Tel. 04-8295361/364 [email protected]
-
Tel. 04-8295361/364 [email protected]
• Introduction to Aspect M2™ MRI• MR safety• Possible applications and examples• MR Basics• “Recipe” for MR Imaging• Useful notes• Scan protocols• Custom MRI Analysis Tool• K‐Pacs Basic Instructions• How to begin
2
-
Tel. 04-8295361/364 [email protected]
Permanentmagnet vs. Standard superconducting coil based
(*illustrative photos)3
-
Tel. 04-8295361/364 [email protected]
Mouse/rat head coil
Mouse/rat body coil
Anastasia
Cradle Heating system 4
-
Tel. 04-8295361/364 [email protected]
• Compact • A new level in MR safety• Virtuallymaintenance‐free• Simple to install• Magnet is always ON, always ready to image
• Located in the SPF at Rappaport : enables follow‐up scanning
5
-
Tel. 04-8295361/364 [email protected]
• Frommouse (or rat) to man• Easy to Learn• 1 Tesla field strength: most suitable for Gadolinium based contrast agent imaging
• Easy to Use• High‐performance: Superior (100µm) resolution, 2D and 3D imaging
• Fast
6
-
Tel. 04-8295361/364 [email protected]
• Scanned size: rat and mouse only, up to 60mm in diameter• Field strength of 1 Tesla: “costs” longer durations or lower resolution• No option for functional imaging or DTI• No option for cardiac gating
7
-
Tel. 04-8295361/364 [email protected]
8
-
Tel. 04-8295361/364 [email protected]
• No external fringe field: metal objects can be in the proximity of the system.
• Safe for PCs or mobile phones• Aspect system is safer to operate
** CAUTION: Metal objects cannot be placed or inserted directly inside Aspect Imaging magnets.Aspect system is still a powerful magnet and placing metal objects inside the bore can create irreparable damage to the system.Metal objects must be kept outside of the magnet’s 5 gauss line.
9
-
Tel. 04-8295361/364 [email protected]
• The Power of 1.5 Tesla (GE)
10
-
Tel. 04-8295361/364 [email protected]
11
-
Tel. 04-8295361/364 [email protected]
• Welding tank
12
-
Tel. 04-8295361/364 [email protected]
13
-
Tel. 04-8295361/364 [email protected]
• Cancer• Cardiovascular• Contrast‐based Molecular Imaging• Diabetes and Obesity• Embryology / Developmental Biology• Nephrology• Neurobiology• Stem Cell / Cell Tracking• Histology (PM / Ex‐vivo)
14
-
Tel. 04-8295361/364 [email protected]
1. Gadolinium (Gd) – Paramagnetic:Increases T1 signal– Extracellular fluid agents– Blood pool agents– Organ‐specific agents2. Iron oxide – Superparamagnetic:Reduces T2 signals3. Iron Platinum – Superparamagnetic:Reduces T2 signals4. Manganese – Paramagnetic:Increases T1 signal
15
-
Tel. 04-8295361/364 [email protected]
FSE, TR/TE=1918/76 ms, 1 mm slice, NEX=2, 176X176, FOV=64 mm
Movie: Photo:
T2W T2W
16
-
Tel. 04-8295361/364 [email protected]
Control (T1w+C): CHF: T1W+C
Zaid Aba
ssi’s
Lab
17
-
Tel. 04-8295361/364 [email protected]
Pathology: Control:
FSE, TR/TE=2273/74 ms, 1 mm slice, NEX=4, 256X256, FOV=64 mm
FSE, TR/TE=1705/80 ms, 1.3 mm slice, NEX=4, 256X256, FOV=64 mm
T2W T2W
18
Yuval Sha
ked’sL
ab
-
Tel. 04-8295361/364 [email protected]
FSE, TR/TE=3000/80 ms, 0.8 mm slice, NEX=4, 200X200, FOV=40 mm
T2W
19
-
Tel. 04-8295361/364 [email protected]
FSE, TR/TE=3400/93 ms, 1 mm slice, NEX=2, 200X200, FOV=64 mm
T2W
20
-
Tel. 04-8295361/364 [email protected]
SE‐DWI, TR/TE=500000/19 ms, 1.5 mm slice, NEX=1, 128X128, FOV=64 mm, BIGDELTA = 10, DIFFMAG = 961, b=1 and b=300
21
T2WDWI
-
Tel. 04-8295361/364 [email protected]
IR‐SE, TR/TE=500/9 ms, TI=100 ms, 1 mm slice, NEX=1, 128X128, FOV=100 mm
Water
Fat
T1W
22
-
Tel. 04-8295361/364 [email protected]
FSE, TR/TE=1700/102, ST=0.7mm, NEX=22, 176X176, FOV=40 mm
Ovary
Kidney
OvaryKidney
23
-
Tel. 04-8295361/364 [email protected]
Pre‐injection Post injection
GRE‐SP, TR/TE=12/3, ST=0.9mm, NEX=6, 128X128, FOV=45 mm
GRE‐SP, TR/TE=12/3, ST=1mm, NEX=4, 128X128, FOV=40 mm
Yuval Sha
ked’sL
ab
Pre‐injection Post injection
24
-
Tel. 04-8295361/364 [email protected]
GRE‐SP, TR/TE=12/3, ST=0.9mm, NEX=5, 160X160, FOV=64 mm 25
-
Tel. 04-8295361/364 [email protected]
Lung tumor:
T2W
FSE, TR/TE=2080/60 ms, 0.9 mm slice, NEX=2, Ext. Ave=3, 256X256, FOV=50 mm
Control:
26
-
Tel. 04-8295361/364 [email protected]
Vlodavsky’s Lab
Tumor imaging
27
FSE, TR/TE=2080/60 ms, 0.9 mm slice, NEX=2, Ext. Ave=3, 256X256, FOV=50 mm
-
Tel. 04-8295361/364 [email protected]
Magnetic
Resonance
Imaging
28
-
Tel. 04-8295361/364 [email protected] 29
• Proton has magnetic moment – behaves like a little magnet
• There are a lot of protons in the body - ~70% water , x% fat
• They create ‘net magnetization’
• Largest Gyro Magnetic Ratio 1T 42.57 MHz
-
Tel. 04-8295361/364 [email protected]
B0 = Big Magnetic Field Produced by Main Magnet(1.5 Tesla ~ 30,000 times earth magnetic field)
Aligns ‘net magnetization’ in the direction of the field
30
-
Tel. 04-8295361/364 [email protected]
• Small B0 produces small net magnetization (M)
• Thermal motions “try” to randomize alignment of proton magnets
• Larger B0 produces larger net magnetization M, lined up with B0
(Reality check: 0.0003% of protons are perfectly aligned per Tesla)
31
-
Tel. 04-8295361/364 [email protected]
32
Excitation: Relaxation:
-
Tel. 04-8295361/364 [email protected] 33
RF energy Excitation
Relaxation
Energy
Energy
Applying RF Energy
Only protons that spin with the same resonance frequency as the RF pulse will respond
The result: A shift of the net magnetization vector away from the direction of the static B0 field in a flip angle
Note: The RF signal intensity/duration determines the flip angle of M.
-
Tel. 04-8295361/364 [email protected] 34
RF energy Excitation
Relaxation
Energy
Energy
The return of net magnetization to the direction of the static B0 field
Accompanied by energy release
Note: Left alone, net magnetization will align itself with B0 in about 2–3 sec
-
Tel. 04-8295361/364 [email protected]
35
Spin‐ lattice relaxation:•Microscopically, the spins exchange energy with the environment (the “lattice"). The time constant of this relaxation type is called T1.
•It is related to Mz recovery.
Spin‐spin relaxation•The spins in the high and low energy state exchange energy but do not loose energy to the surrounding lattice. The time constant of this relaxation type is called T2.
•It is related to Mxy decay.Note: T2 decay is faster than T1 recovery
-
Tel. 04-8295361/364 [email protected]
T1: Recovery of Mz– Usually 500-1000 ms in the brain
T2: Disappearance of Mxy only due to molecular interactions– Usually 50-100 ms in the brain
T2*: Disappearance of Mxy due to molecular interactions and variations in B0– Measured in practice
Note: Drugs containing magnetic impurities can alter T1, T2, and T2* — contrast agents
36
-
Tel. 04-8295361/364 [email protected]
1) Put subject in big magnetic field (leave him there)
2) Excitation: transmit radio waves into subject [about 3 ms]
3) Turn off radio wave transmitter
4) Relaxation: receive radio waves re-transmitted by subject
– Readout interval :10-100 ms - MRI is not a snapshot!
5) Store measured radio wave data vs. time
– Repeat (2) to obtain more data and average
6) Process raw data to reconstruct images
7) Allow subject to leave scanner (optional)37
-
Tel. 04-8295361/364 [email protected]
After the excitation the protons remains with energy in the form of a flip angle.
During relaxation, an oscillating voltage is generated and received by a coil of wires placed around the subject.
This voltage is the RF signal, whose measurements form the raw data for MRI.
– At each instant in time, it is possible to measure only one combined voltage V(t), that includes the sum RF signals.
Must find a way to separate signals from different regions38
-
Tel. 04-8295361/364 [email protected]
• 3D encoding:– Z‐dir: frequency encoding for slice selection– X‐dir: phase encoding– Y‐dir: frequency encoding with readout
• K‐Space ‐ Fourier Transform
39
-
Tel. 04-8295361/364 [email protected]
Extra static Gradient Magnetic Fields (in addition to B0) that linearly vary in intensity across the subject
This allows frequency encoding : the excitation frequency changes with location
x-axis
f
60 KHz
Left = –7 cm Right = +7 cm
Central frequency(63 MHz at 1.5 T)
(* scan noise: turning the gradient fields ON/OFF)
40
-
Tel. 04-8295361/364 [email protected]
z
RF freq. bandwidth
Excited location
Slice profile
Frequency
Applying a gradient field Simultaneous RF excitation corresponding to a narrow range of resonance
frequencies
41
-
Tel. 04-8295361/364 [email protected]
The phase of the signal depend on location This is done by applying a gradient field in the 2nd
direction The phase encoding gradient (GPE) intervenes for a
limited time period. While it is applied, it modifies the spin resonance
frequencies. When turning the gradient off, the spin resonance
frequencies return to the original one, but now with a phase change.
The process has to be for each column of the image (time consuming).
This phase difference lasts until the signal is recorded.
Applying gradient field in the 3rd direction. Simultaneous with acquisition (readout) of re-transmitted signals. Signal is sampled about once every microsecond, digitized, and stored in a
computer. 42
-
Tel. 04-8295361/364 [email protected] 43
-
Tel. 04-8295361/364 [email protected]
Input signal: Output signal:
44Phase encoding
Freq
uency en
coding
X [mm]
Y [m
m]
-
Tel. 04-8295361/364 [email protected]
Echo time, TE, is the time from the RF excitation to the center of the echo being received. Shorter echo times allow less T2signal decay.
Repetition time, TR, is the time between one acquisition and the next. Short TR values do not allow the spins to recover Mz, so the net magnetization available is reduced, allowing T1 weighted imaging.
Short TE and long TR give strong signals.
45
-
Tel. 04-8295361/364 [email protected]
)1(),( 21 //0 TTETTR eeMTETRS
TE TR Image WeightingShort Long Proton Density Short Short T1 Long Long T2
46
-
Tel. 04-8295361/364 [email protected]
T1 ContrastTE = 14 msTR = 400 ms
T2 ContrastTE = 100 msTR = 1500 ms
Proton DensityTE = 14 msTR = 1500 ms
47
-
Tel. 04-8295361/364 [email protected]
48
T1w images T2w imagesDark: ‐ Increased water: edema, tumor,
infarction, inflammation, infection, hemorrhage
‐ Low proton density‐ Calcification‐ Flow void
‐ Low proton density‐ Calcification‐ Fibrous tissue‐ Paramagnetic substances ‐ Protein‐rich fluid‐ Flow void
Bright: ‐ Fat‐ Subacute hemorrhage‐ Melanin‐ Protein‐rich fluid‐ Slowly flowing blood‐ Paramagnetic substances (gadolinium)
‐ Laminar necrosis of cerebral infarction
‐ Increased water: edema, tumor, infarction, inflammation, infection, hemorrhage , subdural collection
‐ Methemoglobin (extracellular) in subacute hemorrhage
-
Tel. 04-8295361/364 [email protected]
MRI has high contrast for different tissue types!
Tissue T1 (ms) T2 (ms)
Grey Matter (GM) 950 100
White Matter (WM) 600 80
Muscle 900 50
Cerebrospinal Fluid (CSF) 4500 2200
Fat 250 60
Blood 1200 100‐200
49
T1 and T2 relaxation time of tissue depend on three factors: a. The inherent energy of the tissueb. How closely packed the molecules arec. How well the molecular tumbling (“self‐movement”) rate matches the Lamar frequency of
hydrogen. The molecule tumbling rate is determined by: Molecular weight, temperature, shape of the molecule
-
Tel. 04-8295361/364 [email protected]
• Higher Signal to Noise Ratio (SNR) results in a better image
–K – constant (related among others to TR,TE, T1,T2…)–FOVx , FOVy - the field-of-view in the x and y directions–Nx , Ny - the number of frequency and phase encoding steps–∆z - the slice thickness–NEX (or NA) - the number of signal averages–BW - the receive bandwidth
50
-
Tel. 04-8295361/364 [email protected]
• Total scan time is given by:
Npe - the number of phase encoding stepsS – number of slices
∝ · · ·
51
-
Tel. 04-8295361/364 [email protected]
• Spin Echo Based Sequences– SE– Fast Spin Echo (FSE)– Inversion Recovery Spin Echo (IR‐SE)– Spin Echo Diffusion Weighted Imaging (SE‐DWI)
• Gradient Echo Based Sequences– Gradient Echo Spoiled (GRE‐SP)– Gradient Echo Spoiled External Averaging (GRE‐EXT)– Gradient Echo Steady State; ESS/FESS (GRE‐SS)
52
-
Tel. 04-8295361/364 [email protected]
Rephase spins with a 180° pulse: Prevents Spins from rotating at different rates in different
locations Overcomes spins dephasing with time
Most sensitive if TE average T2 In Diffusion: refocusing will not be perfect for protons that
have moved
Used for measuring T2:
53
time
e-t/T2*
e-t/T2
MR
sig
nal
Fast Spin Echo:
TETE/2
-
Tel. 04-8295361/364 [email protected]
1. Equilibrium 2. 90 Pulset=0
3. Spin Dephasing
4. 180 Pulset=TE/2
5. Spin echot=TE 54
-
Tel. 04-8295361/364 [email protected]
55
-
Tel. 04-8295361/364 [email protected]
Changing the polarity of the Gradient: Dephase and rephase the
spins Shorter TR/TE decreases
scan time
Most sensitive if TE average T2* time
e-t/T2*
MR sig
nalUsed for measuring T2*
56
TEGradientON
-
Tel. 04-8295361/364 [email protected]
57
-
Tel. 04-8295361/364 [email protected]
Tool features:‐ ROI measurements‐ Volume measurements‐ Series synchronization & comparison‐ T1, T2, T2* and Diffusion mapping‐ Noise filtering
58
-
Tel. 04-8295361/364 [email protected]
GRE, TR/TE [ms]=40/5, 40/8, 40/10, 40/12, 40/15, 1 mm slice, NEX=1, 200X200, FOV=64 mm
T2*W
MATLAB MRI analysis tool for T2* mapping:T2*=18.5689 ms
59
-
Tel. 04-8295361/364 [email protected]
IR‐SE, TR/TE=800/8 ms, TI=400, 300, 200 ms, 1.5 mm slice, NEX=2, 128X128, FOV=64 mm
200 250 300 350 40016
18
20
22
24
26
28
30
32
34
36
TE [msec]
T1 m
agni
tude
T1 mapping
Real DataMAP T1=254.9848
MATLAB MRI analysis tool for T1 mapping:T1=254.98 ms
T1W
60
-
Tel. 04-8295361/364 [email protected]
Installation: • Copy the K‐PACS software from the Tech‐MED ‐MRI server (Aspect folder K‐PACS folder) to your computer and install.
Loading your DICOM files: • Double‐click on the software icon and press accept.• Press Filesystem and select the preferred folder.
• The selected data will be saved in the software database for future viewing.• Mark the folder or scan you would like to view:
• Press viewer 61
-
Tel. 04-8295361/364 [email protected]
Once the viewer is opened, select the scan you would like to see from the scans presented in the upper part of the page:•• Tools: • All tools are presented once you stand on the bottom part of the image or in the upper icon roller.
• Available options: Measure ROI, Zoom, rotate, magnifier, print, export to tiff, Avi etc. 62
-
Tel. 04-8295361/364 [email protected]
• Set a meeting with Edith to discuss the application• Prepare a literature review on other’s MR protocols for similar application• Begin feasibility test (small pilot) with Aspect M2™ – easy to operate• Analyze the results with our MRI tool – continuously developed according to researchers needs
63
-
Tel. 04-8295361/364 [email protected]
• Evert J Blink, “MRI: Physics”• http://www.revisemri.com/questions/basicphysics/• http://www.imaios.com/en/e‐Courses/e‐MRI/• http://www.simplyphysics.com/flying_objects.html• http://quizlet.com/14282869/mri‐physics‐chapter‐2‐flash‐cards/• http://www.tele.ucl.ac.be/PEOPLE/OC/these/node100.html• http://en.wikipedia.org/wiki/Shim_(magnetism)• http://saturn.med.nyu.edu/research/sb/turnbulllab/PDFS/JohnsonMRM99.pdf
• http://www.med.harvard.edu/aanlib/basicsMR.html
64
-
Tel. 04-8295361/364 [email protected]
65