The Determination of 226Ra in Water Samples
Anil H. Thakkar
Eichrom Technologies, Inc.
eichromTECHNOLOGIES INC
![Page 2: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/2.jpg)
Objective
• To design a procedure for radium in water samples using Eichrom’s cation exchange resin and Ln Resin.
eichromTECHNOLOGIES INC
![Page 3: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/3.jpg)
Current Procedure Used
• Barium sulfate precipitation
• Conversion to BaCO3- tedious and time consuming
eichromTECHNOLOGIES INC
![Page 4: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/4.jpg)
eichromTECHNOLOGIES INC
• Sample + Ba carrier + Ba-133 + Conc.H2SO4_+ Heat BaSO4
• Precipitate allowed to settle for several hours or overnight
• Precipitate collected and converted to BaCO3
Current Procedure Used
![Page 5: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/5.jpg)
New Approach
Water samples
Cation exchange resin column
Eichrom Ln Resin column
eichromTECHNOLOGIES INC
![Page 6: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/6.jpg)
Preparation of Water Samples
• 0.5 L / 1.0 L water samples, acidified to
pH 2
• Add 133Ba as a tracer
eichromTECHNOLOGIES INC
![Page 7: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/7.jpg)
Eichrom Cation Exchange Column
• Load solution: 0.5 L/ 1.0 L of sample
• Rinse with 50 mL of 0.1M HNO3
• Strip Ra and Ba with 50 mL of 8M HNO3
eichromTECHNOLOGIES INC
![Page 8: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/8.jpg)
Preparation of Load Solution for Ln Resin Column
• Evaporate eluent from cation exchange column to dryness
• Dissolve residue in 10 ml of 0.095M HNO3 acid
• Load on Eichrom Ln Resin column
eichromTECHNOLOGIES INC
![Page 9: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/9.jpg)
Ln Resin Column Separation
(1) Load: 10 mL of mL 0.095M HNO3,
(3) Strip: 10 mL of 0.35M HNO3
Collect 1) and (2) for Ba & 226Ra
collect (3) for 228Ac
(2) Rinse: 15 mL of 0.095M HNO3
eichromTECHNOLOGIES INC
![Page 10: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/10.jpg)
226Ra Via Alpha Spectrometry• Evaporate 226Ra/133Ba fraction to 10 mL volume
• Add 3 mL of 40% sodium sulfate, 4 drops of 1:1 acetic acid. Swirl
• Add 0.2 mL of seeding suspension. Mix
• Place tubes in cold water bath for 30 minutes
• Filter on 0.1 micron filters, count via alpha spectrometry
eichromTECHNOLOGIES INC
![Page 11: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/11.jpg)
133Ba Via Gamma Counter
• Count 133Ba tracer yield via gamma spectrometry
eichromTECHNOLOGIES INC
![Page 12: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/12.jpg)
228Ac Via Gas-Flow Proportional Counter
• Add 50 micrograms of cerium carrier
• Add 1.0 mL of HF
• Filter on 0.1 micron filters
• Measure for beta radiation
eichromTECHNOLOGIES INC
![Page 13: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/13.jpg)
226Ra % Recoveries-0.5L DI Water
0102030405060708090
100
1 2 3 4 5 6 7 8 9 10 11 12 13 14
Sample #
eichromTECHNOLOGIES INC
![Page 14: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/14.jpg)
226Ra Alpha Spectrum
eichromTECHNOLOGIES INC
![Page 15: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/15.jpg)
133Ba Gamma SpectrumTECHNOLOGIES INCeichrom
![Page 16: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/16.jpg)
226Ra, 224Ra, and Daughters
eichromTECHNOLOGIES INC
226 Ra222Rn
224Ra
218Po220Rn
![Page 17: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/17.jpg)
226Ra % Recoveries in Tap water
0
20
40
60
80
100
120
1 2 3 4 5
Sample #
4 Tap Water (0.5L)5 Tap Water (1.0 L)
eichromTECHNOLOGIES INC
![Page 18: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/18.jpg)
226Ra in EPA SampleSept.18, 1998
Sample # Ba-133% Yield
EichrompCi/L
1 89 1.562 100 1.233 98 1.45
Average 95.7 + 6 1.41 + 0.17
ExpectedEPA Value
1.7 + 0.5pCi/L
eichromTECHNOLOGIES INC
![Page 19: The Determination of 226 Ra in Water Samples Anil H. Thakkar Eichrom Technologies, Inc. eichrom TECHNOLOGIES INC.](https://reader035.fdocuments.us/reader035/viewer/2022072011/56649e215503460f94b0d17a/html5/thumbnails/19.jpg)
Summary
• Cation exchange resin option: – provides a good replacement for the time
consuming BaSO4 precipitation steps
– it eliminates inconsistent metathesis steps– reduces usage of chemicals/reagents– provides consistency and saves time
• Acceptable 226Ra recoveries
eichromTECHNOLOGIES INC