SriniVas Sadda, MDProfessor of OphthalmologyDirector, Medical Retina Unit Ophthalmic Imaging UnitUniversity of Southern CaliforniaLos Angeles, California, USA
Swept Source OCT
![Page 2: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/2.jpg)
Disclosure
Consulting Fee: Allergan; Carl Zeiss Meditec;Genentech; Optos; Regeneron
![Page 3: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/3.jpg)
What is swept source OCT?• Swept Source OCT
– Another Fourier domain OCTtechnology
– 5-10X faster than existing SD-OCT instruments
– Also better sensitivity with lessroll-off
From Potsaid et al, OpticsExpress 2010; 18: 20029.
![Page 4: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/4.jpg)
Improved Speed• Swept Source OCT
– At these speeds, fixation/motionis less of an issue
– Large areas can be scannedquickly and with extensiveaveraging
From Potsaid et al, OpticsExpress 2010; 18: 20029.
![Page 5: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/5.jpg)
Large field OCT Imaging
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
12 mm B-scans are easy to obtain
![Page 6: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/6.jpg)
Large field OCT Imaging
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
Choroid and laminar portions of the optic nerve are well seen
![Page 7: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/7.jpg)
Large field OCT Imaging
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
Choroid and laminar portions of the optic nerve are well seen
![Page 8: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/8.jpg)
Large field OCT imaging
12 x 9 mm volume scans are also easy
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 9: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/9.jpg)
Large field OCT imaging
12 x 9 mm volume scans are also easy
Produce OCTprojectionmaps that trulyresemblefundus images
Weissring
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 10: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/10.jpg)
Large field OCT Imaging
12 mm B-scans – visualize ON and Macula
Geographic atrophy
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 11: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/11.jpg)
Higher Sensitivity• Very little sensitivity loss with depth
with swept source OCT– Better signal-to-noise ratio
• Enhances visualization of outer retinalstructures
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 12: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/12.jpg)
Higher Sensitivity• Very little sensitivity loss with depth
with swept source OCT– Better signal-to-noise ratio
• Enhances visualization of outer retinalstructures, and deep pathologies
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 13: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/13.jpg)
Combine high sensitivity with optimized averaging
• Swept Source OCT– Extensive averaging allows fine
structures to be seen
From Potsaid et al, OpticsExpress 2010; 18: 20029.
![Page 14: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/14.jpg)
A note about image quality…• Though swept source devices feature better
sensitivity than spectral domain devices:– devices do vary in the quality of their optics and
other components– faster scan rates can be associated with degraded
quality
• Averaged B-scans can hide these qualitydifferences
• Advisable to look at the un-averaged nativeimages when comparing between devices
![Page 15: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/15.jpg)
A note about image quality…• Un-averaged native single B-scan image on
patient with media opacity and myopiaEven without averaging,despite scan rates >100K/sec, outer retinalbands are well seen
Note, no apparentsensitivity loss fromanterior to posteriorregions of retina in thiseye with a deepstaphyloma frompathologic myopia
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 16: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/16.jpg)
A note about image quality…• Underlying high quality B-scan data yields
high quality OCT fundus images
Pathologic myopia withtilted nerve and PPA
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
High resolution“Megapixel”
![Page 17: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/17.jpg)
Long Wavelength Imaging
• Most Swept Source OCT devicesfeatures a light source with a 1050nmcenter wavelength
Most SD-OCT devices
![Page 18: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/18.jpg)
OCT and wavelength
• Two “Windows of Opportunity” for retinalOCT imaging
OCT Imaging Windows
1. Visible to near infrared(950nm) -- BROAD
2. 1000nm – 1100nm --NARROWER BANDWIDTH(restricted to 100 nm) andstill more absorption than atshorter wavelengths
![Page 19: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/19.jpg)
Choroidal Visibility: 1050 vs 840Comparison Study at Doheny of 1050nm vs 840nmResults:• Even when the choroid was fully-visible at 840nm, considerable additional
detail was visible at 1050nm
1050 nm 840 nm
(both spectral domain)
![Page 20: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/20.jpg)
Retinitis Pigmentosa
Orbital Fat
![Page 21: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/21.jpg)
1.7mm Choroidal Melanoma
![Page 22: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/22.jpg)
ILM to choroid-sclera ISOS to BM
Choroid 1/3 of choroid 2/3 of choroid
Choroidal imaging with SSOCT
Zeiss Swept Source OCT (prototype, not yet FDA cleared)
![Page 23: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/23.jpg)
En face imaging through retina and choroid
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 24: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/24.jpg)
Serous PED
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 25: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/25.jpg)
En face imaging through CNV
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 26: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/26.jpg)
En face imaging through CNV
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 27: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/27.jpg)
En face imaging through Geographic Atrophy
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 28: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/28.jpg)
En face imaging through Geographic Atrophy
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 29: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/29.jpg)
High-resolution en face OCTEn face slabs can be extracted from any layer of interest
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 30: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/30.jpg)
Non-proliferative Diabetic Retinopathy• Microaneurysms visible on en face images
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 31: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/31.jpg)
Clinical Applications of Swept Source OCT
• Everything!– There is really no advantage of spectral
domain over swept source OCT
– Slightly better axial resolution at 840nm vs1050nm is outweighed by many otheradvantages
![Page 32: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/32.jpg)
Clinical Applications of Swept Source OCT
• But why will SS-OCT be a game changer?
• What are the key new applications that willexpand the purview of OCT?
![Page 33: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/33.jpg)
Penetrating through the sclera
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
• Imaging choroidal tumors may be an exciting new application• Imaging the anterior orbit?? --- remains to be seen
![Page 34: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/34.jpg)
Vitreous imaging with SS-OCT
Better sensitivity means bettervisualization of subtle structures
• Better evaluation of vitreo-macular interface disease andnormal vitreous dynamics
• Possibility of quantification ofvitreous cell
![Page 35: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/35.jpg)
Vitreous imaging with SS-OCT
Wide angle scans allow vitreous relationships between opticnerve and macula to be studied
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 36: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/36.jpg)
![Page 37: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/37.jpg)
Vitreous imaging with SS-OCTWith SS-OCT, both the vitreous and choroid can be imagedsimultaneously with brilliant detail
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 38: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/38.jpg)
Anterior segment imaging with SS-OCT
Deep penetration (1050nm) andhigh sensitivity of SSOCT allowsvisualization of full extent oflens/IOL and angle recess
![Page 39: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/39.jpg)
Anterior segment imaging with SS-OCTIris imaging
![Page 40: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/40.jpg)
Anterior segment imaging with SS-OCTIris imaging
Evaluation of the trabecularoutflow system, Schlemm’scanal, and collector channelsis an area of current interest
![Page 41: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/41.jpg)
Choroidal imaging with SS-OCT
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
![Page 42: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/42.jpg)
Optic Nerve Head Imaging with SSOCT
![Page 43: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/43.jpg)
Glaucomatous nerves with SSOCT
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
Drance hemorrhages areassociated with tears inthe lamina
Moderate to severe glaucoma
![Page 44: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/44.jpg)
Glaucomatous nerves with SSOCT
Zeiss SS-OCT prototype (investigational device, not FDA cleared)
Early glaucoma
![Page 45: Swept Source OCT - Portale OCulistica Amedeo Lucente · – 5-10X faster than existing SD-OCT instruments – Also better sensitivity with less roll-off From Potsaid et al, Optics](https://reader033.fdocuments.us/reader033/viewer/2022051322/601886572966a50eda336a68/html5/thumbnails/45.jpg)
Summary• Swept source OCT represents the next
evolution of OCT technology
• Its higher speed, higher sensitivity, andexpected longer wavelength lightsource, offer important advantages overexisting commercial SD-OCT devices
• SS-OCT will likely further expand theapplications and importance of OCT