  • USD 2145a

    Palltronic AquaWIT XC SystemAutomated integrity test system for sterile gas and vent filters

  • 1

  • 2Introduction ... The Palltronic AquaWIT XC system offers

    major benefits for integrity testing of hydrophobic gas filters. All stages

    of the Water Intrusion test can now be performed fully automatically

    on gas filters - and on line. The operator simply connects the

    Palltronic AquaWIT XC unit to the filter

    system, inputs test parameters on

    the easy to use touchscreen and starts the

    test sequence . The filling , testing ,

    draining and drying are performed without

    further operator involvement and a detailed

    test result is saved and printed.

    With Pall direct flow measurement technology

    and a hydraulic bridge link to the test filter, accurate

    and reproducible results are ensured even during changes in

    environmental temperature. The new XC version provides enhanced

    memory and 21CFR Part 11 features.

    It is expected that there be some program for integrity testing this (vent) filter to assurethat it is intact. There are methods for integrity testing of vent filters in-place.FDA Guidance Document on High Purity Water Systems for Pharmaceutical Use 1996

    The integrity of critical gas and air vent filters should be confirmed after useEuropean GMP

  • Water Breakthrough

    The Water Intrusion Test

    3

    Pharmaceutical regulatory bodies now require

    that critical gas filters are integrity tested

    regularly, preferably in-situ. The Water Intrusion

    test is non-destructive and can be performed

    on a dry hydrophobic filter. Unlike the Forward

    Flow or "Bubble Point" type tests there is no

    need to wet with alcohol. The Water Intrusion test

    is suitable for a wide range of Pall filters including

    Emflon PFR and CPFR filters.

    To perform the test, the upstream side of the

    f i l ter assembly is f i l led with water and a

    pre-determined gas pressure is applied. The

    resulting water flow through the membrane is

    measured directly on the upstream side.

    The Palltronic AquaWIT XC device performs

    all parts of the Water Intrusion test, including

    filling and draining of the filter system, fully

    automatically. The mobile system can easily

    be moved from site to site, enabling automatic

    testing to be performed in-place, conveniently and

    cost effectively on many filter installations, both

    large and small.

    Direct Flow Measurement

    The Palltronic AquaWIT XC unit measures water

    flow during the Water Intrusion test using the

    direct flow method. The benefits of using direct

    flow measurement include:

    Very high measurement accuracy

    Upstream volume measurement and flow

    calculations are eliminated, reducing

    potential measurement errors and

    simplifying validation

    Real time measurement and graphical

    display

    Test times can be up to 50% shorter than

    pressure decay devices

    For further information on the basis of the Water Intrusion test see "Studies onthe Theoretical Basis of the Water Intrusion Test (WIT)," R Jaenchen, J Schubert, S Jafari and A West, European Journal of Parenteral Science 1997,volume 2, No. 2, pages 39 - 45.

    Evaporative Flow

    Pressurized Water

    Integral Filter

    Defective Filter

    Pressurized Water

  • 4Accurate and reliable

    In validation studies flow was measured at

    different environmental temperatures using a

    small filter with a typical water intrusion fiow rate

    of 0.04 mL/min to represent worst case

    conditions. Even under conditions where the

    environmental temperature of the filter was 20C

    (36F) higher than water temperature, an accurate

    test result was obtained.

    Validation work has also shown that accurate

    measurements can be performed using filter

    installations at a distance of at least 10 m (33 ft)

    and a height of 2.5 m (8.2 ft) or more. Pall can

    assist in qualifying the Palltronic AquaWIT XC

    device for increased distances.

    Hydraulic Bridge

    The hydraulic bridge removes all air from the

    upstream fi l ter system and tubing. Flow

    measurement is then performed in a more

    controlled environment within the Palltronic

    AquaWIT XC unit. Stable results can be obtained

    even when environmental temperatures are

    changing by 10C (18F) or more. Direct flow

    and real time measurement further adds to the

    accuracy, speed and reliability of the test result.

    A Versatile System

    The Palltronic AquaWIT XC unit has been

    designed to test vent filters in positions with

    difficult access where environmental temp-

    eratures may be fluctuating. The hydraulic bridge

    permits the Water Intrusion test to be

    performed on:

    Filters in elevated positions, for example,

    product storage and WFI tanks

    Filters which are inside equipment, for

    example lyophilizers or sterilizers which

    are difficult to access

    Several different filter installations testing

    from a fixed point

    Making Water Intrusion Testing Easier

    Wat

    er in

    trus

    ion

    flow

    rat

    e m

    L/m

    in

    Environmental Temperature C

    0.14

    0.12

    0.1

    0.08

    0.06

    0.04

    0.02

    020C

    Water intrusion limit value

    30C 40C

    Direct Flow Measurement using a Hydraulic Bridge

    Sterilizer vent Lyophilizer vent

    WFI or producttank vent

    Variable environmental temperature

    Palltronic AquaWIT XC System

    Palltronic direct flow measurement system

    Measurement chamber

    Air space

    Water tank

    Air space replaced by water

    Hydraulic bridge

    Water temperature 18OC

    Filling machine

    Water Intrusion Testing from a fixed point

    WIT result

  • 5Purpose designed for the Production Environment

    Overview

    The Palltronic AquaWIT XC integrity test device

    is based on Pall's experience with integrity test

    equipment and automated systems over more

    than 20 years. The Palltronic AquaWIT XC

    system is contained within a stainless steel cabinet

    and includes a detachable Palltronic Flowstar

    XC unit. External surfaces are easy to clean and

    ideal for clean room application. The unit is robust

    and splash-proof to IP54 standard. The system

    is CE marked.

    Touch Screen for Easy Operation

    Modern touch screen technology eliminates the

    need for a separate keyboard and means that

    commands can be entered simply with the touch

    of a finger. The large screen clearly displays test

    data and results both graphically and numerically.

    Adjustable contrast means that the display can

    be optimized to suit process conditions and user

    preferences.

    Remote Operation

    An RS232C interface allows the Palltronic

    AquaWIT XC unit to be integrated into a process

    control system. See Pall publication USD2133 for

    more information.

    Integration into Automated Processes

    Addit ional pneumatic ports are f itted as

    standard and can be used to operate process

    valves in the filter installation.

    Pall offers a complete service for integration

    of the Palltronic AquaWIT XC unit into automated

    processes.

  • 6CIP and SIP functions

    The simple to use clean-in-place and steam-

    in-place features help to ensure system cleanliness,

    control of bioburden and prevention of cross-

    contamination in the liquid wetted parts of the system.

    A steam condensate manifold collects condensate and

    minimizes discharge of steam from the instrument.

    Used test water can either be returned to the reservoir

    or drained to waste.

    Detachable Palltronic Flowstar XC Function

    The Palltronic Flowstar XC unit can be removed from

    the Palltronic AquaWIT XC cabinet and used as a

    stand-alone instrument. Standard Forward Flow and

    Bubble Point type tests can be performed, making

    thePalltronicAquaWITXC instrumentextremelyversatile.

    Precision control and engineering

    In addition to the hydraulic bridge, the following

    features ensure that a reliable result is always obtained:

    A temperature sensor in the water tank

    measures the water temperature to ensure

    testing within a validated range

    An internal pressure sensor automatically

    compensates for hydrostatic pressure on

    the measurement chamber

    High level sensor in the water tank for

    automatic filling

    Low level sensor in the water tank gives a

    warning when the tank is empty

    Temperature sensor in the water tank

    monitors the temperature during steaming

    Fully Mobile

    The wheel mounted unit is equipped with a handrail

    ensuring maximum mobility.

  • 7The progress of all test results can be viewed

    as bar charts and line graphs. This provides

    useful information during validation and routine

    production testing.

    Up to 10,000 test results can be stored in

    the memory of the Palltronic AquaWIT XC

    instrument. Test results can also be transferred to

    a computer via an Ethernet port for secure archiving

    and records retention.

    Test results including line graphs are automati-

    cally printed on the internal printer which is silent

    and fast. In addition, the printer casing has been

    designed to protect the printer and make it

    suitable for clean room environments. The

    112 mm (4.4") width paper provides a clear and

    detailed printout that is both permanent

    and waterproof.

    A new notes field allows operators to input

    additional information to all electronic records.

    Electronic Records and Signatures

    The regulatory requirements for the handling of

    electronic records and signatures are well defined

    by the FDA in Code of Federal Regulations 21

    Part 11. To support users of Palltronic AquaWIT

    XC test equipment, the software includes a user

    ID and password system which has been

    developed around the requirements of 21 CFR

    Part 11 and contains the following features:

    System access limited to authorized

    individuals

    Three levels of access: System

    Administrator, Supervisor and Operator

    Electronic signing of all electronic records

    Date and time stamped audit trail of all

    electronic records including: test results,

    test programs, user IDs and system

    settings

    Electronic records of test results cannot be

    altered

    Users are forced to change their password

    at a time interval set by the system

    administrator

    Testing WFI Vent Filters

    The Palltronic AquaWIT XC device is part of the

    Pall system for WFI tank vent filters. See Pall

    publication USD2121 for further information.

    Comprehensive Documentationand 21CFR Part 11 Features

  • Schematic drawing of the Palltronic AquaWIT XC integrity test device

    8

    Palltronic Flowstarintegrity test device

    Filter element

    and housingV2

    V5

    V7

    V6

    L1

    C

    V

    A

    D

    P

    V13

    V3

    V4

    V9E

    F

    B

    V8T

    V1

    V10

    V11

    V12

    AIR INVENT

    TEST

    AIR INFF/BP TEST

    V10V11V12V80V16

    Palltronic Flowstar XCInstrument

    Control Box

    Level sensor

    MeasurementChamber

    Drain test water to wastevalve

    Water tank fill

    Water tank drain

    } Additional pneumatic ports

    Vent

    Hydraulic bridge link

    Hydraulic bridge link

    L1

    T

    P

    Temperature sensor

    Pressure sensor

    L2

    L3

    Water Tank

  • 9Calibration

    The Palltronic AquaWIT XC test device has

    been designed for assurance of instrument

    cal ibrat ion and continuous, trouble

    free operation. An internal f i l ter ensures

    contaminant-free air. Units can be calibrated on

    or off-site by an experienced Pall technician,

    and we recommend calibration annually with a

    full preventative service every 3 years.

    Our international servicing centers are

    purpose designed and fully equipped for

    calibration and repair. All flow and pressure

    measurement equipment is regularly

    re-calibrated against international standards. Full

    traceabi l i ty and documentation is

    maintained, consistent with the highest

    standards in the pharmaceutical industry.

    Documentation

    The Palltronic AquaWIT XC system is supplied

    with comprehensive documentation including:

    Calibration certificate

    System drawings

    Full component list

    CE certificates

    Instruction manual

    Factory acceptance test results

    EN10204.3.1b certification for all wetted

    stainless steel parts

    Calibration

  • Validation, Documentation and Training

    10

    Validation

    The Palltronic AquaWIT XC software has been

    developed in accordance with current

    Good Automated Manufacturing Practice (GAMP)

    guidelines.

    Support documentation available for purchase

    include:

    GAMP documentation

    A generic Operational Qualification (OQ1)

    and system specific IQ and OQ2

    documents

    IQ/OQ documentation can be customized

    to your requirements

    A detailed validation guide is available and GAMP

    documentation can be audited at Pall sites.

    Pall Validation Services can provide on-site or

    off-site IQ/OQ testing and assist with Performance

    Qualification protocols.

    Training Services

    Full training on all aspects of operation is

    avai lable from our training special ists.

    Customized programs can be provided, ranging

    from seminars on the basic principles of

    the integrity test to advanced operator

    training. Certificates are issued to all attendees

    for inclusion into personal training records.

  • 11

    Technical Specifications

    Product

    Part Number AW02XC (Standard)AW02XCH (High Flow)

    Physical Dimensions

    Weight empty 100 kg (220 lb)

    Height, Width, Depth 1100 mm, 580 mm, 840 mm(43 in, 22 in, 33 in)

    Water tank volume 14 Liters (3.7 US gallons)

    Filter Tests

    Water Intrusion Test (fully-automatic)

    Forward Flow Test (semi-automatic)

    Bubble Point Test (semi-automatic)

    Combined FF / BP Test (semi-automatic)

    Function Tests

    Self Test

    Flow Check Test

    Valve Check

    Other Functions

    Clean-in-place

    Steam-in-place

    Level sensors for automatic filling and draining of filter system

    Pressure sensor to measure hydrostatic pressure

    Temperature measurement of water or steam in tank

    Adjustable filter system filling and draining pressures

    Adjustable filter and Palltronic AquaWIT XC system drying pressures and times

    Air Clean

    Test result storage - up to 10,000 tests

    Test programs - up to 500 test programs

    Up to 250 active user IDs

    Standard Languages

    English

    French

    German

    Italian

    Spanish

    US English

    Japanese

    Units

    psi

    mbar

    kPa

    kp/cm2

    Communication Ports

    COM1 RS232C for external printer or PC Ethernet Port

    Measuring Range

    Water Intrusion Test 0.05 - 50 mL/min water

    Forward Flow TestStandard 0.1 - 200 mL/minHigh Flow 0.1 - 2,000 mL/min

    Bubble Point Test 700 - 6000 mbar (10 - 87 psi)

    Water Intrusion Test 0.05 - 50 mL/min

    Resolution

    Water Intrusion Test 0.01 mL/min

    Forward Flow TestMeasured Flow >10 mL/min: 0.1 mL/minMeasured Flow

  • 12

    Electrical Safety

    CE Marked

    Maximum Steam Temperature 125C (257F)

    Design Specifications

    AD Merkblatter

    Pneumatic Specifications

    Maximum gas supply pressure 8 bar (116 psi)

    Minimum gas supply pressure for automatic WIT test 5600 mbar (82 psi)

    Test pressure range 20 - 6000 mbar (0.3 - 87 psi)

    Connections

    Compressed air inlet

    'Stubli' nipple RBE03.6150

    Compressed air outlet

    Stubli coupling RBE03.2904

    Water line to fill filter 1" 'Tri-Clamp'

    Water line to fill tank 1" Tri-Clamp

    Water drain (bottom level) 1" Tri-Clamp

    Environmental Conditions

    Splash proof IP54

    Operating Temperature +5C to +40C (+41F to +104F)

    Storage Temperature -20C to +70C (-4F to +158F)

    Humidity 95% rh (no condensation)

    Touch Screen Specifications

    Size 238 mm (9.4 inch) 640 x 480 pixel

    Features black & white, illuminated background, adjustable contrast

    Materials

    Stainless Steel - product wet DIN 1.4404*

    Stainless Steel - frame DIN 1.4301**

    Seals EPDM

    *DIN 1.4404 element composition limits are comparable to 316L.

    ** DIN 1.4301 element composition limits are comparable to 304.

  • 2200 Northern BoulevardEast Hills, New York 11548-1289

    +1 800.717.7255 toll free

    +1 516.484.5400 phone

    +1 516.625.3610 fax

    [email protected] e-mail

    Europa House, Havant StreetPortsmouth PO1 3PD, United Kingdom

    +44 (0)23 9230 3303 phone

    +44 (0)23 9230 2506 fax

    [email protected] e-mail

    Pall Corporation has offices and plants throughout the world in locations including:Argentina, Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Hong Kong,India, Indonesia, Ireland, Italy, Japan, Korea, Malaysia, Mexico, the Netherlands, New Zealand, Norway, Poland, Puerto Rico, Russia, Singapore, Spain, South Africa, Sweden,Switzerland, Taiwan, Thailand, United Kingdom, United States and Venezuela. Distributors are located in all major industrial areas of the world.

    Copyright 2003, Pall Europe Ltd. , Pall, Palltronic and Emflon are trade marks of Pall Corporation. Stubli is a trademark of Stubli International AG. Tri-Clamp is a trademark of Alfa Laval Inc.Filtration, Separation, Solution is a service mark of Pall Corporation. indicates a Pall trademark registered in the USA.

    PELEH/01-ZP.SH/CS/8.2003

    Visit us on the web at www.pall.com/biopharmaceutical