To Obtain ECG of a Resting Human Subject and Verify the Algebraic Relationship
-
Upload
irfan-ali-khan -
Category
Documents
-
view
34 -
download
0
description
Transcript of To Obtain ECG of a Resting Human Subject and Verify the Algebraic Relationship
-
5/28/2018 To Obtain ECG of a Resting Human Subject and Verify the Algebraic Relationship
1/5
Applied Instrumentation Lab 8EI7
Global Technical Campus, Jaipur Page 1
Experiment No.7
Object: - To obtain ECG of a resting human subject and verify thealgebraic relationship amongst units leads.
Apparatus: - Heart ratemonitor cum ECG Simulator ST2351, ECG Electrodes, PC,RS 232 Cable, Oscilloscope, Banana Connector
Theory: - The heart is basically a hollow muscular pump, which pushes the bloodthroughout the body via the blood vessels. It is located between the lungs and slightly to
the left of center. The heart is an involuntary muscle that has approximately seventy toninety contractions per minute during a restful state. It begins to pump early in the life of
a foetus and will continue unceasingly until death. The heart wall is divided into three
layers: Pericardium, Myocardium and Endocardium. The heart is divided by a partition orseptum into two halves. The halves are in turn divided into chambers. The upper two
chambers of the heart are called atria and the lower two chambers are called ventricles.
Valves allow blood to flow in one direction between the chambers of the heart.The heart
has four distinct chambers.
1. Right atrium is the thin-walled area that receives the venous or "used" blood returning
to the body by the veins.
2. Right ventricle is the "pump" area of the heart's right side. The atrium dumps the blood
into the ventricle where it is then pumped out the pulmonary arteries and to the lungs.3. Left atrium receives the oxygenated blood returning from the lungs.
4. Left ventricle has the thickest walls of all. It is from this chamber the blood is pumped
out of the heart, into the aorta and out to the rest of the body.
Fig. 1.1HumanHeart Construction
-
5/28/2018 To Obtain ECG of a Resting Human Subject and Verify the Algebraic Relationship
2/5
Applied Instrumentation Lab 8EI7
Global Technical Campus, Jaipur Page 2
Heart valves:
1. Tricuspid valve is the one located at the entrance of the right ventricle. It prevents theblood from washing back into the right atrium.
2. Pulmonary semi lunar valve is located between the right ventricle and the pulmonaryartery.
3. Mitral valve is made of very heavy cusps and is located at the entrance of the leftventricle. This is a powerful valve that closes as the left ventricle begins each of its
contractions to ensure the oxygenated blood doesn't re-enter the left atrium.
4. Aortic valve is located, as its name would imply, between the left ventricle's exit andthe aorta itself.
Electo Cardio Gram (ECG): -As the heart undergoes depolarization and repolarization,the electrical currents that are generated spread not only within the heart, but alsothroughout the body. This electrical activity generated by the heart can be measured by
an array of electrodes placed on the body surface. The recorded tracing is called anElectrocardiogram (ECG, or EKG). A "typical" ECG tracing is shown below. Thedifferent waves that comprise the ECG represent the sequence of depolarization and
repolarization of the Atria and Ventricles.
Fig. 1.2ECG Wave form PQRST
The P wave represents the wave of depolarization that spreads from the SA node
throughout the atria, and is usually 0.08 to 0.1 seconds (80-100 ms) in duration.
The QRS complex represents ventricular depolarization. The duration of the QRS
complex is normally 0.06 to 0.1 seconds. This relatively short duration indicates that
ventricular depolarization normally occurs very rapidly.
-
5/28/2018 To Obtain ECG of a Resting Human Subject and Verify the Algebraic Relationship
3/5
Applied Instrumentation Lab 8EI7
Global Technical Campus, Jaipur Page 3
The T wave represents ventricular repolarization and is longer in duration than
depolarization (i.e., conduction of the repolarization wave is slower than the wave ofdepolarization). Typically the total time required for one complete cycle of the heart
electrical activity ranges from approximately 0.4 to 0.6 second. A healthy ECG shows anormal sinus rhythm. This is when each depolarization of the cardiac conduction system
creates a P wave, followed by a QRS complex, followed by a T wave. During a normalsinus rhythm, the atria are contracting first (around the time of the P wave), and the
ventricles contract second (around the time of the QRS complex). Ventricular relaxation
occurs around the time of the T wave.
Electrodes and Standards ECG Leads:- The Bio Electric signal generated on the
surface of body is picked up by using the electrodes. The electrodes make electricalconduction, which is necessary for making measurements. The surface electrodes picked
up potentials from the surface when placed over it without damaging the tissues; whereas
deep-seated electrode senses the potential difference arising inside the live tissue or cell.The most common types of electrodes routinely used for ECG recording are circular or
rectangular surface electrodes. An electrode conductive jelly is applied between the body
and the electrode to reduce contact impedance. Floating are suitable and effective for
long term monitoring they consists of Agcl3metal electrode and flat plastic washer whichdoes not allow metal to make contact with the body directly. This arrangement reduces
the effect of motion artifacts. Floating electrodes are placed on the chest with electrodes
conductive jelly and fixusing sticking plastics. The voltage generated by the pumping
action of the heart is actually a vector whose magnitude, as well as spatial orientation,changes with time. Because the ECG signal is measured from electrodes applied to the
surface of the body, the waveform of this signal is very dependent on the placement of
electrodes. To record the ECG pattern of a subject (human body) it is necessary to applyECG metal electrodes to the patient's limbs in special formats called leads, on each arm
and leg, and six electrodes are placed at defined locations on the chest. These electrode
leads are connected to a device that measures potential differences between selectedelectrodes to produce the characteristic electro-cardio-graphic tracings.
1. Limb Leads (Bipolar)2. Augmented Limb Leads (Unipolar)3. Chest Leads (Unipolar)
-
5/28/2018 To Obtain ECG of a Resting Human Subject and Verify the Algebraic Relationship
4/5
Applied Instrumentation Lab 8EI7
Global Technical Campus, Jaipur Page 4
Fig. 1.3Verify the algebraic relationship amongst units leads
Procedure: -
1. Connect the Heart-rate monitor cum ECG Trainer ST2351 to AC mains.2. Switch ON the trainer by Mains switch.3. Now place the electrodes in Lead I of standard Bipolar lead configuration
Positive electrode (yellow) in left arm of subject, negative electrodes (Red) in
right arm, and common (reference) electrode (Black) in the right leg of subject
properly with ECG Gel and connect the cables to the kit.
4. Observe the ECG waveform in Lead I configuration at test-point Amplifier block(TP3) of trainer. (Adjust gain of the amplifier if required). Or Make a serial
Connection between RS232 port on Training kit and RS232 port (COM1)
available on PC. And observe the waveforms and analyze them with Real time
ECG Analysis Software.
5. Then observe ECG waveforms at outputs of the different blocks of Heart-ratemonitor cum ECG Trainer ST2351.
-
5/28/2018 To Obtain ECG of a Resting Human Subject and Verify the Algebraic Relationship
5/5
Applied Instrumentation Lab 8EI7
Global Technical Campus, Jaipur Page 5
Fig. 1.4Real Time ECG Analysis Software
Heart-rate monitor cum ECG Trainer ST2351 provides in-depth study of
Electrocardiogram), ECG standard leads configuration, and the measurement of heart ratei.e. number of ECG pulses (heartbeats) per minute. ST2351 enables the
Electrocardiogram observation by affixing Biomedical Electrodes to the human body in
standard bipolar leads configuration. The Heart-rate monitor receives each ECG
observation (heartbeat) i.e. detects the electrical action executed by heart of subject(human body) and displays the heart-rate as number of heartbeats per minute. ST2351
also monitors the abnormalities occurs in human cardiovascular system tachycardia(faster heart-rate) and Bradycardia (slower heart-rate) using visible controls. It also
facilitates real time acquisition of raw data (ECG) and their corresponding measurementsby its real time Analysis Software.
Result: - Thus we have obtained Electro Cardio Gram (ECG) of a resting humansubject, verify the algebraic relationship amongst units leads and also monitor heart rate.