Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9 th , 2010 ANALYST TUTORIAL How to operate the API 2000 for a single Mass Spectrum measurement? Is the API 2000 free at a desired slot of time? Booking of the instrument. 1. Sample LOGIN. 2. Instrument standby ready with suitable Method. 3. Background observation at a desired m/z range. 4. DILUTED Sample injection. 5. CLEAN-UP. 6. Data observation and saving. 7. Instrument ready standby .

Transcript of Tutorial Analyst for API 2000 operators -...

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

ANALYST TUTORIAL

How to operate the API 2000 for a single Mass Spectrum

measurement?

Is the API 2000 free at a desired slot of time?

Booking of the instrument.

1. Sample LOGIN.

2. Instrument standby � ready with suitable Method.

3. Background observation at a desired m/z range.

4. DILUTED Sample injection.

5. CLEAN-UP.

6. Data observation and saving.

7. Instrument ready � standby.

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

Analyst Tutorial

Did you book the instrument?

Is the instrument free to use?

If yes, register your sample in the LOGBOOK and start the application “Analyst”.

Then the following text dialog box appears:

1- Click once on “Configure”.

and

2- Double-click on “Hardware Configuration”.

1- Select “MassSpec & Syringe”.

and

2- Click on “Activate Profile”.

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

A “green flag” appears in front of the activated hardware profile chosen, as shown below.

Click on “Close”.

1- Double-click on “Manual Tuning”.

The following dialog box appears

And gas turns on

Press “OK”.

2- Double-click again on “Manual Tuning”.

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

The Tune Method Editor page appears, as shown below.

A new Tune Method Editor page appears, as shown below.

1- Ctrl+O.

Open a File dialog box appears

2- Select Q1MS Pos

or Q1MS Neg method.

3- Press OK.

4- Press No (about saving

method).

The syringe is already loaded with clean-up solvent.

3- Click on “ Start ” (ACQ).

1- Enter target (s) mass ; with 100 of width .

2- Click on “Start syringe pump”.

4- Observation of mass spectrum background at the desired m/z range.

5- Stop ACQ and Syringe pump.

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

3- Click on “Start” (ACQ).

2- Click on “Start syringe pump”.

4- After 20 sec , “Stop syringe pump”

and Reload FULLY with suitable clean-up solvent.

1- Bin the clean-up solvent and half load the syringe with

DILUTED SAMPLE (10-20µg.mL-1

) = (1mg. mL-1

diluted 50-100x).

5- Then “Start syringe pump” again.

6- Stop Syringe pump and ACQ, after observation of ions of interest and clean-up process .

10-20 µg.mL-1

<=> 1-2 ng. µL-1

Example: 10µL of sample solution (1mg. mL-1

) + 740µL of suitable

solvent with 0.1% formic acid.

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

At the end of the run, this workspace appears:

���� HOW TO EXPLORE YOUR DATAS

Mass spectrum pane .

TIC (Total Ionic Current) pane.

Right-Click on “MS pane”.

Select “Open file”.

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

Select MS pane by one left-click in MS pane (blue frame).

Click on for closing MS pane.

- Select the part of TIC with the mouse left-button.

- Then right-click once in the area chosen.

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

Then we obtain the following workspace :

Select “show spectrum”.

An average mass spectrum covering the TIC area chosen appears.

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

we obtain this workspace for DATA EXPLORING.

Save a mass spectrum:

Print-Screen

Open a WORD or PPT new file and paste (+ Crop picture out)

Save the WORD or PPT document in the following link:

C:\QTRAP 2010\Name_of_supervisor\MS requester\name_of_datafile

1- From Tune Editor page, double-click on

2- Browse.

3- Select file of interest among the last files recorded (sorted by date and time) C:\analyst data\Project\API instrument\Tuning cache

Analyst Tutorial Author: Dr. J Bernard Jean-Denis Jun 9th

, 2010

Closing down my session:

- Close panes and windows (WITHOUT saving METHOD CHANGES)

1- Click once on “Configure”.

and

2- Double-click on “Hardware Configuration”.

1- Select “MassSpec & Syringe”.

2- Click on “Deactivate Profile”.

3- Close

and

Close the entire application.

Shut down the monitor.

Leave the computer on.