ROTATION 5: TPS MODELING AND BEAM COMMISSIONING at UIHC Radia9on
Oncology Mentors: John Bayouth, Ph.D.
(04/23/09 – 06/11/09)
By Vibha Chaswal
![Page 2: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/2.jpg)
Goals
• Model 1 photon beam -‐ Know parameters and their effects
• Model 1 electron beam
-‐ Know parameters and their effect
• Model Wedges
• Validate the beams modeled
• Understand the nuances of TPS modeling
![Page 3: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/3.jpg)
Photon beams: data requirements
• Pdds • Profiles • Calibra9on output factor • Rela9ve output factors Data For Valida9on! -‐ A subset of above -‐ Sc and Sp -‐ Dosimetric tests, match computed and measured dose distribu9ons
![Page 4: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/4.jpg)
Data collected for assumed clinical applica9on
![Page 5: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/5.jpg)
TPS modeling: The Machine
![Page 6: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/6.jpg)
TPS Modeling: The machine
![Page 7: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/7.jpg)
TPS Modeling: The Radia9on Beam
• Fine tuning Profiles -‐ Effec9ve source size for Penumbra matching
-‐ Gaussian width and height of FF scacer source for tails
-‐ Transmission factors of Jaw and MLC for tails region
![Page 8: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/8.jpg)
Energy Spectrum
![Page 9: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/9.jpg)
Energy Spectrum: Build-‐up region
![Page 10: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/10.jpg)
In-‐field Region: FF acenua9on using cone model
With Off-‐axis Soeening factor
![Page 11: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/11.jpg)
In-‐field Region: FF acenua9on using arbitrary profile
With Off-‐axis Soeening factor
![Page 12: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/12.jpg)
Off-‐axis soeening
![Page 13: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/13.jpg)
Out-‐of field regions: Penumbra and tails
![Page 14: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/14.jpg)
Out-‐of field modeling parameters
![Page 15: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/15.jpg)
Beam Valida9on: PDDs
Result: All Pdds pass within 2% or 2mm
![Page 16: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/16.jpg)
Beam Valida9on: Profiles
Result: Within 3% or 3mm except outliers in tail regions
![Page 17: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/17.jpg)
Beam Valida9on tests: MU verifica9on
![Page 18: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/18.jpg)
Beam valida9on: R and V import
![Page 19: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/19.jpg)
Rela9ve Decision Maps: error Limit = 3 %, dta Limit = 3 mm
![Page 20: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/20.jpg)
TPS Modeling: The machine, Collimator rota9on incorrect
![Page 21: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/21.jpg)
Electron beams: Data required
![Page 22: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/22.jpg)
Electron Beam Parameters
![Page 23: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/23.jpg)
Two adjustable parameters
- Off-axis ratios, and
- The scatter correction factor (FMCS), dose spread out in media relative to air.
![Page 24: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/24.jpg)
Electron Beam modeling
![Page 25: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/25.jpg)
Electron PDD
![Page 26: TPS modeling and Beam commissioning at UIHC Radiation Oncology](https://reader033.fdocuments.us/reader033/viewer/2022051209/547a742bb4795990098b4a2b/html5/thumbnails/26.jpg)
Thank you!!!!!