The role and the performance of CIRCE lab in the framework of the Isonitrate Italy Project L. Stellato , F. Marzaioli, F. Terrasi, A. D’Onofrio, A. Ricci, M. Rubino, E. Di Fusco, B. Di Rienzo CENTRE FOR ISOTOPIC RESEARCH ON CULTURAL AND ENVIRONMENTAL HERITAGE La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015 E-mail: [email protected]

Transcript of The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass...

Page 1: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

The role and the performance of CIRCE lab

in the framework of the Isonitrate Italy Project

L. Stellato, F. Marzaioli, F. Terrasi, A. D’Onofrio, A. Ricci, M. Rubino,

E. Di Fusco, B. Di Rienzo

CENTRE FOR ISOTOPIC RESEARCH ON

CULTURAL AND ENVIRONMENTAL HERITAGE

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

E-mail: [email protected]

Page 2: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

The CIRCE lab

• A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope ratios (e.g. actinides and 14C). • A mass spectrometry lab (IRMS) for the stable isotope analysis of solid, gaseous and liquids samples. • An environmental radioactivity laboratory allowing also the 222Rn and gamma emitters measurement. • An isotope chemistry lab involved also in methodological developments (e.g AMS measurement of 11B). • An isotope hydrology and cryogenic lab. • A facility for the study of nuclear reactions in the framework of nuclear astrophysics.

CIRCE - FIELDS OF ACTIVITY a) Archaeometry by high-precision

radiocarbon dating; b) Environmental science by 14C-based global

carbon cycle studies; c) Isotope hydrology; d) Nuclear astrophysics by the ERNA recoil

mass separator; e) Nuclear safeguards and contrast to illegal

nuclear fuel use by actinides AMS; f) Forensic applications of AMS; g) Tribology by 7Be implantation.

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

Page 3: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

The IRMS lab

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

De Thermo Finnigan DELTAplus

IRMS with Dual Inlet

FlashEA™ 1112

Thermo Finnigan TC/EA

ConFlo III

GC-C/TC-Unit

ConFlo II

Thermo Finnigan DELTA V IRMS

Page 4: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

Number of isotopic analyses in the Isonitrate Italy project

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

Total: 668 samples

Page 5: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

The environmental samples and isotopic analyses in the Nitrate Project

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

Page 6: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

The samples reception: the Nitrobase

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

Page 7: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

The Silver Nitrate method (Silva et al., 2000) for δ15NNO3 and δ18ONO3 analyses

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

Pretreatment Protocol

Minimum quantity of NO3-:

6 mg (100 µmol)

3 Internal Reference Materials (RM)

in two replicates processed along

with Unknown samples:

1) USGS 34 (δ15N=-1.8±0.2‰;

δ18O=-27.9±0.6‰);

2) SiAl KNO3 (δ15N=26.9±0.8‰;

δ18O=23.8±0.1‰);

3) CIRCE KNO3 3 (δ15N=2.5±0.5‰;

δ18O=24.8±0.5‰).

Page 8: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

The preparation protocol for δ11B analyses

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

NH3 3 M, contaminants removal

Water sample

B(OH)3 + H2O B(OH)-4 + H+

Add NaOH 0.1M, pH 10 adjustment

Eluate/solution Add Mannitol, boron trapping, and drying at 70 °C

11B/10B by AMS

HCl 0.1M, elute

Exchange column HCl 2M, cleaning NH3 3M, seasoning

1 mL Amberlite exchange resin

Minimum quantity of B required:

500 µg

International Reference Materials

(RM) in 5 replicates processed along

with Unknown samples:

NIST951 (δ11B=0 ‰)

Page 9: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

The AMS analysis of δ11B

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

Page 10: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

The isotopic analyses: Measurement performances

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

Measurement precision:

δ15NNO3 = ± 0.7 ‰

δ18ONO3 = ± 1.3 ‰

δ11B = ± 6 ‰

Page 11: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

Throughput of the CIRCE lab for the Isonitrate Italy project

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

Average Measurement Rate: 510 samples per month

Average Measurement Rate:

200 samples per month

Average Measurement Rate:

1100 samples per month

4419

668

Page 12: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

The CIRCE working group

Dr. Mauro Rubino

Dr. Luisa Stellato

Dr. Fabio Marzaioli

Prof. Antonio D’Onofrio

Prof. Filippo Terrasi Egidio

Di Fusco PhD student

Brunella Di Rienzo

PhD student

Page 13: The role and the performance of CIRCE lab in the … · The CIRCE lab • A 3 MV accelerator mass spectrometer (AMS) for the ultrasensitive measurements of rare isotope 14ratios (e.g.

La contaminazione da nitrati nelle acque – Roma, 28 gennaio 2015

(Artwork made of TC/EA consumables)