The Boyles Design as applied in the GE Aisys Anaesthetic Machine Steven Threlfo John Hunter Hospital Anaesthetics Dept.

Transcript of The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all...

Page 1: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

The Boyles Design

as applied in the

GE Aisys

Anaesthetic Machine

Steven Threlfo

John Hunter Hospital Anaesthetics Dept.

Page 2: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Revision

The Classic Boyles Design

Page 3: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 4: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 5: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 6: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 7: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 8: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 9: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 10: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE

Aisys

Page 11: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

AisysFollowingthe same

path

Page 12: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

Aisys

Page 13: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

Aisys

Page 14: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

Aisys

Page 15: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

Aisys

Page 16: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

Aisys

Page 17: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

Aisys

Page 18: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

AisysAdditional

Components

Page 19: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

Aisys

Page 20: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

Aisys

Page 21: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Distribution

Diagram

GE

Aisys

Page 22: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Some important theory

Page 23: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System

Proportional Valve

Page 24: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System

Proportional Valve

Page 25: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System

Proportional Valve

Page 26: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System

Proportional Valve

Page 27: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System

Proportional Valve

Page 28: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Feedback from the flow sensor controls proportional valve drive

Electronic Gas Delivery System

Servo Gas Flow Control

Page 29: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Mixer

GE

Aisys

Page 30: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery Module

Page 31: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram

Page 32: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram

Flow Control Valve

Page 33: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram

Flow Control Valve

Page 34: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram

Flow Sensor

Page 35: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery Hot Wire Anemometers

Page 36: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram

Flow Sensor

Page 37: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram

Page 38: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram

Page 39: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram

Two unique

measurements

of flow are

compared to

ensure safe

gas delivery

Page 40: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram - Pneumatic equivalent

Page 41: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram - Electrical equivalent

Page 42: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Gas Delivery System Diagram

Flow 1 Flow 2

Page 43: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE

Aisys

Alternate

O2 Control

Page 44: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Alternate O2 Control:

• Safety device to allow

administration of 100% O2

from 0.5 - 10 L/min.

• Triggered 3 ways:

• Electronic mixer malfunction,

system reacts in 500 milliseconds.

• LCD screen malfunction; lose

controls.

• No power - mains or battery.

GE

Aisys

Page 45: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Alternate Oxygen Flow Path - Electronic Gas Delivery System

Page 46: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Alternate Oxygen Flow Path - Electronic Gas Delivery System

Page 47: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Alternate O2 Control:

• Requires a cylinder or wall

oxygen supply to be

available.

• Only works when the

power on/off switch is in

the on position.

GE

Aisys

Page 48: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE

Aisys

Alternate

O2 Control

Page 49: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Alternate O2 Control:

• Requires a cylinder or wall

oxygen supply to be

available.

• Only works when the

power on/off switch is in

the on position.

• May also allow agent

delivery, provided the

oxygen flow measurement

system is functional.

GE

Aisys

Page 50: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Alternate Oxygen Flow Path - Electronic Gas Delivery System

Page 51: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Gas

Delivery

GE

AisysIn Summary:

• Mixes O2 with AIR or N2O.

• Has N2O shut off and

anti-hypoxic delivery.

( Computer says NO… )

• May be bypassed with 100%

O2 via the alternate oxygen

control system.

Page 52: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Revision

Detachable mechanical vaporizers

Page 53: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

In a closed system, all volatile

liquids have a unique vapor

pressure which is driven by

temperature.

Page 54: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

In a closed system, all volatile

liquids have a unique vapor

pressure which is driven by

temperature.

Low

Temperature

Page 55: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

In a closed system, all volatile

liquids have a unique vapor

pressure which is driven by

temperature.

Higher

Temperature

Increased

Pressure

Page 56: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Vaporizer - Off - Gas Path

Page 57: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Vaporizer - On - Unit Warm

Page 58: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Vaporizer - On - Unit Warm

Page 59: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Vaporizer - On - Unit Cool

Page 60: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Vaporizer - On - Unit Cool

Page 61: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Anaesthetic agent boiling points at

a pressure of 1 atmosphere:

Sevoflurane 58.6 °C

Isoflurane 48.5 °C

Desflurane 23.5 °C

(Below normal room temperature)

Page 62: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Desflurane Vaporizer

Page 63: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Desflurane Vaporizer

Page 64: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Desflurane Vaporizer

Page 65: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Desflurane Vaporizer

Page 66: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Desflurane Vaporizer

Page 67: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Desflurane Vaporizer

Page 68: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Desflurane Vaporizer

Page 69: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Datex Ohmeda TEC 6

Desflurane Vaporizer

Page 70: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE

Aisys

Electronic

Vaporizer

All the functions of a

mechanical vaporizer

are integrated into one

electronically controlled

subsystem

Page 71: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Vaporizer

Page 72: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Vaporizer

Page 73: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Vaporizer Cassette

Page 74: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Vaporizer Cassette

Page 75: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Vaporizer Schematic Drawing

Page 76: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette

Electronic Vaporizer Schematic Drawing

Page 77: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette

Page 78: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Identification

Selects the software

lookup table relating

saturated vapor

pressure and

temperature.

Page 79: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Identification

Page 80: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Interface Connectors

Liquid level detection

and enhanced

temperature sensing.

Page 81: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Interface Connectors

Page 82: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Temperature Measurement

Central to calculation

of saturated vapor

pressure.

Page 83: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Temperature Measurement

Page 84: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass and Cassette Flow Paths

Cassette

Page 85: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass flow

Bypass and Cassette Flow Paths

Cassette

Page 86: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass flow

Cassette flow

Bypass and Cassette Flow Paths

Cassette

Page 87: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass flow

Cassette flow

Bypass and Cassette Flow Paths

Cassette

Page 88: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Flow Control Valve

Page 89: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Flow Control Valve

Page 90: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Flow Control Valve

This is a Proportional

Valve

Page 91: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Inflow Flow Sensor

Page 92: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Inflow Flow Sensor

Page 93: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Inflow Flow Sensor

If required, the

Cassette Inflow Flow

Sensor is used to

control inlet flow.

Page 94: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Outflow Flow Sensor

Page 95: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Outflow Flow Sensor

Page 96: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Outflow Flow Sensor

If required, the

Cassette Outflow Flow

Sensor is used to

control outlet flow.

Page 97: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Inflow and Outflow Valves

Isolate the vaporizer

cassette from the

machine.

Page 98: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Input Flow Rate

Known from the

electronic gas

delivery system.

Page 99: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Back Pressure Valve

Drives carrier gas

through the vaporizer

cassette.

Page 100: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Inflow Check Valve

Prevents agent

escape upstream of

the Back Pressure

Valve.

Page 101: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Inflow Check Valve

Prevents agent

escape upstream of

the Back Pressure

Valve.

Important if the

cassette pressure

rises to a level

sufficient to open the

Back Pressure Valve.

Page 102: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Page 103: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Components work

together to control the

gas splitting ratio.

Page 104: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Components work

together to control the

gas splitting ratio.

Bypass flow

Page 105: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Components work

together to control the

gas splitting ratio.

Bypass flow

Cassette flow

Page 106: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Components work

together to control the

gas splitting ratio.

Bypass flow

Cassette flow

Page 107: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Components work

together to control the

gas splitting ratio.

Sevoflurane and

Isoflurane delivery.

Bypass flow

Cassette flow

Page 108: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Pressure Transducer

Monitors cassette

pressure.

Page 109: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Pressure Transducer

Monitors cassette

pressure.

Desflurane may have

a saturated vapor

pressure high enough

not to require an

inflow of carrier gas.

Page 110: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Direct Release of Agent

Monitors cassette

pressure.

Desflurane may have

a saturated vapor

pressure high enough

not to require an

inflow of carrier gas.

Outflow is measured

to control Desflurane

concentration.

Page 111: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Direct Release of Agent

Monitors cassette

pressure.

Desflurane may have

a saturated vapor

pressure high enough

not to require an

inflow of carrier gas.

Outflow is measured

to control Desflurane

concentration.

Bypass flow

Page 112: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Direct Release of Agent

Monitors cassette

pressure.

Desflurane may have

a saturated vapor

pressure high enough

not to require an

inflow of carrier gas.

Outflow is measured

to control Desflurane

concentration.

Bypass flow

Cassette flow

Page 113: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Cassette Pressure Transducer

Monitors cassette

pressure.

Once Desflurane has

cooled below its

boiling point, it will no

longer vent into the

gas stream under its

own pressure.

Bypass flow

Page 114: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Cool Desflurane can

be vaporized in the

same manner as

Sevoflurane and

Isoflurane.

Page 115: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Bypass flow

Cool Desflurane can

be vaporized in the

same manner as

Sevoflurane and

Isoflurane.

Page 116: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Bypass flow

Cassette flow

Cool Desflurane can

be vaporized in the

same manner as

Sevoflurane and

Isoflurane.

Page 117: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Bypass flow

Cassette flow

Cool Desflurane can

be vaporized in the

same manner as

Sevoflurane and

Isoflurane.

Page 118: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Bypass Flow Control

Bypass flow

Cassette flow

Cool Desflurane can

be vaporized in the

same manner as

Sevoflurane and

Isoflurane.

This is the key to the

Evap dealing with all

three volatile agents.

Page 119: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Input and Output Flow Comparison

Provides additional

input to the control

software.

Page 120: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Input and Output Flow Comparison

Provides additional

input to the control

software.

Hardware self-checks

and monitoring of

vaporizer gas flows

during use.

Page 121: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Scavenging Valve

Vents excess

cassette pressure.

Page 122: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Scavenging Valve

Vents excess

cassette pressure.

Connected to the

exhaust scavenging

system.

Page 123: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Liquid Flow Prevention Valve

A ball float safety

device to seal the

outlet path if liquid

escapes from the

vaporizer cassette.

Page 124: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Liquid Flow Prevention Valve

A ball float safety

device to seal the

outlet path if liquid

escapes from the

vaporizer cassette.

Only operates if an

overfilled cassette is

installed or agent is

added during

machine operation.

Page 125: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Liquid Flow Prevention Valve

A ball float safety

device to seal the

outlet path if liquid

escapes from the

vaporizer cassette.

Only operates if an

overfilled cassette is

installed or agent is

added during

machine operation.

Gas flow anomaly

causes an immediate

system shutdown.

Page 126: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Vaporizer Control

Electronic Vaporization:

• Automatically detects the agent cassette installed in the machine.

• Monitors agent temperature and adjusts the gas splitting ratio.

Agent Identification and Concentration

Control Button

Page 127: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Electronic Vaporizer Control

Electronic Vaporization:

• Allows accurate volatile agent delivery during low flow anaesthesia.

• Opens the door to End Tidal or “Cruise Control” anaesthesia.

Agent Identification and Concentration

Control Button

Page 128: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Advanced

Breathing

System

(ABS)

GE

Aisys

The ABS is an integrated

circle breathing system and

ventilator bellows assembly.

Page 129: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 130: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 131: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 132: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 133: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 134: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE Aisys - ABS - Hand Bag

Inspiration Exhalation

Page 135: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE Aisys - ABS - Ventilator

Inspiration Exhalation

Page 136: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

The Aisys Flowmeter AssemblyMeasures flow in the

inspiratory and

expiratory limbs of the

circle breathing

system.

Two variable orifice

flow detectors

connect to separate

differential pressure

transducers.

Calibration data, recorded during

manufacture, is retained in an

EPROM located inside the flow

detector connecting plug.

Page 137: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Flowmeter Differential Pressure

Transducers

Page 138: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 139: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 140: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 141: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation

Page 142: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation

Drive calibration is used to explain how components

of the ventilator control system operate.

The phases of operation of the ventilator in volume

control and pressure control modes complete the

picture.

Page 143: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - Major components

Page 144: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - The ventilator control system is reliant on

accurate pressure delivery to the proportional flow control

valve.

Page 145: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - The ventilator control system is reliant on

accurate pressure delivery to the proportional flow control

valve. ( The regulator has a 1% tolerance on adjustment. )

Page 146: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - The control system provides electrical output

to drive the proportional flow control valve.

Page 147: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - The control system provides electrical output

to drive the proportional flow control valve. The control system

also receives feedback from the pressure sensor.

Page 148: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - A calibrated pressure test gauge is

connected to the output.

Page 149: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - A calibrated pressure test gauge is

connected to the output. With no drive to the proportional flow

control valve, the pressure sensor zero is established.

Page 150: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - With the test pressure gauge connected, a leak

via the pneumatic resistance allows the control system to be

manually driven to a prescribed pressure.

Page 151: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - With the test pressure gauge connected, a leak

via the pneumatic resistance allows the control system to be

manually driven to a prescribed pressure. This establishes the

pressure sensor calibration.

Page 152: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - With a stopper to seal the output, the control

system then ramps up the valve drive output.

Page 153: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - With a stopper to seal the output, the control

system then ramps up the valve drive output. The valve drive

versus pressure characteristic of the system is automatically

“mapped” into the control system memory.

Page 154: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - A calibration orifice is connected to the output.

Page 155: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - A calibration orifice is connected to the output.

The size and shape of the orifice is precisely machined to result

in a known pressure versus flow characteristic.

Page 156: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - The control system then ramps up the valve

drive output to increase pressure.

Page 157: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Drive - The control system then ramps up the valve

drive output to increase pressure. By calculation, the valve

drive versus flow characteristic of the system is automatically

“mapped” into the control system memory.

Page 158: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation

Calibration maps the relationship between:

• Valve drive and pressure

• Valve drive and flow

Page 159: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 160: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 161: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

ABS and Ventilator Schematic

Page 162: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation

Page 163: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation

Page 164: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - VCV

Page 165: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - VCV

Page 166: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - VCV

Page 167: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - VCV

Page 168: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - VCV

Page 169: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - VCV

Page 170: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - PCV

Page 171: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - PCV

Page 172: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - PCV

Page 173: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - PCV

Page 174: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Remember - VCV at this point

where drive to the canister is withdrawn

Page 175: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - PCV

Page 176: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - PCV

Page 177: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - PCV

Page 178: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - Summary

• Pure VCV and PCV operation represent

both ends of the spectrum.

Page 179: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - Summary

• Pure VCV and PCV operation represent

both ends of the spectrum.

• All Aisys ventilator modes are variations

or blends of VCV and PCV operation.

Page 180: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - Summary

• Pure VCV and PCV operation represent

both ends of the spectrum.

• All Aisys ventilator modes are variations

or blends of VCV and PCV operation.

• Aisys ventilator capabilities match those

of an advanced ICU ventilator.

Page 181: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - Summary

• Pure VCV and PCV operation represent

both ends of the spectrum.

• All Aisys ventilator modes are variations

or blends of VCV and PCV operation.

• Aisys ventilator capabilities match those

of an advanced ICU ventilator.

• Adult and paediatric patients require a

change of circuit with compliance

determined during the machine self test.

Page 182: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - Summary

• The ventilator control system integrates

inspiratory flow to determine delivered

volume.

Page 183: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - Summary

• The ventilator control system integrates

inspiratory flow to determine delivered

volume. As a result, it can compensate

for gas flow entering the breathing

system during inspiration.

Page 184: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE Aisys - ABS - Ventilator

Inspiration

Page 185: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE Aisys - ABS - Ventilator

Inspiration In volume delivery, the flow toward the patient on inspiration is the sum of flow returned from the ventilator bellows added to the fresh gas flow from the gas delivery system.

Page 186: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE Aisys - ABS - Ventilator

Inspiration In volume delivery, the flow toward the patient on inspiration is the sum of flow returned from the ventilator bellows added to the fresh gas flow from the gas delivery system.

The flow rate measured by the flow sensor is integrated to determine delivered volume.

Page 187: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE Aisys - ABS - Ventilator

Inspiration In volume delivery, the flow toward the patient on inspiration is the sum of flow returned from the ventilator bellows added to the fresh gas flow from the gas delivery system.

The flow rate measured by the flow sensor is integrated to determine delivered volume.

Ventilator drive is continually perfecting inspiratory volume.

Page 188: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

GE Aisys - ABS - Ventilator

Inspiration In volume delivery, the flow toward the patient on inspiration is the sum of flow returned from the ventilator bellows added to the fresh gas flow from the gas delivery system.

The flow rate measured by the flow sensor is integrated to determine delivered volume.

Ventilator drive is continually perfecting inspiratory volume.

Page 189: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - Summary

• At low gas flows the ventilator bellows

will have maximum excursion.

Page 190: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Ventilator Operation - Summary

• At low gas flows the ventilator bellows

will have maximum excursion.

• At high gas flows the ventilator bellows

will have a lower excursion.

Page 191: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Waste Gas Scavenging

Page 192: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Scavenger Assembly

Page 193: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Scavenger Schematic

Scavenger flow is monitored by the pressure drop across the

signal orifice and displayed by a small flowmeter mounted on the

side of the machine.

Page 194: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Scavenger Schematic

Scavenger flow is monitored by the pressure drop across the

signal orifice and displayed by a small flowmeter mounted on the

side of the machine. Scavenger Flow > Machine Exhaust Flow.

Page 195: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

One important

message on

electronically

controlled

equipment.

Page 196: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Keep corrosive solutions away from machines.

Page 197: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Keep corrosive solutions away from machines.

Damage to Aisys

ventilator drive

system.

Page 198: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Keep corrosive solutions away from machines.

Damage to Aisys

ventilator drive

system.

Page 199: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Something for

analysis

Page 200: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 201: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 202: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 203: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 204: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

2 LPM Alternate Oxygen Flow Set

Page 205: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...
Page 206: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

What failed and how?

Page 207: The Boyles Design as applied in the GE Aisys Anaesthetic ... · vaporizers. In a closed system, all volatile liquids have a unique vapor pressure which is driven by temperature. ...

Error Log: 23-Jul-2013, 22:40:08, Mix: FLOW ATTAINMENT FAILURE CH123-Jul-2013, 22:40:09, EGC EXIT ALT O223-Jul-2013, 22:40:10, O2 FLOW CTRL FAIL23-Jul-2013, 22:40:10, EXPIRED GAS CONTROL AUTO EXITED