UCAM-550A · MASTER (When master, lit green, when slave, not lit) TRANSFER (When communications,...

4
3-33 3 -33 Data Loggers MEASURING INSTRUMENTS UCAM-550A Fast Data Logger UCAM-550A Synchronous sampling at 50 Hz of all channels UCAM-550A is a fast data logger that repeatedly measures a maximum of 1000 channels at an interval of 0.02 s. Because this is capable of high-speed synchronous measurement, this unit measures a wide range of phenomena, from static to dynamic phenomena. The following 5 types of measuring units are provided. Strain Unit USS-51B (Potentiometer-type sensor also supported) Voltage Unit USV-51B Thermocouple Unit UST-51B Strain/Voltage/Thermocouple Unit USM-51B, USM-52B They support strain gages, strain-gage transducers, voltage output sensors, potentiometer-type sensors, and thermocouples, measure and collect strain and stress, load, pressure, and displacement, as well as voltage and temperature. Measuring channels is for 1 unit a maximum of 50 channels, and with 20 units cascaded, a maximum of 1000 channels, and this is suited from small-scale to large-scale measurement. Synchronous* sampling of all channels Synchronous measurement of 1000 channels at max. 50/s Synchronous measurement of up to 20 units possible using a LAN cable Control using Dynamic Data Acquisition Software DCS-100A 5 types of measuring units available * Except temperature measurement using USM-51B or USM-52B PC is not included. To Ensure Safe Usage DCS-100A of a standard accessory can be measured up to 300 channels. Measurement up to 1000 channels requires an optional software DCS-106A. See page 4-6. *Requires UCAM-550A firmware version 03.00 or latter. Measuring units Measuring targets USS-51B USM-51B/52B* USV-51B UST-51B 120 Ω 350 Ω Active-dummy Active-active Active opposite-leg Full bridge 1 to 10 kΩ ±20 V K T E J R N Quarter bridge Half bridge 120 to 1 k Ω Full bridge 120 to 1 k Ω Strain gages Strain-gage transducers Potentiometer-type sensors Voltage Temperature Thermocouples Yes Yes Yes Yes Yes Yes* Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Measuring Targets and Measuring Unit

Transcript of UCAM-550A · MASTER (When master, lit green, when slave, not lit) TRANSFER (When communications,...

Page 1: UCAM-550A · MASTER (When master, lit green, when slave, not lit) TRANSFER (When communications, flashing green) Operation Keys UP, Down, Left, Right Data Storage Measurement data

3-33

3-33

Dat

a Lo

gg

ers

MEA

SUR

ING

INST

RU

MEN

TS

UCAM-550AFast Data Logger

UCAM-550A

Synchronous sampling at 50 Hz of all channels

UCAM-550A is a fast data logger that repeatedly measures a maximum of 1000 channels at an interval of 0.02 s. Because this is capable of high-speed synchronous measurement, this unit measures a wide range of phenomena, from static to dynamic phenomena. The following 5 types of measuring units are provided. ●Strain Unit USS-51B (Potentiometer-type sensor

also supported)●Voltage Unit USV-51B●Thermocouple Unit UST-51B●Strain/Voltage/Thermocouple Unit USM-51B,

USM-52BThey support strain gages, strain-gage transducers, voltage output sensors, potentiometer-type sensors, and thermocouples, measure and collect strain and stress, load, pressure, and displacement, as well as voltage and temperature. Measuring channels is for 1 unit a maximum of 50 channels, and with 20 units cascaded, a maximum of 1000 channels, and this is suited from small-scale to large-scale measurement.

●Synchronous* sampling of all channels●Synchronous measurement of 1000 channels at

max. 50/s●Synchronous measurement of up to 20 units

possible using a LAN cable●Control using Dynamic Data Acquisition Software

DCS-100A●5 types of measuring units available* Except temperature measurement using USM-51B or USM-52B

PC is not included.

To Ensure Safe UsageDCS-100A of a standard accessory can be measured up to 300 channels. Measurement up to 1000 channels requires an optional software DCS-106A. See page 4-6.

*Requires UCAM-550A firmware version 03.00 or latter.

Measuring unitsMeasuring targets

USS-51BUSM-51B/52B* USV-51B UST-51B

120 Ω350 ΩActive-dummyActive-activeActive opposite-legFull bridge1 to 10 kΩ±20 VKTEJRN

Quarter bridge

Half bridge 120 to 1 k Ω

Full bridge 120 to 1 k Ω

Strain gages

Strain-gagetransducers

Potentiometer-type sensorsVoltage

Temperature Thermocouples

YesYesYesYesYes

Yes*

Yes

YesYesYesYesYesYesYes

YesYesYesYesYesYesYesYesYesYesYesYesYesYes

Measuring Targets and Measuring Unit

Page 2: UCAM-550A · MASTER (When master, lit green, when slave, not lit) TRANSFER (When communications, flashing green) Operation Keys UP, Down, Left, Right Data Storage Measurement data

3-34

3-34

Dat

a Lo

gg

ers

MEA

SUR

ING

INST

RU

MEN

TS

* Accuracy of the Internal Reference-junction CompensatorWithin ±1.0 °C (When temperature balanced at input terminals) (The ambient temperature is 25 ±10 °C)Within ±2.0 °C (When temperature balanced at input terminals) (The ambient temperature is other than mentioned above.)

±(0.3% of reading + 0.8 ºC)±(0.2% of reading + 0.6 ºC)±(0.3% of reading + 0.8 ºC)±(0.2% of reading + 0.6 ºC)±(0.3% of reading + 0.8 ºC)±(0.2% of reading + 0.6 ºC)±(0.3% of reading + 0.8 ºC)±(0.2% of reading + 0.6 ºC)±(0.6% of reading + 1.2 ºC)±(0.5% of reading + 1.0 ºC)±(0.3% of reading + 0.8 ºC)±(0.2% of reading + 0.6 ºC)

Types Range Accuracy* (Resolution: 0.1 ºC )

-200.0 to 1200.0 ºC

-200.0 to 350.0 ºC

-200.0 to 800.0 ºC

-200.0 to 750.0 ºC

0.0 to 1600.0 ºC

-200.0 to 1250.0 ºC

-200.0 to below -100.0 ºC-100.0 to 1200.0 ºC

-200.0 to below -100.0 ºC -100.0 to 350.0 ºC

-200.0 to below -100.0 ºC-100.0 to 800.0 ºC

-200.0 to below -100.0 ºC-100.0 to 750.0 ºC

0.0 to below 100.0 ºC100.0 to 1600.0 ºC

-200.0 to below -100.0 ºC-100.0 to 1250.0 ºC

K

T

E

N

ModeTargets ResolutionRange Accuracy

0 to ±19 k ×10-6 strain0 to ±300 k ×10-6 strain

-50% to 50%-20 to 20 V

1 ×10-6 strain10 ×10-6 strain

0.01%1 mV

±0.08%FS

±0.1%FS±0.08%FS

Strain LH

PotentiometersVoltage

■Strain/Voltage/Thermocouple Unit USM-51B/USM-52BInput Terminals USM-51B: NDIS connectors, and screw-soldering terminal blocks USM-52B: NDIS connectors, and one-touch terminal blocksChannels 10Measuring Targets Strain gages, strain-gage transducers, potentiometer-type sensors, voltage, and thermocouplesBridge Excitation 2 VDCPower Supply to Sensors 2 VDC, for potentiometer-type sensorsGage Factors 2.00 fixedFrequency Response DC to 7.8 Hz, deviation: 0.5, -3.5dB (Except temperature measurement)Burn-out Check Performing burn-out when checkingTEDS Reads information from TEDS-installed sensor.Strain, Potentiometers, and Voltage

Dedicated Optional Accessories

Thermocouples

AC power cable P-18 (With a 2-pin conversion plug CM-39), ground wire P-72, DVD (DCS-100A, instruction manual)

Standard Accessories

Measuring Functions Original value measurement Measure value measurementInterfaces 10 BASE-T, 100BASE-TX Between PC and UCAM  LAN cable (Straight) Max. 100 m Between UCAM and UCAM  STP straight cable (See notes) Max. 100 m Note: "STP" is the initials of Shield Twisted Pair, and an STP cable is a shielded LAN cableDisplay LCD (20 digits x 2 lines) Status display LED: POWER (When power ON, lit green) MASTER (When master, lit green, when slave, not lit) TRANSFER (When communications, flashing green)Operation Keys UP, Down, Left, RightData Storage Measurement data is saved on a PC (No internal storage)Operating Temperature 0 to 40°COperating Humidity 20 to 85% RH (Non-condensing) Power Supply 100 to 240 VAC Approx. 50 VA (With 5 USS-51B strain units installed, and 120 Ω load on all channels connected)Dimensions 426 W × 132.5 H ×305 D mm (Excluding protrusions)Weight Approx. 7 kg (With 5 USS-51B strain units installed)

Specifications

0.1

Amplitude (dB)

0.2 0.5 1 2 5 10

0−1

−2

−3

−4

−5

−6

Frequencies (Hz)

■UCAM-550AModels UCAM-550A With DCS-100A UCAM-550A-0 Without DCS-100AChannels

Maximum of 50 channels/unit (Possible up to 5 units of the measuring unit)(Each measuring unit measures 10 channels.)Measurement is possible up to 1000 channels at maximum by adding an optional software DCS-106A.*The public command corresponds up to 20 units (Max. 1000 channels).*DCS-100A corresponds up to 6 units (Max. 300 channels).

Sampling Method Synchronous sampling of all channelsSampling Frequencies 1, 2, 10, 20, and 50 Hz*Response frequency depends on the measuring unit. USM-51B/52B*, USS-51B, USV-51B, UST-51B: DC to 7.8 Hz Deviation:0.5 to -3.5 dB*For temperature measurement with USM-51B/52B using scanning mode, the updating rate is approx. 1 s.

Note: Measuring range is indicated when the initial measurement and the original value measurement are performed. In the case of a measure value measurement, the value of the initial measurement is subtracted in advance from the original measurement value.

ModeTarget ResolutionMeasuring Range Accuracy

0 to ±19 k ×10-6 strain0 to ±200 k ×10-6 strain

0 to ±50%

1 ×10-6 strain10 ×10-6 strain

0.01%

±0.05% FS

±0.1% FS

Strain LH

Potentiometers

Terminal cover UT-50A

Terminal cover UM-51B

Optional Accessories

Standard Accessories

■Strain Unit USS-51BChannels 10Measuring Targets Strain gage, strain-gage transducers, potentiometer-type sensorsBridge Excitation 2 VDC constant voltage (Applied constantly)Power Supply to Sensors 2 VDC constant voltage (Applied constantly)Gage Factor 2.00 fixed (Correction is possible at 2.00/Ks with the engineering value conversion function)Measuring Range, Resolution, Accuracy (In static (DC) Inputting)

Measuring Range AccuracyResolution Signal Source Resistance

1 mV ±0.05% FS 50 Ω or less0 to ±20.000 V

■Voltage Unit USV-51BChannels 10Measuring Targets DC voltage, voltage output type sensorsMeasuring Range, Resolution, Accuracy (In static (DC) Inputting)

Terminal cover UT-50AStandard Accessories

* Accuracy of the Internal Reference-junction Compensator, when temperature balanced at input terminals, and the ambient temperature is 25 ±10 ºC.Type K, T, E, J, and R: Within ±0.5 ºC Type N: Within ±1.0 ºCNote: Accuracy does not include internal standard connection accuracy.

Switching between internal and external standard connect compensators is possible. Thermocouple resistance 300 Ω or less (K type)

±0.8°C±2.8°C±0.7°C±0.5°C±1.7°C±0.6°C±2.0°C±2.2°C±(0.4% of reading + 1.0 ºC)±(0.3% of reading + 0.8 ºC) ±(0.4% of reading + 1.2 ºC) ±(0.3% of reading + 1.0 ºC)

Types Measuring Range Accuracy

-200.0 to 437.0°C-200.0 to1200.0°C-200.0 to 350.0°C-200.0 to 260.0°C-200.0 to 800.0°C

0 to 330.0°C0 to 750.0°C

0 to 1600.0°C-200.0 to below -100.0 °C

-100 to 530.0 °C-200.0 to below -100.0 °C

-100 to 1250.0 °C

K

T

E

N

LH

LHLH

Terminal cover UT-50AStandard Accessories

■Thermocouple Unit UST-51BChannels 10Measuring Targets Temperature (Thermocouples)Measuring Range, Resolution, Accuracy (In static (DC) Inputting)

Page 3: UCAM-550A · MASTER (When master, lit green, when slave, not lit) TRANSFER (When communications, flashing green) Operation Keys UP, Down, Left, Right Data Storage Measurement data

3-35

3-35

Dat

a Lo

gg

ers

MEA

SUR

ING

INST

RU

MEN

TS

DCS-100A software for UCAM-550A section*For details of DCS-100A, see page 4-3.

■Dimensions

(11.5)426

(18)

132.5 (15)

305

Controllable Units Max. 6 (Max. 300 channels) Max. 20 (Max.1000 channels), optional software DCS-106 is required.Interfaces LANData Storage The measured data is stored in the PC hard disk as a KS2 file.Sampling Frequencies 1, 2, 10, 20, and 50 HzMeasuring Modes Manual, manual (Data points preset), interval, and analog triggerMeasuring Function Measure, original Measure: Measured value = Sensor output value - Initial value Original: Measured value = Sensor output valueCalibration Factor Calculation ON/OFF setting in all channels of one batch Calibration factor compensation: Measured value × Calibration factor + OffsetChannel Conditions Measurement, mode, range, calibration factor, offset, unit, initial value, CH name, measuring range, decimal digits, upper limit, lower limit, rated capacity, rated output (Selection of any display item is possible.)Initial Value Measurement Measures the initial value of each sensor.Manual Measurement Measurement is made from a press of the REC button to a press of the STOP button or to completion of recording to the data points preset.Interval Measurement Measurement is made automatically at preset intervals from the preset starting time. Analog Trigger Measurement Start and/or stop recording based on specified trigger conditions.

Analog Trigger ConditionsEnd Trigger SettableDelay Both start and end max. 3000 points/channel Trigger Channels Any 1 channelTrigger Level Sets in physical quantity. Trigger Slope Up, down

Changing Stroke Changes the data, before the stroke and after the stroke, when using a displacement transducer. Static Measurement Every time the DCS-100A starts recording data, the DCS-100A additionally saves the moving- averaged measured data in a single CSV format file in manual and interval modes.Burnout Check For USM-51B/52B onlyTEDS Reads sensor's information and sets to channel condition automatically. (USM-51B/52B only) Setting and Loading Parameters Sets and loads the UCAM-550A internal parameters.Environmental Settings Hardware Configuration Setting of connected units, device name, setting for IP address Reading hardware configuration from the UCAM-550A is possible.Communication Status Checked by reading the version of the UCAM

■Connection Cable U-17 to 20 (See page 8-5.)

■Isolation Transformer UPT-300BThis is used to obtain good measurement results under bad power supply conditions (Strong noise, etc.).

■One-touch Terminal Block JT-1AA terminal block that supports one-touch connection of input lead wires, and is used for attaching input terminals. 1 for each lead wire(Sale units: 10).

■Dummy Panel UD-50ACovers the slots of a UCAM-550A that does not have a measuring unit installed.

Page 4: UCAM-550A · MASTER (When master, lit green, when slave, not lit) TRANSFER (When communications, flashing green) Operation Keys UP, Down, Left, Right Data Storage Measurement data

3-36

3-36

Dat

a Lo

gg

ers

MEA

SUR

ING

INST

RU

MEN

TS

Dynamic DataAcquisition Software

DCS-100A

PCFast data logger UCAM-550AConnection of a maximum 20 units, and measurement of 1000 channels possible. With the DCS-100A, measurement with up to 6 units and 300 channels is possible.With the DCS-106A, measurement with upto 20 units and 1000 channels is possible.

LAN cable (Straight)PC-to-UCAM max. 100 m *

If using only a single UCAM-550A, directly connect a LAN cable.

STP straight cable **Max. between devices 100 m *

* Please consult with us if adherence to international standards regarding electromagnetic compatibility is required. ** The STP cable is a shielded LAN cable.

STP straight cable **

Max. between devices 100 m *

Optional softwareDCS-106A

(Capable of controllingup to 1000 channels)

When operating multiple devices synchronously, use a cascade connection between UCAM-550A with a STP straight cable **. No hub is required

Strain gagesStrain-gage transducersPotentiometers

Strain unit Voltage unit Thermocouple unit

Voltage-output type transducers Thermocouples

■System Configurations

■All channels synchronously sampling*

CH000

CH001

CH999

……

Date update rate

Scanning speed

The measurement channel is switched one byone and measured.

→ There is a di erence at the measurement time of each measurement channel, therefore a dynamic measurement is nonconformity.

All channels are synchronously measured(No scanning)

→ All channels are synchronously sampled and the data is updated every constant intervals. Therefore, a dynamic measurement is possible.

CH000

CH001

CH999

……

Date update rate

Scanning methods Synchronous sampling of all channels

* Temperature measurement using USM-51B or USM-52B is scanning mode.

UCAM-550ARecommended

products for combination

→ 4-9

Data Analysis Software DAS-200A