THE GONIOMETERS
Yury Gavrikov25/03/2009
![Page 2: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/2.jpg)
2
Looking back
![Page 3: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/3.jpg)
1 year ago… I met with two very nice g…oniometers
3
Because we decided to adapt their for a new experiment
![Page 4: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/4.jpg)
The idea of mechanicsLinear movement
Gear boxBelt drive + “virtual”
screw
Engine DC-servo motor
Carriage slides on the 2 guides
4
![Page 5: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/5.jpg)
Linear motion
(full range = 90mm)(full range = 90mm)
Inside of beam
Outside of beam
5
![Page 6: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/6.jpg)
The idea of mechanics Angular
movement
1mm~
63m
m
AxisAxis
PusherPusher
Angle Stepper motor
&Belt driveMicroscrew
6
![Page 7: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/7.jpg)
ModernizationUnificated solution for angular & linear
motion:
• Stepper motor for driving
• LVDT for position measurement
• Common Control system for all devices
It was needed the mashining of original parts
7
![Page 8: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/8.jpg)
Mechanical revision
After modernization during last month before installation :
- 4 revisions and full disassembling- A lot of motion tests and measurements- Fixing of sources of the problems
8
![Page 9: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/9.jpg)
9
Measurement & Calibration
![Page 10: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/10.jpg)
Angle measurement Digital Autocollimator specially
designed for the experiment
Range: ~ 8000 mrad (horizontal)~ 6500 mrad (vertical)
Resolution: < 0.5mrad10
![Page 11: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/11.jpg)
Angle measurement
The general Idea of angle measurement
Mirror
11
![Page 12: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/12.jpg)
Angle calibration The Idea of crystal angle calibration
Reference mirror
Crystal
12
![Page 13: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/13.jpg)
Angle calibration Angle vs. LVDT position
-50000 0 50000
-7500
-5000
-2500
0
2500
5000
7500
An
gle
, u
rad
LVDT, uV
Goniometer 2Goniometer 1
-50000 0 50000
-7500
-5000
-2500
0
2500
5000
7500
Ang
le, u
rad
LVDT, uV
13
![Page 14: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/14.jpg)
Nonlinearity of goniometers
Periodical structure corresponds to the motor revolutionNonlinearity : about +/- 50mrad
-30000 -20000 -10000 0 10000 20000 30000-50
-40
-30
-20
-10
0
10
20
30
40
Non
linea
rity,
ura
d
LVDT, V
-30000 -20000 -10000 0 10000 20000-60
-50
-40
-30
-20
-10
0
10
20
30
No
nlin
ear
ity,
rad
LVDT, V
14
![Page 15: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/15.jpg)
Stability
15
The goniometers are very SENSITIVE to any mechanical disturbance
The design is SOFT. “Don’t touch” should be placed everywhere
We can move it only in work direction to avoid all gaps
Mechanical stability
![Page 16: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/16.jpg)
16
Stability Stability of LVDT during 14 hours looks quite good (*)
Gonio2, LVDT
3-5 mrad
(*) with averaging to reduce the noise
![Page 17: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/17.jpg)
Repeatability Preliminary estimation for angular motion
17
-30000 -20000 -10000 0 10000 20000 30000
-70-60-50-40-30-20-10
0102030405060708090
No
nlin
ea
rity
, ra
d
LVDT, V
![Page 18: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/18.jpg)
18
Alignment system
![Page 19: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/19.jpg)
Crystal align andAngle measurement on the tank
T ank
D i g i tal Auto c o l l i m ato r
P e ntapr i zm 1
P e ntapr i zm 2
C r ys tal 1 C r ys tal 2
R e fe r e nc e m i r r o r
19
Crystal 1
Crystal 2
Ref. mirror
![Page 20: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/20.jpg)
Photos of alignment system
Digital autocollimator
Reference mirror
Laser
20
Penta prizm
Viewport
![Page 21: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/21.jpg)
Sources of uncertainty• Initial orientation of the alignment rail to
the tank geometry +/- 350mrad• Nonlinearity of goniometers +/- 50 mrad
21
![Page 22: The Goniometers](https://reader035.fdocuments.us/reader035/viewer/2022081515/5681454f550346895db21faa/html5/thumbnails/22.jpg)
Conclusions
We can use this goniometers with some magic gestures right now
BUT
To reach better results in the future we have to develop a new goniometer
22