TUNABLE DIODE LASER GAS ANALYZERSH2S / CO2 / MOISTURE ANALYSIS
SERIES 6020L & 8800L
[email protected]://www.teledyne-ai.com2
Our Advantage
Wavelength Modulation Spectroscopy(WMS)WMS is a method used to significantly increase the
sensitivity by reducing the low-frequency noise. The laser
light’s intensity is modulated at a high frequency to create a
periodic signal to the detector. The signal is then encoded
and measured at a high frequency, eliminating the effect of
low frequency noise.
Zero BaselineTDLAS technology uses non-absorbing adjacent
wavelengths as a zero point. This is an easier analysis than
detecting a small change in a large signal.
Permanent Wavelength TrackingThe absorption of wavelengths is measured in reference to
two other absorption points to compensate for common
frequency shift. This results in excellent long-term
repeatability to minimize analyzer verifications.
Not resorting to differential measurements means no
switching valves, less hassle disposing of scrubber media,
lower cost of ownership, uncertainty of the CO2, CH
4 and
C2H
4 interferences are zeroed and a much faster respon-
se/reading time.
Direct Measurements
Sample SystemThis has 316 stainless steel tubing and fittings with a
GenieTM filter along with flow control and precision
sample cell pressure regulation.sample cell pressure regulation.
Herriott Sample Cell
Multi-pass cell with an effective path length of ~33 feet
(~10 meters). The sample cell is heated and controlled at
± 50°C 1°C by a wrap around heater.
Output
Source
Product FeaturesFully certified for hazardous area
Precision temperature control of the sample cell
316 stainless steel sample system with GenieTM filter and pressure regulation
Display: IDEC 5.7-inch touchscreen
I/O: 1 x 4-20 mA, 2 x alarm contacts TIA-485/232, USB2, Ethernet (supports MODBUS)
Teledyne’s 6020L and 8800L series of analyzers leverage the latest in Tunable Diode Laser Spectroscopy (TDLAS) to achieve
ultra-low, fast and reliable analysis of hydrogen sulfide (H2S), carbon dioxide (CO
2) and moisture (H
2Ov). By tuning each laser
to a specific absorption wavelength, based on the gas of interest, the attenuation of the light through a sample cell and the
detector can be correlated to the concentration of each gas, per Beer-Lambert Law.
[email protected]://www.teledyne-ai.com3
SPECIFICATIONSModel 6020L Model 6020L Model 8800L
Parameter
Typical Ranges
Certifications
Laser Specification
Repeatability
Linearity
T90 Response Time
Temperature
Ingress Protection
Inlet Pressure
Wetted Materials
Display
Outputs
Power
Dimensions
Inlet Flow
H2S CO
2 Moisture (H2Ov)
0-10 ppm to 0-20% 0-100 ppm to 0-100% 0-5 ppm to 0-2,500ppm
North America Version
CSA Class I Div 1 Groups B,C,D T4 (T3 with opt. freeze-proof heater)
CSA Class I Div 2 Groups B,C,D T4 (T3 with opt. freeze-proof heater)
International Version
ATEX Zone 2 IIB+H2 T4
Class 1
± 2% FS
± 1% FS
4 seconds
NEMA-4/7 explosion-proof
2-4 SCFH (1-2 SLPM)
Tubing/Fittings: 316 stainless steel
Sample Cell: Nickel-plated aluminum
IDEC 5.7-inch touchscreen
22 psig (1.5 bar G) nominal
Anolog Outputs: 1 x 4-20 mADC (isolated)
Discrete Outputs: 2 x alarm contacts, concentration and valid data
Communications: TIA-232 , TIA-485, RJ-45 Ethernet (supports MODBUS)
100-240 VAC 47-63 Hz, 100W
(Optional) 24VDC, 100W
Approximately 25” W x 52” H x 12” D
(635 x 1,320 x 305 mm)
+4 to +140°F (-20 to + 60°C) for most applications
Rev-A 1/17/2019
[email protected] w w.teledyne-ai .com
[email protected] w w.teledyne-ml.com
Middle East OfficeSharjah
UNITED ARAB EMIRATES (UAE)TEL: +971.6557.9727
California OfficeCity of industry, CA
United states of america (USA)TEL: +1 626.934.1500
Southeast Asia Officepetaling jaya
MALAYSIATEL: +603.7805.7712
colorado officeEnglewood, co
United states of america (USA)TEL: +1 303.792.3300