(UK 2723) V(0)a United Kingdom of Great Britain and Northern Ireland Certificate of EC type-approval of a measuring instrument Number: UK 2723 Revision 6 issued by the Secretary for Business, Innovation & Skills Notified Body Number 0126 In accordance with the requirements of the Non-automatic Weighing Instruments Regulations 2000 (SI 2000/3236) which implement, in the United Kingdom, Council Directive 2009/23/EC, this certificate of EC type-approval has been issued to: Avery Weigh-Tronix Ltd Foundry Lane Smethwick West Midlands B66 2LP United Kingdom in respect of a non-automatic weighing instrument, utilising the Avery Weigh-Tronix E1005/E1010 series indicating devices (Test certificate GB-1179) connected to a platform. n 10000 for Class III instruments with single interval n 5000 for Class III instruments with multi-interval operation, maximum two partial weighing ranges. n 1000 for Class IIII instruments with single interval. The necessary data (principal characteristics, alterations, securing, functioning etc) for identification purposes and conditions (when applicable) are set out in the descriptive annex to this certificate. This Revision replaces earlier versions of the certificate. Signatory: G. Glas for Chief Executive National Weights & Measures Laboratory (Part of National Measurement office) Department for Business, Innovation & Skills Stanton Avenue Date: 16 December 2009 Teddington Valid Until: 12 July 2014 Middlesex TW11 0JZ Reference No: T1128/0250 United Kingdom

Transcript of UK 2723 rev 6 - UK Government Web...

Page 1: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

(UK 2723)

V(0)a

United Kingdom of Great Britain and Northern Ireland

Certificate of EC type-approval of a

measuring instrument

Number: UK 2723 Revision 6 issued by the Secretary for Business, Innovation & Skills

Notified Body Number 0126

In accordance with the requirements of the Non-automatic Weighing Instruments

Regulations 2000 (SI 2000/3236) which implement, in the United Kingdom, Council

Directive 2009/23/EC, this certificate of EC type-approval has been issued to:

Avery Weigh-Tronix Ltd

Foundry Lane

Smethwick

West Midlands

B66 2LP

United Kingdom

in respect of a non-automatic weighing instrument, utilising the Avery Weigh-Tronix

E1005/E1010 series indicating devices (Test certificate GB-1179) connected to a platform.

n ≤ 10000 for Class III instruments with single interval

n ≤ 5000 for Class III instruments with multi-interval operation, maximum

two partial weighing ranges.

n ≤ 1000 for Class IIII instruments with single interval.

The necessary data (principal characteristics, alterations, securing, functioning etc) for

identification purposes and conditions (when applicable) are set out in the descriptive annex

to this certificate.

This Revision replaces earlier versions of the certificate.

Signatory: G. Glas

for Chief Executive

National Weights & Measures Laboratory

(Part of National Measurement office)

Department for Business, Innovation & Skills

Stanton Avenue

Date: 16 December 2009 Teddington

Valid Until: 12 July 2014 Middlesex TW11 0JZ

Reference No: T1128/0250 United Kingdom

Page 2: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

2

Descriptive Annex

1 INTRODUCTION

This instrument utilises the digital indicating devices designated the Avery Weigh-Tronix

E1005/E1010 (Figures 1 and 2) connected to a weighing platform to form a Class III or IIII

mains operated self-indicating non-automatic weighing instrument.

2 FUNCTIONAL DESCRIPTION

2.1 Devices

The Avery Weigh-Tronix E1005/E1010 digital weight indicators are fully described in Test

Certificate No. GB-1179.

2.2 Load receptor

The indicator can be connected to a weigh platform to form a complete weighing system

providing that the conditions of Section 2.3 are met.

Load receptor types may be chosen from Table 1 in Section 3.2.

2.3 Load cells

2.3.1 Load cells may be chosen from Table 2 in Section 3.2, or any compatible load

cell(s) may be used providing the following conditions are met:

− There is a respective OIML Certificate of Conformity (R60) or a test

certificate (EN45501) issued for the load cell by a Notified Body responsible

for type examination under Directive 2009/23/EC.

− The certificate contains the load cell types and the necessary load cell data

required for the manufacturer’s declaration of compatibility of modules

(WELMEC 2, Issue 5, 2009, No 11), and any particular installation

requirements. A load cell marked NH is allowed only if humidity testing to

EN45501 has been conducted on this load cell.

− The compatibility of the load cells and indicator is established by the

manufacturer by means of the compatibility of modules calculation, contained

in the above WELMEC 2 document, at the time of verification or declaration

of EC conformity of type.

− The load cell transmission must conform to one of the examples shown in the

WELMEC Guide 2.4, “Guide for Load cells”.

2.3.2 The load cell cable(s) may be either connected internally and permanently

fitted to the indicator, using the watertight glands, or connected externally using waterproof

bayonet connectors, which have provision for sealing, in place of the watertight glands.

Note: Not applicable to Model E1005.

Page 3: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

3

3 TECHNICAL DATA

3.1 Technical data for the indicator is provided in Test Certificate No. GB-1179.

3.2 Load receptors types and load cell list

The load receptors in the following tables are those types intended to be used with the load

cells in Table 2.

The load cells must be mounted correctly according to their mounting drawing.

Table 1

Type Load receptor No. of cells

Weighbridge All concrete platform with or without rails 4

All steel platform with or without rails 4

Steel + concrete platform with or without rails 4

All concrete platform with or without rails. Lever

system 1

All steel platform with or without rails. Lever system 1

Steel + concrete platform with or without rails. Lever

system 1

Platform scale Platform scale with corner load cells 4

“Bar” pair for pallet weighing 4 (2 per bar)

Platform scale with lever system 1

“Bar” pair for pallet weighing with two lever systems 2 per bar

“Single point” type 1

Overhead track

scale

Standard overhead track scale 1 or 2

Hopper Standard 3

Lever system + 1 load cell 1

Page 4: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

4

Table 2

Load Cell Mounting Drawing Emax Vmin nmax

Avery 8701 (1), (2)

& (4)

70712-136 or 70714-162 50000, 100000 lb Emax / 14164 5000

Avery 8707 F3936 or F3937 6 - 30 kg Emax / 3500 3000

Avery 8708 (3) K13767 or

J300000000AA00 or

H500000000A300 or

E1018-1 or CK 13743

750 - 7000 kg Emax / 9259 4000

Avery 8713 (5) F4004, B72164-107 or

H40000000AAAA

125 - 500 kg Emax / 9615 4000

Avery T101 F3938 or C82347-165 60 - 300 kg Emax / 3500 3000

Avery T103 82151-191 or 82151-192 8 - 36 kg Emax / 3500 3000

Avery T103

AVR

E1002-49.IAD 8 - 36 kg Emax / 7500 6000

Avery T109 82151-191 or 82151-192 8 - 36 kg Emax / 3500 3000

Avery T110 70714-302, 82151-196 or

ISD04019

60 - 350 kg Emax / 3500 3000

Avery T201 BRT.171194.1 90 - 1100 kg Emax / 15000 5000

Avery T202 BRT.171195.1 500 - 5000 kg Emax / 10000 3000

Avery T203 (5) H400000000AAAA,

B72164-107, WMA5332,

WMA5320 or WS3244

125 - 500 kg Emax / 12500 6000

Avery T204 K13767 or

J300000000AA00 or

H500000000A300

750 - 3000 kg Emax /15000 4000

Avery T302 ( 2), (4)

70712-136 or 70714-162 50000 or 100000

lb

Emax/15000 5000

Avery or Salter

Weigh Bar

E 1062-1.IAD 220 – 5000 kg Emax / 10000 7000

220 – 34000 kg 5000

Epel SC E1049-1.IAD 750 to 1500 kg Emax/10000 3000

Flintab RC1 4-85105 250, 400 kN Emax/7000 3000

Flintab SB4 4-85105 500 - 10000 kg Emax/11000 3000

HBM BLC, TLC,

HLC, THC

E1018-1.IAD 220 to 4400 kg Emax/10000 3000

HBM C2A 70712-136 or 70714-162 1 to 10 t Emax/10000 4000

HBM

C15AC./..T &

C15BC./..T

E1041-1.IAD 20 to 100 t Emax/10000 4000

HBM C16 AC 70712-136 or 70714-162 12 to 60 t Emax/12000 4000

HBM

PWS/PWSM

70714-302 or 82151-196 7 to 150 kg Emax / 7143 3000

Page 5: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

5

Load Cell Mounting Drawing Emax Vmin nmax

HBM PW2 82151-191 or 82151-192 7.2 kg Emax / 3600 3000

12 kg Emax / 6000 3000

Emax / 12000 3000

13 kg Emax / 6500 3000

18 kg Emax / 3600 3000

Emax / 9000 3000

36 kg Emax / 3600 3000

72 kg Emax / 3600 3000

HBM SP4 82151-191 or 82151-192 7 to 100 kg Emax / 7143 3000

HBM Z6 70714-298 or 70714-297 50 - 250 kg Emax/10000 3000

Huntleigh 355 PLP2 5 - 500 kg Emax/15000 4000

Molen CSP-M 70712-136 or 70714-162 5 to 100 t Emax/5000 1000

Emax/10000 2000

Emax/10000 3000

Emax/11666 3500

Philips

PR6222/..C3

70712-136 or 70714-162 20 to 50 t Emax/10000 3000

Revere BSP E1002-49.IAD 50 to 5000 kg Emax/10000 3000

Revere CSP M E2001-12.IAD or E2001-

13.IAD

10 - 60 t Emax / 8333 3000

Revere HPS E2001-20.IAD 6 - 60 kg Emax / 12000 3000

Revere RLC E2001-14.IAD 500 to 5000 kg Emax/7000 3000

Emax/9333 4000

Emax/11667 5000

Emax/14000 6000

Revere SHB M PLP2 20, 50, 100, 200,

500, 1000 kg

Emax/7143 3000

Revere SSB BJA060843 500 - 5000 kg Emax/10000 3000

Revere S5222

and S5223

E1002-66.IAD 20 to 45 t Emax/10000 3000

Revere 5102 6000LD/LP, 6000DT/P

or H506-6000U

50 to 10000 lb Emax/10000 2000

Emax/10000 3000

Revere 92 70712-136 or 70714-162 50000 to 200000

lb

Emax/8500 3000

Sensortronics

65023C

E1002/41.IAD 1000 - 10000 lb

500 - 5000 kg

Emax/1000 3000

Shering

SCL20SA

70712-136 or 70714-162 10 t to 100 t Emax/8333 3000

Tedea Huntleigh

1040

E1002-30 5 - 100 kg Emax/10000 3000

Tedea Huntleigh

1140

E1002-30 or LDP2 15 to 200 kg Emax/7500 3000

Tedea Huntleigh

1250

LCT2000000BA00 or

LCT3000000CA00

50 - 500 kg Emax/10000 3000

Page 6: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

6

Load Cell Mounting Drawing Emax Vmin nmax

Tedea Huntleigh

1320

E1002-47.IAD 1 to 2 t Emax/10000 3000

Tedea Huntleigh

3410

E2001-14.IAD 450 to 1800 kg Emax/10000 3000

Tedea Huntleigh

3510

PLP1 or E1055-1.1 0.5 to 5 t Emax/4000 1000

Emax/8000 3000

Emax/12000 3000

Tedea Huntleigh

HSB

PLP1 500 - 5000 kg Emax/13333 3000

Thameside T66 LDP 20 - 200 kg Emax/10000 3000

Thameside T95 FS-TS-T95 2, 5, 10 or 20 t Emax/5000 1000

Emax/10000 2000

Emax/10000 3000

Notes:

( 1 ) The weighing system must be linearised by the indicator when 8701 cells are used.

( 2 ) The Avery 8701 load cell approved under this type approval certificate is technically

compatible and freely interchangeable with the Avery T302. Any weighing

instrument verified under this approval may incorporate a mix of both the 8701 and

T302 load cells.

( 3 ) The Avery 8708 load cell approved under this type approval certificate is technically

compatible and freely interchangeable with the Avery T204 load cell.

( 4 ) The Avery 8701 and T302 load cells may be incorporated into a multi-interval

instrument, with two partial ranges, of not more than 4000 divisions per partial range.

( 5 )

The Avery 8713 load cell approved under this type approval certificate is technically

compatible and freely interchangeable with the Avery T203 load cell.

3.3 Having the following load cell:

Load Cell Mounting Drawing No. nmax

T203 B72164-107, WMA5332,

WMA5320 or WS3244 n ≤ 4000 for a multi-interval instrument

with not more than 2 ranges.

Page 7: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

7

3.4 Having an Avery beam type load cell between the transfer lever and the dead

structure of a mechanical instrument converted to use one of the types of indicator described

in this certification and having not more than 3000 digital scale intervals. Approved types are

listed below:-

Load cell Platform/Weighbridge type Drawing No

8713 Avery 8938 F4004

8708 Avery 8940 K13767

8708 Quadrelli ‘Traghetto’ E1001-01 through E1001-52

T203 Avery 8938 F4004

T109 Avery 89S E1047-1.IAD

Having the Avery 8708 load cell with not more than 4000 digital scale intervals.

Having the HBM Z3 load cell with the following characteristics:

Load cell type Emax nmax Y

Z3H2

Z3H3

10 kg – 2000 kg

10 kg – 2000 kg

2000

3000

7000

8400

Z3H2-A

Z3H3-A

50 kg – 2000 kg

50 kg – 2000 kg

2000

3000

7000

8400

3.5 Having an overhead track weighing device.

The device is constructed according to drawing number H100000000AAAA or

4150000000AAOO.

The load cells are Avery type 8713.

The device has a length between and including 300 mm and 1200 mm and a capacity of

either 300 kg by 0.1 kg or 600 kg by 0.2 kg.

3.6 Having an H205 platform using a T103 or T109 load cell, with the following

three alternative AVR (Automatic Variable Resolution) capacities:

capacity 6 kg with e = 1 g from 0 to 3 kg

and e = 2 g from 3 kg to 6 kg (T103 only)

capacity 15 kg with e = 2 g from 0 to 7.5 kg

and e = 5 g from 7.5 kg to 15 kg (T103 only)

capacity 30 kg with e = 5 g from 0 to 15 kg

and e = 10 g from 15 kg to 30 kg

The load cell mounting is shown in Drawing No 82151-191 or 82151-192.

3.7 Having Zenner barriers fitted to the excitation, sense (reference) and signal

lines of the cable, between the load cell and the indicator.

Page 8: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

8

3.8 Having the indicator connected to H5XX series load receptors comprising four

Avery 8708 load cell graded for 6000 V. The load receptor is constructed according to

drawing number H500000000A300.

The system has a capacity of between and including 1200 kg and 6000 kg and has not more

than 6000 scale intervals.

The 8708 load cell can be used for partial weighing operation having two partial ranges

where n ≤ 4000 divisions per range.

3.9 Having an Avery H305 platform, incorporating a T110 load cell, connected to

the indicator. In which case n ≤ 6000 for a single range instrument or ni ≤ 4000 for a multi-

interval instrument with not more than 2 ranges.

3.10 Having an Avery H4xx platform, connected to the indicator, having platform

construction details as shown in the following table:

Construction Drawing Load Cell nmax

E1059-1.IAD

or

E1060-1.IAD

Avery T203 n ≤ 6000 for a single range instrument.

ni ≤ 4000 for a multi-interval instrument with not

more than 2 ranges.

E1059-1.IAD Avery 8708 n ≤ 6000

3.11 Having a pallet truck weigher platform, connected to the indicator, using the

load cells described below:

Load Cell Emax Mounting Drawing vmin nmax

Tedea Huntleigh 3410 450 - 1800 kg E 2001-14.IAD Emax/10000 3000

Tedea Huntleigh 3510 0.5 - 5 t PLP1 or E1055-1.1 Emax/4000 1000

Emax/8000 2000

Emax/12000 3000

The pallet truck weigher platform will have a level indicator.

3.12 Having an H303/281 LC platform, incorporating a T110 load cell, connected

to the indicator, in the following capacities:

Max e Load cell capacity

60 kg 20 g 180 kg

150 kg 50 g 350 kg

Page 9: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

9

3.13 Having any of the range of indicators connected to platforms incorporating the

Avery T204 load cell having the following characteristics:

- for multi-interval operation with not more than two partial weighing ranges

and nmax = 4000 for each partial weighing range. Having the T204 load cells

with Emax of 750 kg, 1500 kg and 3000 kg.

- for single interval operation with nmax = 6000 and using the T204 load cell

with Emax values as listed above.

3.14 Having a mechanical platform incorporating a load cell and lever arrangement

designated Model SH1000. The platform is manufactured from painted mild steel or with a

stainless steel plate and mild steel base and internal components and can have platform

dimensions from 0.80 m x 0.80 m to 1.50 m x 2.00 m.

The platform may have the following capacities utilising either the Avery T201, T203 or

T204 load cells as shown in the following table:-

Platform dimension Capacity Load cell Alternative load cell

0.80 m x 0.80 m 300 kg

600 kg

T203 – 125 kg

T203 – 250 kg

T201 – 90 kg

T201 – 225 kg

1.00 m x 1.00 m 600 kg

1000 kg

T203 – 125 kg

T203 – 250 kg

T201 – 90 kg

T201 – 225 kg

1.00 m x 1.25 m 600 kg

1500 kg

T203 – 125 kg

T203 – 250 kg

T201 – 90 kg

T201 – 225 kg

1.25 m x 1.50 m 600 kg

1500 kg

3000 kg

T203 – 125 kg

T203 – 250 kg

T203 – 500 kg

T201 – 90 kg

T201 – 225 kg

T201 – 450 kg

1.50 m x 1.50 m 600 kg

1500 kg

3000 kg

T203 – 125 kg

T203 – 250 kg

T203 – 500 kg

T201 – 90 kg

T201 – 225 kg

T201 – 450 kg

1.50 m x 2.00 m 1500 kg

3000 kg

6000 kg

T203 – 250 kg

T203 – 500 kg

T204 - 1500

T201 – 225 kg

T201 – 450 kg

T201 – 1100 kg

3.15 Having the indicator connected to an Avery J3xx load receptor comprising

four Avery 8708 load cells graded for 6000v. The load receptor is constructed according to

drawing number J300000000AA00.

The system has a capacity of between and including 1200 kg to 12000 kg and has not more

than 6000 scale intervals.

3.16 Having the indicator connected to an Avery Berkel H4xx, H5xx or J3xx series

load receptor comprising four Revere HCB load cells graded for 6000v. The load cells may

alternatively be designated Avery T205.

The load receptors are constructed according to drawing numbers E1059-1.IAD,

H500000000A300 or J300000000AA00 respectively.

The system has a capacity of, between and including, 1200 kg and 6000 kg with not more

than 6000 scale intervals.

Page 10: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

10

3.17 Having the manufacturer designated as Berkel Prior.

3.18 Having alternative load receptors utilising the following load cells. Where a

reduced operating temperature range is specified, this shall be included in the descriptive

markings.

Manufacture

r Type Capacity

No. of scale

intervals

Max. no.

of load cells

Temperature

range

Revere 363 2,000 & 3,000 lb 3000 1 -10 to +40 °C

Revere CP1 50,000 & 100,000 lb 3000 6 0 to +30 °C

Sensortronics

60001 1,000 & 2,000 lb 3000 1 0 to +30 °C

Toledo CS Capcheck

50,000 lb & 100,000 lb 3000 6 0 to +30 °C

Solidate SB2 35,000 & 45,000 lb 3000 6 0 to +30 °C

BLH/Toledo C2PI 50,000 lb 2500 4 0 to +30 °C

HBM CISF 50,000 & 100,000 lb 2500 4 0 to +30 °C

HBM C3H2 30,000 kg 3000 6 -10 to +40 °C

4 PERIPHERAL DEVICES

4.1 The weighing system may be connected to any peripheral devices that have

been issued with a test certificate by a Notified Body responsible for type approval under

Directive 2009/23/EC in any Member State and bears the CE marking of conformity to the

relevant directives; or

Peripheral devices without a test certificate under the following conditions:

− it bears the CE marking for conformity to the EMC Directive;

− it is not capable of transmitting any data or instruction into the weighing

instrument, other than to release a printout, checking for correct data

transmission or validation;

− it prints weighing results and other data as received from the weighing

instrument without any modification or further processing;

− it complies with the applicable requirements of EN45501, i.e. 4.2, 4.4, 4.6 and

4.7.

Page 11: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

11

4.2 Data storage device (DSD)

A data storage device (DSD) having a Test Certificate (TC) may be connected to The

weighing system if at conformity assessment for putting into service for an Article 1.2(a)

application, it is checked that the requirements 6.1, 6.2, 6.4, 6.5 and 6.6 of WELMEC 2.5,

Issue 2, 2000 are met.

The DSD must have sufficient capacity to store data for a minimum of 90 days after the data

has been used for invoicing.

5 APPROVAL CONDITIONS

The certificate is issued subject to the following conditions:

5.1 Legends and inscriptions

5.1.1 The instrument shall bear the following legends near the display of the

weighing result:

Max

Min

e =

5.1.2 The instrument shall bear the following legends (Figure 3):

CE marking

Verification mark

Green M

Class

Serial number

Manufacture’s mark or name

Certificate number

5.2 The E1005 indicator may be used for “direct sales to the public” if the

“Select” function key is inactive. The E1010 indicator may be used for “direct sales to the

public” if the pre-set tare device and “Select” function key are inactive.

5.3 A levelling device sensitive enough to display a 2/1000 tilt shall be fixed

firmly to the weighing instrument in a place clearly visible to the user, unless the instrument

is freely suspended or installed in a fixed position.

6 LOCATION OF SEALS AND VERIFICATION MARKS

6.1 The CE mark, Green M and Certificate number will be mounted on the rear of

the indicator. The data plate will be mounted in such a manner that it is easily accessible and

clearly visible in its regular operating position. The CE mark shall be impossible to remove

without damaging it. The data plate shall be impossible to remove without it being destroyed.

The markings and inscriptions shall fulfil the requirements of Paragraph 1 of Annex IV of the

Directive 2009/23/EC.

Page 12: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

12

6.2 Access to the configuration and calibration facility is password protected as

detailed in GB-1179.

Alternatively, access can be via a push button located on the rear of the instrument (E1005),

which is accessed by removing the switch cover plate, or on the main PCB (E1010), which is

accessed by removing the front cover of the enclosure. The switch cover plate and the

enclosure(s) can be sealed using a wire and lead seal. N.B. only one of these methods will be

present on an instrument.

Components that may not be dismantled or adjusted by the user (e.g. load cell connection),

and load cell junction boxes where applicable, will be secured by either a wire and lead seal

or tamper evident label and securing mark. The securing mark may be either:

− a mark of the manufacturer and/or manufacturer’s representative, or

− an official mark of a verification officer.

7 ALTERNATIVES

7.1 The indicator may bear the following alternative manufacturers name and

model:

− Salter Brecknell Dp 90 (plastic) or Dp 95 (stainless steel).

7.2 The indicator may be mounted on a simple fold-away stand, a swivel mounted

stand, or an extended stand for wall mounted applications.

7.3 The Avery 8701 load cell approved under this type approval certificate is

technically compatible and freely interchangeable with the Avery T302, T302i and T302x

load cells. Any weighing instrument verified under this approval may incorporate a mix of

8701, T302, T302i and T302x load cells.

8 ILLUSTRATIONS

Figure 1 E1005 indicator general view

Figure 2 E1010 indicator general view

Figure 3 Data plate

Page 13: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

13

9 CERTIFICATE HISTORY

ISSUE NO. DATE DESCRIPTION

UK 2723 13 July 2004 Type approval first issued.

UK 2723 Revision 1 1 September 2004 Revision 1 issued. Increase in the number of

scale intervals to 10000.

UK 2723 Revision 2 25 January 2005 Section 2.3.2 added.

Load cell T302 added to Note (4) at

Section 3.2.

UK 2723 Revision 3 11 May 2005 Partial weighing ranges reduced from 3 to 2

and the maximum number of verification

scale intervals per partial weighing range

increased from 3000 to 5000

UK 2723 Revision 4 17 October 2005 Minor modifications to Table 2.

Note (5) added to Section 3.2.

Sections 3.4 to 3.18 and 4.2 added.

UK 2723 Revision 5 19 April 2006 3 ranges corrected to 2 ranges in Sections 3.3,

3.10 & 3.13.

Partial ranges increased to 4000 divisions for

Avery load cells: 8701, 8708, T110, T202,

T303 & T204.

6 kg crossover capacity changed to 7.5 kg for

15 kg cell in Section 3.6.

UK 2723 Revision 6 16 December 2009 Alternative 7.3 added.

Page 14: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

14

Figure 1 E1005 indicator general view

Page 15: UK 2723 rev 6 - UK Government Web Archivewebarchive.nationalarchives.gov.uk/20100113205514/nmo.bis.gov.uk... · in respect of a non-automatic weighing instrument, utilising the Avery

15

Figure 2 E1010 indicator general view

Figure 3 Data plate

Crown Copyright 2009

NATIONAL WEIGHTS AND MEASURES LABORATORY

(Part of National Measurement Office)

Department for Business, Innovation & Skills