Ultrasonic Scaler QuickStart Guide: Scaler · Ultrasonic Scaler QuickStart Guide: 1. Press S button...

2
OFF ON ZERO MAX Ultrasonic Scaler QuickStart Guide: 1. Press S button to turn ON the Scaler. 2. Tip Working mode: 3. Select scaler function: 4. Water regulation: 5. Foot Pedal Control: 6. Cleaning: Scaler Comprehensive cleaning function: The cleaning function can be started by pressing C. Water or disinfectant cleans the water line and handpiece. The cleaning cycle stops automatically after 80 seconds. With every push of the S button, the function toggles from: Scaling with water Dry mode: scaling without water Irrigation mode: just water, no scaling The unit is in standby mode when the power cord is connected and the green LED is illuminated. or or 100 % 70 % 40 % Depress the foot control pedal in the leftmost position and adjust the water flow with the control ring on the scaler handpiece until the water is dripping. The scaler is activated when the foot control pedal is pressed down. The power is regulated from left/ZERO to right/MAX within the chosen working mode. In ZERO position there are two functions: a) irrigation with no scaling power: hold the footpedal down. b) diagnostic light: press the footpedal once. 1 1 2 2 3 4 3 4 Choose the correct working mode that is marked on the base of each tip by pressing the corresponding working mode button on the keypad: 7. Push the S button for 3 seconds to turn the scaler OFF.

Transcript of Ultrasonic Scaler QuickStart Guide: Scaler · Ultrasonic Scaler QuickStart Guide: 1. Press S button...

Page 1: Ultrasonic Scaler QuickStart Guide: Scaler · Ultrasonic Scaler QuickStart Guide: 1. Press S button to turn ON the Scaler. 2. Tip Working mode: 3. Select scaler function: 4. Water

OFF

ON

ZERO MAX

Ultrasonic Scaler QuickStart Guide:

1. Press S button to turn ON the Scaler.

2. Tip Working mode:

3. Select scaler function:

4. Water regulation:

5. Foot Pedal Control:

6. Cleaning:

Scaler

Comprehensive cleaning function:

The cleaning function can be started by pressing C. Water or disinfectant cleans the water line and handpiece. The cleaning cycle stops automatically after 80 seconds.

With every push of the S button, the function toggles from: Scaling with water Dry mode: scaling without water Irrigation mode: just water, no scaling

The unit is in standby mode when the power cord is connected and the green LED is illuminated.

or or

100 %70 %40 %

Depress the foot control pedal in the leftmost position and adjust the water flow with the control ring on the scaler handpiece until the water is dripping.

The scaler is activated when the foot control pedal is pressed down. The power is regulated from left/ZERO to right/MAX within the chosen working mode. In ZERO position there are two functions: a) irrigation with no scaling power: hold the footpedal

down.b) diagnostic light: press the footpedal once.

1

1

2

2

3

43

4

Choose the correct working mode that is marked on the base of each tip by pressing the corresponding working mode button on the keypad:

7. Push the S button for 3 seconds to turn the scaler OFF.

Page 2: Ultrasonic Scaler QuickStart Guide: Scaler · Ultrasonic Scaler QuickStart Guide: 1. Press S button to turn ON the Scaler. 2. Tip Working mode: 3. Select scaler function: 4. Water

1.

4.

2.

5.

3.

1. Consult a qualified technician to install to water supply if applicable.2. Connect the foot control pedal cable to the foot control pedal and to the rear of the unit.3. Connect the power cord to the unit and the AC power outlet.

6. See Section 4 in the owner’s guide for Operation Instructions.

4. Fill the medicament bottle with water or medicament solution. Screw the bottle connector cap onto the bottle and push it onto the connector on the left side of the unit. When removing the bottle, press first the depressurization button beside the connector.5. Gently slide a handpiece grip onto the scaler handpiece. Place the scaler tip in the torque wrench and use the torque wrench to screw the scaler tip clockwise onto the scaler handpiece. Tighten until the resistance causes the torque wrench to begin to slip.

NOTE! Before installing and using the device please read the owner’s guide.

6.b.

Installation

P/N

20

00

449

9A

coltene.com

Coltène/Whaledent Inc.235 Ascot ParkwayCuyahoga Falls, OH 44223 / USATel. USA & Canada + 1 800 221 3046 + 1 330 916 8800Fax +1 330 645 [email protected]

Coltène/Whaledent AGFeldwiesenstrasse 209450 Altstätten / SwitzerlandTel. +41 (0)71 757 53 00Fax +41 (0)71 757 53 [email protected]

Coltène/Whaledent GmbH + Co. KGRaiffeissenstrasse 3089129 Langenau / GermanyTel. +49 (0)7345 805 0Fax +49 (0)7345 805 [email protected]