Technical specificationFLUXUS WW
TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
Features
• Precise bi-directional and highly dynamic flow measurement with the non-invasive clamp-on technology
• High precision at fast and slow flow rates, high temperature and zero point stability
• Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results
• User-friendly design
• Transducers available for a wide range of inner pipe diameters
• ATEX, IECEx, FM Class I Div. 2 approved transducers for haz-ardous areas available
• Measurement is unaffected by fluid density, viscosity and solid content (max. 10 % of volume)
Applications
• Water and wastewater industry
FLUXUS WW
Water and wastewater ultrasonic flowmeter for permanent installation
Transmitter for permanent outdoor wall or pipe mounting
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu2
Table of contents
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Number of sound paths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Typical measurement setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2" pipe mounting kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Transducer selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Transducer order code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Transducer mounting fixture. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Coupling materials for transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Connection systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Junction box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2" pipe mounting kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Technical specification FLUXUS WW
3TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
Function
Measurement principleThe transducers are mounted on the pipe which is completely filled with the fluid. The ultrasonic signals are emitted alter-nately by a transducer and received by the other. The physical quantities are determined from the transit times of the ultra-sonic signals.
Transit time difference principle
As the fluid where the ultrasound propagates is flowing, the transit time of the ultrasonic signal in flow direction is shorterthan the one against the flow direction.
The transit time difference, ∆t, is measured and allows the flowmeter to determine the average flow velocity along the pro-pagation path of the ultrasonic signals. A flow profile correction is then performed in order to obtain the area averaged flowvelocity, which is proportional to the volumetric flow rate.
Two integrated microprocessors control the entire measuring process. This allows the flowmeter to remove disturbance si-gnals, and to check each received ultrasonic wave for its validity which reduces noise.
HybridTrek
If the gaseous or solid content in the fluid increases occasionally during measurement, a measurement with the transittime difference principle is no longer possible. NoiseTrek mode will then be selected by the flowmeter. This measurementmethod allows the flowmeter to achieve a stable measurement even with high gaseous or solid content.
The transmitter can switch automatically between transit time and NoiseTrek mode without any changes to the measure-ment setup.
Calculation of volumetric flow rate
= kRe · A · ka ·
where
Path of the ultrasonic signal in the flowing fluid
Transit time difference ∆t
- volumetric flow rate
kRe - fluid mechanics calibration factor
A - cross-sectional pipe area
ka - acoustical calibration factor
∆t - transit time difference
tγ - average of transit times in the fluid
∆tt0
V· t2 t-----------
V·
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu4
Number of sound pathsThe number of sound paths is the number of transits of the ultrasonic signal through the fluid in the pipe. Depending on thenumber of sound paths, the following methods of installation exist:
• reflection arrangementThe number of sound paths is even. Both of the transducers are mounted on the same side of the pipe. Correct positio-ning of the transducers is easier.
• diagonal arrangementThe number of sound paths is odd. Both of the transducers are mounted on opposite sides of the pipe. In the case of a high signal attenuation by the fluid, pipe and coatings, diagonal arrangement with 1 sound path will be used.
The preferred method of installation depends on the application. While increasing the number of sound paths increasesthe accuracy of the measurement, signal attenuation increases as well. The optimum number of sound paths for the para-meters of the application will be determined automatically by the transmitter.
As the transducers can be mounted with the transducer mounting fixture in reflection arrangement or diagonal arrange-ment, the number of sound paths can be adjusted optimally for the application.
Typical measurement setup
Reflection arrangement, number of sound paths: 2
a - transducer distance
Diagonal arrangement, number of sound paths: 3
Diagonal arrangement, number of sound paths: 1 Diagonal arrangement, number of sound paths: 1, negative trans-ducer distance
Example of a reflection arrangement
a
a
a > 0 a < 0
transducers
transmitter
Technical specification FLUXUS WW
5TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
Transmitter
Technical data
FLUXUS WW
design standard field deviceapplication flow measurement of water and wastewatermeasurementmeasurement principle
transit time difference correlation principle,automatic NoiseTrek selection for measurements with high gaseous or solid content
flow velocity m/s 0.01...25repeatability 0.15 % of reading ±0.01 m/sfluid watertemperature com-pensation
corresponding to the recommendations in ANSI/ASME MFC-5.1-2011
accuracy1
• with standard cali-bration
±1.6 % of reading ±0.01 m/s
• with advanced cali-bration (optional)
±1.2 % of reading ±0.01 m/s
• with field calibrati-on2
±0.5 % of reading ±0.01 m/s
transmitterpower supply • 100...230 V/50...60 Hz or
• 20...32 V DC or
• 11...16 V DCpower consumption W < 15number of measuring channels
2
damping s 0...100 (adjustable)measuring cycle Hz 100...1000 (1 channel)response time s 1 (1 channel), option: 0.07housing material aluminum, powder coateddegree of protection IP65dimensions mm see dimensional drawingweight kg 3.1fixation wall mounting, optional: 2" pipe mountingambient temperature °C -40...+60 °C (< -20 °C without operation of the display)display 2 x 16 characters, dot matrix, backlightmenu language English, German, French, Dutch, Spanishmeasuring functionsphysical quantities volumetric flow rate, mass flow rate, flow velocitytotalizer volume, masscalculation functions average, difference, sumdiagnostic functions sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit timescommunication interfacesservice interfaces • RS2323
• USB (with adapter)3
process interfaces optional:
• RS485 (ASCII sender)
• Modbus RTUaccessoriesserial data kit• сable RS232• adapter RS232 - USBsoftware • FluxDiagReader: download of measured values and parameters, graphical presentation
• FluxDiag (optional): download of measurement data, graphical presentation, report generationdata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity > 100 000 measured valuesSD card, removable (optional)loggable values all physical quantities and totalized valuescapacity min. 2 GB1 for transit time difference principle, reference conditions and v > 0.15 m/s2 reference uncertainty < 0.2 %3 outside of explosive atmosphere (housing cover open)
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu6
Dimensions
in mm
outputsThe outputs are galvanically isolated from the transmitter.
• switchable current outputAll switchable current outputs are switched to active or passive mode at the same time.
number 2range mA 4...20 (3.2...22)accuracy 0.04 % of reading ±3 μAactive output Rext < 350 Ωpassive output Uext = 8...30 V, depending on Rext, Rext < 1 kΩ• binary outputnumber 3optorelay 26 V/100 mAbinary output as alarm output• functions limit, change of flow direction or errorbinary output as pulse output• functions mainly for totalizing• pulse value units 0.01...1000• pulse width ms 1...1000
FLUXUS WW
1 for transit time difference principle, reference conditions and v > 0.15 m/s2 reference uncertainty < 0.2 %3 outside of explosive atmosphere (housing cover open)
*704
70
10
ø4.5
163
20
0
288
265fixing holes for wall mounting
thread: 6x M20 x 1.5cable gland: max. 6x M20
83
Technical specification FLUXUS WW
7TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
2" pipe mounting kit
*704
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu8
Terminal assignment
*704
power supply
terminal connection (AC) connection (DC)PE earth earth
N(-) neutral -
L(+) phase +
transducers
transducer cable (transducers *****8*, ****LI*), extension cable transducer cable (transducers *****52)measuring channel A measuring channel B measuring chan-
nel Ameasuring chan-nel B
terminal connection terminal connection transducer terminal connectionAV signal BV signal X_AV X_BV SMB connector
AVS shield BVS shield
ARS shield BRS shield X_AR X_BR SMB connector
AR signal BR signal
outputs communication interfaces
terminal connection terminal connection communication interfa-ce
P1+...P2+P1-...P2-
current output A+ signal + • RS485
• Modbus RTUP5a...P7aP5b...P7b
binary output B- signal -
101 shield
Technical specification FLUXUS WW
9TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
Transducers
Transducer selection
transducer order code
FSG 400 500 4000 6500
FSK 100 200 2000 2400
FSM 50 100 1000 1200
FSP 25 50 400 480
FSQ 10 25 150 240
5 10 50 100 500 1000 5000inner pipe diameter [mm]
transducer order code
FSG 11
FSK 5
FSM 2
FSP 1
FSQ 0.6
5 10 15 20 25 30pipe wall thickness [mm]
recommended possible
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu10
Transducer order code
1, 2 3 4 5, 6 7, 8 9...11 no. of character
tra
nsdu
cer
tra
nsdu
cer
freq
ue
ncy
- am
bie
nt t
em
pera
ture
exp
losi
on
pro
tect
ion
con
nect
ion
sys
tem
- ext
en
sio
n ca
ble
/ opt
ion
description
FS set of ultrasonic flow transducers for liquids measurement, shear waveG 0.2 MHzK 0.5 MHzM 1 MHzP 2 MHzQ 4 MHz
N normal temperature rangeNN not explosion proofA2 ATEX zone 2/IECEx zone 2A1 ATEX zone 1/IECEx zone 1S1 inmetro zone 1K1 KOSHA zone 1F2 FM Class I Div. 2
TS direct connection or connection via junction boxXXX 0 m: without extension cable
> 0 m: with extension cableLC long transducer cableIP68 degree of protection IP68OS housing with stainless steel 316
Technical specification FLUXUS WW
11TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
Technical data
Shear wave transducers
Shear wave transducers (zone 2 - FM Class I Div. 2 - nonEx, TS)
order code FSG-N**TS/** FSK-N**TS/** FSM-N**TS/** FSP-N**TS/** FSQ-N**TS/** technical type C(DL)G1N52 C(DL)K1N52 C(DL)M2N52 C(DL)P2N52 C(DL)Q2N52transducer frequency MHz 0.2 0.5 1 2 4inner pipe diameter dmin. extended mm 400 100 50 25 10min. recommended mm 500 200 100 50 25max. recommended mm 4000 2000 1000 400 150max. extended mm 6500 2400 1200 480 240pipe wall thicknessmin. mm 11 5 2 1 0.6materialhousing PEEK with stainless steel cap 304
(1.4301), ***-*****/OS: 316L (1.4404)
PEEK with stainless steel cap 304 (1.4301), ***-*****/OS: 316L (1.4404)
contact surface PEEK PEEKtransducer cabletype 1699 1699length m 5, ***-*****/LC: 9 4, ***-*****/LC: 9 3, ***-*****/LC: 9dimensionslength l mm 129.5 126.5 64 40width b mm 51 51 32 22height h mm 67 67.5 40.5 25.5dimensional drawing
weight (without cable)
kg 0.47 0.36 0.066 0.016
ambient temperaturemin. °C -40 -40max. °C +130 +130temperature com-pensation
x
explosion protection• ATEX/IECExorder code FSG-NA2TS/** FSK-NA2TS/** FSM-NA2TS/** FSP-NA2TS/** FSQ-NA2TS/** explosion protection temperature (pipe surface)• min. °C -55 -55• max. °C gas: +190, dust: +180 gas: +190, dust: +180marking
Ex nA IIC T6...T2 GcEx tb IIIC TX Db
Ex nA IIC T6...T2 GcEx tb IIIC TX Db
certification ATEX IBExU10ATEX1163 X IBExU10ATEX1163 Xcertification IECEx IECEx IBE 12.0005X IECEx IBE 12.0005X• FMorder code FSG-NF2TS/** FSK-NF2TS/** FSM-NF2TS/** FSP-NF2TS/** FSQ-NF2TS/** explosion protection temperature• min. °C -40 -40• max. °C +125 +190marking NI/Cl. I,II,III/Div. 2 /
GP A,B,C,D,E,F,G/Temp. Codes dwg 3860
NI/Cl. I,II,III/Div. 2 /GP A,B,C,D,E,F,G/Temp. Codes dwg 3860
l
hb
l
hb
l
hb
l
hb
II3GII2D0637 II3G
II2D0637
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu12
Shear wave transducers (zone 2 - nonEx, TS, IP68)
order code FSG-N**TS/IP68 FSK-N**TS/IP68 FSM-N**TS/IP68 FSP-N**TS/IP68technical type CDG1LI8 CDK1LI8 CDM2LI8 CDP2LI8transducer frequency MHz 0.2 0.5 1 2inner pipe diameter dmin. extended mm 400 100 50 25min. recommended mm 500 200 100 50max. recommended mm 4000 2000 1000 400max. extended mm 6500 2400 1200 480pipe wall thicknessmin. mm 11 5 2 1materialhousing PEEK with stainless steel cap 316Ti (1.4571)contact surface PEEK
degree of protection IP68transducer cabletype 2550length m 12dimensionslength l mm 130 72width b mm 54 32height h mm 83.5 46dimensional drawing
weight (without cable)
kg 0.43 0.085
ambient temperaturemin. °C -40max. °C +100temperature com-pensation
x
explosion protection• ATEX/IECExorder code FSG-NA2TS/IP68 FSK-NA2TS/IP68 FSM-NA2TS/IP68 FSP-NA2TS/IP68explosion protection temperature (pipe surface)• min. °C -40• max. °C +90marking
Ex nA IIC T6...T2 GcEx tb IIIC TX Db
certification ATEX IBExU10ATEX1163 Xcertification IECEx IECEx IBE 12.0005X1 test conditions: 3 months/2 bar (20 m)/20 °C
1
l
hb
lh
b
II3GII2D0637
Technical specification FLUXUS WW
13TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
Shear wave transducers (zone 1, TS)
order code FSG-N*1TS/** FSK-N*1TS/** FSM-N*1TS/** FSP-N*1TS/** FSQ-N*1TS/**technical type C(DL)G1N81 C(DL)K1N81 C(DL)M2N81 C(DL)P2N81 C(DL)Q2N81transducer frequency MHz 0.2 0.5 1 2 4inner pipe diameter dmin. extended mm 400 100 50 25 10min. recommended mm 500 200 100 50 25max. recommended mm 4000 2000 1000 400 150max. extended mm 6500 2400 1200 480 240pipe wall thicknessmin. mm 11 5 2 1 0.6materialhousing PEEK with stainless steel cap 304 (1.4301), ***-*****/OS: 316L (1.4404)contact surface PEEKdegree of protection IP65 IP66 IP65transducer cabletype 1699length m 5, ***-*****/LC: 9 4, ***-*****/LC: 9 3, ***-*****/LC: 9dimensionslength l mm 129.5 126.5 64 40width b mm 51 51 32 22height h mm 67 67.5 40.5 25.5dimensional drawing
weight (without cable)
kg 0.47 0.36 0.066 0.016
ambient temperaturemin. °C -40max. °C +130temperature com-pensation
x
explosion protection• ATEX/IECExorder code FSG-NA1TS/** FSK-NA1TS/** FSM-NA1TS/** FSP-NA1TS/** FSQ-NA1TS/**explosion protection temperature (pipe surface)• min. °C -55• max. °C +180marking
Ex q IIC T6...T3 GbEx tb IIIC TX Db
certification ATEX IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X• inmetroorder code FSG-NS1TS/** FSK-NS1TS/** FSM-NS1TS/** FSP-NS1TS/** FSQ-NS1TS/**explosion protection temperature (pipe surface)• min. °C -55• max. °C +180marking Ex e q IIC T6...T3 Gb
Ex tb IIIC TX Dbcertification TÜV 15.0616 X
• KOSHAorder code FSG-NK1TS/** FSK-NK1TS/** FSM-NK1TS/** FSP-NK1TS/** FSQ-NK1TS/**marking Ex eq IIC T3...T6
Tamb = -55 °C - +180 °Ccertification 14-AV4BO-0510 14-AV4BO-0509
l
bh
l
bh
lb
hl
bh
0637 II2GII2D
®TÜV Rheinland
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu14
Shear wave transducers (zone 1, TS, IP68)
order code FSG-N*1TS/IP68 FSK-N*1TS/IP68 FSM-N*1TS/IP68 FSP-N*1TS/IP68technical type CDG1LI1 CDK1LI1 CDM2LI1 CDP2LI1transducer frequency MHz 0.2 0.5 1 2inner pipe diameter dmin. extended mm 400 100 50 25min. recommended mm 500 200 100 50max. recommended mm 4000 2000 1000 400max. extended mm 6500 2400 1200 480pipe wall thicknessmin. mm 11 5 2 1materialhousing PEEK with stainless steel cap 316Ti (1.4571)contact surface PEEK
degree of protection IP68transducer cabletype 2550length m 12dimensionslength l mm 130 72width b mm 54 32height h mm 83.5 46dimensional drawing
weight (without cable)
kg 0.43 0.085
ambient temperaturemin. °C -40max. °C +100temperature com-pensation
x
explosion protection• ATEX/IECExorder code FSG-NA1TS/IP68 FSK-NA1TS/IP68 FSM-NA1TS/IP68 FSP-NA1TS/IP68explosion protection temperature (pipe surface)• min. °C -55• max. °C +180marking
Ex q IIC T6...T3 GbEx tb IIIC TX Db
certification ATEX IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X• inmetroorder code FSG-NS1TS/IP68 FSK-NS1TS/IP68 FSM-NS1TS/IP68 FSP-NS1TS/IP68explosion protection temperature (pipe surface)• min. °C -55• max. °C +180marking Ex q IIC T6...T3 Gb
Ex tb IIIC TX Dbcertification TÜV 15.0616 X
1 test conditions: 3 months/2 bar (20 m)/20 °C
1
h
l
b
l
hb
II2GII2D0637
®TÜV Rheinland
Technical specification FLUXUS WW
15TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
Transducer mounting fixture
tension straps and clasps
material: stainless steel 304 (1.4301)
length: 10 m
Variofix C (VC)
material: stainless steel 304 (1.4301), 301 (1.4310)option OS: 316 (1.4571)
inner length:VCK-*L: 500 mmVCK-*S: 350 mmVCM: 400 mmVCQ: 250 mm
dimensions:VCK-*L: 560 x 122 x 102 mm,option IP68: 560 x 126 x 120 mmVCK-*S: 410 x 122 x 102 mm,option IP68: 410 x 126 x 120 mmVCM: 460 x 96 x 80 mmVCQ: 310 x 85 x 62 mm
Variofix L (VLK, VLM, VLQ)
material: stainless steel 304 (1.4301), 301 (1.4310), 410 (1.4006)option OS: 316 (1.4571), 316L (1.4404), 17-7PH (1.4568)
inner length:VLK: 348 mm,option IP68: 368 mmVLM: 234 mmVLQ: 176 mm
dimensions:VLK: 423 x 90 x 93 mmoption IP68: 443 x 94 x 105 mmVLM: 309 x 57 x 63 mmVLQ: 247 x 43 x 47 mm
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu16
Coupling materials for transducers
Technical data
< 100 °C < 170 °C< 24 h coupling compound type N or
coupling foil type VTcoupling compound type E or coupling foil type VT
long time measurement coupling foil type VT1 coupling foil type VT2
1 < 5 years2 < 6 months
type ambient temperature material°C
coupling compound type N -30...+130 mineral grease paste
coupling compound type E -30...+200 silicone paste
coupling foil type VT -10...+200 fluoroelastomer
Technical specification FLUXUS WW
17TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
Connection systems
Cable
Transducer cable
connection system TSconnection with extension cable direct connection transducers
technical type*****8*
****LI*
*****52
x, y - transducer cable length
l - max. length of extension cable
type 1699 2550weight kg/
m0.094 0.035
ambient temperature °C -55...+200 -40...+100properties longitudinal watertightcable jacketmaterial PTFE PURouter diameter mm 2.9 5.2 ±0.2thickness mm 0.3 0.9colour brown greyshield x xsheathmaterial stainless steel 304 (1.4301) -
option OS: 316Ti (1.4571)outer diameter mm 8 -
tran
smitt
er
JB01 x
l
tra
nsm
itter
x
tran
smitt
er
JB01, JBP2, JBP3 x
l
tra
nsm
itte
r
x
tra
nsm
itte
r
JB02, JB03x
l
tra
nsm
itte
r
x
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu18
Extension cable
Cable length
type 2615 5245weight kg 0.18 0.38ambient temperature °C -30...+70 -30...+70properties halogen free
fire propagation test according to IEC 60332-1
combustion test according to IEC 60754-2
halogen free
fire propagation test according to IEC 60332-1
combustion test according to IEC 60754-2
cable jacketmaterial PUR PURouter diameter mm 12 12thickness mm 2 2colour black blackshield x xsheathmaterial - steel wire braid with copolymer
sheathouter diameter mm - 15.6
transducer frequency
F, G, H, K M, P Q S transducerstechnical type
TSx l x l x l x l
cable length m 5 ≤ 300 4 ≤ 300 3 ≤ 90 2 ≤ 40 *****8******5*
cable length (option LC)
m 9 ≤ 300 9 ≤ 300 9 ≤ 90 - *****8******5*
cable length m 12 ≤ 300 12 ≤ 300 - - - - ****LI*
Technical specification FLUXUS WW
19TSFLUXUS_WWV1-1EN_Leu, 2017-05-12
Junction box
Technical data
Dimensions
2" pipe mounting kit
JB01S4E3M JB02 JB03 JBP2 JBP3weight kg 1.2 kg 1.2 kg 1.2 kg 1.2 kg 1.2 kgfixation wall mounting
optional: 2" pipe mounting
wall mountingoptional: 2" pipe mounting
wall mountingoptional: 2" pipe mounting
wall mountingoptional: 2" pipe mounting
wall mountingoptional: 2" pipe mounting
materialhousing stainless steel 316L
(1.4404)stainless steel 316L (1.4404)
stainless steel 316L (1.4404)
stainless steel 316L (1.4404)
stainless steel 316L (1.4404)
gasket silicone silicone silicone silicone siliconedegree of protection IP67 IP67 IP67 IP67 IP67ambient temperaturemin. °C -40 -40 -40 -40 -40max. °C +80 +80 +80 +80 +80explosion protection• ATEX/IECExzone 1 2 - 2 -marking
Ex e mb IIC (T6)...T4 GbEx tb IIIC T 100 °C DbTa -40...+(70)80 °C
II3G Ex nA IIC (T6)...T4 GcII3D Ex tc IIIC T 100 °C DcTa -40...+(70)80 °C
-
II3G Ex nA IIC (T6)...T4 GcII3D Ex tc IIIC T 100 °C DcTa -40...+(70)80 °C
-
certification ATEX IBExU06ATEX1161 - - - -certification IECEx IECEx IBE 08.0006 - - - -type of protection gas: increased safety
decoupled network: encapsulation
dust: protection by enclosure
gas: non sparkingdust: protection by enclosure
- gas: non sparkingdust: protection by enclosure
-
JBxx
in mm
JBxx
II2GII2D0637
17
4
119
2
wall mounting holder
70
163.5
156
Ø 9 thread: 3x M20 x 1.5cable gland: max. 2x M20
FLUXUS WW Technical specification
2017-05-12, TSFLUXUS_WWV1-1EN_Leu20
Terminal assignment
JB01transducersterminal strip terminal connection transducerKL1 V signal
VS internal shieldRS internal shieldR signal
cable gland external shield
extension cableterminal strip terminal connectionKL2 TV signal
TVS internal shieldTRS internal shieldTR signal
shield terminal external shield
JB02, JB03transducers
terminal connection transducerXV SMB connector
XR SMB connector
cable gland external shield
extension cableterminal strip terminal connectionKL2 TV signal
TVS internal shieldTRS internal shieldTR signal
shield terminal external shield
JBP2, JBP3transducersterminal strip terminal connection transducerKL1 TV signal
TVS internal shieldTRS internal shieldTR signal
cable gland external shield
extension cableterminal strip terminal connectionKL2 TV signal
TVS internal shieldTRS internal shieldTR signal
shield terminal external shield
shield terminal
equipotential bonding terminal(at wall mounting holder)
shield terminal
equipotential bonding terminal(at wall mounting holder)
shield terminal
equipotential bonding terminal(at wall mounting holder)