Total Skin Electron Therapy_PRESENTED_10!12!11

download Total Skin Electron Therapy_PRESENTED_10!12!11

of 42

Transcript of Total Skin Electron Therapy_PRESENTED_10!12!11

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    1/42

    Total Skin Electron

    Therapy

    Wednesday, October 12th, 2011

    Michael S. Curry, M.S.

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    2/42

    TSET

    TG-30

    Introduction

    Requirements

    Techniques

    LINAC Operating Conditions

    Dosimetry and Instrumentation

    Patient Considerations

    Commissioning

    TSET at MDACCO

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    3/42

    Introduction to TSET

    Diseases most commonly treated (require shallow Tx depths)

    Cutaneous T-cell Lymphomas

    Micosis Fungoides

    Sezary syndrome (less so than Micosis Fungoides)

    Less commonly treated diseases Kaposis sarcoma

    Dose Scheme (TG-30):

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    4/42

    Requirements

    Beam

    Specification of:

    Field size, penetration, energy, depth, dose rate, field flatness in

    treatment plane, X-ray background, and the need and nature of boost

    fields

    Room

    Careful consideration of:

    Space, shielding, ventilation, electron ranges

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    5/42

    Beam Requirements

    Field size

    200 cm high by 80 cm wide

    Penetration depth

    Varies with stage, type of disease, and over the body surface

    Typically 515mm or more at the 50% isodose line

    Energy

    3 to 7 MeV at patient treatment plane

    4 to 10 MeV at beam exit window

    Dose rate

    In order to reduce Tx time high dose rates are desirable

    Range from 0.25 Gy/min to several Gy/min

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    6/42

    Beam Requirements

    Field flatness in the treatment plane

    Vertical uniformity of +-8%, Horizontal Uniformity of +-4% over the central

    160 cm x 60 cm area

    X-ray background

    Penetrating and forward directed

    Exposes most of body and should be ALARA

    Can be reduced by angling beams such that the x-ray peaks lie outside the

    body

    Desired to be 1% or less averaged over the entire body (typically 1-4%)

    EORTC spec < 5%

    Need and nature of boost fields

    Some body areas are unintentionally shielded by other body sections or

    inadequately exposed due to limitations of beam geometry

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    7/42

    Profiles

    Left: 6E 10x10 surface

    Right: 6E HDTSE (no cone, 10x10 XY jaws)

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    8/42

    Room Requirements

    Space

    In order to provide good dose uniformity large distances are needed

    Typically 2-7m from scatterer and patient (depending on technique)

    Ventilation

    TSET involves significant ozone production from ionizing large

    volumes of air in the treatment room

    Frequent exchange of air is essential for confining ozone exposure

    Shielding

    Typically MV X-ray shielding is adequate, but should measure Electron Ranges

    Max track length range of ~0.5 g/cm2 per MeV (~4 m/MeV) in air

    Range is typically not in direction of beam, most

    stop far short of this

    Bremsstrahlung

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    9/42

    Techniques

    Why do we need special techniques, are large AP/PA fields

    not sufficient?

    No, they are not sufficient:

    Patients body shape is not flat

    Dose uniformity impossible with AP/PA

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    10/42

    Techniques

    Flat = good

    Round = bad

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    11/42

    Techniques

    How do you treat curved surfaces? Multiple fields

    Electron arcs

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    12/42

    Techniques

    Prior to use of LINACs

    Beta Particles

    Beta-particle beams used Sr/Y-90 with Emax=2.18 MeV

    Use 10% isodose line and deliver 2Gy/fx in 15 min fx by scannins over a

    patient surface 60 cm x 180 cm

    Narrow rectangular beams

    Use Van de Graaf accelerators in fixed positions with vertically

    downward beams

    Patient translated horizontally on a motor driven couch under the 1.5 to

    4.5 MeV beam

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    13/42

    Techniques

    Used with LINACs

    Scattered single beam

    6.5 MeV beam with 0.15mm thick titanium scattering foil placed 10

    cm from accelerator window

    Shaped polystryrene beam-flattening filter mounted on front of

    treatment head

    Producing a flattened 4 MeV beam at the Tx plane

    Pair of parallel beams

    2 horizontal parallel beams with axes contained in a vertical plane

    (axes seperation=150cm) with Tx distance of 2m

    8MeV linac with carbon energy degraders just beyond exit window

    Adjusting thicknesses adjusts depth of penetration in patient

    X-ray background=2%

    Pair of angled beams

    Stanford Technique

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    14/42

    Stanford Technique

    Most commonly used technique

    Utilizes 6 dual fields

    Each dual field is comprised of two angled beams

    Typically 18 to 20 from 270 or 90 gantry position

    The 6 fields correspond to 6 different patient treatmentpositions

    Provides acceptable dose uniformity

    3 sets of dual fields treated per day

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    15/42

    Stanford Technique

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    16/42

    Stanford Technique

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    17/42

    Stanford Technique

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    18/42

    Mayo Clinic Jacksonville TSET Technique

    IMPAC Rx and Tx fields

    Each day 6 fields (upper and lower of 3 fields),

    e.g. 1U,1D,2U,2D,3U,3D

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    19/42

    Mayo Clinic Jacksonville TSET Technique

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    20/42

    University of Iowa TSET Technique

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    21/42

    University of Iowa TSET Technique

    Tray allows 40 x 40cm field size

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    22/42

    University of Iowa TSET Technique

    EORTC: R80 > 4mm, R20 < 2cm

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    23/42

    University of Iowa TSET Technique

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    24/42

    University of Iowa TSET Technique

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    25/42

    Techniques

    Used with LINACs

    Pendulum-arc

    8 MeV beam with the gantry rotated continuously during treatment

    in a 50 arc (six fields)

    Arc starts from an initial angle with CAX above the head and ends

    with CAX below the feet

    Degrader: large 1 cm thick plexiglass sheet 5cm from the patient

    Provides large angle electron scattering near the patient

    Patient Rotation

    Use single horizontal 6 MeV beam (3.5 MeV at Tx plane) with a

    scatterer near the exit window and a 7 m treatment distance

    X-ray background = 2.2%

    Reduced setup and Tx times as well as simplified beam matching

    Self shielding by limbs is unavoidable

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    26/42

    Pendulum Arc

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    27/42

    Patient Rotation

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    28/42

    LINAC Operating Conditions

    LINAC operating perameters

    Stable, Repeatable operating energy is essential

    Energy changes can shift large SSD fields laterally and change dose

    calibration and uniformity

    High average beam current (100 times normal electron beam) for highdose rate (>1Gy/min) at Tx plane

    Beam scatterer-energy degraders

    Scatterers are thin materials used to spread out the beam

    Energy degraders are thick materials used to reduce beam energy at

    the Tx plane

    Can be placed internally in the Tx head or externally

    Location and Materials used are important in determining dose rate

    X-ray background is least when scatterer-

    degrader is close to patient (~15mm from pt)

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    29/42

    LINAC Operating Conditions

    Beam Monitoring

    Typically monitor electron fluence rate or dose rate at Dmax

    Monitor response should be directly proportional to parameter of

    interest

    Acceptable monitors include: Built-in transmission ion chambers, secondary electron emission

    monitors, and electromagnetic induction monitors

    Should be placed where beam exits accelerator

    A common combination for TSET monitoring:

    a full-beam transmission ionization chamber at or within the treatmenthead

    a sampling chamber or electron collector placed at or near the patient Tx

    plane but not in line with the patient

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    30/42

    Dosimetry and Instrumentation

    Methods

    Acceptable detectors:

    ion chambers, film, TLD, Fricke dosimeters, electron collectors, and

    faraday cups

    Ion chambers (small thimble or small volume/thin window p-p) are

    recommended for scanning in a water tank

    If film is used with solid water, no air gaps should be present

    Might be difficult if film is in a package

    Phantoms

    Square water phantoms are recommended for depth dose data

    Otherwise, layered, flat phantoms made of conducting plastics or thin

    laminae of polystyrene with conductive graphite coatings are

    recommended for depth dose and buildup data

    Elliptical, oval or cylindrical phantoms are

    recommended for simulating a patients body

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    31/42

    Dosimetry and

    Instrumentation

    Measurements

    Energy

    Determined from depth ionization curve and the range-energy

    relationship given in section 2.1 of TG-30:

    Fluence

    Evacuated Faraday cup with collimator placed over the aperture

    Electron fluence is determined from the charge collected and the area of

    the collimator

    This fluence can be used to estimate entrance surface dose

    Depth dose

    Using a parallel-plate chamber overlayed with varying thicknesses of

    polystyrene

    EORTC: R80 > 4mm, R20 < 2cm

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    32/42

    Dosimetry and Instrumentation

    Calibration Point Dose Measurements

    TG-30 recommends that TSET absorbed dose be evaluated at a point

    (0,0,0) see figure

    The TG-21 protocol should be followed for calibration using data for

    electron energy of0 determined from R50 ( )

    A p-p chamber with known Ngas is recommended

    The chamber surface should be placed at dmax using polystyrene

    The rest of the chamber should also be surrounded with polystyrene

    1 cm posteriorly and 5 cm radially

    Expose the chamber to a single dual-field

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    33/42

    Dosimetry and Instrumentation

    Treatment Skin Dose Measurements

    Treatment skin dose is defined as the dose along a circle at or near the

    surface of a cylindrical polystyrene phantom 30 cm in diameter and

    30 cm high which has been irradiated as a hypothetical patient with

    all 6 dual fields.

    During irradiation the phantom is outfitted with appropriate dosimeters

    These dosimeters are calibrated with a single dual field

    Therefore calibration point dose is related to treatment skin dose by

    a factor B (typical values between 2.5 and 3.1)

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    34/42

    Patient Considerations

    Positioning

    Patient should be positioned to minimize self shielding

    Support devices

    Patients may have trouble standing

    Should be prepared with multiple alternativepositioning methods

    Shielding

    Lenses of the eyes can be shielded by placing high Z material either

    over or under the eye lids Finger and toe nails can be shielded with thin sheets of lead cut to size

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    35/42

    Shielding

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    36/42

    University of Iowa TSET Shielding

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    37/42

    Patient Considerations

    Boost Fields

    Typically the Soles of the feet, the perineal area, the dorsal surface of

    the penis, peri-anal skin, and the inframammary region of large

    breasted women

    Areas requiring a boost are determined by in-vivo dosimetry

    In-vivo dose measurements

    Important for 2 reasons:

    1. determination of the distribution of dose to the patient's skin

    2. verifying that the prescribed dose to the patient's skin is correct

    In-vivo dosimeters include:

    Ion chambers, diodes, film, TLDs, OSLs, MOSFETs

    Ion chambers and diodes are impractical due to the number required (at

    least 40)

    TG-30 recommends TLDs for In-vivo

    dosimetry (could argue for OSL or MOSFET too)

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    38/42

    Boost Fields

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    39/42

    Commissioning

    LINAC must be capable

    Room must be large enough

    Determine a Beam monitoring approach

    Decide on a technique

    Build or purchase patient support system

    Create a written procedure for changing from conventional

    modalities to TSET and back to conventional

    Determine an In-vivo dosimetry protocol

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    40/42

    TSET at MDACCO

    IX Vault

    Patient Support System

    Stanford Technique

    In-Vivo Dosimetry system

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    41/42

    In-Vivo Dosimetry system

    Landauer OSL system (InLight microStar)

    Carbon-doped aluminum oxide (Al2O3:C)

    Similar to TLD except use LED light for stimulation instead of heat

    The light used is of a specific wavelength

    The stimulation process doesnt anneal the OSL OSL can be read multiple times and or stored as record of delivered dose

    Independent of energy for 6 and 18 MV beams

    Response increases linearly with dose rate

    Time resolution = 0.1s, Spatial resolution

  • 7/30/2019 Total Skin Electron Therapy_PRESENTED_10!12!11

    42/42

    Review

    What is TSET e- beam energy? Nominal? 9 MeV

    At pt surface? 4 MeV

    What are differences between clinical e- beam and TSET

    beam? Higher dose rates; Larger Field Sizes; No Cones What are common Tx indications for TSET?

    T-cell Lymphoma (micosis fungoides)

    Doses? 36Gy in 9 fx

    Describe one technique for delivering TSET. Stanford Technique