The New Visian ICL V4c with KS- Aquaport TM and CentraFLOW TM Technology

42
The New Visian ICL V4c with KS-Aquaport TM and CentraFLOW TM Technology

description

The New Visian ICL V4c with KS- Aquaport TM and CentraFLOW TM Technology. COMPARISON V3 & V4 and MODELS. TICMV4 =. Design features : V4 model offered increased lens vault. Comparison of V4, V4b & V4c Models of the Visian ICL. - PowerPoint PPT Presentation

Transcript of The New Visian ICL V4c with KS- Aquaport TM and CentraFLOW TM Technology

Page 1: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

The New Visian ICL V4c

with KS-AquaportTM and CentraFLOWTM Technology

Page 2: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

COMPARISON V3 & V4 and MODELS

TICMV4 =

Design features:V4 model offered increased lens vault

Page 3: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Comparison of V4, V4b & V4c Models of the Visian ICL

Page 4: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Visian ICL V4 Visian ICL V4B Visian ICL V4c

FEATURE ICL (V4) VISIAN (V4b) VISIAN (V4c)with CentraFLOW Technology

Storage solution NaCl (equivalent to BSS once hydrated in the eye)

BSS Same as V4b

Axis alignment marks

2 diamond-shaped marks 2 straight lines Same as V4b

Full thickness holes None 2 peri-optic (aid in alignment & facilitate visco removal)

Same as V4b with addition of central optic port eliminating need for PI

Lens orientation landmarks

2 squares diagram on foot-plates (distal & proximal)

Full-thickness holes on foot-plates (distal & proximal)

Same as V4b

Diopter range(label power)

ICM -3 to -23 D VICM -3.0 to -18.0 (0.5D steps)VICM -0.5 to -3.0 (0.25D steps

Same as V4b

ICH +3 to +21 D VICH +3.0 to +10.0 (0.5D steps)VICH +0.5 to +3 D (0.25D steps)

N/A

VTICM -3 to -23 Sph. +1 to +6 CylNo MIXED

VTICM -0.5 to -18 Sph +0.5 to +6 CylIncluding MIXED astigmatism

VTICMO -0.5 to -18 Sph +0.5 to +6 CylIncluding MIXED astigmatism

None VTICH +0.5 to +10 Sph +0.5 to +6 Cyl Including plano-postive toric

N/A

Page 5: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Labelling

V4 V4b V4cICM VICM VICMOICH VICHTICM VTICM

(incl.mixed) VTICMO

VTICH (incl.plano+toric)

Model Names

Page 6: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Labelling

ICL V4 ICL V4b & V4c

11.0 11.611.5 12.112.0 12.612.5 13.213.0 13.7

Lens Length

Page 7: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

V4b Design

Exactly the same design as the V4 Also has Penetrating peri-optic holes Also has Extended Toric alignment markings Lens stored in BSS (different storage medium)

Page 8: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

V4b: New Penetrating Holes

Spherical Lens

Toric Lens

Located on either side of OZ, these holes will assist in viscoelastic removal. These holes are

present on all models

Page 9: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

ICL Toric V4b: New Alignment Marks

Available on Myopic and Hyperopc versions

New alignment marks will assist in proper positioning of the Toric ICL during and after

surgery

Page 10: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Orientation Marks:

Directionof

Implantation

Leading RightLeading Right

Trailing LeftTrailing Left

Page 11: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

•The V4c is the same design as the V4b with the addition of a central port

•The KS- Aquaport has been added in the central topic and is the same size as the peri-optic ports (360 microns)

•V4c model available for the myopic spheric and myopic toric lens versions

V4c

Page 12: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

VICMO

VTICMO

Visian ICL V4c Available Powers•Available only in the myopic range•Low diopter range expanded down to -0.5D•0.25 D increments up to -3.0D

Page 13: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Results with the Visian ICL with CentraFLOW•Erik Mertens, MD (BE)

▫130 eyes of 68 patients▫Mean age 31.66 years ± 7.4 years▫Mean Pre-op SEQ = -6.15D ± 2.72D▫Mean Pre-op Cyl = -1.11D ± 1.24D▫No PIs performed

Page 14: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

•Erik Mertens, MD (BE)▫At 6 months Post op UCVA = 1.10 ±0.14▫Mean BCVA = 1.11 ±0.14▫Efficacy index = 1.05▫Safety index = 1.06▫Mean SEQ decreased from -6.15 ± 2.72 to

-0.01 ±0.06 D ▫96.2% of eyes were within ±0.50 D of

attempted correction and 100% were within ±1.00 D 

Results with the Visian ICL with CentraFLOW

Page 15: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Results with the Visian ICL with CentraFLOW

•Erik Mertens, MD (BE)▫IOP was maintained and stable throughout

follow-up. ▫Mean postop ICL-to-lens distance (vault) was

560 μm (range, 400–800)▫No eye experienced pupillary block or

elevated IOP▫No eye required peripheral iridotomy or

explantation.

Page 16: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

evolution over timeIntraocular Pressure

Pre 1 Day 1 Month

3 Months

6 Months

MeanMedian

Page 17: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

measured with OCT AC angles measured preop and postop

Vault

1 D 1 Mo 3 Mo 6 Mo

Page 18: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

New Visian ICL Product Innovation

Saves Time and Less Risk of Errors

ICL V5

Page 19: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

ONLINE & ORDERING SYSTEM OCOS https://ocos.staarag.ch/

Page 20: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

STAAR will provide the URLs of the Online Calculator and the PasswordThe User Id will be the Email AddressThe Link to the Online Calculator is: https://ocos.staarag.ch/

Login

Page 21: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

A New Calculation

Page 22: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Walkthrough

Enter the pre-op data

Page 23: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

A drop down menu will appearFor the DOB entry

Pop-up messages will guide the userto enter the data correctly

Walkthrough

Page 24: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Walkthrough

Click ‚Calculate‘ to calculate the power of the lens and choose length

Page 25: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Walkthrough

Select the lens power according to the desired post-op refraction the surgeon wants for this eye.

Page 26: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

The length is selected automatically but it can be changed by ticking this box

Walkthrough

Page 27: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Walkthrough

Page 28: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Click ‚Submit‘ to look up lens availability

Walkthrough

Page 29: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Walkthrough

Select a lens from inventory by clicking the radio button

Page 30: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Once a lens is slected, the Manufacture Button will change to Reserve. Click reserve to reserve the lens for your order

Page 31: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology
Page 32: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology
Page 33: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Printing an implantaion orientation diagram (IOD)

Once a calculation has been done and a lens is tagged (Toric) the IOD can be generated. Go back to main screen, find your eye using the search function and click on the IOD icon.

Page 34: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology
Page 35: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

What Happens after a surgeon reserves a lens?•An email will be sent automatically to the

person who made the calculation and the distributor.

•The distributor will then need to process the order.

Page 36: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Online Calculation & Ordering System 2010

When you reserve the lens, you will immediately get an emailconfirmation from customer service. You should continueto order the lens as you normally do through your distributor.

What Happens after a surgeon reserves a lens?

WHAT HAPPENS AFTER A SURGEON RESERVES A LENS?

Page 37: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

What Happens after a surgeon reserves a lens?

Page 38: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Clicking on „show only my reserved items“ will show you all lenses that you have reserved..

YOU CAN REVIEW ALL LENSES THAT YOU’VE RESERVED

Page 39: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Online Calculation & Ordering System 2010

Walkthrough

Please check your patient‘s name, eye and the reserved lens. You can also check the reservation number from the email you receive from STAAR Customer Service. You can also export your calculations to an excel file.

Page 40: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

h

If you want to cancel your reservation, click on the small box next to the lens and click on „cancel reserved items“. Please use this option with caution since no more actions may be taken once the reservation is cancelled.

Page 41: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

You can now search for the calculation by any of the following methods:Date of the CalculationPatient ID„Search“ to see all stored calculations

WALKTHROUGH

Page 42: The New Visian ICL V4c with KS- Aquaport TM   and   CentraFLOW TM  Technology

Questions???

Visian ICL

THANK YOU