Update on Hemodynamic Monitoring

66
Gozde Demiralp, MD Assistant Professor of Anesthesiology Department of Anesthesiology Division of Critical Care Medicine University of Oklahoma, College of Medicine Oklahoma City, Oklahoma Update on Hemodynamic Monitoring

description

hemodynamic monitoring

Transcript of Update on Hemodynamic Monitoring

Page 1: Update on Hemodynamic Monitoring

Gozde Demiralp, MD Assistant Professor of Anesthesiology

Department of Anesthesiology

Division of Critical Care Medicine

University of Oklahoma, College of Medicine

Oklahoma City, Oklahoma

Update on Hemodynamic Monitoring

Page 2: Update on Hemodynamic Monitoring

Disclosure I have no affiliations to disclose.

There will be display of new technologies and monitors

during this presentation. All the opinions that will be displayed here reflects scientific data that is supported by literature.

Page 3: Update on Hemodynamic Monitoring

Outline: Reeeee ee Beeeee: “What have we been doing and Why?” Ceeeeee Oeeeee Meeeeee Invasive vs. Noninvasive Methods

Oee Meeeeeee e Nee Teeeeeeeee

Arterial Wave Form Analysis Veeeee Oeeeeeee Why does it matter? How can we monitor it?

Nee Ceeeeeee ee Heeeeeeeeee Meeeeeeeeee

Page 4: Update on Hemodynamic Monitoring

Hemodynamic Monitoring Monitoring critically ill patients: What are we really worried about? “Tissue Hypoperfusion” What do we really want to monitor? “Adequate Oxygen Delivery” Definition of Shock: “Inadequate tissue perfusion affecting multiple organ systems.”

Page 5: Update on Hemodynamic Monitoring

Hemodynamic Monitoring: Review of Basics

Oxygen Delivery (DaO2) = CaO2 x CO x 10 = 1000 ml/min Arterial Oxygen Content (CaO2) = (0.0138 x Hgb x SaO2) + (0.0031 x PaO2 )

= 20.1 ml/dl Cardiac Output (CO) = SV x HR =4-8 L/min

Page 6: Update on Hemodynamic Monitoring
Page 7: Update on Hemodynamic Monitoring

Assessment of Adequate Oxygen Delivery • Tachycardia Diaphoresis

• Tachypnea Cool Peripheries • Prolonged Capillary Refill • Increased Temperature Difference • Reduced Level of Consciousness • Decreased Urine Output

Clinical Estimation

• Lactate • Base Excess Lab Values

• Invasive & Noninvasive

• New vs. Old

Measurement of CO and/or

SvO2%

Page 8: Update on Hemodynamic Monitoring

Basics of Hemodynamic Monitoring Heart Rate

Blood Pressure ( Noninvasive vs. Invasive) Central Venous Pressure

Page 9: Update on Hemodynamic Monitoring

Basics of Hemodynamic Monitoring

Patient is hypotensive, now what? Aline, CVP, ECG, Sato2% in place… Why is he hypotensive? What is the first step?

Page 10: Update on Hemodynamic Monitoring

Figure taken from Reference 1.

Why is he hypotensive?

(Contractility)

(Afterload)

(Preload)

Page 11: Update on Hemodynamic Monitoring

Basics of Hemodynamic Monitoring IS PATIENT VOLUME RESPONSIVE?

Primary resuscitation question is whether the patient will

increase their cardiac output in response to intravascular volume infusion.

That is why almost always the first step is to give fluids and see patient’s response, in case of a hypotension.

It is important to determine “ volume responsiveness” !!!

Page 12: Update on Hemodynamic Monitoring

Volume Responsiveness: Frank Starling Curve…

Frank starling curve

Page 13: Update on Hemodynamic Monitoring

How to asses volume responsiveness?

Volume Responsiveness was defined as increase in CO by 15% or more, by 500 ml fluid bolus.

Passive Leg Raising*

CVP-MAP Relationship CVP vs. EDV/EDVI?

Stroke Volume Variations Pulse Pressure Variations New Generation Monitors

Systolic Pressure Variations

Page 14: Update on Hemodynamic Monitoring

CVP Central venous pressure (CVP) =Peeeeee e

CVP can reflect a volume increase in RA pressures or decrease in RV contractility can be both. Need to be monitored in conjunction with other monitors.(CVP&MAP)

The main limitations of CVP monitoring: (a) it does not allow to measure cardiac output (b) it does not provide reliable information on the status of the pulmonary circulation in the presence of left ventricular dysfunction.

Page 15: Update on Hemodynamic Monitoring

Figure taken from Reference 1.

Why is he hypotensive?

(Contractility)

(Afterload)

CVP (Preload)

Page 16: Update on Hemodynamic Monitoring

Measurement of Cardiac Output:

Thermo Dilution

Dye/Indicator Dilution

Arterial Pulse Pressure Analysis**

Page 17: Update on Hemodynamic Monitoring

Measurement of Cardiac Output: Thermodilution ( TD): Dilution of Temperature (Cold NS) Area Under the Curve (AUC) Pulmonary Artery Catheter ( PACs) or Newer Generation CVL w sensors Intermittent vs. Continuous TD Dye/Indicator Dilution: Same Technique ( Dilution of dye or indicator instead of NS)

Dye(indocyanine green) vs. Indicator( Lithium) A line vs. CVL , no need for PACs.

Page 18: Update on Hemodynamic Monitoring

Thermodilution(TD)- Measurement of Cardiac Output

Intermittent Bolus Pulmonary Artery Thermodilution Technique:

1. 1870-oldest defined CO measurement system

2. Cold solution – proximal port/RA – distal port- thermistor

3. Measures RV outflow, not systemic CO.

4. Not perfect because

Volume & temp of cold solution may vary.

Baseline blood temperature changes

Intracardiac shunts ( R_L ; L_R)

Mechanical and Spontaneous Resp. Cycle effects.

Page 19: Update on Hemodynamic Monitoring

Thermodilution(TD)- Measurement of Cardiac Output

Figure taken from reference # 2.

Page 20: Update on Hemodynamic Monitoring

Thermodilution(TD)- Measurement of Cardiac Output Continuous Pulmonary Artery Thermodilution Technique: Same principle, New technology. Electric filament ( vigilance) vs. thermal coil(optiQ) Intermittently heating R heart Temp change is being detected at tip of PAC. Every 30 – 60 second data collection but… Displayed value is average of 5-15 minutes of data collection.

Advantage: 1. Continuous trend of CO 2. Decrease operator workload 3. Reduce infection risk ( no more boluses into system)

Page 21: Update on Hemodynamic Monitoring

Pulmonary Artery Catheters (PACs): Both TD techniques to calculate CO requires PACs. Standard of reference for cardiac filling pressures.

KEY advantage: Simultaneous measurements of other hemodynamic parameters: CO Pulmonary Artery Pressures ( PAD, PAOP) Right Sided Filling Left Sided Filling SvO2%

Page 22: Update on Hemodynamic Monitoring

PACs3-6: Studies overall the years had conflicting results: Increased Mortality ? Multiple RCTs were conducted in several RCTs. Conclusion:

No difference in LOS in the ICU No difference in Mortality

No benefit, no harm

“There is no guided therapy tailored towards PAC use.” “PAC is a diagnostic tool, not a therapeutic one.”

Page 23: Update on Hemodynamic Monitoring

Other ways of measuring CO by dilution technique?

• PiCCO & PiCCO2 (Pulsion Medical Systems) • VolumeView (Edwards Life Sciences)

Transpulmonary Thermodilution Methods:

• LiDCO /LiDCOplus/LiDCOrapid ( LiDCO limited)

Lithium Dilution Technique:

• COstatus (Transonic Systems, Inc.)

Ultrasound Indicator Dilution

Page 24: Update on Hemodynamic Monitoring

VolumeView

Taken from website http://www.edwards.com/eu/products/mininvasive/Pages/volumeviewsetup.aspx?volumeview=1

Page 25: Update on Hemodynamic Monitoring
Page 26: Update on Hemodynamic Monitoring

Taken from http://www.transonic.com/ICU/COstatusSetup

Page 27: Update on Hemodynamic Monitoring

LiDCO and PiCCO ***

Page 28: Update on Hemodynamic Monitoring

Less Invasive Methods for CO Calculation

Arterial Waveform Analysis - Waveform Derived CO Measurements.

Page 29: Update on Hemodynamic Monitoring

Arterial Pressure Waveform- Derived CO Measurements

“BP is a product of SV ( CO) and Vascular Resistance.” Advantages: Non invasiveness Works through an already existing aline catheter Continuous CO monitoring May require calibration for arterial compliance and resistance: 1. Lithium Dilution ( LiDCO) 2. Thermodilution (VolumeView, PiCCO)

Page 30: Update on Hemodynamic Monitoring

Arterial Pressure Waveform- Derived CO Measurements

Art

eria

l Pre

ssur

e Wav

efor

m

Ana

lysis

Pressure Recording Analytical Method

(PRAM)

FloTrac Vigileo

Pulse Power Analysis

LiDCOplus

LiDCOrapid

Pulse Contour Analysis PiCCO2

Page 31: Update on Hemodynamic Monitoring

PiCCO2

Pulse Contour Analysis – Aline Wave Form matters ! Area under systolic area ( until dicrotic notch) on arterial

waveform is detected. **

External manual calibration for CO is necessary. This is done via transpulmonary TD every 8 hours or Hourly calibration is required, if patient is hemodynamically

unstable.

Taken from http://www3.pulsion.de/fileadmin/pulsion_share/Products_Flyer/PiCCO_Broschure_E_MPI810205_R02_270208.pdf

Page 32: Update on Hemodynamic Monitoring

PiCCO2

System Requires:

Thermistor Tipped Aline Catheter –Femoral Aline ( longer cath, better signal)

Radial, Axillary, Brachial ok as well

Central Venous Catheter ( thermo indicator solution injection)

Provides:

Dynamic parameters of Preload: SVV, PPV, SPV.

Other Volumetric parameters : GEDV, EVLW, ITBV.

Global end diastolic volume

Extravascular lung water

Intrathoracic Blood Volume

Taken from http://www3.pulsion.de/fileadmin/pulsion_share/Products_Flyer/PiCCO_Broschure_E_MPI810205_R02_270208.pdf

Page 33: Update on Hemodynamic Monitoring

picco2 Foto koy

Taken from http://www.pulsion.com/fileadmin/pulsion_share/Products_Flyer/PiCCO_DE_EN/PiCCO2Device_Datasheet_EN_MPI850305_R05_081110.pdf

Page 34: Update on Hemodynamic Monitoring

LiDCOplus Combination of two monitors LiDCO (Indicator dilution CO calculation & monitor) PulseCO (Software ,CO from aline waveform. )

Unlike PiCCO2, aline wave form is not important in pulse power analysis. (This is why damping effect is also minimized in LiDCO system.) (No need to detect dicrotic notch.) Requires: Aline Peripheral or Central Access (to insert a lithium sensitive sensor) External CO calibration with lithium every 8 hours is necessary. Neuromuscular blockers within 30 min. (due to interactivity w lithium on sensor level)

Lithium therapy receiving patients*

Page 35: Update on Hemodynamic Monitoring

LiDCOrapid

No need for CO calibration with lithium. Replaced by nomogram which is derived form in vivo data to

estimate CO.

Page 36: Update on Hemodynamic Monitoring

lidco Foto koy

Page 37: Update on Hemodynamic Monitoring

Taken from http://www.lidcoplc.com/documents/Brochure.pdf

Page 38: Update on Hemodynamic Monitoring

FloTrac Vigileo

FloTrac Algorithm integrates multiple characteristics of arterial pressure waveform with patient specific demographic data.

No external CO calibration is required. ( Unlike LiDCO, PiCCO)

An arterial aline is adequate to use the monitor. No CVL is required.

Page 39: Update on Hemodynamic Monitoring

Taken from http://ht.edwards.com/sci/edwards/sitecollectionimages/products/mininvasive/flotracbrochurear05917.pdf

Page 40: Update on Hemodynamic Monitoring

FloTrac Vigileo: Advantages Requires No Manual Calibration for CO Calculation. User enters Patient (Pt) Specific Data. ( age, gender, height, weight to initiate the monitoring.) Advanced Arterial Waveform Analysis ( PRAM) by FloTrac Sensor. Pt to pt differences in vasculature. Real time changes in vascular tone. Different arterial sites are acceptable. Central Venous Oximetry (Scvo2) is available. (if used in conjunction with appropriate oximetry CVL.)*

Page 41: Update on Hemodynamic Monitoring

FloTrac Vigileo Monitor - Disadvantages Aline Wave form is important !

Good Arterial Signal Quality is critical for accurate CO calculation.

Still Not Reliable During Arrhythmias For Hemodynamically Unstable Patients Intra Aortic Balloon Pump in use. Ventricular Assist Devices in use.

Page 42: Update on Hemodynamic Monitoring

FloTRac Vigileo

FloTrac Sensor

http://www.edwards.com/products/mininvasive/Pages/flotracsystemanimation.aspx?wt.ac=flotrac

Page 43: Update on Hemodynamic Monitoring

Vigileo Monitor: Can display: CCO CCI SV SVI SVV Every 20 seconds When used with FloTrac Sensor.

Page 44: Update on Hemodynamic Monitoring

New Concepts in Hemodynamic Monitoring

Stroke Volume Variations (SVV) Pulse Pressure Variations (PPV) Systolic Pressure Variations (SPV)

Page 45: Update on Hemodynamic Monitoring

How to asses volume responsiveness?

Volume Responsiveness was defined as increase in CO by 15% or more, by 500 ml fluid bolus.

Passive Leg Raising*

CVP-MAP Relationship CVP vs. EDV/EDVI?

Stroke Volume Variations Pulse Pressure Variations New Generation Monitors

Systolic Pressure Variations

Page 46: Update on Hemodynamic Monitoring

FloTrac Vigileo – Stroke Volume Variation

Page 47: Update on Hemodynamic Monitoring

Can be used as a tool for volume responsiveness in low CO states. SVV > 13% = Volume Responsive

“SVV and PPV are more effective indicators for Volume

Responsiveness then static indicators of preload (CVP, PAOP).”7

Limitations: Pt needs to be on 100% Controlled Mechanical Ventilation. Spontaneous Ventilation & SVV ? Arrhythmias can affect SVV.

Page 48: Update on Hemodynamic Monitoring

FloTrac Pocket Guide

Page 49: Update on Hemodynamic Monitoring

Venous Oximetry

Page 50: Update on Hemodynamic Monitoring

Fick Principle

Oxygen Delivery (DaO2): 1000 ml/min Oxygen Consumption (VO2): 200-250 ml/min Oxygen Extraction Ratio: 25% Normal SvO2%: 65-75%. SvO2< 60% More O2 Extraction Svo2 > 75% Inadequate O2 Utilization

Page 51: Update on Hemodynamic Monitoring

Increased VO2 will be compensated by increased CO.

If this is not adequate, elevated oxygen extraction occurs in the peripheral tissues

SvO2 will drop

SvO2 thus reflects the balance between oxygen delivery and oxygen demand

Page 52: Update on Hemodynamic Monitoring

Supranormal SvO2 (or ScvO2) values do not guarantee adequate tissue oxygenation.( SvO2%>80%)

If tissue is not capable of extracting oxygen, venous return may have a high oxygen content despite persistent cellular hypoxia.

E.g.: Sepsis, Burn Patients, Shunts etc.

Page 53: Update on Hemodynamic Monitoring
Page 54: Update on Hemodynamic Monitoring

Continuous SvO2/ScvO2 Monitoring

Page 55: Update on Hemodynamic Monitoring

How do we measure SvO2% & ScvO2% ?:

Swan Ganz Pulmonary Artery Catheter: “Old Method, New Technology” Advanced Technology Catheters:

Swan-Ganz CCOmbo Pulmonary Artery Catheter: (Combo = Continuous CO + Venous Oximetry)

Swan-Ganz CCOmbo VIP Catheters: CCO &Venous Oximetry Provide additional lumen.

Page 56: Update on Hemodynamic Monitoring
Page 57: Update on Hemodynamic Monitoring
Page 58: Update on Hemodynamic Monitoring

The Vigilance II Monitor

Page 59: Update on Hemodynamic Monitoring

Wwwwwwwwwwwwwwwwwwwwwwwwww wwwwwwwwwwww Wwwwwwwwwwwwwwwwwwwwwwww

Page 60: Update on Hemodynamic Monitoring

Continuous ScvO2 Monitoring – PreSep Catheters & Vigileo Monitors

Taken from http://www.edwards.com/presentationvideos/2007annualreport/prodPreSepOxi.html

Page 61: Update on Hemodynamic Monitoring
Page 62: Update on Hemodynamic Monitoring

Taken from http://ht.edwards.com/sci/edwards/sitecollectionimages/edwards/products/presep/presepbrochure2.pdf

Ref#2

Page 63: Update on Hemodynamic Monitoring

Other Hemodynamic Monitors for ICU2:

Echocardiography (TTE, TEE.)

Tissue Oxygen Saturation Measurements

Thoracic Bioimpedance

Thoracic Bioreactance

Endotracheal CO Monitor

The NICO System (Fick’s Principle for CO2)

Esophageal Doppler

Page 64: Update on Hemodynamic Monitoring

Other Hemodynamic Monitors for ICU2

Page 65: Update on Hemodynamic Monitoring

References: 1) Bigatello,L.M., George, E. Hemodynamic Monitoring. Minerva Anesthesiol 2002;68:219-25

2) Lee, A.J., Cohn Hochman J., Ranasinghe, J. Cardiac output assessed by invasive and minimally invasive techniques. Anesthesiology Research and Practice Volume 2011, Article ID 475151, 17 pages.

3) Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K: Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomized controlled trial. Lancet 2005, 366:472-477.

4) Connors AF, Jr., Speroff T, Dawson NV, Thomas C, Harrell FE, Jr., Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ, Jr., Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA: The effectiveness of right heart atheterization in the initial care ofcritically ill patients. SUPPORT Investigators. JAMA1996, 276:889-897.

5) Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, Laporta DP, Viner S, Passerini L, Devitt H, Kirby A, Jacka M: A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med 2003, 348:5-14.

6) Richard C, Warszawski J, Anguel N, Deye N, Combes A, Barnoud D, Boulain T, Lefort Y, Fartoukh M, Baud F, Boyer A, Brochard L, Teboul JL: Early use of the pulmonary artery catheter and outcomes in patients with shock and acute respiratory distress syndrome: a randomized controlled trial. JAMA 2003, 290:2713-2720

7) Michard, F. Changes in Arterial Pressure During Mechanical Ventilation, Anesthesiology 2005; 103:419-28.

8) Rivers et al. Early Goal-Directed Therapy in the Treatment of Severe Sepsis and Septic Shock. New England Journal of Medicine 2001; 345(19): 1368-77

9) J. Mayer, J. Boldt, R. Poland, A. Peterson, and G. R. Manecke, “Continuous arterial pressure waveform-based cardiac output using the FloTrac/Vigileo: a review and metaanalysis,” Journal of Cardiothoracic and Vascular Anesthesia, vol. 23, no. 3, pp. 401–406, 2009

10) S. Sundar and P. Panzica, “LiDCO systems,” International Anesthesiology Clinics, vol. 48, no. 1, pp. 87–100, 2010

11) Vincent, J., Rhodes,A., Perel, A.,Martin,G., Rocca, D., Vallet, B.,Pinsky,M., Hofer,C., Tebou,J., Pieter de Boode, W., Scolletta,S.,Vieillard-Baron,A.,De Backer,D., Walley,K., Maggiorini,M.,Singer,M. “Clinical review: Update on hemodynamic monitoring - a consensus of 16.” Critical Care 2011, 15:229

12) Van Beest et al. “Clinical review: use of venous oxygen saturations as a goal – a yet unfi nished puzzle”. Critical Care 2011, 15:232

Page 66: Update on Hemodynamic Monitoring