TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS ® G70x Ultrasonic Gas Flowmeters for Permanent Installation Designed for wall mounting or for installation in 19'' rack systems Features Non-invasive measurement using the clamp-on method for precise bi-directional, highly dynamic flow measure- ment Transducers approved for the use in hazard areas available: Automatic loading of calibration data and transducer detection, reduces set-up times and provides precise, long-term stable results Proven clamp-on method; transducers available for a wide range of inner pipe diameters and temperatures in the range of ; resistant to dust and humidity Measurement is unaffected by gas density, viscosity and composition, dust, humidity, temperature or pres- sure User-friendly design Applications Designed for industrial use in harsh environments, in gas processing and natural gas extraction, chemical in- dustry and in the petroleum industry. Practical applica- tions: - Measurement on natural gas pipelines and in natural gas storage installations - Measurement of synthesized gas and injection gas - Measurement for the gas supply industry ATEX, FM (7…1600 mm) -40...+200 °C FLUXUS FLUXUS G704 G709

Transcript of Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 ·...

Page 1: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1

Technical Specification

FLUXUS® G70x

Ultrasonic Gas Flowmeters for Permanent Installation

Designed for wall mounting or for installation in 19'' rack systems

Features

• Non-invasive measurement using the clamp-on method for precise bi-directional, highly dynamic flow measure-ment

• Transducers approved for the use in hazard areas available:

• Automatic loading of calibration data and transducer detection, reduces set-up times and providesprecise, long-term stable results

• Proven clamp-on method; transducers available for a wide range of inner pipe diameters and temperatures in the range of ; resistant to dust and humidity

• Measurement is unaffected by gas density, viscosity and composition, dust, humidity, temperature or pres-sure

• User-friendly design

Applications

• Designed for industrial use in harsh environments, in gas processing and natural gas extraction, chemical in-dustry and in the petroleum industry. Practical applica-tions:

- Measurement on natural gas pipelines and in natural gas storage installations

- Measurement of synthesized gas and injection gas

- Measurement for the gas supply industry

ATEX, FM

(7…1600 mm)-40...+200 °C

FLUXUS

FLUXUS

G704

G709

Page 2: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

FLUXUS® G70x Technical Specification

List of Contents

2 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

Function .......................................................................................................................................................... 3

Measuring Principle........................................................................................................................................... 3

Calculation of the Volume Flow......................................................................................................................... 3

Number of Sound Paths.................................................................................................................................... 4

Typical Measurement Setup ............................................................................................................................ 5

Standard Volume Flow ..................................................................................................................................... 5

Flowmeter ....................................................................................................................................................... 6

Technical Data ................................................................................................................................................. 6

Dimensions ...................................................................................................................................................... 8

2 " Pipe Mounting Kit (option) ........................................................................................................................... 9

Terminal Assignment ..................................................................................................................................... 10

Transducers .................................................................................................................................................. 12

Transducer Selection ..................................................................................................................................... 12

Order Code Key for Transducers ................................................................................................................... 15

Technical Data ............................................................................................................................................... 16

Transducer Mounting Fixtures ................................................................................................................... 33

Coupling Materials for Transducers ........................................................................................................... 34

Damping Mats (Option)................................................................................................................................. 35

Selection of Damping Mats ............................................................................................................................ 35

Length of Pipe Damping Mat - Type B ............................................................................................................ 35

Connection Systems .................................................................................................................................... 36

Transducer Cables ......................................................................................................................................... 37

Junction Box ................................................................................................................................................ 38

Technical Data ............................................................................................................................................... 38

Dimensions .................................................................................................................................................... 38

2 " Pipe Mounting Kit (option) ......................................................................................................................... 39

Terminal Assignment ..................................................................................................................................... 39

Temperature Probes (Option) ..................................................................................................................... 41

Page 3: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 3

Technical Specification FLUXUS® G70x

Function

Measuring Principle

For the flow measurement of a medium in a pipe, ultrasonic signals are used, employing the transit time dif-ference principle. Ultrasonic signals are emitted by a transducer installed on one side of a pipe, reflected onthe opposite side and received by a second transducer. These signals are emitted alternatively in flow direc-tion and against it.

As the medium in which the signals propagate is flowing, the transit time of the ultrasonic signals in flow di-rection is shorter than against the flow direction.

The transit time difference t is measured and allows to determine the average flow velocity on the propaga-tion path of the ultrasonic signals. A flow profile correction is then performed in order to obtain the area aver-age of the flow velocity, which is proportional to the volume flow.

The received ultrasonic signals will be checked for their usefulness for the measurement and the plausibilityof the measured values will be evaluated. The complete measuring cycle is controlled by the integrated mi-croprocessors. Disturbance signals will be eliminated by statistical signal processing.

Path of the ultrasonic signal Transit time difference t

Calculation of the Volume Flow

Q = kRe . A . k . t/(2 . tt)

with:

Q - volume flowkRe- fluid mechanics correction factorA - cross-sectional area of the pipek - flowmeter constantt - transit time differencett - transit time the medium

��

��

��

Page 4: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

4 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

.

Number of Sound Paths

The number of sound paths is the number of transits of the ultrasonic signals through the medium in the pipe.

reflection mode: number of sound paths = even, the transducers are mounted on the same side of the pipe,correct positioning of the transducers easier

diagonal mode: number of sound paths = odd, the transducers are mounted on opposite sides of the pipe

The mode to be used depends on the application. If the number of sound paths is increased, the accuracy ofthe measurement will be better, but the signal attenuation is increased.

In case of a high signal attenuation by medium, pipe and coatings, diagonal mode with 1 sound path will beused.

The optimum number of sound paths for the parameters of the application will be detemined automatically bythe flowmeter

As the transducers can be mounted with the transducer mounting fixture (option) in reflection mode or diago-nal mode the number of sound paths can be adjusted optimally to the application.

Diagonal mode, number of sound paths: 1 Diagonal mode, number of sound paths: 1,negative transducer distance

a - transducer distance

Reflex mode, number of sound paths: 2

a > 0 a < 0

a

Page 5: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 5

Technical Specification FLUXUS® G70x

Typical Measurement Setup

Example for a measurement setup in reflection mode with connection of the inputsto an external process pressure and process temperature measurement for standard volume flow calculation

Standard Volume Flow

The standard volume flow can be selected as physical quantity to be measured. It will be calculated internallyby:

VN = V . p/pN . TN/T . 1/K

with:

VN - standard volume flow V - operational volume flow pN - standard pressure (absolute value) p - operational pressure (absolute value) TN - standard temperature in K T - operational temperature in K K - gas compressibility factor

The operational pressure p and the operational temperature T of the medium will be entered directly as fixedvalues into the flowmeter.

Or:

If inputs are installed (option), pressure and temperature can be measured by the customer and fed in theflowmeter.

The gas compressibility factor K will be entered in the flowmeter:

• as fixed value or

• as approximation according to e.g. AGA8 or GERG

� � � � � � � � � � � � � � � �

P T

transducers

damping mat

flowmeter

RS232

analog/binary outputs analog inputs

power supply

temperature probe e.g. external pressure sensor

RS485(option)

Page 6: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

6 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Flowmeter

Technical Data

FLUXUS G704 G704 A2 G709design standard field device field device for ATEX zone 2 19 " module

measurement measuring principle transit time difference correlation principle flow velocity , pipe diameter dependent repeatability 0.15 % of reading

accuracy- volume flow ± 1...3 % of reading depending on application

± 0.5 % of reading with field calibrationmedium gases with a ratio of the characteristic acoustic impedances of pipe wall and gas < 3000,

e.g. nitrogen, air, oxygen, hydrogen, argon, helium, ethylene, propaneflowmeter power supply 100...240 V/50...60 Hz or

20...32 V DCpower consumption < 15 Wnumber of flow measuring channels 1, option: 2signal damping 0...100 s, adjustablemeasuring cycle (1 channel) 100...1000 Hzresponse time 1 s (1 channel), option: 70 mshousing material aluminum, powder coated aluminum degree of protectionaccording to EN 60529

IP 65 IP 20

dimensions see dimensional drawing 42TE x 3HE(without back panel)

see dimensional drawing weight fixation wall mounting, option: 2 " pipe mounting 19 " rack mounting operating temperature -20...+60 °C display 2 x 16 characters, dot matrix, backlitmenu language English, German, French, Dutch, Spanishexplosion protection

ATEX

zone - 2 -marking - II3G

Ex nA II T4Ta -20...+60 °C

II3D Ex tD A22 IP65 T100 °C

-

measuring functions physical quantities operational volume flow, standard volume flow, mass flow, flow velocitytotalizers volume, mass calculation functions average, difference, sum data logger loggable values all physical quantities and totalized values capacity > 100 000 measured valuescommunication interface -process integration: option: RS485 (Modbus, sender) or HART

-diagnosis: RS232

0.01...35 m/s±0.01 m/s

±0.01 m/s

±0.01 m/s

2.8 kg 1.7 kg

Page 7: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 7

Technical Specification FLUXUS® G70x

serial data kit (option)

software (all WindowsTM versions) -FluxData: download of measured data, graphical presentation,conversion to other formats (e.g. for ExcelTM)

-FluxKoeff: creating medium data setscable RS232

adapter RS232 - USBoutputs (option)

The outputs are galvanically isolated from the flowmeter. number request

current output current output

- range 0/4...20 mA - accuracy 0.1 % of reading ±15 A- active output Rext < 500 - passive output Uext = 4...24 V, dependent on Rext, Rext < 1 k

current output I1 in HART mode - range 4...20 mA- passive output Uext = 10...24 V

voltage output range 0...1 V or 0...10 Vaccuracy 0...1 V: 0.1 % of reading ±1 mV

0...10 V: 0.1 % of reading ±10 mVinternal resistance Ri = 500

frequency output range 0...1 kHz or 0...10 kHzopen collector 24 V/4 mA

binary output Reed relay - 48 V/0.25 Aopen collector - 24 V/4 mAoptorelay 26 V/100 mA -binary output as alarm output

- functions limit, change of flow direction or error limit, change of flow direction or error

binary output as pulse output - pulse value 0.01...1000 units 0.01...1000 units- pulse width 1...1000 ms 80...1000 ms

inputs (option) The inputs are galvanically isolated from the flowmeter.

number max. 4, on requesttemperature input

designation Pt100/Pt1000 Pt100connection 4-wire 4-wire range -150...+560 °C -50...+400 °C resolution 0.01 K 0.1 Kaccuracy ±0.01 % of reading ±0.03 K ±0.1 % of reading ±0.2 K

current input range active: 0...20 mA

passive: -20...+20 mAaccuracy 0.1 % of reading ±10 Aactive input Ui = 24 V, Ri = 50 , Pi < 0.5 W, not short circuit proofpassive input Ri = 50 , Pi < 0.3 W

voltage input range 0...1 V 0...1 V or 0...10 Vaccuracy 0.1 % of reading ±1 mV 0...1 V: 0.1 % of reading

±1 mV0...10 V: 0.1 % of reading ±10 mV

internal resistance Ri = 1 M Ri = 1 M

FLUXUS G704 G704 A2 G709

Page 8: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

8 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Dimensions

FLUXUS G704

FLUXUS G709

in mm

70

10Ø 4.5

163

200

287

M20 (6x)

265

screw holes for wall mounting

198

208

129

222

15

123

110

M2.

544

213

129

170

Page 9: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 9

Technical Specification FLUXUS® G70x

2 " Pipe Mounting Kit (option)

FLUXUS G704

for vertical and horizontal pipes

Page 10: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

10 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Terminal Assignment

FLUXUS G704

2 The number, type and terminal assignment of the outputs and inputs will be customized.

Power Supply

terminal strip KL3terminal connection AC connection DC

PE earth earth N(-) neutral - DCL(+) phase + DC

Transducers

terminal strip KL1extension cable for connection system TS

transducer cable for connection system TS (ATEX zone 1 )

transducer cable for connection system TS, AS(ATEX zone 2, FM or without explosion protection)

measuring channel A measuring channel B measuring channel A measuring channel Bterminal connection terminal connection terminal connection

AV signal BV signal X_AV X_BV SMB connector AVS shield BVS shield X_AR X_BR SMB connector ARS shield BRS shield X1 X2 AMP-Quick connector1

AR signal BR signal 1 connection system AS

Outputs2 RS485 (option)

terminal strip KL4 terminal strip KL4terminal connection terminal connection

P1+...P4+, P1-...P4- current output, voltage output, frequency output or binary output (optorelay)

A+ signal +B- signal -

P5a...P7a, P5b...P7b binary output (optorelay) 101 shield

Inputs2

terminal strip KL2temperature probe passive current source active current source

terminal connection connection withextension cable

connection connection

T1a...T4a red red not connected not connected T1A...T4A red/blue gray - +T1b...T4b white/blue blue + not connected T1B...T4B white white not connected -S1...S4 shield shield not connected not connected

� � � � ! " � !� � � � � � � � � � � � � � � � � � � � # � � # � #

� � � � � � � � � � � � � � � � � � $ � � $ � $

� � #� � $ � � # � � � � $ � � # � � $ � � # � � � � $

� � � � � � � � � � � � � � � � � � � � � � � � � � � �

� " �

� " �

� " �

� � �� � � � � � � � � � � � � � � � � � � � �

� % � % � � & � � � � � � � % � % � � & � � � � � �

� " �

' � %

' � �

' � %

' � �

� �

Page 11: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 11

Technical Specification FLUXUS® G70x

FLUXUS G709

2 The number, type and terminal assignment of the outputs and inputs will be customized.

Power Supply

terminal strip KL2terminal connection AC terminal connection DC

PE earth PE earth N neutral L- DC-L1 phase L+ DC+

Transducers

terminal strip KL6, KL8extension cable for connection system TS

transducer cable for connection system TS (ATEX zone 1 )

transducer cable for connection system TS, AS(ATEX zone 2, FM or without explosion protection)

measuring channel A measuring channel B measuring channel A measuring channel Bterminal connection terminal connection terminal connection

AV signal BV signal X6AV X8BV SMB connector AVS shield BVS shield X6AR X8BR SMB connector ARS shield BRS shield X5 X7 AMP-Quick connector1

AR signal BR signal 1 connection system AS

Outputs2 RS485 (option)

terminal strip KL1 terminal strip KL4terminal connection terminal connection

P1+...P7+, P1-...P7- current output, voltage output, frequency output or binary output (open collector)

4A+ signal +4B- signal -

P5a...P7a, P5b...P7b binary output (Reed relay) герконовое реле 43 shield

Inputs2

terminal strip KL3temperature probe passive current source active current source

terminal connection connection withextension cable

connection connection

T1a...T4a red red not connected not connected T1A...T4A red/blue gray - +T1b...T4b white/blue blue + not connected T1B...T4B white white not connected -S1...S4 shield shield not connected not connected

� " � " �� " �� " �

� " �� " �

� � % � %

� � � � �

��

� " � � "

� ( � � � � "

� � �

� � �

� � �

� � �

� � �

� � �

� � �

� � �

� % �

� %

� � �

� & �

� �

� % �

� %

� � �

� & �

� �

� � �

� � �

� �

� �

� �

" �

" �

" �

� �

� � $

� $

� �

� � $

� � �

� � �

� � �

� � �

� � �

� � �

� � #

� #

� �

� � #

� � �

� � �

� � �

� � �

� � �

� � �

� � �

� � �

� � �

� �

� � �

� � �

� � �

� � �

� �

� � �

� � $

� � #

� � #

� �

� � $

� � #

� � #

� � $

� �

� � $

Page 12: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

12 29.9.2009 TSFLUXUS_G7V1-3EN_LEU

FLUXUS® G70x Technical Specification

Transducers

Transducer Selection

Step 1a:

select a Lamb wave transducer:

Step 1b:

If the pipe wall thickness is not in the range of the Lamb wave transducers, select a shear wave transducer:

Step 2:

inner pipe diameter d dependent on the flow velocity v of the medium in the pipe

The transducers are selected from the characteristics (see next page). Lamb wave transducers are selectedfrom the left column, shear wave transducers from the right column.

Lamb wave transducers: If the values d and v are not in the range, diagonal mode with 1 sound path may beused, i.e. the same characteristics can be used with doubling the inner pipe diameter. If the values are still notin the range, shear waves transducers regarding the pipe wall thickness have to be selected in step 1b.

order code

GLG 11 23

GLH 7 15

GLK 4 9

GLM 2 5

GLP 1...3

GLQ 0.5...1

5 10 15 20 25pipe wall thickness [mm]

order code

GSG 14 23

GSK 5 9

GSM 2.5 5

GSP 1.5 3

5 10 15 20 25pipe wall thickness [mm]

recommended possible

Page 13: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3EN_LEU 29.9.2009 13

Technical Specification FLUXUS® G70x

Lamb wave transducers1 shear wave transducers1

GLG GSG

GLH

GLK GSK

GLM GSM

GLP

GSP

GLQ

1 inner pipe diameter and max. flow velocity for a typical application with natural gas, N2, O2 in reflection mode with 2 sound paths(Lamb wave transducers)/1 sound path (shear wave transducers)

0 5 10 15 20 25 30 35v�m�s�

200

400

600

800

1000

1200D�mm�

0 5 10 15 20 25 30 35v�m�s�

200

400

600

800

1000

1200D�mm�

0 5 10 15 20 25 30 35v�m�s�

200

400

600

800

1000

1200D�mm�

0 5 10 15 20 25 30 35v�m�s�

200

400

600

800

1000

1200D�mm�

0 5 10 15 20 25 30 35v�m�s�

200

400

600

800

1000

1200D�mm�

0 5 10 15 20 25 30 35v�m�s�

20

40

60

80

100

120

140

D�mm�

0 5 10 15 20 25 30 35v�m�s�

20

40

60

80

100

120

140

D�mm�

0 5 10 15 20 25 30 35v�m�s�

20

40

60

80

100

120

140

D�mm�

0 5 10 15 20 25 30 35v�m�s�

20

40

60

80

100

120

140

D�mm�GDP

0 5 10 15 20 25 30 35v�m�s�

20

40

60

80

100

120

140

D�mm�

Page 14: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

14 29.9.2009 TSFLUXUS_G7V1-3EN_LEU

FLUXUS® G70x Technical Specification

Step 3:

min. medium pressure

Examples

Step 4:

for determination of character 4...11 of the transducer order code (temperature, explosion protection, connec-tion system, extension cable) see page 15

Step 5:

for the technical data of the selected transducer see page 16 et seqq.

Lamb wave transducers shear wave transducers order code medium pressure [ ] order code medium pressure [ ]

metal pipe plastic pipe metal pipe plastic pipemin. min. extended min. min. min. extended min.

GLG 15 10 1 GSG 30 20 1

GLH 15 10 1 GSK 30 20 1

GLK 1 GSM 30 20 1

GLM - - GSP 30 20 1

GLP - -

GLQ - -

d - inner pipe diameter

step 1 pipe wall thickness 12 12 12 30

selected transducer GLG or GLH GLG or GLH GLG or GLH GS2 inner pipe diameter 800 600 800 300

max. flow velocity 15 15 30 15

selected transducer GLG GLG or GLH values not in the range of the characteristics, but by using diagonal mode with 1 sound path, the inner pipe diameter in the char-acteristics is doubled:GLG

GSK

3 min. medium pressure 17 17 17 35

selected transducer GLG GLG or GLHinfluence of noise is reduced with increased transducer frequency, thus rec-ommended:GLH

GLG GSK

bar bar

15 (d > 120 mm)10 (d < 120 mm)

10 (d > 120 mm)5 (d < 120 mm)

10 (d > 60 mm)5 (d < 60 mm)10 (d > 35 mm)5 (d < 35 mm)10 (d > 15 mm)5 (d < 15 mm)

mm

mm

m/s

bar

Page 15: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 15

Technical Specification FLUXUS® G70x

Order Code Key for Transducers

1, 2 3 4 5, 6 7, 8 9...11 no. of character

tra

nsdu

cer

tra

nsdu

cer

fre

quen

cy

- tem

pera

ture

expl

osi

onpr

otec

tion

conn

ectio

nsy

stem

- exte

nsio

n ca

ble

description

GL set of ultrasonic flow transducers for gas measurement, Lamb wave

GS set of ultrasonic flow transducers for gas measurement, shear wave

G 0.2 MHz

H 0.3 MHz (Lamb wave only)

K 0.5 MHz

M 1 MHz

P 2 MHz

Q 4 MHz (Lamb wave only)

N normal temperature range

A1 ATEX zone 1 (with connection system TS)

A2 ATEX zone 2 (with connection system TS)

F2 FM Class I Div. 2 (G704 with connection system TS)

NN not explosion proof

AS with Amphenol connector

TS direct connection or connection via junction box

XXX cable length in m, for max. length of extension cable see page 36

example

GL K - N A1 TS - 030 Lamb wave transducer 0.5 MHz, normal temperature range, zone 1, connec-tion system TS with junction box JB01 and 30 m extension cable

- -

connection system TS:

0 m: without junction box

> 0 m: with junction box JB01 (ATEX zone 1), JB02 (ATEX zone 2, FM), JB03 (not explosion proof), JBP2 (IP 68 transducers for ATEX zone 2), JBP3 (not explosion proof IP 68 transducers)

Page 16: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

16 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Technical Data

Shear Wave Transducers (ATEX zone 1)

technical type GDG1N31 GDK1N31order code GSG-NA1TS GSK-NA1TStransducer frequency MHz 0.2 0.5

medium pressure1

min. extended metal pipe: 20 metal pipe: 20min. metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended 250 70min. recommended 380 80max. recommended 810 500max. extended 1100 720pipe wall thickness min. 14 5max. - -material housing PEEK with stainless steel

cap 304 (1.4301)PEEK with stainless steel cap 304 (1.4301)

contact surface PEEK PEEKdegree of protection according to EN 60529

IP 65 IP 65

transducer cable type 2549 2549length 5 5dimensions length l 129.5 126.5width b 50 50height h 64 53.5dimensional drawing

operating temperature min. °C -40 -40max. °C +130 +130explosion protection

ATEX

transducer GSG-NA1TS GSK-NA1TSzone 1 1explosion protection temperature min. °C -40 -40max. °C +180 +180marking 0044; II2G

Ex q II T6...T3Ta -40...+180 °C

II2D Ex tD A21 IP65 TX

0044; II2GEx q II T6...T3Ta -40...+180 °C

II2D Ex tD A21 IP65 TXcertification IBExU04ATEX1011 X IBExU04ATEX1011 Xtype of protection powder filling powder filling transducer shoe necessary

- -

1 depending on application, typical value for natural gas, N2, compressed air 2 shear wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a flow velocity of

barbar

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

15 m/s

Page 17: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 17

Technical Specification FLUXUS® G70x

Shear Wave Transducers (ATEX zone 1)

technical type GDM2N81 GDP2N81order code GSM-NA1TS GSP-NA1TStransducer frequency MHz 1

medium pressure1

min. extended metal pipe: 20 metal pipe: 20min. metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended 30 15min. recommended 40 20max. recommended 80 40max. extended 120 60pipe wall thickness min. 2.5 1.5max. - -material housing PEEK with stainless steel

cap 304 (1.4301)PEEK with stainless steel cap 304 (1.4301)

contact surface PEEK PEEKdegree of protection according to EN 60529

IP 65 IP 65

transducer cable type 1699 1699length 4 4dimensions length l 69.5 69.5width b 32.5 28height h 61 40dimensional drawing

operating temperature min. °C -40 -40max. °C +130 +130explosion protection

ATEX

transducer GSM-NA1TS GSP-NA1TSzone 1 1explosion protection temperature min. °C -55 -55max. °C +180 +180marking 0044; II2G

Ex eq II T6...T3Ta -55...+180 °C

II2D Ex tD A21 IP65 TX

0044; II2GEx eq II T6...T3Ta -55...+180 °C

II2D Ex tD A21 IP65 TXcertification IBExU07ATEX1168 X IBExU07ATEX1168 Xtype of protection powder filling powder filling transducer shoe necessary

x x

1 depending on application, typical value for natural gas, N2, compressed air 2 shear wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a flow velocity of

barbar

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

15 m/s

Page 18: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

18 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Shear Wave Transducers (ATEX zone 1, IP 68)

technical type GDG1LI1 GDK1LI1 GDM2LI1 GDP2LI1order code GSG-NA1TS/IP68 GSK-NA1TS/IP68 GSM-NA1TS/IP68 GSP-NA1TS/IP68transducer frequency MHz 0.2 0.5 1 2

medium pressure1

min. extended metal pipe: 20 metal pipe: 20 metal pipe: 20 metal pipe: 20min. metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended 250 70 30 15min. recommended 380 80 40 20max. recommended 810 500 80 40max. extended 1100 720 120 60pipe wall thickness min. 14 5 2.5 1.5max. - - - -material housing PEEK with stainless steel

cap 316Ti (1.4571)PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

contact surface PEEK PEEK PEEK PEEKdegree of protection according to EN 60529

IP 68 IP 68 IP 68 IP 68

transducer cable type 2550 2550 2550 2550length 12 12 12 12dimensions length l 128.5 128.5 70 70width b 50 50 28 28height h 75 72 42 42dimensional drawing

operating temperature min. °C -40 -40 -40 -40max. °C +100 +100 +100 +100explosion protection

ATEX

transducer GSG-NA1TS/IP68 GSK-NA1TS/IP68 GSM-NA1TS/IP68 GSP-NA1TS/IP68zone 1 1 1 1explosion protection temperature min. °C -55 -55 -55 -55max. °C +180 +180 +180 +180marking 0044; II2G

Ex q II T6...T3Ta -55...+180 °C

II2D Ex tD A21 IP68 TX

0044; II2GEx q II T6...T3Ta -55...+180 °C

II2D Ex tD A21 IP68 TX

0044; II2GEx q II T6...T3Ta -55...+180 °C

II2D Ex tD A21 IP68 TX

0044; II2GEx q II T6...T3Ta -55...+180 °C

II2D Ex tD A21 IP68 TXcertification IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xtype of protection powder filling powder filling powder filling powder filling transducer shoe necessary

x x x x

1 depending on application, typical value for natural gas, N2, compressed air 2 shear wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a flow velocity of

barbar

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

hb

lh

bl

15 m/s

Page 19: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 19

Technical Specification FLUXUS® G70x

Shear Wave Transducers (ATEX zone 1/2 (gas/dust), extended temperature range)

technical type GDM2E85 GDP2E85order code GSM-EA1TS GSP-EA1TStransducer frequency MHz 1 2

medium pressure1

min. extended metal pipe: 20 metal pipe: 20min. metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended 30 15min. recommended 40 20max. recommended 80 40max. extended 120 60pipe wall thickness min. 2.5 1.5max. - -material housing PI with stainless steel

cap 304 (1.4301)PI with stainless steel cap 304 (1.4301)

contact surface PI PIdegree of protection according to EN 60529

IP 56 IP 56

transducer cable type 6111 6111length 4 4dimensions length l 69.5 69.5width b 32.5 32.5height h 61 61dimensional drawing

operating temperature min. °C -30 -30max. °C +200 +200explosion protection

ATEX

transducer GSM-EA1TS GSP-EA1TSzone 1 1explosion protection temperature min. °C -45 -45max. °C +225 +225marking 0044; II2G

Ex eq II T6...T2Ta -45...+225 °C

II3D Ex tD A22 IP56 TX

0044; II2GEx eq II T6...T2Ta -45...+225 °C

II3D Ex tD A22 IP56 TXcertification IBExU07ATEX1168 X IBExU07ATEX1168 Xtype of protection powder filling powder filling transducer shoe necessary

x x

1 depending on application, typical value for natural gas, N2, compressed air 2 shear wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a flow velocity of

barbar

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

15 m/s

Page 20: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

20 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Shear Wave Transducers (ATEX zone 2, FM, without explosion protection)

technical type GDG1N52 GDK1N52 GDM1N52 GDP1N52order code GSG-NA2TS

GSG-NF2TSGSG-NNNTS

GSK-NA2TSGSK-NF2TSGSK-NNNTS

GSM-NA2TSGSM-NF2TSGSM-NNNTS

GSP-NA2TSGSP-NF2TSGSP-NNNTS

transducer frequency MHz 0.2 0.5 1 2medium pressure1

min. extended metal pipe: 20 metal pipe: 20 metal pipe: 20 metal pipe: 20min. metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended 250 70 30 15min. recommended 380 80 40 20max. recommended 810 500 80 40max. extended 1100 720 120 60pipe wall thickness min. 14 5 2.5 1.5max. - - - -material housing PEEK with stainless steel

cap 304 (1.4301)PEEK with stainless steel cap 304 (1.4301)

stainless steel 304 (1.4301)

stainless steel 304 (1.4301)

contact surface PEEK PEEK PEEK PEEKdegree of protection according to EN 60529

IP 67 IP 67 IP 67 IP 67

transducer cable type 1699 1699 1699 1699length 5 5 4 4dimensions length l 129.5 126.5 60 60width b 47 47 30 30height h 66.4 55.9 33.5 33.5dimensional drawing

operating temperature min. °C -40 -40 -40 -40max. °C +130 +130 +130 +130explosion protection

ATEX

transducer GSG-NA2TS GSK-NA2TS GSM-NA2TS GSP-NA2TSzone 2 2 2 2explosion protection temperature min. °C -55 -55 -20 -20max. °C +190 +190 +130 +130marking II3G

Ex nA II T6...T3Ta -55...+190 °C

II3D Ex tD A22 IP67 TX

II3GEx nA II T6...T3Ta -55...+190 °C

II3D Ex tD A22 IP67 TX

II3GEx nA II T6...T4Ta -20...+130 °C

II3D Ex tD A22 IP67 TX

II3GEx nA II T6...T4Ta -20...+130 °C

II3D Ex tD A22 IP67 TXcertification - - - -type of protection non sparking, protection

by enclosure non sparking, protection by enclosure

non sparking, protection by enclosure

non sparking, protection by enclosure

transducer shoe necessary

x x - -

FM

transducer GSG-NF2TS GSK-NF2TS GSM-NF2TS GSP-NF2TSexplosion protection temperature min. °C -40 -40 -40 -40max. °C +125 +125 +125 +125marking NI/Cl. I, II, III/Div. 2/

Gp A-G/T4 Ta = 125 °C NI/Cl. I, II, III/Div. 2/

Gp A-G/T4 Ta = 125 °C NI/Cl. I, II, III/Div. 2/

Gp A-G/T4 Ta = 125 °C NI/Cl. I, II, III/Div. 2/

Gp A-G/T4 Ta = 125 °Ctype of protection non incendive non incendive non incendive non incendive

1 depending on application, typical value for natural gas, N2, compressed air 2 shear wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a flow velocity of

barbar

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

hb

l

hb

l

15 m/s

Page 21: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 21

Technical Specification FLUXUS® G70x

Shear Wave Transducers (ATEX zone 2, without explosion protection, IP 68)

technical type GDG1LI8 GDK1LI8 GDM2LI8 GDP2LI8order code GSG-NA2TS/IP68

GSG-NNNTS/IP68GSK-NA2TS/IP68GSK-NNNTS/IP68

GSM-NA2TS/IP68GSM-NNNTS/IP68

GSP-NA2TS/IP68GSP-NNNTS/IP68

transducer frequency MHz 0.2 0.5 1 2

medium pressure1

min. extended metal pipe: 20 metal pipe: 20 metal pipe: 20 metal pipe: 20min. metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended 250 70 30 15min. recommended 380 80 40 20max. recommended 810 500 80 40max. extended 1100 720 120 60pipe wall thickness min. 14 5 2.5 1.5max. - - - -material housing PEEK with stainless steel

cap 316Ti (1.4571)PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

contact surface PEEK PEEK PEEK PEEKdegree of protection according to EN 60529

IP 68 IP 68 IP 68 IP 68

transducer cable type 2550 2550 2550 2550length 12 12 12 12dimensions length l 128.5 128.5 70 70width b 50 50 28 28height h 75 72 42 42dimensional drawing

operating temperature min. °C -40 -40 -40 -40max. °C +100 +100 +100 +100explosion protection

ATEX

transducer GSG-NA2TS/IP68 GSK-NA2TS/IP68 GSM-NA2TS/IP68 GSP-NA2TS/IP68zone 2 2 2 2explosion protection temperature min. °C -55 -55 -55 -55max. °C +90 +90 +90 +90marking II3G

Ex nA II T6...T5Ta -55...+90 °C

II3D Ex tD A22 IP68 TX

II3GEx nA II T6...T5Ta -55...+90 °C

II3D Ex tD A22 IP68 TX

II3GEx nA II T6...T5Ta -55...+90 °C

II3D Ex tD A22 IP68 TX

II3GEx nA II T6...T5Ta -55...+90 °C

II3D Ex tD A22 IP68 TXcertification - - - -type of protection non sparking, protection

by enclosure non sparking, protection by enclosure

non sparking, protection by enclosure

non sparking, protection by enclosure

transducer shoe necessary

x x x x

1 depending on application, typical value for natural gas, N2, compressed air 2 shear wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a flow velocity of

barbar

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

hb

l

hb

l

15 m/s

Page 22: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

22 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Shear Wave Transducers (without explosion protection and with connection system AS)

technical type GDG1NZ7 GDK1NZ7 GDM1NZ7 GDP1NZ7order code GSG-NNNAS GSK-NNNAS GSM-NNNAS GSP-NNNAStransducer frequency MHz 0.2 0.5 1 2

medium pressure1

min. extended metal pipe: 20 metal pipe: 20 metal pipe: 20 metal pipe: 20min. metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended 250 70 30 15min. recommended 380 80 40 20max. recommended 810 500 80 40max. extended 1100 720 120 60pipe wall thickness min. 14 5 2.5 1.5max. - - - -material housing PEEK with stainless steel

cap 304 (1.4301)PEEK with stainless steel cap 304 (1.4301)

stainless steel 304 (1.4301)

stainless steel 304 (1.4301)

contact surface PEEK PEEK PEEK PEEKdegree of protection according to EN 60529

IP 67 IP 67 IP 67 IP 67

transducer cable type 1699 1699 1699 1699length 5 5 4 4dimensions length l 129.5 126.5 60 60width b 47 47 30 30height h 66.4 55.9 33.5 33.5dimensional drawing

operating temperature min. °C -40 -40 -40 -40max. °C +130 +130 +130 +1301 depending on application, typical value for natural gas, N2, compressed air 2 shear wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a flow velocity of

barbar

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

hb

lh

bl

15 m/s

Page 23: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 23

Technical Specification FLUXUS® G70x

Shear Wave Transducers (extended temperature range, without Explosion Protection and withConnection System AS)

technical type GDM2EZ7 GDP2EZ7order code GSM-ENNAS GSP-ENNAStransducer frequency MHz 1 2

medium pressure1

min. extended metal pipe: 20 metal pipe: 20min. metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended 30 15min. recommended 40 20max. recommended 80 40max. extended 120 60pipe wall thickness min. 2.5 1.5max. - -material housing PI with stainless steel

cap 304 (1.4301)PI with stainless steel cap 304 (1.4301)

contact surface PI PIdegree of protection according to EN 60529

IP 65 IP 65

transducer cable type 6111 6111length 4 4dimensions length l 69.5 69.5width b 32.5 32.5height h 61 61dimensional drawing

operating temperature min. °C -30 -30max. °C +200 +2001 depending on application, typical value for natural gas, N2, compressed air 2 shear wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a flow velocity of

barbar

mmmmmmmm

mmmm

m

mmmmmm

hb

l

15 m/s

Page 24: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

24 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Shear Wave Transducers (extended temperature range, ATEX zone 2 or without explosion protection)

technical type GDM2E52 GDP2E52order code GSM-EA2TS

GSM-ENNTSGSP-EA2TSGSP-ENNTS

transducer frequency MHz 1 2

medium pressure1

min. extended metal pipe: 20 metal pipe: 20min. metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended 30 15min. recommended 40 20max. recommended 80 40max. extended 120 60pipe wall thickness min. 2.5 1.5max. - -material housing PI with stainless steel

cap 304 (1.4301)PI with stainless steel cap 304 (1.4301)

contact surface PI PIdegree of protection according to EN 60529

IP 56 IP 56

transducer cable type 6111 6111length 4 4dimensions length l 69.5 69.5width b 32.5 32.5height h 61 61dimensional drawing

operating temperature min. °C -30 -30max. °C +200 +200explosion protection

ATEX

transducer GSM-EA2TS GSP-EA2TSzone 2 2explosion protection temperature min. °C -45 -45max. °C +235 +235marking II3G

Ex nA II T6...T2Ta -45...+235 °C

II3D Ex tD A22 IP56 TX

II3GEx nA II T6...T2Ta -45...+235 °C

II3D Ex tD A22 IP56 TXcertification - -type of protection non sparking non sparking transducer shoe necessary

x x

1 depending on application, typical value for natural gas, N2, compressed air 2 shear wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal mode and for a flow velocity of

barbar

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

15 m/s

Page 25: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 25

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (ATEX zone 1)

technical type GRG1N33 GRH1N33 GRK1N33order code GLG-NA1TS GLH-NA1TS GLK-NA1TStransducer frequency MHz 0.2 0.3 0.5

medium pressure1

min. extended metal pipe: 10 metal pipe: 10 metal pipe:

min. metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:

plastic pipe: 1

inner pipe diameter d2

min. extended 190 120 60min. recommended 220 140 80max. recommended 900 600 300max. extended 1600 1000 500pipe wall thickness min. 11 7 4max. 23 15 9material housing PPSU with stainless steel

cap 304 (1.4301)PPSU with stainless steel cap 304 (1.4301)

PPSU with stainless steel cap 304 (1.4301)

contact surface PPSU PPSU PPSUdegree of protection according to EN 60529

IP 65 IP 65 IP 65

transducer cable type 2549 2549 2549length 5 5 5dimensions length l 128.5 128.5 128.5width b 50 50 50height h 67.5 67.5 67.5dimensional drawing

operating temperature min. °C -40 -40 -40max. °C +170 +170 +170explosion protection

ATEX

transducer GLG-NA1TS GLH-NA1TS GLK-NA1TSzone 1 1 1explosion protection temperature min. °C -40 -40 -40max. °C +140 +140 +140marking 0044; II2G

Ex q II T6...T3Ta -40...+140 °C

II2D Ex tD A21 IP65 TX

0044; II2GEx q II T6...T3Ta -40...+140 °C

II2D Ex tD A21 IP65 TX

0044; II2GEx q II T6...T3Ta -40...+140 °C

II2D Ex tD A21 IP65 TXcertification IBExU04ATEX1011 X IBExU04ATEX1011 X IBExU04ATEX1011 Xtype of protection powder filling powder filling powder filling transducer shoe necessary

- - -

1 depending on application, typical value for natural gas, N2, compressed air 2 Lamb wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity of pipe diameter max. extended: in diagonal mode and for a flow velocity of

bar10 (d > 120 mm)5 (d < 120 mm)

bar15 (d > 120 mm)10 (d < 120 mm)

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

hb

l

15 m/s25 m/s

Page 26: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

26 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Lamb Wave Transducers (ATEX zone 1)

technical type GRM1N83 GRP1N83 GRQ1N83order code GLM-NA1TS GLP-NA1TS GLQ-NA1TStransducer frequency MHz 1 2 4

medium pressure1

min. extended - - -min. metal pipe: metal pipe: metal pipe:

inner pipe diameter d2

min. extended 30 15 7min. recommended 40 20 10max. recommended 90 50 22max. extended 150 70 35pipe wall thickness min. 2 1 0.5max. 5 3 1material housing PPSU with stainless steel

cap 304 (1.4301)PPSU with stainless steel cap 304 (1.4301)

PPSU with stainless steel cap 304 (1.4301)

contact surface PPSU PPSU PPSUdegree of protection according to EN 60529

IP 65 IP 65 IP 65

transducer cable type 1699 1699 1699length 4 4 3dimensions length l 74 74 42width b 28 28 18height h 42.9 42.9 25.5dimensional drawing

operating temperature min. °C -40 -40 -40max. °C +170 +170 +170explosion protection

ATEX

transducer GLM-NA1TS GLP-NA1TS GLQ-NA1TSzone 1 1 1explosion protection temperature min. °C -55 -55 -55max. °C +140 +140 +140marking 0044; II2G

Ex eq II T6...T3Ta -55...+140 °C

II2D Ex tD A21 IP65 TX

0044; II2GEx eq II T6...T3Ta -55...+140 °C

II2D Ex tD A21 IP65 TX

0044; II2GEx eq II T6...T3Ta -55...+140 °C

II2D Ex tD A21 IP65 TXcertification IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xtype of protection powder filling powder filling powder filling transducer shoe necessary

x x x

1 depending on application, typical value for natural gas, N2, compressed air 2 Lamb wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity of pipe diameter max. extended: in diagonal mode and for a flow velocity of

barbar

10 (d > 60 mm)5 (d < 60 mm)

10 (d > 35 mm)5 (d < 35 mm)

10 (d > 15 mm)5 (d < 15 mm)

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

15 m/s25 m/s

Page 27: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 27

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (ATEX zone 1, IP 68)

technical type GRG1LI3 GRH1LI3 GRK1LI3order code GLG-NA1TS/IP68 GLH-NA1TS/IP68 GLK-NA1TS/IP68transducer frequency MHz 0.2 0.3 0.5

medium pressure1

min. extended metal pipe: 10 metal pipe: 10 metal pipe:

min. metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:

plastic pipe: 1

inner pipe diameter d2

min. extended 190 120 60min. recommended 220 140 80max. recommended 900 600 300max. extended 1600 1000 500pipe wall thickness min. 11 7 4max. 23 15 9material housing PPSU with stainless steel

cap 316Ti (1.4571)PPSU with stainless steel cap 316Ti (1.4571)

PPSU with stainless steel cap 316Ti (1.4571)

contact surface PPSU PPSU PPSUdegree of protection according to EN 60529

IP 68 IP 68 IP 68

transducer cable type 2550 2550 2550length 12 12 12dimensions length l 143.5 143.5 143.5width b 50 50 50height h 84 84 84dimensional drawing

operating temperature min. °C -40 -40 -40max. °C +100 +100 +100explosion protection

ATEX

transducer GLG-NA1TS/IP68 GLH-NA1TS/IP68 GLK-NA1TS/IP68zone 1 1 1explosion protection temperature min. °C -55 -55 -55max. °C +140 +140 +140marking 0044; II2G

Ex q II T6...T3Ta -55...+140 °C

II2D Ex tD A21 IP68 TX

0044; II2GEx q II T6...T3Ta -55...+140 °C

II2D Ex tD A21 IP68 TX

0044; II2GEx q II T6...T3Ta -55...+140 °C

II2D Ex tD A21 IP68 TXcertification IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xtype of protection powder filling powder filling powder filling transducer shoe necessary

x x x

1 depending on application, typical value for natural gas, N2, compressed air 2 Lamb wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity of pipe diameter max. extended: in diagonal mode and for a flow velocity of

bar10 (d > 120 mm)5 (d < 120 mm)

bar15 (d > 120 mm)10 (d < 120 mm)

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

hb

l

15 m/s25 m/s

Page 28: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

28 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Lamb Wave Transducers (ATEX zone 2, FM or without explosion protection)technical type GRG1N52 GRH1N52 GRK1N52order code GLG-NA2TS

GLG-NF2TSGLG-NNNTS

GLH-NA2TSGLH-NF2TSGLH-NNNTS

GLK-NA2TSGLK-NF2TSGLK-NNNTS

transducer frequency MHz 0.2 0.3 0.5medium pressure1

min. extended metal pipe: 10 metal pipe: 10 metal pipe:

min. metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:

plastic pipe: 1inner pipe diameter d2

min. extended 190 120 60min. recommended 220 140 80max. recommended 900 600 300max. extended 1600 1000 500pipe wall thickness min. 11 7 4max. 23 15 9material housing PPSU with stainless steel cap 304

(1.4301)PPSU with stainless steel cap 304 (1.4301)

PPSU with stainless steel cap 304 (1.4301)

contact surface PPSU PPSU PPSUdegree of protection according to EN 60529

IP 67 IP 67 IP 67

transducer cable type 1699 1699 1699length 5 5 5dimensions length l 128.5 128.5 128.5width b 47 47 47height h 69.9 69.9 69.9dimensional drawing

operating temperature min. °C -40 -40 -40max. °C +170 +170 +170explosion protection

ATEX

transducer GLG-NA2TS GLH-NA2TS GLK-NA2TSzone 2 2 2explosion protection temperature min. °C -55 -55 -55max. °C +150 +150 +150marking II3G Ex nA II T6...T3

Ta -55...+150 °C II3D Ex tD A22 IP67 TX

II3G Ex nA II T6...T3Ta -55...+150 °C

II3D Ex tD A22 IP67 TX

II3G Ex nA II T6...T3Ta -55...+150 °C

II3D Ex tD A22 IP67 TXcertification - - -type of protection non sparking, protection by enclo-

sure non sparking, protection by enclo-sure

non sparking, protection by enclo-sure

transducer shoe necessary

x x x

FM

transducer GLG-NF2TS GLH-NF2TS GLK-NF2TSexplosion protection temperature min. °C -40 -40 -40max. °C +125 +125 +125marking NI/Cl. I, II, III/Div. 2/Gp A-G/T4

Ta = 125 °C NI/Cl. I, II, III/Div. 2/Gp A-G/T4

Ta = 125 °C NI/Cl. I, II, III/Div. 2/Gp A-G/T4

Ta = 125 °Ctype of protection non incendive non incendive non incendive

1 depending on application, typical value for natural gas, N2, compressed air 2 Lamb wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity of pipe diameter max. extended: in diagonal mode and for a flow velocity of

bar10 (d > 120 mm), 5 (d < 120 mm)

bar15 (d > 120 mm), 10 (d < 120 mm)

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

lh

bl

15 m/s25 m/s

Page 29: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 29

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (ATEX zone 2 or without explosion protection)

technical type GRM1N52 GRP1N52 GRQ1N52order code GLM-NA2TS

GLM-NNNTSGLP-NA2TSGLP-NNNTS

GLQ-NA2TSGLQ-NNNTS

transducer frequency MHz 1 2 4

medium pressure1

min. extended - - -min. metal pipe: metal pipe: metal pipe:

inner pipe diameter d2

min. extended 30 15 7min. recommended 40 20 10max. recommended 90 50 22max. extended 150 70 35pipe wall thickness min. 2 1 0.5max. 5 3 1material housing PPSU with stainless steel

cap 304 (1.4301)PPSU with stainless steel cap 304 (1.4301)

PPSU with stainless steel cap 304 (1.4301)

contact surface PPSU PPSU PPSUdegree of protection according to EN 60529

IP 65 IP 65 IP 65

transducer cable type 1699 1699 1699length 4 4 3dimensions length l 74 74 42width b 28 28 18height h 42.9 42.9 25.5dimensional drawing

operating temperature min. °C -40 -40 -40max. °C +170 +170 +170explosion protection

ATEX

transducer GLM-NA2TS GLP-NA2TS GLQ-NA2TSzone 2 2 2explosion protection temperature min. °C -55 -55 -55max. °C +150 +150 +150marking II3G

Ex nA II T6...T3Ta -55...+150 °C

II3D Ex tD A22 IP67 TX

II3GEx nA II T6...T3Ta -55...+150 °C

II3D Ex tD A22 IP67 TX

II3GEx nA II T6...T3Ta -55...+150 °C

II3D Ex tD A22 IP67 TXcertification - - -type of protection non sparking, protection

by enclosure non sparking, protection by enclosure

non sparking, protection by enclosure

transducer shoe necessary

x x x

1 depending on application, typical value for natural gas, N2, compressed air 2 Lamb wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity of pipe diameter max. extended: in diagonal mode and for a flow velocity of

barbar

10 (d > 60 mm)5 (d < 60 mm)

10 (d > 35 mm)5 (d < 35 mm)

10 (d > 15 mm)5 (d < 15 mm)

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

15 m/s25 m/s

Page 30: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

30 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Lamb Wave Transducers (ATEX zone 2 or without explosion protection, IP 68)

technical type GRG1LI8 GRH1LI8 GRK1LI8order code GLG-NA2TS/IP68

GLG-NNNTS/IP68GLH-NA2TS/IP68GLH-NNNTS/IP68

GLK-NA2TS/IP68GLK-NNNTS/IP68

transducer frequency MHz 0.2 0.3 0.5medium pressure1

min. extended metal pipe: 10 metal pipe: 10 metal pipe:

min. metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:

plastic pipe: 1inner pipe diameter d2

min. extended 190 120 60min. recommended 220 140 80max. recommended 900 600 300max. extended 1600 1000 500pipe wall thickness min. 11 7 4max. 23 15 9material housing PPSU with stainless steel cap

316Ti (1.4571)PPSU with stainless steel cap 316Ti (1.4571)

PPSU with stainless steel cap 316Ti (1.4571)

contact surface PPSU PPSU PPSUdegree of protection according to EN 60529

IP 68 IP 68 IP 68

transducer cable type 2550 2550 2550length 12 12 12dimensions length l 143.5 143.5 143.5width b 50 50 50height h 84 84 84dimensional drawing

operating temperature min. °C -40 -40 -40max. °C +100 +100 +100explosion protection

ATEX

transducer GLG-NA2TS/IP68 GLH-NA2TS/IP68 GLK-NA2TS/IP68zone 2 2 2explosion protection temperature min. °C -40 -40 -40max. °C +90 +90 +90marking II3G Ex nA II T6...T5

Ta -40...+90 °C II3D Ex tD A22 IP68 TX

II3G Ex nA II T6...T5Ta -40...+90 °C

II3D Ex tD A22 IP68 TX

II3G Ex nA II T6...T5Ta -40...+90 °C

II3D Ex tD A22 IP68 TXcertification - - -type of protection non sparking, protection by enclo-

sure non sparking, protection by enclo-sure

non sparking, protection by enclo-sure

transducer shoe necessary

x x x

1 depending on application, typical value for natural gas, N2, compressed air 2 Lamb wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity of pipe diameter max. extended: in diagonal mode and for a flow velocity of

bar10 (d > 120 mm)5 (d < 120 mm)

bar15 (d > 120 mm)10 (d < 120 mm)

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

hb

l

15 m/s25 m/s

Page 31: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 31

Technical Specification FLUXUS® G70x

Lamb Wave Transducers (without explosion protection and with connection system AS)

technical type GRG1NC3 GRH1NC3 GRK1NC3order code GLG-NNNAS GLH-NNNAS GLK-NNNAStransducer frequency MHz 0.2 0.3 0.5

medium pressure1

min. extended metal pipe: 10 metal pipe: 10 metal pipe:

min. metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:

plastic pipe: 1

inner pipe diameter d2

min. extended 190 120 60min. recommended 220 140 80max. recommended 900 600 300max. extended 1600 1000 500pipe wall thickness min. 11 7 4max. 23 15 9material housing PPSU with stainless steel

cap 304 (1.4301)PPSU with stainless steel cap 304 (1.4301)

PPSU with stainless steel cap 304 (1.4301)

contact surface PPSU PPSU PPSUdegree of protection according to EN 60529

IP 65 IP 65 IP 65

transducer cable type 1699 1699 1699length 5 5 5dimensions length l 128.5 128.5 128.5width b 47 47 47height h 69.9 69.9 69.9dimensional drawing

operating temperature min. °C -40 -40 -40max. °C +170 +170 +1701 depending on application, typical value for natural gas, N2, compressed air 2 Lamb wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity of pipe diameter max. extended: in diagonal mode and for a flow velocity of

bar10 (d > 120 mm)5 (d < 120 mm)

bar15 (d > 120 mm)10 (d < 120 mm)

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

hb

l

15 m/s25 m/s

Page 32: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

32 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Lamb Wave Transducers (without explosion protection and with connection system AS)

technical type GRM1NC3 GRP1NC3 GRQ1NC3order code GLM-NNNAS GLP-NNNAS GLQ-NNNAStransducer frequency MHz 1 2 4

medium pressure1

min. extended - - -min. metal pipe: metal pipe: metal pipe:

inner pipe diameter d2

min. extended 30 15 7min. recommended 40 20 10max. recommended 90 50 22max. extended 150 70 35pipe wall thickness min. 2 1 0.5max. 5 3 1material housing PPSU with stainless steel

cap 304 (1.4301)PPSU with stainless steel cap 304 (1.4301)

PPSU with stainless steel cap 304 (1.4301)

contact surface PPSU PPSU PPSUdegree of protection according to EN 60529

IP 65 IP 65 IP 65

transducer cable type 1699 1699 1699length 4 4 3dimensions length l 74 74 42width b 28 28 18height h 42.9 42.9 25.5dimensional drawing

operating temperature min. °C -40 -40 -40max. °C +170 +170 +1701 depending on application, typical value for natural gas, N2, compressed air 2 Lamb wave transducers:

typical values for natural gas, N2, O2, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection mode and for a flow velocity of pipe diameter max. extended: in diagonal mode and for a flow velocity of

barbar

10 (d > 60 mm)5 (d < 60 mm)

10 (d > 35 mm)5 (d < 35 mm)

10 (d > 15 mm)5 (d < 15 mm)

mmmmmmmm

mmmm

m

mmmmmm

hb

l

hb

l

hb

l

15 m/s25 m/s

Page 33: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 33

Technical Specification FLUXUS® G70x

Transducer Mounting Fixtures

Tension Straps and Clasps

material: stainless steel 304 (1.4301), 303 (1.4305)

length:

Tension Straps, Clasps and Mounting Shoes

material: stainless steel 304 (1.4301), 303 (1.4305)

length:

Variofix Rail VFX with Tension Straps and Clasps1

material: stainless steel 304 (1.4301), 303 (1.4305)

dimensions:

(b - transducer width + h - dependent on transducer)

length:

1 on request

10/20 m

10/20 m

220/320/520/1020 x b x h mm30 mm

10/20 m

Page 34: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

34 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Coupling Materials for Transducers

Technical Data

normal temperature range(4th character of transducer order code = N)

extended temperature range(4th character of transducer order code = E)

< 100 °C 100...170 °C < 150 °C 150...200 °C

< 2 h coupling com-pound type N

coupling com-pound type E

coupling com-pound type E

coupling com-pound type E or H

< 24 h coupling com-pound type N

coupling com-pound type E

coupling com-pound type E

coupling foiltype VT

long time measure-ment

indoor coupling com-pound type N

coupling com-pound type E

coupling foiltype VT1

coupling foiltype VT2

outdoor coupling foiltype VT

coupling foiltype VT

coupling foiltype VT1

coupling foiltype VT2

1 < 5 years2 < 6 months

type order code temperature material remark °C

coupling compoundtype N

990739-1 -30...+130 mineral grease paste

coupling compoundtype E

990739-2 -30...+200 silicone paste

coupling compoundtype H

990739-3 -30...+250 fluoropolymer paste

coupling foil type VT 990739-0 -10...+150,peak max. 200

fluoroelastomer for transducers with transducer frequency G, H, K

990739-6 for transducers with transducer frequency M, P

990739-5 for transducers with transducer frequency Q

990739-10 for transducers with transducer frequency S

Page 35: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 35

Technical Specification FLUXUS® G70x

Damping Mats (Option)Damping mats will be used for the gas measurement to reduce noise influences on the measurement.

Transducer damping mats will be installed below the transducers.

Pipe damping mats will be installed at reflection points, e.g. flange, weld.

Selection of Damping Mats type description outer pipe

diameterdimensions

l x b x htransducerfrequency

(3rd character of transducer order

code)

techni-cal type

temperature remark

G H K M P °Ctransducer damping mat C self-adhesive, for stationary

installation < 80 450 x 115 x 0.5 - - - x x C20S3 -25...+60 80 900 x 230 x 0.5 - - x x - C20S2

900 x 230 x 1.3 x x - - - C50S2pipe damping mat B self-adhesive, for stationary

installation l x 100 x 0.9 x x x x x B35R2 -35...+50 l - see table below

Length of Pipe Damping Mat - Type B(length l depending on transducer frequency and outer pipe diameter)

outer pipe diameter D transducer frequency G, H K, M, P

100 2 1200 6 3300 12 6500 32 161000 126 63

b

lD

transducer damping mat pipe damping mat

reflection mode diagonal mode

D - outer pipe diameter

reflection point

mm mm

mm m m

Page 36: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

36 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Connection Systems

x, y - transducer cable length

l - max. length of extension cable

Connection System AS

Connection System TS

connection via junction box direct connection

(only G704, G704 A2)

(not explosion proof transducers)

transducer frequency(3rd character of transducer

order code)

G, H, K M, P Q S

x y l x y l x y l x y l

cable length 2 3 100 2 2 100 2 1 50 1 1 20m

FLU

XU

S

xl y

transducer frequency(3rd character of transducer

order code)

G, H, K M, P Q S

x l x l x l x l

cable length m 5 300 4 300 3 90 2 40

ATEX zone 2, FM, without explosion protection

LX

FLU

XU

S

l

xJB02, JB03

X

FLU

XU

S

x

ATEX zone 1 ATEX zone 2 (order code ***-NA2TS/IP68)

not explosion proof (order code ***-NNNTS/IP68)

L

X

FLU

XU

S

l

xJB01, JBP2, JBP3

X

FLU

XU

S

x

Page 37: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 37

Technical Specification FLUXUS® G70x

Transducer Cables

Technical Data

transducer cable extension cable item number 1699 2549 2550 6111 2551 2552 2615connection system AS TS TSstandard length see table

above see table above

12 see table above 10

- -

max. length - - - - see table above

see table above

see table above

temperature °C -55...+200 -100...+200 -40...+100 -100...+225 < 115 < 80 -40...+70properties longitudinal

water tight halogen free

fire propaga-tion test according to IEC 60332-1

combustion test accord-ing to IEC 60754-2

sheathmaterial stainless steel

304 (1.4301)- - stainless steel

304 (1.4301)- - -

outer diameter 8 - - 8 - - -

cable jacket material PTFE PTFE PUR PFA TPE-O TPV PURouter diameter 2.9 5.3 5.2 ±0.2 2.7 8 12 12

thickness 0.3 0.5 0.9 0.5 2

color brown black gray white black black black shield x x x x x x x

m 1

m

mm

mm

mm

Page 38: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

38 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Junction Box

Technical Data

Dimensions

technical type JB01S4E3M JB02 JB03 JBP2 JBP3dimensions see dimensional

drawing see dimensional drawing

see dimensional drawing

see dimensional drawing

see dimensional drawing

fixation wall mountingoption: 2 " pipe mounting

wall mountingoption: 2 " pipe mounting

wall mountingoption: 2 " pipe mounting

wall mountingoption: 2 " pipe mounting

wall mountingoption: 2 " pipe mounting

material housing stainless steel

316L (1.4404)stainless steel 304 (1.4301)

stainless steel 304 (1.4301)

stainless steel 316L (1.4404)

stainless steel 316L (1.4404)

gasket silicone silicone silicone silicone silicone degree of protection according to EN 60529

IP 67 IP 67 IP 67 IP 67 IP 67

cable gland M20 M20 M20 M20 M20 operating temperature min. °C -40 -40 -40 -40 -40max. °C +80 +80 +80 +80 +80explosion protection

ATEX

zone 1 2 - 2 -marking 0044

II2G Ex e mb II (T6)...T4Ta -40...+(70) 80 °C

II2D Ex tD A21 IP 67 T 100 °C

II3G Ex nA II T6...T4Ta -40...+80 °C

II3D Ex tD A22 IP67 T 100 °C

- II3G Ex nA II T6...T4Ta -40...+80 °C

II3D Ex tD A22 IP67 T 100 °C

-

certification IBExU06ATEX1161 - - - -type of protection junction box:

increased safety

decoupled network: encapsulation

non sparking, pro-tection by enclo-sure

- non sparking, pro-tection by enclo-sure

-

174

1192

wall mounting holder

70

163.5

156

Ø 9

Page 39: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 39

Technical Specification FLUXUS® G70x

2 " Pipe Mounting Kit (option)

Terminal Assignment

JB01

Transducers

terminal strip KL1terminal connection

V signal VS shield RS shield R signal

Extension Cable (Flowmeter)

terminal strip KL2terminal connection

TV signal TVS shield TRS shield TR signal

�%

�%�

�&

�&

���

��

% %�

��

� " � � " �

Page 40: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

40 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

JB02, JB03

JBP2, JBP3

Transducers

terminal connection XV SMB connector XR SMB connector

Extension Cable (Flowmeter)

terminal strip KL2terminal connection

TV signal TVS shield TRS shield TR signal

% �

� " �

�%

�%�

�&

�&

���

��

Transducers

terminal strip KL1terminal connection

TV signal TVS shield TRS shield TR signal

Extension Cable (Flowmeter)

terminal strip KL2terminal connection

TV signal TVS shield TRS shield TR signal

�%

�%�

�&

�&

���

��

�%

�%�

���

��

� " � � " �

Page 41: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 41

Technical Specification FLUXUS® G70x

Temperature Probes (Option)

Technical Data order code 670415-1

770415-1670414-1770414-1

770415-1A2 770414-1A2 670415-2 670414-2

type Pt100 Pt100 paired according to DIN 1434-1

Pt100 Pt100 paired according to DIN 1434-1

Pt100 Pt100 paired according to DIN 1434-1

design 4-wire 4-wire 4-wire measuring range °C -30...+250 -30...+250 -50...+250

accuracy T ±(0.15 °C + 2 . 10-3 . T), class A ±(0.15 °C + 2 . 10-3 . T), class A ±(0.15 °C + 2 . 10-3 . T), class Aaccuracy T - ≤ 0.1 K

(3K < T < 6 K), more corre-sponding to EN 1434-1

- ≤ 0.1 K(3K < T < 6 K), more corre-sponding to EN 1434-1

- ≤ 0.1 K(3K < T < 6 K), more corre-sponding to EN 1434-1

response time s 50 50 8housing aluminum aluminum PEEK, stainless steel 304

(1.4301), Cudegree of protection according to EN 60529

IP 66 IP 66 IP 66

cable gland M12 M12 M12 weight (without connector) 0.25 0.5 0.25 0.5 0.32 0.64

fixation clamp-on clamp-on clamp-on accessories - - plastic protection plate, isolation

foam dimensions length l 15 15 15 14

width b 15 15 15 30

height h 20 20 20 27

dimensional drawing 670415-1: A770415-1: C

670414-1: A770414-1: C

C B

explosion protection

ATEX

zone - 2 -explosion protection temperature min. - -30 -max. - +250 -marking - II3G

Ex nA II T6...T2Ta -30...+250 °C

-

kg

mm

mm

mm

extension cable h

l

junction box

C

B

h

l

A

C

extension cable

Page 42: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

42 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

Connection

Temperature Probe

Connector

Cables

pin cable of temperature probe extension cable 1 white/blue blue 2 red/blue gray 3, 4, 5 not connected 6 red red 7 white white 8 not connected

cable of temperature probe extension cable

type 4 x 0.25 mm2 black or white LIYCY 8 x 0.14 mm2 gray standard length 3 5/10/25

max. length - 200

cable jacket PTFE PVC

Junction Box

technical type JBT2 JBT3dimensions see dimensional drawing see dimensional drawing fixation wall mounting

option: 2 " pipe mountingwall mountingoption: 2 " pipe mounting

material housing stainless steel 316L (1.4404) stainless steel 316L (1.4404)gasket silicone silicone degree of protection according to EN 60529

IP 67 IP 67

cable gland M12 M12 operating temperature min. °C -40 -40max. °C +80 +80explosion protection

ATEX

zone 2 -marking II3G Ex nA II T6...T4

Ta -40...+80 °C II3D Ex tD A22 IP67 T 100 °C

-

certification - -type of protection non sparking, protection by enclo-

sure -

red/blue red white/blue white

� �

m

m

Page 43: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 43

Technical Specification FLUXUS® G70x

JBT2, JBT3

Transducers

terminal strip KL1terminal connection

1 red 2 red/blue 3 white 4 white/blue

Extension Cable (Flowmeter)

terminal strip KL2terminal connection

1 red 2 gray 3 white 4 blue

� � � � � � � � � �

� " � � " �

Page 44: Ultrasonic Gas Flowmeters for Permanent Installationdynamicflow.net/8a.pdf · 2010-04-22 · TSFLUXUS_G7V1-3-1EN_LEU 29.9.2009 1 Technical Specification FLUXUS® G70x Ultrasonic Gas

45 29.9.2009 TSFLUXUS_G7V1-3-1EN_LEU

FLUXUS® G70x Technical Specification

,

FLEXIM GmbH

Wolfener Str. 36

12681 Berlin

Germany

Tel.: +49 (30) 93 66 76 60

Fax: +49 (30) 93 66 76 80

internet: www.flexim.com

e-mail: [email protected]

Subject to change without notification. Errors excepted.

FLUXUS® is a registered trademark of FLEXIM GmbH.