Update of Analysis of the Test Beam Experiment of ScECAL...
Transcript of Update of Analysis of the Test Beam Experiment of ScECAL...
Update of Analysis of the Test Beam Experiment of ScECAL1
Physics Prototype2
The CALICE-ScECAL group3
October 7, 20124
Abstract5
Analysis of the test beam at Fermilab of ScECAL physics prototype (2009) is updated on two6
issues: 1) Temperature correction via the ADC-photon conversion factor is applied for each chan-7
nel of each event. 2) Systematic uncertainties are estimated and evaluated. The response of ScE-8
CAL for the electron beams increase 3% with this temperature correction. The intrinsic energy9
resolution of the ScECAL is determined as 12.9±0.1(stat)±0.4(sys)% and 1.2±0.1(stat)±+0.4/-10
1.2(sys)% for the stochastic term and constant term, respectively.11
1 Intruduction12
The ScECAL measures energy deposit of a particle by using scintillator strips where each strip13
is read out with a pixelated photon detector (PPD). A PPD has saturation behavior due to14
its mechanism, and the saturation is corrected using measured properties of PPD toward the15
incident photons; the number of detected photons as a function of the number of incident photons16
to the PPD. Therefore, the ADC-photon conversion factor is required for each channel to convert17
the ADC output into the number of detected photons. The sensitivity of a PPD depends on18
temperature. Most of temperature dependence of PPD is comprehensively corrected by using19
the ADC-MIP conversion factor depending on temperature in the previous CALICE note [1].20
Although the effect of temperature correction on ADC-photon conversion factor is expected not21
to be large as comparing with the effect of ADC-MIP conversion factor, it is required to be22
measured and evaluated as a second order correction.23
Systematic uncertainties including also the uncertainty from the ADC-photon conversion24
factor are estimated and evaluated in this update.25
2 Temperature correction on the ADC-photon conversion26
factor27
2.1 PPD saturation correction28
The property of PPD saturation is seen with the number of detected photons (fired pixels) as afunction of the number of incident photons in PPD:
Nfired = Npix
(1 − exp
(−ϵNin
Npix
)), (1)
1
where Nfired is the number of photons detected by PPD, Npix is the number of pixels on the29
PPD, ϵ is the photon detection efficiency and Nin is the number of photons incident into the30
PPD sensor. The revers function of Eq. 1 is used to apply the PPD saturation correction by31
inputting the number of detected photons so that the output is the number of incident photons32
into the PPD sensor. The number of detected photons is determined from ADC counts of each33
hit by using ADC-photon conversion factor for each channel. The ADC-photon conversion factor34
for each channel is determined in the following section.35
The only parameter of the revers function of Eq. 1, Npix is determined by fitting Eq. 1 to a36
bench test data in the previous note [2]. One of the update issues is that Npix is determined37
by measuring 72 channels of ScECAL prototype in this note. Therefore, the fluctuation of Npix38
can be evaluated to estimate the uncertainty from the PPD saturation correction. Figure 139
shows the distribution of measured Npix having the mean value of 2428.39±28.88 pixels and the40
standard deviation of 245.07 pixels.
Figure 1: Distribution of the number of pixels, Npix, measured with 72 strips.41
2.2 ADC-photon conversion factor42
An ADC-photon conversion factor is determined by using few-photon spectra of LED righttaken often between beam data taking [2]. The ADC counts corresponding to one photon isan ADC-photon conversion factor for the channel (l, s). The ADC conversion factor is a linearfunction of temperature as the same as the case of the ADC-MIP conversion factor. Therefore,the ADC-photon conversion factor, d(T ; l, s) is expressed as the following:
d(T ; l, s) = d(T0; l, s) + sdl,s · (T − T0), (2)
where l and s is layer number and strip number of this channel, respectively, T0 is a certain43
temperature to give the offset of the function, and sdl,s is slope of this function for the channel44
l, s. Figure 2, left shows the distribution of d(T0 = 20◦C; l, s) and right shows the distribution45
of sdl,s/d(T0 = 20◦C; l, s). The number of entries are ∼70% due to some problems to take LED46
right spectra1. To apply temperature correction, individual d(T0 = 20◦C; l, s) are used for the47
succeeded channels to measure d where it is required to have the d(T0 = 20◦C; l, s) between 17048
and 260 ADC/photon and the uncertainty between 0.2 to 50 ADCs/photon, while the average49
value of these succeeded channels are used for the failed channels. With regard to sdl,s, mean50
of gaussian fit in Fig. 2 right is used for all channels.51
1Some part of channels have large noise when LED right on, and some part of channels have two top pedestalduring LED data taking.
2
Figure 2: Distribution of d(T0 = 20◦C; l, s) (left) and sdl,s/d(T0 = 20◦C; l, s) (right) with the fitting resultof gaussian function. The small peak of d(T0 = 20◦C; l, s) near 235 ADCs/photon is well understood due tothe difference of the product lot of PPD. The mean value by gaussian fit (-1.520± 0.012) is consistent withthe mean of all data within its uncertainty.
2.3 Results of the temperature correction on the ADC-photon con-52
version factor53
Figure 3, left shows the linear behavior of the energy deposit and deviations from the result54
of linear fit and right shows the energy resolution analyzed with d(T ; l, s) as a function of55
temperature measured by two thermocouples put on the detector surface at the data taking.56
Although effect of this temperature correction is not large as comparing with the case of ADC-57
MIP conversion factor, the linear relation between the detector response measured in the number58
of MIPs and incident beam momentum, MIP/GeV/c increase from 127.6 MIPs/GeV/c [1] to59
131.3 MIPs/GeV/c and the linearity is a little improved from the result in [1] except 30 GeV/c60
beam.61
The energy resolutions with higher energy except 30 GeV/c show a little degradation from62
the case in [1]. This phenomenon can be explained as the following: All LED data were taken in63
∼ 20◦C in 2009, and some beam data takings are in higher temperature. Therefore, the detected64
number of photons with PPD are corrected to more than the case without this temperature65
correction, since the value of sdl,s is negative. In such case with the fact that the PPD response66
as the function of the number of incident photons decrease as the number of incident photons67
increase, the energy resolution looks better in the case without this temperature correction.68
Fact that the LED data have been taken in lower temperature than average of the beam69
data taking also explains the increasing the MIP/GeV/c (3%) by this temperature correction .70
3 Systematic uncertainties71
After temperature correction of on ADC-photon conversion factor is implemented, the system-72
atic uncertainties of not only from ADC-photon conversion factor but also from other source of73
the systematic uncertainties are studied and evaluated in this section.74
3.1 Beam momentum fluctuation75
The MTest beam has momentum spread, ∆p/p = 2% as the designed value for 1 - 60 or76
90GeV/c [3]. A calorimetry test for Muon g-2 experiment at the MTest estimates 2.7± 0.3%77
of the beam momentum spread for 1 - 4 GeV/c using Pb/Glass calorimeter [4]. The other78
3
Figure 3: Distribution of d(T0 = 20◦C; l, s) (left) and sdl,s/d(T0 = 20◦C; l, s) (right) with the fitting resultof gaussian. The small peak of d(T0 = 20◦C; l, s) near 235 ADCs/photon is well understood due to thedifference of the product lot. The mean value by gaussian fit (-1.520± 0.012) is consistent with the mean ofall data within its uncertainty.
experiment for a SiFi Calorimeter with tungsten estimates 2.3± 0.3% for 8 GeV/c by using their79
own detector and the results of the previous one [5]. Preceding this study, they has estimated80
2.3% in the range 1.5 - 3.5 GeV/c [6]. From these measurements we estimate the MTest beam81
momentum spread in two incidental beam momentum ranges, 2.7± 0.3% for 2 - 4 GeV/c, and82
2.3± 0.3% for 8 - 32 GeV/c. To estimate the intrinsic energy resolution of ScECAL, these83
momentum spread should be quadratically subtracted from the energy resolution estimated in84
2.3.85
The result of the intrinsic energy resolution evaluated with other systematic uncertainties86
will be shown in 3.8 as the summary of this section.87
3.2 Event selection88
As discussed in [2], six event selections has bean implemented to reduce out of fiducial events. To89
estimate the systematic uncertainties from these selection cuts, the cut variations are measured90
and evaluated.91
The mean value uncertainties from the variation of the cut ranges except for the cut of92
the gravitational center of energy are less than 0.05%. Uncertainties due to the different cut93
variations on the x and y position of gravitational center of energy are listed in Table 1.94
Contributions from the cut variations for the energy resolution are negligibly small (< 0.5%)95
compare to the 3% of the uncertainty comes from the beam momentum spread.96
3.3 ADC-MIP conversion factor97
Individual strip response is calibrated using MIP signals as discussed in [1]. The uncertainties98
on the mean of the measured energy deposit and energy resolution come from the uncertainty99
of the ADC-MIP conversion factor are estimated.100
An ADC-MIP conversion factor is a linear function of temperature of detector. Therefore,101
it is expressed with two parameters; the value of ADC-MIP conversion factor at a certain102
temperature (c(T0; l, s)), and the slope of ADC-MIP conversion factor (scl,s). Propagation of103
the statistical uncertainties of these parameters are studied by using pseudo-experiments in104
which each parameter is randomly fluctuated by a gaussian function within its uncertainty.105
Deviations of recreated mean and resolution of the energy deposit from the nominal value in106
4
Table 1: Uncertainty from the event selection (%).
Energy center cut in;Beam momentum (GeV/c) x y
2 +0.23 +0.164 +0.28 +0.098 +0.14 +0.0312 +0.13 +0.0215 +0.10 +0.0220 +0.43 +0.0130 +0.14 +0.0132 +0.03 +0.01
20 trials are taken as the systematic uncertainties from the ADC-MIP conversion factor. Mean107
value of each energy is fluctuated in 0.09 - 0.24% and 0.02 - 0.06% of the value due to the108
uncertainties of c(T0; l, s) and scl,s, respectively, as listed in Table 2.109
Table 2: Fluctuation of mean of measured energy deposit created with twenty times pseudo-experiments.
Deviation (%) from;Beam momentum (GeV/c) c(20◦C; l, s) scl,s
2 0.23 0.034 0.09 0.028 0.21 0.0312 0.16 0.0315 0.13 0.0420 0.13 0.0430 0.12 0.0632 0.23 0.04
Figure 4 shows the distributions of stochastic term and the constant term of energy resolution110
varying c(T0; l, s). The systematic uncertainties come from c(T0; l, s) are 0.08% and 0.07% for111
stochastic term and the constant term, respectively. It is 0.01% for both stochastic and the112
constant term from the variation of scl,s.113
3.4 ADC-photon conversion factor114
An ADC-photon conversion factor for each channel is also a linear function of temperature of115
the detector and it is used to convert ADC counts to the number of photons (fired pixels of116
PPD). The number of detected photons are required to apply the PPD saturation correction as117
discussed in 2. Therefore, the propagation of uncertainties of these parameters are also studied118
by using pseudo-experiments as well as the ADC-MIP conversion factors.119
Systematic uncertainties of mean and energy resolution of the measured energy deposit due120
to the uncertainty of these parameters are negligible.121
5
Figure 4: Distribution of the stochastic term (left) and the constant term (right) of the energy resolutionin 20 pseudo-experiments.
3.5 Inter calibration constant122
Systematic uncertainty due to the uncertainty of inter calibration constant is also studied by123
using pseudo-experiment method. Although most of uncertainties of gain inter calibration con-124
stant for each channel are taken as the sigma of gaussian used to make the variation, the125
standard deviation of the measured gain constants is used for the channels which are not suc-126
ceeded to estimate the inter calibration constant due to the same reason for the measurement of127
the ADC-photon conversion factor. Mean value of each energy is fluctuated less than 0.02% and128
the uncertainty of both stochastic and constant term from the uncertainty of inter calibration129
constant is less than 0.01%.130
3.6 The number of effective pixels of PPD131
The mean value of the number of effective pixels of PPD, Npix measured for 72 strips is used132
as an input of the PPD saturation correction as discussed in section 2. Therefore, the standard133
deviation of distribution of Npix is taken as the uncertainty of Npix to create the pseudo-134
experiments to estimate the contribution from the variation of Npix. Mean value of each energy135
is fluctuated in 0.01 - 0.16%. The uncertainties are 0.07% and 0.06% for the stochastic term136
and the constant term, respectively.137
3.7 Difference of mean value of deposit energy among runs138
After temperature correction and saturation correction, there are larger variations of mean values139
of measured energy deposit among runs than the variation expected from the uncertainties of140
respective runs. Therefore, uncertainties from these discrepancies are implemented from the141
standard deviation of the expected value, and listed them for respective beam momenta in142
Table 3.143
3.8 Summary of uncertainties144
Total of systematic uncertainties estimated in this section and the statistic uncertainties for145
each beam momentum are listed in Table 4. Figure 5, left shows the deposit energy with the146
result of linear fitting with those uncertainties, and and deviation of each data from the fitting.147
The intrinsic energy resolution with systematic uncertainties discussed in this section is148
shown in Fig. 5, right as a function of the reverse of the square root of incident beam momentum.149
6
Table 3: Uncertainty estimated from the deviations of the expected values of measured energy depositamong runs.
Beam momentum (GeV/c) Deviation (%)2 0.314 0.238 0.2712 0.8115 0.4520 0.6630 0.1032 0.10
Table 4: Total of systematic and statistical uncertainties of measured deposit energy.
Uncertainty (%)Beam momentum (GeV/c) statistical systematic
2 0.030 0.494 0.022 0.388 0.013 0.3812 0.014 0.8415 0.012 0.4820 0.012 0.8030 0.014 0.2732 0.018 0.29
Figure 5: Response linearity (left) and intrinsic energy resolution (right) of ScECAL.
The curve shows the result of fit to the data after subtracting beam momentum spread contri-150
bution with a quadratical parametrization of the resolution, resulting 12.9±0.1(stat)±0.4(sys)%151
and 1.2±0.1(stat)±+0.4/-1.2(sys)% for the stochastic term and constant term, respectively. The152
systematic uncertainties are estimated considering the case that the beam momentum spread is153
7
varied coherently for all beam momenta.154
4 Discussions155
The temperature correction for the ADC-photon conversion factor is applied in 2. With this156
correction, the deviation from the result of linear fit is less than 2%2. Therefore, we can conclude157
that this temperature correction does not affect on the linearity. However, the 3% of increase158
of the MIP/GeV/c shows that this correction is also necessary.159
Comprehensive study of the systematic uncertainties are also discussed in 3. To explain the160
1.2±0.1(stat)±+0.4/-1.2(sys)% of the constant term of the energy resolution, the Monte Carlo161
simulation study is ongoing.162
References163
[1] CALICE collaboration, CALICE note 16a.164
[2] CALICE collaboration, CALICE note 16.165
[3] C. Johnstone, Proc.,EPAC 2006, Edinburgh, Scotland.166
[4] T. Tsai, Special Reports of All Experimenters meetings FNAL, (2012)167
http://www.fnal.gov/directorate/program planning/all experimenters meetings/special reports/168
Tsai T1018 01 30 12.pdf.169
[5] C. Polly, Special Reports of All Experimenters meetings FNAL, (2010),170
http://www.fnal.gov/directorate/program planning/all experimenters meetings/special reports/171
Polly T1005 08 23 10.pdf.172
[6] R. McNabb, et al, NIM, A 601, 396-402(2009).173
2Some of 8 GeV/c runs have been found to have DAQ problem after [1]. By removing these runs from the analysis,maximum amplitude of the deviation is changed from 1.5% to 2.0%.
8