B

An optic fiber link

Page 5: User Guide KI 734x Series Two way Loss & ORL Tester .

Source off

Source Return Loss Power Meter Auto test

- 52.73 dBm

1310nm

Source on

Auto test Menu Abs / Rel Hold

- 4.01 dBm

1310nm - +

-4.00 dBm

Source on

Auto test Menu Set Ref Abs / Rel Hold

- 0.00 dB R

1310nm - +

-4.00 dBm

Source off

Source Return Loss Power Meter Auto test

- 52.73 dBm

1310nm

Pushing “AUTOTEST”

Pushing “Abs / Rel”

Alternates between 1310 and 1550 nm if 1310 & 1550 nm lasers are installed

Holding “Set Ref” for 3 seconds to set reference for multi-

Pushing “Menu” to be back to the turn-on display

Set Reference Key Press Sequence

Page 6: User Guide KI 734x Series Two way Loss & ORL Tester .

An optic fiber link

Step 2 - Measure Insertion And Return Loss

End A

End B

Source off

Source Return Loss Power Meter Auto test

- 52.73 dBm

1310nm

Pushing “Auto test” at one End.

Page 7: User Guide KI 734x Series Two way Loss & ORL Tester .

Source on

Auto test Menu Hold

- 34.011310nm - +

-4.00 dBm

Alternates between 1310 and 1550 if 1310 & 1550 nm lasers are installed

- +

Source on

Auto test Menu Hold

- 54.17 - +

-4.00 dBm

1310nm

Testing is completed for the link.

A ◄► BdB

A ◄ BdB RL

STORE

Pushing “STORE” to store average Loss, A-to-B Loss, B-to-A

Loss, ORL at end A, ORL at end B of multi-

Displays the average bi-directional link loss of multi-

Displays Loss/Optical Power from A-to-B (‘A ►B’)

Displays Loss/Optical Power from B-to-A (‘A◄ B’)

Displays ORL at A end (‘A ►’)

Displays ORL at B end (‘◄ B’)

Back to the average bi-directional link loss display

Pre

ss

‘Sele

ct’

Pre

ss

‘Men

u’

Measure Insertion And Return Loss - Continue

Page 8: User Guide KI 734x Series Two way Loss & ORL Tester .

End A

End B

An optic fiber link

Communication During Testing

Source on

Auto test Menu Hold

- 34.011310nm - +

-4.00 dBm

A ◄► BdB

After completing one link test,

pushing “Hold” will freeze the

display on both units and the unit

at End B beeps twice. This tells

End B that a test has been

completed.

Source on

Auto test Menu Hold

- 34.01 - +

-4.00 dBm

A ◄► BdB

Then, End B connects to the next

link and pushes “Hold”. This releases

the display on both units and the unit

at End A beeps once. This tells End A

that he is ready for the next test.

Increases the productivity in high fiber-count systems.

Page 9: User Guide KI 734x Series Two way Loss & ORL Tester .

Set Ref In Autotest

Source on

Auto test Menu Set Ref Hold

- 34.011310nm - +

A ◄►BdB

Displays average link loss for multi-.

Hold “Set Ref” for 3 seconds will set references on both units for multi-.

Page 10: User Guide KI 734x Series Two way Loss & ORL Tester .

View Ref Value In Autotest

Source on

Auto test Menu Abs / Rel Hold

- 34.011310nm - +

A ►BdB R

Displays link loss and reference value for multi- for A-to-B.

Source on

Auto test Menu Abs / Rel Hold

- 32.241310nm - +

A ◄ BdB R

Displays link loss and reference value for multi- for B-to-A.

-8.20 dBm

-7.80 dBm

Displays the average bi-directional link loss of multi-

Displays Loss/Optical Power from A-to-B (‘A ►B’)

Displays Loss/Optical Power from B-to-A (‘A◄ B’)

Displays ORL at A end (‘A ►’)

Displays ORL at B end (‘◄ B’)

Back to the average bi-directional link loss display

Pre

ss

‘Sele

ct’

Pre

ss

‘Men

u’

- +

Unit A reference value

Unit B reference value

Page 11: User Guide KI 734x Series Two way Loss & ORL Tester .

Pass/Fail Assessment

In Two-way testing mode, press ‘Level’ to enter Pass/Fail display mode.

To exit press ‘Level’ again, or press ‘Select’.

Source on

<<< Auto test Menu Hold

PASS1310nm - +

A

Level

Select

Page 12: User Guide KI 734x Series Two way Loss & ORL Tester .

Insertion Loss Pass/Fail Threshold Set Up

Multi- of Insertion Loss Pass/Fail criteria can be set.

Shift Default Level

Cancel Set

Shift Source off

SEt

Menu

15.731310nm - +

dB

Select another two-way operational wavelength to set up Average Insertion Loss Pass/Fail threshold.

Enter Power Meter mode.Select two-way operational wavelength.

Use ‘Shift’ + ‘Level’ to set Average Insertion Loss Pass/Fail threshold, push ‘Set’ to store.

- +

Page 13: User Guide KI 734x Series Two way Loss & ORL Tester .

Return Loss Pass/Fail Threshold Set Up

Multi- of Return Loss Pass/Fail criteria can be set.

Shift Default Level

Cancel Set

Select another wavelength to set up Return Loss Pass/Fail threshold.

Use ‘Shift’ + ‘Level’ to set Return Loss Pass/Fail threshold, push ‘Set’ to store.

- +

Shift Source off

SEt

Menu

-10.571310nm - +

dB RL

Enter Return Loss mode.Select wavelength.

Page 14: User Guide KI 734x Series Two way Loss & ORL Tester .

The Contents of User Memory

Average loss for multi-

Reference of both units for multi-

ORL for multi-

Nominal output power for multi-

Location info by flashing symbol of either A or B

Page 15: User Guide KI 734x Series Two way Loss & ORL Tester .

KITS™ testing software