The Boyles Design
as applied in the
GE Aisys
Anaesthetic Machine
Steven Threlfo
John Hunter Hospital Anaesthetics Dept.
Revision
The Classic Boyles Design
GE
Aisys
Gas
Distribution
Diagram
GE
AisysFollowingthe same
path
Gas
Distribution
Diagram
GE
Aisys
Gas
Distribution
Diagram
GE
Aisys
Gas
Distribution
Diagram
GE
Aisys
Gas
Distribution
Diagram
GE
Aisys
Gas
Distribution
Diagram
GE
Aisys
Gas
Distribution
Diagram
GE
Aisys
Gas
Distribution
Diagram
GE
AisysAdditional
Components
Gas
Distribution
Diagram
GE
Aisys
Gas
Distribution
Diagram
GE
Aisys
Gas
Distribution
Diagram
GE
Aisys
Some important theory
Electronic Gas Delivery System
Proportional Valve
Electronic Gas Delivery System
Proportional Valve
Electronic Gas Delivery System
Proportional Valve
Electronic Gas Delivery System
Proportional Valve
Electronic Gas Delivery System
Proportional Valve
Feedback from the flow sensor controls proportional valve drive
Electronic Gas Delivery System
Servo Gas Flow Control
Gas
Mixer
GE
Aisys
Electronic Gas Delivery Module
Electronic Gas Delivery System Diagram
Electronic Gas Delivery System Diagram
Flow Control Valve
Electronic Gas Delivery System Diagram
Flow Control Valve
Electronic Gas Delivery System Diagram
Flow Sensor
Electronic Gas Delivery Hot Wire Anemometers
Electronic Gas Delivery System Diagram
Flow Sensor
Electronic Gas Delivery System Diagram
Electronic Gas Delivery System Diagram
Electronic Gas Delivery System Diagram
Two unique
measurements
of flow are
compared to
ensure safe
gas delivery
Electronic Gas Delivery System Diagram - Pneumatic equivalent
Electronic Gas Delivery System Diagram - Electrical equivalent
Electronic Gas Delivery System Diagram
Flow 1 Flow 2
GE
Aisys
Alternate
O2 Control
Alternate O2 Control:
• Safety device to allow
administration of 100% O2
from 0.5 - 10 L/min.
• Triggered 3 ways:
• Electronic mixer malfunction,
system reacts in 500 milliseconds.
• LCD screen malfunction; lose
controls.
• No power - mains or battery.
GE
Aisys
Alternate Oxygen Flow Path - Electronic Gas Delivery System
Alternate Oxygen Flow Path - Electronic Gas Delivery System
Alternate O2 Control:
• Requires a cylinder or wall
oxygen supply to be
available.
• Only works when the
power on/off switch is in
the on position.
GE
Aisys
GE
Aisys
Alternate
O2 Control
Alternate O2 Control:
• Requires a cylinder or wall
oxygen supply to be
available.
• Only works when the
power on/off switch is in
the on position.
• May also allow agent
delivery, provided the
oxygen flow measurement
system is functional.
GE
Aisys
Alternate Oxygen Flow Path - Electronic Gas Delivery System
Gas
Delivery
GE
AisysIn Summary:
• Mixes O2 with AIR or N2O.
• Has N2O shut off and
anti-hypoxic delivery.
( Computer says NO… )
• May be bypassed with 100%
O2 via the alternate oxygen
control system.
Revision
Detachable mechanical vaporizers
In a closed system, all volatile
liquids have a unique vapor
pressure which is driven by
temperature.
In a closed system, all volatile
liquids have a unique vapor
pressure which is driven by
temperature.
Low
Temperature
In a closed system, all volatile
liquids have a unique vapor
pressure which is driven by
temperature.
Higher
Temperature
Increased
Pressure
Vaporizer - Off - Gas Path
Vaporizer - On - Unit Warm
Vaporizer - On - Unit Warm
Vaporizer - On - Unit Cool
Vaporizer - On - Unit Cool
Anaesthetic agent boiling points at
a pressure of 1 atmosphere:
Sevoflurane 58.6 °C
Isoflurane 48.5 °C
Desflurane 23.5 °C
(Below normal room temperature)
Desflurane Vaporizer
Desflurane Vaporizer
Desflurane Vaporizer
Desflurane Vaporizer
Desflurane Vaporizer
Desflurane Vaporizer
Desflurane Vaporizer
Datex Ohmeda TEC 6
Desflurane Vaporizer
GE
Aisys
Electronic
Vaporizer
All the functions of a
mechanical vaporizer
are integrated into one
electronically controlled
subsystem
Electronic Vaporizer
Electronic Vaporizer
Electronic Vaporizer Cassette
Electronic Vaporizer Cassette
Electronic Vaporizer Schematic Drawing
Cassette
Electronic Vaporizer Schematic Drawing
Cassette
Cassette Identification
Selects the software
lookup table relating
saturated vapor
pressure and
temperature.
Cassette Identification
Cassette Interface Connectors
Liquid level detection
and enhanced
temperature sensing.
Cassette Interface Connectors
Temperature Measurement
Central to calculation
of saturated vapor
pressure.
Temperature Measurement
Bypass and Cassette Flow Paths
Cassette
Bypass flow
Bypass and Cassette Flow Paths
Cassette
Bypass flow
Cassette flow
Bypass and Cassette Flow Paths
Cassette
Bypass flow
Cassette flow
Bypass and Cassette Flow Paths
Cassette
Cassette Flow Control Valve
Cassette Flow Control Valve
Cassette Flow Control Valve
This is a Proportional
Valve
Cassette Inflow Flow Sensor
Cassette Inflow Flow Sensor
Cassette Inflow Flow Sensor
If required, the
Cassette Inflow Flow
Sensor is used to
control inlet flow.
Cassette Outflow Flow Sensor
Cassette Outflow Flow Sensor
Cassette Outflow Flow Sensor
If required, the
Cassette Outflow Flow
Sensor is used to
control outlet flow.
Cassette Inflow and Outflow Valves
Isolate the vaporizer
cassette from the
machine.
Input Flow Rate
Known from the
electronic gas
delivery system.
Back Pressure Valve
Drives carrier gas
through the vaporizer
cassette.
Inflow Check Valve
Prevents agent
escape upstream of
the Back Pressure
Valve.
Inflow Check Valve
Prevents agent
escape upstream of
the Back Pressure
Valve.
Important if the
cassette pressure
rises to a level
sufficient to open the
Back Pressure Valve.
Bypass Flow Control
Bypass Flow Control
Components work
together to control the
gas splitting ratio.
Bypass Flow Control
Components work
together to control the
gas splitting ratio.
Bypass flow
Bypass Flow Control
Components work
together to control the
gas splitting ratio.
Bypass flow
Cassette flow
Bypass Flow Control
Components work
together to control the
gas splitting ratio.
Bypass flow
Cassette flow
Bypass Flow Control
Components work
together to control the
gas splitting ratio.
Sevoflurane and
Isoflurane delivery.
Bypass flow
Cassette flow
Cassette Pressure Transducer
Monitors cassette
pressure.
Cassette Pressure Transducer
Monitors cassette
pressure.
Desflurane may have
a saturated vapor
pressure high enough
not to require an
inflow of carrier gas.
Direct Release of Agent
Monitors cassette
pressure.
Desflurane may have
a saturated vapor
pressure high enough
not to require an
inflow of carrier gas.
Outflow is measured
to control Desflurane
concentration.
Direct Release of Agent
Monitors cassette
pressure.
Desflurane may have
a saturated vapor
pressure high enough
not to require an
inflow of carrier gas.
Outflow is measured
to control Desflurane
concentration.
Bypass flow
Direct Release of Agent
Monitors cassette
pressure.
Desflurane may have
a saturated vapor
pressure high enough
not to require an
inflow of carrier gas.
Outflow is measured
to control Desflurane
concentration.
Bypass flow
Cassette flow
Cassette Pressure Transducer
Monitors cassette
pressure.
Once Desflurane has
cooled below its
boiling point, it will no
longer vent into the
gas stream under its
own pressure.
Bypass flow
Bypass Flow Control
Cool Desflurane can
be vaporized in the
same manner as
Sevoflurane and
Isoflurane.
Bypass Flow Control
Bypass flow
Cool Desflurane can
be vaporized in the
same manner as
Sevoflurane and
Isoflurane.
Bypass Flow Control
Bypass flow
Cassette flow
Cool Desflurane can
be vaporized in the
same manner as
Sevoflurane and
Isoflurane.
Bypass Flow Control
Bypass flow
Cassette flow
Cool Desflurane can
be vaporized in the
same manner as
Sevoflurane and
Isoflurane.
Bypass Flow Control
Bypass flow
Cassette flow
Cool Desflurane can
be vaporized in the
same manner as
Sevoflurane and
Isoflurane.
This is the key to the
Evap dealing with all
three volatile agents.
Input and Output Flow Comparison
Provides additional
input to the control
software.
Input and Output Flow Comparison
Provides additional
input to the control
software.
Hardware self-checks
and monitoring of
vaporizer gas flows
during use.
Scavenging Valve
Vents excess
cassette pressure.
Scavenging Valve
Vents excess
cassette pressure.
Connected to the
exhaust scavenging
system.
Liquid Flow Prevention Valve
A ball float safety
device to seal the
outlet path if liquid
escapes from the
vaporizer cassette.
Liquid Flow Prevention Valve
A ball float safety
device to seal the
outlet path if liquid
escapes from the
vaporizer cassette.
Only operates if an
overfilled cassette is
installed or agent is
added during
machine operation.
Liquid Flow Prevention Valve
A ball float safety
device to seal the
outlet path if liquid
escapes from the
vaporizer cassette.
Only operates if an
overfilled cassette is
installed or agent is
added during
machine operation.
Gas flow anomaly
causes an immediate
system shutdown.
Electronic Vaporizer Control
Electronic Vaporization:
• Automatically detects the agent cassette installed in the machine.
• Monitors agent temperature and adjusts the gas splitting ratio.
Agent Identification and Concentration
Control Button
Electronic Vaporizer Control
Electronic Vaporization:
• Allows accurate volatile agent delivery during low flow anaesthesia.
• Opens the door to End Tidal or “Cruise Control” anaesthesia.
Agent Identification and Concentration
Control Button
Advanced
Breathing
System
(ABS)
GE
Aisys
The ABS is an integrated
circle breathing system and
ventilator bellows assembly.
ABS and Ventilator Schematic
ABS and Ventilator Schematic
ABS and Ventilator Schematic
ABS and Ventilator Schematic
ABS and Ventilator Schematic
GE Aisys - ABS - Hand Bag
Inspiration Exhalation
GE Aisys - ABS - Ventilator
Inspiration Exhalation
The Aisys Flowmeter AssemblyMeasures flow in the
inspiratory and
expiratory limbs of the
circle breathing
system.
Two variable orifice
flow detectors
connect to separate
differential pressure
transducers.
Calibration data, recorded during
manufacture, is retained in an
EPROM located inside the flow
detector connecting plug.
Flowmeter Differential Pressure
Transducers
ABS and Ventilator Schematic
ABS and Ventilator Schematic
ABS and Ventilator Schematic
Ventilator Operation
Ventilator Operation
Drive calibration is used to explain how components
of the ventilator control system operate.
The phases of operation of the ventilator in volume
control and pressure control modes complete the
picture.
Ventilator Drive - Major components
Ventilator Drive - The ventilator control system is reliant on
accurate pressure delivery to the proportional flow control
valve.
Ventilator Drive - The ventilator control system is reliant on
accurate pressure delivery to the proportional flow control
valve. ( The regulator has a 1% tolerance on adjustment. )
Ventilator Drive - The control system provides electrical output
to drive the proportional flow control valve.
Ventilator Drive - The control system provides electrical output
to drive the proportional flow control valve. The control system
also receives feedback from the pressure sensor.
Ventilator Drive - A calibrated pressure test gauge is
connected to the output.
Ventilator Drive - A calibrated pressure test gauge is
connected to the output. With no drive to the proportional flow
control valve, the pressure sensor zero is established.
Ventilator Drive - With the test pressure gauge connected, a leak
via the pneumatic resistance allows the control system to be
manually driven to a prescribed pressure.
Ventilator Drive - With the test pressure gauge connected, a leak
via the pneumatic resistance allows the control system to be
manually driven to a prescribed pressure. This establishes the
pressure sensor calibration.
Ventilator Drive - With a stopper to seal the output, the control
system then ramps up the valve drive output.
Ventilator Drive - With a stopper to seal the output, the control
system then ramps up the valve drive output. The valve drive
versus pressure characteristic of the system is automatically
“mapped” into the control system memory.
Ventilator Drive - A calibration orifice is connected to the output.
Ventilator Drive - A calibration orifice is connected to the output.
The size and shape of the orifice is precisely machined to result
in a known pressure versus flow characteristic.
Ventilator Drive - The control system then ramps up the valve
drive output to increase pressure.
Ventilator Drive - The control system then ramps up the valve
drive output to increase pressure. By calculation, the valve
drive versus flow characteristic of the system is automatically
“mapped” into the control system memory.
Ventilator Operation
Calibration maps the relationship between:
• Valve drive and pressure
• Valve drive and flow
ABS and Ventilator Schematic
ABS and Ventilator Schematic
ABS and Ventilator Schematic
Ventilator Operation
Ventilator Operation
Ventilator Operation - VCV
Ventilator Operation - VCV
Ventilator Operation - VCV
Ventilator Operation - VCV
Ventilator Operation - VCV
Ventilator Operation - VCV
Ventilator Operation - PCV
Ventilator Operation - PCV
Ventilator Operation - PCV
Ventilator Operation - PCV
Remember - VCV at this point
where drive to the canister is withdrawn
Ventilator Operation - PCV
Ventilator Operation - PCV
Ventilator Operation - PCV
Ventilator Operation - Summary
• Pure VCV and PCV operation represent
both ends of the spectrum.
Ventilator Operation - Summary
• Pure VCV and PCV operation represent
both ends of the spectrum.
• All Aisys ventilator modes are variations
or blends of VCV and PCV operation.
Ventilator Operation - Summary
• Pure VCV and PCV operation represent
both ends of the spectrum.
• All Aisys ventilator modes are variations
or blends of VCV and PCV operation.
• Aisys ventilator capabilities match those
of an advanced ICU ventilator.
Ventilator Operation - Summary
• Pure VCV and PCV operation represent
both ends of the spectrum.
• All Aisys ventilator modes are variations
or blends of VCV and PCV operation.
• Aisys ventilator capabilities match those
of an advanced ICU ventilator.
• Adult and paediatric patients require a
change of circuit with compliance
determined during the machine self test.
Ventilator Operation - Summary
• The ventilator control system integrates
inspiratory flow to determine delivered
volume.
Ventilator Operation - Summary
• The ventilator control system integrates
inspiratory flow to determine delivered
volume. As a result, it can compensate
for gas flow entering the breathing
system during inspiration.
GE Aisys - ABS - Ventilator
Inspiration
GE Aisys - ABS - Ventilator
Inspiration In volume delivery, the flow toward the patient on inspiration is the sum of flow returned from the ventilator bellows added to the fresh gas flow from the gas delivery system.
GE Aisys - ABS - Ventilator
Inspiration In volume delivery, the flow toward the patient on inspiration is the sum of flow returned from the ventilator bellows added to the fresh gas flow from the gas delivery system.
The flow rate measured by the flow sensor is integrated to determine delivered volume.
GE Aisys - ABS - Ventilator
Inspiration In volume delivery, the flow toward the patient on inspiration is the sum of flow returned from the ventilator bellows added to the fresh gas flow from the gas delivery system.
The flow rate measured by the flow sensor is integrated to determine delivered volume.
Ventilator drive is continually perfecting inspiratory volume.
GE Aisys - ABS - Ventilator
Inspiration In volume delivery, the flow toward the patient on inspiration is the sum of flow returned from the ventilator bellows added to the fresh gas flow from the gas delivery system.
The flow rate measured by the flow sensor is integrated to determine delivered volume.
Ventilator drive is continually perfecting inspiratory volume.
Ventilator Operation - Summary
• At low gas flows the ventilator bellows
will have maximum excursion.
Ventilator Operation - Summary
• At low gas flows the ventilator bellows
will have maximum excursion.
• At high gas flows the ventilator bellows
will have a lower excursion.
Waste Gas Scavenging
Scavenger Assembly
Scavenger Schematic
Scavenger flow is monitored by the pressure drop across the
signal orifice and displayed by a small flowmeter mounted on the
side of the machine.
Scavenger Schematic
Scavenger flow is monitored by the pressure drop across the
signal orifice and displayed by a small flowmeter mounted on the
side of the machine. Scavenger Flow > Machine Exhaust Flow.
One important
message on
electronically
controlled
equipment.
Keep corrosive solutions away from machines.
Keep corrosive solutions away from machines.
Damage to Aisys
ventilator drive
system.
Keep corrosive solutions away from machines.
Damage to Aisys
ventilator drive
system.
Something for
analysis
2 LPM Alternate Oxygen Flow Set
What failed and how?
Error Log: 23-Jul-2013, 22:40:08, Mix: FLOW ATTAINMENT FAILURE CH123-Jul-2013, 22:40:09, EGC EXIT ALT O223-Jul-2013, 22:40:10, O2 FLOW CTRL FAIL23-Jul-2013, 22:40:10, EXPIRED GAS CONTROL AUTO EXITED
Conclusion
Photograph - Gandhi Smriti
Delhi India - 2013