1.14 Tonometry in General Practice- It's Use in Early Etection of Primary Open-Angle Glaucoma, h.
Tonometry
-
Upload
chittranjan-kumar -
Category
Education
-
view
500 -
download
1
description
Transcript of Tonometry
By. Chittranjan kumar Enternship
Definition :- It is the procedure performed to determine the intraocular pressure (IOP)
Def. of Glaucoma It is the characteristic and progressive condition of the eye which is raised the intraocular pressure occurs progressively damage the optic nerve fibers due to gradually loss of vision.
Direct Indirect
Manometer
DIGITAL INDETATION(Schiotz)
APPLANATION
(Goldman)
NCT
Types of Tonometer
Perkin’s
DirectTonometery
Procedure:- -Cannula inserted into the the
Eye -Then connected to a tube of
calibrated in cm of water or mercury(mmhg)
Disadvantage :- -Not clinical method - Needs Anaesthesia
Manometry
Indirect Tonometery
Principle:-
-Based on Palpation By the Examiner
Procedure:- -Pt looks Down -Index finger of both hand used one
finger push eye ball above tarsal plate that time other finger palpation the process.
Digital Tonometry
Advantage :- -Easiest -No equipment -No anaesthesia -No staining Disadvantage:- -Reading not proper -Only depend on examiner -Miner IOP can’t judged
properly
Indentation tonometer (Schiotz)Principle:- - Amount of the indentation depends upon IOP
-The higher IOP the lesser
Indentation -The lesser theIOP the deeper the indentation
-Less reading on scale ---- High IOP
-Scale reading have to be converted to IOP by the help of chart (Friedenwald conversion table)
HENCE
Plunger of tonometer Indenting the cornea
Base plate of tonometer resting on cornea
Fig.1 In this cornea is indented
PART OF SCHIOTZ TONOMETER
scale
Needle
Weight 5.5g plunger
Holder
Foot plate
lever
Additional weights 7.5,10,15g
Pseudocornea
Process :- -Patient in supine position - 4%lignocaine/proparacaine -looking up at a fixation targetAdvantage:- - Fast - Very good to camp screening
- Portable - Cheep - Easy to cleaning &
maintenance - No required Slit lamp /
electricity
Disadvantage :- - Reading effect by corneo scleral rigidity
- Can not be done in injury cases/early post-op
Cleaning& Mantanance:- - Cleaning after the removed from
barrel with
> Alcohal > Methylated spirt
Calibration:- This instrument should be calibrated before each use by placing on pseudo cornea and be sure checking scale reading is zero.
Applanation tonometer Principle:- -Applanation tonometer based on IMBERT FICK LAW
- Biprism tip flattens - A Biprism convert applantion area into half alignment of 2 half give the correct IOP reading
- Applanates a small area of cornea central 3.06mm.
F- ForceA- Flattening of the areaP- Pressure in side the Eye
IMBERT FICK LAW
Parts of Applanation tonometer
Biprism (measuring prism)Feeder arm
Housing
Adjusting knob
Connects to the slit lamp
Control weight insert
Equipment:- -Housing control the prism
movement-Neck-Biprism plastic Biprism with
a mirror to slit corneal applantion image to 2 halves
Process:--slit lamp-4% xylocaine -Dark room- Flurosceine stain of tear
film-Cobalt blue light -Biprism applanation cornea
with knobs reading at (1)-Reading read on knobs
scale multiplied by 10 ====IOP
Advantage:–
- Most accurate . -No indentation so not
much force applied on cornea. -Does not effect by
cornea scleral rigidity . -Reading directly read
from knob . -Can be done on post op case /injury cases .
Disadvantage :-
-Need slit lamp -Dark room .
- Need staining cobalt light
Calibration:-◦ Appl, tonometer is calibrated with calibration bar.
◦ It should be done once in a weekly or monthly depend upon uses.
NON CONTACT TONOMETER
Principle :-
A puff of air creates a constant force that momentarily flattens the cornea and internal reference point to the movement of flattening is measured and converted into IOP .Directly reading on display.
Fig:- NCT
Process:-
ADVANTAGE ;- -Quick . -No anesthetic required
. -May be delegated . -Nothing contact to the
Eye .
DISADVANTAGE :- -Expensive . -Difficult on to obtain reading
on scarred cornea , soft Eye -Not partable ..
Perkins TonometerIt uses the same biprism as the Goldmann
applanator.
The light source is powered by battery.
The readings are consistent and compare quite well with the Goldmann applanator.