Tests Operacionalesen Radiografía
Mauricio VergaraClínica del Country, 2015
Electronesacelerados Tungsteno
Positivo(anodo)
Negativo(catodo)
FOTONESCalorelectrones
Ampolla de vidrio
Circuito deAlta tensión
Circuito delfilamento
100 mA200 mA300 mA
Reloj seg
Funcionamiento
FOTONESEnergía: circuito kV
N°: circuito mAs
22 cm
28cm
28cm
110 kV, 5 mAs
110 kV, 5 mAs
110 kV, 12 mAs
DetectorIMAGEN
AEC
Tubo RX
fotones
Paciente
fotones CI
tiempo mAs
N° fotonesemergentes(ajuste densidad)
Control de Exposición Automático
Tests Operacionales en RX
• Tests Operacional: certificar que el equipo opera “normalmente” y genera imágenes en rangos de calidad clínica
• Test Aceptación: certificar que las especificaciones técnicas actuales del equipo son compatibles con las originales
RX
Grilla AEC Detector
generador mAskV
gain DOSIMETROkVmAsGain
VISUALColimadorGrilla
kVDosis : mGy, Gy, nGyTiempo: sec, msec
Generador kVp
Tubo RX
dosímetro
Separador10 cm
mesa
nominal
medido difercia%
50 kVp
80 kVp
110 kVp
Filamento mAs vs Dosis
Tubo RX
dosímetro
Separador10 cm
mesa mAs Gy
80 kVp
6
80 kVp
20
80 kVp
30
Colimación
campo Luz15 cm x 15 cm
Campo radiación 16 cm x 16 cm máximo
delimitadordetector
mesa
Tubo RX 50 kV, 6 mAs
Colimación
mesa
detector
Alineacion Tubo RX - Detector
Histograma FantomaColumna Lumbar
81 kV 1.87 mAs
OBSERVACIONES
• kVp y mAs con registros OK. Equipo RX OK• Ajuste de Densidad con valores asimétricos• Ajuste ganancias AEC: valores muy altos y
diferentes en mesa y mural (detector central)
• Histograma con forma OK. mAs muy bajo. Imagen OK
• No hay antecedentes de registros de kVp, mAs, AEC e Histogramas para comparar
• No hay especificaciones de grillas y detector
CIRS mammography phantom
TEST OPERACIONAL - CASSETTES RAYOS X Institución: Clinica del Country Fecha: Dic,2012 Equipo Modelo Sede Fabricante CR CAPSULA X CLINICA Fujifilm cassette 14x17 TAMAÑO (IP 35x43 cm) Fantoma mm =180 nominal medido ajuste mm 180 180 0 180 VALOR S para 1 mR 1 Sv = 100 mR exposición kVp 80 mA 100 mAs 2 dosis mR 1 1.014 FCR S Value Control Table 2 1.026 average 1.0324 3 1.057 4 1.038 5 1.027 Exponer IP 35x43 cm (leer monitor) S nominal 200 S min 154 S medido 279 S max 260 S corregido 200
gamma is the relationship between the brightness of a pixel as it appears on the screen, and the numerical value of that pixel. You probably already know that a pixel can have any 'value' of Red, Green, and Blue between 0 and 255, and you would therefore think that a pixel value of 127 would appear as half of the maximum possible brightness, and that a value of 64 would represent one-quarter brightness, and so on. Well, that's just not the case,
Cathode-ray tubes, such as the screen you're probably reading this on at the moment, have a peculiar relationship between the voltage applied to them, and the amount of light emitted. It isn't linear, and in fact it follows what's called by mathematicians and other geeks, a 'power law' (a number raised to a power). The numerical value of that power is what we call the gamma of the monitor or system.
DAC
Imagen Matriz
numeros
Señaleselectricas
monitor
Señaleselectricas
impresora
DACSeñales
electricasMonitor curve
Señaleselectricas
Impresora curve
LUT
LUT
curve
curve
Procesador
Resultados de curvas Gamma para sistema radiografía digital tipo CR. La recta es la Compuesta ideal; la curva roja es la Compuesta medida.
TEST OPERACIONAL
Institución : ONCOLOGICO
Fecha : 7/19/2012
Equipo Modelo
Fabricante
impresora DryPix FUJIFILM
8 x 10
TAMAÑOnominal mm
medido mm diferencia
A 156.00 155.0 1%
B 202.00 201.0 0%
DENSIDADES REFERENCIA
referencia OD OD medida diferencia OD interno
0 0.20 0.19 5% 0.17
1 0.48 0.53 -10% 0.48
2 1.20 1.27 -6% 1.19
3 1.92 2.02 -5% 1.93
SMPTE FILM 17 steps
celda OD medida step OD OD medida diferencia
100% 0.23 92% 5 0.48 0.51 -6%
90% 0.53 82% 9 1.20 1.24 -3%
80% 0.82 71% 13 1.92 1.99 -4%
70% 1.10 62%
60% 1.38 52% RECTA
50% 1.67 42% origen 2.990
40% 1.97 31% pendiente -2.710
30% 2.26 21% R2 0.996
20% 2.49 13%
10% 2.70 6%
0% 2.87 0%
CONCLUSIONES
• Tests Operacionales se ejecutan en 60 minutos y requieren dosímetro (Unfors) y fantomas (CIRS en mamografía y fantoma columna de densitómetro)
• Entregan información de Generador (kVp y mAs) y calidad de imagen
• Frecuencia cada 4 meses
PROYECTO EN CURSOHISTOGRAMAS EN ECOGRAFIA
Performance Measurements:Dead ZoneHorizontal DistanceVertical DistanceDepth of PenetrationImage UniformityAxial ResolutionLateral ResolutionAnechoic Mass ResolutionGray Scale Contrast ResolutionElasticity Image Evaluation
Transductor convexo
Transductor OKCon postprocesamiento
Transductor OKSin postprocesamiento
Transductor OKSin postprocesamiento
Daño en cristalesSin postprocesamiento
Transductor convexo
Transductor lineal
Sin postprocesamiento