Page 1: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Template Effective Date: 27-May-2019 | HOOD05162003010284

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC™ Data Management SystemPEP to Go

Page 2: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC –Table of Contents

System Overview 3

Routine Operations 19

Result Management 37

Manage Reports 43

Administration 45

2

Page 3: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

User Log In Screen

1. Start the UniPOC software using the assigned URL address

2. Enter your User ID and Password3. Select the domain from the dropdown list, if applicable4. Select Log In

Note: The UniPOC session expires after 20 minutes of no activity within the application.

3

UniPOC System Overview

Note: The URL address can be obtained for your Point of Care Coordinator.

Page 4: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Main Screen Key Areas

A. Menu BarB. Location Tree (also known

simply as Tree)C. Breadcrumb TrailD. Total Alarm CountE. Details Working AreaF. Action ButtonsG. Current Logged in UserH. Log Off

F

B

A C

G

H

D

E

4

UniPOC System Overview

Note: Menu bar and Location Tree options are dependent upon permission granted by the Administrator or Point of Care Coordinator (POCC).

Page 5: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Main Screen Details

A. Menu Bar• Displays a list of available menu options• When selected, menu items drop down and sub-

menus appearB. Location Tree

• Displays an expandable and collapsible list of the available Facilities, Departments and Locations within the Organization

• Displays device types and serial numbers assigned to specific locations

• Can remain visible or be hiddenC. Breadcrumb Trail

• Displays the current menu selection, followed by the current Tree selection

B

CA

5

UniPOC System Overview

Page 6: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Main Screen Details (continued)

D. Total Alarm Count• Displays the total count of unacknowledged alarms

for the Location Tree scope assigned to the user, if enabled

• Includes all locations accessible to user regardless of the current Tree selection

• LIS: LIS/test alarms; No: notifications; Op: operator alarms

E. Details Working Area• Displays screens based on the selected Tree

location and Menu Bar option• Displays Search Criteria to help locate or filter

information

D

E

6

UniPOC System Overview

Page 7: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Main Screen Details (continued)

F. Action Buttons• Utilize to perform actions for selected items in the

table• Examples include: Exclude, Review, Edit, View, Re-

Route, Acknowledge, Email, etc. G. Current Logged In User

• Displays the name of the current userH. Log Off

• Logs off the system

F

G

H

7

UniPOC System Overview

Page 8: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

The Menu Bar

The Menu Bar displays the following eight options:• Dashboard• Review• Reports• Operators• Instruments• Lots• Administrative• Help

Note: Menu bar options are dependent upon permission granted by the Administrator or Point of Care Coordinator (POCC).

8

UniPOC System Overview

Page 9: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Dashboard

Displays an overview of summary and alarm information including: A. LIS/Test AlarmsB. NotificationsC. OperatorD. Instrument SummaryE. My Views

9

A

E

B

C

D

UniPOC System Overview

Page 10: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Review

• Summary➢ View instrument summary information on the connection status,

the last good “QC” performed and the last upload time and location

• Alarms➢ View, acknowledge and email alarm records

• All Results➢ Exclude, review, edit, view, re-route, acknowledge and email

patient, linearity, proficiency, control test results and electronic QC

• Patient Results➢ View and email patient test results and view test panel

information for multi analyte instruments• Patient Lookup➢ List of patients selected to be sent to the UniPOC database➢ Retrieve current patient data for any Facility in the Organization, if

enabled

10

UniPOC System Overview

Page 11: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Reports

Provides access to all standard and user-defined custom reports grouped into four categories:• Test Data Reports• Operator and User Reports• Instrument Reports• I-STAT Reports

All reports are suitable for archiving on paper or online.

11

UniPOC System Overview

Page 12: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Operators

Manage operators and instrument certifications using two sub-menus:• Manage➢ View and add a new operator and instrument

certification• Certification➢ View and update instrument certifications

12

UniPOC System Overview

Page 13: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Instruments

Provides access to three sub-menus used to manage instrument settings and two sub-menus with log files:• Configuration➢ Configure instrument settings, if applicable

• Comments➢ Define comment codes by instrument type

• Ports➢ Manage port assignments for instruments with a serial

connection to the server• Tracking Log➢ Review events relating to instrument location assignments

• Upload Log➢ Review instrument communication session results

13

UniPOC System Overview

Page 14: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Lots

Provides access to four sub-menus for use in managing reagent lots:• Control➢ Manage Quality Control lot information and control

ranges• Linearity➢ Manage lot information for Linearity Panels and

linearity analyte ranges• Proficiency➢ Manage lot and proficiency analyte ranges for

Proficiency Testing• Reagent➢ Manage reagent lot information and enter QC

ranges per analyte

14

UniPOC System Overview

Page 15: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Administrative

Allows access to managing the Location Tree, User Accounts, the Database and Instrument Test Panels:• Organization Setup➢ Add or edit Organization, Facility, Department and

Location information to the Tree• Facility Settings➢ Manage facility-specific settings such as LIS result

forwarding, notifications, warnings and review comments

• User Setup➢ Create a new user or find and edit existing UniPOC user

accounts and permissions• My Profile➢ Set your preferred startup screen and default

instrument, adjust alarm display types or change your password

• Database➢ Import operator information or check the database

status• Panels➢ Manage multi analyte instrument test panels

15

UniPOC System Overview

Page 16: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Help

• Displays available User Manuals➢ Table of contents provides links to chapters and sections of the

selected User Manual➢ PDF viewer must be installed

16

UniPOC System Overview

Page 17: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

The Location Tree

The Location Tree is made up of one or more Facilities and associated Departments and Locations. Following a Parent-Child relationship, each level of the Tree is linked to the level above:A. Organization: parent to all Facilities within the OrganizationB. Facility: parent to all Departments within the Facility

• Child to the OrganizationC. Department: parent to all Locations within the Department

• Child to the associated FacilityD. Location: child to the associated Department

17

A

B

C

D

UniPOC System Overview

Page 18: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Location Tree Functions:

Symbol : expands locations in the Tree

Symbol : collapses locations in the Tree

Refresh: refreshes the Tree with current information from the databaseHide: hides the Tree and replaces it with a vertical Show Tree button to enable users to display the Tree again

18

UniPOC System Overview

Page 19: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Summary Screen

The Summary screen allows you to view instrument summary information on the connection status, the last good “QC” performed and the last upload time and location. 1. Select Review > Summary2. Select a Facility, Department or Location in the Tree3. Enter desired Search Criteria options4. Select Display Records

19

2

1

3

4

Page 20: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Summary Screen (continued)

Table Columns:• Connect Status

• OK: terminal server or STLB terminal is responding to network requests. IP Address with red background if the terminal server or STLB terminal failed to respond to a ping request on the network

• N/A: instrument communicates via a local serial COM port connection

• LIS Transfer• OK: last data transfer to the LIS was successful• FAIL: last data transfer to the LIS failed• Pending: data transfer to the LIS is still pending• N/A: there is no interface to the LIS

20

Page 21: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Summary Screen (continued)

• Last Good QC Date/Time• The date and time of the last successful (passing) QC done on the

instrument• A red highlighted date of the last known successful QC if the

specified QC interval requirement is exceeded• Last Upload Location

• The last upload location column displays the name of the last upload location for that instrument

• The location name will appear in red if the instrument’s last upload location is different than its home location in the Tree

• Last Upload Date/Time• The date and time of the last upload• A red highlighted date and time if the specified time interval for the

Meter Download Warning in Facility Settings screen is exceeded

21

Page 22: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Email a Summary Record

1. From the Summary screen, select one or more records in the table • Select All to email all summary records

2. Select Email

22

Note: Email is the only functionality of the Summary screen. Instrument summary information is displayed in the Summary table but cannot be edited.

Page 23: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Common Dashboard LIS/Test Alarms

Alarm Type Description

Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient Action/Critical Range or QC Range)

Invalid Patient- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Patient ID associated to test result was not accepted by LIS/HIS. Patient ID should be corrected and result resubmitted

Out-of-Range QC Test Alarm is generated based on the QC Range set at time of test. Alarm is created when the test result isuploaded

Out-of-Range Result Test Alarm is generated based on Action/Critical Range set at time of test. Alarm is created when the test result is uploaded

Test Rejected- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted was not accepted by the LIS/HIS

Note: Refer to Operator’s Manual for further information on additional LIS/Test Alarms and their descriptions. 23

Page 24: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Common Dashboard Notification Alarms

Alarm Type Description

New Reagent Lot New Alarm is generated when a new reagent lot is used. Alarm is created upon initial upload if the reagent lot is not currently listed

Expired Reagent Lot Lot Alarm is generated when a reagent lot used to run a test is past its expiration date. Alarm is createdbased on the expiration date entered

New Control Lot New Alarm is generated when a new control lot is used. Alarm is created upon initial upload if the control lotis not currently listed

Expired Control Lot Lot Alarm is generated when a QC lot used to run a test is past its expiration date. Alarm is created based onthe expiration date entered

System Alert Contact Siemens Support or No

Alert is set to “Contact Siemens Support” when UniPOC software requires service from Siemens. Please contact Siemens Support. Please note that System Alert applies to facilities only. Hence, the alert will be set only when a Facility is selected

Note: Refer to Operator’s Manual for further information on additional Notification Alarms and their descriptions. 24

Page 25: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Dashboard Operator Alarms

Alarm Type Description

Expired Op. Certification Operator Alarm is generated when individual operator certification has expired for a specific instrument type. Alarm is generated on the day the certification status changes from ‘due to expire’ (Op Exp Warning) to ‘expired’ because the certification was not updated

New Operator Operator Alarm is generated when a test is completed using an Operator ID that is not listed (i.e. this can occur if glucose meter is set to Allow Test or Warn for the Operator Certification requirement). Alarm is created upon upload if the Operator ID is not currently on the Operator list

Operator Exp. Warning Operator Alarm is generated when individual Operator certification is due to expire for a specific instrument type. Alarm is generated based on Expiration Date Notification setting for Operator Certification under Administration > Facility Settings. Once the Operator is expired, the status changes from due to expire to expire, and an ‘Expired Op Certification’ alarm is created for this Operator

25

Page 26: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Dashboard Instrument Summary Alarms

Alarm Description

Not Connected Total number of unreachable terminals (IP address added to a location) via a ping command

Upload Required Total number of instruments that have exceeded the Meter Download Warning Interval defined under Facility Settingsand as such require upload (i.e. last instrument upload is older than the number of hours specified for the MeterDownload Warning)

Upload Location Warning Total number of instruments uploaded from a location other than their assigned home location

26

Page 27: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Manage Dashboard Alarms

If there are unacknowledged alarms and user has permission to access linked screens, the Alarm Type will provide a hyperlink displaying further information for the selected alarm(s).

Unacknowledged results are highlighted red and data is filtered based on Tree selection.A. LIS/Test Alarms

• A hyperlinked LIS/Test Alarm row will display the All Results screen

• The All Results screen allows users to exclude, review, edit, view, re-route, acknowledge and email patient, linearity, proficiency, control test results and electronic QC

• Follow your laboratory’s policy for editing and resolving LIS alarms

27

A

Page 28: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Manage Dashboard Alarms (continued)

B. Notification Alarms• A hyperlinked Notification row will display the Alarms screen• No hyperlink is available for the System Alert and Reject Result Alert

row. Contact Siemens Support for assistanceC. Operator Alarms

• A hyperlinked Operator row will display the Operator Certification screen

• New Operator Alarm: i. Multiple rows will be present for each operator, one row for

each instrument type certificationii. Select the checkbox next to the operator for each device that

requires certification. Select Edit and enter Operator demographic information, including Home Location selected in the Tree (required once per operator)

iii. Select Recertify and enter operator certification information into the Edit Certification screen

28

C

Page 29: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Manage Dashboard Alarms (continued)

• Expired Op. Certification Alarm: i. Select the checkbox next to the operator for each device

that requires recertification. Select Recertify and enter the updated certification information

ii. Select Edit to update Operator demographic informationD. Instrument Summary Alarms

• Selecting Not Connected, Upload Required or Upload Location Warning under the Instrument Summary category will display the Summary screen with default Search Criteria

E. My Views• A custom View is a specific search criteria saved as a link for quick

viewing from the Dashboard• A hyperlinked View will display the All Results screen• If user does not have Review All Results Menu Permissions, the

My Views category is not displayed

29

D

E

Page 30: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Enter Reagent Lot Information

1. Select Lots > Reagent2. Locate the desired lot code and select Edit

• If the lot code has not been recognized by the UniPOC software, users must select Add and manually enter the new lot code and reagent type

3. Enter the Lot number• The lot code and reagent type cannot be edited once a lot code as been

recognized by the UniPOC software4. Enter the additional required entry fields: In-Service Date and Expiration

Date5. Select the Download to Instruments checkbox to allow the Reagent Lot to

be downloaded to instruments, if applicable• Only download reagent lots currently in use to ensure instruments

receive the appropriate reagent download list6. Select Save

• Reagent Lot screen appears with the new lot displayed

30

3

4

5

6

Page 31: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Enter Reagent Ranges

1. Select a desired Lot No. checkbox2. Select Edit3. Select Add4. Select an analyte from the Analyte dropdown list, if applicable

• Reagent Type is a read-only field and cannot be edited5. Select Low from the Level dropdown list6. Enter the lower and upper values in the Critical Range fields7. Select the corresponding unit from the Unit dropdown list8. Select Save9. Repeat steps 3-8 to add normal and high level ranges

31

4

Note: For multi-analyte reagent entry, ranges must be entered per lot number per analyte in order to generate statistical reports.

5

6

7

8

Note: Refer to vendor documentation provided with the installed instrument to determine what information is required to successfully enter reagent lot information and reagent ranges.

Page 32: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Enter Control Lot Information

1. Select Lots > Control2. Select a Facility in the Tree3. Select Add

• Add Control Lot screen appears4. Select a Reagent Type from the dropdown list5. Lot No., In-Service Date and Expiration Date are required entry

fields6. Select Save

• Control Lots screen appears with the new lot displayed

32

4

5

6

Page 33: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Enter Control Ranges

1. Select Lots > Control2. Select a Facility, Department or Location in the Tree3. Select Reagent Type from the dropdown list4. Select Display Records5. Select desired Lot No. checkbox6. Select Edit

• Edit Control Lot screen appears with Reagent Type and Lot No. displayed

*Procedure is continued on the next slide.

33

1

2 3

4

5

6

Page 34: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Enter Control Ranges (continued)

7. Select Add• Add Control Range screen appears

8. Select Analyte from the dropdown list9. Select Level from the dropdown list10. Enter the ranges for the selected Level

• Range fields completed are dependent upon device type. Refer to specific device’s user manual for additional information

11. Select Unit from dropdown list12. Select Save13. Repeat steps 7-12 to add additional levels and their associated ranges

34

9

8

10

11

12

Page 35: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Enter Linearity Lot Information

1. Select Lots > Linearity2. Select a Facility in the Tree3. Select Reagent Type from the dropdown list4. Select Display Records5. Select Add

• Add Linearity Lot screen appears6. Enter linearity lot information

• Lot No. and Expiration Date are required entry fields7. Select Save

• Linearity Lots screen appears with the new lot displayed

35

6

7

Page 36: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Routine Operations

Enter Linearity Ranges

1. In the Linearity Lot table, select desired linearity lot by selecting the checkbox in the Analyte column

2. Select Edit• The Edit Analyte Ranges screen appears with

stored information3. Edit the range fields (L1-L10) as applicable4. Select Save

36

Note: Ranges must be entered per analyte to generate a Linearity report.

1

2 3

4

Page 37: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Result Management

All Results Screen

Review > All ResultsAllows users to exclude, review, edit, view, re-route, acknowledge and email patient, linearity, proficiency, control test results and electronic QC.

Utilize Search Criteria and a location highlighted in the Location Tree to display filtered results.

37

Page 38: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Result Management

View Results

1. Select Review > All Results2. Filter for desired information in the Search Criteria and

select a location in the Location Tree3. Select Display Records4. Select one or more checkbox(es) in the Location

column• To view all records select the All checkbox in the

Location column header5. Select View6. The View Result Record screen appears with the data

of the first selected record• Result records cannot be modified from this screen

7. Select the previous and next page navigation buttons to view additional data records

38

3

4

5

7

Page 39: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Result Management

Edit Results

1. Select Review > All Results2. Filter for desired information in the Search Criteria and

select a location in the Location Tree3. Select Display Records4. Select one or more checkbox(es) in the Location column to

edit• To edit all records select the All checkbox in the Location

column header5. Select Edit6. The Edit Result Record screen appears with the data of the

first selected record7. Edit the result information accordingly 8. Select Save to save the edited record or Cancel to cancel

the edit action• The next selected record is then displayed

39

Page 40: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Result Management

Evaluate QC Results

1. Select Review > All Results2. Select Control from Result Type dropdown list

• To evaluate QC results by specific alarm types or statuses filter Search Criteria accordingly

3. Select Display Records• To view additional QC result information

select the View or Edit action buttons4. QC results appear in the result table below

• White letters on a red background indicate data related to an alarm

• Black letters on a yellow background indicate previously acknowledged alarms

40

2

4

3

Page 41: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Result Management

Evaluate Patient Results

1. Select Review > All Results2. Select Patient from Result Type dropdown list3. Select Display Records4. Patient results appear in the result table below

• Utilize the Edit and View action buttons to evaluate and manage patient results

• The Review action button can be utilized to add User ID and timestamp with or without a comment to results for regulatory purposes

41

2

34

Page 42: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Result Management

Patient Results Screen

The Patient Results screen allows users to view and email patient test results and view test panel information for multi analyte instruments.1. Select Review > Patient Results2. Filter for desired information in the Search

Criteria and select a location in the Location Tree

3. Select Display Records

42

Page 43: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Manage Reports

View Available Reports

1. On the Reports screen, select the report category • Test Data Reports• Operator and User Reports• Instrument Reports• I-STAT Reports

2. Select the report • Select + to display any sub-reports• Customized reports display as sub-reports under

the parent report selected3. Select a Facility, Department or Location in the Tree to

customize the report for a specific location4. On the Create Report screen, select any desired filters

under Report Options5. Select View to preview the report6. Utilize the Report toolbar to view, navigate, print and

export reports

Note: Refer to the UniPOC User Manual for additional information on individual report types.

43

1

2

3

4

5

Page 44: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Manage Reports

Create a Custom Report

Create Report screen allows users to customize and save reports that include selected:• Report Location• Report Options

1. On the Reports screen, select the report category 2. Select the report to be customized3. Select a Facility, Department or Location in the Tree to customized

the report for a specific location AND/OR select available Report Options

• Report Options include filter and date range options4. Select Save5. Enter a Report Name 6. Select Save

• The report is listed as a sub-report under the default report name

44

3

4

5

6

Note: Refer to the UniPOC User Manual for additional information on individual report types.

Page 45: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Add and Certify a New Operator

1. Select Operators > Manage2. Select Add

• The Add Operator screen appears3. Under Operator Information, enter the First Name,

Last Name and Operator ID4. Highlight a location in the Location Tree to assign the

operator to a Home Department and Home Location

45

3

1

4

*Procedure is continued on the next slide.

Page 46: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Add and Certify a New Operator (continued)

5. To add operator instrument certification, select Add under Certifications

6. Enter the appropriate operator certification information

7. Select Save Certification8. Select Save9. Repeat steps 5-8 to add additional certifications

Note: To add and certify a new operator, separate permission must be assigned under Administrative > User Setup. To do this, select Edit Operator Certification and Edit Operator Information under UniPOC Function Permissions.

46

5

7

8

6

Page 47: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Operator Certification Screen

The Operator Certification screen allows users to search operators by Last Name, ID or recertification information and to recertify operators.1. Select Operators > Certification

• The Operator Certification screen appears2. Use Search Criteria and Location Tree to display filtered

operators3. Select Display Records4. Use Actions Buttons in the table to apply selected actions to

selected operators• Recertify: certify or recertify operator(s) from the Edit

Certification screen• Expire: expire certification(s) from the Operator Certification

screen• Delete Operator: delete an operator to hide all records for

the selected operator, including the operator’s identity and all certification information for the operator

47

Page 48: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Edit Operator Records

1. From the Operator Certification screen, select one or more records from the table

• Select All to edit all records2. Select Edit

• Edit Operator screen appears3. Edit Operator Information

• Operator ID is not an editable field• To assign an operator to a different location

select a location in the Location Tree4. Select Save

48

Note: An operator cannot be moved to a different Facility. Instead, create a New Operator record to add the operator to a different Facility. The same ID number can be utilized, provided the ID number is not currently assigned to another operator. Certifications will be kept separate per facility.

Page 49: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Delete an Operator

The delete action button will hide all records for the selected operator(s), including the operator’s identity and all certification information for the operator(s).1. From the Operator Certification screen, select one or more records

from the table2. Select Delete3. The following dialogue box appears:

4. Select OK to delete the record(s)

49

Page 50: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Recertify an Operator

1. From the Operator Certification screen, select one or more operators from the table

• Select All to select all operators2. Select the desired instrument type certification from

the Instrument dropdown list 3. Select Recertify4. Complete the appropriate operator certification fields5. Select Save to certify the selected operator(s) for the

selected instrument

50

1

3

2

4

5

Page 51: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Instruments and the Location Tree

New Instruments:The first time a new instrument uploads data to the system the instrument serial number appears in the Tree under an instrument type icon in the Location from which the instrument was uploaded.

Instruments are grouped by Location.

Instruments visible on the Tree can be moved, renamed, deactivated and deleted.• User must be assigned Instrument Operations Function Permission

51

Page 52: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Add Instruments to the Location Tree

1. Right click a Location name in the Tree and select Add Instrument2. Select the Instrument Type from the dropdown list3. Enter the serial number for the instrument4. Select Add Instrument

52

12

34

Page 53: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Move Instruments within the Location Tree

Instruments can be moved to a different Location within the same Facility or a Location within a different Facility via drag and drop.1. Place the mouse cursor over the serial number or name2. Click and drag the instrument name to a new location

• The cursor appears as symbol3. When the cursor changes from to the instrument can be

moved to that location4. Release the mouse to display the instrument in its new location

53

Page 54: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Rename Instruments within the Location Tree

1. Select the desired instrument from the Tree2. Right click and select Rename Instrument

• Serial number or name text changes to an editable text box3. Type the new instrument name and click the mouse anywhere outside

the text box to display the new instrument name in the Tree4. To cancel the rename action, press Esc on the keyboard

54

Page 55: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Deactivate Instruments within the Location Tree

If a device is no longer required (ex. device out for repairs) it can be deactivated. • All data is retained in UniPOC database• Information is synchronized and can be restored after reactivating the

device1. Select the desired instrument from the Tree2. Right click and select Deactivate Instrument3. Select OK to deactivate the instrument

• The serial number or name text will now appear as italicized and greyed out in the Tree

Note: Re-docking the instrument will reactive the instrument. All test results will be accessible to users and operators.

55

Page 56: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Delete Instruments from the Location Tree

1. Select the instrument to delete from the Tree2. Right click and select Delete Instrument3. Select OK to delete the instrument

• Instrument is removed from the Tree

Note: If the instrument flagged as deleted uploads data at a later time, the instrument will reappear in the Tree in the location from which it was uploaded. The Instrument name and serial number are both saved in the database.

56

Page 57: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Create and Manage the Location Tree

1. Organization SetupA. Select Administrative > Organization Setup > Organization

• The default for the Organization Name appears in the Tree as Organization

• Only one Organization can be supported by UniPOC softwareB. Enter the required Organization Name and any additional detailsC. Select Save

• Information is stored in the database and the new Organization name will appear in the Tree

D. To edit information for an Organization• Right click the Organization name in the Tree and select Edit

Organization• Add or edit fields in the Organization Details table• Select Save

*Procedure is continued on the next slide.

57

A

B

C

Page 58: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Create and Manage the Location Tree (continued)

2. Facility SetupA. Right click the Organization name in the Tree and select Add

FacilityB. Enter the required Facility Name and any additional detailsC. Select Save to save or select a menu bar option to exit the

facility setup• Information is stored in the database and the Facility name

will appear in the TreeD. To edit a Facility in the Organization

• Right click the Organization name in the Tree and select Edit Facility

• Add or edit fields in the Facility Details table• Select Save

Note: If a Facility is interfaced please contact Siemens support before editing name of the Facility. Changing the name may impact the interface. 58

C

B

A

*Procedure is continued on the next slide.

Page 59: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Create and Manage the Location Tree (continued)

3. Department SetupA. Right click the Facility name in the Tree and select Add

DepartmentB. Enter the required Department Name and any additional

detailsC. Select Save

• Information is stored in the database and the Department name will appear in the Tree

D. To edit a Department in a Facility• Right click the Department name in the Tree and

select Edit Department• Add or edit fields in the Department Details table• Select Save

*Procedure is continued on the next slide.

59

C

B

A

Page 60: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Create and Manage the Location Tree (continued)

4. Location SetupA. Right click the Department name in the Tree and select Add

LocationB. Enter the required Location Name and any additional detailsC. Select Save

• Information is stored in the database and the Location name will appear in the Tree

D. To edit a Location • Right click the Location name in the Tree and select

Edit Location• Add or edit fields in the Location Details table• Select Save

Note: A Location can only be deleted if there is no data associated with the specified Location in the active database.

60

C

B

A

Page 61: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Add a Terminal Server to a Location

1. Select Administrative > Organization Setup > Location2. Select a Location name in the Tree3. Select Add4. Enter the Hostname, IP Address or MAC Address for the specified

Location• DHCP environment: only enter Hostname. IP Address should

update automatically• IP Addresses statically assigned: IP Address and Hostname should

be added • Comment is an optional field

5. Select Save

61

1

3

2

4

5

Page 62: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Enter Instrument Configurations

Instrument configuration, like the Location Tree, follows the Parent-Child Relationship. Settings saved at the top level of the Tree affect all sub-levels. • Select Instruments > Configuration• It is recommended that instrument configuration settings be

made at the Facility level when possible

Note: For complete information on configuration options, refer to documentation provided with the installed instrument.

62

Note: The above image displays an example referring to the CLINITEK Status® + Analyzer. Refer to documentation provided with the instrument for further information.

Page 63: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Enter Instrument Configurations (continued)

Breaking the Parent-Child Relationship:• Saving instrument configuration for a lower level of the Tree

(i.e. Department or Location) will sever the relationship between the child and its parent level on the Tree

• Breaking the Parent-Child relationship is necessary to enable configuration settings different from the global Organization or Facility settings

• Once a location is disassociated from its parent location, any changes to instrument configuration will need to be made independent of the parent location

63

Note: The above image displays an example referring to the I-STAT 1 Analyzer. Refer to documentation provided with the instrument for further information.

Page 64: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Define Instrument Comment Codes

Each Facility in the Organization maintains a separate list of user-defined comment codes. 1. Select Instruments > Comments2. Select a Facility name in the Tree3. Select an Instrument from the dropdown list4. Select Add5. Enter a Code number 0 to 99,999 and a Description

• Refer to manufacturer for number of digits required or used by device

• For reporting purposes only, instrument code list does not download to specified instrument

6. Select Save

64

1

2 3

4

5

6

Note: If an undefined comment code is used on a device, the code will be automatically entered into the UniPOCdatabase. The description can then be entered by a UniPOC user with appropriate permissions.

Page 65: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Create UniPOC User Accounts

The New User screen allows the Administrator or POCC to add users to UniPOC software. A user is also granted permission to access locations, menu options and specific functions. 1. Select Administrative > User Setup2. Select New3. In the Identification section, enter First Name, Last Name,

Employee ID, User ID and any additional information4. Select the Domain from the dropdown list, if applicable

*Procedure is continued on the next slide.

65

1

2

3

4

Page 66: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Create UniPOC User Accounts (continued)

5. In the Permit Access to Locations section, select All, None or Selected to determine locations displayed in the Tree

• For Selected, select the + button to display associated Departments and Locations for the selected Facility

6. In the UniPOC Menu Permissions section, grant access to menu and sub-menu options

7. In the UniPOC Function Permissions section, grant access to specific functions in the system which correspond to Action Buttons

8. Select Save

66

5

6 7

8

Page 67: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Manage UniPOC User Accounts

The User Search screen allows the user to find, edit and delete users.1. Select Administrative > User Setup2. Enter First Name, Last Name, User ID and/or

Employee ID3. Select Display Records4. Edit or delete users displayed in the User table

Note: A UniPOC user should be POC personnel utilizing the UniPOC software. Personnel operating POC instrumentation from a specific location are classified as Operators.

67

Page 68: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Manage My Profile

The My Profile screen allows users to customize the behavior of the system for their User ID. 1. Select Administrative > My Profile2. Select a Startup Screen and Default Instrument3. Customize your alarm preferences by enabling or disabling

Total Alarm Counts and displayed alarm types4. Change your password

• By default the password is the user ID until changed by the user

5. Select Save

68

1

2

3

4

5

Page 69: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Manage Facility Settings

If more than one Facility exists within the Organization, each Facility must be configured separately.1. Select Administrative > Facility Settings2. Select a Facility name in the Tree3. Enter desired settings for the specified Facility

• Result Forwarding: forward test results to LIS/HIS or hold test results in queue, if applicable

• Expiration Date Notifications: number of days for advance warning notification• Performance Target Levels: establishes QC performance target levels used in

reports• Meter Download Warning: required number of hours allowed between

instrument uploads• Facility-Defined Review Comment: specify up to 10 Review comments to be

displayed in the Predefined Review Comment list of the Edit Result screen4. Select Save

69

1

2

3

4

Page 70: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Create Test Panels

Creating a test panel allows the user to view multiple analytes in a summary format via the Review Patient Results screen and are accessible to the entire Organization.1. Select Administrative > Panels2. Select Add3. Enter the Panel Name and select the corresponding analytes4. Select Save5. Panel information is stored and available in the Panel

dropdown list of the Patient Results screen

70

Note: Create test panels configuration is optional. This feature benefits UniPOC users with multi analyte devices.

1

2

3

4

Page 71: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

UniPOC Administration

Create a My View

My Views are customized views saved as links for quick viewing from the Dashboard. Creating a view saves the current table column configuration and search criteria. However, the date range filter and current Location Tree selection are not saved.1. Select Review > All Results2. Select desired Search Criteria options3. In the All Results table, select Customize to choose which

columns are displayed in the table4. Select Save to My Views 5. Enter a View Name6. Select Save

• The customized view now appears in the My Views section of the Dashboard

71

1

2

3

4

5

6

Page 72: UniPOC™ Data Management System · Out-of-Range Result- LIS LIS Alarm is generated based on information returned from the LIS/HIS. Test result submitted is out of range (patient

Unrestricted © Siemens Healthineers, 2019Effective Date: 12-Nov-2019 | HOOD05162003049935

Please note that the learning material is for training purposes only!

For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training.

The Operator's Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions.

Note: Some functions shown in this material are optional and might not be part of your system.

Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information.

The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages.

All names and data of patients, parameters and configuration dependent designations are fictional and examples only.

All rights, including rights created by patent grant or registration of a utility model or design, are reserved.

UniPOC, CLINITEK Status and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc., or its affiliates. All other trademarks are the property of their respective owners.

Copyright © Siemens Healthcare GmbH 2019

Siemens Healthineers HeadquartersSiemens Healthcare GmbH Henkestr. 12791052 Erlangen, GermanyTelephone: +49 9131 84-0siemens.com/healthineers

72