SYNTHESIS AND CHARACTERIZATION OF SOME...
Transcript of SYNTHESIS AND CHARACTERIZATION OF SOME...
Chapter-3
A validated stability-indicating normal phase LC method for Clopidogrel bisulfate and its impurities in bulk drug and pharmaceutical dosage form
___________________________________________
3.1 Introduction of Clopidogrel bisulfate and survey of analytical
methods
Clopidogrel bisulfate is a thienopyridine class inhibitor of P2Y12 ADP
platelet receptors. Chemically it is methyl (+)-(S)-α-(2-chlorophenyl)-6,7-
dihydrothieno[3,2-c]pyridine-5(4H)-acetate sulfate. Clopidogrel bisulfate
is a white to off-white powder. It is practically insoluble in water at
neutral pH but freely soluble at pH 1. It also dissolves freely in methanol,
dissolves sparingly in methylene chloride, and is practically insoluble in
ethyl ether. The empirical formula of Clopidogrel bisulfate is
C16H16ClNO2S•H2SO4. The molecular weight of Clopidogrel bisulfate is
419.9.
Fig: 3.1.F1: Chemical structure of Clopidogrel bisulfate
methyl (+)-(S)-α-(2-chlorophenyl)-6,7-dihydrothieno[3,2-c]pyridine-5(4H)-
acetate sulfate
Molecular formula: C16H16ClNO2S•H2SO4, Molecular weight : 419.9
Clopidogrel is a potent anti-platelet and anti-thrombotic drug. It is a
dihydro thieno pyridine derivative pro-drug which is inactive in vitro. In
vivo, it selectively and irreversibly inhibits the binding of adenosine
diphosphate (ADP) to its platelet receptors [1]. Although the majority of
the drug is hydrolyzed by esterase to an inactive carboxylic acid
metabolite, the full anti-aggregating activity of the drug is achieved by
biotransformation to 2-oxo-Clopidogrel by cytochrome P450-1A. This
intermediate metabolite is hydrolyzed and generates an active form
which reacts as thiol reagent with the ADP receptor on platelets thus
preventing the binding of ADP. Clopidogrel has an absolute S-
configuration at carbon 7; the corresponding R-enantiomer is totally
devoid from anti-aggregating activity [2]. Both the carboxylic acid
hydrolysis product and the R-enantiomer are stated by the United States
Pharmacopeia (USP) as impurities A and C in the monograph, besides
impurity B which is a racemic mixture of regioisomer of Clopidogrel
(Impurity B1 and B2) [3]. USP method uses a glycoprotein based
ovomucoid column. The resolution of impurity B1 and B2 from
Clopidogrel is going bad in less number of injections in the USP method
(Fig: 3.1.F2). Apart from the monograph listed impurities, two more
process related impurities D and E are included in this study.
Few other methods have been reported for the estimation of
Clopidogrel by spectrophotometric [4, 5], by LC [6, 7], by CZE [8] and by
TLC [9]. A bio-analytical method for the determination of Clopidogrel in
plasma using LC–MS/MS was reported [10]. Clopidogrel was determined
by LC in the presence of its impurities using a chiral column and its
impurities were determined at low levels [11]. The carboxylic acid
metabolite (impurity A) was determined in human plasma by GC–MS
[12], by LC with UV detection [13] and by LC–MS [14]. The major
objective of the present work is to develop a single method for the
separation of related substance of Clopidogrel bisulfate as well as the
enantiomeric impurities with a short run time in drug substance and
drug product. Further the use of carbohydrate based chiral HPLC
column ensures the column life unlike glycoprotein based columns. So
far to our knowledge there was no method reported using normal phase
chromatography.
Fig: 3.1.F2: Typical chromatogram from USP Method
Hence a sensitive reproducible simple isocratic stability indicating
normal phase HPLC method was developed for quantitative
determination Clopidogrel bisulfate and its five impurities in
pharmaceutical dosage forms. The developed method was validated as
per ICH guidelines.
3.2. Development of stability-indicating normal phase HPLC method
for Clopidogrel bisulfate and its impurities
3.2.1 Materials
Tablets and standards of Clopidogrel bisulfate and its five impurities
namely impurity A, impurity B (racemic mixture of B1 and B2), impurity
C, impurity D and impurity E were supplied by Dr. Reddy’s laboratories
limited, Hyderabad, India. The HPLC grade n-hexane, ethanol and
analytical grade diethyl amine were purchased from Merck, Darmstadt,
Germany. High purity water was prepared by using Millipore MilliQ Plus
water purification system (Millipore, Milford, MA, USA).
3.2.2 Equipment
The Waters HPLC system (Waters, Milford, USA) used consists of a
pump, auto sampler and a PDA detector. The output signal was
monitored and processed using empower-2 software. Cintex digital water
bath was used for hydrolysis studies. Photo stability studies were carried
out in a photo stability chamber (Sanyo, Leicestershire, UK). Thermal
stability studies were performed in a dry air oven (Cintex, Mumbai, India)
3.2.3. Chromatographic conditions
The method was developed using Chiralcel OJ-H (250 mm x 4.6 mm,
5 μm) column. n- Hexane, Ethanol and diethyl amine in 95:5:0.05 v/v/v
ratio was used as mobile phase. The flow rate of the mobile phase was
1.0 ml/min. The column temperature was maintained at 25 °C and the
wavelength was monitored at 240 nm. The injection volume was 10 µl.
Fig: 3.2.F1: Chemical structures of impurities of Clopidogrel
Impurity-A (USP Related compound A)
Molecular formula: C15H14Cl NO2S
Molecular weight : 307.80
Impurity-D
Molecular formula :C7H9NS
Molecular weight : 139.22
Impurity-B (USP Related compound B)
Molecular formula: C16H16Cl NO2S
Molecular weight: 321.82
Impurity-E
Molecular formula:
C15H16ClNO2S
Molecular weight : 309.81
Impurity-C (USP Related compound C)
Molecular formula : C16H16ClNO2S Molecular weight : 321.82
3.2.4 Sample preparation
Stock solution of Clopidogrel (2.0 mg mL-1) was prepared by dissolving
appropriate amount of Clopidogrel bisulfate in ethanol. Working
solutions of 1000 and 100 µg mL-1in mobile phase were prepared from
the above stock solution for the related substance determination and
assay determination, respectively. A stock solution of impurity (mixture
of impurities A, B, C, D and E) at 0.5 mg mL-1 was also prepared in
ethanol.
3.2.5 Preparation of Tablets Sample Solution:
Twenty Clopidogrel bisulfate 75 mg tablets were weighed, transferred
to a clean and dry mortar and ground into a fine powder. Tablet powder
equivalent to 200 mg drug was then transferred to a 100 ml volumetric
flask, 70 ml of ethanol was added, and the flask was attached to a rotary
shaker for 10 min to disperse the material completely. The mixture was
then sonicated for 10 min and diluted to volume with ethanol to give a
solution containing 2000 μg mL-1. This solution was centrifuged at
3,000 RPM for 5 min and supernatant was diluted using mobile phase to
give test solutions containing 1000 μg mL-1 and 100 μg/ml. These
solutions were filtered through a 0.45 μm pore size Nylon 66 membrane
filter
3.2.6 Generation of stress samples
One lot of Clopidogrel bisulfate tablets was selected for stress testing.
From the ICH Stability guideline: “stress testing is likely to be carried out
on a single batch of material. Different kinds of stress conditions (i.e.,
heat, humidity, acid, base, oxidative and light) were employed on one lot
of Clopidogrel bisulfate tablets based on the guidance available from ICH
Stability Guideline (Q1AR2). The details of the stress conditions applied
were as follows:
a) Acid hydrolysis: Drug solution in 0.5 N HCl was exposed at at 80°C
for 24 h. b) Base hydrolysis: Drug solution in 0.5 N NaOH was exposed
at 80°C for 24 h. c) Oxidative stress: Drug solution in 3% H2O2 v/v was
exposed at room temperature for 24 h. d) Water hydrolysis: Drug
solution in water was exposed at 80°C for 24 h. e) Thermal stress:
Tablets were subjected to dry heat at 60°C for 10 days. f) Photolytic
degradation: The photo degradation was carried out by exposing the
Clopidogrel bisulfate tablets in solid state to light providing an overall
illumination of not less than 1.2 million lux hours and an integrated near
ultraviolet energy of not less than 200 W h/m2, which took about 10
days period in photo stability chamber .
3.3 Method development and optimization of chromatographic conditions
Forced degradation studies were performed to develop a stability
indicating HPLC method for the quantitative determination and purity
evaluation of Clopidogrel bisulfate. Stressed samples obtained during
forced degradation studies including the samples of Impurity-A,
Impurity-B, Impurity-C, Impurity-D and Impurity-E were utilized in the
HPLC method development.
3.3.1 Selection of wavelength
Impurities and Clopidogrel bisulfate solutions were prepared in
diluent at a concentration of 100 µg/ml and scanned in UV-Visible
spectro photometer; both the impurities and Clopidogrel bisulfate was
having UV maxima at around 240 nm (Fig: 3.3.F1). Hence detection at
240 nm was selected for method development purpose.
Fig: 3.3.F1: Typical UV spectrums of Clopidogrel and its impurities
A
b
so
r
b
an
c
e
Impurity-A
Impurity-B
Clopidogrel
Impurity-D
3.3.2 Column Selection
The objective of the chromatographic separation was to separation
impurity-A, impurity-D and impurity-E along with separation of
impurity-B (impurity-B1), enatiomer of impurity-B (impurity-B2), and
enatiomer Clopidogrel (impurity-C) from Clopidogrel. Since the method
involves the separation of chiral and non chiral impurities, initially trials
were conducted using reverse phase chromatography using different C18
columns, but separation was poor for chiral impurities.
Initially trials were conducted using different chiral stationary phases
like carbohydrates (cellulose and amylose) based columns. Trials were
conducted using 250 x 4.6 mm, 5 µm CHIRALPAK®AD column, 250 x
4.6 mm, 5 µm CHIRALCEL® OD column and 250 x 4.6 mm, 5 µm CHIRALCEL® OJ-H columns. n-Hexane and Ethanol in the ratio 95:5
(v/v) was used as mobile phase at flow rate 1.0 ml/min. For
CHIRALPAK®AD (amylose based) column, the separation of impurities namely impurity-A, impurity-B (B1 and B2), impurity-C and impurity-D
was not adequate and impurity-E was co- eluted with Clopidogrel. For
CHIRALCEL® OD and CHIRALCEL® OJ-H (cellolose based) columns, the separation was satisfactory for impurities namely impurity-A, impurity-B
(B1 and B2), impurity-C and impurity-D, but impurity-E was eluted very
closely with Clopidogrel and retention of Clopidogrel was about 21 min. But separation of impurity-E from Clopidogrel was better in
CHIRALCEL® OJ-H column. Hence CHIRALCEL® OJ-H was chosen for
further method development.
Wavelength
(nm)
Impurity-C Impurity-E
Table 3.3.T1: Experimental results on different stationary phases
Column Parameter Imp-
A Imp-
D Imp-B1
Imp-C
Imp-B2
Imp-E
Clopidogrel
250 x 4.6mm,
CHIRALPAK®AD, 5 µm
Retention time
6.8 7.5 8.9 9.6 23.9 - 29.9
USP
Tailing 1.2 1.4 1.1 0.9 0.8 - 1.1
250 x 4.6mm,
CHIRALCEL® OD, 5 µm
Retention time
4.8 8.5 10.3 12.4 15.6 20.4 21.2
USP
Tailing 1.1 1.2 1.1 1.2 0.9 - 1.2
250x4.6mm,
CHIRALCEL® OJ-H
Retention time
4.9 8.7 10.8 12.9 16.3 21.2 22.3
USP
Tailing 1.3 1.2 1.1 1.2 0.9 1.2 1.3
CHIRALCEL® OD = Tris(3,5-dimethylphenylcarbamate) of cellulose
CHIRALPAK® AD= Tris(3,5-dimethylphenylcarbamate) of amylase
CHIRALCEL® OJ-H= tris(4-methyl benzoate) of cellulose
3.3.3 Effect of additive
To improve the resolution between the impurities, basic additives
triethyl ammine and diethyl amine (0.05%) were added to above selected
mobile phase (n-Hexane, Ethanol in 95:5 (v/v)). When diethyl amine was
used in mobile phase, the resolution between impurities was improved
when compared to addition of triethyl amine to the mobile phase. Hence
n- Hexane, Ethanol and diethyl amine in 90:10:0.05, v/v/v; was chosen
as mobile phase with 1.0 mL min-1 flow rate
3.3.4 Effect of Column Temperature
When the column temperature was at 25°C, the symmetry of
Clopidogrel peak was 1.3. When Temperature increased to 40°C, there
was no improvement in symmetry of Clopidogrel peak. Hence column
oven temperature was finalized as 25°C
3.3.5 Effect of Organic solvent
When isopropyl alcohol was used in place of ethanol (n- Hexane,
isopropyl alcohol and diethyl amine in the ratio 95:5:0.05 (v/v/v),
Clopidogrel peak was strongly retained (about 30 min) and also impurity-
E was eluted as a broad peak. When ethanol ratio was increased (polarity
of mobile phase was increased) to 10% from 5%, the resolution between
impurity-E and Clopidogrel was poor. Hence ethanol ratio in mobile
phase was optimized at 5% (n- Hexane, ethanol and diethyl amine in
95:5:0.05, v/v/v).
3.3.6 Optimized chromatographic conditions for the determination
of related substances and assay of Clopidogrel
Column : Chiralcel OJ-H (250 x 4.6) mm with 5 µm
particle size
Elution : Isocratic
Mobile phase : Hexane, Ethanol and diethyl amine in
95:5:0.05 (v/v/v) ratio
Flow rate : 1.0 ml/min
Column temperature : 25°C
Wavelength of detection : 240 nm
Injection volume : 10 l
Run time : 35 min
Diluent : Mobile phase
Retention time : Clopidogrel about 22 min
Relative Retention Time : Impurity-A - about 0.22
Impurity-B1 - about 0.50
Impurity-B2 - about 0.75
Impurity-C - about 0.59
Impurity-D- about 0.40
Impurity-E- about 0.90
The interference from excipients Mannitol, Colloidal Silicon Dioxide, Croscarmellose Sodium, Hydroxy Propyl Cellulose, Hydroxy Propyl
Methyl Cellulose, Hydrogenated Castor Oil Powder, microcrystalline
cellulose, Opadry Pink was also checked by injecting sample solutions of
excipients. There was no interference from excipients found at retention times of impurities (impurity-A, impurity-B1& B2, impurity-C, impurity-
D and impurity-E) and Clopidogrel peak.
3.4 Degradation behavior
HPLC studies on Clopidogrel bisulfate under different stress
conditions suggested the following degradation behavior.
3.4.1 Degradation in acidic solution
Clopidogrel bisulfate tablets powder in solution was exposed to 0.5 N
HCl at 80°C for 24 h, degradation was observed leading to the formation
Impurity-A (Fig: 3.4.F1 (a) to Fig: 3.4.F1 (b)).
Fig: 3.4.F1 (a): Typical HPLC chromatogram of Acid hydrolysis
Fig: 3.4.F1 (b): Peak purity plot of Clopidogrel under Acid hydrolysis
Purity Angle Purity
Threshold Purity Flag Peak Purity
0.274 0.353 No Pass
3.4.2 Degradation in Basic solution
Clopidogrel bisulfate tablets powder in solution was exposed to 0.5 N
NaOH at 80°C for 24 h, degradation was observed leading to the formation Impurity-A (Fig: 3.4.F2 (a) to Fig: 3.4.F2 (b)).
Fig: 3.4.F2 (a): Typical HPLC chromatogram of alkali hydrolysis
Fig: 3.4.F2 (b): Peak purity plot of Clopidogrel under alkali
hydrolysis
Purity Angle Purity
Threshold Purity Flag Peak Purity
0.341 0.355 No Pass
3.4.3 Oxidative conditions
Clopidogrel bisulfate tablets powder in solution was exposed to 3% hydrogen peroxide at room temperature for 24 h. The drug gradually
undergone degradation with time in 3 % hydrogen peroxide and mild
degradation was observed (0.6%) (Fig: 3.4.F3 (a) to Fig: 3.4.F3 (b)).
Fig: 3.4.F3 (a): Typical HPLC chromatogram of oxidative degradation
Fig: 4.4.F3 (b): Peak purity plot of Clopidogrel under oxidative
degradation
Purity Angle Purity
Threshold Purity Flag Peak purity
0.133 1.442 No Pass
3.4.4 Degradation in Neutral (Water) solution
No major degradation products were observed after 24 h at room
temperature. The drug was also stable in water on heating at 80°C for
24h. The drug was stable to water hydrolysis (Fig: 3.4.F4 (a) to Fig: 3.4.F4 (b)).
Fig: 3.4.F4 (a): Typical HPLC chromatogram of Water hydrolysis
Fig: 3.4.F4 (b): Peak purity plot of Clopidogrel under Water
hydrolysis
Purity Angle Purity
Threshold Purity Flag Peak purity
0.513 0.591 No Pass
3.4.5 Photolytic conditions
When the drug powder was exposed to light for an overall illumination of 1.2 million lux hours and an integrated near ultraviolet
energy of 200-watt hours/square meter (w/mhr) (in photo stability
chamber), no degradation of the drug was observed. The drug was stable to photolysis (Fig: 3.4.F5 (a) to Fig: 3.4.F5 (b)).
Fig: 3.4.F5 (a): Typical HPLC chromatogram of Photo degradation
Fig: 3.4.F5 (b): Peak purity plot of Clopidogrel under Photo
degradation
Purity Angle Purity
Threshold Purity Flag Peak purity
0.460 0.597 No Pass
3.4.6 Thermal Degradation:
The drug was stable to the effect of temperature. When the drug
powder exposed to dry heat at 60°C for 10 days, mild decomposition of the drug was observed (Fig: 3.4.F6 (a) to Fig: 3.4.F6 (b)).
Fig: 3.4.F6 (a): Typical HPLC chromatogram of thermal degradation
Fig: 3.4.F6 (b): Peak purity plot of Clopidogrel under Thermal
degradation
Peak purity test performed for the Clopidogrel peak using photodiode
array (PDA) detector data confirmed the peak purity of the peak for all
the forced degradation samples. % Assay of all stressed samples were
calculated using qualified standard of Clopidogrel. Considering the
purities from the respective chromatograms of forced degradation
samples, mass balance (%assay + % degradents + % impurities) was
calculated for all stress samples. The mass balance of stressed samples
was close to 99%. This clearly demonstrates that the developed normal
phase HPLC method was specific for Clopidogrel in presence of its
impurities (Impurity-A, Impurity-B, Impurity-C, Impurity-D and
Impurity-E) and degradation products.
3.5 Analytical method Validation and its results
Purity Angle Purity
Threshold Purity Flag Peal purity
0.038 1.123 No Pass
The developed and optimized HPLC method was taken up for method
validation. The analytical method validation was carried out in
accordance with ICH guidelines [15, 16].
3.5.1 System Suitability Test
Clopidogrel bisulfate standard, impurity-A, impurity-B, impurity-C,
Impurity-D and impurity-E were spiked at 0.20% level with respect to the
concentration of Clopidogrel in diluent and injected for five times into
HPLC system. Resolution between impurities and Clopidogrel, tailing
factor, theoretical plates for impurities and Clopidogrel, RSD% for the
areas of impurities and Clopidogrel was calculated. Good resolution was
obtained between impurities and Clopidogrel (Fig: 3.5.F1). System
suitability results were tabulated (Table 3.5.T1).
Fig: 3.5.F1: Typical chromatogram of system suitability
Table 3.5.T1: System Suitability results
Compound
(n=5)
Resolution (Rs) USP
Tailing factor (T)
% RSD for
peak area
Impurity-A
- 1.2 0.5
Impurity-D
11.8 1.2 0.2
Impurity-B1
4.6 1.2 0.9
Impurity-C
4.3 1.2 0.5
Impurity-B2
6.0 1.1 0.3
Impurity-E
5.0 1.0 0.5
Clopidogrel
2.4 1.3 0.4
n = Number of determinations
3.5.2 Precision
Assay method precision (repeatability) study was evaluated by carrying
out six independent assays of Clopidogrel bisulfate test sample against
qualified reference standard and %RSD of six consecutive assays was
0.81% (Table 3.5.T2). Results showed insignificant variation in measured
response which demonstrated that the method was repeatable with RSDs
below 0.9%.
Table 3.5.T2: Precision results of the assay method
Preparation %Assay
1 99.2
2 100.5
3 99.9
4 99.4
5 100
6 98.2
Mean 99.5
SD 0.80
% RSD 0.81
95% Confidence interval of
mean 98.9 to 100.1
The precision of the related substance method was evaluated by
injecting six individual preparations of Clopidogrel (1.0 mg mL-1) spiked
with 0.20% of Impurity-A, Impurity-B, Impurity-C, Impurity-D and
impurity-E with respect to Clopidogrel analyte concentration. The % RSD for area of Impurity-A, Impurity-B, Impurity-C, Impurity-D and impurity-
E for six consecutive determinations was below 3.0% (Table 3.5.T3).
Results showed insignificant variation in measured response which demonstrated that the related substances method was repeatable with
RSDs below 3.0%.
Table 3.5.T3: Precision results of the RS method
Prepa
ration Impurity-A Impurity-B* Impurity-C Impurity-D Impurity-E
1 225053 48870 260280 188212 128917
2 221536 51509 266775 186777 129955
3 219594 48852 264130 187569 131354
4 218822 49369 269611 181313 126631
5 214859 49500 267868 188662 130236
6 218808 50550 258616 185682 134204
Mean 219778.6 49775 264546.7 186369.2 130216
SD 3376.4 1050.9 4363.3 2694.2 2522.1
%RSD 1.53 2.11 1.64 1.44 1.93
* Impurity B1+ B2
Intermediate precision for assay and related substances method was
performed by carrying out six independent assays of Clopidogrel bisulfate
tablets test sample against qualified reference standard and RSD of six
consecutive assays was calculated. Related substances method was
performed by injecting six individual preparations of Clopidogrel (1.0 mg
mL-1) spiked with 0.20% of Impurity-A, Impurity-B, Impurity-C,
Impurity-D and impurity-E with respect to Clopidogrel analyte
concentration over different days, different instruments and with
different analysts (Table 3.5.T4).
Table 3.5.T4: Results of Intermediate precision
S.No
Parameter
Variation
%RSD
for Assay
%RSD
for Related
Substances
1
Repeatability
(a) Analyst-1
(b) Waters 2695 Alliance system.
(c) Day-1
0.81%
< 3.0%
2
Intermediate
precision
(a) Analyst-2
(b) Agilent 1100 series
VWD system.
(c) Day-2
0.75%
< 3.0%
3.5.3 Limit of quantification (LOQ) and limit of detection (LOD)
LOQ and LOD were established for Impurity-A, Impurity-B, Impurity-
C, Impurity-D and impurity-E based on signal to noise ratio method.
3.5.3.1 Limit of quantification (LOQ)
LOQ was established for Clopidogrel and for impurity-A, impurity-B,
impurity-C, impurity-D and impurity-E based on signal to noise ratio method (Table 3.5.T5).
Table 3.5.T5: LOQ values of Clopidogrel and its impurities
S.No Impurity
name Concentration
Signal to noise ratio
1 Impurity-A 0.410 µg mL-1 11.2
2 Impurity-B 0.320 µg mL-1 9.1
3 Impurity-C 0.310µg mL-1 9.5
4 Impurity-D 0.360 µg mL-1 10.8
5 Impurity-E 0.360 µg mL-1 9.8
3.5.3.2 Limit of detection (LOD) The detection limits were established by S/N methodology and
established concentration values are reported in Table 3.5.T6.
Table 3.5.T6: LOD values of Clopidogrel and its impurities
S.No Impurity name Concentration Signal to noise
ratio
1 Impurity-A 0.160 µg mL-1 3.3
2 Impurity-B 0.140 µg mL-1 2.8
3 Impurity-C 0. 090µg mL-1 2.6
4 Impurity-D 0.110 µg mL-1 3.1
Impurity-E 0.130 µg mL-1 2.9
3.5.4 Precision at Limit of quantification
The precision of the related substance method was also checked by injecting six individual preparations of Impurity-A, Impurity-B, Impurity-
C, Impurity-D and impurity-E at their LOQ level with respect to
Clopidogrel analyte concentration. Results showed insignificant variation
in measured responses which demonstrate that the method was
repeatable at LOQ level with RSD below 4.0% (Table 3.5.T7).
Table 3.5.T7: Precision results of the RS method at LOQ level.
Preparati
on
Impurity-
A
Impurity-
B
Impurity-
C
Impurity-
D
Impurity-
E
1 29536 15961 20010 37004 27932
2 29761 15652 19604 37431 26795
3 30287 15342 20910 36660 28018
4 29974 15857 19835 36287 27583
5 29174 14407 20859 36627 27367
6 29959 15088 20322 37382 27814
Average 29782 15385 20257 36899 27585
Std Dev 388.29 578.34 540.10 454.53 454.31
%RSD 1.3 3.7 2.6 1.2 1.6
3.5.5 Accuracy at LOQ level
The recovery studies for Impurity-A, Impurity-B, Impurity-C, Impurity-D and impurity-E were carried out in triplicate at LOQ level of
the Clopidogrel target analyte concentration (1000 µg mL-1). The
percentage recovery of Impurity-A, Impurity-B, Impurity-C, Impurity-D and impurity-E was calculated (Table 3.5.T8). The method showed
consistent and high absolute recoveries at LOQ level with mean absolute
recovery ranging from 98.1 % to 99.3%. The obtained absolute recoveries were normally distributed around the mean with uniform RSD values.
The method was found to be accurate with low % bias (< 2.0).
Table 3.5.T8: Recovery at LOQ level
S.No
Impurity name
Mean recovery (%)
(n = 3 )
%RSD
1 Impurity-A 98.1 0.31
2 Impurity-B 98.5 0.22
3 Impurity-C 98.9 0.33
4 Impurity-D 99.1 0.35
5 Impurity-E 99.3 0.18
3.5.6 Linearity
3.5.6.1 Linearity of the assay method
The linearity of the assay method was established by injecting test
sample at 50%, 75%, 100%, 125% and 150% of Clopidogrel bisulfate
assay concentration (i.e.100 µg mL-1). Each solution was injected in to
HPLC system (Table 3.5.T9). Calibration curve obtained by analysis
between peak area and the concentration showed (Fig: 3.5.F2) linear
relationship with a regression coefficient 0.9999. The best fit linear
equation obtained was y = 10651 x con + 93892.
Analysis of residuals indicated that the residuals were normally (Table
3.5.T10) distributed around the mean with uniform variance across all
concentrations suggesting the homoscedastic nature of data.
Table 3.5.T9: Linearity results of the assay method
Concentration (%) Peak area
50 5433303
75 8044168
100 10819223
125 13497932
150 16050284
Correlation
Coefficient(r) 0.99991
Slope 106751
Intercept 93892
Standard error 644278.0
Fig: 3.5.F2: Linearity plot for Assay Method
Table 3.5.T10: Residual summary of Linearity results of the assay
method
%conc
Mean area
Response achieved
Response calculated
thru Trend
line equation
Residual
(Response
practical -Response
theoretical)
Residual
square
50 5433303 5431436.80 -1866 3482702.4
75 8044168 8100209.40 56041 3140638514.0
100 10819223 10768982.00 -50241 2524158081.0
125 13497932 13437754.60 -60177 3621319470.8
150 16050284 16106527.20 56243 3163297546.2
Residual sum of squares
12452896314.4
Trend line equation y = 10675x + 93892
3.5.6.2 Linearity of the related substances method
Linearity experiments were carried out by preparing the Clopidogrel
bisulfate sample solutions containing Impurity-A, Impurity-B, Impurity-C, Impurity-D and impurity-E from LOQ to 150% (i.e. LOQ, , 50%, 75%,
100% and 150%) with respect to specification limit 0.20% (Table
3.5.T11).
Calibration curve was drawn by plotting peak area of the impurity on
the Y-axis and concentration on the X-axis (Fig: 3.5.F3 to Fig: 3.5.F7)
which showed linear relation ship with a regression coefficient of greater
than 0.998 for all impurities. Analysis of residuals showed (Table 3.5.T12 to Table 3.5.T16) that the residuals were normally distributed around the
mean with uniform variance for all concentrations.
Table 3.5.T11: Linearity results of the RS method
S.No
% of impurity
w.r.t
1000 µg/ml
Impurity-A (Peak area)
Impurity-B (Peak area)
Impurity-
C (Peak
area)
Impurity-
D (Peak
area)
Impurity-E (Peak area)
1 LOQ 25477 15772 18786 37490 27451
2 0.10 64056 29005 46188 85817 65263
3 0.15 93814 43700 75471 134565 103809
4 0.20 134091 57814 100288 177444 135244
6 0.30 198079 85053 148587 264424 200325
Correlation
coefficient (r) 0.9992 0.9999 0.9991 0.9995 0.9989
intercept -7213 1694 -5254 -4910 -3773
Slope 689479 278588 519493 905366 688696
Fig: 3.5.F3: Linearity plot for impurity-A
Fig: 3.5.F4: Linearity plot for impurity-B
Fig: 3.5.F5: Linearity plot for impurity-C
Fig: 3.5.F6: Linearity plot for impurity-D
Fig: 3.5.F7: Linearity plot for impurity-E
Table 3.5.T12: Residual summary of Linearity results of impurity-A
% of
impurity
w.r.t 1000
µg/ml
Mean area
Response achieved
Response
calculated
thru Trend line
equation
Residual
(Response
practical -Response
theoretical)
Residual
square
0.05 25477 27260.69 1783.7 3181547.1
0.10 64056 61734.65 -2321.4 5388672.1
0.15 93814 96208.61 2394.6 5734148.0
0.20 134091 130682.57 -3408.4 11617411.6
0.30 198079 199630.49 1551.5 2407110.3
Residual sum of squares 28328889.2
Trend line equation y = 689479.189x - 7213.270
Table 3.5.T13: Residual summary of Linearity results of impurity-B
% of
impurity w.r.t
1000
µg/ml
Mean area Response
achieved
Response
calculated thru Trend
line
equation
Residual
(Response practical -
Response
theoretical)
Residual
square
0.05 15772 15624.11 -147.9 21872.0
0.10 29005 29553.51 548.5 300867.1
0.15 43700 43482.92 -217.1 47124.2
0.20 57814 57412.32 -401.7 161343.3
0.30 85053 85271.14 218.1 47582.9
Residual sum of squares
578789.6
Trend line equation y = 278588.108x + 1694.703
Table 3.5.T14: Residual summary of Linearity results of impurity-C
% of
impurity
w.r.t 1000
µg/ml
Mean area
Response achieved
Response
calculated
thru Trend line
equation
Residual
(Response
practical -Response
theoretical)
Residual
square
0.05 18786 20719.74 1933.7 3739362.9
0.10 46188 46694.41 506.4 256446.4
0.15 75471 72669.07 -2801.9 7850825.4
0.20 100288 98643.73 -1644.3 2703624.7
0.30 148587 150593.05 2006.1 4024252.9
Residual sum of squares 18574512.3
Trend line equation y = 519493.243x - 5254.919
Table 3.5.T15: Residual summary of Linearity results of impurity-D
% of
impurity
w.r.t 1000
µg/ml
Mean area
Response achieved
Response
calculated
thru Trend line
equation
Residual
(Response
practical -Response
theoretical)
Residual
square
0.05 37490 40357.72 2867.7 8223796.3
0.10 85817 85626.03 -191.0 36470.7
0.15 134565 130894.34 -3670.7 13473760.7
0.20 177444 176162.65 -1281.4 1641861.3
0.30 264424 266699.27 2275.3 5176854.8
Residual sum of squares
28552743.8
Trend line equation y = 905366.216x - 4910.595
Table 3.5.T16: Residual summary of Linearity results of impurity-E
% of
impurity
w.r.t 1000
µg/ml
Mean area
Response achieved
Response
calculated
thru Trend line
equation
Residual
(Response
practical -Response
theoretical)
Residual
square
0.05 27451 30661.76 3210.8 10308959.0
0.10 65263 65096.59 -166.4 27690.8
0.15 103809 99531.43 -4277.6 18297584.3
0.20 135244 133966.27 -1277.7 1632593.3
0.30 200325 202835.95 2510.9 6304849.5
Residual sum of squares
36571676.8
Trend line equation y = 688696.757x - 3773.081
3.5.7 Accuracy/Recovery
3.5.7.1 Accuracy of the assay method
Accuracy of the assay method was established by injecting three
preparations of test sample using Clopidogrel bisulfate tablets 75 mg at
50%, 100% and 150% of analyte concentration (i.e.100 µg mL-1). Each solution was injected trice (n=3) into HPLC system and the mean peak
area of Clopidogrel peak was calculated. Assay of test solution was
determined against qualified Clopidogrel bisulfate reference standard (Table 3.5.T17).
The method showed consistent and high absolute recoveries at all
three concentration (50, 100 and 150%) levels with mean absolute recovery ranging from 100.3 % to 100.8%. The obtained absolute
recoveries were normally distributed around the mean with uniform RSD
values.
3.5.7.2 Accuracy of the RS method
Accuracy of the related substances method established at 50%, 100%
and 150% of the impurities specification limit (0.20%). Test solution prepared in triplicate (n=3) with impurities (impurity-A, impurity-B,
impurity-C, Impurity-D and impurity-E) at 0.1%, 0.20% and 0.30% level
w.r.t. analyte concentration (i.e. 1.0 mg mL-1). Mean % recovery of
impurities calculated in the test solution using the area of impurities standard at 0.20% level with respect to analyte (Table 3.5.T17).
Table 3.5.T17: Recovery results
Amount
spike
d
% Recovery Mean ± % RSD for three determinations
Clopidogrel
Impurity-A
Impurity -B
Impurity -C
Impurity –D
Impurity –E
50%
100.8±
0.35
99.1±
0.28
98.1±
0.23
98.5±
0.11
100.5±
0.15
99.9±
0.37
100
%
100.3±
0.27
97.5±
0.42
99.9±
0.33
100.1±
0.17
99.5±
0.23
98.6±
0.39
150 %
100.5± 0.33
99.1± 0.28
102.1± 0.36
98.7± 0.45
98.9± 0.29
100.5± 0.19
The related substances method showed consistent and high absolute recoveries of all five impurities at all three concentration (50,100 and
150%) levels with mean absolute recovery ranging from 97.5 % to 102.1%
in drug product.
3.5.8 Solution state stability
The solution state stability of Clopidogrel in diluent (i.e. mobile phase)
in the assay method was carried out by leaving the test solutions of
sample in tightly capped volumetric flasks at room temperature for two days. The same sample solution were assayed for every 12 hours interval
up to the study period, each time freshly prepared reference standard
was used to estimate the assay of test sample. The % RSD of assay of Clopidogrel during solution stability experiments was with in 1%.
The solution stability of Clopidogrel in the related substance method
was carried out by leaving sample solution in tightly capped volumetric flask at room temperature for two days. Content of Impurity-A, Impurity-
B, Impurity-C, Impurity-D and impurity-E were checked for every 12
hours interval up to the study period. No significant change was observed in the impurity content during solution stability experiments
from the initial values up to the study period. Hence Clopidogrel sample
solutions were stable for at least 48 hours in the developed method.
Table 3.5.T18: Solution stability results of the assay method
S.No Interval %Assay
Solution stability
1 0 h 100.3
2 12 h 100.4
3 24 h 99.6
4 36 h 99.8
5 48 h 100.2
% RSD 0.3
3.5.9 Robustness
To determine the robustness of the developed method experimental
conditions were purposely altered (flow rate 0.9 mL min-1 and 1.1 mL
min-1, varying ethanol content in mobile phase by +/- 3%, varying
column temperature from 25°C to 300C) and the resolution between all
pairs of compounds was greater than 2.0 and tailing factor for
Clopidogrel and its impurities was less than 1.2. The assay variability of
Clopidogrel was within ±1%. The variability in the estimation of
Clopidogrel bisulfate impurities was within ±10%.
3.5.10 Mass balance
The mass balance is very closely linked to the development of
stability indicating assay method as it acts as an approach to establish its validity. The stressed samples of Clopidogrel bisulfate tablets were
assayed against the qualified reference standard and the results of mass
balance obtained were very close to 99%. The results of mass balance obtained in each condition are presented below (Table 3.5.T19).
Table 3.5.T19: Mass balance of the assay method
Stress condition Time
~% Degradation
% Assay of
active
substance
Mass balance
(%Assay +
%impurities +% Degradation
products)
Acid hydrolysis
(0.5 N HCl, 80°C ) 24 h 0.6 98.6 99.2
Base hydrolysis
(0.5 N NaOH,
80°C )
24 h 7.1 92.2 99.3
Oxidation
(3% H202) 24 h 12.8 86.1 98.9
Water hydrolysis
(80°C) 24 h 0.5 98.6 99.1
Thermal (60°C) 10 days 1.3 98.2 99.5
Photolytic
degradation 10 days 0.2 99.6 99.8
3.6 Summary and Conclusions
A new normal phase liquid chromatographic method was developed
for assay and impurities estimation of Clopidogrel bisulfate with stability-indicating power for pharmaceutical dosage forms. The proposed
developed method is capable of separating all chiral and non chiral
impurities of Clopidogrel bisulfate in presence degradation impurities formed under forced degradation studies and also proposed method can
be employed for assay determination of Clopidogrel bisulfate in
pharmaceutical dosage forms. The column used for development of this
method was Chiralcel OJ-H 250 mm x 4.6 mm, 5μm. Ethanol and n- Hexane and diethyl amine in 5:95:0.05, v/v/v; ratio was employed as a
mobile phase. The detection wavelength was to monitor eluted
components was 240 nm. Clopidogrel bisulfate tablets were exposed to
degradation conditions of acid, oxidative, base, hydrolytic, photolytic and
thermal degradation. The degradation impurities were well separated from Clopidogrel peak and its related compounds, showed the stability-
indicating power of the developed method. The developed method was
validated as per ICH guidelines for precision, accuracy, linearity, specificity, limit of detection, limit of quantification, system suitability
and robustness and found to be precise, accurate, linear rugged, robust
and specific. The developed can be employed assay and purity estimation
of Clopidogrel bisulfate drug substance (API). Thus the developed and validated normal phase liquid chromatographic method can be used for
stability analysis and production samples analysis of Clopidogrel
bisulfate in API and in pharmaceutical dosage forms.
Table 3.6.T1 Summary of analytical method validation
Test
parameter Related Substance method
Assay
method
Imprity-A
Impurity-B
Impurity-C
Impurity-D
Impurity-
E
Precision
(RSD) 1.53 2.11 1.64 1.44 1.93 0.81
LOQ
(µg mL-1) 0.410 0.320 0.310 0.360 0.360 0.320
LOD
(µg mL-1) 0.160 0.140 0.090 0.110 0.130 0.120
Precision
at LOQ(%RSD)
1.3 3.7 2.6 1.2 1.6
Linearity
(Corre
coefficient)
0.9992 0.9999 0.9991 0.9995 0.9989 0.9999
Accuracy %
at 100%
(Drug product)
97.5 99.9 100.1 99.5 98.6 100.3
Robustness
Resolution between all pairs of compounds was greater than
2.0 and tailing factor for Clopidogrel and its impurities was less than 1.2.
Solution Stability
Stadard and test solutions are stable for 48 hrs at room
temperature
Specificity
For all stress conditions, purity angle is less than the purity
threshold and there is no purity flag observed for purity
results
Reference:
[1] T. Richter, T.E. Mürdter, G. Heinkele, J. Poleis, S. Tatzel, M.
Schwab, M. Eichelbaum, U.M. Zanger, J. Pharmacol. Exp. Ther.
308 (2004) 189–197.
[2] J.M. Pereillo, M. Maftouh, A. Andrieu, M.F. Uzabiaga, O. Fedeli, P.
Savi, M. Pascal, J.M. Herbert, J.P. Maffarand, C. Picard, Drug
Metab. Dispos. 30 (2002) 1288–1295.
[3] United States Pharmacopeia, 28th ed., The United States
Pharmacopeial Convention, Rockville, Maryland, USA, 2005, p.
516–517.
[4] P. Mishara, A. Dolly, Indian J. Pharm. Sci. 67 (2005) 491–493.
[5] P. Mishara, A. Dolly, Indian J. Pharm. Sci. 68 (2006) 365–368.
[6] A. Mitakos, I. Panderi, J. Pharm. Biomed. Anal. 28 (2002) 431–438.
[7] H.Y. Aboul-Enein, H. Hoenen, A. Ghanem, M. Koll, J. Liq.
Chromatogr. Relat.Technol. 28 (2005) 1357–1365.
[8] A.S. Fayeda, S.A. Weshahy, M.A. Shehata, N.Y. Hassan, J.
Pauwels, J. Hoogmartens, A.V. Schepdael, J. Pharm. Biomed.
Anal. 49 (2009) 193–200.
[9] H. Agrawal, N. Kaul, A.R. Paradkar, K.R. Mahadik, Talanta 61
(2003) 581–586.
[10] A. Robinson, J. Hillis, C. Neal, A.C. Leary, J. Chromatogr. B. 848
(2007) 344–347.
[11] Y. Gomez, E. Adams, J. Hoogmartens, J. Pharm. Biomed. Anal. 34
(2004) 341–348.
[12] P. Lagorce, Y. Perez, J. Ortiz, J. Necciari, F. Bressolle, J.
Chromatogr. B. Biomed. Sci. Appl. 720 (1998) 107–117.
[13] H. Ksycinska, P. Rudzki, M. Bukowsk-Kiliszek, J. Pharm. Biomed.
Anal. 41 (2006) 533–539.
[14] E. Souri, H. Jalalizadeh, A. Kebriaee-Zadeh, M. Shekarchi, A.
Dalvandi, Biomed. Chromatogr. 20 (2006) 1309–1314.
[15] ICH, Validation Of Analytical Procedures: Text And Methodology
(Q2(R1)): International Conference on Harmonization, IFPMA,
Geneva, 2005.
[16] ICH, Stability Testing of New Drug Substances and Products
(Q1AR): International Conference on Harmonization, IFPMA,
Geneva, 2000.