Transferring and Scaling Methods among a Variety of … · 2016-09-11 · F=F Superficially porous...
Transcript of Transferring and Scaling Methods among a Variety of … · 2016-09-11 · F=F Superficially porous...
Superficially porous particles offer improved
efficiency and performance over similarly sized
traditional totally porous particles. Higher efficiency
leads to improved resolution and possible time
savings with superficially porous particles, hence
their growing popularity for HPLC analyses. Columns
using superficially porous particles are currently
available in a wide variety of particle sizes, pore sizes
and stationary phase chemistries to meet most
analysts’ needs. This work will address method
transfer and scalability from totally porous particles,
as well as among different varieties of superficially
porous particles and when to use which particle and
column configurations. Overcoming common barriers
of method transfer with high efficiency columns,
including instrument configuration, method settings,
and proper connections will also be addressed.
SPP Columns, like Agilent Poroshell 120, have a Lower
Reduced Plate Height and Better C-term than 1.8 µm
TPP Columns
Significant Care is Taken to Ensure the Wide Variety of
Agilent TPP and SPP Columns Provide Similar
Selectivity for Predictable and Easy Method Transfer
and Scaling Selectivity Comparisons with 4.6 x 50 mm Columns, 5-95% CH3CN in 2
min with 0.1% formic acid, 2 mL/min
Selectivity plots, using a simple gradient show highly correlated
selectivity among these TPP and SPP Agilent C18 columns for 140
compounds (including acids, bases and neutrals), with R2>0.99 and slopes
close to 1.
Because of the strongly correlated selectivity, transferring and scaling
methods among these Agilent C18 columns is predictive, often requiring
no further method development.
When High pH Mobile Phases are Necessary, the
Agilent Poroshell HPH-C18 Column can Appreciably
Improve Column Lifetime
Lifetime of SPP columns in phosphate buffer, pH 8, at elevated
temperature. Mobile phase: Premixed 60% 30 mM sodium phosphate
buffer at pH 8 and 40% methanol; Flow rate 0.4 mL/min; UV absorbance
254 nm; 65 °C; Columns: 2.1 x 50 mm, 2.7 µm; Analyte: Naphthalene
The HPH SPP particles are synthesized by chemically modifying the
surface of the silica SPP particles with an organic layer, resulting in 15x
longer lifetime under these stressful chromatographic conditions
(columns are considered dead after a 10% loss of efficiency).
1.50
2.00
2.50
3.00
3.50
4.00
4.50
5.00
0.00 5.00 10.00 15.00 20.00
Red
uced
pla
te h
eigh
t, h
Reduced interstitial velocity, v
van Deemter Plot, k'(naphthalene)=3, 2.1 x 50 mm columns
Transferring and Scaling Methods among a Variety of
Superficially Porous Particle Columns Anne Mack, William Long, Jason Link, Xiaoli Wang, Maureen Joseph, Bill Barber; Agilent Technologies, 2850 Centerville Rd, Wilmington, DE
Introduction
Conclusions
High Efficiency Generated by SPP Columns can
Improve Resolution, Making the Difference between
Separating Analytes or Not
The 5 µm TPP column cannot separate these 8
benzodiazepines; however a 2.7 µm SPP column with the same
selectivity has enough efficiency to drive baseline resolution
for all analytes.
High Efficiency Generated by SPP Columns can also
Save Substantial Time and Solvent, Allowing More
Analyses per Day and Lower Costs per Analysis
The 2.7 µm SPP column gives a more sensitive, faster, better
resolved separation of 10 sulfa drugs, compared to a longer 5
µm TPP column, while only generating 325 bar pressure.
Analysis time is reduced by 73% with the SPP column.
Mobile phase consumption is reduced by 47% with the SPP
column.
Recent Revisions to USP General Chapter 621 Allow
For Method Flexibility with SPP Columns of Various
Particle Sizes
Allowable Column Adjustments per USP General Chapter 621
before and after August 1, 2014
Example: SPP Column Alternatives for USP Naproxen Assay
Additional Allowable Adjustments for Method Parameters
Instrument Dispersion Can Have a Considerable Negative Impact on the Performance of Low Volume, High Efficiency SPP Columns Instrument Dispersion is Improved by 60% (Reduced from 9.7 to 3.9 µL) using Short, Small id Connecting Capillaries to Reduce System Extra-Column Volume
Performance of Isocratic Analyses is Significantly Improved by Minimizing LC System Extra-Column Volume; Efficiency is
Improved up to 65%, Resolution is Improved up to 39% Gradient Performance is also Improved by Minimizing System Extra-Column Volume; Peak Capacity is Improved by 27%, Resolution is Improved by >20%
Spring-Loaded Agilent A-Line Fittings Improve Performance by Ensuring Ideal Connections Every Time a Connection is Made The Agilent A-Line Quick Connect fitting won The Analytical Scientist Innovation Award (TASIA) for its highly innovative approach in successfully simplifying the connection of fittings in UHPLC.
Monitoring of Tailing Factor and Theoretical Plates Number over 200 Reconnections of an A-Line Quick Connect Fitting using a Naphthalene QC Analysis
There is no visible change in chromatographic performance after 200 reconnections; tailing factors and theoretical plates stayed constant within the experimental allowance through the procedure. The spring-loaded A-Line Quick Connect Fitting kept the connection dead-volume free and leak-free after 200 times of reuse.
Method transfer and scaling is predictable and
easy with SPPs Similar selectivities are available on both SPPs and
TPPs SPP’s improved efficiency can increase resolution
and save significant time New USP guidelines allow flexibility with SPP
columns Low volume, high efficiency columns are very
sensitive to instrument dispersion Spring-loaded fittings paired with high efficiency
SPP columns ensure ideal connections
Results and Discussion Results and Discussion
1.8 µm
TPP
2.7 µm
SPP
A 1.26 1.06
B 5.83 3.74
C 0.08 0.05
— ZORBAX Eclipse Plus C18, 2.1 x 50 mm, 1.8 µm (p/n 959757-902) —Poroshell 120 EC-C18, 2.1 x 50 mm, 2.7 µm (p/n 699775-902)
y = 1.0163x - 0.0024
R² = 0.9994
0
0.5
1
1.5
2
2.5
3
0 1 2 3
RT
on 2
.7 µ
m P
oros
hell
12
0 E
C-C
18
(m
in)
RT on 4 µm Poroshell 120 EC-C18
(min)
y = 1.0296x + 0.012
R² = 0.9992
0
0.5
1
1.5
2
2.5
3
0 1 2 3
RT
on 5
µm
ZO
RB
AX
Ecl
ipse
Plu
s C
18
(m
in)
RT on 4 µm Poroshell 120 EC-C18
(min)
y = 1.0459x + 0.0608
R² = 0.9989
0
0.5
1
1.5
2
2.5
3
0 1 2 3
RT
on 2
.7 µ
m P
oros
hell
12
0 E
C-C
18
(m
in)
RT on 2.7 µm Poroshell HPH-C18
(min)
50
60
70
80
90
100
110
0 1000 2000 3000
% In
itial
Eff
icie
ncy
mL Stress Buffer
Silica SPP-C18
HPH SPP-C18
—Poroshell 120 EC-C18 2.1 x 50 mm, 2.7 µm (p/n 699775-902) —Poroshell HPH-C18 2.1 x 50 mm, 1.8 µm (p/n 699775-702)
Parameters for
System Suitability
USP34 USP37-NF32S1
Isocratic/Gradient Isocratic Gradient
Particle Size -50% L/dp: -25% to +50%
or N: -25% to +50%
No Changes allowed
Column Length ±70%
Column Inner
Diameter
Flexible, w/ constant
linear velocity
Flexible, w/ constant linear
velocity
No Changes allowed
Parameters for
System Suitability
USP34 USP37-NF32S1
Isocratic/Gradient Isocratic Gradient
Flow rate Based on column
dimension: F2=F1×[(l2×d2
2)/(l1×d12)]
Additional adjustments:
±50%
Based on dp: F2=F1×[(dc2
2×dp1)/(dc12×dp2)]
Additional adjustments:
±50%, provided N
decreases ≤20%
No Changes allowed
Injection volume May be reduced, as far as
is consistent with
precision and detection
limits;
increase not permitted
May be adjusted, as far as
is consistent with precision
and detection limits
May be adjusted, as far as
is consistent with precision
and detection limits
Column Temp ±10oC ±10oC ±10oC
Mobile phase pH ±0.2 units ±0.2 units ±0.2 units
Salt Concentration within ±10% if the
permitted pH is met
within ±10% if the
permitted pH is met
within ±10% if the
permitted pH is met
Ratio of
Components in
Mobile Phase
Minor component
(≤50%): ±30% relative,
cannot exceed ±10%
absolute; may only adjust
1 component in ternary
mixtures
Minor component (≤50%):
±30% relative, cannot
exceed ±10% absolute;
may only adjust 1
component in ternary
mixtures
No Changes allowed *
* Not specified in 621,
assume no changes are
allowed
Wavelength of UV-
Vis Detector
No changes allowed No changes allowed No changes allowed
min 0.5 1 1.5 2 2.5 3
mAU
0 25 50 75
100 125 150
min 0.5 1 1.5 2 2.5 3 3.5
mAU
0 25 50 75
100 125 150
min 0.5 1 1.5 2 2.5 3 3.5
mAU
0 25 50 75
100 125 150
A: 0.1% formic acid in water, B: 0.1% formic acid in acetonitrile, 2 mL/min, 32%B isocratic, 25C, 254nm, 80 Hz
5 µm ZORBAX Eclipse Plus C18 4.6 x 50 mm (p/n 959946-902)
4 µm Poroshell 120 EC-C18 4.6 x 50 mm (p/n 699970-902)
2.7 µm Poroshell 120 EC-C18 4.6 x 50 mm (p/n 699975-902)
1. alprazolam 2. clonazepam 3. diazepam 4. flunitrazepam 5. lorazepam 6. nitrazepam 7. oxazepam 8. temazepam
5 10 15 20
mAU
0 20 40 60 80
100
5 10 15 20 25
mAU
0 50
100 150 200 250
1. sulfadiazine 2. sulfathiazole 3. sulfapyridine 4. sulfamerazine 5. sulfamethazine 6. sulfamethazole 7. sulfamethoxypyridazine 8. sulfachloropyridazine 9. sulfamethoxazole 10. sulfadimethoxine
5 µm ZORBAX Eclipse Plus C18 4.6 x 250 mm (p/n 959990-902)
2.7 µm Poroshell 120 EC-C18 4.6 x 100 mm (p/n 695975-902) 25
A. 0.1% Formic Acid in H2O, B. CH3CN, 1 or 2 mL/min, 8-33% B in 30 or 8 min, 25 C, 254 nm
min 0 1 2 3 4 5 6 7 8 9
mAU
0 10 20 30 40 50 60
USP Prescribed Column: 5 µm Eclipse Plus C18 4.6 x 150 mm
L/dp=30,000
min 0 1 2 3 4 5 6 7 8 9
mAU
0 10 20 30 40 50 60
System Suitability Method Requirement: N>4000, Rs>11.5
min 0 1 2 3 4 5 6 7 8 9
mAU
0 10 20 30 40 50 60
1 2 3 4 5 6 7
mAU
0 5 10 15 20 25 30
1 2 3 4 5 6 7
mAU
0 5 10 15 20 25 30
4 µm Poroshell 120 EC-C18 4.6 x 150 mm
25%↑L/dp L/dp=37,500
4 µm Poroshell 120 EC-C18 4.6 x 100 mm
17%↓L/dp L/dp=25,000
2.7 µm Poroshell 120 EC-C18 4.6 x 100 mm
23%↑L/dp L/dp=37,037
2.7 µm Poroshell 120 EC-C18 4.6 x 50 mm
38%↓L/dp L/dp=18,518
N=10639
Rs=13.7
N=19054 79%↑N
Rs=16.9
N=13186 24%↑N
Rs=14.2
N=21046 98%↑N
Rs=17.0
N=11281 6%↑N
Rs=12.6
50:49:1 CH3CN:H2O:CH3COOH 1.2 mL/min, All Pressures < 300 bar Sample: Naproxen, Butyrophenone
Binary Pump
Binary Pump
Autosampler
Diode Array Detector
Diode Array Detector
Solvent Tray Solvent Tray
Default 1290 LC Optimized 1290 LC
Default 1290 LC:
Needle Seat Capillary: 0.12 x 100 mm = 1.1 µL ALSTCC Capillary: 0.12 x 340 mm = 3.8 µL
TCCDAD Capillary: 0.12 x 220 mm = 2.5 µL
Flow Cell V(σ)1.0 µL = 2.3 µL
2.1 x 50 mm Column = 172.3 µL
Void Volume of Column = 103.9 µL
Optimized 1290 LC:
Needle Seat Capillary: 0.11 x 100 mm = 0.9 µL ALSTCC Capillary: 0.08 x 220 mm = 1.1 µL
TCCDAD Capillary: 0.08 x 220 mm = 1.1 µL
Flow Cell V(σ)0.6 µL = 0.8 µL
2.1 x 50 mm Column = 172.3 µL
Void Volume of Column = 103.9 µL
Autosampler
Column Compartment
Column Compartment
x102
0 0.2 0.4 0.6 0.8 1
x102
0 0.2 0.4 0.6 0.8 1
0.4 0.8 1 1.4 1.8 2.2 2.6 3 3.4 3.8 4.2
Optimized LC/UV System, 3.9 µL extra column volume, Pmax = 240 bar RRHD Eclipse Plus C18, 1290 Infinity LC
0.4 mL/min, 60% CH3CN
Agilent Publication 5990-9502EN
Default LC/UV System, 9.7 µL extra column volume, Pmax = 230 bar RRHD Eclipse Plus C18, 1290 Infinity LC
0.4 mL/min, 60% CH3CN
Agilent Publication 5990-9502EN
4.6 5 5.4 5.8
Rs5,6=2.36 N4=5450 N8=9714 N9=10068
Rs5,6=2.78 N4=7630 N8=11196 N9=10967
x102
0 0.2 0.4 0.6 0.8 1
x102
0 0.2 0.4 0.6 0.8 1
0.1 0.2 0.3 0.4 0.5 0.6
Optimized LC/UV System, 3.9 µL extra column volume, Pmax = 310 bar RRHD Eclipse Plus C18, 1290 Infinity LC
0.4 mL/min, 25-95% CH3CN in 1.2 min
Agilent Publication 5990-9502EN
Default LC/UV System, 9.7 µL extra column volume, Pmax = 300 bar RRHD Eclipse Plus C18, 1290 Infinity LC
0.4 mL/min, 25-95% CH3CN in 1.2 min
Agilent Publication 5990-9502EN
0.7 0.8 0.9 1 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8
Rs5,6=1.77 nC=44.0
Rs5,6=2.25 nC=55.7
A-Line Quick Connect Assembly p/n
A-Line Quick Connect Assy ST
0.12x105mm 5067-5957