Tech Guide Pt Scint

download Tech Guide Pt Scint

of 44

Transcript of Tech Guide Pt Scint

  • 7/28/2019 Tech Guide Pt Scint

    1/44

    Parathyroid Scintigraphy

    A Technologists Guide

    European Association of Nuclear Medici

    ne

  • 7/28/2019 Tech Guide Pt Scint

    2/44

    Contributors

    Nish Fernando

    Chie Technologist

    Department o Nuclear Medicine

    St. Bartholomews Hospital, London, UK

    Dr. Eli Hindi, MD, PhD

    Service de Mdecine NuclaireHpital Saint-Antoine, Paris, France

    Sue Huggett

    Senior University Teacher

    Department o Radiography

    City University, London, United Kingdom

    Jos Pires Jorge

    Proesseur HES-S2

    Ecole Cantonale Vaudoise de Techniciens enRadiologie Mdicale (TRM)

    Lausanne, Switzerland

    Regis Lecoultre

    Proesseur HES-S2

    Ecole Cantonale Vaudoise de Techniciens en

    Radiologie Mdicale (TRM)

    Lausanne, Switzerland

    Sylviane Prvot

    Chair, EANM Technologist Committee

    Chie Technologist

    Service de Mdecine Nuclaire

    Centre Georges-Franois Leclerc, Dijon, France

    Domenico Rubello*, MDDirector

    Nuclear Medicine Service - PET Unit

    S. Maria della Misericordia Hospital

    Istituto Oncologico Veneto (IOV), Rovigo, Italy

    Audrey Taylor

    Chie Technologist

    Department o Nuclear Medicine

    Guys and St. Thomas Hospital, London, UK

    Linda Tutty

    Senior Radiographer

    St. Jamess Hospital

    Dublin, Ireland

    * Domenico Rubello, MD

    is coordinator of the National Study Group on parathyroid scintigraphy of AIMN (Italian Association of Nuclear

    Medicine) and is responsible for developing study programmes on minimally invasive radioguided surgery in patients

    with hyperparathyroidism for GISCRIS (Italian Study Group on Radioguided Surgery and Immunoscintigraphy)

    Acknowledgement for the photo:

    P. Lind, Department of Nuclear Medicine, LKH Klagenfurt, Austria

  • 7/28/2019 Tech Guide Pt Scint

    3/44

    EA

    NM

    Contents

    Foreword 4

    Sylviane Prvot

    Introduction 5

    Sue Huggett

    Chapter 1 Applications o parathyroid imaging 612

    Eli Hindi

    Chapter 2 Radiopharmaceuticals 1317

    Linda Tutty

    Chapter 3 Imaging equipment preparation and use 1823

    Jos Pires Jorge and Regis Lecoultre

    Chapter 4 Patient preparation 2428

    Audrey Taylor and Nish Fernando

    Chapter 5 Imaging protocols 2932

    Nish Fernando and Sue Huggett

    Chapter 6 Technical aspects o probe-guided surgery or parathyroid adenomas 3339

    Domenico Rubello

    Reerences 4043

    This booklet was sponsored by an educational grant from Bristol-Myers Squibb Medical Imaging.

    The views expressed are those of the authors and not necessarily of Bristol-Myers Squibb Medical

    Imaging.

  • 7/28/2019 Tech Guide Pt Scint

    4/44

    ForewordSylviane Prvot

    Today the notion o competence is at the

    heart o proessional development. Technol-

    ogists specic proessional skills o working

    eciently and knowledgeably are essential

    to ensure high-quality practice in nuclear

    medicine departments.

    Since they were ormed, the EANM Tech-

    nologist Committee and Sub-committee onEducation have devoted themselves to the

    improvement o nuclear medicine technolo-

    gists (NMTs) proessional skills.

    Publications that will assist in the setting o

    high standards or NMTs work throughout

    Europe have been developed. A series o bro-

    chures, technologists guides, was planned in

    early 2004. The rst o these was dedicated tomyocardial perusion imaging and the current

    volume, the second in the planned series, ad-

    dresses parathyroid imaging.

    Renowned authors with expertise in the eld

    have been selected to provide an inormative

    and truly comprehensive tool or technolo-

    gists that will serve as a reerence and improve

    the quality o daily practice.

    I am grateul or the hard work o all the con-

    tributors, who have played a key role in ensur-

    ing the high scientic content and educational

    value o this booklet. Many thanks are due to

    Sue Huggett, who coordinated the project,

    to the members o the EANM Technologist

    Sub-committee on Education and particularly

    to Bristol-Myers Squibb Imaging or their con-

    dence and generous sponsorship.

    Eorts to image the parathyroid gland date

    back many years. I hope this brochure will be

    useul to technologists in the management

    o patients with hyperparathyroidism and will

    benet these patients by optimising care and

    welare.

    Sylviane Prvot

    Chair, EANM Technologist Committee

  • 7/28/2019 Tech Guide Pt Scint

    5/44

    EA

    NM

    IntroductionSue Huggett

    The rst publication o the EANM Technolo-

    gist Committee sponsored by Bristol Myers

    Squibb in 2004 was a book on myocardial

    perusion imaging or technologists. We

    are very grateul that they have sponsored

    us again this year to produce this book on

    parathyroid imaging, the second book in

    what we hope will be a series.

    We hope that we have combined the theory

    and rationale o imaging with the practicalities

    o patient care and equipment use. I think that

    certain things I wrote or the last book bear re-

    peating, and so I will do so here or the benet

    o those or whom this is their rst book.

    Knowledge o imaging theory provides a deeper

    understanding o the techniques that is satisy-ing or the technologist and can orm the basis

    or wise decision making. It also allows the tech-

    nologist to communicate accurate inormation

    to patients, their carers and other sta. Patient

    care is always paramount, and being able to

    explain why certain oods must be avoided or

    why it is necessary to lie in awkward positions

    improves compliance as well as satisaction.

    Awareness o the rationales or using certain

    strategies is needed in order to know when

    and how various protocol variations should be

    applied, in acquisition or analysis, e.g. or the

    patient who cannot lie fat or long enough

    or subtraction and may need to be imaged

    with another protocol or when we may need

    a dierent lter i the total counts are low.

    Protocols will vary between departments,

    even within the broader terms o the EANM

    Guidelines. This booklet is not meant to sup-

    plant these protocols but will hopeully sup-

    plement and explain the rationales behind

    them, thereby leading to more thoughtul

    working practices.

    The authors are indebted to a number osources or inormation, not least local proto-

    cols, and reerences have been given where

    original authors are identiable. We apolo-

    gise i we have inadvertently used material

    or which credit should have been, but was

    not, given.

    We hope that this booklet will provide help-

    ul inormation as and when it is needed sothat the integration o theory and practice is

    enabled and encouraged.

  • 7/28/2019 Tech Guide Pt Scint

    6/44

    Applications o parathyroid imagingEli Hindi

    Primary hyperparathyroidism

    Primary hyperparathyroidism (pHPT) is a

    surgically correctable disease with the third

    highest incidence o all endocrine disorders

    ater diabetes mellitus and hyperthyroidism

    (Al Zahrani and Levine 1997). Through their

    secretion o parathyroid hormone (PTH), the

    two pairs o parathyroid glands, located in the

    neck posterior to the thyroid gland, regulateserum calcium concentration and bone me-

    tabolism. PTH promotes the release o calcium

    rom bone, increases absorption o calcium

    rom the intestine and increases reabsorp-

    tion o calcium in the renal tubules. In turn,

    the serum calcium concentration regulates

    PTH secretion, a mechanism mediated via a

    calcium-sensing receptor on the surace o

    the parathyroid cells. pHPT is caused by thesecretion o excessive amounts o PTH by

    one or more enlarged diseased parathyroid

    gland(s). Patients with pHPT may suer rom

    renal stones, osteoporosis, gastro-intestinal

    symptoms, cardiovascular disease, muscle

    weakness and atigue, and neuropsychologi-

    cal disorders. The highest prevalence o the

    disease is ound in post-menopausal women.

    A prevalence o 2% was ound by screening

    post-menopausal women (Lundgren).

    In the past, pHPT was characterised by severe

    skeletal and renal complications and apparent

    mortality. This may still be the case in some

    developing countries. The introduction o

    calcium auto-analysers in the early 1970s

    led to changes in the incidence o pHPT and

    deeply modied the clinical spectrum o the

    disease at diagnosis (Heath et al. 1980). Most

    new cases are now biologically mild without

    overt symptoms (Al Zahrani and Levine 1997).

    Parathyroidectomy is the only curative treat-

    ment or pHPT. In the recent guidelines o the

    US National Institute o Health (NIH), surgery

    is recommended or all young individuals and

    or all patients with overt symptoms (Bilezikianet al. 2002). For patients who are asymptom-

    atic and are 50 years old or older, surgery is

    recommended i any o the ollowing signs

    are present: serum calcium greater than 10

    mg/l above the upper limits o normal; 24-h

    total urine calcium excretion o more than 400

    mg; reduction in creatinine clearance by more

    than 30% compared with age-matched per-

    sons; bone density more than 2.5 SDs belowpeak bone mass: T score < -2.5. Surgery is also

    recommended when medical surveillance

    is either not desirable or not possible. Ater

    complete baseline evaluation, patients who

    are not operated on need to be monitored

    twice yearly or serum calcium concentration

    and yearly or creatinine concentration; it is

    also recommended that bone mass measure-

    ments are obtained on a yearly basis (Bilezikian

    et al. 2002). Some authors recommend para-

    thyroidectomy or all patients with a secure

    diagnosis o pHPT (Utiger 1999).

    Successul parathyroidectomy depends on

    recognition and excision o all hyperunc-

    tioning parathyroid glands. pHPT is typically

    caused by a solitary parathyroid adenoma, less

  • 7/28/2019 Tech Guide Pt Scint

    7/44

    EA

    NM

    Chapter 1: Applications of parathyroid imaging

    requently (about 15% o cases) by multiple

    parathyroid gland disease (MGD) and rarely

    (about 1% o cases) by parathyroid carcino-

    ma. Patients with MGD have either double

    adenomas or hyperplasia o three or all our

    parathyroid glands. Most cases o MGD are

    sporadic, while a small number are associated

    with hereditary disorders such as multiple en-

    docrine neoplasia type 1 or type 2a or amilialhyperparathyroidism (Marx et al. 2002). Con-

    ventional surgery consists in routine bilateral

    exploration with identication o all our para-

    thyroid glands.

    Imaging is mandatory beore reoperation

    For several decades, preoperative imaging was

    not used beore rst-time surgery. Unguided

    bilateral exploration, dissecting all potentialsites in the neck, achieved cure in 9095% o

    patients (Russell and Edis 1982). The two main

    reasons or ailed surgery are ectopic glands

    (retro-oesophageal, mediastinal, intrathyroid,

    in the sheath o the carotid artery, or unde-

    scended) and undetected MGD (Levin and

    Clark 1989). Repeat surgery is associated with a

    dramatic reduction in the success rate and an

    increase in surgical complications. Imaging is

    thereore mandatory beore reoperation (Sosa

    et al. 1998). 99mTc-sestamibi scanning (Coakley

    et al. 1989) has been established as the imag-

    ing method o choice in reoperation o persis-

    tent or recurrent hyperparathyroidism (Weber

    et al. 1993). In these patients it is necessary to

    have all inormation concerning the rst inter-

    vention, including the number and location o

    parathyroid glands that have been seen by the

    surgeon and the size and histology o resected

    glands. Whichever 99mTc-sestamibi scanning

    protocol is used, it is necessary to provide the

    surgeon with the best anatomical inormation

    by using both anterior and lateral (or oblique)

    views o the neck, and SPECT whenever use-

    ul, especially or a mediastinal ocus. It is the

    authors opinion that99m

    Tc-sestamibi resultsshould be conrmed with a second imaging

    technique (usually ultrasound or a neck ocus

    and CT or MRI or a mediastinal image) beore

    proceeding to reoperation.

    Scanning with 99mTc-sestamibi is increasingly

    ordered on a routine basis or rst-time

    parathyroidectomy

    The rst exploration is the best time to curehyperparathyroidism. Most surgeons would

    now appreciate having inormation concern-

    ing whereabouts in the neck to start dissec-

    tion and the possibility o ectopic parathyroid

    glands (Sosa et al. 1998; Liu et al. 2005). When

    the rare cases (2-5%) o ectopic parathyroid

    tumours are recognised preoperatively, the

    success o bilateral surgery can now reach

    very close to 100% (Hindi et al. 1997). In the

    case o a mediastinal gland, the surgeon can

    proceed directly with rst-intention thoracos-

    copy, avoiding unnecessary initial extensive

    neck surgery in the search or the elusive

    gland (Liu et al. 2005). Preoperative imaging

    would also shorten the duration o bilateral

    surgery (Hindi et al. 1997). By allowing the

    surgeon to nd the oending gland earlier in

  • 7/28/2019 Tech Guide Pt Scint

    8/44

    the operation, the time necessary or rozen

    section examination can be used by the sur-

    geon or inspection o the other parathyroid

    glands, also reducing surgeon anxiety.

    Important points to know when proceeding

    with parathyroid imaging

    Imaging is not or diagnosis. The increase

    in plasma levels o calcium (normal value88105 mg/l) and PTH (normal value 1058

    ng/l) establishes the diagnosis.

    Imaging does not identiy normal parathy-

    roid glands, which are too small (2050 mg)

    to be seen.

    Imaging should detect abnormal para-

    thyroid(s) and indicate the approximate sizeand the precise relationship to the thyroid

    (the level o the thyroid at which the para-

    thyroid lesion is seen on the anterior view;

    and whether it is proximal to the thyroid or

    deeper in the neck on the lateral or oblique

    view or SPECT) (Fig. 1).

    Imaging should identiy ectopic glands (add

    SPECT in cases o a mediastinal ocus, and

    ask or additional CT or MRI or conrmation

    and anatomical landmarks) (Fig. 2).

    Imaging should be able to dierentiate pa-

    tients with a single adenoma rom those

    with MGD (Fig. 3).

    Imaging should identiy thyroid nodules

    that may require concurrent surgical re-

    section.

    The choice o imaging technique

    The most common preoperative localisation

    methods are radionuclide scintigraphy and

    ultrasound. As stated beore, the two main

    reasons or ailed surgery are ectopic glands

    and undetected MGD (Levin and Clark 1989).Because high-resolution ultrasound would,

    even in skilled hands, ail to detect the major-

    ity o these cases, it is not optimal or preop-

    erative imaging as a single technique. In the

    study by Haber et al. (2002), ultrasound missed

    six o eight ectopic glands and ve o six cases

    o MGD. Ultrasound may, however, be useul

    in combination with 99mTc-sestamibi imaging

    (Rubello et al. 2003).

    99mTc-sestamibi scanning is now considered

    the most sensitive imaging technique in pa-

    tients with pHPT (Giordano et al. 2001; Mullan

    2004). Whatever the protocol used, 99mTc-ses-

    tamibi scanning will usually meet the require-

    ment o detecting ectopic glands (all eight

    were detected in the study by Haber et al.).

    With regard to the recognition o MGD, how-

    ever, the protocol in use will determine the

    sensitivity. When 99mTc-sestamibi is used as a

    single tracer with planar imaging at two time

    points -- the dual-phase (or washout) method

    the sensitivity or primary hyperplasia is very

    low (Tailleer et al. 1992; Martin et al. 1996). Bet-

    ter results can be obtained by adding SPECT.

    Subtraction scanning, using either

    123

    I (Borley

  • 7/28/2019 Tech Guide Pt Scint

    9/44

    EA

    NM

    Chapter 1: Applications of parathyroid imaging

    et al. 1996; Hindi et al. 2000; Mullan 200)

    or 99mTc-pertechnetate (Rubello et al. 2003)

    in addition to 99mTc-sestamibi, improves the

    sensitivity or hyperplastic glands. One di-

    culty with subtraction imaging is keeping the

    patient still or the time necessary to scan the

    thyroid, to inject 99mTc-sestamibi and to record

    images o this second tracer. Simultaneous

    recording o123

    I and99m

    Tc-sestamibi can be asimple answer to these diculties. It prevents

    arteacts on subtraction images due to pa-

    tient motion, and shortens the imaging time

    (Hindi et al. 1998; Mullan 2004).

    Preoperative imaging has opened a new era

    o minimally invasive parathyroid surgery

    Conventional bilateral exploration is still

    considered the gold standard in parathyroidsurgery. However, the introduction o99mTc-

    sestamibi scanning, the availability o intra-

    operative adjuncts such as the gamma probe

    and intraoperative monitoring o PTH to help

    detect MGD have challenged the dogma o

    routine bilateral exploration. When preopera-

    tive imaging points to a single well-dened

    ocus, unequivocally suggesting a solitary

    adenoma, the surgeon may now choose o-

    cussed surgery instead o bilateral exploration.

    Focussed excision can be made by open sur-

    gery through a mini-incision, possibly under

    local anaesthesia, or by video-assisted endo-

    scopic surgery under general anaesthesia (Lee

    and Inabnet 2005). Compared with patients

    who undergo bilateral surgery, those in whom

    ocussed parathyroid surgery is successully

    completed enjoy a shorter operation time,

    the possibility o local anaesthesia, a better

    cosmetic scar, a less painul postoperative

    course, less proound postoperative transient

    hypocalcaemia and an earlier return to normal

    activities. The act that many clinicians now

    use a lower threshold or surgery is partly due

    to the perception that parathyroid surgery is

    easier than in the past (Utiger 1999).

    Patients at specic risk o ailure o minimal

    surgery are those with unrecognised MGD.

    Thereore, when choosing minimal surgery,

    the surgeon is committed to distinguishing

    cases o MGD either preoperatively, through

    an appropriate imaging protocol, or by intra-

    operative monitoring o PTH plasma levels, or

    by a combination o both. The true sensitivityo intraoperative PTH or MGD is still under

    debate. What raises concern is that studies

    relying solely on intraoperative measurements

    report a low percentage o MGD, only 3% (Mo-

    linari et al. 1996), which is three to our times

    lower than is generally observed during rou-

    tine bilateral surgery. Whether this will lead to

    higher rates o late recurrence is not known. It

    is thus important that imaging methods used

    to select patients or ocussed surgery have a

    high sensitivity or detecting MGD.

    In this new era o ocussed operations, the

    success o parathyroid surgery depends not

    only on an experienced surgeon but also on

    excellent interpretation o images. A localisa-

    tion study with high accuracy is mandatory to

  • 7/28/2019 Tech Guide Pt Scint

    10/44

    10

    avoid conversion o the surgery to a bilateral

    exploration under general anaesthesia ater

    minimal surgery has been started. It is impor-

    tant to avoid conusion with a thyroid nod-

    ule, and precise anatomical description is also

    important. With enlargement and increased

    density, superior parathyroid adenomas can

    become pendulous and descend posteriorly.

    A lateral view (or an oblique view or SPECT)should indicate whether the adenoma is close

    to the thyroid or deeper in the neck (tracheo-

    oesophageal groove or retro-oesophageal).

    This inormation is useul, because visualisa-

    tion through the small incision is restricted.

    Moreover, the surgeon may choose a lateral

    approach to excise this gland instead o an

    anterior approach. To achieve a high sensitiv-

    ity in detecting MGD with subtraction tech-niques, the degree o subtraction should be

    monitored careully. Progressive incremental

    subtraction with real-time display is a good

    way to choose the optimal level o subtraction

    (residual 99mTc-sestamibi activity in the thyroid

    area should not be lower than in surround-

    ing neck tissues). Oversubtraction could easily

    delete additional oci o activity and in some

    patients provide a alse image suggestive o

    a single adenoma.

    Secondary hyperparathyroidism

    Secondary hyperparathyroidism is a common

    complication in patients with chronic renal

    ailure. Hypocalcaemia, accumulation o phos-

    phate and a decrease in the active orm o

    vitamin D lead to increased secretion o PTH.

    With chronic stimulation, hyperplasia o para-

    thyroid glands accelerates and may develop

    into autonomous adenomas. The extent o

    parathyroid growth then becomes a major

    determinant o PTH hypersecretion. Second-

    ary hyperparathyroidism leads to renal bone

    disease, the development o sot tissue calci-

    cations, vascular calcications and increased

    cardiovascular risk, among other complica-tions. When medical therapy ails, surgery

    becomes necessary. Surgery can be either

    subtotal parathyroidectomy, with resection

    o three glands and partial resection o the

    ourth gland, or total resection with grating

    o some parathyroid tissue into the sot tissues

    o the orearm in order to avoid permanent

    hypoparathyroidism.

    Preoperative imaging

    Surgery o secondary hyperparathyroidism

    requires routine bilateral identication o all

    parathyroid tissue. Moreover, early studies

    based on single-tracer 99mTc-sestamibi scan-

    ning have reported a very low sensitivity

    o about 050% in detecting hyperplastic

    glands. Ineciency o single-tracer techniques

    both in secondary hyperparathyroidism and

    in primary hyperplasia is possibly due to more

    rapid washout o tracer rom hyperplastic

    glands than rom parathyroid adenomas. For

    those reasons, preoperative imaging has not

    yet gained wide acceptance among surgeons.

    Dual-tracer subtraction imaging, planar or

    SPECT, provides substantial improvement in

    the rate o detection o hyperplastic glands in

  • 7/28/2019 Tech Guide Pt Scint

    11/44

    EA

    NM

    11

    Chapter 1: Applications of parathyroid imaging

    patients with renal ailure (Hindi et al. 1999;

    Peri et al. 2005)

    What inormation can be obtained?

    The preoperative map may acilitate rec-

    ognition o the position o aberrant para-

    thyroid glands, also reducing the extent o

    dissection (Hindi et al. 1999).

    Parathyroid glands with major ectopia

    would be missed without preoperative

    imaging.

    Although the usual number o parathyroid

    glands is our, some individuals (about 10%)

    have a supernumerary th gland (Aker-

    strm et al. 1984). When this inormation is

    provided by preoperative imaging, it mayprevent surgical ailure or late recurrence

    (Hindi et al. 1999).

    Imaging ndings in patients with persistent

    or recurrent secondary hyperparathyroidism

    Immediate ailure and delayed recurrence are

    not unusual, occurring in 1030% o patients.

    Imaging is mandatory beore reoperation.

    Knowledge o all details concerning the initial

    intervention is necessary or interpretation. As

    with primary hyperparathyroidism, we recom-

    mend that lesions seen on the 99mTc-sestamibi

    scan be matched with a second radiological

    technique (ultrasound or MRI) or conrmation

    and identication o anatomical landmarks

    beore reoperation.

    Some aspects specic to patients reoperated

    or secondary hyperparathyroidism need to

    be emphasised:

    Specic views o the orearm should be

    obtained in patients who have had a para-

    thyroid grat.

    It is not unusual or imaging in these pa-tients to show two oci o activity, one

    corresponding to recurrent disease at the

    subtotally resected gland (or grated tissue)

    and the other corresponding to an ectopic

    or th parathyroid, missed at initial inter-

    vention (unpublished data).

    Figure 1

    Parathyroid subtraction scintigraphy, with simul-

    taneous acquisition o99mTc-sestamibi and 123I

    in a patient with primary hyperparathyroidism.

    The anterior view and the lateral view show a

    solitary adenoma located at the lower right pole

    o the thyroid. At surgery, an adenoma o 1.9 g

    was ound at the predicted site.

  • 7/28/2019 Tech Guide Pt Scint

    12/44

    1

    Figure 2This patient with a previous history o thyroid surgery (right lobectomy) was reerred with a recent

    diagnosis o primary hyperparathyroidism. Ultrasound examination suggested the presence o

    a parathyroid adenoma at the side o previous thyroid lobectomy, which was a alse-positive

    image. The large eld o view 99mTc-sestamibi acquisition shows a mediastinal ocus (arrow).

    The suspected ectopic parathyroid was conrmed by MRI (arrow). A mediastinal parathyroid

    adenoma o 0.59 g was resected.

    Figure 3

    Parathyroid 99mTc-sestamibi/123I subtraction scintigraphy in a patient with primary hyperpara-

    thyroidism. The computed subtraction images show two sites o preerential 99mTc-sestamibi

    uptake: one at the lower third o the let thyroid lobe and the second lateral to the lower pole

    o the right thyroid lobe. Two adenomas were excised: a let parathyroid adenoma weighing

    2.3 g and a right adenoma weighing 0.07 g.

  • 7/28/2019 Tech Guide Pt Scint

    13/44

    EA

    NM

    1

    RadiopharmaceuticalsLinda Tutty

    Radiopharmaceuticals used or

    parathyroid scintigraphy

    Details o the photo peak energy, hal-lie, e-

    ective dose and standard dose or radiophar-

    maceuticals commonly used or parathyroid

    scintigraphy are shown in Table 1.

    201Tl-chloride201

    Tl-chloride has a physical hal-lie o 73.1 h.Its main photo peak is due to characteristic x-

    rays o mercury, which have an energy range

    o 6983 keV. In addition, gamma rays are pro-

    duced at 167 keV (8%) and 135 keV (2%). The

    administered activity is 80 MBq and it is given

    intravenously. 201Tl-chloride is taken up by ab-

    normal parathyroid tissue and thyroid tissue in

    proportion to blood fow.

    99mTc-pertechnetate99mTc-pertechnetate has a hal-lie o 6 h and a

    gamma energy o 140 keV. 99mTc-pertechnetate is

    used to delineate the thyroid gland because unc-

    tioning thyroid parenchyma traps it. This image is

    then subtracted rom the 201Tl or 99mTc-sestamibi

    images, and what remains is potentially a parathy-

    roid adenoma. When utilising 201Tl, the adminis-

    tered activity is usually 75150 MBq, depending on

    the administered radioactivity o201Tl and which o

    the two radiopharmaceuticals is administered rst.

    I using 99mTc-sestamibi, the amount o pertechne-

    tate administered is usually 185370 MBq, because99mTc-sestamibi has a higher total activity in the

    thyroid tissue than 201Tl.

    99m

    Tc-sestamibiThe range o intravenously administered radio-

    activity is 185-900 MBq; the typical dose is 740

    MBq. This radiotracer localises in both parathyroid

    gland and unctioning thyroid tissue, and usually

    washes out o normal thyroid tissue more rap-

    idly than out o abnormal parathyroid tissue. The

    exact mechanism o uptake remains unknown

    (Farley 2004). 99mTc-sestamibi uptake depends on

    numerous actors, including perusion, cell cycle

    phase and unctional activity (Beggs and Hain2005). The nal cellular localisation o99mTc-sesta-

    mibi is within the mitochondria. It accumulates in

    the mitochondria o many tissues but particularly

    in normal cardiac and thyroid cells; it is especially

    prominent in overactive parathyroid glands and

    is held there preerentially (Farley 2004).

    123I-sodium iodide

    123I has a hal-lie o 13 h and emits a photonwith an energy o 159 keV. It has been used

    particularly with 99mTc-sestamibi as a thyroid-

    imaging agent in subtraction studies. The

    administered dose, given orally, ranges rom

    7.5 to 20 MBq.

    99mTc-tetroosmin99mTc-tetroosmin use in parathyroid imaging

    is described in the literature (Smith and Oates

    2004). Its manuacturers do not license it or

    use as a parathyroid scintigraphy agent. 99mTc-

    sestamibi and 99mTc-tetroosmin have similar

    imaging characteristics (Smith and Oates

    2004). The typical dose o administered activ-

    ity is 740 MBq.

    11

    C-methionine11C-methionine has a hal-lie o 20 min. It is

  • 7/28/2019 Tech Guide Pt Scint

    14/44

    1

    cyclotron produced. Its uptake reects amino

    acid refux into stimulated parathyroid tissue

    (Otto et al. 2004; Beggs and Hain 2005). Up-

    take in infammatory conditions may pose

    a problem and should be considered when

    interpreting images. The typical radioactivity

    dose ranges between 240 and 820 MBq, with

    an average intravenous dose o 400 MBq.

    18F-FDG18F-FDG has a hal-lie o 110 min and is cy-

    clotron produced. 18F-FDG allows glucose

    metabolism to be assessed and evaluated

    using PET. There is dierential concentra-

    tion o FDG in abnormal parathyroid tissue

    and this dierence is used to demonstrate

    the abnormal gland. FDG also accumulates

    in other malignant and benign tissues, andin infamed or inected tissue; this potentially

    limits its useulness. The typical intravenous

    dose is 400 MBq.

    Radiopharmaceutical eatures

    Multiple radiopharmaceuticals have been

    described or the detection o parathyroid le-

    sions. Thallium, sestamibi and tetroosmin are

    the three most commonly used (Ahuja et al.

    2004). All these agents were originally devel-

    oped or cardiac scanning. In the 1980s, 201Tl

    was the most commonly used agent, but it

    has a longer hal-lie and delivers a higher ra-

    diation dose to the patient (Kettle 2002). Con-

    sequently, 201Tl is no longer commonly used,

    and most recent literature reers to the use o99m

    Tc-sestamibi. However, or the subtraction

    method it is probable that each radiopharma-

    ceutical would provide the same diagnostic

    inormation (Kettle 2002).

    Subtraction agents

    Thyroid-specic imaging with 123I or 99mTc-

    pertechnetate may be employed using a sub-

    traction method to dierentiate parathyroid

    rom thyroid activity (Clark 2005).

    The two main agents used or imaging the

    thyroid are 123I (sodium iodide) and 99mTc-

    pertechnetate. There is a slight preerence or

    the use o123I, as it is organied and thereore

    provides a stable image. The pertechnetate

    washes out rom the thyroid gland with time,

    and i there is some delay in imaging there

    may be a reduction in the quality o thethyroid image (Kettle 2002). However, both

    agents may be aected i the patient is taking

    thyroxine or anti-thyroid medications or has

    recently received iodine contrast agents.

    Thyroid-specic radiopharmaceuticals may

    aid delineation o the thyroid parenchyma i

    required ater dual-phase imaging. This may

    be helpul as a second-line visual subtraction

    procedure when no parathyroid adenoma is

    visible on dual-phase parathyroid imaging

    (Clark 2005).

    Activities given or imaging the thyroid

    and parathyroid glands are as ollows: 99mTc

    pertechnetate, 80 MBq; 123I, 0 MBq; 201Tl,

    80 MBq;

    99m

    Tc-sestamibi, 900 MBq. I a

    99m

    Tc-

  • 7/28/2019 Tech Guide Pt Scint

    15/44

    EA

    NM

    1

    Chapter 2: Radiopharmaceuticals

    pertechnetate/99mTc-sestamibi combination

    is used then the radiation dose or the com-

    bined study is 11.6 mSv. I123I and 201Tl are used,

    this rises to 18.3 mSv.

    Dual-phase agents99mTc-sestamibi and 99mTc-tetroosmin are com-

    monly used agents or dual-phase parathyroid

    scintigraphy. The washout technique relies onthe act that while 99mTc-sestamibi and 99mTc-tet-

    roosmin are taken up by both the thyroid gland

    and the parathyroid at a similar rate, there is a

    aster rate o washout rom the thyroid gland.

    These tracers localise in the thyroid gland as

    well as in parathyroid adenomas. This makes

    correlation o the adenoma in relation to the

    thyroid gland possible on planar as well as earlySPECT imaging. 99mTc-sestamibi is released rom

    the thyroid with a hal-lie o about 30 min but

    is usually retained by abnormal parathyroid

    glands (Smith and Oates 2004). 99mTc-tetroos-

    min may clear more slowly rom the thyroid

    gland. This dierential washout improves the

    target-to-background ratio so that abnormal

    parathyroid tissue should be more visible on

    delayed images (Smith and Oates 2004; Clark

    2005). However, thyroid adenomas and carcino-

    mas can coexist and may retain 99mTc-sestamibi

    or 99mTc-tetroosmin, resulting in alse positive

    results (Smith and Oates 2004).

    99mTc-sestamibi and 99mTc-tetroosmin have

    comparable imaging characteristics. Usually,

    the choice o imaging agent depends on its

    availability and the experience o the nuclear

    medicine radiologist.

    The dual-phase subtraction method with ad-

    junctive thyroid-selective imaging (99mTc or 123I)

    may be helpul, or even essential, in patients

    with goitres or other conounding underly-

    ing thyroid disease, ater thyroid surgery or in

    those patients with a palpable mass (Smithand Oates 2004).

    PET imaging agents

    Use o18F-fuorodeoxyglucose (FDG) positron

    emission tomography (PET) and 11C- methio-

    nine PET or parathyroid imaging has been de-

    scribed (Otto et al. 2004; Beggs and Hain 2005).

    Initial studies with PET have shown conficting

    results when using FDG as a tracer to image theparathyroid glands (Beggs and Hain 2005; Otto

    et al. 2004). It has been shown that 11C-methio-

    nine PET holds more promise than FDG PET im-

    aging o the parathyroid localisation (Beggs and

    Hain 2005). 11C-methionine PET scanning is o

    value in cases o primary hyperparathyroidism

    in which conventional imaging techniques have

    ailed to localise the adenoma beore proceed-

    ing to surgery, or in patients in whom surgery

    has been perormed but has ailed to correct the

    hyperparathyroidism (Beggs and Hain 2005).

    Adverse reactions to

    radiopharmaceuticals

    Table 2 shows side-eects and reactions to

    radiopharmaceuticals used or parathyroid

    scintigraphy.

  • 7/28/2019 Tech Guide Pt Scint

    16/44

    1

    Table 1

    Radiopharmaceuticals used or parathyroid scintigraphy

    201Tl- 99mTc 99mTc 11C 18F

    chloride sestamibi tetroosmin methionine fuorodeoxy-

    glucose

    Photo peak 69-80 (98% 140 140 511 511

    energy abundance)

    (keV) 135 (2%)

    167 (8%)Hal-lie 73.1 hours 6 hours 6 hours 20 min 110 min

    Cyclotron Always Always Cyclotron Cyclotron

    product available available produced produced

    to be (24-month (6-month

    ordered shel lie at shel lie at

    ready or room 2-8C

    use temperature)

    Eectivedose adult

    (mSv)

    18 11 9 2 10

    Standard 80 900 900 400 400

    dose*(MBq)

    *Allowable upper limits o radiotracers may dier rom country to country. Please reer to the

    Summary o Product Characteristics in each European country. Doses given here are quoted

    rom ARSC, December 1998.

  • 7/28/2019 Tech Guide Pt Scint

    17/44

    EA

    NM

    1

    Chapter 2: Radiopharmaceuticals

    Table 2

    Adverse reactions to radiopharmaceuticals used or parathyroid scintigraphy (as printed in J

    Nucl Med 1996;37:185192, 10641067)

    Radiopharmaceutical Side-eects, reactions201TI-chloride Fever, erythema, fushing, diuse rash,

    pruritis, hypotension99mTc-pertechnetate Chills, nausea, vomiting, diuse rash,

    pruritis, hives/urticaria, chest pain,

    tightness or heaviness, hypertension,

    dizziness, vertigo, headache, diaphoresis,

    anaphylaxis

    99mTc-sestamibi Nausea, erythema, fushing, diuse rash,

    pruritis, seizures, headache, metallic taste,

    tingling

    123I-Sodium iodide Nausea, vomiting, diuse rash,

    pruritus, hives/urticaria, chest pain, tightness

    or heaviness, respiratory reaction,tachycardia, syncope or aintness and

    headache, tachypnea, parosmia

    99mTc-tetroosmin Angina, hypertension, torsades de

    pointes (these three probably occurred

    because o underlying heart disease);

    vomiting, abdominal discomort,

    cutaneous allergy, hypotension,

    dyspnoea, metallic taste, burning o

    mouth, unusual odour, mild leucocytosis

    18F-FDG None

  • 7/28/2019 Tech Guide Pt Scint

    18/44

    1

    Imaging equipment preparation and useJos Pires Jorge and Regis Lecoultre

    Quality control procedures that must be

    satisactorily perormed beore imaging

    Ater acceptance testing, a QC protocol must

    be set up in each department and ollowed

    in accordance with national guidelines. The

    ollowing routine quality control test schedule

    is typical:

    a) Daily energy peaking

    b) Daily food uniormity tests

    c) Daily gamma camera sensitivity measure-

    ment

    d) Weekly linearity and resolution assess-

    ment

    e) Weekly centre-o-rotation calibration

    A routine quality control programme or a

    SPECT gamma camera includes quality control

    procedures appropriate to planar scintillation

    cameras (ad) and specic SPECT quality con-

    trols (e). Further, more complex tests should be

    undertaken on a less requent basis.

    Energy peaking

    This quality control procedure consists in

    peaking the gamma camera or relevant

    energies prior to obtaining food images. It

    is mandatory that the energy peaking is un-

    dertaken on a daily basis and or each radio-

    nuclide used.

    Checking the peaking is needed to ascertain

    that:

    The camera automatic peaking circuitry is

    working properly

    The shape o the spectrum is correct

    The energy peak appears at the correctenergy

    There is no accidental contamination o the

    gamma camera

    It is recommended that the spectra obtained

    during peaking tests are recorded.

    Daily food uniormity testsAter a successul peaking test it is recom-

    mended that a uniormity test is perormed

    on a daily basis. Flood elds are acquired and

    evaluation o camera uniormity can be made

    on a visual assessment. Quantitative param-

    eters should also be computed regularly and

    recorded in order both to demonstrate sud-

    den variations rom normal and to alert the

    technologist to progressive deterioration in

    the equipment. On cameras that have inter-

    changeable uniormity correction maps, it

    is vital that one is used that is for the correct

    nuclide, accurate and up to date.

    Daily gamma camera sensitivity

    measurement

    A practical means o measuring sensitivity is

  • 7/28/2019 Tech Guide Pt Scint

    19/44

    EA

    NM

    1

    Chapter 3: Imaging equipment - preparation and use

    by recording the time needed to acquire the

    food eld using the known activity. It should

    not vary by more than a ew percent rom one

    day to another.

    Weekly linearity and resolution assessment

    Linearity and resolution should be assessed

    weekly. This may be done using transmission

    phantoms.

    Centre o rotation calibration

    The centre o rotation measurement deter-

    mines the oset between the axis o rotation

    o the camera and the centre o the matrix

    used or reconstruction, as these do not cor-

    respond automatically.

    The calibration o the centre o rotation ismade rom the reconstruction o a tomo-

    graphic acquisition o a point source placed

    slightly oset rom the mechanical centre o

    rotation o the camera. A sinogram is ormed

    rom the projections and is used to t the

    maximum count locations to a sine wave. De-

    viations between the actual and tted curves

    should not exceed 0.5 pixels.

    Collimator

    The choice o a collimator or a given study

    is mainly determined by the tracer activity.

    This will infuence the statistical noise con-

    tent o the projection images and the spatial

    resolution. The number o counts needs to be

    maximised, possibly at the expense o some

    resolution and taking into account that in

    parathyroid imaging the dierence in tracer

    activity between 99mTcO4 (thyroid only) and

    any 99mTc-labelled agent (thyroid and parathy-

    roid) must be signicant.

    Collimators vary with respect to the relative

    length and width o the holes. The longer the

    hole length, the better the spatial resolution

    obtained, but at the expense o a lower countsensitivity. Conversely, a larger hole gives a

    better count sensitivity but with a loss o spa-

    tial resolution.

    When using 201Tl, the available counts are

    greatly reduced owing to the long hal-lie

    o the isotope and the consequent limited

    dose; so traditionally a low-energy general-

    purpose collimator is recommended. With99mTc-pertechnetate and 99mTc-labelled agents,

    count rate is no longer a major limitation, and

    urthermore, the resolution o a high-resolu-

    tion collimator decreases less with distance

    rom the source than does that o a general-

    purpose collimator. Thus a high-resolution

    collimator is currently recommended or

    SPECT imaging, despite the lower sensitivity.

    Although the choice o collimator is crucial, it

    should be borne in mind that other technical

    aspects play an important role in determining

    optimal spatial resolution, such as the matrix

    size, the number o angles and the time per

    view.

    Matrix and zoom actor

    The SPECT images (or projections rom the

  • 7/28/2019 Tech Guide Pt Scint

    20/44

    0

    angles round the patient) create multiple raw

    data sets containing the representation o the

    data in one projection. Each o these is stored

    in the computer in order to process them later

    on and extract the inormation.

    Matrix

    Each projection is collected into a matrix.

    These are characterised by the number opicture elements or pixels. Pixels are square

    and organised typically in arrays o 66,

    128128 or 256256.

    In act, the choice o matrix is dependent on

    two actors:

    a) The resolution: The choice should not de-

    grade the intrinsic resolution o the object. Thecommonly accepted rule or SPECT (Groch

    and Erwin 2000) is that the pixel size should be

    one-third o the ull-width at hal-maximum

    (FWHM) resolution o the organ, which will

    depend on its distance rom the camera ace.

    The spatial resolution o a SPECT system is o

    the order o 1825 mm at the centre o rota-

    tion (De Puey et al. 2001). Thus a pixel size o

    6-8 mm is sucient, which, or a typical large

    eld o view camera, leads to a matrix size o

    6464.

    b) The noise: This is caused by the statistical

    fuctuations o radiation decay. The lower the

    total counts, the more noise is present and, i

    the matrix size is doubled (128 instead o 64),

    the number o counts per pixel is reduced by

    a actor o 4. 128128 matrices produce ap-

    proximately three times more noise on the

    image ater reconstruction than do 6 x 6

    matrices (Garcia et al. 1990).

    The planar images (or static projections) do

    not have the reconstruction problem and can

    be acquired over longer times so a 256 256

    matrix is commonly used.

    Zoom actor

    The pixel size is dependent on the camera

    eld o view (FOV). When a zoom actor o

    1.0 is used, the pixel size (mm) is the useul

    FOV (UFOV, mm) divided by the number o

    pixels in one line. When a zoom actor is used,

    the number o pixels per line should rst be

    multiplied by this actor beore dividing it intothe FOV.

    Example:

    Acquisition with matrix 128, zoom 1.0 and

    UFOV 400 mm. Pixel size: 400/128=3.125 mm.

    The same acquisition with a zoom actor o 1.5.

    Pixel size: 400/(1.5128)=2.08 mm.

    It is important to check this parameter be-

    ore the acquisition, as it is very oten used in

    parathyroid imaging, especially i a subtraction

    technique is used.

    Preerred orbit

    Either circular or elliptical orbits can be used

    in SPECT imaging (Fig. 1). A circular orbit (Fig.

  • 7/28/2019 Tech Guide Pt Scint

    21/44

    EA

    NM

    1

    Chapter 3: Imaging equipment - preparation and use

    With a circular orbit, the camera is distant rom

    the body at some angles, causing a reduction

    in spatial resolution in these projections. This

    will reduce the resolution o the reconstructed

    images.

    With an elliptical orbit, spatial resolution will be

    improved as the camera passes closer to the

    body at all angles. Nevertheless, the distance

    rom the organ to the detector varies more

    signicantly with an elliptical orbit than with

    a circular orbit. This may generate arteacts

    simulating small photopenic areas when re-

    constructing using ltered back projection.

    Programmes that allow the camera to learn

    and closely ollow the contours o the body

    are available and improve resolution, although

    at the expense o computing power to modiy

    the data beore reconstruction.

    The loss o spatial resolution with a circular

    orbit has to be oset against the potential ar-

    teacts that may be generated by an elliptical

    or contoured orbit.

    Filtered back projection image

    reconstruction: some considerations

    The main goal o nuclear medicine parathy-

    roid imaging procedures is to identiy the

    site o parathyroid hormone production,

    usually a single parathyroid adenoma. How-

    ever, parathyroid adenomas can be ound in

    diverse locations: alongside, beside or within

    the thyroid, or in anatomical regions distantrom thyroid, such as high or low in the neck

    and mediastinum. The diversity o these ana-

    tomical locations makes SPECT a useul tool

    in parathyroid imaging.

    Furthermore, parathyroid adenomas are small

    structures with increased uptake oten close

    to normal thyroid activity. The choice o an

    optimum lter when using ltered back pro-

    jection or reconstruction is crucial (Pires Jorge

    et al. 1998).

    A lter in SPECT is a data processing algorithm

    that enhances image inormation, without

    signicantly altering the components o the

    input data, creating arteacts or losing inor-

    mation. It should produce results that lead to

    Figure 1a circular orbit

    Figure 1b elliptical orbit

    Figure 1a Figure 1b

    1a) is dened by a xed distance rom the axis

    o rotation to the centre o the camera surace

    or all angles. Elliptical orbits (Fig. 1b) ollow

    the body outline more closely.

  • 7/28/2019 Tech Guide Pt Scint

    22/44

    a correct diagnosis. Incorrect or over-ltering

    may produce adverse eects by reducing ei-

    ther resolution or contrast, or by increasing

    noise.

    A lter in SPECT, being a processing algorithm,

    operates in the requency-amplitude domain,

    which is obtained rom the spatial domain by

    the Fourier transorm. In the spatial domain,the image data obtained can be expressed by

    proles o any matrix row or column show-

    ing the activity distribution (counts) as a unc-

    tion o distance (pixel location). The Fourier

    method assumes that this prole is the sum

    o several sine and cosine unctions o dier-

    ent amplitudes and requencies. The Fourier

    transorm o the activity distribution o a given

    prole is a unction in which the amplitude othe sine or cosine unctions is plotted against

    the corresponding requency o each. This rep-

    resentation is also called the image requency-

    amplitude domain.

    In input data, the highest requency that

    can be measured is named the Nyquist re-

    quency, which is determined by the matrix

    size as well by the scintillation detector size

    and is expressed by the ormula: fn=1/(2d)

    where fn is the Nyquist requency and dis the

    acquisition pixel size. For example, when using

    a 6464 matrix with a 41-cm gamma camera

    UFOV, the pixel size (d) is 0.64 cm. Thereore

    the Nyquist requency is 0.78. This means that

    any input data where the requency is higher

    than 0.78 cannot be measured.

    The input data plotted in the image requen-

    cy-amplitude domain present three com-

    ponents that are partially superposed: the

    low-requency background, useul or target

    data and the high-requency noise. Here back-

    ground does not mean surrounding natural

    radiation or surrounding non-tissue activity

    but rather the low-requency waves gener-

    ated by the reconstruction process, such asthe well-known star arteact that appears in

    an unltered back projection. The high-re-

    quency noise is related to background and

    scatter radiation or statistical count fuctua-

    tions during SPECT acquisition, which may

    induce image distortions.

    Usually SPECT ltered back projection couples

    a ramp lter with an additional ltering (e.g.Hann, Hamming, Parzen). The ramp lter is

    so called as its shape looks like a ramp and

    it will eliminate an important portion o the

    unwanted low-requency background. How-

    ever, the ramp lter amplies the contribution

    o the high-requency noise to the image. This

    is why it is recommended that an additional

    lter be coupled with the ramp lter in order

    to smooth an image where some details could

    appear very noisy. The degree o smoothing

    or each additional lter is under the control

    o the user, as s/he has to decide the cut-

    o requency at which the lter will be ap-

    plied. The cut-o requency is the requency

    value that denes the maximum requency

    acceptable (which may contain useul data)

    while ignoring the higher requency noise.

  • 7/28/2019 Tech Guide Pt Scint

    23/44

    EA

    NM

    Chapter 3: Imaging equipment - preparation and use

    Obviously the maximum value o the cut-o

    requency or a given additional lter is the

    Nyquist requency.

    As parathyroid adenomas appear as small hot

    spots, requently within normal thyroid activ-

    ity, the optimum choice o lter is a high-pass

    type lter with a cut-o requency value close

    to the Nyquist requency. A high-pass typelter will be applied in order to eliminate the

    background image components (low requen-

    cy) and conserve target data, although some

    noise (high requency) will have to be toler-

    ated because o the low image smoothing.

  • 7/28/2019 Tech Guide Pt Scint

    24/44

    Patient preparationAudrey Taylor and Nish Fernando

    Patient identication

    To minimise the risk o a misadministration:

    Establish the patients ull name and other

    relevant details prior to administration o

    any drug or radiopharmaceutical.

    Corroborate the data with inormation pro-

    vided on the diagnostic test reerral.

    I the inormation on the reerral orm does

    not match the inormation obtained by the

    identication process, then the radiopharma-

    ceutical/drug should not be administered to

    the patient. This should be explained to the

    patient and clarication sought as soon as

    possible by contacting the reerral source.

    The patient/parent/guardian/escort shouldbe asked or the ollowing inormation, which

    should then be checked against the request

    orm and ward wristband in the case o an

    in-patient:

    Full name (check any spellings as appropri-

    ate, e.g. Steven vs Stephen)

    Date o birth

    Address

    I there are any known allergies or previous

    reactions to any drug, radiopharmaceutical,

    iodine-based contrast media or products

    such as micropore or Band-Aids

    A minimum o TWO corroborative details

    should be requested and conrmed as cor-

    rect.

    The ollowing inormation should be checked

    with the patient/parent/guardian/escort

    where appropriate:

    Reerring clinician/GP/hospital

    Any relevant clinical details

    Conrmation that the patient has complied

    with the dietary and drug restrictions

    Conrmation that the results o correlative

    imaging (e.g. echocardiography, angiogra-

    phy, etc.) are available prior to the study, andnoting o any recent interventions

    I in doubt, do not administer the radiophar-

    maceutical or drug and seek clarication.

    Specic patient groups

    This is a guide only. Patients who are unable to

    identiy themselves or any o a variety o rea-

    sons should wear a wrist identication band.

    Hearing diculties: Use written questions

    and ask the patient to supply the inormation

    verbally or to write their responses down.

    Speech diculties: Ask the patient to write

    down their name, date o birth and address

    and other relevant details.

  • 7/28/2019 Tech Guide Pt Scint

    25/44

    EA

    NM

    Chapter 4: Patient preparation

    Language diculties: I an accompanying

    person is unable to interpret the questions,

    then the study should be rebooked when

    a member o sta or relative with the ap-

    propriate language skills or an interpreter

    is available.

    Unconscious patient: Check the patients

    ID wristband or the correct name and dateo birth. I no wristband is attached, ask the

    nurse looking ater the patient to positively

    conrm the patients ID.

    Conused patient: I the patient is an in-pa-

    tient, check the patients ID wristband or

    the correct name and date o birth. I no

    wristband is attached, ask the nurse looking

    ater the patient to positively conrm thepatients ID. I the patient is an out-patient,

    ask the person accompanying the patient

    to positively conrm the patients ID.

    I a relative, riend or interpreter provides in-

    ormation re the patients name, date o birth

    etc., it is advisable or them to sign so as to

    provide written evidence conrming the rel-

    evant details.

    Patients can be required to send in a list o

    medications, approximate height, weight and

    asthma status so that stressing drugs can be

    chosen in advance. They should be advised to

    contact the department i they are diabetic so

    as to ensure that the appropriate guidance is

    given with regard to eating, medication etc.

    A ull explanation o the procedure should be

    given, including, risks, contraindications and

    side-eects o stress agents used, time taken

    or scan, the need to remain still etc.

    I the patients are phoned prior to appoint-

    ment, it acts as a reminder o the test and

    gives the patient an opportunity to discuss

    any concerns.

  • 7/28/2019 Tech Guide Pt Scint

    26/44

    QUESTIONNAIRE FOR ALL FEMALE PATIENTS OF CHILD BEARING AGE

    (12 55 YEARS)

    We are legally obliged under The Ionising Radiation (Medical Exposure) Regulations 2000

    to ask emales o child bearing age who are having a nuclear medicine procedure whether

    there is any chance they may be pregnant or breasteeding.

    Prior to your test, please answer the ollowing questions in order or us to comply with

    these regulations:

    PATIENT NAME ................................................................................................................................... D.O.B

    1. Have you started your periods? (please tick appropriate box)

    Y What is the date o your last period ...................................................................

    N Please sign below and we can then proceed with your test

    OR Have you nished your periods / had a hysterectomy (please tick appropriate box)

    Y Please sign below and we can then proceed with your test

    N What is the date o your last period

    2. Is there any chance you may be pregnant (please tick appropriate box)

    Y We will need to discuss your test with you beore we proceed

    Not sure We will need to discuss your test with you beore we proceed

    N Please sign below and we can then proceed with your test

    3. Are you breasteeding? (please tick appropriate box)

    Y We will need to discuss your test with you beore we proceed

    N Please sign below and we can then proceed with your test

    Pregnancy

    Women o childbearing potential should have their pregnancy status checked using a orm

    such as the example below:

  • 7/28/2019 Tech Guide Pt Scint

    27/44

    EA

    NM

    Chapter 4: Patient preparation

    I have read and understood the questions above and conrm that I am not pregnant or

    breasteeding and that I am aware that ionising radiation could damage a developing

    baby.

    Signed: ____________________________________ Date: _____________________

    (Patient)

    For all patients under 16 years o age

    I have read and understood the question above and conrm that the patient named is not

    pregnant or breasteeding

    Signed: ____________________________________ Date: _____________________

    Parent Guardian (please tick appropriate box)

    THIS FORM WILL BE CHECKED / DISCUSSED PRIOR TO THE START OF THE TEST

    The operator administering the radiopharma-

    ceutical should advise the patient on minimis-

    ing contact with pregnant persons and chil-

    dren. In addition, the operator administering

    the radiopharmaceutical should check that

    any accompanying person is not pregnant

    (e.g. escort nurse)

    Parathyroid patient preparation

    I possible, and under guidance rom the

    reerring clinician, the patient should be o

    any thyroid medication or 46 weeks prior

    to imaging.

    Establish whether the patient has had any

    imaging procedure using iodine contrast

  • 7/28/2019 Tech Guide Pt Scint

    28/44

    within the last 6 weeks (CT with contrast,

    IVU etc). Allow a period o 6 weeks between

    these procedures and thyroid imaging.

    Iodine-containing medications may have to

    be withdrawn, and the reerring clinicians

    advice should be sought. These medica-

    tions include: propylthiouracil, mepro-

    bamate, phenylbutazone, sulphonamides,corticosteroids, ACTH, perchlorate, anti-

    histamines, enterovioorm, iodides, Lugols

    solution, vitamin preparations, iodine oint-

    ments and amiodarone.

    Beore any pharmaceuticals are ordered,

    check whether the patient has had a total

    thyroidectomy. I this is the case, then the

    subtraction technique should not be car-ried out and consideration should be given

    to undertaking a dual-phase 99mTc-sestamibi

    study.

    Ask the patient whether he or she has any

    thyroid disorders such as thyrotoxicosis,

    hypothyroidism, thyroid nodules or thy-

    roid goitre. These conditions can increase

    instances o alse-positive 99mTc-sestamibi

    uptake and also aect 123I sodium iodide

    uptake. In the case o hypothyroidism, do

    not carry out the subtraction technique and

    consider undertaking a dual-phase 99mTc-

    sestamibi study.

    Ask the patient whether he or she is able to

    lie supine or the duration o the study and

    also whether he or she is claustrophobic.

    Consider another imaging modality i the

    patient cannot lie still or the duration o the

    study owing to discomort or anxiety.

    Although 123I sodium iodide contains little

    carrier-ree iodide, it is important to ask

    the patient about any adverse reactions

    to iodide in the orm o contrast media ormedication. I positive, seek the advice o

    the lead clinician.

  • 7/28/2019 Tech Guide Pt Scint

    29/44

    EA

    NM

    Imaging protocolsNish Fernando and Sue Huggett

    There are many variations in the imaging pro-

    tocols used. For dual-isotope studies where

    images are acquired sequentially with the sec-

    ond nuclide being injected ater the rst set o

    images, consideration must be given to timing

    o uptake and downscatter rom the higher

    energy nuclide when considering which

    nuclide to use rst. O course, simultaneous

    imaging, although aected by downscatter,obviates problems o image registration.

    One subtraction and one washout technique

    are described, including SPECT imaging, as ex-

    amples only. Explanations have been given or

    the choices so that adaptations can be made

    with knowledge o their eects.

    SPECT/CT has been suggested as a suitabletechnique to increase the sensitivity o de-

    tection (Gayed et al. 2005) but is beyond the

    scope o this booklet.

    123I sodium iodide/99mTc-sestamibi

    subtraction

    Give a ull explanation o the procedure to the

    patient. In particular, stress the importance

    o keeping still during the acquisition.

    Ensure good venous access. A venon with

    a three-way tap system into a vein in the

    patients arm or the back o the hand is

    more convenient than a butterfy needle

    as veins more requently collapse around a

    buttery needle than around the plastic o a

    venon. Also, the patient has more reedom

    o movement i a venfon is inserted.

    Inject 123I sodium iodide ollowed by a saline

    ush o 10 ml. Wrapping a bandage around

    the arm or hand where the venfon is sited

    will protect it during the period o delay.

    At 0 min post 123I injection, ask the pa-

    tient to empty the bladder. Ask the pa-tient whether he or she understands the

    procedure. Again, stress the importance

    o keeping still.

    At 50 min post 123I injection, position the pa-

    tient supine on the gamma camera couch.

    Ensure the neck is extended by positioning

    his/her shoulders on a pillow. Use sand-

    bags and a strap to immobilise the headand neck. Ensure the patient is comortable

    and understands the need to keep still. A

    pillow placed underneath the knees can

    reduce back discomort.

    Position the patient so that an anterior image

    o the thyroid and mediastinum can be ob-

    tained, allowing any ectopic tissue to be in-

    cluded in the image. Place the patients arm

    that has the venfon and three-way system

    onto an arm rest. Ensure patency o the ven-

    fon by fushing through with saline. Re-site

    the venfon i the vein has collapsed.

    Start acquisition at 60 min post 123I sodium

    iodide injection using a dual-isotope dy-

    namic acquisition with non-overlapping

  • 7/28/2019 Tech Guide Pt Scint

    30/44

    0

    windows or 99mTc (-10% to +5% about

    the peak at 140 keV) and 123I (-5% to +10%

    about the 159-keV peak).

    Acquire 2-min rames or 20 min using a

    zoom o .0 and a matrix o 128128. A

    dynamic acquisition is preerable to a static

    one as movement correction, provided it is

    in the x- or y-direction, can be applied to theimages. The large zoom is chosen in order

    to increase spatial resolution. However, this

    will increase noise in the image and so the

    acquisition must be o sucient duration

    to compensate or this.

    Pure iodide images may be acquired or 10

    min. As well as being critical or processing,

    these pure iodide images can be benecialin reducing alse-positive cases due to thy-

    roid disease.

    Without any patient movement, inject 99mTc-

    sestamibi ollowed by a 10-ml saline fush,

    between the 11th and 12th minute and

    continue the acquisition or 30 min.

    As a large zoom has been used, it is advis-

    able at the end o the dynamic acquisition

    to carry out an unzoomed image to include

    the salivary glands and the heart. This will

    ensure the detection o any ectopic para-

    thyroid glands, which can occur in the re-

    gion o the unzoomed image. Acquire this

    image on the same dual-isotope settings or

    300 s onto a matrix o 256256.

    Processing

    In order to detect the increased uptake o99mTc-sestamibi in the parathyroid tissue it is

    necessary to subtract the 123I image rom the

    99mTc-sestamibi image.

    The precise computer protocol will vary rom

    centre to centre and even rom camera system

    to camera system. However, all protocols willollow the same basic steps.

    Movement correction

    As the images have been acquired simultane-

    ously, there will be no need to match the posi-

    tions by shiting either image, but the images

    should be checked or movement and any

    correction algorithms applied beore com-

    mencing. This can be as simple as checkingall the rames and rejecting any with blurring

    beore the rames are summed. This will not

    help i the patient has moved to a dierent

    position rather than having coughed or swal-

    lowed deeply and returned to the original

    position.

    I overall movement has occurred, the situa-

    tion may be rescued i the subsequent rames

    can be shited by reerence to some standard

    point (sometimes the hottest pixel) or even

    by eye.

    The subtraction

    Both sets o rames (corrected i necessary)

    are summed to make one 99mTc and one 123I

    image.

  • 7/28/2019 Tech Guide Pt Scint

    31/44

    EA

    NM

    1

    Chapter 5: Imaging protocols

    There will be more counts in the thyroid on the123I image, so it must be matched to the 99mTc

    image beore subtraction and a scaling actor is

    used so that the counts in the subtraction im-

    age are reduced by this actor, pixel by pixel.

    The simplest technique reduces the image

    to be subtracted to, or example, 30%, 0%,

    50%, 60% and 70% o its original values, andthese images are subtracted in turn rom the99mTc-sestamibi image, that which gives the

    best result or eliminating the thyroid tissue

    being chosen by eye.

    A more automated system will draw a region

    o interest around the normal thyroid on the123I image by allowing the operator to choose

    the count contour line which best represents itsedges. The counts in this region are then com-

    pared against the counts in the same region on

    the 99mTc image. The scaling actor is calculated

    rom the ratio o these two values. This adjusted

    image is then subtracted rom the 99mTc image

    and the results displayed as a new image. Again,

    there are usually two or three options oered or

    the operator to choose the best result.

    SPECT imaging

    Additional SPECT imaging gives increased

    sensitivity and more precise anatomical lo-

    calisation. Acquiring both early and delayed

    SPECT can be a useul addition to either the

    dual-phase 99mTc-sestamibi method or the

    dual-isotope 123I/99mTc-sestamibi subtraction

    method.

    Early SPECT should be acquired 1030 min

    ater the 99mTc-sestamibi injection and de-

    layed SPECT at around 3 h post injection.

    The camera is peaked or both 99mTc and123I as beore.

    An 180o acquisition optimises the time close

    to the area o interest and attenuation isnot a problem with structures so close to

    the body surace.

    Acquisition should start with the camera

    head at 270o; proceed in a clockwise rota-

    tion and stop at 90o.

    I the acquisition is carried out on a double-

    headed camera, a 90o L-Mode SPECT will beuseul and more counts will be collected as

    both heads are used or the acquisition.

    Contouring can be used i available, al-

    though the patient should be warned i

    the camera will move closer during the

    acquisition.

    It should be ensured that the zoom chosen

    allows adequate coverage o the mediasti-

    num to locate any ectopic glands there.

    Using a zoom o 2 will ensure better spatial

    resolution than no zoom, and there should

    be enough coverage o the mediastinum.

    A 6464 matrix is sucient or the expected

  • 7/28/2019 Tech Guide Pt Scint

    32/44

    resolution and will optimise counts per pixel

    and hence reduce noise.

    30 60-s rames is manageable or the patient

    and will maximise counts in the image.

    For delayed SPECT, increasing the time

    per projection to 90 s will restore the total

    counts.

    Processing

    The data may be reconstructed using the

    methods o ltered back projection or itera-

    tive reconstruction. Streak arteacts may be

    seen in data reconstructed by ltered back

    projection and these will not occur or the

    iterative method.

    The raw data can be viewed as a rotating

    image and the limits or the region to be

    reconstructed chosen.

    The region should extend rom the parotid

    glands to the mediastinum to locate any

    ectopic tissue.

    The reconstruction programme with cho-

    sen lters is initiated.

    Once the reconstruction is completed, the

    images are viewed in the transverse, coronal

    and sagittal planes.

    Datasets can be viewed as volumetric dis-

    plays as well as tomographic slices.

    99mTc-sestamibi washout technique

    I 99mTc-sestamibi is used alone, the two sets

    o images (early and delayed) are inspected

    visually.

    740 MBq 99mTc-sestamibi is injected using the

    same protocols or patient preparation, i.e.

    ID and LMP checks/explanation to patient as

    beore and imaging typically at 10 min and23 h

    The camera is peaked or technetium and

    a low-energy high-resolution collimator

    can be used as images can be taken or

    a sucient time to avoid statistical noise

    problems and pinhole collimators are used

    in some centres. Zoom can be used but

    remember the possibility o ectopic tissue.

    Neck and mediastinum views are taken with

    patient positioning as beore. Again, a single

    view o 600 s counts can be taken or a series

    o 1060-s rames acquired so that move-

    ment arteacts can be corrected.

    The same parameters and positioning must

    be used or the 10-min and late views

    Right and let anterior oblique views can be

    obtained i required.

    SPECT can be used in this case also.

    All lms should be correctly annotated with L,

    R and anatomical markers and labelled.

  • 7/28/2019 Tech Guide Pt Scint

    33/44

    EA

    NM

    Technical aspects o probe-guided surgery orparathyroid adenomasDomenico Rubello

    Bilateral neck exploration (BNE) still represents

    the gold standard approach in patients with

    primary hyperparathyroidism (pHPT). Howev-

    er, surgical approaches to pHPT patients have

    altered signicantly in many surgical centres

    during the past decade, with the development

    o minimally invasive parathyroidectomy using

    endoscopic surgery or radio-guided surgery

    (MIRS). This development can be attributed totwo main reasons: (a) the consciousness that

    pHPT is due to a single parathyroid adenoma

    in the majority o patients (at least 85%), and

    (b) the technical improvements introduced

    into surgical practice with the availability o

    microsurgery instruments, endoscopes, intra-

    operative measurements o quick parathyroid

    hormone (QPTH) and gamma probes.

    New approaches to minimally invasive para-

    thyroidectomy consisting in the removal o a

    solitary parathyroid adenoma via a small 12

    cm skin incision have been widely adopted. O

    course, in contrast to BNE, minimally invasive

    parathyroidectomy always requires accurate

    preoperative imaging in order (a) to establish

    whether the parathyroid adenoma is eec-

    tively solitary and (b) to locate precisely the

    enlarged gland. The present chapter ocusses

    mainly on technical aspects o the MIRS tech-

    nique. Moreover, the MIRS technique devel-

    oped in our centre is based on the injection

    o a very low 99mTc-sestamibi dose 37 MBq

    compared with the traditional MIRS tech-

    nique, which uses a high 99mTc-sestamibi

    dose 740925 MBq.

    Selection criteria or ofering MIRS

    When planning MIRS (unlike when perorm-

    ing BNE), strict inclusion criteria need to be

    ollowed: (a) evidence at 99mTc-sestamibi scin-

    tigraphy o a solitary parathyroid adenoma;

    (b) intense 99mTc-sestamibi uptake in the para-

    thyroid adenoma; (c) absence o concomitant

    thyroid nodules at 99mTc-sestamibi scintigraphy

    and high-resolution (10 MHz) neck ultrasound;(d) no history o amilial HPT or multiple endo-

    crine neoplasia; and (e) no history o irradia-

    tion to the neck. O note, previous thyroid or

    parathyroid surgery is not a contraindication

    to MIRS. When these inclusion criteria are

    adopted, approximately 6070% o pHPT pa-

    tients can be oered MIRS. The main reason or

    exclusion is the presence o 99mTc-sestamibi-

    avid thyroid nodules, which, by mimicking aparathyroid adenoma, can cause alse posi-

    tive results during surgery. Figure 1 shows a

    patient scheduled or MIRS while Fig. 2 shows

    a patient excluded rom MIRS.

    Preoperative imaging protocol

    In our protocol, preoperative imaging proce-

    dures include single-session 99mTc-sestamibi

    scintigraphy and neck ultrasound (Norman

    and Chheda 1997; Costello and Norman

    1999; Mariani et al. 2003; Rubello et al. 2000).

    In patients with concordant 99mTc-sestamibi

    and ultrasound results (both positive or nega-

    tive), no urther imaging is perormed, while

    in cases with discrepant ndings (99mTc-sesta-

    mibi positive and US negative) a tomographic

    (SPECT) examination is obtained to investi-

  • 7/28/2019 Tech Guide Pt Scint

    34/44

    gate a possible ectopic or deep position o the

    parathyroid adenoma. SPECT is obtained just

    ater the completion o planar 99mTc-sestamibi

    scintigraphy, thus using the same radiotracer

    dose; in this way, 99mTc-sestamibi re-injection is

    not necessary, thus avoiding additional radia-

    tion exposure to the patient and personnel.

    In other centres, preoperative 99mTc-sestamibi

    scintigraphy alone is considered a sucienttool or the planning o MIRS.

    Intra-operative MIRS protocol

    Table 1 shows the steps in the MIRS protocol

    used in our centre.

    A collimated gamma probe is recommended

    with an external diameter o 111 mm. A

    non-collimated probe, which can be used orsentinel lymph node biopsy, is not ideal or

    parathyroid surgery owing to the relative com-

    ponent o diuse and scatter radioactivity de-

    riving rom the anatomical structures located

    near to the parathyroid glands, mainly related

    to the thyroid gland. Probes utilising either a

    NaI scintillation detector or a semiconductor

    detector have proved adequate or MIRS.

    A learning curve o at least 2030 MIRS op-

    erations is recommended or an endocrine

    surgeon. During these, the presence in the

    operating theatre o a nuclear medicine phy-

    sician is usually considered mandatory. In

    the opinion o the writer, the presence o a

    nuclear medicine technician, with expertise

    in probe utilisation, is very useul in helping

    the surgeon to become more skilled in the

    use o the probe.

    The probe is usually handled by the surgeon,

    who should measure radioactivity in dier-

    ent regions o the thyroid bed and neck in an

    attempt to localise the site with the highest

    count rate beore commencing the operation.

    This site is likely to correspond to the parathy-roid adenoma. Then, during the operation, the

    surgeon should measure the relative activity

    levels in the parathyroid adenoma, thyroid

    bed and background. Moreover, a check o

    the empty parathyroid bed ater removal

    o the parathyroid adenoma is a very useul

    parameter to veriy the completeness o re-

    moval o hyperunctioning parathyroid tissue.

    Ex vivo measurement o any removed surgicalspecimen should be done to veriy the total

    clearance o the parathyroid adenoma. The

    calculation o tissue ratios parathyroid to

    background (P/B) ratio, thyroid to background

    (T/B) ratio, parathyroid to thyroid (P/T) ratio

    and the empty parathyroid bed to background

    (empty-P/B) ratio can be useul in evaluating

    the ecacy o MIRS. The tissue ratios obtained

    in a large series o 355 pHPT patients operated

    on in our centre are reported in Table 2.

    Attention has to be given to avoidance o

    intra-operative alse negative results due to99mTc-sestamibi-avid thyroid nodules and to

    stagnation o the radiotracer within vascular

    structures o the neck and thoracic inlet: in this

    regard, the careul acquisition and evaluation

  • 7/28/2019 Tech Guide Pt Scint

    35/44

  • 7/28/2019 Tech Guide Pt Scint

    36/44

    Table 1. Steps in minimally invasive radioguided surgery (MIRS) o parathyroid adenomas using

    the low 99mTc-sestamibi dose protocol developed in our centre

    Blood samples are drawn rom a peripheral vein both beore commencing surgery and 10 min

    ollowing the removal o the parathyroid adenoma to measure intra-operative QPTH levels

    37 MBq o99mTc-sestamibi is injected in the operating theatre 10 min beore the start o

    surgery

    Prior to surgical incision, the patients neck is scanned with an 11-mm collimated probe to

    localise the site with the highest count rate, corresponding to the cutaneous projection o

    the parathyroid adenoma

    A transverse midline neck access (approximately 1 cm above the sternal notch) is preerred

    because conversion to BNE is easily obtained i necessary

    An 11-mm collimated probe is repeatedly inserted through a 2-cm skin incision, guiding the

    surgeon to the maximum count rate area corresponding to the parathyroid adenoma

    In some patients with a parathyroid adenoma located deep in the neck, ligature o the middle

    thyroid vein and o the inerior thyroid artery is required

    Radioactivity o the parathyroid adenoma, thyroid gland and background is measured with

    the probe

    Radioactivity is measured ex vivo to conrm successul removal o parathyroid tissue

    Radioactivity o the empty operation site is checked to evaluate the completeness o para-

    thyroid tissue removalTissue ratios are calculated (P/B, P/T etc.)

    QPTH, quick parathyroid hormone; BNE, bilateral neck exploration; P/B, parathyroid to back-

    ground ratio; P/T, parathyroid to thyroid ratio

  • 7/28/2019 Tech Guide Pt Scint

    37/44

    EA

    NM

    Chapter 6: Technical aspects of probe-guided surgery for parathyroid adenomas

    tamibi dose protocol, with a success rate in

    the intra-operative detection o parathyroid

    adenoma o approximately 9698%, without

    major intra-operative surgical complications. It

    is likely that Normans single-day protocol will

    be preerable in patients with a low likelihood

    o nodular goitre whilst our dierent-day pro-

    tocol appears preerable in areas with a higher

    prevalence o nodular goitre.

    Irrespective o the type o MIRS protocol used,

    the principal advantages o MIRS over tradi-

    tional BNE can be summarised as: (a) a small

    skin incision with avourable cosmetic results,

    (b) a shorter operating time, (c) the possibility

    o perorming MIRS under local anaesthesia,

    (d) a shorter hospital recovery time, (e) the

    possibility o same-day hospital discharge, ()

    lower post-surgical time and (g) lower costs.

    Table 2. Probe tissue ratios calculated during MIRS or parathyroid adenoma removal (n=355

    pHPT patients)

    P/B ratio = 1.64.8 (mean 2.60.5)

    P/T ratio = 1.12.8 (mean 1.50.4)

    T/B ratio = 1.51.8 (mean 1.60.1)

    Empty-P bed/B ratio = 0.91.1 (mean 1.00.03)

    TN/P ratio = 0.51.5 (mean 1.00.4)

    P=parathyroid

    T=thyroid

    B=background

    TN=thyroid nodule

  • 7/28/2019 Tech Guide Pt Scint

    38/44

    Table 3. Radiation dose to operating theatre personnel during MIRS with the low (37 MBq)99mTc-sestamibi dose protocol used in our centre

    Gy/hour Gy/year*

    Surgeons body 1.2 120

    Surgeons hands 5.0 500

    Anaesthesiologist 0.7 70

    Instrument nurse 1.1 110

    Other nurses 0.1 10

    *Estimated or 100 interventions, each lasting 60 min

    Figure 1. Preoperative dual-tracer parathyroid subtraction scintigraphy. Left image: 99mTc-pertech-

    netate scan showing a normal thyroid gland. Middle image: 99mTc-sestamibi scan showing an

    area o radiotracer uptake juxtaposed to the lower pole o the let thyroid lobe. Right image:

    Subtraction (99mTc-sestamibi-99mTc-pertechnetate) image, clearly showing a let inerior para-

    thyroid adenoma. This patient was oered MIRS.

  • 7/28/2019 Tech Guide Pt Scint

    39/44

    EA

    NM

    Chapter 6: Technical aspects of probe-guided surgery for parathyroid adenomas

    Figure 2. Preoperative dual-tracer parathyroid subtraction scintigraphy. Left image: 99mTc-pertech-

    netate scan showing a multinodular goitre with some hyperunctioning nodules (three in the

    right thyroid lobe, one in the let thyroid lobe). Middle image: 99mTc-sestamibi scan showing

    a picture similar to the 99mTc-pertechnetate image plus a let superior area o exclusive 99mTc-

    sestamibi uptake. Right image: subtraction (99mTc-sestamibi-99mTc-pertechnetate) image clearly

    showing a let superior parathyroid adenoma. This patient was excluded rom MIRS due to the

    coexistence o a solitary parathyroid adenoma and multinodular goitre with multiple99m

    Tc-sestamibi-avid thyroid nodules in both thyroid lobes

  • 7/28/2019 Tech Guide Pt Scint

    40/44

    0

    Chapter 1

    Reerences1. Akerstrm G, Malmaeus J, Bergstrm R. Surgical anato-

    my o human parathyroid glands. Surgery 1984;95:1421.

    2. Al Zahrani AA, Levine MA. Primary hyperparathyroidism.

    Lancet 1997; 349:12331238.

    3. Bilezikian JP, Potts JT Jr, Fuleihan Gel-H, Kleerekoper M,

    Neer R, Peacock M, et al. Summary statement rom a work-

    shop on asymptomatic primary hyperparathyroidism: a

    perspective or the 21st century [review]. J Clin EndocrinolMetab 2002;87:53535361.

    4. Borley NR, Collins REC, ODoherty M, Coakley A. Tech-

    netium 99m-sestamibi parathyroid localization is accurate

    enough or scan-directed unilateral neck exploration. Br J

    Surg 1996;83:989991.

    5. Coakley AJ, Kettle AG, Wells CP, ODoherty MJ, Collins

    REC. 99mTc sestamibi - a new agent or parathyroid ima-

    ging. Nucl Med Commun 1989;10:791794.

    6. Giordano A, Rubello D, Casara D. New trends in parathyroid

    scintigraphy [review]. Eur J Nucl Med 2001;28:14091420.

    7. Haber RS, Kim CK, Inabnet WB. Ultrasonography

    or preoperative localization o enlarged parathyroid

    glands in primary hyperparathyroidism: comparison with

    (99m)technetium sestamibi scintigraphy. Clin Endocrinol

    (Ox ) 2002;57:241249.

    8. Heath H, Hodgson SF, Kennedy MA. Primary hyperpara-

    thyroidism. Incidence, morbidity, and potential economic

    impact in a community. N Engl J Med 1980;302:189193.

    9. Hindi E, Melliere D, Perlemuter L, Jeanguillaume C,

    Galle P. Primary hyperparathyroidism: Higher success rate

    o rst surgery ater preoperative Tc-99m-sestamibi/I-123

    subtraction scanning. Radiology 1997;204:221-228.

    10. Hindi E, Mellire D, Jeanguillaume C, Perlemuter L,

    Galle P. Parathyroid imaging using simultaneous double-

    window recording o Tc-99m-sestamibi and 123I. J Nucl

    Med 1998;39:11001105.

    11. Hindi E, Urea P, Jeanguillaume C, Melliere D, Berthelot

    JM, Menoyo-Calonge V, et al.. Preoperative imaging o pa-

    rathyroid glands with technetium-99m-labelled sestamibi

    and iodine-123 subtraction scanning in secondary hyper-parathyroidism. Lancet 1999;353:22002204.

    12. Hindi E, Melliere D, Jeanguillaume C, Urena P, deLabri-olle-Vaylet C, Perlemuter L. Unilateral surgery or primary

    hyperparathyroidism on the basis o technetium Tc 99m

    sestamibi and iodine 123 subtraction scanning. Arch Surg

    2000;135:14611468.

    13. Lee JA, Inabnet WB 3rd. The surgeons armamentarium

    to the surgical treatment o primary hyperparathyroidism

    [review]. J Surg Oncol 2005;89:130135.

    14. Levin KE, Clark OH. The reasons or ailure in parathyroid

    operations. Arch Surg 1989;124:911915.

    15. Liu RC, Hill ME, Ryan JA Jr. One-gland exploration or me-

    diastinal parathyroid adenomas: cervical and thoracoscopic

    approaches. Am J Surg 2005;189:601605.

    16. Lundgren E, Rastad J, Thurfell E, Akerstrm G, Ljunghall

    S. Population-based screening or primary hyperparathyro-

    idism with serum calcium and parathyroid hormone values

    in menopausal women. Surgery 1997; 121:287294.

    17. Martin D, Rosen IB, Ichise M. Evaluation o single isoto-

    pe technetium-99m-sestamibi in localization eciency or

    hyperparathyroidism. Am J Surg 1996;172:633636.

    18. Marx SJ, Simonds WF, Agarwal SK, Burns AL, Weinstein

    LS, Cochran C, et al. Hyperparathyroidism in hereditary

    syndromes: special expressions and special managements

    [review]. J Bone Miner Res 2002;17 Suppl 2:N37N43.

    19. Molinari AS, Irvin GL 3rd, Deriso GT, Bott L. Incidence

    o multiglandular disease in primary hyperparathyroidism

    determined by parathyroid hormone secretion. Surgery

    1996;120:934937.

    20. Mullan BP. Nuclear medicine imaging o the parathyroid

    [review]. Otolaryngol Clin North Am 2004;37:909939, xixii.

    21. Peri S, Fessi H, Tassart M, Younsi N, Poli