SUPPLEMENTAL OXYGEN MANUAL FOR MODELS … · SUPPLEMENTAL OXYGEN MANUAL FOR MODELS 100, 200, 300...
Transcript of SUPPLEMENTAL OXYGEN MANUAL FOR MODELS … · SUPPLEMENTAL OXYGEN MANUAL FOR MODELS 100, 200, 300...
SUPPLEMENTAL OXYGEN
MANUAL FOR MODELS
100, 200, 300 ANALYZERS
Instruments, Inc.California Analytical
1312 West Grove Avenue, Orange, CA 92865 - 714 974-5560 - Fax 714 921-2531 Website: www.gasanalyzers.com
Section 1 Model 100/200/300 Supplement
Preface This instruction manual is supplemental to the Model 100/200/300 Operations and Maintenance Manual for infrared analyzers. It is sent whenever oxygen is one of the measured parameters of the analyzer. The oxygen measurement information is divided into two parts:
Part 1 – Oxygen by chemical fuel cell sensor techniques.
Part 2 – Oxygen by paramagnetic oxygen sensor techniques.
Please refer to the appropriate part of this supplemental manual according to the method of oxygen determination for your particular analyzer. Thank you.
Supplemental Manual Oxygen- V7.1.doc 1-2
Section 1 Model 100/200/300 Supplement
Table of Contents
Part 1 – Oxygen by Chemical Fuel Sensor Techniques ...................................................................5 1. FUEL CELL DETECTOR OPTION............................................................................................6
1.1. Description Fuel Cell Detector Option ................................................................................6 1.2. Product Specifications (Galvanic Fuel Cell Detector).........................................................6 1.3. Principle of operation..........................................................................................................7 1.4. Oxygen Fuel Cells ..............................................................................................................7
2. Preparations for Operation ........................................................................................................8 2.1. Power On............................................................................................................................8 2.2. Zero Adjustment .................................................................................................................8 2.3. Span Adjustment ................................................................................................................8 2.4. Start-Up & Routine Maintenance........................................................................................8 2.5. Zero and Span Calibration Frequency................................................................................8
3. ADJUSTMENTS CHECKS AND REPAIRS...............................................................................9 3.1. Electrical Zero Adjustments................................................................................................9 3.2. Electrical Span Adjustments...............................................................................................9 3.3. The Oxygen Sensor..........................................................................................................10
3.3.1. Coarse-Span-Adjustment ..........................................................................................10 3.3.2. Sensor Checks..........................................................................................................10 3.3.3. Sensor Replacement.................................................................................................10
4. Chemical Fuel Cell Sample Flow (Supplemental) ...................................................................11 5. Electrical Drawings ..................................................................................................................12 Part 2 – Oxygen by Paramagnetic Oxygen Sensor Techniques.....................................................16 6. PARAMAGNETIC OXYGEN OPTION.....................................................................................17
6.1. Description........................................................................................................................17 6.2. Product Specifications Paramagnetic Oxygen Detector...................................................17 6.3. Principle of operation........................................................................................................18
7. Preparations for Operation ......................................................................................................20 7.1. Power On..........................................................................................................................20 7.2. Zero Adjustment ...............................................................................................................20 7.3. Span Adjustment ..............................................................................................................20 7.4. Zero and Span Calibration Frequency..............................................................................20
8. Start-Up & Routine Operation..................................................................................................21 8.1. Sampling System..............................................................................................................21 8.2. Cross sensitivity of gases.................................................................................................22
8.2.1. Example 1 .................................................................................................................22 8.2.2. Example 2 .................................................................................................................22
9. Mechanical Zero Adjustment (Not required with some sensor models) ..................................24
Supplemental Manual Oxygen- V7.1.doc 1-3
Section 1 Model 100/200/300 Supplement
Table of Figures
Figure 2-1 AC Power Switch, Connector, and Fuse .........................................................................8 Figure 4-1 Chemical fuel cell sample flow ......................................................................................11 Figure 5-1 100F O2 analyzer wiring diagram ..................................................................................13 Figure 5-2 100F O2 electrical schematic.........................................................................................14 Figure 5-3 100F O2 PCB component locator ..................................................................................15 Figure 6-1 Magnetic Susceptibility of gases ...................................................................................18 Figure 6-2 The Measuring cell in theory .........................................................................................18 Figure 6-3 Principle of operation.....................................................................................................19 Figure 7-1 AC Power Switch, Connector, and Fuse .......................................................................20 Figure 8-1 Paramagnetic Oxygen Sample flow ..............................................................................21 Figure 9-1 Oxygen sensor adjustment............................................................................................24
Tables
Table 8-1 Cross Sensitivity of gases...............................................................................................23
Supplemental Manual Oxygen- V7.1.doc 1-4
Section 1 Model 100/200/300 Supplement
Part 1 – Oxygen by Chemical Fuel Sensor Techniques
Supplemental Manual Oxygen- V7.1.doc 1-5
Section 1 Model 100/200/300 Supplement
1. FUEL CELL DETECTOR OPTION
1.1. Description Fuel Cell Detector Option The fuel cell option utilizes a low cost replaceable fuel cell to determine the percent level of oxygen contained in the sample gas. A fuel cell analyzer features linear output with low noise and excellent repeatability from 0 to 100% O2.
It is available in two versions, one for oxygen levels up to 25 % and the second for oxygen up to 100%.
Warning: This is a general-purpose analyzer, not suitable for hazardous areas. High-pressure oxygen is very dangerous. Virtually any material will burn in it, possibly explosively. It is essential that persons using this analyzer are aware of the dangers of oxygen, and take all appropriate precautions.
1.2. Product Specifications (Galvanic Fuel Cell Detector)
SAMPLE CONTACT MATERIAL: Stainless steel, Teflon* and Tygon**
DISPLAY: 3 ½" digit panel meter
RANGES: Standard fixed ranges, either A or B
OUTPUTS: 0 to 10 VDC and 4 to 20 mA (0 to 20 mA)
A) Range 1: 0 to 5%; Range 2: 0 to 10%; Range 3: 0 to 25%
B) Range 1: 0 to 25%; Range 2: 0 to 40%; Range 3: 0 to 100%
RESPONSE TIME: 90% full scale in 5 seconds
AMBIENT TEMPERATURE: 5 to 45° C
SAMPLE TEMPERATURE: 0 to 50° C
SAMPLE CONDITION: Particles < 1µ, non-corrosive dry gas
FITTINGS: ¼" tube
NOISE: Less than 1% full scale SAMPLE FLOW RATE: 0.5 –2.0 L/Min
LINEARITY: Better than 1% full scale Power Requirements: 115/230 (± 10%) VAC, 50/60 Hz, 70 watts/channel
REPEATABILITY: Better than 1% full scale Relative Humidity: Less than 90% R.H.***
ZERO SPAN DRIFT: Less than 1% full scale in 24 hours
ZERO & SPAN ADJUSTMENT: Ten turn potentiometer
Specifications subject to change without notice *Teflon is a registered trademark of DuPont.
**Tygon is a registered trademark of the Norton Performance Plastics Corporation
***Non-condensing
Supplemental Manual Oxygen- V7.1.doc 1-6
Section 1 Model 100/200/300 Supplement
1.3. Principle of operation The fuel cell option is designed to measure percent levels of oxygen. The analyzer uses a galvanic fuel cell, which is an electrochemical transducer. The fuel cell contains a cathode, anode, and an electrolyte. A permeable membrane holds the electrolyte in the cell. The sample flows over the membrane and oxygen diffuses in to the fuel cell where it reacts with the electrolyte. This reaction produces an electrical current. This current is directly proportional to the concentration of oxygen in the gaseous mixture surrounding the cell. The current output is linear with an absolute zero. The fuel cell produces no current in the absence of oxygen.
1.4. Oxygen Fuel Cells The standard oxygen sensor (type PSR-39-11) has a shelf life of approximately three years. It is designed to measure oxygen in ambient air. The other gases contained in ambient air are low enough concentrations so as not to be a factor of concern. When used for ambient air measurement the manufacturer estimates the life of the sensor to be approximately three years.
If the standard PSR type sensor is used to measure pure oxygen, oxygen in flue gas or oxygen in stack emissions the life of the standard PSR type sensor will be shorter.
One of the most common emission gases is CO2 and is usually found in concentrations of between 4-20 percent by volume. High levels Of CO2 change the pH of the gel and thus shorten the life of the standard PSR type sensor. The life expectancy dramatically shortens from three years to approximately four weeks at CO2 concentrations of 20% volume.
For the reason stated above, sensor type XLT-39-11 was developed. The XLT type sensor has an electrolytic gel that is strongly buffered and highly caustic. A caustic gel is required to combat the effects of high CO2 concentrations. The average life of this type of sensor is approximately eight to ten months at CO2 levels of 20% by volume; however, the caustic nature of the electrolytic gel creates vapors that clog the cell membrane within a very short time if the cell is not put into immediate use. The shelf life of this type of sensor is approximately two weeks if not stored at a cold, constant temperature. A shelf life of 30 days or so can be achieved if the sensor is kept refrigerated.
As previously stated the XLT-39-11 sensor was developed for use in sample gases containing high levels Of CO2 (approximately 20% volume. or greater). It provides a longer life at high CO2 levels, but the linearity of response has been sacrificed. Typical linearity with this type of sensor is within ± 1% Volume O2.
Because of the non-linearity issue, a third type of sensor was developed. Type XLT-39-11-1 provides a signal that is linear to within ± 0.25% volume O2. This sensor has an average life of eight to ten months at medium CO2 levels of 10% by volume. Levels of CO2 at 20% by volume will shorten the life of this sensor to approximately 3-4 months. The shelf life of this sensor is similar to the XLT-39-11.
Supplemental Manual Oxygen- V7.1.doc 1-7
Section 2 Model 100/200/300 Supplement
2. Preparations for Operation Warning: This is a general-purpose analyzer, not suitable for hazardous areas. High-pressure oxygen is very dangerous. Virtually any material will burn in it, possibly explosively. It is essential that any persons using this analyzer are aware of the dangers of oxygen, and take all appropriate precautions.
2.1. Power On Turn ON the power switch located on the rear panel. The digital panel meter should illuminate. Allow the instrument to warm up for approximately 10 minutes. It is preferable, but not essential, that zero gas flow through the instrument during warm-up.
0
I
Figure 2-1 AC Power Switch, Connector, and Fuse
2.2. Zero Adjustment Zero Adjustment: After the 10 minute warm-up period, flow nitrogen or other zero gas through the instrument at a rate of about 1 liter/min. After the reading stabilizes, adjust the zero control on the front panel until the digital meter (or analog output) is exactly at zero. Nitrogen is the preferred zero Gas.
2.3. Span Adjustment Span Adjustment: Flow span gas through the instrument at about 1 liter/min. After the reading stabilizes, adjust the span control on the front panel until the digital meter or analog output is reading the value corresponding to the span gas concentrations.
Note: Span gas concentration should not be less than 80% of the range to be spanned. Note: On the 0-25% range of the analyzer ambient air may be used as span gas. While ambient air is flowing to the analyzer, adjust the span potentiometer to 20.9% O2.
2.4. Start-Up & Routine Maintenance Prepare and check the sample system. Adjust the flow of sample gas to about 1 liter/min. Select an operating range that is suitable for the oxygen concentration of the sample. The instrument should show a meter indication. The analyzer is designed for extended operation and may be left switched ON continuously.
2.5. Zero and Span Calibration Frequency The zero and span levels should be checked and/or calibrated daily or as often as mandated.
Supplemental Manual Oxygen- V7.1.doc 2-8
Section 3 Model 100/200/300 Supplement
3. ADJUSTMENTS CHECKS AND REPAIRS Note: A millivolt generator is required for this procedure.
3.1. Electrical Zero Adjustments a) Disconnect the plug (P3) from J3 on the printed circuit board (PCB). (P3 is on the end of
the cable assembly coming from the oxygen sensor.)
b) Connect a millivolt generator between pin no. 1 (-) and pin no. 2 (+) at J3 on the PCB.
c) Adjust simulator for 000.00 mv ± 0.01 mv input as measured at TP-8 (+) and common (-).
d) Adjust Front panel zero potentiometer for 000.00 mv ± 0.01 mv as measured between TP-9 and common
e) Set front-panel range switch to Range 1.
f) Adjust R8 for 000.00 mv ± 0.01 mv as measured at TP-4.
g) Adjust R12 for 0.000 VDC ± 0.001 VDC as measured between the 0-10 VDC (+) output terminal and common.
h) Adjust R20 for 4 mA ± 0.01 mA across the 4-20 mA output terminals.
i) Verify O2 display reads 000.0 for all ranges.
j) Repeat steps a) through j) until no further adjustment is required.
3.2. Electrical Span Adjustments Note: Electrical zero adjustments must be made before electrical span adjustments. a) Adjust the millivolt generator and the front-panel span-potentiometer (as necessary) for
exactly 50.00 mv as measured between TP-9 and common.
b) Set front-panel range switch to range 1.
c) Set O2 display selector switch (SW2) to percent (%) O2 mode. Display should read 5.0.
d) Adjust R4 for 1.000 VDC ± 0.001 VDC as measured at TP-4.
e) Set range switch to Range 2 and adjust R3 for 0.500 VDC ± 0.001 VDC at TP-4.
f) Set range switch to Range 3 and adjust R32 for 0.200 VDC ± 0.001 VDC at TP-4.
g) Set range selector to range 1 and verify 1.000 VDC ± 0.00 VDC at TP-4.
h) Adjust R33 for 10.00 VDC ± 0.01 VDC as measured across the 0-10 VDC output terminals.
i) Adjust R24 for 20 mA ± 0.01 mA as measured between the 4-20 mA output terminals. at TP-4
j) Repeat steps a) through j) until no further adjustment is required.
Note: the voltages given in steps e) and f) assume the analyzer is configured for the standard ranges of 0-5, 10, and 25% O2. Consult the factory for the correct voltages required for other non-standard range configurations.
Supplemental Manual Oxygen- V7.1.doc 3-9
Section 3 Model 100/200/300 Supplement
3.3. The Oxygen Sensor The output of the oxygen sensor is a DC output of approximately –10 millivolts at 20.9 % O2. As the sensor is used, it gradually becomes consumed and its voltage output slowly diminishes towards 0 millivolts. Periodic zero and span calibration should be preformed to insure accurate oxygen measurement over the sensor’s life span. Whenever the front panel span potentiometer runs out of adjustment capability it will be necessary to perform a coarse-span-adjustment.
3.3.1. Coarse-Span-Adjustment a) Lift the cover from the analyzer by first removing the retaining screws that secure the cover
to the chassis.
b) Adjust the front panel potentiometer to its mid-setting. The span dial should indicate 5.0.
c) Flow N2 into the analyzer at a flow rate of 1 L/MIN (± 0.5). After the reading stabilizes, adjust the front panel zero potentiometer for an indication of 0.0% O2.
d) Flow an oxygen span gas into the analyzer at a flow rate of 1.0 L/MIN (± 0.5).
e) Locate R27 on the printed circuit board and adjust it as required to obtain a digital display that is equivalent to the value of the span gas being used.
3.3.2. Sensor Checks a) Connect a DC voltmeter to TP-8 (+) and TP-2 common.
b) Introduce a span gas of 21%
c) The measured voltage at TP-8 should be between –3 mv and –13 mv.
d) Whenever the measured voltage is less than –3 mv (or whenever the signal stability becomes an issue) the sensor should be replaced.
Note: A new sensor has a specification of –10 mv (± 3 mv) when measuring 21% O2 at sea level. A correction factor (cf) must be applied to this specification whenever the measuring site is at higher elevations.
mercuryofincheshgWherehg
hgpressurebarometricsitecfWherecfmvoutputCorrected
=
=×−=
""30
)(":10
3.3.3. Sensor Replacement a) Refer to section 2.1 of this manual for selection of the proper type of replacement sensor
b) After replacing sensor it will be necessary to perform a coarse-span-adjustment (refer to paragraph 5.3.1).
Supplemental Manual Oxygen- V7.1.doc 3-10
Section 4 Model 100/200/300 Supplement
4. Chemical Fuel Cell Sample Flow (Supplemental)
O2 SENSOR
NDIR SAMPLE CELL(s)
OUT
P1
IN
NV1
MODEL 100FWITH OPTIONAL FLOWMETER & SAMPLE PUMP
Fl1
MF140
SAMPLEINLET
OUT
IN
NV1MF140
SAMPLEINLET
MODEL 100FWITH OPTIONAL FLOWMETER & SAMPLE PUMP
O2 SENSOR
Fl1
4. IN-LINE FILTER (MF1) CUSTOMER SUPPLIED.
3. INLET & OUTLET CONNECTIONS ARE 1/4” OD TUBE.
2. SAMPLE FLOW RATE: 0.5 L/min to 2.0 L/min.
1. SAMPLE PRESSURE: 1-5 PSIG (NOT REQUIRED WITH OPTIONAL SAMPLE PUMP).
UNLESS OTHERWISE SPECIFIED.NOTES:
NDIR SAMPLE CELL(s)
Figure 4-1 Chemical fuel cell sample flow
Supplemental Manual Oxygen- V7.1.doc 4-11
Section 5 Model 100/200/300 Supplement
5. Electrical Drawings Figure 5-1 100F O2 analyzer wiring diagram
Figure 5-2 100F O2 electrical schematic
Figure 5-3 100F O2 PCB component locator
Supplemental Manual Oxygen- V7.1.doc 5-12
Section 5 Model 100/200/300 Supplement
FRO
NT
PAN
ELPA
NE
L M
ETE
RJ1
0P
101 2 3 4 5 6 7 8 9 10
HI
LOG
ND
CO
MH
OLD 10
3
102
101
0V +5V
DE
CIM
ALPO
INT
POW
ER
10 C
ND
CT
RIB
BO
N C
ABLE
RA
NG
E S
WIT
CH
13
57
92
46
810
S4
R1 R
2 R3
RM
T
1211 10 98
7
23 4 5
61
AC
VI WH
BL
GR
GR
RD
RD
ZER
OR
38
YL BR
GYP
21 2 3 4 5 6 7 8
RA
NG
E
J2
20k Ω
RM
T
P4
J4
MET
ER
1 2 3 4 5 6
BR
CW
CC
W S
GY
OR
BL
YL
CW
CC
W
SPA
NR
3710
0k Ω
S
PU
MP
SW
ITC
H (O
PTI
ON
AL)
ZER
O/S
PAN
S5
NC
NC
CC
+
BK
BK
GR
RD
NO
NO
P8
J8 1 2P
UM
P S
WIT
CH
P9
J9
PUM
P L
ED
1 2
P5 J5
P3
BKRD
J3S
EN
SO
R1
2
1 2 3J1P
1
115
VAC
12
3
SE
NS
OR
BK
WH
GR
OXY
GEN
AN
ALY
ZER
PC
B11
0010
J61 2 1 2 3 4
PU
MP
OU
TPU
TS
BK
P7
J7PU
MP
(OPT
ION
AL)
GR
BK
P6
BK
BR
RD
OR
YL GY
BR
RD
GROU
TPU
T C
ON
NEC
TOR
CH
ASSI
S G
ND
GR
GR
0–10
VD
C
4–20
mA
1 2 3 4 5 6 7 8 9 10 11 12R
MT
J10
+ – + – R2
R3
GN
D+1
5VG
ND
REA
R P
AN
ELPO
WE
R E
NTR
Y M
OD
ULE
H G N
FJ
F1 1AS
3
GC
EA
DH
B
Figure 5-1 100F O2 analyzer wiring diagram
Supplemental Manual Oxygen- V7.1.doc 5-13
Section 5 Model 100/200/300 Supplement
2
11 3
F
J1
115V
AC
R1
1 M
EG
HO
TN
EU
TRA
LG
RO
UN
D2
F
J7 21
C28 .01
200V
SH
OW
N IN
115
VAC
PO
SIT
ION
SW
1
1 3 6 4
2 5
J8 1 2
+15V
17
T114
A–1
0–36
43 6
109 12
F
D1
21
4 3
+–
F
C3
.1 C4
.1
C1
330
UF
50V
C2
100
UF
50V
+ +
1 2
C5
.1 C6
.1
TP
3
TP
4
R28
1.5K
R29
1.5K
U1
LM34
0AT–
15T
P1
3 3
V1
V0
G N D G N D
LM12
0T–1
5U
2
V1
V0
CW
R3
20K
R2
162K
CW
R5
91K
R4
20K
F
–15V
+15V
CW
R8
10K
C25
.1R
92K
R30
6.8K
1 6U
6
84
+–2 3
R6
10K
R7
10K
1 2 3
4
R31
30K
K1
+15V
F
C9
.17
OP
07
F+1
5V
+15V
CW
C17
.1
R22
6.81
KR
24 2KD
4LM
336–
5.0
U3
LM78
05C
T
V1
V0
G N DC
7.11
TP2
C8
.12
3
F
–15V
+15V
CW
18K
R11
R12
10K
C26
.1C
W R33
100K
LED
1
LED
2
GR
EE
N
YE
LLO
W
+5V
+15V
–15V 1 2 3 4J5
FO
UTP
UTS
1 6U
7
84
2 3+–
C10
.17
OP
07
F
F
+15V
R10 2K
C11
.1
C12
.1C
11.1
F
–15V
18
6U
82 3
+–
74
OP
07
C13
.1
F
CW
R13 402
R14
499
R15 100
TP5
TP6
FR
218.
2K
12
J5
R20 2K
M E T E R
1 2 3 4 5 6 7 8 9 10
F
R23 3K
F+1
5V+1
5V
F
R17 33
R34
124
C16 .1
+15V
67
8
1
1
2
OU
T
A 1 O U T A 2 O U T
O F A D J
O F A D J
+ V C C – V C CC O M
– IN
GS
EN
1
GS
ET
1
GS
ET
2
GS
EN
2
+ IN
3 4 5 10 11 12
R18 1K 1%
A D JV
OV
I2 C
24 .1
U9
LM31
7T
3
C14 .1
FF
R19
10K
1%
ZER
O/S
PAN
9U
10IN
A10
1C
11.1
1 41 3
F
F
–15V
D5
5.6V
+15V
+5V
F
F
J4
C18
.1
F
1 2 3 4 5 6
CW CT
CC
W
FF
TP
9C
22.1
C23
.01
C21
.1
500K
R27
F
F
7
OP
07
C20
.1
+15V
R26
10K
1 6U
11
84
2 3+–
J3
FC19
.1
–15V
TP8
R25
10K
F
2
DR
OD
RR
DR
1R
TN
IGN
D+1
5VR
OR
1
J9
F
J2R
AN
GE
+15V
R35
1.5K
1 2 3 4 5 6 7 81 2
F
K2
D2
1N40
02
D3
1N40
02
CW
R32
10K
1 24
3
02FS
%02
1 3
TP1
SW
22
120 Ω
Figure 5-2 100F O2 electrical schematic
Supplemental Manual Oxygen- V7.1.doc 5-14
Section 5 Model 100/200/300 Supplement
J111
5
VAC
115 VAC230 VAC
31
5W1
T1
TP2
TP1
TP3
LED
1LE
D2
R28
R1
R29
C8
C7
C6
U3
C24
C14
C10
C25
U9
U7
R33
R30
R9
R6
R25
R26
J3
1 1 6
2
2
J2J5
81
K2
K1
2
SEN
SO
R
J4ZE
RO
/SPA
N
C21
C23
C20
C22
C9
C19
U11
U6
R34
R11
R10
R12
1
4
C26
TP5
TP4
TP6
TP7
TP8
TP9
J6OUTPUTS
R19
R14
R16
R8
R7 R4
R3
R5
R2
D2
D5
D3
R32
R27
R31
OXY
GEN
FU
EL C
ELL
4101
75
RAN
GE
MET
ER
C15
C18
R24
R23
R13
C13
C16
C12
R18
U8
C11
R20
C5
C17
R22
R21
D4
U1
C3
C4
U2
C2
U10
R35
C1
J9J8
J7
2
21
I +
+
21
1
C28
C27
17
39 10 12
4 6
D1
SW2
FS
Figure 5-3 100F O2 PCB component locator
Supplemental Manual Oxygen- V7.1.doc 5-15
Section 6 Model 100/200/300 Supplement
Part 2 – Oxygen by Paramagnetic Oxygen Sensor Techniques
Supplemental Manual Oxygen- V7.1.doc 6-16
Section 6 Model 100/200/300 Supplement
6. PARAMAGNETIC OXYGEN OPTION
6.1. Description The Paramagnetic Oxygen detector is a thermostated device designed primarily for but not necessarily limited to stationary use. It is a ‘19” rack mount analyzer that is also suitable for bench top use. The operation of the analyzer is based upon the principle of the magneto-dynamic oxygen cell, which is the most accurate and reliable cell for determining the oxygen content of a gas mixture from 0-100 volume percent oxygen.
Warning: This is a general-purpose analyzer, not suitable for hazardous areas. High-pressure oxygen is very dangerous. Virtually any material will burn in it, possibly explosively. It is essential that any persons using this analyzer are aware of the dangers of oxygen, and take all appropriate precautions.
6.2. Product Specifications Paramagnetic Oxygen Detector
SAMPLE CONTACT MATERIAL: Platinum. glass, stainless steel and Viton, Teflon* and Tygon**
DISPLAY: 3 ½" digit panel meter
RANGES: Standard fixed ranges, either A or B or C
OUTPUTS: 0 to 10 VDC and 4 to 20 mA (0 to 20 mA)
A) Range 1: 0 to 1%; Range 2: 0 to 15%; Range 3: 0 to 25%
B) Range 1: 0 to 5%; Range 2: 0 to 10%; Range 3: 0 to 25%
C) Range 1: 0 to 25%; Range 2: 0 to 40%; Range 3: 0 to 100%
RESPONSE TIME: 90% full scale in 2 seconds
AMBIENT TEMPERATURE: 5 to 45° C
SAMPLE TEMPERATURE: 0 to 50° C
SAMPLE CONDITION: Particles < 1µ, non-corrosive dry gas
FITTINGS: ¼" tube
NOISE: Less than 1% full scale SAMPLE FLOW RATE: 0.5 –2.0 L/MIN
LINEARITY: Better than 1% full scale Power Requirements: 115/230 (± 10%) VAC, 50/60 Hz, 70 watts/channel
REPEATABILITY: Better than 1% full scale Relative Humidity: Less than 90% R.H.***
ZERO SPAN DRIFT: Less than 1% full scale in 24 hours
ZERO & SPAN ADJUSTMENT: Ten turn potentiometer
Specifications subject to change without notice *Teflon is a registered trademark of DuPont.
**Tygon is a registered trademark of the Norton Performance Plastics Corporation
***Non-condensing
Supplemental Manual Oxygen- V7.1.doc 6-17
Section 6 Model 100/200/300 Supplement
6.3. Principle of operation The paramagnetic susceptibility of oxygen is significantly greater than that of other common gases, and consequently, the molecules of oxygen are attracted much more strongly by a magnetic field than the molecules of the other gases. Most of the other gases are slightly diamagnetic which means that their molecules are then repelled by a magnetic field.
Figure 6-1 Magnetic Susceptibility of gases
The principle of the magneto-dynamic cell is based upon Faraday's method of determining the magnetic susceptibility of a gas. The cell consists of two nitrogen-filled quarts spheres arranged in the form of a dumbbell. A single turn of platinum wire is placed around the dumbbell that is suspended in a symmetrical non-uniform magnetic field.
Figure 6-2 The Measuring cell in theory
Supplemental Manual Oxygen- V7.1.doc 6-18
Section 6 Model 100/200/300 Supplement
When the surrounding gas contains oxygen, the dumbbell spheres are pushed out of the magnetic field by the change in the field that is caused by the relatively strong paramagnetic oxygen. The torque acting on the dumbbell will be proportional to the paramagnetism of the surrounding gas and, therefore, it can be used as a measure of the oxygen concentration.
The distortion of the dumbbell is sensed by a light beam and projected on a mirror attached to the dumbbell whereof it is reflected to a pair of photocells. When both photocells are illuminated equally, the output will be zero. The output from the photocells is connected to an amplifier, which in turn is fed to the feedback coil of the measuring cell. If the oxygen content of the gas sample changes, the corresponding current output of the amplifier, which is proportional to the oxygen content, produces a magnetic field in the feedback coil opposing the forces and thereby causing the dumbbell to rotate.
Figure 6-3 Principle of operation
Since the feedback current from the amplifier is proportional to the oxygen content of the gas sample, the output signals that are produced by the amplifier will be accurate and linear. The paramagnetic susceptibility of oxygen varies inversely as the square of the absolute temperature. To provide compensation for changes in analyzer temperature, a temperature sensitive element in contact with the measuring cell assembly, is included in the feedback current circuit.
Supplemental Manual Oxygen- V7.1.doc 6-19
Section 7 Model 100/200/300 Supplement
7. Preparations for Operation Warning: This is a general-purpose analyzer, not suitable for hazardous areas. High-pressure oxygen is very dangerous. Virtually any material will burn in it, possibly explosively. It is essential that any persons using this analyzer are aware of the dangers of oxygen, and take all appropriate precautions.
7.1. Power On Turn ON the power switch on the rear panel. The digital panel meters should illuminate. Allow the instrument to warm up for approximately one hour. It is preferable, but not essential, that zero gas flow through the instrument during warm-up.
0
I
Figure 7-1 AC Power Switch, Connector, and Fuse
Note: DO NOT introduce the sample gas UNTIL the analyzer has warmed-up. This will help prevent condensation from forming in the sample cell.
7.2. Zero Adjustment After the one-hour warm-up period, flow zero gas (see Section - Gases) through the instrument at a rate of about 1 liter/min. Adjust the zero control on the front panel until the digital meter (or analog output) is exactly at zero. To achieve final stability, the analyzer may require some additional warm-up period/up to four hours (depending on variables in the analyzer's environment).
7.3. Span Adjustment Flow span gas through the instrument at about 1 liter/min. Adjust the span control on the front panel until the digital meter or analog output is reading the value corresponding to the span gas concentrations.
Note: Span gas concentration should not be less than 80% of the range to be spanned. Note: On the 0-25% range of the analyzer ambient air may be used as span gas. While ambient air is flowing to the analyzer, adjust the span potentiometer to 20.9% O2.
7.4. Zero and Span Calibration Frequency
The zero and span levels should be checked and/or calibrated daily or as often as mandated.
Supplemental Manual Oxygen- V7.1.doc 7-20
Section 8 Model 100/200/300 Supplement
8. Start-Up & Routine Operation Prepare and check the sample system. Adjust the flow of sample gas to about 1 L/min. The instrument should show a meter indication. The paramagnetic oxygen analyzer is designed for extended operation and may be left switched ON continuously.
8.1. Sampling System Note: High-pressure oxygen is very dangerous. Virtually any material will burn in it, possibly explosively. It is essential that any person using this analyzer is aware of the dangers of oxygen, and take all appropriate precautions. The analyzer's sampling system consists of:
1. An internally mounted in line particulate filter
2. A sample pump and flow meter (optional)
3. A sample capillary that controls the sample flow rate to the sensor at 0.2 LPM (or 0.5 depending upon model of O2 sensor.
4. A precision controlled relief valve.
The relief valve maintains a constant inlet pressure to the sample capillary and reduces response time by providing a bypass loop to the exhaust for excess sample.
The analyzer is designed to measure a clean dry sample gas that has been conditioned to remove moisture to prevent condensation in the analyzer. Some applications may require additional sample conditioning, dependent upon the specifications of the sample gas to be measured.
4.IN-LINE FILTER (MF2) CUSTOMER SUPPLIED.
3.INLET & OUTLET CONNECTIONS ARE 1/4” OD TUBE.
2.SAMPLE FLOW RATE: 0.5 L/min to 2.0 L/min.
1.SAMPLE PRESSURE: 3-5 PSIG (NOT REQUIRED WITH OPTIONAL SAMPLE PUMP).
UNLESS OTHERWISE SPECIFIED.NOTES:
SAMPLEINLET
SAMPLEINLET
IN
MF240µ
IN
MF240µ
OUT
OUT
RV1
RV1
MF15µ NV1
P1 C1O2 SENSOR
Fl1
MODEL 100PWITH OPTIONAL FLOWMETER & SAMPLE PUMP
MF15µ NV1
C1O2 SENSOR
Fl1
MODEL 100PWITH OPTIONAL FLOWMETER
NDIR SAMPLE CELL(s)
NDIR SAMPLE CELL(s)
Figure 8-1 Paramagnetic Oxygen Sample flow
Supplemental Manual Oxygen- V7.1.doc 8-21
Section 8 Model 100/200/300 Supplement
8.2. Cross sensitivity of gases The paramagnetic measuring principle is based on the very high magnetic susceptibility of oxygen. In comparison to oxygen, other gases have such a minor susceptibility that most of them are insignificant. Exceptions to this are the nitrogen oxides. However, as this gas is in most cases present in a very low concentration, the error is still negligible.
8.2.1. Example 1 The residual oxygen percentage is measured in a closed carbon dioxide (CO2) atmosphere. The "zero calibration" is done by means of nitrogen (N2).
According to the list of cross-sensitivities, the error for 100 % CO2 at 200°C is -0.27%. In order to obtain a higher accuracy, this means for the zero calibration that the reading should be adjusted at +0.27% with N2, in order to compensate the error of CO2.
Since the values of cross-sensitivities are based on 100 Volume percentage of that particular gas, the error at 50% by volume CO2 and 50% by volume N2 is - 0.135%.
8.2.2. Example 2 Given the following gas composition at a temperature of 20° C:
5% volume Oxygen (O2) +100.00 x 10-2 x 5 = +5.0000
40% volume Carbon Dioxide(CO2) -0.27 x 10-2 x 40 = -0.1080
1% volume Ethane(C21-14) -0.43 x 10-2 x l = -0.0043
54% volume Nitrogen (N2) 0.00 X 10-2 x 54 = 0.0000
Gives a reading (%by volume) of: +4.8877
As this example shows, the total error (5.000 minus 4.8877) is - 0.1123.
Supplemental Manual Oxygen- V7.1.doc 8-22
Section 8 Model 100/200/300 Supplement
Table 8-1 Cross Sensitivity of gases
All values based on nitrogen 0% / oxygen 100% Gas Formula 20oC 50oC
Argon Ar -0.23 -0.25 Acetylene C2H2 -0.26 -0.28 Acetone C3H60 -0.63 -0.69 Acetaidehyde C2H4O -0.31 -0.34 Ammonia N3 -0.17 -0.19 Benzene C6H4 -1.24 -1.34 Bromine Br2 -1.78 -1.97 Butadiene C4H6 -0.85 -0.93 Isobutylene (CH3)2CH=CH2 -0.94 -1.06 n-Butane C4H10 -1.10 -1.22 Chlorine CL2 -0.83 -0.91 Hydrogen Chloride HCL -0.31 -0.34 Nitrous Oxide N2O -0.20 -0.22 Diacetylene (CHCl)2 -1.09- -1.20 Ethane C2H4 -0.43 -0.47 Ethylene Oxide C2H4O2 -0.54 -0.60 Ethylene C2H4 -0.20 -0.22 Ethylene Glycol CH2OHCH2OH -0.78 -0.88 Ethylbenzene C8H10 -1.89 -2.08 Hydrogen Fluoride HF +0.12 +0.14 Furan C4H40 -0.90 -0.99 Helium He +0.29 +0.32 n-Hexane C6H14 -1.78 -1.97 Krypton Kr -0.49 -0.54 Carbon Monoxide CO -0.06 -0.07 Carbon Dioxide CO2 -0.27 -0.29 Methane CH4 -0.16 -0.17 Methanol CH4O -0.27 -0.31 Methylene Chloride CH2Cl2 -1.00 -1.10 Neon Ne +0.16 +0.17 n-Octane C8H18 -2.45 -2.70 Phenol C6H6O -1.40 -1.54 Propane C3H8 -0.77 -0.85 Propylene C3H6 -0.57 -0.62 Propene CH3CH=CH12 -0.58 -0.64 Propylene Oxide C3H6O -0.90 -1.00 Propylene Chloride C3H7Cl -1.42 -1.44 Silane SiH4 -0.24 -0.27 Styrene C7H6=CH2 -1.63 -1.80 Nitogen N2 -0.00 -0.00 Nitrogen Monoxide NO +42.70 +43.00 Nitrogen Dioxide NO2 +5.00 +16.00 Oxygen O2 +100.00 +100.00 Sulfur Dioxide SO2 -0.18 -0.20 Sulfur Fluoride SF6 -0.98 -1.05 Hydrogen Sulfide H2S -0.41 -0.43 Toluene C7H8 -1.57 -1.73 Trichloroethylene C2HCl3 -1.56 -1.72 Vinyl Chloride C2H3CI -0.68 -0.74 Vinyl.Fluoride CH3F -0.49 -0.54 Water H2O -0.03 -0.03 Hydrogen H2 +0.23 +0.26 Xenon Xe -0.95 -1.02
Supplemental Manual Oxygen- V7.1.doc 8-23
Section 9 Model 100/200/300 Supplement
9. Mechanical Zero Adjustment (Not required with some sensor models) This adjustment depends upon sensor model and may be periodically required over the life span of the analyzer, or whenever the front panel zero adjustment has reached its limit. Note: Always check for gas leaks prior to making this adjustment, especially when the zero potentiometer is at its counter-clockwise limit. a) Place the front panel zero adjustment to its mid-setting (5.0 on the dial.)
b) Introduce N2 into the analyzer at a flow rate of 0.5-2.0 L/MIN.
c) Remove the screws securing the top to the chassis and lift off the cover of the analyzer
d) Locate and remove the rubber ‘boot’ that covers the O2 sensor.
e) Loosen the locking screw that secures the adjusting screw for positing the photo-sensor on the detector assembly. Only loosen the locking screw enough to permit the necessary adjustment (see Figure 8-1.)
f) Turn the adjusting screw as required to give an indication of approximately 0.00 on the digital display.
g) Carefully tighten the locking screw and replace the rubber ‘boot’ removed in step d).
h) Observe the digital display and turn the front panel zero adjustment to achieve reading of 0.00 in the display.
i) The zero adjustment should be between 4.0 and 6.0 on its dial. If not, repeat steps d) though h) until no further adjustment is required.
Adjusting Screw
Locking Screw
Loosen
Photo DetectorsCircuit Board
Photo DetectorMounting Block
Adjust
Figure 9-1 Oxygen sensor adjustment
Supplemental Manual Oxygen- V7.1.doc 9-24