Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine...
Transcript of Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine...
![Page 1: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/1.jpg)
www.rfb.bio
Use of Target Values in EQA
Dr. Anja Kessler Bonn, Germany
1
![Page 2: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/2.jpg)
www.rfb.bio
2
Example: Progesterone 580 participants Measurement principles:
Luminescence detection Radioactivity detection Fluorescence detection Mass spectrometry ...
Which result is accurate? Are the results comparable within certain limits?
External Quality Assessment
![Page 3: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/3.jpg)
www.rfb.bio
Traceability in Laboratory Medicine
SI-Unit (definition)
Primary calibrator
Secondary calibrator
Mf.´s working(master) calibrator
Mf.´s productcalibrator
Secondary referencemeasurement
procedure
Mf.´s selectedmeasurement
procedure
Mf.´s standingmeasurement
procedure
End-user´s routinemeasurement
procedure
Routine sample
RESULT
Trac
eabi
lity
BIPM:International Institute of Weightsand Measures
NMI:National MetrologicalInstitute
ACL:AccreditedCalibrationLaboratory
MCL:Manufacturer´sCalibrationLaboratory
ML:Manufacturer´sLaboratory
Mf:Manufacturer
Primary referencemeasurement procedure
Uncertaintyµc(y)
Adapted from ISO 17511
3
![Page 4: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/4.jpg)
www.rfb.bio
Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO2, pO2 Valproic acid …
Traceability for Measurands in EQA - present state-
4
![Page 5: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/5.jpg)
www.rfb.bio
Traceability in Laboratory Medicine
SI-Unit (definition)
Primary calibrator
Secondary calibrator
Mf.´s working(master) calibrator
Mf.´s productcalibrator
Secondary referencemeasurement
procedure
Mf.´s selectedmeasurement
procedure
Mf.´s standingmeasurement
procedure
End-user´s routinemeasurement
procedure
Routine sample
RESULT
Trac
eabi
lity
BIPM:International Institute of Weightsand Measures
NMI:National MetrologicalInstitute
ACL:AccreditedCalibrationLaboratory
MCL:Manufacturer´sCalibrationLaboratory
ML:Manufacturer´sLaboratory
Mf:Manufacturer
Primary referencemeasurement procedure
Uncertaintyµc(y)
Adapted from ISO 17511
5
![Page 6: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/6.jpg)
www.rfb.bio
RiliBÄK – Table B1a
......
![Page 7: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/7.jpg)
www.rfb.bio
Kit Dependent Target Values
7
![Page 8: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/8.jpg)
www.rfb.bio
Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO2, pO2 Valproic acid
…
ALT ALP AST CK GGT LDH HbA1c …
8
Traceability for Measurands in EQA - present state-
target value dependent on method/kit
![Page 9: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/9.jpg)
www.rfb.bio
Traceability in Laboratory Medicine
SI-Unit (definition)
Primary calibrator
Secondary calibrator
Mf.´s working(master) calibrator
Mf.´s productcalibrator
Secondary referencemeasurement
procedure
Mf.´s selectedmeasurement
procedure
Mf.´s standingmeasurement
procedure
End-user´s routinemeasurement
procedure
Routine sample
RESULT
Trac
eabi
lity
BIPM:International Institute of Weightsand Measures
NMI:National MetrologicalInstitute
ACL:AccreditedCalibrationLaboratory
MCL:Manufacturer´sCalibrationLaboratory
ML:Manufacturer´sLaboratory
Mf:Manufacturer
Primary referencemeasurement procedure
Uncertaintyµc(y)
Adapted from ISO 17511
9
![Page 10: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/10.jpg)
www.rfb.bio
Traceability in Laboratory Medicine
SI-Unit (definition)
Primary calibrator
Secondary calibrator
Mf.´s working(master) calibrator
Mf.´s productcalibrator
Secondary referencemeasurement
procedure
Mf.´s selectedmeasurement
procedure
Mf.´s standingmeasurement
procedure
End-user´s routinemeasurement
procedure
Routine sample
RESULT
Trac
eabi
lity
BIPM:International Institute of Weightsand Measures
NMI:National MetrologicalInstitute
ACL:AccreditedCalibrationLaboratory
MCL:Manufacturer´sCalibrationLaboratory
ML:Manufacturer´sLaboratory
Mf:Manufacturer
Primary referencemeasurement procedure
Uncertaintyµc(y)
Adapted from ISO 17511
10
![Page 11: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/11.jpg)
www.rfb.bio
RiliBÄK – Table B1a
......
![Page 12: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/12.jpg)
www.rfb.bio
Enzymes – Situation before Standardization
12
![Page 13: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/13.jpg)
www.rfb.bio
Enzymes – Situation after Standardization
13
![Page 14: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/14.jpg)
www.rfb.bio
Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO2, pO2 Valproic acid
…
ALT ALP AST CK GGT LDH HbA1c …
Creatinine Glucose Total cholesterol Cortisol 17ß-estradiol Progesterone Testosterone …
14
Traceability for Measurands in EQA - present state-
target value dependent on method/kit
reference measurement procedure target value independent on method/kit
![Page 15: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/15.jpg)
www.rfb.bio
Traceability in Laboratory Medicine
SI-Unit (definition)
Primary calibrator
Secondary calibrator
Mf.´s working(master) calibrator
Mf.´s productcalibrator
Secondary referencemeasurement
procedure
Mf.´s selectedmeasurement
procedure
Mf.´s standingmeasurement
procedure
End-user´s routinemeasurement
procedure
Routine sample
RESULT
Trac
eabi
lity
BIPM:International Institute of Weightsand Measures
NMI:National MetrologicalInstitute
ACL:AccreditedCalibrationLaboratory
MCL:Manufacturer´sCalibrationLaboratory
ML:Manufacturer´sLaboratory
Mf:Manufacturer
Primary referencemeasurement procedure
Uncertaintyµc(y)
Adapted from ISO 17511
15
![Page 16: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/16.jpg)
www.rfb.bio
RiliBÄK – Table B1a
......
![Page 17: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/17.jpg)
www.rfb.bio
Survey ‚Hormones‘
17
![Page 18: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/18.jpg)
www.rfb.bio
http://www.bipm.org : JCTLM database
REFERENCE MEASUREMENT SYSTEMS
• Reference materials
• Reference measurement procedures
• Services of reference measurement laboratories
Joint Committee for Traceability in Laboratory Medicine
18
![Page 19: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/19.jpg)
www.rfb.bio
The laboratory has to be accreditated according to ISO/IEC 17025 and ISO 15195.
The laboratory has to participate regularly in collaborative surveys for calibration laboratories.
The laboratory has to make use of a reference measurement procedure approved according to ISO 15193 and reference materials approved according to ISO 15194.
Reference Measurement Systems
19
![Page 20: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/20.jpg)
www.rfb.bio
20
EQA for Calibration Laboratories
![Page 21: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/21.jpg)
www.rfb.bio
21
EQA for Calibration Laboratories
![Page 22: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/22.jpg)
www.rfb.bio
22
EQA for Calibration Laboratories
![Page 23: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/23.jpg)
www.rfb.bio
23
680 routine labs
3 calibration labs
Linking Routine and Calibration Laboratories
Limits of Acceptance: 35%
HM 1/16 vs. RELA2014
Testosterone
![Page 24: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/24.jpg)
www.rfb.bio
The implementation of quality assurance for clinical measurands requires the support of
• National Metrology Institutes • Scientific Societies of Clinical Chemistry and Laboratory Medicine • Accreditation Bodies
• Suppliers of Certified Reference Materials • Calibration Laboratories (also known as: Reference Measurement Laboratories)
• IVD Industry • External Quality Assessment Organizers • Legal Authorities
• Medical Laboratories themselves
Implementation of Traceability
24
![Page 25: Use of Target Values in EQA€¦ · Adapted from ISO 17511 3 . Alpha-fetoprotein Carbamazepine C-reactive protein Ferritin IgA, IgG,IgM pCO 2, pO 2 Valproic acid … Traceability](https://reader035.fdocuments.us/reader035/viewer/2022071511/612ff5471ecc51586943c91e/html5/thumbnails/25.jpg)
www.rfb.bio
About 40 reference measurement systems are available to assign
independent target values for interlaboratory comparisons. (most
frequently analysed measurands)
However, many measurands still have to be evaluated by method/kit
dependent target values.
Scientific work is necessary to develop reference measurement systems for
further measurands in laboratory medicine.
The implementation of traceability in laboratory medicine and the
improvement of comparability of EQA results depend on the contribution of
many parties.
Summary and Outlook
25