Variable-length accelerometer features and electromyography to improve accuracy of fetal kicks detection during pregnancy using a single wearable device BHI, 2017 bloomlife.com Marco Altini, Elisa Rossetti, Michiel Rooijakkers, Julien Penders, Dorien Lanssens, Lars Grieten and Wilfried Gyselaers

Transcript of Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

Page 1: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

Variable-length accelerometer features and electromyography to improve accuracy of fetal kicks detection during pregnancy using a single wearable device

BHI, 2017 bloomlife.com

Marco Altini, Elisa Rossetti, Michiel Rooijakkers, Julien Penders, Dorien Lanssens, Lars Grieten and Wilfried Gyselaers

Page 2: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

2 |

FETAL MOVEMENT

BHI, 2017 bloomlife.com

Monitoring fetal movement during pregnancy is the mostpractical and widespread method to assess fetal wellbeing, one of the most important and complex tasks of modern obstetrics.!!As birth outcomes are strongly linked to the development of fetal conditions during pregnancy, several techniques have been developed to monitor fetal movement up to date!

Page 3: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

3 |

CURRENT CLINICAL PRACTICE

BHI, 2017 bloomlife.com

Ultrasound: relies on high frequency sound!waves being used to generate an image of the fetus and!can be used only for a limited amount of time due to!safety concerns. Require hospital stays or trained personnel.!!Continuous cardiotocography: require cumbersome infrastructure and hospital visits, also involving trained personnel to set up the device and process the produced information.!

Page 4: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

4 |

CURRENT CLINICAL PRACTICE

BHI, 2017 bloomlife.com

Ultrasound: relies on high frequency sound!waves being used to generate an image of the fetus and!can be used only for a limited amount of time due to!safety concerns. Require hospital stays or trained personnel.!!Continuous cardiotocography: require cumbersome infrastructure and hospital visits, also involving trained personnel to set up the device and process the produced information.!

-> Only sporadic checks in the hospital environment

Page 5: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

5 |

NEW PASSIVE SOLUTIONS

BHI, 2017 bloomlife.com

Accelerometers: Most studies to date involved one single accelerometer placed on the abdomen and reported rather low sensitivity and specificity.!!Other researchers added a reference accelerometerwith the rationale that by monitoring maternal movement!artifacts using an accelerometer placed outside of the abdominal area, fetal movement should be separable from!maternal movement and therefore detected more accurately. !

Page 6: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

6 |

NEW PASSIVE SOLUTIONS: REFERENCE ACCELEROMETER

BHI, 2017 bloomlife.com

0

100

200

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

sen

sor 1

labelnothingkick

Multi − data sample

0

50

100

150

200

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

sen

sor 3

0

10

20

30

40

50

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

sen

sor 6

Reference accelerometer on the back

Fetal movements do not appear on the reference accelerometer, while maternal movements typically do!

Page 7: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

7 |

NEW PASSIVE SOLUTIONS

BHI, 2017 bloomlife.com

-> Promising results, still limited practical applicability

Page 8: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

8 |

SINGLE SENSOR

BHI, 2017 bloomlife.com

Performance: consistently lower with respect to multiple sensors and reference accelerometers outside of the abdomen area. Higher false positives (harder to discriminate between maternal movements / artifacts and fetal movements)!! !

How do we reduce false positives?

Page 9: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

9 |

VARIABLE-LENGTH ACCELEROMETER FEATURES AND PHYSIOLOGICAL DATA

BHI, 2017 bloomlife.com

The proposed techniques aim at reducing false positives by!providing more contextual information related to maternal!movement while still using a single wearable device to cope!with the absence of a reference accelerometer or a more!obtrusive system.!!To account for different dynamics in maternal and fetal movement, we computed features over two time windows of 0.5 and 4 seconds. The rationale is that short fetal movements should be averaged out over longer time windows but captured over short ones, while maternal movements should appear over windows of both durations.!

Page 10: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

0.0

0.1

0.2

0.3

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

labelnothingkick

Single sensor, short time window

0.0

0.1

0.2

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

Single sensor, long time window

0

100

200

300

0 5 10 15Timestamp (minutes)

EHG

inte

nsity

Single sensor, EHG data

10 |

VARIABLE-LENGTH ACCELEROMETER FEATURES AND PHYSIOLOGICAL DATA

BHI, 2017 bloomlife.com

Page 11: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

0.0

0.1

0.2

0.3

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

labelnothingkick

Single sensor, short time window

0.0

0.1

0.2

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

Single sensor, long time window

0

100

200

300

0 5 10 15Timestamp (minutes)

EHG

inte

nsity

Single sensor, EHG data

11 |

VARIABLE-LENGTH ACCELEROMETER FEATURES AND PHYSIOLOGICAL DATA

BHI, 2017 bloomlife.com

Maternal movements appear on both traces!

Page 12: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

0.0

0.1

0.2

0.3

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

labelnothingkick

Single sensor, short time window

0.0

0.1

0.2

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

Single sensor, long time window

0

100

200

300

0 5 10 15Timestamp (minutes)

EHG

inte

nsity

Single sensor, EHG data

12 |

VARIABLE-LENGTH ACCELEROMETER FEATURES AND PHYSIOLOGICAL DATA

BHI, 2017 bloomlife.com

Fetal movements appear on the short window trace only!

Page 13: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

0.0

0.1

0.2

0.3

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

labelnothingkick

Single sensor, short time window

0.0

0.1

0.2

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

Single sensor, long time window

0

100

200

300

0 5 10 15Timestamp (minutes)

EHG

inte

nsity

Single sensor, EHG data

0.0

0.1

0.2

0.3

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

labelnothingkick

Single sensor, short time window

0.0

0.1

0.2

0 5 10 15Timestamp (minutes)

Mot

ion

inte

nsity

Single sensor, long time window

0

100

200

300

0 5 10 15Timestamp (minutes)

EHG

inte

nsity

Single sensor, EHG data

13 |

VARIABLE-LENGTH ACCELEROMETER FEATURES AND PHYSIOLOGICAL DATA

BHI, 2017 bloomlife.com

Maternal movements are more likely to trigger EMG activity!

Page 14: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

14 |

STUDY DESIGN

BHI, 2017 bloomlife.com

Twenty-two recordings of about 60 minutes duration were collected from 22 pregnant women at different gestational ages during pregnancy, all from week 30 onwards. !!Fetal movements ranged between 0 for inactive babies to 315 for hiccups cases. !!Measurements were performed using two devices. A research version of the Bloomlife wearable device, configured to acquire two channels EMG at 4096 Hz and triaxial accelerometer data at 128 Hz and the TMSi system including 6 accelerometers, five placed on the abdomen and one on the back.!

Page 15: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

15 |

FEATURES, CLASSIFIER AND CLASS-IMBALANCE

BHI, 2017 bloomlife.com

Features: low-complexity time domain features (mean, standard!deviation, interquartile range, correlation between axis, sum,!min, max and magnitude).!!Classification: Random forests. We set the number of features to select at each iteration to the square root of the total number of features.!!Class imbalance: small number of kicks with respect to the total available data. The optimal ratio between reference class (kicks) and majority class (non-kicks) was determined by cross-validating and optimizing for F-score. Our optimal balance included all data from the minority class and one fifth of the majority class data !

Page 16: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

16 |

COMPARISONS AND CROSS-VALIDATION

BHI, 2017 bloomlife.com

We compared four feature sets associated to the two systems used in this study in order to highlight the impact of the novel methods proposed to improve accuracy of a single wearable device:!1.  TMSi (6 accelerometer system) and variable-length features.!2.  Bloomlife (single wearable sensor) and features computed

over a short time window only Bloomlife and features computed over both short and long time windows. !

3.  Bloomlife and features computed over both short and long time windows plus EMG features. !

All models were derived and validated using leave one!participant out cross-validation!

Page 17: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

17 |

RESULTS: SENSITIVITY

BHI, 2017 bloomlife.com

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

Sensitivity

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

PPV

Page 18: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

18 |

RESULTS: SENSITIVITY

BHI, 2017 bloomlife.com

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

Sensitivity

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

PPV

Page 19: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

19 |

RESULTS: SENSITIVITY

BHI, 2017 bloomlife.com

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

Sensitivity

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

PPV

Sensitivity does not change much by introducing variable-length and EMG features as the aim of these features is to reduce false positives. !

Page 20: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

20 |

RESULTS: SENSITIVITY

BHI, 2017 bloomlife.com

Sensitivity does not change much by introducing variable-length and EMG features as the aim of these features is to reduce false positives. !

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

Sensitivity

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

PPV

Page 21: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

21 |

RESULTS: POSITIVE PREDICTIVE VALUE

BHI, 2017 bloomlife.com

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

Sensitivity

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

PPV

Page 22: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

22 |

RESULTS: POSITIVE PREDICTIVE VALUE

BHI, 2017 bloomlife.com

On the other hand, PPV was 0.75 for the 6 sensors system and increased between 0.65 to 0.75 when including variable-length and EMG features

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

Sensitivity

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

PPV

Page 23: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

23 |

RESULTS: POSITIVE PREDICTIVE VALUE

BHI, 2017 bloomlife.com

On the other hand, PPV was 0.75 for the 6 sensors system and increased between 0.65 to 0.75 when including variable-length and EMG features

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

Sensitivity

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

PPV

Page 24: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

24 |

RESULTS: POSITIVE PREDICTIVE VALUE

BHI, 2017 bloomlife.com

On the other hand, PPV was 0.75 for the 6 sensors system and increased between 0.65 to 0.75 when including variable-length and EMG features

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

Sensitivity

0

25

50

75

Multi SL Single S Single SLSingle SLEModel

Perc

enta

ge (%

)

PPV

Page 25: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

25 |

RESULTS: TOTAL NUMBER OF KICKS PER RECORDING

BHI, 2017 bloomlife.com

●●●●●●● ●

●●

●●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Multi SL

●●●●●●●●

●●

●●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single S

●●●●●●●●

●●●

●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single SL

●●● ●●●●●

●●●

●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single SLE

Page 26: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

26 |

RESULTS: TOTAL NUMBER OF KICKS PER RECORDING

BHI, 2017 bloomlife.com

●●●●●●● ●

●●

●●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Multi SL

●●●●●●●●

●●

●●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single S

●●●●●●●●

●●●

●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single SL

●●● ●●●●●

●●●

●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single SLE

Page 27: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

27 |

RESULTS: TOTAL NUMBER OF KICKS PER RECORDING

BHI, 2017 bloomlife.com

●●●●●●● ●

●●

●●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Multi SL

●●●●●●●●

●●

●●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single S

●●●●●●●●

●●●

●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single SL

●●● ●●●●●

●●●

●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single SLE

Page 28: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

28 |

RESULTS: TOTAL NUMBER OF KICKS PER RECORDING

BHI, 2017 bloomlife.com

Less overdetections at the recording level as well

●●●●●●● ●

●●

●●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Multi SL

●●●●●●●●

●●

●●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single S

●●●●●●●●

●●●

●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single SL

●●● ●●●●●

●●●

●●

0

100

200

300

0 100 200 300Detected kicks

Actu

al k

icks

Actual vs Detected kicks, Single SLE

Page 29: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

29 |

CONCLUSIONS

BHI, 2017 bloomlife.com

We proposed a method to improve the accuracy of fetal kicks detection during pregnancy using a single wearable device placed on the abdomen. !!Including variable-length accelerometer features, short fetal movement is averaged out over longer time windows but captured over short ones, while maternal movements of greater intensity appear over windows of both durations. As a result, a single wearable device can be used to better discriminate fetal and maternal movement without the need for a reference accelerometer (11% improvement in PPV).!

Page 30: Talk at the International Conference on Biomedical and Health Informatics (BHI 2017)

Thank you

Marco Altini, PhD Head of Data Science | [email protected]