The value of microscopy analysis
Garry BurdettHealth and Safety Laboratory
Harpur Hill, Buxton, UK, SK17 9JN
![Page 2: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/2.jpg)
Asbestos Hazard
• Mortality (excess deaths) due to:– Fibrosis of the lung (asbestosis)– Cancer of the lung (bronchial cancer etc)– Cancer of the lung lining (e.g. mesothelioma)
• Morbidity (lung changes/dysfunction)– Pleural plaques;– Diffuse pleural effusions
![Page 3: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/3.jpg)
Fibre Related Risk:
• Depends on:– Dimensions– Durability / Type– Dose / Exposure– Time
(Measurable by microscopy)
![Page 4: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/4.jpg)
EU Directive: Dimensions
• For the purposes of measuring in the air, only fibres with a length of more than five micrometres, a breadth of less than three micrometres and a length to breadth ratio greater than 3:1 shall be taken into consideration.
(2003/18/EC amending to EU asbestos worker protection directive (83/477/EEC) article 7.6 )
![Page 5: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/5.jpg)
Durability / Type
500
100
1
0
100
200
300
400
500
Ris
k fa
cto
r
Crocidolite Amosite Chrysotile
Asbestos type
Fig 1: Mesothelioma risk by fibre type
![Page 6: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/6.jpg)
Dose / Exposure
• Exposure is used as a surrogate for dose by sampling a known volume of air onto a membrane filter and counting the numbers of fibres present in a known area.
• This is used to calculate the fibre concentration in f/ml or f/cm3 of air.
• Uses size criteria to determine ability to reach pulmonary region of the lung.
![Page 7: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/7.jpg)
Time
• Dose = exposure x duration (f/ml/years)
• Risk cumulative dose (assume linear);
• Risk time since first exposure (mesothelioma related by power law).
• Lag time 15 – 60 years from exposure to disease symptoms
![Page 8: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/8.jpg)
EU directive: Fibre Counting
• “Fibre counting shall be carried out whenever possible by PCM (Phase contrast microscope) in accordance with the 1997 (World Health Organisation) recommended method (16) or any other method giving equivalent results.”
![Page 9: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/9.jpg)
EU Directive: Requirements
• The 2003 amendment to 83/477/EEC abandoned action levels and different control levels for different types of asbestos and introduced a single limit of 0.1 f/ml over an 8-hour time weighted average. (Article 8)
• Therefore the EU regulation for controlling the risk to workers, require only PCM fibre number information over a broadly defined size range.
• Article 4 requires the type and quantities of asbestos handled to be notified
![Page 10: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/10.jpg)
Main types of microscopy for fibre counting.
• Phase contrast light microscopy PCM
• Scanning electron microscopy SEM
• Transmission electron microscopy TEM.
• All have ISO or WHO standard methods
• Polarised light microscopy PLM for bulk analysis.
![Page 11: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/11.jpg)
![Page 12: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/12.jpg)
PCM (X500 magnification)
![Page 13: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/13.jpg)
SEM
![Page 14: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/14.jpg)
TEM
![Page 15: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/15.jpg)
SEM higher Mag & 3D
![Page 16: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/16.jpg)
TEM high mag chrystotile
![Page 17: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/17.jpg)
PLM : Dispersion staining
![Page 18: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/18.jpg)
Discrimination of fibre type
• PCM: No discrimination or identification ;• PLM Uses optical properties + refractive
index/ dispersion colours to identify fibres; >0.8µm width;
• SEM uses Energy Dispersive X-ray analysis to semi-quantitatively measure chemistry;
• TEM uses quantitative EDX and electron diffraction to identify fibres.
![Page 19: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/19.jpg)
![Page 20: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/20.jpg)
Comparison of methods for fibre counting
Operating
Mag
Area (mm2) examined
Time taken
(mins.)
Rate area examinedmm2/hr
PCM X500 (200 fov) 1.5
20 4.5
SEM X2000 (50 fov)
0.15
30 0.3
TEM x5000 (50 go)
0.5
40 0.75
![Page 21: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/21.jpg)
Likely upper limit of detection (f/ml) by method and volume
WORK ENVIRONMENTAL
COST PER HOUR
VOLUME
(litres)
500 3000
PCM 0.01* 0.0017* 30-50
SEM 0.015 0.0025 80 - 120
TEM 0.005 0.0007 80 - 120
![Page 22: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/22.jpg)
Accuracy & Precision
• Accuracy not usually known.
• Precision includes:– Random “Poisson” counting error– Instrumental errors (QA / calibration)– Human errors (QA / Proficiency testing)
![Page 23: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/23.jpg)
Random counting errors (95%)
No. of fibres
Lower limit
Upper limit
No. of fibres
Lower limit
Upper limit
0 0 2.99 5 1.624 11.669
1 0.025 5.572 6 2.202 13.060
2 0.242 7.225 10 4.795 18.39
3 0.619 8.767 20 12.217 30.889
4 1.090 10.242 50 37.112 65.919
![Page 24: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/24.jpg)
Quality and accreditation
• Quality assurance and proficiency testing is essential, particularly if results from different analysts and laboratories are being used or compared.
• Accreditation of Laboratories to ISO 17025 mandated by the EU directive.
![Page 25: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/25.jpg)
Type of analysis
PCM SEM TEM PLM
Fibre counting of air samples
Belgium (25)France (?)Netherlands(20)Spain (20)UK (200)
Germany*
France (20) Not applicable
Fibre Identification of bulk samples
Not applicable UK (5) UK (17) UK (250)
Fibre counting and identification of air samples
Not applicable Germany*
France (20) Not applicable
* PT scheme run for one or two rounds in past a new scheme was plannedNote: UK schemes also include a number of EU countries who participate
![Page 26: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/26.jpg)
RICE: Fibre counting PT
• RICE - 187 labs.
• MMMF – 12 labs.
• ERM – WHO major changeChart 1. RICE Ratings
196 197 195 192 190 193 194 186 189 190 184 184 183 184 184
0
50
100
150
200
250
52 53 54 55 56 57 58 59 60 61 62 63 64 65 66
Round
Nu
mb
ers
of
lab
ora
tori
es
Rating 1
Rating 2
Rating 3
![Page 27: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/27.jpg)
AIMS: Fibre Identification PT
Chart 4 : Distribution of cumulative scores for rounds 22 - 24 (UK Laboratories)
25%
33%
19%
22%
1%
0 (No Errors)
7 (1 Minor Error)
8 - 20
21 - 47
> 47
![Page 28: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/28.jpg)
Value of PCM counting
1. Workplace exposure can be controlled using the WHO - PCM method;
2. Counts “regulated” WHO fibres & meets EU criteria;
3. Relatively quick (on-site) and cheap;
4. Environmental exposure can be screened by PCM.
![Page 29: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/29.jpg)
PCM Limitations
1. PCM has limited detection due to background counts on blank filters (0.01 f/ml);
2. No knowledge of fibre types;
3. Only an “index of exposure”
![Page 30: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/30.jpg)
Value of analytical SEM
• Can be adjusted to give regulated fibre counts,
• Can give fibre size information,
• Classifies fibre types present using EDX spectra data.
![Page 31: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/31.jpg)
SEM limitations
• Limited availability;• Off site analysis;• Costly • Can only be used for regulated fibre
counting,• No definitive identification of fibres,• Lowest area of filter scanned per field of
view.
![Page 32: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/32.jpg)
Value of TEM
• Can count regulated fibres + all fibre sizes.
• Accurate size information,
• EDX chemical analysis of all fibre sizes
• Crystal structure information (SAED)
• Identification of fibre type,
• Can give complete information on all fibre sizes if needed
![Page 33: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/33.jpg)
Limitations of TEM
• Limited availability;
• Off site;
• Costly
![Page 34: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/34.jpg)
Recommendations 1
• Use PCM for workplace control and checking after clean –up (with disturbance sampling).
• Indoor and near source environmental use PCM screening and SEM or TEM analysis (as necessary) on selected half-filters.
• Ambient air: Analytical TEM is best, SEM is more limited, PCM not very useful.
![Page 35: The value of microscopy analysis Garry Burdett Health and Safety Laboratory Harpur Hill, Buxton, UK, SK17 9JN.](https://reader036.fdocuments.us/reader036/viewer/2022062322/56649e545503460f94b4a936/html5/thumbnails/35.jpg)
Recommendations 2
• Use Polarised light microscopy (PLM) for analysis of bulk samples for asbestos present and type of asbestos.
• Use PLM to screen minerals and soils for asbestos content.
• Asbestos in water usually analysed by TEM