Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

19
Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett

Transcript of Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Page 1: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums

X. Zhang, K. Triplett

Page 2: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Summary

• Using Gaussmeter, the peak field measured near the two GPMS buses on the wall is about 25Gauss.

• The peak file inside the reference magnet box for the proposed new transductor location is about 7 Guass (combined, likely due to the bus), which is satisfying the recommend external field limit 50 Gauss for precision measurement.

• The 720Hz signal are near the noise level of about 80dB lower than 15Hz, except for the Bdot signal monitor.

Page 3: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Hall Probe Inside Ref. Magnet Box

Page 4: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Hall Probe Inside Ref. Magnet Box

Page 5: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Hall Probe Inside Ref. Magnet Box

Page 6: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Hall Probe Inside Ref. Magnet Box

Page 7: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Hall Probe Inside Ref. Magnet Box

Page 8: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

TRANSDUCTOR/BDOT BUFFER MONITOR

Page 9: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Transductor Buffer Monitor

Page 10: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Transductor Buffer Monitor

Page 11: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Bdot Buffer Monitor

Page 12: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Bdot Buffer Monitor

Page 13: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Reference Magnet Field Measurement

Page 14: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.
Page 15: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Gaussmeter By Analog Output

Page 16: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Gaussmeter By Spectrum

Peak Field:2645 Gs

Page 17: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Gaussmeter By Spectrum

Peak Field:2645 Gs

Page 18: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Gaussmeter Bx Spectrum

Peak Field:45.53 Gs

Page 19: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett.

Gaussmeter Bx Spectrum

Peak Field:45.53 Gs