Technical Data Sheet Thermo Scientific Revco Ultra-Low ....pdf · Industrial-Rated Two Stage...
Transcript of Technical Data Sheet Thermo Scientific Revco Ultra-Low ....pdf · Industrial-Rated Two Stage...
Thermo Fisher Scientific TDS_PEEK.xlsm 7/11/2011
Thermo Fisher Scientific, Asheville, North Carolina
Model Number
Application Storage of General (non-flammable) Laboratory Materials Storage Volume 548 liters / 19.4 cu. ft., 400 Standard 2" Boxes
Temperature Rating -50°C to -85°C @ 32 °C(90°F) AmbientElectrical Power
Instrument Rated Current 9.0 FLABuilding Supply Rating
Power Plug/Power Cord LengthAgency Listings
Sound Pressure Level (dBA) <= 66 dBA Indoor/Outdoor Usage Indoor Use Only
Application Environment Non-Corrosive, Non-Flammable, Non-Explosive, Good Air Ventilation, 15° C - 32° C (59° F - 90° F)Lifetime of Product
Refrigeration ConfigurationRefrigeration System Industrial-Rated Two Stage Cascade SystemCompressor / Number 1 HP Hermetic Compressor for Low Temperature Application / 2
Condenser Type/Number Enhanced Finned-Tube and Forced-Air Cooled / 1Expansion Device Capillary Tube On Both Cascade StagesEvaporator Type Cold Wall With Enhanced Heat Transfer TreatmentDefrost Method Manual Defrost
Refrigerant Charge/Flammability CFC/HCFC-Free Enviromentally Safe Refrigerant Mixtures / Non-Flammable in both stagesController/Electrical System Configuration and Features
Controller LevelPower Switch
Controller Type Microprocessor Control with Touch Screen Input and Display. Includes USB System Data RetreivalSetpoint Security
Compressor Safe GuardControl Sensor Single RTD (1000 ohm Platinum RTD)
RS232/Remote Alarm TerminalsAdjustable Warm/Cold Alarms
Auto-Voltage Safeguard Buck/Boost System Dimensions and Construction
Interior Dimensions (H x D x W) 1.30 x 0.72 x 0.59 m (51.2 x 28.3 x 23.1 in.)Exterior Dimensions (H x D x W) 1.98 x 0.96 x 0.82 m (78.0 x 37.6 x 32.4 in.)
InsulationPerimeter Heater Silicone-Based High Performance Seal Gasket with Electrical Door Perimeter Heater
Shelves / Capacity 3 or 4 Stainless Steel Shelves Adjustable In 1" Increments. Max. Cap. per Shelf: 73.4 kg (165 lbs.)All-Direction Casters Standard with Locks
Ship Weight Approximately 332 kg (730 lbs.)Other Options
Typical Performance Characteristics in 25 ° C AmbientC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\400_PD_WU.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\400_Cycle.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\400_perf.jpg
1) Performance is nominal and individual units may vary. 2) Freezer performance will differ due to product amount, product size and operating conditions.3) Continuous product enhancements may, without notice, result in amendments or ommisions to this specification. Thermo Scientific cannot accept responsibility for damage, injury, loss or expenses resulting from misapplication of the information herein.
© 2011 Thermo Scientific Inc. All trademarks are the property of Thermo Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject
to change. Not all products are available in all countries. Please consult your local sales representative for details.
LN2 or CO2 Back Up System, HID Controlled Access, SMS Text, Chart Recorder, 4 or 5 Inner Doors
Fully Adjustable
YesHigh Temperature Warning/Current and Temperature Protection/Logic Protection
RS485/4-20mA output
High R-value Vacuum Insulation Panels and High Density Water-Blown Polyurethane Foam
SpecificationsThermo Scientific Revco UxF40086D
208V/230V, 60 Hz, 1 Phase
On-Off with Circuit Breaker
Technical Data SheetThermo Scientific Revco Ultra-Low Temperature Upright Freezer
Revision-0
10 years
Eye Level
Application, Rating and Electrical Data
UL, cULNEMA 5-20P or IEC Cord, 10 Feet or 3.0 Meters
20.0A dedicated grounded circuit. Protected by circuit breaker rated for inductive loads
57.61463.2
18.2463.4
23.0584.1
19.9504.5
3.177.5
1.024.5
2.152.6
78.01981.3
17.2
SHELF WIDTH436.3
51.21301.1
FRONT
USER INTERFACE
35.9911.2
36.2919.2
1.436.1
10.3262.1
10.3262.1
9.9251.2
10.3262.1
2.974.8
27.0
SHELF DEPTH686.6
10.4263.8
2.666.3
4X ADJUSTABLESHELVES(3X SHELVES OPTIONAL)
49.21248.8
11.5292.1
25.1638.0
FLOOR REFERENCE
SERIALCOMMUNICATION
POWER PLUG
BACK
REMOTE ALARMCONTACT
POWER SWITCH
1.0[25.4]ACCESS PORTS
.38.0
27.0686.1
37.6955.4
31.7806.3
51.1
DOOR OPEN @ 1801299.0
2.666.3
2.666.3
58.6
DOOR OPEN @901489.2
17.8
TANK WIDTH451.4
28.3718.8
SYMBOLS, DESIGNATIONS AND1.GENERAL DRAWING METHODSENGWI002ALL COMPONENTS SHOULD BE2.RoHS COMPLIANTBASIC DIMENSION TOLERANCE3.UNLESS OTHERWISE SPECIFIED:.X = 0.100"DUAL DIMENSION IS INCH OVER4.METRIC.
- 60347 2011-06-30 NEW RELEASEREV. ECN # DATE DESCRIPTION
REVISIONS
12 11 10
J
I
H
G
A
B
300
THE INFORMATION CONTAINED HEREIN IS THE PROPRIETARY DATA OF THERMO FISHER SCIENTIFIC. IT IS NOT TO BE
REPRODUCED OR DISCLOSED IN WHOLE OR IN PART WITHOUT PRIOR WRITTEN
CONSENT OF THERMO FISHER SCIENTIFIC.DO NOT SCALE THIS DRAWING.
275 AIKEN ROADASHEVILLE, NC 28804
TITLE FREEZER UNIT
SIZE SHEET 1 OF 1
DWG. NO.3rd ANGLEPROJECTION
D12 SolidWorks 2007-10-25
C
D
E
F
3456789
K
L
M
13141516171819
M
L
K
F
E
D
C
B
A
G
H
I
J
19 18 17 16 15 14 13 9 8 7 6 5 4 3 2 1101112
ULT
315100G01
Thermo Fisher Scientific TDS_PEEK.xlsm 7/11/2011
Thermo Fisher Scientific, Asheville, North Carolina
Model Number
Application Storage of General (non-flammable) Laboratory Materials Storage Volume 548 liters / 19.4 cu. ft., 400 Standard 2" Boxes
Temperature Rating -50°C to -85°C @ 32 °C(90°F) AmbientElectrical Power
Instrument Rated Current 9.0 FLABuilding Supply Rating
Power Plug/Power Cord LengthAgency Listings
Sound Pressure Level (dBA) <= 66 dBA Indoor/Outdoor Usage Indoor Use Only
Application Environment Non-Corrosive, Non-Flammable, Non-Explosive, Good Air Ventilation, 15° C - 32° C (59° F - 90° F)Lifetime of Product
Refrigeration ConfigurationRefrigeration System Industrial-Rated Two Stage Cascade SystemCompressor / Number 1 HP Hermetic Compressor for Low Temperature Application / 2
Condenser Type/Number Enhanced Finned-Tube and Forced-Air Cooled / 1Expansion Device Capillary Tube On Both Cascade StagesEvaporator Type Cold Wall With Enhanced Heat Transfer TreatmentDefrost Method Manual Defrost
Refrigerant Charge/Flammability CFC/HCFC-Free Enviromentally Safe Refrigerant Mixtures / Non-Flammable in both stagesController/Electrical System Configuration and Features
Controller LevelPower Switch
Controller Type Microprocessor Control with Touch Screen Input and Display. Includes USB System Data RetreivalSetpoint Security
Compressor Safe GuardControl Sensor Single RTD (1000 ohm Platinum RTD)
RS232/Remote Alarm TerminalsAdjustable Warm/Cold Alarms
Auto-Voltage Safeguard Buck/Boost System Dimensions and Construction
Interior Dimensions (H x D x W) 1.30 x 0.72 x 0.59 m (51.2 x 28.3 x 23.1 in.)Exterior Dimensions (H x D x W) 1.98 x 0.96 x 0.82 m (78.0 x 37.6 x 32.4 in.)
InsulationPerimeter Heater Silicone-Based High Performance Seal Gasket with Electrical Door Perimeter Heater
Shelves / Capacity 3 or 4 Stainless Steel Shelves Adjustable In 1" Increments. Max. Cap. per Shelf: 73.4 kg (165 lbs.)All-Direction Casters Standard with Locks
Ship Weight Approximately 332 kg (730 lbs.)Other Options
Typical Performance Characteristics in 25 ° C AmbientC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\400_PD_WU.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\400_Cycle.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\400_perf.jpg
1) Performance is nominal and individual units may vary. 2) Freezer performance will differ due to product amount, product size and operating conditions.3) Continuous product enhancements may, without notice, result in amendments or ommisions to this specification. Thermo Scientific cannot accept responsibility for damage, injury, loss or expenses resulting from misapplication of the information herein.
© 2011 Thermo Scientific Inc. All trademarks are the property of Thermo Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject
to change. Not all products are available in all countries. Please consult your local sales representative for details.
LN2 or CO2 Back Up System, HID Controlled Access, SMS Text, Chart Recorder, 4 or 5 Inner Doors
Fully Adjustable
YesHigh Temperature Warning/Current and Temperature Protection/Logic Protection
RS485/4-20mA output
High R-value Vacuum Insulation Panels and High Density Water-Blown Polyurethane Foam
SpecificationsThermo Scientific Revco UxF40086D
208V/230V, 60 Hz, 1 Phase
On-Off with Circuit Breaker
Technical Data SheetThermo Scientific Revco Ultra-Low Temperature Upright Freezer
Revision-0
10 years
Eye Level
Application, Rating and Electrical Data
UL, cULNEMA 5-20P or IEC Cord, 10 Feet or 3.0 Meters
20.0A dedicated grounded circuit. Protected by circuit breaker rated for inductive loads
19.9504.8
18.2463.4
57.61463.2
3.077.2
28.3720.0
1.024.5
78.01981.3
2.152.6
22.5
SHELF WIDTH572.1
51.21301.1
FRONT
FOR GRILLE SEEDETAIL VIEW
USER INTERFACE
WHITE BOARD
1.436.1
35.9911.036.2919.3
10.3262.1
10.3262.1
9.9251.2
10.3262.1
2.974.8
27.0
SHELF DEPTH686.6
2.666.3
10.4263.8 4X ADJUSTABLE
SHELVES(3X SHELVES OPTIONAL)
49.21250.8
11.5292.1
25.0636.0
FLOOR REFERENCE
SERIALCOMMUNICATION
POWER PLUGBACK
REMOTE ALARM CONTACT
POWER SWITCH
1.0[25.4]ACCESS PORTS
.38.2
37.6955.4
32.4821.9
31.7806.3
61.8
DOOR OPEN @1801570.6
2.666.3
64.0
DOOR OPEN @901626.2
28.3718.8
2.666.3
23.1
TANK WIDTH587.4
OPTIONAL CHART RECORDER
SYMBOLS, DESIGNATIONS AND1.GENERAL DRAWING METHODSENGWI002ALL COMPONENTS SHOULD BE2.RoHS COMPLIANTBASIC DIMENSION TOLERANCE3.UNLESS OTHERWISE SPECIFIED:.X = 0.100"DUAL DIMENSION IS INCH OVER4.METRIC.
DETAIL VIEW OF GRILLE
- 60347 2011-06-30 NEW RELEASEREV. ECN # DATE DESCRIPTION
REVISIONS
12 11 10
J
I
H
G
A
B
400
THE INFORMATION CONTAINED HEREIN IS THE PROPRIETARY DATA OF THERMO FISHER SCIENTIFIC. IT IS NOT TO BE
REPRODUCED OR DISCLOSED IN WHOLE OR IN PART WITHOUT PRIOR WRITTEN
CONSENT OF THERMO FISHER SCIENTIFIC.DO NOT SCALE THIS DRAWING.
275 AIKEN ROADASHEVILLE, NC 28804
TITLE FREEZER UNIT
SIZE SHEET 1 OF 1
DWG. NO.3rd ANGLEPROJECTION
D12 SolidWorks 2007-10-25
C
D
E
F
3456789
K
L
M
13141516171819
M
L
K
F
E
D
C
B
A
G
H
I
J
19 18 17 16 15 14 13 9 8 7 6 5 4 3 2 1101112
ULT
315100G02
Thermo Fisher Scientific TDS_PEEK.xlsm 7/11/2011
Thermo Fisher Scientific, Asheville, North Carolina
Model Number
Application Storage of General (non-flammable) Laboratory Materials Storage Volume 682 liters / 24.1 cu. ft., 500 Standard 2" Boxes
Temperature Rating -50°C to -85°C @ 32 °C(90°F) AmbientElectrical Power
Instrument Rated Current 9.0 FLABuilding Supply Rating
Power Plug/Power Cord LengthAgency Listings
Sound Pressure Level (dBA) <= 66 dBA Indoor/Outdoor Usage Indoor Use Only
Application Environment Non-Corrosive, Non-Flammable, Non-Explosive, Good Air Ventilation, 15° C - 32° C (59° F - 90° F)Lifetime of Product
Refrigeration ConfigurationRefrigeration System Industrial-Rated Two Stage Cascade SystemCompressor / Number 1 HP Hermetic Compressor for Low Temperature Application / 2
Condenser Type/Number Enhanced Finned-Tube and Forced-Air Cooled / 1Expansion Device Capillary Tube On Both Cascade StagesEvaporator Type Cold Wall With Enhanced Heat Transfer TreatmentDefrost Method Manual Defrost
Refrigerant Charge/Flammability CFC/HCFC-Free Enviromentally Safe Refrigerant Mixtures / Non-Flammable in both stagesController/Electrical System Configuration and Features
Controller LevelPower Switch
Controller Type Microprocessor Control with Touch Screen Input and Display. Includes USB System Data RetreivalSetpoint Security
Compressor Safe GuardControl Sensor Single RTD (1000 ohm Platinum RTD)
RS232/Remote Alarm TerminalsAdjustable Warm/Cold Alarms
Auto-Voltage Safeguard Buck/Boost System Dimensions and Construction
Interior Dimensions (H x D x W) 1.30 x 0.72 x 0.73 m (51.2 x 28.3 x 28.8 in.)Exterior Dimensions (H x D x W) 1.98 x 0.96 x 0.97 m (78.0 x 37.6 x 38.0 in.)
InsulationPerimeter Heater Silicone-Based High Performance Seal Gasket with Electrical Door Perimeter Heater
Shelves / Capacity 3 or 4 Stainless Steel Shelves Adjustable In 1" Increments. Max. Cap. per Shelf: 91.7 kg (205 lbs.)All-Direction Casters Standard with Locks
Ship Weight Approximately 355 kg (782 lbs.)Other Options
Typical Performance Characteristics in 25 ° C AmbientC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\500_PD_WU.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\500_Cycle.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\500_perf.jpg
1) Performance is nominal and individual units may vary. 2) Freezer performance will differ due to product amount, product size and operating conditions.3) Continuous product enhancements may, without notice, result in amendments or ommisions to this specification. Thermo Scientific cannot accept responsibility for damage, injury, loss or expenses resulting from misapplication of the information herein.
© 2011 Thermo Scientific Inc. All trademarks are the property of Thermo Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject
to change. Not all products are available in all countries. Please consult your local sales representative for details.
LN2 or CO2 Back Up System, HID Controlled Access, SMS Text, Chart Recorder, 4 or 5 Inner Doors
Fully Adjustable
YesHigh Temperature Warning/Current and Temperature Protection/Logic Protection
RS485/4-20mA output
High R-value Vacuum Insulation Panels and High Density Water-Blown Polyurethane Foam
SpecificationsThermo Scientific Revco UxF50086D
208V/230V, 60 Hz, 1 Phase
On-Off with Circuit Breaker
Technical Data SheetThermo Scientific Revco Ultra-Low Temperature Upright Freezer
Revision-0
10 years
Eye Level
Application, Rating and Electrical Data
UL, cULNEMA 5-20P or IEC Cord, 10 Feet or 3.0 Meters
20.0A dedicated grounded circuit. Protected by circuit breaker rated for inductive loads
1.436.1
35.9911.2
36.2919.3
2.666.3
2.974.8
27.0
SHELF DEPTH686.6
9.9251.2
10.3262.1
10.3262.1
10.3262.1
10.4263.8
4X ADJUSTABLESHELVES
(3X SHELVES OPTIONAL)
49.21250.8
25.0636.0
11.5292.1
1.0[25.4]ACCESS PORTS
REMOTE ALARMCONTACTS
SERIALCOMMUNICATION
POWER SWITCH
POWER PLUG
2.152.6
18.2463.4
57.61463.0
34.0862.8
3.077.2
19.9504.6
1.024.5
28.2
SHELF WIDTH715.0
51.21301.2
78.01981.3
FRONT
USER INTERFACE
WHITE BOARD
FOR GRILLESEE DETAIL VIEW
.38.1
38.0964.8
37.6955.4
31.8808.1
73.1
DOOR OPEN @1801856.4
2.666.3
2.666.3
28.8
TANK WIDTH730.9
69.6
DOOR OPEN @901769.1
28.3718.8
OPTIONAL CHART RECORDER
BACK
FLOOR REFERENCE
GRILLE DETAIL VIEW
1.SYMBOLS, DESIGNATIONS AND GENERAL DRAWING METHODS ENGWI0022. ALL COMPONENTS SHOULD BE RoHS COMPLIANT3. BASIC DIMENSION TOLERANCE UNLESS OTHERWISE SPECIFIED: .X = 0.100" 4. DUAL DIMENSION IS INCH OVER METRIC.
- 60347 2011-06-30 NEW RELEASEREV. ECN # DATE DESCRIPTION
REVISIONS
12 11 10
J
I
H
G
A
B
500
THE INFORMATION CONTAINED HEREIN IS THE PROPRIETARY DATA OF THERMO FISHER SCIENTIFIC. IT IS NOT TO BE
REPRODUCED OR DISCLOSED IN WHOLE OR IN PART WITHOUT PRIOR WRITTEN
CONSENT OF THERMO FISHER SCIENTIFIC.DO NOT SCALE THIS DRAWING.
275 AIKEN ROADASHEVILLE, NC 28804
TITLE FREEZER UNIT
SIZE SHEET 1 OF 1
DWG. NO.3rd ANGLEPROJECTION
D12 SolidWorks 2007-10-25
C
D
E
F
3456789
K
L
M
13141516171819
M
L
K
F
E
D
C
B
A
G
H
I
J
19 18 17 16 15 14 13 9 8 7 6 5 4 3 2 1101112
ULT
315100G03
Thermo Fisher Scientific TDS_PEEK.xlsm 7/11/2011
Thermo Fisher Scientific, Asheville, North Carolina
Model Number
Application Storage of General (non-flammable) Laboratory Materials Storage Volume 682 liters / 24.1 cu. ft., 500 Standard 2" Boxes
Temperature Rating -50°C to -85°C @ 32 °C(90°F) AmbientElectrical Power
Instrument Rated Current 9.0 FLABuilding Supply Rating
Power Plug/Power Cord LengthAgency Listings
Sound Pressure Level (dBA) <= 66 dBA Indoor/Outdoor Usage Indoor Use Only
Application Environment Non-Corrosive, Non-Flammable, Non-Explosive, Good Air Ventilation, 15° C - 32° C (59° F - 90° F)Lifetime of Product
Refrigeration ConfigurationRefrigeration System Industrial-Rated Two Stage Cascade SystemCompressor / Number 1 HP Hermetic Compressor for Low Temperature Application / 2
Condenser Type/Number Enhanced Finned-Tube and Forced-Air Cooled / 1Expansion Device Capillary Tube On Both Cascade StagesEvaporator Type Cold Wall With Enhanced Heat Transfer TreatmentDefrost Method Manual Defrost
Refrigerant Charge/Flammability CFC/HCFC-Free Enviromentally Safe Refrigerant Mixtures / Non-Flammable in both stagesController/Electrical System Configuration and Features
Controller LevelPower Switch
Controller Type Microprocessor Control with Touch Screen Input and Display. Includes USB System Data RetreivalSetpoint Security
Compressor Safe GuardControl Sensor Single RTD (1000 ohm Platinum RTD)
RS232/Remote Alarm TerminalsAdjustable Warm/Cold Alarms
Auto-Voltage Safeguard Buck/Boost System Dimensions and Construction
Interior Dimensions (H x D x W) 1.30 x 0.72 x 0.73 m (51.2 x 28.3 x 28.8 in.)Exterior Dimensions (H x D x W) 1.98 x 0.96 x 0.97 m (78.0 x 37.6 x 38.0 in.)
InsulationPerimeter Heater Silicone-Based High Performance Seal Gasket with Electrical Door Perimeter Heater
Shelves / Capacity 3 or 4 Stainless Steel Shelves Adjustable In 1" Increments. Max. Cap. per Shelf: 91.7 kg (205 lbs.)All-Direction Casters Standard with Locks
Ship Weight Approximately 355 kg (782 lbs.)Other Options
Typical Performance Characteristics in 25 ° C AmbientC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\500_PD_WU.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\500_Cycle.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\500_perf.jpg
1) Performance is nominal and individual units may vary. 2) Freezer performance will differ due to product amount, product size and operating conditions.3) Continuous product enhancements may, without notice, result in amendments or ommisions to this specification. Thermo Scientific cannot accept responsibility for damage, injury, loss or expenses resulting from misapplication of the information herein.
© 2011 Thermo Scientific Inc. All trademarks are the property of Thermo Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject
to change. Not all products are available in all countries. Please consult your local sales representative for details.
LN2 or CO2 Back Up System, HID Controlled Access, SMS Text, Chart Recorder, 4 or 5 Inner Doors
Fully Adjustable
YesHigh Temperature Warning/Current and Temperature Protection/Logic Protection
RS485/4-20mA output
High R-value Vacuum Insulation Panels and High Density Water-Blown Polyurethane Foam
SpecificationsThermo Scientific Revco UxF50086D
208V/230V, 60 Hz, 1 Phase
On-Off with Circuit Breaker
Technical Data SheetThermo Scientific Revco Ultra-Low Temperature Upright Freezer
Revision-0
10 years
Eye Level
Application, Rating and Electrical Data
UL, cULNEMA 5-20P or IEC Cord, 10 Feet or 3.0 Meters
20.0A dedicated grounded circuit. Protected by circuit breaker rated for inductive loads
1.436.1
35.9911.2
36.2919.3
2.666.3
2.974.8
27.0
SHELF DEPTH686.6
9.9251.2
10.3262.1
10.3262.1
10.3262.1
10.4263.8
4X ADJUSTABLESHELVES
(3X SHELVES OPTIONAL)
49.21250.8
25.0636.0
11.5292.1
1.0[25.4]ACCESS PORTS
REMOTE ALARMCONTACTS
SERIALCOMMUNICATION
POWER SWITCH
POWER PLUG
2.152.6
18.2463.4
57.61463.0
34.0862.8
3.077.2
19.9504.6
1.024.5
28.2
SHELF WIDTH715.0
51.21301.2
78.01981.3
FRONT
USER INTERFACE
WHITE BOARD
FOR GRILLESEE DETAIL VIEW
.38.1
38.0964.8
37.6955.4
31.8808.1
73.1
DOOR OPEN @1801856.4
2.666.3
2.666.3
28.8
TANK WIDTH730.9
69.6
DOOR OPEN @901769.1
28.3718.8
OPTIONAL CHART RECORDER
BACK
FLOOR REFERENCE
GRILLE DETAIL VIEW
1.SYMBOLS, DESIGNATIONS AND GENERAL DRAWING METHODS ENGWI0022. ALL COMPONENTS SHOULD BE RoHS COMPLIANT3. BASIC DIMENSION TOLERANCE UNLESS OTHERWISE SPECIFIED: .X = 0.100" 4. DUAL DIMENSION IS INCH OVER METRIC.
- 60347 2011-06-30 NEW RELEASEREV. ECN # DATE DESCRIPTION
REVISIONS
12 11 10
J
I
H
G
A
B
500
THE INFORMATION CONTAINED HEREIN IS THE PROPRIETARY DATA OF THERMO FISHER SCIENTIFIC. IT IS NOT TO BE
REPRODUCED OR DISCLOSED IN WHOLE OR IN PART WITHOUT PRIOR WRITTEN
CONSENT OF THERMO FISHER SCIENTIFIC.DO NOT SCALE THIS DRAWING.
275 AIKEN ROADASHEVILLE, NC 28804
TITLE FREEZER UNIT
SIZE SHEET 1 OF 1
DWG. NO.3rd ANGLEPROJECTION
D12 SolidWorks 2007-10-25
C
D
E
F
3456789
K
L
M
13141516171819
M
L
K
F
E
D
C
B
A
G
H
I
J
19 18 17 16 15 14 13 9 8 7 6 5 4 3 2 1101112
ULT
315100G03
Thermo Fisher Scientific TDS_PEEK.xlsm 7/11/2011
Thermo Fisher Scientific, Asheville, North Carolina
Model Number
Application Storage of General (non-flammable) Laboratory Materials Storage Volume 815 liters / 28.8 cu. ft., 600 Standard 2" Boxes
Temperature Rating -50°C to -85°C @ 32 °C(90°F) AmbientElectrical Power
Instrument Rated Current 9.0 FLABuilding Supply Rating
Power Plug/Power Cord LengthAgency Listings
Sound Pressure Level (dBA) <= 66 dBA Indoor/Outdoor Usage Indoor Use Only
Application Environment Non-Corrosive, Non-Flammable, Non-Explosive, Good Air Ventilation, 15° C - 32° C (59° F - 90° F)Lifetime of Product
Refrigeration ConfigurationRefrigeration System Industrial-Rated Two Stage Cascade SystemCompressor / Number 1 HP Hermetic Compressor for Low Temperature Application / 2
Condenser Type/Number Enhanced Finned-Tube and Forced-Air Cooled / 1Expansion Device Capillary Tube On Both Cascade StagesEvaporator Type Cold Wall With Enhanced Heat Transfer TreatmentDefrost Method Manual Defrost
Refrigerant Charge/Flammability CFC/HCFC-Free Enviromentally Safe Refrigerant Mixtures / Non-Flammable in both stagesController/Electrical System Configuration and Features
Controller LevelPower Switch
Controller Type Microprocessor Control with Touch Screen Input and Display. Includes USB System Data RetreivalSetpoint Security
Compressor Safe GuardControl Sensor Single RTD (1000 ohm Platinum RTD)
RS232/Remote Alarm TerminalsAdjustable Warm/Cold Alarms
Auto-Voltage Safeguard Buck/Boost System Dimensions and Construction
Interior Dimensions (H x D x W) 1.30 x 0.72 x 0.87 m (51.2 x 28.3 x 34.4 in.)Exterior Dimensions (H x D x W) 1.98 x 0.96 x 1.11 m (78.0 x 37.6 x 43.6 in.)
InsulationPerimeter Heater Silicone-Based High Performance Seal Gasket with Electrical Door Perimeter Heater
Shelves / Capacity 3 or 4 Stainless Steel Shelves Adjustable In 1" Increments. Max. Cap. per Shelf: 110 kg (245 lbs.)All-Direction Casters Standard with Locks
Ship Weight Approximately 388 kg (854 lbs.)Other Options
Typical Performance Characteristics in 25 ° C AmbientC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\600_PD_WU.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\600_Cycle.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\600_perf.jpg
1) Performance is nominal and individual units may vary. 2) Freezer performance will differ due to product amount, product size and operating conditions.3) Continuous product enhancements may, without notice, result in amendments or ommisions to this specification. Thermo Scientific cannot accept responsibility for damage, injury, loss or expenses resulting from misapplication of the information herein.
© 2011 Thermo Scientific Inc. All trademarks are the property of Thermo Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject
to change. Not all products are available in all countries. Please consult your local sales representative for details.
SpecificationsThermo Scientific Revco UxF60086D
208V/230V, 60 Hz, 1 Phase
On-Off with Circuit Breaker
Technical Data SheetThermo Scientific Revco Ultra-Low Temperature Upright Freezer
Revision-0
10 years
Eye Level
Application, Rating and Electrical Data
UL, cULNEMA 5-20P or IEC Cord, 10 Feet or 3.0 Meters
20.0A dedicated grounded circuit. Protected by circuit breaker rated for inductive loads
LN2 or CO2 Back Up System, HID Controlled Access, SMS Text, Chart Recorder, 4 or 5 Inner Doors
Fully Adjustable
YesHigh Temperature Warning/Current and Temperature Protection/Logic Protection
RS485/4-20mA output
High R-value Vacuum Insulation Panels and High Density Water-Blown Polyurethane Foam
18.2463.4
57.61463.2
19.9504.6
3.077.2
39.61005.7
1.024.5
2.152.6
78.01981.3
33.8
SHELF WIDTH857.9
51.21301.2
FLOOR REFERENCE
USER INTERFACE
WHITE BOARD
FRONT
FOR GRILL SEEDETAIL VIEW
.38.0
43.61107.7
37.6955.4
1.436.1
36.2919.3
35.9911.2
2.974.8
27.0
SHELF DEPTH686.6 2.9
73.9
9.9251.2
10.3262.1
10.3262.1
10.3262.1
10.4263.8
4X ADJUSTABLESHELVES(3X SHELVESOPTIONAL)25.0
636.0
49.21250.8
2X 11.5292.9
ACCESS PORTS1.0[25.4]
POWER PLUG
POWER SWITCH
SERIALCOMMUNICATION
REMOTE ALARMCONTACTS
BACK
31.8808.1
84.3
DOOR OPEN @1802142.1
75.2
DOOR OPEN @901910.8
2.666.3
2.666.3
34.4
TANK WIDTH874.5
28.3718.8
OPTIONAL CHART RECORDER
GRILLE DETAIL VIEW
SYMBOLS, DESIGNATIONS 1.AND GENERAL DRAWING METHODS ENGWI002ALL COMPONENTS SHOULD 2.BE RoHS COMPLIANTDUAL DIMENSION IS INCH 3.OVER METRICBASIC DIMENSION 4.TOLERANCE UNLESS OTHERWISE SPECIFIED:.X = .100"
- 60347 2011-06-30 NEW RELEASEREV ECN# DATE DESCRIPTION
REVISIONS
12 11 10
J
I
H
G
A
B
600
THE INFORMATION CONTAINED HEREIN IS THE PROPRIETARY DATA OF THERMO FISHER SCIENTIFIC. IT IS NOT TO BE
REPRODUCED OR DISCLOSED IN WHOLE OR IN PART WITHOUT PRIOR WRITTEN
CONSENT OF THERMO FISHER SCIENTIFIC.DO NOT SCALE THIS DRAWING.
275 AIKEN ROADASHEVILLE, NC 28804
TITLE FREEZER UNIT
SIZE SHEET 1 OF 1
DWG. NO.3rd ANGLEPROJECTION
D12 SolidWorks 2007-10-25
C
D
E
F
3456789
K
L
M
13141516171819
M
L
K
F
E
D
C
B
A
G
H
I
J
19 18 17 16 15 14 13 9 8 7 6 5 4 3 2 1101112
ULT
315100G04
Thermo Fisher Scientific TDS_PEEK.xlsm 7/11/2011
Thermo Fisher Scientific, Asheville, North Carolina
Model Number
Application Storage of General (non-flammable) Laboratory Materials Storage Volume 949 liters / 33.5 cu. ft., 700 Standard 2" Boxes
Temperature Rating -50°C to -85°C @ 32 °C(90°F) AmbientElectrical Power
Instrument Rated Current 9.0 FLABuilding Supply Rating
Power Plug/Power Cord LengthAgency Listings
Sound Pressure Level (dBA) <= 66 dBA Indoor/Outdoor Usage Indoor Use Only
Application Environment Non-Corrosive, Non-Flammable, Non-Explosive, Good Air Ventilation, 15° C - 32° C (59° F - 90° F)Lifetime of Product
Refrigeration ConfigurationRefrigeration System Industrial-Rated Two Stage Cascade SystemCompressor / Number 1 HP Hermetic Compressor for Low Temperature Application / 2
Condenser Type/Number Enhanced Finned-Tube and Forced-Air Cooled / 1Expansion Device Capillary Tube On Both Cascade StagesEvaporator Type Cold Wall With Enhanced Heat Transfer TreatmentDefrost Method Manual Defrost
Refrigerant Charge/Flammability CFC/HCFC-Free Enviromentally Safe Refrigerant Mixtures / Non-Flammable in both stagesController/Electrical System Configuration and Features
Controller LevelPower Switch
Controller Type Microprocessor Control with Touch Screen Input and Display. Includes USB System Data RetreivalSetpoint Security
Compressor Safe GuardControl Sensor Single RTD (1000 ohm Platinum RTD)
RS232/Remote Alarm TerminalsAdjustable Warm/Cold Alarms
Auto-Voltage Safeguard Buck/Boost System Dimensions and Construction
Interior Dimensions (H x D x W) 1.30 x 0.72 x 1.02 m (51.2 x 28.3 x 40.0 in.)Exterior Dimensions (H x D x W) 1.98 x 0.96 x 1.25 m (78.0 x 37.6 x 49.2 in.)
InsulationPerimeter Heater Silicone-Based High Performance Seal Gasket with Electrical Door Perimeter Heater
Shelves / Capacity 3 or 4 Stainless Steel Shelves Adjustable In 1" Increments. Max. Cap. per Shelf: 128 kg (285 lbs.)All-Direction Casters Standard with Locks
Ship Weight Approximately 432 kg (951 lbs.)Other Options
Typical Performance Characteristics in 25 ° C AmbientC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\700_PD_WU.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\700_Cycle.jpgC:\Documents and Settings\karen.wisniewski\My Documents\ULT\Ranier\Technical Data Sheets\TDS\Pictures\700_perf.jpg
1) Performance is nominal and individual units may vary. 2) Freezer performance will differ due to product amount, product size and operating conditions.3) Continuous product enhancements may, without notice, result in amendments or ommisions to this specification. Thermo Scientific cannot accept responsibility for damage, injury, loss or expenses resulting from misapplication of the information herein.
© 2011 Thermo Scientific Inc. All trademarks are the property of Thermo Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject
to change. Not all products are available in all countries. Please consult your local sales representative for details.
SpecificationsThermo Scientific Revco UxF70086D
208V/230V, 60 Hz, 1 Phase
On-Off with Circuit Breaker
Technical Data SheetThermo Scientific Revco Ultra-Low Temperature Upright Freezer
Revision-0
10 years
Eye Level
Application, Rating and Electrical Data
ULNEMA 5-20P or IEC Cord, 10 Feet or 3.0 Meters
20.0A dedicated grounded circuit. Protected by circuit breaker rated for inductive loads
LN2 or CO2 Back Up System, HID Controlled Access, SMS Text, Chart Recorder, 4 or 5 Inner Doors
Fully Adjustable
YesHigh Temperature Warning/Current and Temperature Protection/Logic Protection
RS485/4-20mA output
High R-value Vacuum Insulation Panels and High Density Water-Blown Polyurethane Foam
1.436.1
35.9911.2
36.2919.2
2.974.8
27.0
SHELF DEPTH686.6 2.6
66.3
9.9251.2
10.3262.1
10.3262.1
10.3262.1
10.4263.8
4X ADJUSTABLESHELVES
(3 SHELVES OPTIONAL)25.0
636.0
49.21250.8
11.5292.9
SERIAL COMMUNICATION
REMOTE ALARMCONTACTS
POWER PLUG
1.0[25.4]ACCESS PORTS
POWERSWITCH
45.21148.6
1.024.5
3.077.2
19.9505.0
18.2463.4
2.152.6
57.61463.2
2.666.3
51.21301.1
39.4
SHELF WIDTH1000.8
78.01981.3
FOR GRILLE SEE DETAIL VIEW
USER INTERFACE
WHITE BOARD
31.8808.1
95.6
DOOR OPEN @ 1802427.9
2.665.6
2.665.6
40.0
TANK WIDTH1016.0
80.9
DOOR OPEN @ 902054.8
28.3718.8
.38.0
49.21250.6
37.6955.4
OPTIONAL CHARTRECORDER
BACK
1.SYMBOLS, DESIGNATIONS AND GENERAL DRAWING METHODS ENGWI0022. ALL COMPONENTS SHOULD BE RoHS COMPLIANT3. BASIC DIMENSION TOLERANCE UNLESS OTHERWISE SPECIFIED: .X= 0.100" 4. DUAL DIMENSION IS INCH OVER METRIC.
GRILLE DETAIL VIEW
FLOOR REFERENCE
- 60347 2011-06-30 NEW RELEASEREV. ECN # DATE DESCRIPTION
REVISIONS
12 11 10
J
I
H
G
A
B
700
THE INFORMATION CONTAINED HEREIN IS THE PROPRIETARY DATA OF THERMO FISHER SCIENTIFIC. IT IS NOT TO BE
REPRODUCED OR DISCLOSED IN WHOLE OR IN PART WITHOUT PRIOR WRITTEN
CONSENT OF THERMO FISHER SCIENTIFIC.DO NOT SCALE THIS DRAWING.
275 AIKEN ROADASHEVILLE, NC 28804
TITLE FREEZER UNIT
SIZE SHEET 1 OF 1
DWG. NO.3rd ANGLEPROJECTION
D12 SolidWorks 2007-10-25
C
D
E
F
3456789
K
L
M
13141516171819
M
L
K
F
E
D
C
B
A
G
H
I
J
19 18 17 16 15 14 13 9 8 7 6 5 4 3 2 1101112
ULT
315100G05