TASTEPROBE Type DTP-1 Pre-amplifier for recording from ...
Transcript of TASTEPROBE Type DTP-1 Pre-amplifier for recording from ...
TASTEPROBE Type DTP-1 Pre-amplifier for recording from contact chemosensilla
INSTRUCTIONS
© SYNTECH 2006
Kirchzarten, Germany
Reproduction of text and/or drawings is permitted for personal use. The use of reproductions in publications is only allowed if correct reference is made to SYNTECH
TastePROBE
Type DTP-1
with high input impedance headstage and electrode holder
SYNTECH TastePROBE DTP-02
CLIP INPUT
HEADSTAGE
READY
RUNNING
CLIP
MENU ENTERUP
(MODE)
DOWN
(RUN)
TOTALSYSTEM GAIN
10 x
(GAIN 1 x)
IDACSIGNAL / POWER
INTERFACE
TRIGGERENABLE
PEDALSWITCH
TRIGGEROUT
SIGNALOUT
12 V DCEXT. POWER
SUPPLY
+
TastePROBEDTP-02
FRONT PANEL
REAR PANEL
Fig. 1 TASTEPROBE FRONT and REAR PANEL
1 INTRODUCTION
The SYNTECH Tasteprobe, Type DTP-1, is a further development of the prototype of the special amplifier described by Marion-Poll and Van der Pers (1996) that permits reliable DC recordings from insect taste receptors using the Hodgson (1955) technique. The present model can operate in DC or in AC mode. In DC mode the complete signal, slow (DC) and fast (AC) fluctuations are recorded and amplified. In AC mode the slow (DC) fluctuations are out-filtered and only the fast (AC) components of the signal (like Spikes) are amplified. In both recording modes the DC initial offset at the onset of the recording (before contact is made between the recording electrode and the preparation) is cancelled automatically. DESCRIPTION (The numbers refer to the encircled numbers in Fig. 1) FRONT PANEL 1 Input receptacle for the headstage (Probe) containing the high input
impedance pre-amplifier. The pipette electrode holder is mounted directly on the headstage.
2 Display showing the settings and the modes of operation 3 LED indicating the 'ready' status of the device 4 LED showing the 'running' status 5 LED indicating a positive over-voltage at the input 6 LED indicating a negative over-voltage at the input 7 MENU button 8 Selection button ; also: ENABLE button 9 Selection button ; also: PASS-THROUGH operation 10 Enter button to leave the menu and to return to ready REAR PANEL 11 Receptacle to connect the Tasteprobe with a Syntech IDAC interface 12 12 V Power input (not needed with Syntech IDAC interface) 13 Signal output (10 x amplified) 14 Trigger signal output 15 Enable switch (pedal switch) input
Probe
PEDAL SWITCH
GROUND of SET UP
Electrode holder with pipette electrode
IDACSIGNAL / POWER
INTERFACE
TRIGGERENABLE
PEDALSWITCH
TRIGGEROUT
SIGNALOUT
12 V DCEXT. POWER
SUPPLY
TastePROBEDTP-02
Fig. 2 CONNECTIONS FOR USE WITH SYNTECH IDAC-4
SYNTECH Tast ePROBE DTP-02
CLIP INPUT
HEADSTAGE
READY
RUNNING
CLIP
MENU ENTERUP
(MODE)
DOWN
(RUN)
TOTALSYSTEM GAIN
10 x
(G A IN 1 x )
Syntech NLTasteProbe V2.0
FRONT REAR
INTELLIGENT DATA ACQUISITION CONTROLLER PROGRAMMABLE OUTPUT DRIVERSIDAC 4
INPUTS / OUTPUTS
DIG ITAL INPUTS
DIG ITAL IN / OUT
INDICATORS
DIGITAL OUTPUTS
1 12 23 34 45 56 67 78 8
ANALOGSIGNAL INPUTS
DIGITALINPUTS
1 - 4
M AINS
ACTIVE
AUDIO OUT
2 OPERATION USING THE INITIAL SETTINGS The instrument can be operated without changing the settings as entered by the manufacturer. These setting are suitable for a first test and to become familiar with the system. After the device has been connected the settings can be checked by pushing the MODE button repeatedly; push ENTER to return to the READY state. CONNECTIONS to Syntech USB-IDAC (Fig. 2) 1 Connect the headstage (probe) 2 Connect the pedal switch 3 Make the connection between the IDAC interface using the special
cable supplied (see figure). Connect to the Analog 1 input of the USB-IDAC (Fig. 2). (Power is supplied by the USB-IDAC through this cable).
4 Make the connection for the Trigger : connect the Trigger out at the rear of the Tasteprobe to the Digital 1 input of the USB-IDAC.
5 Configure the Autospike program to record at least Channel 1, enter the ext. ampl. factor to 10; set the high Filter at 3000 Hz, the low Filter to 100 Hz, and set the Digital channel 1 to trigger at the highest sample rate (Fig. 7). Connect an Audio monitor amplifier to the Audio output of the USB-IDAC, and enable the Audio output for Channel 1 (Fig. 7).
CONNECTIONS to any DATA ACQUISTION SYSTEM (fig. 3) 1 Connect the headstage (probe) 2 Connect the pedal switch 3 Connect the SIGNAL OUT receptacle at the rear of the Tasteprobe to
an analog signal input of a suitable DATA ACQUISITION system. 4 Make the connection for the Trigger : connect to the Trigger out at the
rear of the Tasteprobe to the trigger input of the DATA ACQUISITION system
5 Provide the 12V power at the rear of the Tasteprobe using a suitable power supply (12 V DC, regulated ; min. 200 mA)
6 Configure the Data Acquisition software for recording of the analog signal, and select a NEGATIVE trigger input. (If only a Positive trigger input is available, change the Trigger setting in the Tasteprobe menu to Positive (Fig. 6).
Probe
PEDAL SWITCH
GROUND of SET UP
DATA ACQUISITION SYSTEM
12 V POWER SUPPLY
ANALOG SIGNAL INPUT
TRIGGERINPUT
Electrode holder with pipette electrode
SYNTECH Tast ePROBE DTP-02
CLIP INPUT
HEADSTAGE
READY
RUNNING
CLIP
MENU ENTERUP
(MODE)
DOWN
(RUN)
TOTALSYSTEM GAIN
10 x
(G A IN 1 x)
Syntech NLTasteProbe V2.0
IDACSIGNAL / POWER
INTERFACE
TRIGGERENABLE
PEDALSWITCH
TRIGGEROUT
SIGNALOUT
12 V DCEXT. POWER
SUPPLY
+
TastePROBEDTP-02
Fig. 3 CONNECTIONS FOR USE WITH DATA ACQUISITION SYSTEM
REAR
12 VDC
FRONT
3
INPUT CONFIGURATION (Fig. 4) 7 Prepare the pipette electrode (Fig. 5). 8 Select and prepare a suitable insect preparation.
Connect the preparation to the central (common) GROUND of the recording set up.
9 Connect the shaft of the headstage also to the central GROUND 10 Connect the Faraday cage and other metal devices to the central
GROUND PASS-THROUGH RECORDING 11 Start the Syntech Auto spike program. Check the Recording properties and the Filter settings (Fig. 7) or start another suitable Data Acquisition program. 12 Press the ENTER button of the Tasteprobe. The READY Led should go ON. 13 Check the baseline on the display of the recording system. Adjust the vertical scale (sensitivity) of the recording display
between 1 mV - 10 mV. 14 Press the UP (MODE) button: The Tasteprobe is now in PASS-THROUGH mode. In this mode the system operates as a normal 10x amplifier. 15 Establish contact with the preparation. Record the resulting signal trace.
If the data acquisition system and software settings all are properly adjusted the signal from the preparation should be visible on the signal display. The characteristics of the recorded signal should be the same as expected from a normal 10x amplifier with a high-pass filter of 100 Hz. Offset voltages are NOT compensated in the PASS-THROUGH mode! (Signals from taste sensilla may NOT come through!)
GROUND of SET UP
Faraday cage
To USB port
IDACSIGNAL / POWE R
INT ERFACE
T RI G G EREN AB L E
PE D A LS W I T CH
T RI G G E RO U T
SI G NA LO U T
12 V DCEXT. POWER
SUPPLY
TastePROBE
DTP-02
SYNTECH TastePROBE DTP-02
C L IP INP UT
HEADSTAGE
R EA DY
R UN N IN G
C L IP
MENU ENTE RUP
(MODE)
DOWN
(RUN)
TO TALSYST EM G AIN
10 x
( G AI N 1 x)
Syntech NLTasteProbe V2.0
USB CABLE
IDAC-4
Preparation
Pedal switch
Probe
Micro Manipulator
REARFRONT
Fig. 4 GENERAL RECORDING ARRANGEMENT
IN T ELLIG EN T D ATA A C QU ISIT IO N C ONT R O LL ER PR O GRA M M A BLE OU TPU T D RIV ER SIDAC 4
INPU TS / O UT PU TS
DIG ITA L IN PUT S
DIG ITA L IN / OU TIND ICATO R S
DIG ITA L O UT PUT S
1 12 23 34 45 56 67 78 8
AN AL OG
SIG NAL IN P UTS
DIG ITA L
INP U TS1 - 4
MAI NS
AC TIVE
AU DIO O UT
CROSS SECTION of ELECTRODE HOLDER
SILVER WIRE 0.3 - 0.5 mm
make t iny bend
at end of wi re
inse r t wi re into ho lder
GLASS PIPETTE ELECTRODE diam. 1.5 mm
screw ho lder onto probe
inse r t p ipette into e lectrode holder
Do NOT connect e lectrode holder to ground!
s ta in le ss s tee l body
wi r e gu ide
ins u la t io n
p ipet te gu ide
Fig. 5 Electrode Preparation
Fig. 6 MENU OPTIONS and SETTINGS
DC - - - - -100 Hz
-2V, -1V, GND, + 1V, + 2V , Open
Continue, 1 - - - - 100 Sec
0 - - 1 - - - - 30 Sec
0.5 - - - - 100 ms
Negative - - - Positive
1 - - - - - 12
0 - - - - - 60
ON - - - - OFF
ON - - - - OFF
OFF - - ON
Fig. 7 AutoSpike RECORDING and FILTER SETTINGS
Recording Properties / Settings
Recording Properties / Digital I/O
Filter Settings
* Enable Channel 1
* Select Sample rate at 11904 or higher
* Set External Amplification to 10
* Set Digital Sample rate at 1904 (the maximum)
* Enable Channel 1
* Set Trigger to input 1
* Set A/B Channel to None
* Set High cutoff at 3kHz
* Set Low cutoff at 100Hz
* Enable Audio out
Fig. 8 AutoSpike FILTER SETTINGS and OFFSET CONTROL for DC RECORDING
* Set High cutoff at 3kHz
* Set Low cutoff at DC
* Enable Audio out
Use SLIDER for coarse
adjustmentUP - DOWNbuttons for
fineadjustment
Filter Settings
4
NORMAL (Offset-compensated) RECORDING 16 Disconnect the pipette electrode from the preparation 17 Press the ENTER button (The READY Led goes ON) if needed.
This is only necessary if the system is NOT already in the READY state.
18 Press the PEDAL SWITCH (or the DOWN (RUN) button).
The display now shows: WAITING FOR CONTACT... and the RUNNING Led is blinking
It is possible that the RECORDING is triggered, indicated on the display and by a BEEP. If this is the case, the TRIGGER threshold must be increased. Press the MENU button several times until the Trig. Threshold appears; use the UP button to increase the value (max. is 12;Fig. 6)
19 Make contact with the preparation. A BEEP should be audible, and the display shows RECORDING,
indicating that the offset has been compensated and the recording is started. During recording the RUNNING Led is ON. The display counts down the recording time (in seconds) and the end of the recording time is indicated by another BEEP.
If NO Beep is produced, and the system does NOT go into the Recording state, the TRIGGER threshold might be too high; Decrease the Trig. Threshold in the appropriate MENU box (Fig. 6).
The system returns to the READY state automatically.
20 Check the recorded signal on the display of the signal acquisition system used.
Adjust scale and timing values if necessary 21 To make another recording disconnect the electrode from the
preparation, and press the PEDAL switch (or the DOWN (RUN) button) just before making contact again.
5 DC RECORDING For DC recording it is necessary to set all signal conditioning circuits in DC mode and to adjust the output baseline level exactly to zero. The signal may contain large DC fluctuations; therefore, the signal recording interface (Syntech IDAC or any other acquisition system) output scale should be adjusted accordingly (10 - 20 mv) . If full DC recording results in too large fluctuations the signal can be filtered (High-Pass 0.1 Hz - 10 Hz; see Fig. 6) until the signal fits in the recording amplitude window. The DC component of the signal is stronger suppressed at higher filter values. For DC recording, the following settings and adjustments must be made: * In the Tasteprobe: High-Pass : DC (Fig. 6) * In the IDAC Filter settings: Low cutoff: DC (Fig. 8)
The filter settings in any other acquisition system must be set to DC * Set the signal display scale to 10 or 1 mV and bring the signal trace to
the zero level (Fig. 8) Operation of the Tasteprobe in DC mode is otherwise the same as for AC recording ( steps 16- 21) ZEROING TIME In DC recording the signals may show large irregular fluctuations at the onset , which are a result of instabilities during the initial contact with the sensillum. These large fluctuations may be suppressed by increasing the ZEROING TIME. However, any actionpotentials generated during the zeroing time will not be recorded. Therefore, the zeroing time should be as short as possible. PROBE OFFSET and TRIGGER THRESHOLD Normally the Probe Offset (Fig. 6) is kept at ground (= zero) potential by a shunt resistor of 1 Gohm inside the headstage. This offset voltage can be adjusted from -2 to +2 V (Fig. 6) to test the sensitivity of the preparation to input offset transients or the improve the activation of the trigger. The input may also be left in the open state. The trigger threshold is adjustable in 12 steps, and should be adjusted such that no 'false triggering' occurs before the pipette has been brought in contact with the preparation, but that the trigger does respond immediately after contact has been made. The trigger can only be activated in the WAITING FOR CONTACT.... state.
6 AUTOMATIC INITIALISATION (AUTO RESTART) It is not always necessary to use the pedal switch to enable the WAITING FOR CONTACT... state. If the recording situation is free of interference and there is no risk for activation of the trigger before contact has been made, or if repeated contact need to be made, the Tasteprobe can be set to Auto Restart by changing the appropriate setting (Fig. 6) In the Auto Restart mode the system automatically returns to the WAITING FOR CONTACT.... after an adjustable delay. The delay can be entered in the Restart Delay box (Fig. 6). During the restart delay time the trigger cannot be activated, and the count-down time is indicated by a BEEP every second. A few seconds before the end of the restart delay time the beep rate is increased. The restart delay time allows replacement of the pipette without the risk of 'false triggering'. For repeated stimulations the delay time can be set to zero. TRIGGER OUTPUT The trigger output is a TTL signal, which changes either from Low (0) to High (5V) during the recording time if set to Positive, or from High (5V) to Low (0) if set to Negative (Fig. 6). The Syntech USB-IDAC responds by default to a Negative trigger; however, this can be inverted in the Digital I/O properties. Make sure that for any other signal acquisition system the trigger polarity is set according to the trigger requirements of that system. REFERENCES Hodgson, E.S., J.Y. Lettvin & K.D. Roeder, 1955. Physiology of a primary chemoreceptor unit. Science 122: 417-418 Marion-Poll, F. & J. van der Pers, 1996. Un-filtered recordings from insect taste sensilla. Entomologia Experimentalis et Applicata 80: 113-11 January 2003