![Page 1: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/1.jpg)
Technological Advances in Radiotherapy
Samuel ChiuDepartment of Clinical Oncology
Prince of Wales HospitalHong Kong
HA Convention
May 2007
![Page 2: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/2.jpg)
Discovery of X ray1895• German Physics Professor
Wilhem Conrad Roentgen discovered X ray when he noticed that only the bones of his hand was shown on the fluorescent screen placed at the path of a beam produced by the Crooke’s tube
1901• Get the Nobel Prize
![Page 3: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/3.jpg)
The beginning of Radiotherapy
1904• Thomas Edison’s assistant died of
repeated exposure to x rays causing severe burns with several amputations
• Noticed the harmful effects of x rays and marks the beginning of the potential use of x rays in radiotherapy
![Page 4: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/4.jpg)
Development of Radiotherapy Machines
Early half of twentieth century
• Orthovoltage machines • low energy (200-300kv) x-rays
machines for treatment of cancer
• Low penetration• High toxicities
![Page 5: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/5.jpg)
Development of Radiotherapy Machines
1951-The first high energy Cobalt 60 machines
• gamma rays, high energy.
• Deep penetration• Allows more widely
use of x rays in treatment of cancers
![Page 6: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/6.jpg)
Development of Radiotherapy Machines
1970’s• Development of Linear
accelerators • accelerate electrons,
then slow them down through their collision with a metal target, causing high energy x-rays to be released
• Marks the beginning of technological development of teletherapy and widely use of external beam irradiation in treatment of cancers
![Page 7: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/7.jpg)
Two dimensional radiotherapy in 80’s
![Page 8: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/8.jpg)
Two dimensional RTTwo fields Three fields
HIGH DOSE to SURROUNDING NORMAL TISSUE
HIGH TOXICITIES LIMITING RADIATION DOSE
![Page 9: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/9.jpg)
RT late toxicityBrain as an example
• Focal injury Necrosis (related to the high dose region)– Optic neuropathy– Pituitary dysfunction
• Neuropsychologic/cognitive (related to the whole brain dose)– Volume, fraction, size dose, age
dependent
![Page 10: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/10.jpg)
How to decrease late toxicities
• Focal injury (Better Conformity to decrease volume and dose to critical organs)
• Neuropsychologic/cognitive (Increase conformity and number of beams to decrease volume of high dose zone)
![Page 11: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/11.jpg)
Three dimensional Conformal Radiotherapy
(3DCRT)
Computer Planning
CT scan
MultileafCollimator
PWH RTPWH RT
ImmobilisationDevice
![Page 12: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/12.jpg)
3 dimensional conformal RT (3DCRT) – 90’s
•Imaged guided target localisation•Computer 3D planning•Multiple beams•High conformity•Spare surrounding critical organs•Less dose to surrounding brain
![Page 13: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/13.jpg)
2DCRT 3DCRT
High conformity decrease risk of focal injuries
Decrease high dose volume
Decrease cognitive impact
![Page 14: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/14.jpg)
3-DCRTConventional 2-D
3D CRT treatment Dosimetric advantage over 2D
LimitationSuboptimal for irregular Targets (Concave shape)
Brainstem
![Page 15: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/15.jpg)
60Gy Tumour
50Gy Chiasm
54Gy brainstem
3DCRT cannot produce an isodose distribution to exactly to what we want ?
70Gy Highly proliferativetumour
![Page 16: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/16.jpg)
Limitation
The planner starts with a set of beam weights and profiles to obtain a plan by trial-and-error process.
3DCRT - Conventional (Forward) Planning
![Page 17: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/17.jpg)
Based on two new principles:
1. Inverse treatment planning
2. Intensity modulated treatment beams
Intensity Modulated Radiotherapy (IMRT)
![Page 18: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/18.jpg)
•The planner define the required dose & dose distribution
•The computer can calculate and optimized the beam intensity patterns of the individual IMRT beams.
Principle of Inverse Planning
70Gy to tumor
54Gy to Brainstem
![Page 19: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/19.jpg)
90Gy to visible tumour by imaging
81Gy to subclinicaldisease
60Gy to urethra to prevent stenosis
![Page 20: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/20.jpg)
Intensity modulated radiotherapy (IMRT)-the use of optimised multiple non uniform radiation beam intensities with multiple small beamlets
![Page 21: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/21.jpg)
IMRTDynamic Multileaf Collimator
![Page 22: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/22.jpg)
Methods of IMRT Treatment Delivery
![Page 23: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/23.jpg)
3-DCRT IMRTConventional 2-D
IMRT Dosimetric advantage over 2D and 3D CRT treatment
High conformity
![Page 24: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/24.jpg)
IMRT Dosimetric AdvantageNon uniform dose
Higher dose to Tumor
Lower dose to subclinical diseae
![Page 25: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/25.jpg)
Potential Benefits of IMRT
1 Deliver treatments with better dose conformity and coverage to the target,hence reducing the probability of in-field recurrence.
2 Better sparing and protection of normal tissue, hence minimise the degree of morbidity associated with treatment.
3 Facilitating the escalation of dose to improve local control.
![Page 26: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/26.jpg)
Gy
2DRT IMRT
Gy 1. 2D-Ho’s2. 3DCRT3. IMRT
Cumulative Dose Volume HistogramStructure :Rt Parotid
Parotid Gland
Kam et al, Asco 2006
Clinical Oncology, PWH
Parotid Sparing
![Page 27: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/27.jpg)
RTOG/EORTC Xerostomia Grade
Grade ≧2 (%) Chi-Square
2DRT IMRT p-value
6 week 85.7 46.4 0.0019
6 month 92.9 75 0.0689
1 year 82.1 39.3 0.0010
85.7
46.4
92.9
7582.1
39.3
0102030405060708090
100
Xer
osto
mia
Gra
de ≧
2(%
)
6 week 6 month 1yearTime after RT
2DRTIMRT
![Page 28: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/28.jpg)
Dose Escalation For Ca prostate
Decrease in biochemical failure with increase dose
Decrease Grade 2 rectal complications using IMRT
Alan Pollack et al.
World J Urol (2003)
![Page 29: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/29.jpg)
Problems & Challenges - IMRT
1. High precision & sophisticated treatment planning procedures
2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral dose- Increases the probability
of radiation induced secondary cancer in the irradiated volumes.
5. The effect of geometrical errors on treatment dosimetry is more serious then conventional treatments.
![Page 30: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/30.jpg)
Geometric Errors in RadiotherapyInter-fraction treatment errors1 Set up errors2 Organ positional changes3 Integrity of immobilization cast4 Changes in patient physical conditions 5 Machine errors
Intra-fraction treatment errors 1 Patient movement2 Organ movement caused by e.g. breathing
![Page 31: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/31.jpg)
Conventional setup only relies on laser to line up surface markers on skin/cast disregard the positions of internal organs and the targetInter-fraction treatment errors
![Page 32: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/32.jpg)
Better Immobilization DeviceHelp to decrease setup errors but not the ultimate solution
PWH RTPWH RT
![Page 33: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/33.jpg)
Irradiated VolumeTreated Volume
PTV
ITV
CTV
GTV
ICRU Report 50 & 62
Internal margin
Setup margin
Penumbra
Geometrical Uncertainties
Definition of Target Volumes Definition of Critical Normal Structures
OAR
PRV
![Page 34: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/34.jpg)
Image-Guided Radiation Therapy (IGRT)
![Page 35: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/35.jpg)
What is IGRT?
• Measure and correct target and critical structure positional errors immediately prior to or duringtreatment delivery
• To reduce ITV and / or Setup inaccuracy thus decrease RT related toxicity and allow tumourdose escalation
![Page 36: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/36.jpg)
IGRT
• Positioning of patient based on internal bone landmark or implanted marker using X rays
![Page 37: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/37.jpg)
IGRT – Acquisition of CT scan images
Allow reconstruction of images in different planes
![Page 38: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/38.jpg)
IGRT
1st set of X ray images
2nd set of X ray images
![Page 39: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/39.jpg)
IGRT
•Fusion of x ray images to the digitally reconstructed images from planning CT
•Patient aligned to correct position based on bony landmarks/implanted markers
![Page 40: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/40.jpg)
IGRTPositioning of patient based on internal bone landmark or
implanted marker using X rays
Courtesy of BrainLab
![Page 41: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/41.jpg)
IGRT
Positioning of patient with CT guided
Helical Tomotherapy
![Page 42: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/42.jpg)
LINAC and CT Scan in the same machine
Multileafcollimator with IMRT
No limitation on field size and field matching
![Page 43: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/43.jpg)
On-line treatment verification by matching of the planning CT image (bottom left) with the tomotherapy treatment set up image (top left) immediately before treatment delivery without moving the patient.
TomotherapyCT image guided treatment delivery
Courtesy of Peter Teo, HKS & Hospital
![Page 44: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/44.jpg)
Intra-fraction Errors
Tumour moves with respiration. Range: 0.5 ~3cm
![Page 45: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/45.jpg)
Convention RT: Need extra margin for tumour motion – Internal Target Volume (ITV)
![Page 46: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/46.jpg)
What is IGRT?
• Measure and correct target and critical structure positional errors immediately prior to or duringtreatment delivery
• To reduce ITV and / or Setup inaccuracy thus decrease RT related toxicity and allow tumourdose escalation
![Page 47: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/47.jpg)
Respiratory Gated RT: Less extra margin for tumour motion – Decreased ITV
![Page 48: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/48.jpg)
IGRT Equipments – PWH – Respiratory Gating
• Real-time Positioning Management (RPM) system (Varian)
• 4D-CT
• Respiratory Gated RT
![Page 49: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/49.jpg)
Gating of radiotherapy treatment beams by respiratory motion waveform to compensate for target movement
Beam On
Beam On Window
Beam off
Breathing signal
![Page 50: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/50.jpg)
Liver Volume: 1010 cm3
GTV= 119 cm3
![Page 51: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/51.jpg)
Liver Volume: 1010 cm3
GTV= 119 cm3
PTV = 684 cm3
without Gating Txmargins: SI=2.5cm, circumferential=1.5cm
![Page 52: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/52.jpg)
Liver Volume: 1010 cm3
GTV= 119 cm3
PTV = 684 cm3
without Gating Txmargins: SI=2.5cm, circumferential=1.5cm
PTV = 286 cm3
with Gating Txmargins: 0.75cm for all
![Page 53: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/53.jpg)
Liver Volume: 1010 cm3
GTV= 119 cm3
PTV = 684 cm3
without Gating Txmargins: SI=2.5cm, circumferential=1.5cm
PTV = 286 cm3
with Gating Txmargins: 0.75cm for all
PTV VolumeReduction: 58%
![Page 54: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/54.jpg)
Gating of respiratory motionExternal infrared markers and internal implanted markers
Courtesy of BrainLab
![Page 55: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/55.jpg)
REAL-TIME tumor-tracking - CyberknifeCorrect for respiratory tumor motion
Robot motion is steered by external fiducial motion
![Page 56: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/56.jpg)
Internal tumor motionX-ray images
The Synchrony™ Respiratory Tracking SystemBuilding the correspondence model
Implanted gold seeds
![Page 57: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/57.jpg)
External fiducial motionInfrared LED-camera system
The Synchrony™ Respiratory Tracking System
Building the correspondence model
![Page 58: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/58.jpg)
The Synchrony™ Respiratory Tracking System
Robot follows internal fiducials using a correspondence modelbetweenmeasured internal fiducial motion (implanted gold markers)
andmeasured external fiducial motion (LEDs)Correspondence model is build prior to the start of treatmentCorrespondence model is verified and updated during the treatment
LED position
Inte
rnal
fidu
cial
x,y,
z (m
m)
Correspondencemodel
![Page 59: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/59.jpg)
Surgery
Multidisciplinary Management of Cancers
Radiotherapy
Department of Clinical Oncology, PWH
Chemotherapy Biological Therapy
Pathology
Radiology
Clinical Psychology Palliative Medicine
![Page 60: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/60.jpg)
Conclusion
• The rapid technological advancement brings radiotherapy to a new era
• Development of radiotherapy from two dimensional to three dimensional conformal radiotherapy achieve high conformity of radiation to the target with sparing of surrounding normal tissue and allow dose escalation with improvement in tumor control
![Page 61: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/61.jpg)
Conclusion• Intensity modulated radiotherapy (IMRT)
further improves the conformity of radiation especially for the irregular targets, allows the delivery of differential doses within the target according to the tumor characteristics and dose limitations of critical organs
• The development of these highly conformed radiotherapy further facilitated by the development of imaged guided radiotherapy (IGRT) with or without respiratory gating and assures accurate setup, decrease margins and decrease volume of irradiation
![Page 62: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/62.jpg)
Conclusion
• Treatment of cancers require multidisciplinary management. Good coordination among the different disciplines and careful selection of patient will allow provision of best services to our patients.
![Page 63: Technological Advances in Radiotherapy€¦ · 1. High precision & sophisticated treatment planning procedures 2. High cost 3. Radiation safety- higher x ray exposure 4. Higher integral](https://reader035.fdocuments.us/reader035/viewer/2022071008/5fc68cbf689dad5df56e21c0/html5/thumbnails/63.jpg)