Dräger Evita Infinity® V500Train-the-Trainer
Operational Concepts
Title of presentation | Author of presentation | 05/01/2010| 30
Table of contents
1. Operating Concept2. Start Ventilation3. Alarm Management
Title of presentation | Author of presentation | 05/01/2010| 30
Table of contents
1. Operating Concept2. Start Ventilation3. Alarm Management
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Touch – Turn – Confirm
1.Touch the screen at the required button button turns yellow, ring of the rotary knob turns
orange2.Turn the rotary knob to adjust the setting3.Confirm the selection by pressing the rotary
knob (Note: Orange light goes off)
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Touch – Turn – Confirm
1.Touch the screen at the required button button turns yellow, ring of the rotary knob turns
orange2.Turn the rotary knob to adjust the setting3.Confirm the selection by pressing the rotary
knob orange light goes off
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Touch – Turn – Confirm
1.Touch the screen at the required button button turns yellow, ring of the rotary knob turns
orange2.Turn the rotary knob to adjust the setting3.Confirm the selection by pressing the rotary
knob orange light extinguishes
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Direct setting of a parameter (QuickSet)
Set a therapy control value without confirming the setting. The changes to the setting will immediately become effective for the patient.
Direct setting of ventilation parameters can be performed in all ventilation modes.
Only the therapy controls O2 and Flow cannot be set directly. Apart from that it is applicable for all the other ventilation parameters.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Direct setting of a parameter (QuickSet)
1.Touch the corresponding therapy control2.Press and hold the rotary knob for approx. 3 seconds
Therapy control now turns dark green withyellow edge.3.Press and hold the rotary knob and turn to set the value
The setting will immediately apply.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Direct setting of a parameter (QuickSet)
1.Touch the corresponding therapy control2.Press and hold the rotary knob for approx. 3 seconds
Therapy control now turns dark green withyellow edge.3.Press and hold the rotary knob and turn to set the value
The setting will immediately apply.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Direct setting of a parameter (QuickSet)
1.Touch the corresponding therapy control2.Press and hold the rotary knob for approx. 3 seconds
Therapy control now turns dark green withyellow edge.3.Press and hold the rotary knob and turn to set the value
The setting will immediately apply.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Linked setting of two parameters: PEEP/Pinsp and RR/TiSet the two therapy controls PEEP and Pinspsimultaneously with the pressure difference (ΔP) remaining constant when adjusting eitherparameter.
Set the two therapy controls RR and Tisimultaneously with the I:E ratio remainingconstant when adjusting either parameter.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Linked setting of two parameters
Example: PEEP/Pinsp1.Touch one of the two therapy controls2.Touch the link buttonThe two linked therapy controls turn yellow.3.Turn the rotary knob to adjust the two values with the pressure difference (ΔP) remaining constant and confirm the setting
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Linked setting of two parameters
Example: PEEP/Pinsp1.Touch one of the two therapy controls2.Touch the link buttonThe two linked therapy controls turn yellow.3.Turn the rotary knob to adjust the two values with the pressure difference (ΔP) remaining constant and confirm the setting
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Linked setting of two parameters
Example: PEEP/Pinsp1.Touch one of the two therapy controls2.Touch the link buttonThe two linked therapy controls turn yellow.3.Turn the rotary knob to adjust either value and note the pressure difference ΔP remains constant and confirm the setting
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Lock function for therapy bar
1.Press “System setup”2.Select “Screen layout”, chose “Therapy bar”
and press “On”
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Lock function for therapy bar
1.Press “System setup”2.Select “Screen layout”, chose “Therapy bar” and press “On”
Now, the therapy controls may only bechanged in the Ventilation settings dialogwindow, indicated by: (1)the missing edge of the therapy controls(2)the advisory message on top of the therapy bar
(2)
(1)
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Night mode
When setting Evita V500 to the night mode an automatic changeover of the followingsettings will apply:
Color scheme of the screenVolume of the alarm toneAutomatic increase of the alarm tone is activated
The switch to night mode can either be doneautomatically (by configuring a certain time inSystem setup) or manually by pressing“Day/Night” (the “Day/Night” button has to beconfigured onto the main screen).
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Freeze waveforms
For better analysis, the current graphs,waveforms, loops, trends may be stopped fora short amount of time
Prerequisite:
the “Freeze waveforms” button is configured onto the main menu bar
1.To freeze a graph: press “Freeze waveforms” at the desired time
2.Move the cursor by turning the rotary knob to analyze data at a certain point of time
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Freeze waveforms
For better analysis, the current graphs,waveforms, loops, trends may be stopped fora short amount of time
Prerequisite:
the “Freeze waveforms” button is configured onto the main menu bar
1.To freeze a graph: press “Freeze waveforms” at the desired time
2.Move the cursor by turning the rotary knob to analyze data at a certain point of time
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Export a screenshot
Screenshots can be exported directly to aUSB storage medium.
Prerequisite:
the “Export screenshot” button is configured onto the main menu bar
1.Insert a USB memory stick into one of the USB ports of the Infinity C500
2.To export a screenshot: press “Export screenshot” button at the desired point of time
Only Dräger USB memory sticks are approved for usage with Infinity C500!
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Export a screenshot
Screenshots can be exported directly to aUSB storage media.
Prerequisite:
the “Export screenshot” button is configured onto the main menu bar
1.Insert a USB memory stick into one of the USB ports at C500
2.To export a screenshot: press “Export screenshot” at the desired point of time
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Activating StandbyThe V500 can be placed in “Standby” if temporary cessation of ventilation is desired (i.e. transport not utilizing the V500), and must be placed in “Standby” in order to change “Therapy Type” or “Patient Category”
1.Press “Start/Standby” on the main screen2.Select “Standby”3.Press the Rotary knob to confirmEvita V500 is now in standby and gives a highpriority alarm.4. To deactivate the alarm: Press “Alarm reset” and confirm
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Activating Standby
For example to change the application modeor patient category the “Standby mode” needsto be activated because these items can onlybe changed when Evita V500 is in standby.
1.Press “Start/Standby” on the main screen2.Select “Standby”3. Press the Rotary knob to confirmEvita V500 is now in standby and gives a highpriority alarm.4. To deactivate the alarm: Press “Alarm reset” and confirm
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Activating Standby
For example to change the application modeor patient category the “Standby mode” needsto be activated because these items can onlybe changed when Evita V500 is in standby.
1.Press “Start/Standby” on the main screen2.Select “Standby”3.Press the Rotary knob to confirmEvita V500 is now in standby and gives ahigh priority alarm.4. To deactivate the alarm: Press “Alarm reset” and confirm by pressing the Rotary knob
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Calibrating the Infinity C500 screen
Prerequisite:Evita V500 must be in “Standby”
1.Keep the “Audio paused” button and the rotary knob pressed for approx. 20 seconds
After approx. 10 seconds an alarm will appear2.Continue pressing the two buttonsAfter approx. 20 seconds the alarm message
resolves itself in preparation for calibration.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Calibrating the Infinity C500 screen
PrerequisiteEvita V500 in in standby mode
1.Keep the “Audio paused” and the rotary knob pressed for approx. 20 seconds
After approx. 10 seconds a medium priorityalarm will appear.
2.Continue pressing the two buttonsAfter approx. 20 seconds the alarm message resolves itself in preparation for calibration.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Calibrating the Infinity C500 screen
PrerequisiteEvita V500 in in standby mode
1.Keep the “Audio paused” and the rotary knob pressed for approx. 20 seconds
After approx. 10 seconds a medium priorityalarm will appear.
2.Continue pressing the two buttonsAfter approx. 20 seconds the alarmresolves itself in preparation for calibration.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Calibrating the Infinity C500 screen
3.Touch “Calibrate”A red and white colored target appears on thescreen.
4.Touch the target within the given amount of time at three different positions and confirm calibration by pressing the green checkmark appearing in the center of the screen.
Screen calibration is now successfullycompleted.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Calibrating the Infinity C500 screen
3.Touch “Calibrate”A red and white colored target appears onthe screen.
4.Touch the target within the given amount of time at three different positions and confirm calibration by pressing the green checkmark appearing in the center of the screen.
Screen calibration is now successfullycompleted.
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Adjusting the angle of the Infinity C500 screen
1.Push the handle on the back of the screen to the right
2.Keeping the handle pressed in, grasp and tilt the screen to the desired position
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Adjusting the angle of the Infinity C500screen
1.Push the handle on the back of the screen to the right
2.Keeping the handle pressed in, grasp and tilt the screen to the desired position
Title of presentation | Author of presentation | 05/01/2010| 30
Device operation Operating concept
Adjusting the angle of the Infinity C500screen
1.Push the handle on the back of the screen to the right
2.Keeping the handle pressed in, grasp and tilt the screen to the desired position
Title of presentation | Author of presentation | 05/01/2010| 30
Table of contents
1. Operating Concept2. Start Ventilation3. Alarm Management
Title of presentation | Author of presentation | 05/01/2010| 30
Ventilation Start Ventilation
Using the settings of the previous patient
Prerequisites: Evita V500 is switched on The start dialog is opened
Start up:1. Select “Current patient”2. Chose therapy3. Select application mode4. Select breathing circuit and humidifier5. If necessary: adjust ventilation settings6. Start ventilation
Title of presentation | Author of presentation | 05/01/2010| 30
Ventilation Start Ventilation
Using the settings of the previous patient
Prerequisites: Evita V500 is switched on The start dialog is opened
Start up:1. Select “Current patient”2. Chose therapy (ventilation or O2 Therapy)3. Select application mode4. Select breathing circuit and humidifier5. If necessary: adjust ventilation settings6. Start ventilation
Title of presentation | Author of presentation | 05/01/2010| 30
Ventilation Start Ventilation
Using the settings of the previous patient
Prerequisites: Evita V500 is switched on The start dialog is opened
Start up:1. Select “Current patient”2. Chose therapy3. Select application mode (Tube or NIV)4. Select breathing circuit and humidifier5. If necessary: adjust ventilation settings6. Start ventilation
Title of presentation | Author of presentation | 05/01/2010| 30
Ventilation Start Ventilation
Using the settings of the previous patient
Prerequisites: Evita V500 is switched on The start dialog is opened
Start up:1. Select “Current patient”2. Chose therapy3. Select application mode4. Select breathing circuit and humidifier type5. If necessary: adjust ventilation settings6. Start ventilation
Title of presentation | Author of presentation | 05/01/2010| 30
Ventilation Start Ventilation
Using the settings of the previous patient
Prerequisites: Evita V500 is switched on The start dialog is opened
Start up:1. Select “Current patient”2. Chose therapy3. Select application mode4. Select breathing circuit and humidifier5. If necessary: adjust ventilation settings6. Start ventilation
Title of presentation | Author of presentation | 05/01/2010| 30
Ventilation Start Ventilation
Using the settings of the previous patient
Prerequisites: Evita V500 is switched on The start dialog is opened
Start up:1. Select “Current patient”2. Chose therapy3. Select application mode4. Select breathing circuit and humidifier5. If necessary: adjust ventilation settings6. Start ventilation
Title of presentation | Author of presentation | 05/01/2010| 30
Ventilation Start Ventilation
Admitting a new patient
Prerequisites: Evita V500 is switched on The start dialog is opened
Start up:1. Select desired patient category2. Chose therapy3. Select application mode4. Select breathing circuit and humidifier5. If necessary: adjust ventilation settings6. Start ventilation
Title of presentation | Author of presentation | 05/01/2010| 30
Table of contents
1. Operating Concept2. Start Ventilation3. Alarm Management
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm BehaviorAlarm limits
Adjustable alarm limits
Expiratory Minute Volume: MVe high/low Delay time for MVe alarms: MV delay
high/low Airway pressure: PAW high Respiratory Rate: RR high Apnea alarm delay time: Tapn Disconnection time in NIV: Tdisconnect Tidal volume: VT high Endexpiratory CO2 concentration: etCO2
high/low
Alarm limits that cannot be set manually (PAW low, RR low, VT low) are linked to the respective setting and will adjust automatically.
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm BehaviorAlarm limits
Displayed alarm limitsThe following alarm limits are displayed on the
main screen in the respective parameter box:
PAW high
MVe high/low
RR high
VT high/low
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm BehaviorAlarm limits
Set alarm limits
1. Touch “Alarms”2. Touch “Limits”3. Adjust desired alarm limit/s
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm limits
Set alarm limits
1. Touch “Alarms”2. Touch “Limits”3. Adjust desired alarm limit/s
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm limits
Set alarm limits
1. Touch “Alarms”2. Touch “Limits”3. Adjust desired alarm limit
Configure start settings for the alarm limits inSystem setup.
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm limits
Deactivate alarm limits
in the neonatal patient category the MVe low alarm limit may be deactivated
in NIV the MVe low as well as the VT high and low alarm may be deactivated
1. Touch the required alarm limit2. Turn the rotary knob counter clockwise to
turn a lower alarm limit off or clockwise to turn a high alarm limit off until “Off” is displayed
3. Confirm with rotary knob
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm limits
Deactivate alarm limits
in the neonatal patient category the MVe low alarm limit may be deactivated
in NIV the MVe low as well as the VT high and low alarm may be deactivated
1. Touch the required alarm limit2. Turn the rotary knob counter clockwise
to turn a lower alarm limit off or clockwise to turn a high alarm limit off until “Off” is displayed
3. Confirm with rotary knob
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm limits
Deactivate alarm limits
in the neonatal patient category the MVe low alarm limit may be deactivated
in NIV the MVe low as well as the VT high and low alarm may be deactivated
1. Touch the required alarm limit2. Turn the rotary knob counter clockwise to
turn a lower alarm limit off or clockwise to turn a high alarm limit off until “Off” is displayed
3. Confirm with rotary knob
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm concept
Indication of an alarm
Visually: via the alarm message in the header barvia highlighting the affected parameter in the monitoring area (if configured)
via the 360° alarm light on Infinity C500
Acoustically:via the alarm tone (Dräger sequence or ICE sequence)
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm concept
Alarm priorities
CAUTION: Medium priority alarm message indicated by the yellow background prompt action is necessary in order to avert a danger
WARNING: High priority alarm message indicated by the red background immediate action is necessary in order to avert an acute danger
NOTE: Low priority alarm message indicated by the cyan background attention is necessary, but a delayed response is sufficient
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm concept
Display current alarms
In the header bar no more than 2 activealarms can be displayed.If there are more then two active alarmsthe “More…” button is displayed.
To view all active alarms: Touch “More…”or select “Alarms” ► “Current alarms” or touch displayed alarm in header bar
The Current alarms page in the Alarmsmenu is opened displaying Date, Time,Duration and Priority of the current alarms.
if the cause of the alarm is solved: Press “Reset all” to delete Alarms from the overview
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm concept
Alarm history
On the Alarm history page all alarms are savedand displayed with Date, Time, Duration andPriority. The entries in the alarm history arealso retained after the device has beenswitched off and on again.
To view the Alarm history page: select “Alarms” ► “Alarm history”
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm concept
Alarm silence
The Alarm silence function mutes acurrent alarm for a maximum of 120 s.
To suppress the acoustic alarm for 120 s: Press “Audio paused”To reactivate the alarm tone before the suppression time has elapsed: Press “Audio paused” again
Even if the current alarm is silenced Evita V500will give an acoustic alarm if a alarm with ahigher priority occurs!
Alarm silence activated
Remaining alarm suppression time
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm – Cause – Remedy
The Alarm – Cause – Remedy helpfunction
With the Alarm – Cause – Remedy help functionpotential causes and solutions for currentalarms are available directly on the screen forquickly solving the problem.
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm – Cause – Remedy
Use the Alarm – Cause – Remedy helpfunction
Prerequisite
the Current alarms or the Alarm history page is open
1.Press “↑” at the top of the page to display the Alarm – Cause – Remedy page
2.Use the cursor to display cause and remedy for a current alarm
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Alarm – Cause – Remedy
Use the Alarm – Cause – Remedy helpfunction
Prerequisite
the Current alarms or the Alarm history page is open
1.Press “↑” at the top of the page to display the Alarm – Cause – Remedy page
2.Use the cursor to display Cause and Remedy for a current alarm
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Troubleshooting
In the event of an alarm
1.Touch the displayed alarm in the header bar
2.Touch “↑” to display the Alarm – Cause – Remedy help function
3.To reset no longer active alarms: press “Reset all” on the Current alarms page or “Alarm reset” in the header bar and confirm
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Troubleshooting
In the event of an alarm
1.Touch the displayed alarm in the header bar2.Touch “↑” to display the
Alarm – Cause – Remedy help function3.To reset no longer active alarms: press “Reset
all” on the Current alarms page or “Alarm reset” in the header bar and confirm
Title of presentation | Author of presentation | 05/01/2010| 30
Alarm Behavior Troubleshooting
In the event of an alarm
1.Touch the displayed alarm in the header bar2.Touch “↑” to display the
Alarm – Cause – Remedy help function3.To reset no longer active alarms: press
“Reset all” on the Current alarms page or “Alarm reset” in the header bar and confirm