Surface-activated chemical ionization ion trap mass spectrometry in the analysis of amphetamines in...
-
Upload
simone-cristoni -
Category
Documents
-
view
212 -
download
0
Transcript of Surface-activated chemical ionization ion trap mass spectrometry in the analysis of amphetamines in...
RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Rapid Commun. Mass Spectrom. 2004; 18: 1847–1855
Published online in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/rcm.1558
Surface-activated chemical ionization ion trap mass
spectrometry in the analysis of amphetamines in
diluted urine samples
Simone Cristoni1*, Luigi Rossi Bernardi1, Piermario Gerthoux2, Elisabetta Gonella2
and Paolo Mocarelli2
1University of Milan, Centre for Bio-molecular Interdisciplinary Studies and Industrial Applications (CISI), Via Fratelli Cervi 93,
20090 Segrate Milano, Italy2University Department of Laboratory Medicine, University of Milano-Bicocca, Hospital of Desio, Via Mazzini 1, 20033 Desio, Milan, Italy
Received 9 March 2004; Revised 17 June 2004; Accepted 18 June 2004
A new ionization method, named surface-activated chemical ionization (SACI), was employed for
the analysis of five amphetamines (3,4-methylenedioxyamphetamine (MDA), 3,4-methylenedioxy-
methamphetamine (MDMA), 3,4-methylenedioxyethylamphetamine (MDE), amphetamine and
methamphetamine) by ion trap mass spectrometry. The results so obtained have been compared
with those achieved by using atmospheric pressure chemical ionization (APCI) and electrospray
ionization (ESI) using the same instrument, clearly showing that SACI is the most sensitive of
the three. The limit of detection and linearity range for SACI were compared with those obtained
using APCI and ESI, showing that the new SACI approach provides the best results for both cri-
teria. SACI was used to analyze MDA, MDMA MDE, amphetamine and methamphetamine in four
urine samples, and the quantitation results are compared with those achieved using ESI. Copyright
# 2004 John Wiley & Sons, Ltd.
Amphetamines are compounds able to produce hallucino-
genic and euphoric effects.1 They are potent sympathomi-
metic amines causing stimulation of the central nervous
system. Street use of amphetamines may lead to a syndrome
called amphetamine psychosis, which is indistinguishable
from paranoid schizophrenia.2–5 The unfortunately wide dif-
fusion of amphetamines has dramatically increased in the last
years on the European illegal market. This trend is confirmed
by the European Monitoring Centre for Drugs and Drug
Addiction (EMCDDA).2
Amphetamines can be usually detected in urine, plasma
samples, nails or hair samples.5–7 Several different analytical
techniques are routinely employed for the analysis of these
compounds, based on the immunochemical approach (radio-
immunoassay (RIA), enzyme immunoassay (EIA), and
fluorescence immunoassay polarization (FPIA)).5 Mass spec-
trometry is usually employed as a confirmative assay. The
immunochemical approaches suffer, in practice, from a
number of limitations: (1) high cut-off values permit positive
detection only for recent ingestion of high doses, resulting in
the risk of false negative analyses;8,9 (2) low cut-off values
lead to a high number of false positive results, due to presence
of substances employed for therapeutic purposes such as
ephedrine and ranitidine, as well as to interfering substances
present in the biological matrix;10,11 and (3) immunochemical
assays are cost-effective only when many samples are
analyzed.5
Mass spectrometry is a powerful technique, in terms of
sensitivity and specificity, for the detection of drugs. In
particular gas chromatography/mass spectrometry (GC/MS)
has been widely employed for the analysis of amphetamines
and other drugs, using both positive- and negative-ion modes
and using various derivatization procedures;5,12–29 in
most cases a derivatization step is essential to achieve valid
results, even if it is time-consuming and often leads to sample
loss.
Other methods employed to analyze amphetamines and
other drugs are based on the use of capillary electrophoresis/
mass spectrometry (CE/MS)30–34 and particularly on liquid
chromatography/mass spectrometry (LC/MS).35–42 The
ionization technique mainly used to analyze amphetamines
and other drugs are electrospray ionization (ESI)35–39 and
atmospheric pressure chemical ionization (APCI).40,41
Another interesting technique successfully employed for
drug analysis is sonic spray ionization (SSI).42 This method is
similar to ESI but it does not use a high potential on the
nebulizer spray needle; the ionization and vaporization of the
sample are achieved by using a high nitrogen nebulizer gas
flow. It must be emphasized that, in most cases, a solid-phase
extraction (SPE) step is usually employed before the
analysis;36,37,39 on one hand this approach strongly increases
the sensitivity of the developed methods, but on the other it is
time-consuming.
Copyright # 2004 John Wiley & Sons, Ltd.
*Correspondence to: S. Cristoni, Universita degli Studi di Milano(CISI), Via Fratelli Cervi 93, 20090 Segrate Milano, Italy.E-mail: [email protected]
Scheme 1. Structural formulas of the five amphetamines analyzed.
Figure 1. Full scan mass spectra of a mixture of MDMA, MDA, MDE, amphetamine and
methamphetamine obtained using (a) APCI, (b) ESI, and (c) SACI. The concentration was 50 ng/
mL for each compound. The direct infusion sample flow was 30 mL/min. The counts/s value and the
S/N ratio of the most abundant peak in each spectrum are also reported. The S/N ratio was
calculated using the RMS algorithm.
Copyright # 2004 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2004; 18: 1847–1855
1848 S. Cristoni et al.
As shown by Sato et al.,35 an interesting approach for drug
analysis is based on the direct injection of urine samples into
the LC/MS system. However, it must be emphasized that,
although this method is fast (due to the absence of sample
pre-treatments), the direct injection of urines into the
chromatographic column can lead to serious decrease of
chromatographic performance due to the urine matrix that
contains salts and other compounds present at high
concentration which can precipitate inside the chromato-
graphic column. Moreover, some matrix compounds can
exhibit negative effects with respect to sample ionization
efficiency in both ESI and APCI, leading to instability of the
analyte signal and thus to quantitation errors.
Surface-activated chemical ionization (SACI) has been
recently developed in our laboratory.43 This ionization
technique is an improvement of the no-discharge ionization
approach.44–46 As in the case of SSI, it does not use a high
potential in order to ionize the sample molecules. It is
composed of a heated vaporization chamber into which a
metallic surface is inserted, leading to high ionization
efficiency when no or low potential (generally lower than
150 V) is applied to it. The new device has been tested in the
analysis of some high molecular weight standard peptides
(peptide YY fragment 13-36, 3014 Da; diabetes associated
peptide fragment 8-37, 3200 Da; gastrin-releasing peptide
(human), 2859 Da; phospholipase A2 activating peptide,
2330 Da; and vasoactive intestinal peptide fragment 6-28,
2816 Da) and also of peptides obtained by tryptic digestion of
cytochrome C, introduced by direct infusion. This new
approach exhibits high performance in term of selectivity
and sensitivity, and this encouraged us to employ it also for
the analysis of amphetamines in order to verify the
performance of SACI in the analysis of these compounds.
The aim of this work is to compare the data obtained
using SACI, APCI and ESI techniques in term of sensitivity,
linearity range and limit of detection (LOD) in the analysis
of five amphetamines (3,4-methylenedioxyamphetamine
(MDA), 3,4-methylenedioxymethamphetamine (MDMA),
3,4-methylenedioxyethylamphetamine (MDE), ampheta-
mine and methamphetamine). After the optimization of
the instrumental parameters for a mixture of the pure
compounds, four urine samples were analyzed; they were
strongly diluted (1:100) in order to prevent either matrix-
sample interaction effects taking place during ionization, or
column damage due to the matrix composition. Further-
more, the diluted urine samples were directly analyzed,
without pre-treatment, using LC/MS and LC/MS/MS; the
results obtained are reported and discussed.
EXPERIMENTAL
ChemicalsStandard MDMA, MDA, MDE, amphetamine and metham-
phetamine were purchased from SALARS (Como, Italy).
Acetonitrile was purchased from J. T. Baker (Deventer, The
Netherlands). Formic acid was purchased from Sigma
Aldrich (Milan, Italy).
Sample preparationThe urine samples obtained from four drug addict subjects, in
which the amphetamine composition had been previously
qualitatively and quantitatively evaluated using GC/MS,
were diluted using two dilution ratios (1:20 and 1:100). These
samples were directly analyzed by LC/MS and LC/MS/MS
using different ionization sources.
Scheme 2. Fragmentation pathways of the [MþH]þ ions of MDMA, MDA and MDE.
SACI-LC/MS analysis of amphetamines in urine 1849
Copyright # 2004 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2004; 18: 1847–1855
ChromatographyA Surveyor LC system (ThermoFinnigan, San Jose, CA, USA)
was used. The chromatographic column was a reverse-phase
C18 (150� 1 mm, 5mm, 300 A). The HPLC gradient used two
eluents: (A) H2Oþ 0.05% formic acid and (B) CH3CNþ0.05% formic acid: 20% of B was maintained for 2 min, then
a linear gradient was used passing from 20% of B to 80% of
B in 10 min. Then 80% of B was maintained for 3 min, and
then over the next 2 min the initial conditions were reached.
Thus, 17 min of chromatographic analysis time were used,
but the mass chromatogram acquisition cycle time was set
to 20 min in order to make it possible to re-equilibrate the chro-
matographic column. The eluent flow rate was 100mL/min.
Mass spectrometryThe APCI mass spectra were obtained using an LCQ-DecaXP
(ThermoFinnigan, San Jose, CA, USA) ion trap. The vaporizer
temperature was 3508C and the entrance capillary tempera-
ture was 1508C. The corona discharge voltage was 5 kV. The
flow rate of nebulizing sheath gas (nitrogen) was 2.00 L/min.
The He pressure inside the trap was kept constant; the pres-
sure directly read by ion gauge (in the absence of N2 stream)
was 2.8� 10�5 Torr. The maximum injection scan time was
200 ms, five microscans were used and the automatic gain
control (AGC) was turned on.
The same instrument was used to obtain the ESI mass
spectra. The needle voltage was 5 kV. The entrance capillary
temperature was 2408C. The flow of nebulizing gas (nitrogen)
was 1.5 L/min. The He pressure inside the trap (in the
absence of N2 stream), maximum injection scan and micro-
scan times were the same as those used to obtain the APCI
spectra. The AGC was turned on.
SACI spectra were obtained using a gold surface held at a
potential of 150 V. The vaporizer temperature was 4008C and
the entrance capillary temperature was 1508C. The surface
temperature, monitored using an optical pyrometer, was
808C. The flow of nebulizing gas (nitrogen) was 2.5 L/min.
The He pressure inside the trap (in the absence of N2 stream),
maximum injection scan and microscan times were the same
as those used in order to obtain the APCI spectra. The AGC
was turned on.
LC/ESI, APCI and SACI mass chromatograms were
obtained using both full scan MS, and by MS/MS with single
reaction monitoring (SRM) (only for analyzing MDA); the full
Figure 2. LC/APCI extracted mass chromatograms of (a) MDMA, (b) MDA (obtained
using SRM), (c) MDE, (d) amphetamine, and (e) methamphetamine, obtained by injecting
20 mL of a 50 ng/mL standard mixture (corresponding to 1 ng injected on column for
each drug).
1850 S. Cristoni et al.
Copyright # 2004 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2004; 18: 1847–1855
scan range was m/z 110–220, and the scan time and peak
width (FWHM) were 200 ms and 0.6 Da, respectively. In these
conditions the instrumental peak width was 1 Da. In the case
of SRM experiments, the isolation width of the precursor ion
was 3 Da and the peak width for the monitored fragment ion
was 3 Da. The collision energy was 30% of its maximum value
(5 V peak to peak); five microscans were used and the
microscan time in SRM mode was 20 ms.
The mass spectra were acquired using positive-ion mode.
The direct infusion mass spectra were obtained using a
sample flow of 30mL/min.
Data analysisThe signal/noise (S/N) ratio was calculated using the
root mean square (RMS) algorithm. The chromatographic
data were elaborated using Xcalibur qualbrowser and Excel
software.
RESULTS AND DISCUSSION
Preliminary results were obtained by direct infusion of a
mixture of the five selected drugs (Scheme 1), with a concen-
tration of 50 ng/mL for each compound, using APCI, ESI and
SACI methods; the mass spectra obtained are shown in
Figs. 1(a), 1(b) and 1(c), respectively. The counts/s values
and the S/N ratios of the most abundant peaks in the spectra
are also reported. It is emphasized that an ion at m/z 163 is
present at high relative abundance in both APCI and ESI
spectra (Figs. 1(a) and 1(b)), but the intensity of this peak is
very low in the spectrum obtained using SACI. This ion
could have originated through ‘in-source’ fragmentation of
[MþH]þ ions of MDMA, MDA and MDE, leading to the
structure shown in Scheme 2. To verify this hypothesis the
MS/MS spectra of the [MþH]þ ions of these compounds,
obtained by injection of 200 ng/mL solutions of pure com-
pounds, were obtained (data not shown); they exhibit an
abundant peak at m/z 163 and this clearly shows that this
fragmentation pathway is strongly favored. These results
indicate that partial fragmentation of the [MþH]þ ions of
these amphetamines takes place using both APCI and ESI,
while in SACI conditions it does not occur, indicating that
this approach is the most suitable ionization technique for
the analysis of these compounds. Furthermore, the lowest
chemical noise and the best S/N ratio were obtained in
Figure 3. LC/ESI ion extracted mass chromatograms of (a) MDMA, (b) MDA (obtained
using SRM), (c) MDE, (d) amphetamine, and (e) methamphetamine obtained by
injecting 20mL of a 50 ng/mL standard mixture (corresponding to 1 ng injected on
column for each drug). The counts/s values and the S/N ratios are also reported. The S/
N ratio was calculated using the RMS algorithm.
SACI-LC/MS analysis of amphetamines in urine 1851
Copyright # 2004 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2004; 18: 1847–1855
SACI conditions, even though the counts/s value is lower
than those achieved using ESI and APCI (Figs. 1(a), 1(b)
and 1(c)). It must be emphasized that, at this analyte concen-
tration level, no [MþH]þ signals were detected in APCI con-
ditions (Fig. 1(a)); by using the ESI source these species
became detectable (Fig. 1(b)), but in the presence of abundant
chemical noise. The chemical noise present in the ESI spec-
trum, higher than that observed in SACI conditions, is prob-
ably due to the high potential (5 kV) at which the spray
needle of the ESI source operated; this high potential gives
rise to high production of solvent ion clusters that lead to
increases in the chemical noise.
An LC method was also developed in order to analyze the
five drugs. The compound that gave rise to the lowest signal
intensity was MDA, detected at m/z 180 (see Figs. 1(b)
and 1(c)). The collision spectra (MS/MS) of [MþH]þ of MDA,
obtained in both ESI and SACI conditions, show an abundant
peak at m/z 163 corresponding to loss of NH3 (Scheme 2,
fragmentation pathway b). Thus, this fragmentation reaction
can be used to perform SRM analysis in order to increase the
instrumental sensitivity via increased selectivity (decrease in
chemical noise); use of SRM to analyze this drug resulted in
an LOD lower by a factor 10. Thus, two acquisition methods
were alternatively used during the chromatographic analy-
sis. The first was the full scan analysis used to detect MDMA,
MDE, amphetamine and methamphetamine, while SRM was
used to detect MDA only.
The LC/APCI ion extracted mass chromatograms
obtained by injecting 20 mL of a 50 ng/mL standard mixture
solution (corresponding to 1 ng injected on-column) are
shown in Figs. 2(a)–2(e); in the case of APCI no chromato-
graphic peaks were detected at this analyte concentration
level. Better results were achieved using ESI (Figs. 3(a)–3(e))
in which all compounds except MDA (Fig. 3(b)) were clearly
detected. Each compound exhibits a characteristic retention
time (MDMA 10.60 min, MDE 11.04 min, amphetamine
9.66 min, methamphetamine 10.39 min). The S/N ratio was
between 11 and 33.
Figure 4. LC/SACI ion extracted mass chromatograms of (a) MDMA, (b) MDA (obtained
using SRM), (c) MDE, (d) amphetamine, and (e) methamphetamine obtained by injecting 20 mLof a 50 ng/mL standard mixture (corresponding to 1 ng injected on column for each drug). The
counts/s values and the S/N ratios are also reported. The S/N ratio was calculated using the RMS
algorithm.
1852 S. Cristoni et al.
Copyright # 2004 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2004; 18: 1847–1855
Using the SACI source a strong improvement in terms of S/
N ratio (between 102 and 371, Figs. 4(a)–4(e)) was achieved. It
is emphasized that, by using this new ionization method,
MDA was clearly detected (retention time 10.07 min).
Calibration curves for MDMA, MDA, MDE, amphetamine
and methamphetamine were also obtained using the three
ionization sources, to obtain and compare the linearity ranges
achieved for each compound (Table 1). The R2 values of the
calibration curves were in the range 0.9545–0.9948. The lower
limit of the linearity range for SACI is about a factor 15–100
lower than that achieved using ESI and APCI, and the upper
limit is a factor of 10 lower with respect to that obtained with
ESI and APCI. In fact, in the case of SACI, if the upper
linearity range limit is exceeded, the calibration curve flattens
to reach a plateau. This is probably due to a saturation effect
of the ionizing surface that leads to a decrease in ionization
efficiency. However, the upper to lower linearity limit ratio is
definitely better for SACI (between 20–50) than for ESI
(between 13–20) and APCI (between 2–5). Thus, all these
effects lead to a good linearity range at lower concentration
levels with respect to ESI and APCI. Moreover, this fact
makes it possible to strongly dilute the urine samples before
executing the direct injection analysis (see below).
As expected the worst LOD was achieved using APCI
(200 ng/mL for MDMA, MDE, amphetamine and metham-
phetamine injecting 20mL on-column corresponding to 4 ng,
and 500 ng/mL for MDA injecting 20 mL on-column corre-
sponding to 10 ng). The LOD values obtained using ESI were
lower than those for APCI (50 ng/mL for MDMA, MDE,
amphetamine and methamphetamine injecting 20 mL on-
column corresponding to 1 ng, and 75 ng/mL for MDA
injecting 20 mL on-column corresponding to 1.5 ng). Again,
however, the best performance was obtained using SACI
(LOD 2 ng/mL for MDMA, MDE, amphetamine and
methamphetamine injecting 20mL on-column corresponding
to 0.04 ng, and 5 ng/mL for MDA injecting 20mL on-column
corresponding to 0.1 ng). The high sensitivity of SACI can be
accounted for, at least in part, by a low degree of in-source
fragmentation of the [MþH]þ ions and better ion focusing
conditions of this ionization approach. It is believed that the
ionizing surface can act as an electrostatic mirror after
formation of the ions to better direct the ions to the mass
analyzer. Some experiments are currently in progress using
molecules of different m/z ratios to study the effect of the
surface potential on both ionization and focusing efficiency.
From these data it was clear that the ESI and SACI
approaches gave the best performance, and thus they were
chosen to analyze four urine samples containing some of the
selected drugs. Two dilution ratios (1:20 and 1:100) were used
in order to perform direct analyses of the urine samples. The
quantitation values for these drugs, quoted with reference to
the undiluted urine samples, detected using both ESI (urine
dilution ratio 1:20) and SACI (urine dilution ratio 1:100), are
reported in Table 2. To reduce the quantitation errors, three
injection replicates were used for each sample. The standard
deviations reported in Table 2 were calculated considering
the standard deviation of the linearity calibration curves. The
agreement between the results obtained using the two
ionization sources is quite good; the percent variations of
the quantitation values of the detected drugs using SACI with
respect to those obtained using ESI are between 0.1–8%. This
is the range expected based on the instrumental precision.
The ion trap analyzer employed to obtain the experimental
data is affected by space charge effects47 that must be
accounted for by the AGC, and leads to lower precision in
quantitative analyses. Better results should be obtained using
other mass analyzers such as a triple quadrupole that can
provide high quantitation precision.48
It must be emphasized that ESI was able to detect drugs
when the sample dilution ratio was 1:20, but no drugs were
detected using ESI when the dilution ratio was 1:100. As an
example, Figs. 5(a)–5(c) show the extracted mass chromato-
grams for the [MþH]þ ion of MDMA at m/z 194, obtained by
injecting 20mL of urine 2 (Table 2) diluted 1:100 and using the
APCI, ESI and SACI techniques, respectively. The only
technique able to detect the analyzed drug at this concentra-
tion level was SACI. Furthermore, some drugs were detected
only using the SACI source and not with ESI even when using
Table 1. Linearity ranges obtained using APCI, ESI and
SACI for analysis of MDMA, MDA, MDE, amphetamine and
methamphetamine. The injection volume was 20 mL
Compound
Linearity ranges
APCI (ng/mL) ESI (ng/mL) SACI (ng/mL)
MDMA 200–1000 50–1000 2–100MDA 500–1000 75–1000 5–100MDA 200–1000 50–1000 2–100Amphetamine 200–1000 50–1000 2–100Methamphetamine 200–1000 50–1000 2–100
Table 2. Quantitation values, referred to the undiluted urine
samples, for MDA, MDMA, MDE, amphetamine and
methamphetamine detected in urine samples using both
ESI (urine dilution ratio 1:20) and SACI (urine dilution ratio
1:100). Three injection replicates were used for each sample.
The injection volume was 20 mL
Sample Compounds ESI (ng/mL) SACI (ng/mL)
Urine 1 MDMA — —MDA N.D.* 688� 10MDE — —Amphetamine — —Methamphetamine — —
Urine 2 MDMA 1236� 28 1138� 23MDA N.D.* 512� 10MDE — —Amphetamine — —Methamphetamine — —
Urine 3 MDMA 1803� 32 1800�30MDA — —MDE — —Amphetamine N.D.* 434� 17Methamphetamine N.D.* 797� 14
Urine 4 MDMA — —MDA — —MDE — —Amphetamine N.D.* 750� 10Methamphetamine 1825� 36 1810� 22
* The compounds classified as N.D. were detected using SACI but notusing ESI.
SACI-LC/MS analysis of amphetamines in urine 1853
Copyright # 2004 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2004; 18: 1847–1855
the dilution ratio 1:20 (Table 2). This observation corresponds
to the higher LODs of ESI, mentioned above.
Finally, an experiment was performed in order to evaluate
whether SACI is affected by the signal suppression phenom-
enon due to endogenous components in the biological matrix
that typically affects liquid ionization sources.49 A urine
sample that did not contain the previous mentioned drugs
was spiked in three aliquots of 1 mL. MDMA was added to
the three urine samples at concentrations of 500 (sample A),
1000 (sample B) and 1500 (sample C) ng/mL. These samples
were then diluted 1:100 and analyzed by LC/SACI-MS. The
MDMA external calibration curve (obtained using clean
solutions of standards) was used to measure the MDMA
quantitation values in the three samples. The measured and
real quantitation values, referred to the undiluted urine
samples, are reported in Table 3. The calculated amounts of
MDMA are about 39–47% lower with respect to those
predicted by calibration using the standard solutions. Thus,
it can be concluded that SACI is also affected by the matrix
effect. In future work a detailed study of this and other
instrumental aspects of the SACI source will be performed.
CONCLUSIONS
The new SACI method provides performance in terms of sen-
sitivity, limit of detection and linearity range in the analysis of
MDMA, MDA, MDE, amphetamine and methamphetamine,
that are superior to those achieved by the usually employed
APCI and ESI techniques. The high sensitivity available using
this technique allows direct detection of these drugs in
strongly diluted (1:100) urine samples. Using this approach
it is possible to directly inject the dilute urines on the chroma-
tographic column without any noticeable column damage
caused by the urine matrix (salts, proteins, etc.). However it
must be emphasized that SACI, like ESI and APCI, is affected
by the matrix suppression effect.49 In future work the behavior
of SACI under various experimental conditions will be better
evaluated and also the matrix effect will be investigated.
Future developments will be focused on applying this
method to the analysis of other street drugs and of
compounds of clinical interest. Also, new surface materials
will be tried in order to further improve the instrumental
performance in terms of selectivity and sensitivity.
Figure 5. Extracted mass chromatograms for the [MþH]þ ion of MDMA at m/z 194 obtained by
injecting 20 mL of urine 2 diluted 1:100 and using (a) APCI, (b) ESI, and (c) SACI techniques. The
counts/s value and the S/N ratio are reported in the chromatogram obtained using SACI. The S/N
ratio was calculated using the RMS algorithm.
Table 3. Measured and real quantitation values of MDMA
added to a urinary matrix solution. The quantitation values
were obtained using the MDMA external calibration curve
obtained using clean standard solutions. The % differences
of the measured values with respect to the real ones are also
reported. Three injection replicates were used for each
sample. The injection volume was 20 mL
Urinesample
MDMAmeasured
concentration(ng/mL)
MDMA realconcentration
(ng/mL) % Difference
A 304� 20 500 39B 552� 15 1000 45C 796� 14 1500 47
1854 S. Cristoni et al.
Copyright # 2004 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2004; 18: 1847–1855
AcknowledgementsThe authors thank Dr. Pietro Traldi for useful advice. The
authors also thank Dr. Remo Cristoni, Mrs. Maria Florio
and Mrs. Karim Amaya Mendoza for their support.
REFERENCES
1. Matz LM, Hill HH. Anal. Chem. 2002; 700: 420.2. EMMCDA. Annual Report on the State of Drugs problem in the
European Union, European Monitorino Centre for Drugs andDrug addiction, 2000.
3. Angrist BM, Gershon S. Biol. Psychiat. 1970; 2: 95.4. Griffith JD, Cavanaugh J, Held J, Oates JA. Arch. Gen.
Psychiat. 1972; 26: 97.5. Pellegrini M, Rosati F, Pacifici R, Zuccaio R, Romolo FS,
Lopez A. J. Chromatogr. B 2002; 769: 243.6. March C, Karnes HT, McLean A, Mukherjee PS. Biomed.
Chromatogr. 2001; 15: 100.7. Uhl M. Forensic Sci. Int. 2000; 107: 169.8. Hensley D, Cody JT. Anal. Toxicol. 1999; 23: 518.9. Jurado C, Gimenez MP, Soriano T, Menendez M, Repetto
M. J. Anal. Toxicol. 2000; 24: 11.10. Wu AH, Onigbinde TA, Wong SS, Johnson KG. J. Anal.
Toxicol. 1992; 16: 137.11. Kataoka H, Lord HL, Pawliszyn J. J. Anal. Toxicol. 2000; 24:
257.12. Suzuki O, Hattori H, Asano M. Forensic Sci. Int. 1984; 24: 9.13. Leis HJ, Rechberger GN, Fauler G, Windischhofer W. Rapid
Commun. Mass Spectrom. 2003; 17: 569.14. Wang SM, Giang YS, Ling YC. J. Chromatogr. 2001; 759: 17.15. Lim HK, Su Z, Foltz RL. Biol. Mass Spectrom. 1993; 22: 403.16. Maurer HH, Kraemer T, Kratzsch C, Peters FT, Weber AA.
Ther. Drug Monit. 2002; 24: 117.17. Peters FT, Kraemer T, Maurer HH. Clin. Chem. 2002; 48:
1472.18. Singh AK, Jang Y, Mishra U, Granley K. J. Chromatogr. 1991;
568: 351.19. Cone EJ, Presley L, Lehrer M, Seiter W, Smith M, Kardos
KW, Fritch D, Salamone S, Niedbala RS. J. Anal. Toxicol.2002; 26: 541.
20. Brotherton HO, Yost RA. Am. J. Vet. Res. 1984; 45: 2436.21. Reimer ML, Mamer OA, Zavitsanos AP, Siddiqui AW,
Dadgar D. Biol. Mass Spectrom. 1993; 22: 235.22. Gaillard Y, Vayssette F, Peppin G. Forensic Sci. Int. 2000;
107: 361.
23. Pizarro N, Ortuno J, Segura J, Farre M, Mas M, Cami J, De laTorre R. J. Pharm. Biomed. Anal. 1999; 21: 739.
24. Ugland HG, Krogh M, Rasmussen KE. J. Pharm. Biomed.Anal. 1999; 19: 463.
25. Beck O, Kraft M, Moeller MR, Smith BL, Schneider S,Wennig R. Ann. Clin. Biochem. 2000; 37: 199.
26. Powers KH, Ebert MH. Biomed. Mass Spectrom. 1979; 6: 187.27. Marde Y, Ryhage R. Clin. Chem. 1978; 24: 1720.28. Matin SB, Wan SH, Knight JB. Biomed. Mass Spectrom. 1977;
4: 118.29. Anggard E, Hankey A. Acta Chem. Scand. 1969; 23: 3110.30. Ramseier A, Siethoff C, Caslavska J, Thormann W. Electro-
phoresis 2000; 21: 380.31. Tsai JL, Wu WS, Lee HH. Electrophoresis 2000; 21: 1580.32. Geiser L, Cherkaoui S, Veuthey JL. J. Chromatogr. A 2000;
895: 111.33. Iwata YT, Kanamori T, Ohmae Y, Tsujikawa K, Inoue H,
Kishi T. Electrophoresis 2003; 24: 1770.34. Lazar IM, Naisbitt G, Lee ML. Analyst 1998; 123: 1449.35. Sato M, Hida M, Nagase H. Forensic Sci. Int. 2002; 128:
146.36. Sato M, Mitsui T, Nagase H. J. Chromatogr. B 2001; 751: 277.37. Katagi M, Tatsuno M, Miki A, Nishikawa M, Nakajima K,
Tsuchihashi H. J. Chromatogr. B 2001; 759: 125.38. Katagi M, Tatsuno M, Miki A, Nishikawa M, Tsuchihashi
H. J. Anal. Toxicol. 2000; 24: 354.39. Kostiainen R, Kotiaho T, Kuuranne T, Auriola S. J. Mass
Spectrom. 2003; 38: 357.40. Slawson MH, Taccogno JL, Foltz RL, Moody DE. J. Anal.
Toxicol. 2002; 26: 430.41. Bogusz MJ, Kala M, Maier RD. J. Anal. Toxicol. 1997; 21: 59.42. Mortier KA, Dams R, Lambert WE, De Letter EA,
Van Calenbergh S, De Leenheer AP. Rapid Commun. MassSpectrom. 2002; 16: 865.
43. Cristoni S, Bernardi LR, Biunno I, Tubaro M, Guidugli F.Rapid Commun. Mass Spectrom. 2003; 17: 1973.
44. Cristoni S, Bernardi LR, Biunno I, Guidugli F. Rapid Com-mun. Mass Spectrom. 2002; 16: 1686.
45. Cristoni S, Bernardi LR, Biunno I, Guidugli F. RapidCommun. Mass Spectrom. 2002; 16: 1153.
46. Cristoni S, Bernardi LR. Mass Spectrom. Rev. 2003; 22:369.
47. Dobson G, Murrell J, Despeyroux D, Wind F, Tabet JC.Rapid Commun. Mass Spectrom. 2003; 17: 1657.
48. Qi M, Wang P, Liu L. J. Chromatogr. B Anal. Technol. Biomed.Life Sci. 2004; 805: 7.
49. Mallet CR, Lu Z, Mazzeo JR. Rapid Commun. Mass Spectrom.2004; 18: 49.
SACI-LC/MS analysis of amphetamines in urine 1855
Copyright # 2004 John Wiley & Sons, Ltd. Rapid Commun. Mass Spectrom. 2004; 18: 1847–1855