Thermo-regulation of pediatric patients has many applications, and the Tecotherm Neo has been
designed to meet medical conditions that need effective temperature management. With the flexibility
to cool, warm or maintain normothermia of patients up to 50 kg, the Tecotherm Neo can be used
anywhere from the NICU, to pediatric emergency rooms for short or long term applications.
The innovative servo controlled design has instant feedback, monitoring the infant’s temperature every 2
seconds and making minute changes to the cooling fluid to ensure that the infant’s temperature remains
stable. Simple alarms are on hand, should the temperature deviate more than 0.5° C from the set temperature.
The large color display shows the temperature clearly will change color to alert the nursing staff to any
problem.
The Tecotherm Neo can be programmed for a complete treatment cycle. The user can set the target
temperature, duration, and even the rate of warming / cooling, making it easy and simple to set up and
use. Changes can be made at any time and all changes set, and measured parameters are recorded on to a
memory card every minute for later analysis.
Tecotherm Neo
Features
Modes of Operation
The Tecotherm Neo has 3 modes of operation:
Servo-Control Complete Treatment ModeThe Clinician can set the duration of cooling, the rate of rewarming, as well as the target temperature. The Tecotherm Neo will then complete the whole cycle without interruption.
Servo-Control Mode(Constant Rectal Temperature)The Clinician can set the target temperature, the time to get to the target temperature and the length of time to maintain the target temperature. The Tecotherm Neo will complete the whole cycle without interruption.
Rate of change can also be set to allow even warming and cooling.
Constant Mattress Temperature ModeNo Servo Control: useful for research work when control of the mattress needs to be stable, or where experienced staff are available to monitor the patient.
Examples of patient data collected from Tecotherm Neo
Infant transfer temperature 35.3°C
Target temperature 33.5°C (72 hours)
Target temperature range 33.5°C (+/- 0.5°C)
Target rewarming temperature 37°C(12 hours/0.29°C per hour)
3 modes of operation for maximum flexibility in use
Clinician can set and adjust parameters at the start and throughout treatment
Data storage for allowing simple analysis of temperature profiles
USB output
Large color screen is easy to see across the room
Internal battery restarts from same settings in event of power failure
Compatible with reusable or disposable rectal probes
Optional use of skin probes
Simple sterile water
Tecotherm Specifications
Dimensions: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . H 210 mm (8¼ in), W 373 mm (14¾ in), D 310 mm (12¼ in)Weight (Reservoir Full): . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.2 kg (15.9 lbs)Weight (Reservoir Empty: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.95 kg (15.3 lbs)Screen Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 640 x 480 pixelsElectrical Power Consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.5 A/ 230 VAC / 3.1 A/ 110 VAC Max. 345 WNoise Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 - 54.5dB (Mean Value 1m Distance)Fluid Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .250 ml (inner tank) 500 ml (fill up set)Minimum Fluid Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12° CMaximum Fluid Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39° CMaximum Patient Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50kg / 110.2 lbs (For larger patients, please contact your local representative)Data Storage Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2GB (50,000 hours) All set and measures parameters
Spare Parts and Accessories
Aqua Wrap Around Mattress - Reusable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P02-003Aqua Wrap Mattress - Disposable 3/pk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P02-004Mattress Connecting Hoses 1m . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P03-001Mattress Connecting Hoses 2m . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P03-002Felt Ties (Case of 20) for use with Wrap Around Mattress . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P05Inter Layer Foils (Case of 10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P06Reusable Skin Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P07-001Disposable Rectal / Esophageal Probe Non Sterile (case of 20) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P08Reusable Adapter Cable for use with Disposable Rectal Probe (single) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P09Trolley for Tecotherm Neo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .R505P10Fill Up Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P11Mattress Repair Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .R505P12
Starter Kit Includes: Tecotherm NeoMain CablesDisposable Aqua Wrap (Case of 3)2m Mattress Connecting HosesFill Up Set1 x Reusable Skin Probe1 x Reusable Adapter Cable (Disposable Probe Kit only)Disposable Rectal / Esophageal Probe (Case of 20)Instruction Manual
2305 South 1070 WestSalt Lake City, Utah 84119
500-113(1 US) The details given in this leaflet are correct at time of going to press. The company reserves the right to improve the products shown.
Manufactured By:
Exclusive US Distributor for Tecotherm Neo:
Part Number
Ordering Information
Disposable Starter Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R505P01-001
Part Number
866.4.Maxtec www.maxtec.com
ML#269 Rev.D