Therapeutic ultrasound: Personalised dose management Jean-François Aubry Institut Langevin (CNRS UMR 7587), INSERM U979, ESPCI ParisTech, Paris 7 University, France Invited Associate Prof, Univ of Virginia, Department of Radiation Oncology President of the International Society for Therapeutic Ultrasound (www.ISTU.org)

Transcript of Therapeutic ultrasound: Personaliseddose management Jean ... · Therapeutic ultrasound:...

Therapeutic ultrasound:Personalised dose management

Jean-François Aubry

Institut Langevin (CNRS UMR 7587), INSERM U979, ESPCI ParisTech, Paris 7 University, France

Invited Associate Prof, Univ of Virginia, Department of Radiation Oncology

President of the International Society for Therapeutic Ultrasound (www.ISTU.org)

Applications of Ultrasound in Medicine

• Diagnostic Ultrasound

• Therapeutic ultrasound:- Thrombolysis- Lithotripsy- Histotripsy- Bone repair- Drug delivery- Fat removal- Cancer treatments

Courtesy of the Focused Ultrasound Foundation

Therapeutic Ultrasound

-12MPaduring 1µs

skin

transducer Mechanical effectKidney stones destruction(Lithotripsy)

Biological effects of ultrasound

-12MPaduring 1µs

skin

transducer

Cavitation:Histotripsy

Biological effects of ultrasound

~60°Cduring 3s4MPa

skin

Normal Cells

Dead Cells

Frontier of the lesion

transducer

liver

Thermal effect:Tissue ablation

1mm

Biological effects of ultrasound

Example of MR-guided Therapeutic Ultrasound

Courtesy of Y. Medan and A. Hananel, Insightec

(Uterine Fibroid)

What does MR guidance provide?

• - 3D anatomic information for tumor

targeting

• - Beam path representation for safe

treatment

Example of MR-guided Therapeutic Ultrasound

Courtesy of Y. Medan and A. Hananel, Insightec

Example of MR-guided Therapeutic Ultrasound

Courtesy of Y. Medan and A. Hananel, Insightec

• - Post treatment T1 images

• (+ Gd injection)

• - Thermal Dose evaluation

Example of MR-guided Therapeutic Ultrasound

What does MR guidance provide?

43 ( ),43

0

tT t

eqt R dt-= ò {0.25 430.5 43

if T CR if T C< °= > °

Equivalent exposure time at 43°C(necrosis if >240eq min in muscle)

Courtesy of Y. Medan and A. Hananel, Insightec

Similar to radiation dose (D. Schlesinger et al, "Equivalence of cell survival data for radiation dose and thermal dose in ablative treatments“, Int. J. of Hyperthermia (2017)

MR-guided Feed back loop control

Temporal analysis of the temperature control with the minimum, mean, and maximum temperature of 195 voxels regulated at 14°C during 100 s.

Mougenot et al, Three-dimensional spatial and temporal temperature control with MR thermometry-guided focused ultrasound, MRM, 2009

MR-guided Feed back loop control

Profound Medical, real-time Magnetic Resonance Imaging with transurethral, robotically-driven therapeutic ultrasound and closed-loop thermal feedback control

MR-guided Feed back loop control

Profound Medical, real-time Magnetic Resonance Imaging with transurethral, robotically-driven therapeutic ultrasound and closed-loop thermal feedback control

5

Operator Console

Treatment Bed

Example of Ultrasound-guided FUS (USgFUS; or USgHIFU)

Haifu Model JC Focused Ultrasound Tumour Therapeutic System

Courtesy of J. Kennedy, F. Wu

Integrated Treatment Transducer (diagnostic US probe and HIFU transducer) in degassed water reservoir

Therapeutic frequency: 0.8MHz—2.4MHz

Courtesy of J. Kennedy, F. Wu

Example of Ultrasound-guided FUS (USgFUS; or USgHIFU)

Example of change seen on ultrasound image during treatment(Left = Before, Right = After)

Courtesy of J. Kennedy, F. Wu

Example of Ultrasound-guided FUS (USgFUS; or USgHIFU)

Ultrasonic Backscattering-Based Monitoring

Principle: detecting bubbles(cavitation or boiling)

CourtesyV. Khokhlova

Ultrasonic Thermometry

))()(1)(())(,( 0 zTzzczTzc db+=

zztczT ¶

¶-=D )()(2)(

0 dba

R. Seip et al, IEEE UFFC,1997C. Simon et al., IEEE UFFC 45(4),1998 Tissue dependent parameter

Ultrasonic speckle motion detection :

• Sound Speed variations– Apparent displacement of scatterers

• Thermal Expansion– Physical displacement of scatterers

a : thermal expansion coefficient

• Temperature versus scatterers displacement

(Bamber et al. 1979)

0>b0<b

In muscleIn fat

Principle of Ultrasonic Thermometry

Temperature Monitoring during HIFU

n HIFU Beam- 10 s- 1500 W.cm-2

n 1 image/s

20

22

24

26

28

30

32

20 40 60 80 100 120

10

20

30

40

50

60

70

80

90

100

110

21°C

32 °C

DT

M. Pernot, M. Tanter, J. Bercoff, K.R. Waters and M. Fink IEEE-UFFC, 2004, 51, pp 606-615

Limitations of US thermometry

n A priori Knowledge of :

n Motion Artifacts: scaterrer displacement can be hidden by tissue motionè High frame rate imaging can help to reduce motion artifacts

)(20ba-

c

Courtesy E. Ebbini

High sensitive cavitation detectionwith ultrafast imaging

Conventional Imaging Ultrafast Imaging

RAM

Parallel Processing

Principles of Ultrafast Ultrasonic Imaging

In vivo experiments on sheep brain

Trepanated sheep

Gateau et al, Phys. Med. Biol. 2011

In vivo experiments on sheep brain

Bmode image

Skull base

Trepanated sheep

Gateau et al, Phys. Med. Biol. 2011

In vivo experiments on sheep brain

Coherent subtractionNo nucleation Nucleation

Ultrafast images 550 µs after the high amplitude excitation

Gateau et al, Phys. Med. Biol. 2011

In vivo experiments on sheep brain

Coherent subtractionNo nucleation Nucleation

Ultrafast images 550 µs after the high amplitude excitation

Gateau et al, Phys. Med. Biol. 2011

In vivo experiments on sheep brain

Ultrafast imaging of bubble cloud formation

Currently impossible to image such events with MRI !

Ultrasound-based feedback loop control ?

20

22

24

26

28

30

32

20 40 60 80 100 120

10

20

30

40

50

60

70

80

90

100

110

20

3 mm

0

T (°C)

US- thermal map

Not accurrate enough for in vivo feedback control

Ultrasound-based feedback loop control ?

But cavitation can be used as a feedback control

Ultrasound-based feedback loop control ?

But cavitation can be used as a feedback control

A.Poizat,Phd,UniversitédeLyon,2016

Cavitation dose

Frequency

No cavitationCavitation (stable)

Ultrasound-based feedback loop control ?

But cavitation can be used as a feedback control

A.Poizat,Phd,UniversitédeLyon,2016

Ultrasound-based feedback loop control ?

But cavitation can be used as a feedback control

Gerold etal,Acoustic cavitationsub-millisecond signalprocessing forultrasound surgery ,JASA,2016

Currently being implementedon the EchoPuse®

(Theraclion, Malakof, France)

MR monitoring

Take home message

Real time thermal mapping

Possibility of feedback loop control

US monitoring

Real time cavitation mapping

Possibility of feedback loop control