Troubleshooting and Correcting Sensor Problems

4
1 1 Troubleshooting and correcting sensor problems The Calibrate branch of the command tree contains a set of consequent displays that can be used to troubleshoot sensor problems. Each display contains some important values than can give you clues to where the problem might be. 1.1 LED value Check the LED value that tells you how the light source in the sensor is doing. The LED value is given in the Optical image display, available through the commands Calibrate / Optical image. A B C D RAW SENSOR DATA RMN: 2 RMX: 216 LED: 96 Scans: 2 A/D: 314 Sts: 00h HT: 143 HH: 10 Scaled image Normal operation Figure 1.1 Checking the LED value A good LED value is in the range 20-150. The most common reason to a LED value higher than 150 is that prism is coated. It is also possible the prism has been destroyed with an overactive steam wash. Check the wash times and (especially if the wash time exceeds 3s) the prism surface. Replace the prism if the surface is damaged. If the process is very hot, it is also possible that the light source LED wears out after many years’ constant use and needs replacing. Also leakage into the sensor can destroy LEDs (and everything else inside the sensor housing). However, normally the LEDs last as long as the sensor and high LED values indicate a prism problem. It is very unusual for the LED value to be lower than 20, but in theory this can be caused by light from outside reaching the prism e.g. because the pipe is translucent. 1.2 Sensor temperature and humidity Checking the conditions inside the sensor head will give you an idea whether there’s is a real physical problem. The Scaled image (Calibrate / Optical image / Scaled image) shows you both the internal sensor temperature and humidity. The HT value gives you the sensor temperature and should be below 60 C. If the value is over this limit, the measurement may not be accurate and the sensor needs to be cooled more effectively. The HH value gives the humidity level inside the sensor and should be lower than 10 %. Most commonly a higher humidity level indicates a leak in the sensor. If the sensor is subjected to big temperature changes,

Transcript of Troubleshooting and Correcting Sensor Problems

1

1 Troubleshooting and correcting sensor problems

The Calibrate branch of the command tree contains a set of consequent displays that can be used totroubleshoot sensor problems. Each display contains some important values than can give you clues towhere the problem might be.

1.1 LED value

Check the LED value that tells you how the light source in the sensor is doing. The LED value is given inthe Optical image display, available through the commands Calibrate / Optical image.

A B C D

RAWSENSORDATA

RMN: 2RMX: 216LED: 96Scans: 2A/D: 314Sts: 00hHT: 143HH: 10

Scaledimage

Normal operation

Figure 1.1 Checking the LED value

A good LED value is in the range 20-150. The most common reason to a LED value higher than 150 isthat prism is coated. It is also possible the prism has been destroyed with an overactive steam wash. Checkthe wash times and (especially if the wash time exceeds 3 s) the prism surface. Replace the prism if thesurface is damaged.

If the process is very hot, it is also possible that the light source LED wears out after many years’ constantuse and needs replacing. Also leakage into the sensor can destroy LEDs (and everything else inside thesensor housing). However, normally the LEDs last as long as the sensor and high LED values indicate aprism problem.

It is very unusual for the LED value to be lower than 20, but in theory this can be caused by light fromoutside reaching the prism e.g. because the pipe is translucent.

1.2 Sensor temperature and humidity

Checking the conditions inside the sensor head will give you an idea whether there’s is a real physicalproblem. The Scaled image (Calibrate / Optical image / Scaled image) shows you boththe internal sensor temperature and humidity.

The HT value gives you the sensor temperature and should be below 60 ◦C. If the value is over this limit,the measurement may not be accurate and the sensor needs to be cooled more effectively.

The HH value gives the humidity level inside the sensor and should be lower than 10 %. Most commonlya higher humidity level indicates a leak in the sensor. If the sensor is subjected to big temperature changes,

2

A B C D

SCALEDOPTICALIMAGE

TEST: 115.951.2 °CEndp: 21L(: 9.8R): 25.9HT: 20 °CHH: 3%

SlopeNormal operation

Figure 1.2 Scaled image

condensate can cause humidity problems. If this is the case, the problem can be solved by replacing thedryer unit inside the sensor.

1.3 Slope value

The Slope display (Calibrate / Optical image / Scaled image / Slope) shows youthe slope image and also gives the Slope value as a number. A good slope value is above 1.5. If the slopevalue drops near 1.0, something is wrong with the prism:

− Prism is coated or− Prism surface is not good or− Prism gaskets leak and process liquid gets in by the prism sides

A B C D

Diagnos-tics

SLOPE

TEST: 115.731.2 °CMax: 208Slope: 2.7Endp: 21L): 10.1R(: 25.8

Normal operation

Figure 1.3 Slope display

Prism coating can be prevented by good process flow, high process temperature or prism wash. If yoursystem already has prism wash, the wash interval can be decreased.Important: The maximum wash times are given in Table 1.1. Do not use a longer continuous wash time,instead you can try washing more often.

3

Wash type Time Pressure Recovery Interval

Steam wash 3-5 sec. 5-6 bar (psi) 20-30 s 20-30 minHigh pressure water wash 10-15 sec. 20-30 bar (psi) 20-30 s 10-20 min

Table 1.1 Recommended prism wash times for the PR-03 sensors

If prism surface is damaged, for example something in the process has abraded the prism or the wash timehas been long, replace the prism and adjust the prism wash or other conditions so that the prism doesn’t getdamaged.

If prism gaskets leak, replace them and all the sensor parts that have been damaged by the process liquidinside the sensor.

1.4 Image diagnostics

The last tool in this branch of the selection tree is the Image diagnostics display, available through Cali-brate / Optical image / Scaled image / Slope / Diagnostics. If the answer toall the questions in this screen is Yes (Figure 1.4), the sensor head temperature and humidity are below thelimits and the message is Normal operation, your refractometer system should be in working order.

Note: If the system is offline, some of the image diagnostics questions will get No for an answer even in aworking system (Figure 1.4).

A B C D

IMAGE DIAGNOSTICS

Max intensity OK (above 100)?Endpoint below 75%?Image below corner?Slope OK (above 1)?Left curve L) OK (above 1)?Right curve R) OK (above 1)?

YesYesYesYesYesYes

SENSOR HEADHead temperature: 28 °CHead humidity: 3%

Normal operation.

Inline image diagnostics (= process).

A B C DA B C D

IMAGE DIAGNOSTICS

Max intensity OK (above 100)?Endpoint below 75%?Image below corner?Slope OK (above 1)?Left curve L) OK (above 1)?Right curve R) OK (above 1)?

YesYesNoYesNoNo

SENSOR HEADHead temperature: 28 °CHead humidity: 3%

Normal operation.

Offline image diagnostics (= in air).

Figure 1.4 Image diagnostics display

4

IMAGE DIAGNOSTICS

Max intensity OK (above 100)?Endpoint below 75%?Image below corner?Slope OK (above 1)?Left curve L) OK (above 1)?Right curve R) OK (above 1)?

YesYesYesYesYesYes

SENSOR HEADHead temperature: 28 °CHead humidity: 3%

Normal operation.

A B C D

CONC 68.0%PROCESS TEMPERATURE: 31.2 °CSTANDARD RI(25°C): 1.4194

TEST: 115.7Normal operation

Calibrate Display

A B C D

A B C D

CONC

PROCESS TEMPERATURE: 32.5 °CSTANDARD RI(25°C): 1.4190

TEST: 117.2Normal operation

Opticalimage

Para-meters

68.0%

A B C D

RAWSENSORDATA

RMN: 2RMX: 216LED: 96Scans: 2A/D: 314Sts: 00hHT: 143HH: 10

Scaledimage

Normal operation

A B C D

SCALEDOPTICALIMAGE

TEST: 115.951.2 °CEndp: 21L(: 9.8R): 25.9HT: 20 °CHH: 3%

SlopeNormal operation

Diagnos-tics

SLOPE

TEST: 115.731.2 °CMax: 208Slope: 2.7Endp: 21L): 10.1R(: 25.8

Normal operation

A B C D

Good LED value 20-150

Good slope valuemore than 1.5

Good HT lower than 60 ºCGood HH lower than 10 %

Figure 1.5 Troubleshooting your refractometer system with IT-R