Ultrasonic Flow Meter-Clamp-On type

Transcript of Ultrasonic Flow Meter Clamp-On type - Eletta · Pipe Strap Section. DC Power Relay OCT Transducer...

  • Ultrasonic Flow Meter-Clamp-On type

    ManeeshTypewritten text EUC

  • 150mm 68mm

    130m

    m

    155m

    m

    1rate

    4coolcap.

    7flowvel.

    .cal.

    5heatcap.

    8soundvel.

    0zero

    total2 3

    start/stop

    6temp

    <diag.

    9signal

    Menu

    /^ +

    Enter

    /^

    -

    ANBIENT TEMPERATURE(MAX):80℃CERTIFICATION:Ex ia II BT6ENCLOSURE:IP68

    Ultrasonic Transducer

    Part No.CP035

    FLOW 31m

    m

    54mm

    25m

    m

    24m

    m

    419AAF Series Ultrasonic Flowmeter is a state-of-the-art universal transit-time flowmeter designed using FPGA chip and low-voltage broadband pulse transmission.Comparing with other traditional flowmeter or ultrasonic flowmeter, it has distinctive features such as high precision, high reliability, high capability and low cost, the flowmeter features other advantages:TVT technology designed. Less hardware components, low voltage broadband pulse transmission, low consumption power. Clear, user-friendly menu selections make flowmeter simple and convenient to use.Daily, monthly and yearly totalized flowParallel operation of positive, negative and net flow totalizes with scale factor (span) and 7 digit display, while the output of totalize pulse and frequency output are transmitted via relay and open collector.

    Description :

    Dimensions :

    Transmitter

    Transducer

  • Flow

    Flow

    Flow

    Downstream TransducerFlow

    Pipe Strap

    Section

    DC Power Relay OCT Transducer 4-20mA RS485

    V method measuring pipe size : 25mm-400mm

    Upstream Transducer

    Downstream Transducer Pipe Strap

    Top View

    Transducer space

    Z method measuring pipe size: 100mm-800mm

    Side View

    Upstream Transducer Transducer

    Top View

    Transducer space

    Wiring Diagram:

    Transducer Installation Methods:

    Side View Section

    Transducer

  • Valve

    L≥30DL≥10D

    Reduce

    L≥10D L≥5D

    Diffuser

    L≥30DL≥5D

    90° Bend

    L≥10D

    L≥5D

    Vertical

    L10D

    L5D

    L≥50D

    OPNE:ERxA

    iTaU IIR BET(M

    6AX):80 ℃ENCLOSURE

    Ultrasonic T

    ransducer

    Part No.

    L≥10D

    Tee

    L≥50D

    Ultrasonic Transducer

    Ultrasonic Transduc

    er

    When selecting a measurement site, it is important to select an area where the fluid flow profile is fully developed to guarantee a highly accurate measurement. Use the following guidelines to select a proper installation site:1.Choose a section of pipethat is always full of liquid, such as a verticalpipe with flow in theupward direction or a fullhorizontal pipe.

    2.Ensure enough straightpipe length at least equalto the figure shown belowfor the upstream anddownstream transducersinstallation.3.Ensure that the pipesurface temperature at themeasuring point is withinthe transducertemperature limits.

    4.Consider the insidecondition of the pipecarefully. If possible,select a section of pipewhere the inside is free ofexcessive corrosion orscaling.

    Straight length of upstream piping Straight length of downstream piping

    Installation Site Selection :

  • Eletta Flow ABP.O. Box 5084SE-141 05 Kungens KurvaSweden

    Tel: +46 8 – 603 07 [email protected]

    ±0.03ft/s ±0.01m/s

    3

    74

    8

    OCT output, Relay output, RS232, 4-20mA output, RTD inputOCT output, Relay output, RS485, 4-20mA output

    OCT output, Relay output, RS485, 4-20mA output, RTD input

    Code Output mode

    Code

    CP035

    W210

    Code

    030

    xxx

    Type of transducers

    Clamp on transducer,Operating temperature:32℉~ +140℉(0℃~+60℃)

    Insertion transducer,Operating temperature:-40℉~+176℉(-40℃~+80℃)

    Transducer Cable LengthStandard 30ft (9m)

    Maximum lengthen to 305m(1000ft), per 5m is a lengthen unit.

    CodePT1000

    output, Relay output, RS485, 9m cable.

    Type of Temperature sensorPT1000 Temperature sensor

    419AAF

    Standard Model: 419AAF-4-CP035-030Description: standard flowmeter with Clamp-on transducers, OCT pulse

    DescriptionModel

    Digital Correlation Transit Time FlowmeterInstallation method:wall mountTransmitter:Flow Range: ~ ±16ft/s ( ~ ±5m/s)Accuracy: ±1.0% of measured valueRepeatability: 0.3%Pipe Size Range:1"~48" (25mm ~ 1200mm)Keyboard:16 (4×4) touch keysDisplay:20×2,alphanumeric,backlit LCDPower supply:10-36V DC@1AmaxTransmitter enclosure:IP65,ABS/PC enclosureTemperature:-20℃~50℃Output: OCT pulse output 0-10KHz, Relay output, 4-20mA optionalCommunication: RS232, Modbus ProtocolTemperature: -40℉~+140℉ (-40℃~60℃)

    Ordering Information:

    OCT output, Relay output, RS232, 4-20mA output

    EUC4