THE USE OF A TRIPLE DETECTION SYSTEM (UV, … be formed, including those found in forced degradation...

1
TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2016 Waters Corporation PDA Spectrum (210-400)nm 0.381 - 374.23 0.546 - 432.21 2.958 - 513.27 AU 0.00 0.10 0.20 0.30 QDa Positive Scan CODA SM 5 MCQ 0.75 (QDa Positive(+) Scan (100.00-600.00)Da, Centroid, CV=5) Intensity 0 2x10 6 4x10 6 6x10 6 GlimepirideB XIC - 0.381 - QDa Positive Scan: 374.23 m/z GlimepirideC XIC - 0.546 - QDa Positive Scan: 432.21 m/z Glimepiride XIC - 2.958 - QDa Positive Scan: 513.27 m/z Intensity 0 2x10 6 4x10 6 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 Calibration Plot Name: Glimepiride C; Fit Type: Linear (1st Order); R^2: 0.999140; Equation Y = 1.67e+000 X + 2.14e+000 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 LogAmount 0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 Calibration Plot Name: Glimepiride C; Fit Type: Quadratic (2nd Order); R^2: 0.996621; Equation Y = 1.62e+001 X^2 + 1.65e+003 X - 2.00e+004 -2.0x10 5 0.0 2.0x10 5 4.0x10 5 6.0x10 5 8.0x10 5 1.0x10 6 1.2x10 6 1.4x10 6 1.6x10 6 Amount 0.00 20.00 40.00 60.00 80.00 100.00 120.00 140.00 160.00 180.00 200.00 220.00 240.00 INTRODUCTION In the pharmaceutical industry, there is a need to fully understand the stability of an active pharmaceutical ingredient (API) and characterize any impurities that may be formed, including those found in forced degradation studies. Reversed-phase liquid chromatography UV-based techniques are often used for these types of analyses. However, separating and detecting the related impurities and other components can be challenging. By combining a UV detector, an evaporative light scattering detector, and a mass spectrometer, it is possible to detect compounds with different chemical properties. In the following study, we will conduct a forced degradation study which will also include mass balance studies using triple detection. Mass balance can be affected by the differences in response factors of the API and related impurities. We will perform two sets of experiments to determine the relative response factors (RRF) of the impurities. This process will include either: 1) using the ratio of the slopes of the API and related impurities or 2) comparing the UV peak area to the log of the ELSD peak area, the latter which has a response based on unit mass. Using the relative response factors we will then evaluate the mass balance of the acid hydrolysis of glimepiride. In addition, the QDa mass detector will provide added information for peak confirmation and peak purity. THE USE OF A TRIPLE DETECTION SYSTEM (UV, ELSD, MS) FOR PHARMACEUTICAL DEGRADATION STUDIES Paula Hong , Aaron Phoebe, and Patricia R. McConville Waters Corporation, 34 Maple Street, Milford, MA, USA, 01757 Figure 1. ACQUITY UPLC H-Class system with triple detection including ACQUITY PDA, ELSD and QDa detectors. The triple detection system includes an isocratic solvent manager (ISM) which provides make-up solvent to the QDa detector and houses the splitter required for the ELSD and the QDa. All of the flow is directed to the splitter from the PDA detector. A portion is then sent to the ELSD detector. Make-up solvent is introduced in the next port and the flow is then mixed as it flows into the QDa or mass detector. The composition and flow rate of the make-up solvent impact the split ratio to the ELSD and the QDa. METHODS Conditions System: ACQUITY UPLC H-Class with Column Manager Column: ACQUITY UPLC BEH C18, 1.7 μm, 2.1 x 50 mm Mobile phase A: 0.1% (v/v) Formic acid in Water Mobile phase B: 0.1% (v/v) Formic acid in Acetonitrile Column Temperature: 30 C Injection volume: 2 uL Flow rate: 0.8 mL/min Isocratic:60% A: 40%B Gradient: ACQUITY PDA Detector Wavelength range: 210-400 nm Resolution: 3.6 nm Selected wavelengths: 228 nm, 4.8 nm resolution Time Constant: Normal Sampling rate: 20 pts/s ACQUITY Evaporative Light Scattering Detector (ELSD) Gas: 25 psi Neubilizer Mode: Cooling Nebulizer Temperature: 55 ˚C Gain: 350 Data rate: 10 pps Isocratic Solvent Manager Flow rate: 0.3 mL/min Solvent: 0.1% (v/v) formic acid in methanol ACQUITY QDa Detector (mass detector) Mass range: 100-600 Cone voltage: 5V Sampling Rate: 5 pps Capillary Voltage: 1.4 kV Sample Preparation: Glimepiride and related compounds B and C were purchased from the USP. All standards were dissolved in 55:45 methanol:water and sonicated. The drug substance glimepiride was obtained from an outside source. Acid hydrolysis was conducted at 40 ˚C for 0-7 days . The concentration of acid was 0.1M HCl in the degradation reaction. References 1. Chapter <621> CHROMATOGRAPHY United States Pharmacopeia and National Formulary (USP 37-NF 32 S1) Baltimore, MD: United Book Press, Inc.; 2014. p. 6376-85. 2. Mark AN, Andreas K, Patrick JJ. Role of Mass Balance in pharmaceutical stress testing. Pharmaceutical Stress Testing: CRC Press; 2011. p. 233-53. 3. Bansal G, Singh M, Jindal KC, Singh S. LCUVPDA and LCMS studies to characterize degradation products of glimepiride. Journal of pharmaceutical and biomedical analysis. 2008; 48:788-95. Ensuring Peak Purity using Mass and UV detection CONCLUSIONS Triple detection system in combination with Empower 3 FR2 provides various tools to assist in mass balance, including: Determination of relative response factors by using the ratio of UV peak and the log of ELSD peak responses The ability to input relative response values into Empower 3 FR2 to determine corrected area values for impurities for mass balance determinations Using orthogonal detection (UV and mass) to confirm peak purity and to detect the presence of co-elutions that could impact mass balance. Mass Balance for Forced Degradation Studies Time %A %B %C %D Curve Initial 95.0 5.0 0.0 0.0 Initial 5.00 5.0 95.0 0.0 0.0 6 6.50 5.0 95.0 0.0 0.0 6 6.51 95.0 5.0 0.0 0.0 6 Samples Conditions API Control RT and 40°C for 1,3,5, 7 days Acid control RT and 40°C for 1,3,5, 7 days Acid Hydrolysis RT and 40°C for 1,3,5, 7 days Table 3. Mass balance determinations for forced degradation of glimepiride. The calculations were performed using RRF de- termined with the ELSD method. The RRF were entered into Empower 3 FR 2 for corrected values of the related impurities. All mass balance values were within 2%. AU 0.00 0.20 0.40 374.3 432.2 513.2 AU 0.00 0.20 0.40 AU 0.00 0.10 0.20 0.30 0.40 0.50 Minutes 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 t=0 t= 3 days t= 5 days Figure 7. UV chromatograms of forced degradation of glimepiride drug substance with base mass labels. The drug substance was exposed to acidic hydrolysis conditions at 40 C over a period of days. Over the course of the study the two impurity peaks ( related compound C and B) increased in peak area. Figure 2. Forced degradation of glimepiride drug substance. AU 0.00 0.05 0.10 0.15 Minutes 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 LSU 0.00 10.00 20.00 Minutes 1.00 2.00 3.00 4.00 5.00 Intensity 0.0 5.0x10 5 1.0x10 6 1.5x10 6 2.0x10 6 Minutes 1.00 2.00 3.00 4.00 5.00 PDA @ 228 nm ELSD MS XIC ESI+ Figure 3. Separation of standards of active pharmaceutical ingredient (API), related compound B and related compound C under isocratic conditions. The overlay of standards at 250 μg/mL for the API and 10 μg/mL for related compounds B and C shows the differences in relative response among the detectors. The UV and ELSD give similar relative response for the three compounds. In the mass detector related compound C has a greater peak area than related compound B. Figure 5. Overlay of glimepiride related compound C standards (10-250 μg/mL) in PDA and ELSD. The UV detector produces a linear response for standards. Evaluating the peak areas in the ELSD, a non-linear or logarithmic response is observed. For example, at 10 μg/mL the response in the ELSD (pink trace) is significantly lower than that observed in the PDA. Wavelength Separation Conditions RRF Rel Compound B RRF Rel Compound C 228 Isocratic 1.36 1.10 Name: GlimepirideB; Fit Type: Linear (1st Order); R^2: 0.999945; Equation Y = 1.06e+004 X - 1.78e+001 Name: Glimepiride; Fit Type: Linear (1st Order); R^2: 0.999469; Equation Y = 7.72e+003 X - 1.56e+003 Name: Glimepiride C; Fit Type: Linear (1st Order); R^2: 0.998663; Equation Y = 8.86e+003 X - 1.76e+003 -4.0x10 5 -2.0x10 5 0.0 2.0x10 5 4.0x10 5 6.0x10 5 8.0x10 5 1.0x10 6 1.2x10 6 1.4x10 6 1.6x10 6 1.8x10 6 2.0x10 6 2.2x10 6 ppm 0.00 20.00 40.00 60.00 80.00 100.00 120.00 140.00 160.00 Area -2.0x10 5 0.0 2.0x10 5 4.0x10 5 6.0x10 5 8.0x10 5 1.0x10 6 1.2x10 6 ppm 0.00 10.00 20.00 30.00 40.00 50.00 60.00 70.00 Figure 4. Overlay of UV linearity curves at 228nm of glimepiride and related impurities B and C. The API linearity curve covers the range of 1-250 μg/mL for the API and 1-50 μg/mL for related compounds B and C. All R 2 values were >0.998. AU 0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 Minutes 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 PDA @ 228 nm 250 μg/mL 125 μg/mL 50 μg/mL 10 μg/mL LSU 0.00 100.00 200.00 300.00 400.00 500.00 600.00 700.00 800.00 900.00 Minutes 0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00 250 μg/mL 125 μg/mL 50 μg/mL 10 μg/mL ELSD Figure 6. ELSD calibration curves for glimepiride related compound C. The ELS detector has a quadratic fit to the calibration curve (top) for peak area vs. the amount. If the values are converted to the logarithmic functions (inset), the calibration curve fit is linear (bottom). The R 2 value for this curve is 0.999140. Standard RRF Rel Compound B RRF Rel Compound C 50 1.25 1.10 125 1.24 1.09 ELSD calibration curve Quadratic fit ELSD calibration curve Linear fit Table 1. Relative Response Factors for Related compound B and C using the ratio of the slope of the API/slope of the impurity. The value for related compound B is outside of 0.8-1.2 range and, therefore, should be applied, as specified by the USP Chapter <621>. 1 RESULTS AND DISCUSSION Multi-detection of API and Related Compounds Determination of Relative Response Factors Using Slope of Calibration Curves in UV = [ ] [ ] Determination of Relative Response Factors Using Ratio of UV to ELSD (log) Peak Area Y-axis X-axis LogArea LogAmount Table 2. Relative Response Factors for related compound B and C using the ratio of the UV peak area to the log of the ELSD peak area. RRF can be calculated using the response of the UV detector to a mass concentration dependent detector. 2 This assumes a linear relationship for both detectors. To convert the ELSD calibration response to a linear function, the log of both x and y values can be used. Thereby, using the log of the ESLD peak area, we can calculate RRF factors for both impurities. These values have good correlation with those obtained using the slopes of the calibration curves in the UV. = log ( / = / n = 3 Reference T=0 1 day 3 days 5 days 7 days Amount 238.4 236.1 239.0 242.7 239.1 244.1 % Recovery 99.0 100.2 101.8 100.3 102.4 Figure 8. Mass analysis window in Empower 3 FR2 of forced degradation at 3 days. In forced degradation studies, mass imbalance can be caused by co-elutions of the API and any impurities. To evaluate peak purity, mass detection can provide valuable information. In this view, the UV, Total Ion Chromatogram (TIC) and Extracted Ion Chromatograms (XIC) are shown in a stacked plot. For each integrated peak in the UV chromatogram the leading, apex and trailing portion of the peak can be viewed (top portion of figure). Using the mass and UV detectors the peak purity can be evaluated to ensure no co- elutions. Pre-configured Splitter in the Isocratic Solvent Manager ACQUITY UPLC H-Class System with Column Managers and Triple Detection (UV-ELSD-PDA) API Rel Cmpd B Rel Cmpd C API Rel Cmpd B Rel Cmpd C GlimepirideC - 0.546 - QDa Positive Scan Leading Apex Trailing 228.3 294.5337.2359.0 228.3 228.3 300.1 316.7355.2 432.26 432.21 432.30 UV TIC, ESI+ XIC, ESI+ Peak purity view of glimepiride related compound C + HCl Δ Flow diagram

Transcript of THE USE OF A TRIPLE DETECTION SYSTEM (UV, … be formed, including those found in forced degradation...

Page 1: THE USE OF A TRIPLE DETECTION SYSTEM (UV, … be formed, including those found in forced degradation studies. Reversed-phase liquid ... Mass balance determinations for forced degradation

TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2016 Waters Corporation

Project Name: Mass Balance Degrad Nov15Reported by User: System

Report Method: Untitled Date Printed:

125 12/1/2015Report Method ID: 125

11:17:57 AM US/Eastern

GlimepirideB - 0.381 - QDa Positive Scan

Leading Apex Trailing

222.8

300.1339.0381.3

226.5 226.5

298.2 360.8

374.05

491.19

514.96

374.23 374.12

374.72

432.44

GlimepirideC - 0.546 - QDa Positive Scan

Leading Apex Trailing

228.3

294.5337.2359.0

228.3 228.3

300.1316.7355.2

432.26 432.21 432.30

PDA Spectrum (210-400)nm

0.3

81

- 3

74

.23

0.5

46

- 4

32

.21

2.9

58

- 5

13

.27

AU

0.00

0.10

0.20

0.30

QDa Positive Scan CODA SM 5 MCQ 0.75 (QDa Positive(+) Scan (100.00-600.00)Da, Centroid, CV=5)

Inte

ns

ity

0

2x106

4x106

6x106

GlimepirideB XIC - 0.381 - QDa Positive Scan: 374.23 m/z

GlimepirideC XIC - 0.546 - QDa Positive Scan: 432.21 m/z

Glimepiride XIC - 2.958 - QDa Positive Scan: 513.27 m/z

Inte

ns

ity

0

2x106

4x106

Minutes

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00

Calibration curve ELSD stack

Project Name: MassBalance Glimep 14Oct15Reported by User: System

Report Method: Calibration curve ELSD stack Date Printed:

24851 2/26/2016Report Method ID: 24851

9:49:19 AM US/Eastern

Calibration Plot

Name: Glimepiride C; Fit Type: Linear (1st Order); R^2: 0.999140; Equation Y = 1.67e+000 X + 2.14e+000

1.50

2.00

2.50

3.00

3.50

4.00

4.50

5.00

5.50

6.00

6.50

7.00

LogAmount

0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

Point Information ' Peak: GlimepirideB' table contains no data.

1

2

3

4

5

6

7

8

9

10

Name Level X Value Response Calc. Value % Deviation Manual Ignore

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

1.000000

1.000000

1.000000

1.698970

1.698970

1.698970

2.096910

2.096910

2.096910

2.397940

3.838315

3.805043

3.820003

4.921575

4.942371

4.964010

5.629349

5.648075

5.678314

6.122215

1.018281

0.998353

1.007314

1.667098

1.679553

1.692515

2.091018

2.102234

2.120345

2.386219

1.828

-0.165

0.731

-1.876

-1.143

-0.380

-0.281

0.254

1.118

-0.489

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

Peak: Glimepiride C

Calibration curve ELSD stack

Project Name: MassBalance Glimep 14Oct15Reported by User: System

Report Method: Calibration curve ELSD stack Date Printed:

24851 2/26/2016Report Method ID: 24851

9:50:31 AM US/Eastern

Calibration Plot

Name: Glimepiride C; Fit Type: Quadratic (2nd Order); R^2: 0.996621; Equation Y = 1.62e+001 X̂ 2 + 1.65e+003 X - 2.00e+004

-2.0x105

0.0

2.0x105

4.0x105

6.0x105

8.0x105

1.0x106

1.2x106

1.4x106

1.6x106

Amount

0.00 20.00 40.00 60.00 80.00 100.00 120.00 140.00 160.00 180.00 200.00 220.00 240.00

Point Information ' Peak: GlimepirideB' table contains no data.

1

2

3

4

5

6

7

8

9

10

Name Level X Value Response Calc. Value % Deviation Manual Ignore

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

Glimepiride C

10.000000

10.000000

10.000000

50.000000

50.000000

50.000000

125.000000

125.000000

125.000000

250.000000

6891.517641

6383.265905

6606.982471

83478.507006

87573.059729

92047.180381

425941.030337

444707.880031

476775.136351

1324998.044518

14.265562

14.025072

14.131038

43.764007

45.086203

46.510456

122.439136

125.740811

131.244015

241.342736

42.656

40.251

41.310

-12.472

-9.828

-6.979

-2.049

0.593

4.995

-3.463

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

Peak: Glimepiride C

INTRODUCTION

In the pharmaceutical industry, there is a need to fully

understand the stability of an active pharmaceutical ingredient (API) and characterize any impurities that

may be formed, including those found in forced degradation studies. Reversed-phase liquid

chromatography UV-based techniques are often used for these types of analyses. However, separating and

detecting the related impurities and other components can be challenging. By combining a UV detector, an

evaporative light scattering detector, and a mass

spectrometer, it is possible to detect compounds with different chemical properties.

In the following study, we will conduct a forced

degradation study which will also include mass balance studies using triple detection. Mass balance can be

affected by the differences in response factors of the API and related impurities. We will perform two sets of

experiments to determine the relative response factors (RRF) of the impurities. This process will include either:

1) using the ratio of the slopes of the API and related

impurities or 2) comparing the UV peak area to the log of the ELSD peak area, the latter which has a response

based on unit mass. Using the relative response factors we will then evaluate the mass balance of the acid

hydrolysis of glimepiride. In addition, the QDa mass detector will provide added information for peak

confirmation and peak purity.

THE USE OF A TRIPLE DETECTION SYSTEM (UV, ELSD, MS) FOR PHARMACEUTICAL DEGRADATION STUDIES

Paula Hong, Aaron Phoebe, and Patricia R. McConville

Waters Corporation, 34 Maple Street, Milford, MA, USA, 01757

Figure 1. ACQUITY UPLC H-Class system with triple detection

including ACQUITY PDA, ELSD and QDa detectors. The triple

detection system includes an isocratic solvent manager (ISM)

which provides make-up solvent to the QDa detector and houses

the splitter required for the ELSD and the QDa. All of the flow is

directed to the splitter from the PDA detector. A portion is then

sent to the ELSD detector. Make-up solvent is introduced in the

next port and the flow is then mixed as it flows into the QDa or

mass detector. The composition and flow rate of the make-up

solvent impact the split ratio to the ELSD and the QDa.

METHODS

Conditions

System: ACQUITY UPLC H-Class with Column Manager

Column: ACQUITY UPLC BEH C18, 1.7 µm, 2.1 x 50 mm

Mobile phase A: 0.1% (v/v) Formic acid in Water

Mobile phase B: 0.1% (v/v) Formic acid in Acetonitrile

Column Temperature: 30 C

Injection volume: 2 uL

Flow rate: 0.8 mL/min

Isocratic:60% A: 40%B

Gradient:

ACQUITY PDA Detector

Wavelength range: 210-400 nm

Resolution: 3.6 nm

Selected wavelengths: 228 nm, 4.8 nm resolution

Time Constant: Normal

Sampling rate: 20 pts/s

ACQUITY Evaporative Light Scattering Detector (ELSD)

Gas: 25 psi

Neubilizer Mode: Cooling

Nebulizer Temperature: 55 ˚C

Gain: 350

Data rate: 10 pps

Isocratic Solvent Manager

Flow rate: 0.3 mL/min

Solvent: 0.1% (v/v) formic acid in methanol

ACQUITY QDa Detector (mass detector)

Mass range: 100-600

Cone voltage: 5V

Sampling Rate: 5 pps

Capillary Voltage: 1.4 kV

Sample Preparation:

Glimepiride and related compounds B and C were purchased from the USP.

All standards were dissolved in 55:45 methanol:water and sonicated.

The drug substance glimepiride was obtained from an outside source. Acid

hydrolysis was conducted at 40 ˚C for 0-7 days . The concentration of acid

was 0.1M HCl in the degradation reaction.

References

1. Chapter <621> CHROMATOGRAPHY United States Pharmacopeia and National Formulary (USP 37-NF 32

S1) Baltimore, MD: United Book Press, Inc.; 2014. p. 6376-85.

2. Mark AN, Andreas K, Patrick JJ. Role of Mass Balance in pharmaceutical stress testing. Pharmaceutical

Stress Testing: CRC Press; 2011. p. 233-53.

3. Bansal G, Singh M, Jindal KC, Singh S. LC–UV–PDA and LC–MS studies to characterize degradation

products of glimepiride. Journal of pharmaceutical and biomedical analysis. 2008; 48:788-95.

Ensuring Peak Purity using Mass and UV detection

CONCLUSIONS

Triple detection system in combination with Empower

3 FR2 provides various tools to assist in mass

balance, including:

Determination of relative response factors by using

the ratio of UV peak and the log of ELSD peak responses

The ability to input relative response values into

Empower 3 FR2 to determine corrected area values for impurities for mass balance determinations

Using orthogonal detection (UV and mass) to

confirm peak purity and to detect the presence of

co-elutions that could impact mass balance.

Mass Balance for Forced Degradation Studies

Time %A %B %C %D Curve

Initial 95.0 5.0 0.0 0.0 Initial

5.00 5.0 95.0 0.0 0.0 6

6.50 5.0 95.0 0.0 0.0 6

6.51 95.0 5.0 0.0 0.0 6

Samples Conditions

API Control RT and 40°C for 1,3,5, 7 days

Acid control RT and 40°C for 1,3,5, 7 days

Acid Hydrolysis RT and 40°C for 1,3,5, 7 days

Table 3. Mass balance determinations for forced degradation

of glimepiride. The calculations were performed using RRF de-termined with the ELSD method. The RRF were entered into

Empower 3 FR 2 for corrected values of the related impurities. All mass balance values were within 2%.

AU

0.00

0.20

0.40

374.3 432.2

513.2

AU

0.00

0.20

0.40

AU

0.00

0.10

0.20

0.30

0.40

0.50

Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50

t=0

t= 3 days

t= 5 days

Figure 7. UV chromatograms of forced degradation of glimepiride

drug substance with base mass labels. The drug substance was exposed to acidic hydrolysis conditions at 40 C over a period of

days. Over the course of the study the two impurity peaks ( related compound C and B) increased in peak area.

Figure 2. Forced degradation of glimepiride drug substance.

AU

0.00

0.05

0.10

0.15

Minutes

0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00

LS

U

0.00

10.00

20.00

Minutes

1.00 2.00 3.00 4.00 5.00

Inte

nsity

0.0

5.0x105

1.0x106

1.5x106

2.0x106

Minutes

1.00 2.00 3.00 4.00 5.00

PDA @ 228 nm

ELSD

MS XIC

ESI+

Figure 3. Separation of standards of active pharmaceutical ingredient (API), related compound B and related compound C under

isocratic conditions. The overlay of standards at 250 µg/mL for the

API and 10 µg/mL for related compounds B and C shows the

differences in relative response among the detectors. The UV and ELSD give similar relative response for the three compounds. In the mass detector related compound C has a greater peak area than

related compound B.

Figure 5. Overlay of glimepiride related compound C standards

(10-250 µg/mL) in PDA and ELSD. The UV detector produces a

linear response for standards. Evaluating the peak areas in the

ELSD, a non-linear or logarithmic response is observed. For

example, at 10 µg/mL the response in the ELSD (pink trace) is

significantly lower than that observed in the PDA.

Wavelength

Separation

Conditions

RRF Rel

Compound B

RRF Rel

Compound C

228 Isocratic 1.36 1.10

Calibration curve

Project Name: MassBalance Glimep 14Oct15Reported by User: System

Report Method: Calibration curve Date Printed:

18823 12/2/2015Report Method ID: 18823

1:42:05 PM US/Eastern

Calibration Plot

Name: GlimepirideB; Fit Type: Linear (1st Order); R^2: 0.999945; Equation Y = 1.06e+004 X - 1.78e+001

Name: Glimepiride; Fit Type: Linear (1st Order); R^2: 0.999469; Equation Y = 7.72e+003 X - 1.56e+003

Name: Glimepiride C; Fit Type: Linear (1st Order); R^2: 0.998663; Equation Y = 8.86e+003 X - 1.76e+003

Are

a

-4.0x105

-2.0x105

0.0

2.0x105

4.0x105

6.0x105

8.0x105

1.0x106

1.2x106

1.4x106

1.6x106

1.8x106

2.0x106

2.2x106

ppm

0.00 20.00 40.00 60.00 80.00 100.00 120.00 140.00 160.00 180.00 200.00 220.00 240.00

1

2

3

4

5

6

7

8

9

10

11

12

Name Level X Value Response Calc. Value % Deviation Manual Ignore

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

1.000000

1.000000

1.000000

5.000000

5.000000

5.000000

10.000000

10.000000

10.000000

50.000000

50.000000

50.000000

10427.014939

10322.367898

10382.729387

53820.208641

54051.671831

53601.392108

106798.664119

107004.997494

107492.919871

529969.685196

529309.806651

528377.102249

0.983969

0.974110

0.979797

5.071877

5.093682

5.051263

10.062776

10.082214

10.128179

49.928110

49.865945

49.778079

-1.603

-2.589

-2.020

1.438

1.874

1.025

0.628

0.822

1.282

-0.144

-0.268

-0.444

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

Peak: GlimepirideB

Calibration curve

Project Name: MassBalance Glimep 14Oct15Reported by User: System

Report Method: Calibration curve Date Printed:

18823 12/2/2015Report Method ID: 18823

1:45:01 PM US/Eastern

Calibration Plot

Name: GlimepirideB; Fit Type: Linear (1st Order); R^2: 0.999945; Equation Y = 1.06e+004 X - 1.78e+001

Name: Glimepiride; Fit Type: Linear (1st Order); R^2: 0.999469; Equation Y = 7.72e+003 X - 1.56e+003

Name: Glimepiride C; Fit Type: Linear (1st Order); R^2: 0.998663; Equation Y = 8.86e+003 X - 1.76e+003

Are

a

-2.0x105

0.0

2.0x105

4.0x105

6.0x105

8.0x105

1.0x106

1.2x106

ppm

0.00 10.00 20.00 30.00 40.00 50.00 60.00 70.00

1

2

3

4

5

6

7

8

9

10

11

12

Name Level X Value Response Calc. Value % Deviation Manual Ignore

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

GlimepirideB

1.000000

1.000000

1.000000

5.000000

5.000000

5.000000

10.000000

10.000000

10.000000

50.000000

50.000000

50.000000

10427.014939

10322.367898

10382.729387

53820.208641

54051.671831

53601.392108

106798.664119

107004.997494

107492.919871

529969.685196

529309.806651

528377.102249

0.983969

0.974110

0.979797

5.071877

5.093682

5.051263

10.062776

10.082214

10.128179

49.928110

49.865945

49.778079

-1.603

-2.589

-2.020

1.438

1.874

1.025

0.628

0.822

1.282

-0.144

-0.268

-0.444

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

No

Peak: GlimepirideB

Figure 4. Overlay of UV linearity curves at 228nm of glimepiride

and related impurities B and C. The API linearity curve covers

the range of 1-250 µg/mL for the API and 1-50 µg/mL for

related compounds B and C. All R2 values were >0.998.

AU

0.00

0.20

0.40

0.60

0.80

1.00

1.20

1.40

1.60

1.80

Minutes

0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00

PDA @

228 nm

250 µg/mL 125 µg/mL 50 µg/mL 10 µg/mL

LS

U

0.00

100.00

200.00

300.00

400.00

500.00

600.00

700.00

800.00

900.00

Minutes

0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00

250 µg/mL 125 µg/mL 50 µg/mL 10 µg/mL

ELSD

Figure 6. ELSD calibration curves for glimepiride related

compound C. The ELS detector has a quadratic fit to the

calibration curve (top) for peak area vs. the amount. If the

values are converted to the logarithmic functions (inset), the

calibration curve fit is linear (bottom). The R2 value for this

curve is 0.999140.

Standard

RRF Rel Compound B

RRF Rel Compound C

50 1.25 1.10

125 1.24 1.09

ELSD calibration curve

Quadratic fit

ELSD calibration curve Linear fit

Table 1. Relative Response Factors for Related compound B and

C using the ratio of the slope of the API/slope of the impurity.

The value for related compound B is outside of 0.8-1.2 range

and, therefore, should be applied, as specified by the USP

Chapter <621>.1

RESULTS AND DISCUSSION

Multi-detection of API and Related Compounds

Determination of Relative Response Factors Using Slope

of Calibration Curves in UV

𝑅𝑅𝐹 = [𝑆𝑙𝑜𝑝𝑒 𝑜𝑓 𝐼𝑚𝑝𝑢𝑟𝑖𝑡𝑦]

[𝑆𝑙𝑜𝑝𝑒 𝑜𝑓 𝐴𝑃𝐼]

Determination of Relative Response Factors Using Ratio of UV to ELSD (log) Peak Area

Calibration curve

Project Name: MassBalance Glimep 14Oct15Reported by User: System

Report Method: Calibration curve Date Printed:

23805 12/28/2015Report Method ID: 23805

8:36:43 PM US/Eastern

Calibration Plot

Name: Glimepiride; Fit Type: Linear (1st Order); R^2: 0.996948; Equation Y = 1.84e+000 X + 1.38e+000

Name: GlimepirideB; Fit Type: Linear (1st Order); R^2: 0.999125; Equation Y = 1.69e+000 X + 2.25e+000

Name: GlimepirideB; Fit Type: Linear (1st Order); R^2: 0.999125; Equation Y = 1.69e+000 X + 2.25e+000

Name: Glimepiride; Fit Type: Linear (1st Order); R^2: 0.996948; Equation Y = 1.84e+000 X + 1.38e+000

Name: Glimepiride C; Fit Type: Linear (1st Order); R^2: 0.999140; Equation Y = 1.67e+000 X + 2.14e+000

Name: Glimepiride; Fit Type: Linear (1st Order); R^2: 0.996948; Equation Y = 1.84e+000 X + 1.38e+000

Lo

gA

rea

1.00

2.00

3.00

4.00

5.00

6.00

7.00

ppm

0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40

1

2

3

4

5

Name Y-axis X-axis Fit Type R^2 Equation

GlimepirideB

Glimepiride C

Glimepiride

Glimepiride C

Glimepiride

LogArea

LogArea

LogArea

LogAmount

LogAmount

LogAmount

Linear (1st Order)

Linear (1st Order)

Linear (1st Order)

0.999125

0.999140

0.996948

Y = 1.69e+000 X + 2.25e+000

Y = 1.67e+000 X + 2.14e+000

Y = 1.84e+000 X + 1.38e+000

Table 2. Relative Response Factors for related compound B and

C using the ratio of the UV peak area to the log of the ELSD

peak area. RRF can be calculated using the response of the UV

detector to a mass concentration dependent detector.2 This

assumes a linear relationship for both detectors. To convert the

ELSD calibration response to a linear function, the log of both x

and y values can be used. Thereby, using the log of the ESLD

peak area, we can calculate RRF factors for both impurities.

These values have good correlation with those obtained using

the slopes of the calibration curves in the UV.

𝑅𝑅𝐹 = 𝑈𝑉 𝐴𝑟𝑒𝑎𝐼𝑚𝑝𝑢𝑟𝑖𝑡𝑦

𝐸𝐿𝑆𝐷 log(𝐴𝑟𝑒𝑎 𝑖𝑚𝑝𝑢𝑟𝑖𝑡𝑦/

𝑈𝑉 𝐴𝑟𝑒𝑎𝐴𝑃𝐼𝐸𝐿𝑆𝐷 𝐴𝑟𝑒𝑎 𝐴𝑃𝐼

𝑅𝑅𝐹 = 𝑈𝑉 𝐴𝑟𝑒𝑎𝐼𝑚𝑝𝑢𝑟𝑖𝑡𝑦

𝐶𝑜𝑛𝑐 𝐷𝑒𝑡𝑒𝑐𝑡𝑜𝑟 𝑟𝑒𝑠𝑝𝑜𝑛𝑠𝑒 𝑖𝑚𝑝𝑢𝑟𝑖𝑡𝑦/

𝑈𝑉 𝐴𝑟𝑒𝑎𝐴𝑃𝐼𝐶𝑜𝑛𝑐 𝐷𝑒𝑡𝑒𝑐𝑡𝑜𝑟 𝑟𝑒𝑠𝑝𝑜𝑛𝑠𝑒 𝐴𝑃𝐼

n = 3

Reference

T=0

1 day

3 days

5 days

7 days

Amount

238.4 236.1 239.0 242.7 239.1 244.1

% Recovery

99.0 100.2 101.8 100.3 102.4

Figure 8. Mass analysis window in Empower 3 FR2 of forced

degradation at 3 days. In forced degradation studies, mass

imbalance can be caused by co-elutions of the API and any

impurities. To evaluate peak purity, mass detection can provide

valuable information. In this view, the UV, Total Ion

Chromatogram (TIC) and Extracted Ion Chromatograms (XIC)

are shown in a stacked plot. For each integrated peak in the UV

chromatogram the leading, apex and trailing portion of the peak

can be viewed (top portion of figure). Using the mass and UV

detectors the peak purity can be evaluated to ensure no co-

elutions.

Pre-configured Splitter in the

Isocratic Solvent Manager

ACQUITY UPLC H-Class System

with Column Managers and Triple

Detection (UV-ELSD-PDA)

API Rel Cmpd B

Rel Cmpd C

API

Rel Cmpd B

Rel Cmpd C

Project Name: Mass Balance Degrad Nov15Reported by User: System

Report Method: Untitled Date Printed:

125 12/1/2015Report Method ID: 125

11:17:57 AM US/Eastern

GlimepirideB - 0.381 - QDa Positive Scan

Leading Apex Trailing

222.8

300.1339.0381.3

226.5 226.5

298.2 360.8

374.05

491.19

514.96

374.23 374.12

374.72

432.44

GlimepirideC - 0.546 - QDa Positive Scan

Leading Apex Trailing

228.3

294.5337.2359.0

228.3 228.3

300.1316.7355.2

432.26 432.21 432.30

PDA Spectrum (210-400)nm

0.3

81

- 3

74

.23

0.5

46

- 4

32

.21

2.9

58

- 5

13

.27

AU

0.00

0.10

0.20

0.30

QDa Positive Scan CODA SM 5 MCQ 0.75 (QDa Positive(+) Scan (100.00-600.00)Da, Centroid, CV=5)

Inte

ns

ity

0

2x106

4x106

6x106

GlimepirideB XIC - 0.381 - QDa Positive Scan: 374.23 m/z

GlimepirideC XIC - 0.546 - QDa Positive Scan: 432.21 m/z

Glimepiride XIC - 2.958 - QDa Positive Scan: 513.27 m/z

Inte

ns

ity

0

2x106

4x106

Minutes

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00

UV

TIC, ESI+

XIC, ESI+

Peak purity view of glimepiride related

compound C

+ HCl Δ

Flow diagram