The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form...
Transcript of The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form...
![Page 1: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/1.jpg)
The Biological Evaluation Plan (BEP) A crucial first step in the biocompatibility evaluation of a medical device
05 MAR 2020
![Page 2: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/2.jpg)
Biological Safety Evaluation
2
Biological Evaluation Plan (BEP): What are your risks and how do you plan
to mitigate them?
Testing and risk assessments
Biological Evaluation Report (BER): Is the
device safe?
![Page 3: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/3.jpg)
Introduction
![Page 4: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/4.jpg)
Introduction
4
Device contact
Contact time
Perform tests
![Page 5: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/5.jpg)
Introduction
5
Didn’t understand
Materials
Testing
![Page 6: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/6.jpg)
Introduction
6
• ISO 10993-1 (2018): Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
• What is risk ?
ISO 14971: Combination of the probability of occurrence of harm and the severity of that harm.
![Page 7: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/7.jpg)
Introduction
7
Sen
siti
zati
on
![Page 8: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/8.jpg)
Introduction
8
Sen
siti
zati
on
X means prerequisite information needed for a risk assessment. E means endpoints to be evaluated in the risk assessment (either through the use of existing data, additional endpoint-specific testing, or a rationale for why assessment of the endpoint does not require an additional data set). If a medical device is manufactured from novel materials, not previously used in medical device applications, and not toxicological data exists in the literature, additional endpoints beyond those marked “E” in this table should be considered. For particular medical devices, there is a possibility that it will be appropriate to include additional or fewer endpoints than indicated.
![Page 9: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/9.jpg)
Introduction
9
• ISO 10993-1 (2018): The biological evaluation of any material or medical device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.
This risk management process involves identification of biological hazards, estimation of the associated biological risks, and determination of their acceptability.
• Annex B.2.2: Since biological evaluation is a risk management activity, a
Biological Evaluation Plan is required, and this forms part of the Risk Management Plan. It is emphasized that simply planning to conduct testing against all of the aspects of biocompatibility identified in Annex A does not meet the requirements of ISO 14971 or this document.
![Page 10: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/10.jpg)
Introduction
Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process“ - Section III
“Such a process should generally begin with assessment of the device, including the material components, the manufacturing processes, the clinical use of the device… Considering this information, the potential risks from a biocompatibility perspective should be identified. Considering the potential biological impact, a plan should be developed…either by biocompatibility testing or other evaluations that appropriately address the risks.”
10
![Page 11: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/11.jpg)
BEP – the practice
![Page 12: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/12.jpg)
BEP – the practice
12
ISO 10993-1 (2018) Clause 4
Configuration
Raw materials
Historical data
Packaging
Literature
Manufacturing
Test strategy
![Page 13: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/13.jpg)
1. Applicable guidelines
Guidance on ISO 10993
ISO 10993
13
ISO 18562 ISO 19227
USP 1663/1664 FDA guidance on Container
Closure Systems
![Page 14: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/14.jpg)
2. Device description
• Dimensions
• Intended purpose
14
Patient contact (invasive)
Practioner contact (non-invasive)
![Page 15: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/15.jpg)
2. Device description
• Dimensions
• Intended purpose
• Frequency of use
15
Sen
siti
zati
on
![Page 16: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/16.jpg)
2. Device description
• Dimensions
• Intended purpose
• Frequency of use
• Patient population
16
![Page 17: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/17.jpg)
2. Device description
• Dimensions
• Intended purpose
• Frequency of use
• Patient population
• Off-label use
17
![Page 18: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/18.jpg)
2. Device description
• Off-label use
18
IFU Intact skin contact ≤ 30 days contact Adult use
Risk evaluation !
![Page 19: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/19.jpg)
2. Device description
• Dimensions
• Intended purpose
• Frequency of use
• Patient population
• Off-label use
• ...
Device Categorization
19
![Page 20: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/20.jpg)
3. Material characterisation
List of all materials used: standard materials ?
Cleaning validation as per ISO 19227 ?
20
![Page 21: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/21.jpg)
3. Material characterisation
21
“The extent of characterization required is determined by the invasiveness and duration of
clinical exposure in the intended use…”
ISO 10993-18
Limited contact: identify materials and processing;
use biocompatibility testing to support safety.
Prolonged contact: Use biocompatibility testing to
support safety. Maybe chemical characterization
if materials are new.
Permanent contact : Perform chemical
characterization testing with a toxicological risk
assessment.
![Page 22: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/22.jpg)
3. Material characterisation
• Chemical characterisation needed ? ISO 10993-1: “The choice of test procedures shall take into account that
certain biological tests (i.e. those designed to assess systemic effects) are not justifiable where the presence of leachable chemicals has been excluded (in accordance with ISO 10993-18), or where chemicals have a known and acceptable toxicity profile, allowing the safe use by evaluation in accordance with ISO 10993-17 and risk assessment in accordance with ISO 14971”
22
Sen
siti
zati
on
![Page 23: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/23.jpg)
3. Material characterisation
23
“The extent of characterization required is determined by the invasiveness and duration of
clinical exposure in the intended use…”
ISO 10993-18
Limited contact: identify materials and processing;
use biocompatibility testing to support safety.
Prolonged contact: Use biocompatibility testing to
support safety. Maybe chemical characterization
if materials are new.
Permanent contact : Perform chemical
characterization testing with a toxicological risk
assessment.
![Page 24: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/24.jpg)
3. Material characterisation
• Chemical characterisation needed ?
Which strategy ?
Which solvents ?
Which techniques ?
Toxicological evaluation of the detected compounds
24
![Page 25: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/25.jpg)
4. Biocompatibility test selection and rationale
25
• Selection of each test
Cytotoxicity needed ?
MEM Elution
L929 MTT/XTT
L929 MTT/XTT with dilutions
V79 colony assay
Sen
siti
zati
on
![Page 26: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/26.jpg)
4. Test selection and rationale
26
• Justification out of testing
![Page 27: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/27.jpg)
5. Test sample selection
27
Representative product
Coupons Family
grouping “Monster”
product
![Page 28: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/28.jpg)
5. Test sample selection
28
Representative product
Coupons Family
grouping “Monster”
product
![Page 29: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/29.jpg)
5. Test sample selection
29
Representative product
Coupons Family
grouping “Monster”
product
Exclusion of components
No patient contact
Removal of electronic
components
![Page 30: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/30.jpg)
5. Test sample selection
30
Representative product
Coupons Family
grouping “Monster”
product
Exclusion of components
No patient contact
Removal of electronic
components
Contacting manner
![Page 31: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/31.jpg)
6. Test sample extraction
31
• Solvent: polar and apolar
• Ratios: surface area has preference
• Time and temperature
37°C for 24 hours
37°C for 72 hours
50°C for 72 hours
70°C for 24 hours
121°C for 1 hours
Cytotoxicity (72 hrs for implants)
Typical recommended extraction condition for other in-vitro/in-vivo tests
![Page 32: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/32.jpg)
7. Extra considerations ?
32
• Biodegradation
?
![Page 33: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/33.jpg)
7. Extra considerations ?
33
• Biodistribution
![Page 34: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/34.jpg)
7. Extra considerations ?
34
• Drug/device interaction
• ...
![Page 35: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/35.jpg)
Conclusion
![Page 36: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/36.jpg)
Conclusion
Identify applicable guidelines
Understand contact (manner and time)
=> Device categorization
Material characterisation
Select tests Define tested sample
and extraction conditions
Extra considerations ?
36
![Page 37: The Biological Evaluation Plan (BEP) · 2020-03-10 · device intended for use in humans shall form part of a structured biological evaluation plan within a risk management process.](https://reader033.fdocuments.us/reader033/viewer/2022042114/5e911b12c76d083f155dc50f/html5/thumbnails/37.jpg)
Biological Safety Evaluation
37
Biological Evaluation Plan (BEP): What are your risks and how do you plan
to mitigate them?
Testing and risk assessments
Biological Evaluation Report (BER): Is the
device safe?