Stereoscopic Analysis of Trabecular Bone Orientation in ...

9
Cells and Materials Cells and Materials Volume 2 Number 1 Article 2 1991 Stereoscopic Analysis of Trabecular Bone Orientation in Proximal Stereoscopic Analysis of Trabecular Bone Orientation in Proximal Human Tibias Human Tibias K. N. Bachus VA Medical Center, Salt Lake City M. K. Harman VA Medical Center, Salt Lake City R. D. Bloebaum VA Medical Center, Salt Lake City Follow this and additional works at: https://digitalcommons.usu.edu/cellsandmaterials Part of the Biological Engineering Commons Recommended Citation Recommended Citation Bachus, K. N.; Harman, M. K.; and Bloebaum, R. D. (1991) "Stereoscopic Analysis of Trabecular Bone Orientation in Proximal Human Tibias," Cells and Materials: Vol. 2 : No. 1 , Article 2. Available at: https://digitalcommons.usu.edu/cellsandmaterials/vol2/iss1/2 This Article is brought to you for free and open access by the Western Dairy Center at DigitalCommons@USU. It has been accepted for inclusion in Cells and Materials by an authorized administrator of DigitalCommons@USU. For more information, please contact [email protected].

Transcript of Stereoscopic Analysis of Trabecular Bone Orientation in ...

Page 1: Stereoscopic Analysis of Trabecular Bone Orientation in ...

Cells and Materials Cells and Materials

Volume 2 Number 1 Article 2

1991

Stereoscopic Analysis of Trabecular Bone Orientation in Proximal Stereoscopic Analysis of Trabecular Bone Orientation in Proximal

Human Tibias Human Tibias

K. N. Bachus VA Medical Center, Salt Lake City

M. K. Harman VA Medical Center, Salt Lake City

R. D. Bloebaum VA Medical Center, Salt Lake City

Follow this and additional works at: https://digitalcommons.usu.edu/cellsandmaterials

Part of the Biological Engineering Commons

Recommended Citation Recommended Citation Bachus, K. N.; Harman, M. K.; and Bloebaum, R. D. (1991) "Stereoscopic Analysis of Trabecular Bone Orientation in Proximal Human Tibias," Cells and Materials: Vol. 2 : No. 1 , Article 2. Available at: https://digitalcommons.usu.edu/cellsandmaterials/vol2/iss1/2

This Article is brought to you for free and open access by the Western Dairy Center at DigitalCommons@USU. It has been accepted for inclusion in Cells and Materials by an authorized administrator of DigitalCommons@USU. For more information, please contact [email protected].

Page 2: Stereoscopic Analysis of Trabecular Bone Orientation in ...

CELLS & MATERIALS, Vol. 2, No. 1, 1992 (Pages 13-20) 1051 -6794/92$3. 00 + . 00 Scanning Microscopy International , Chicago (AMF O'Hare), IL 60666 USA

STEREOSCOPIC ANALYSIS OF TRABECULAR BONE ORIENTATION IN PROXIMAL HUMAN TIBIAS

K.N. Bachus,* M.K. Harman, and R.D. Bloebaum

Bone & Joint Research Laboratories (151F) VA Medical Center, 500 Foothill Blvd.

Salt Lake City, UT 84148 U.S.A.

(Received for publication November 19, 1991, and in revised form March 9, 1992)

Abstra~t

The three-dimensional orientation of trabeculae is a key factor in determining the load carrying capabilities of cancellous bone. Previous biomechanical studies have shown that proximal tibias resected parallel to the articulating surface are stronger and stiffer than the contralateral tibias resected perpendicular to the long axis of the bone. However, morphologic evidence was not provided to help explain the mechanical differences.

To determine the orientation of the trabeculae in the medial condyle for both parallel cut and perpendicular cut specimens, a scanning electron microscope and stereoscopic techniques were used. Data showed that tibias cut parallel to the articular surface had trabeculae oriented nearly vertical with a mean angle of 4.5° ± 14.7° (range, 0° to 56.3°). The contralateral tibias cut perpendicular to the long axis of the tibia had trabeculae oriented at a mean angle of 36.0° ± 12.2° (range, 16.1° to 67.4 °) from vertical. The differences between the two resection techniques were shown to be significant (p s. 0.01) using an Analysis of Variance.

This study provided morphologic evidence to explain why previous specimens cut parallel to the articular surface had stronger and stiffer cancellous bone than the contralateral specimens cut perpendicular to the long axis of the tibia. This information is important in understanding the load carrying capabilities of cancellous bone and how it may be applied to improving the clinical results of primary total knee arthroplasty.

Key Words: 3-D stereoscopy, trabecular orientation, cancellous bone, total knee arthroplasty, joint replacement, tibia, bone morphology, parallax, resection technique, scanning electron microscope

*Address for correspondence: Kent N. Bachus Bone & Joint Research Laboratories (151F) VA Medical Center, Salt Lake City, UT 84148 U.S.A.

Phone : (801 )-582-1565, Ext. 4696

13

Introduction

It has been suggested that the trabeculae in weight bearing regions of cancellous bone develop along the axes of principal stress [9, 12, 26, 28, 29, 39]. This contributes to the structural anisotropic (direction dependent) properties of human cancellous bone [11, 34, 38]. Studies have shown that the stiffness and strength of cancellous bone is dependent upon the orientation . of the trabeculae [5, 11, 29, 34, 38]. Therefore, trabecular orientation is a key factor in determining the load carrying capabilities of cancellous bone.

A recent investigation by Hofmann, et al. [19], determined that the orientation of the tibial resection during total knee arthroplasty significantly affects the load carrying capabilities of the underlying subchondral bone. Clinical evidence revealed that the tibias resected parallel to the anatomic posterior slope of the articular surface (surface-parallel) during primary total knee arthroplasty experienced a lower incidence of tibial component subsidence when compared to tibias resected perpendicular to the long axis of the bone (axis-perpendicular) (Figure 1). Results of mechanical testing indicated that the surface­paraJlel resected bone exhibited 40% greater load carrying capacity and 70% greater stiffness than axis­perpendicular resected bone. It was speculated by the investigators [ 19] that the trabecular orientation may be a factor in the mechanical differences. However, morphological data was not provided to explain the observed differences in the mechanical properties of the cancellous bone.

To understand the mechanical properties of cancellous bone, various attempts have been made to analyze the trabecular structure [5, 7, 11, 29, 32, 36, 38]. However, three dimensional structural analysis is difficult due to the anisotropy of cancellous bone [5, 10, 35, 37]. Pugh, et al. [29] recognized that trabeculae are not all oriented in ideal columns. From studies using longitudinal thin sections through human knees, they [29] observed that some trabeculae had an angled orientation, but most trabeculae were oriented normal to the articular surface of the bone . . Takechi [32] used

Page 3: Stereoscopic Analysis of Trabecular Bone Orientation in ...

K.N. Bachus, Ph.D., M.K. Harman, M.S. and R.D. Bloebaum, Ph .D.

Articular Surface

Anterior

Surface-Parallel Axis-Perpendicular

Figure 1. Diagram representing two angles currently used to surgically resect the proximal tibias in total knee replacement.

microradiographs of the proximal tibia to relate trabecular arrangement to mechanical function. While quantitative results were not given, it was reported that the trabeculae oriented perpendicular to the articular surface of the tibial condyles help support the compressive loads from the condyles of the femur. [32] A crossing trabecular pattern near the tibial tuberosity was also observed, serving to resist the tension forces from the hamstrings and the quadriceps muscles attaching to the proximal tibia [32].

Mechanical testing by other investigators [1, 2, 13, 24] has shown that the anteromedial condylar region of the bone found in the proximal tibia is stronger than the posteromedial and lateral condylar regions when tested in compression. However, the morphologic analysis related to these studies was either not reported, or it was restricted to qualitative descriptions from two-dimensional images of the bone. Williams, et al. [38] reported a qualitative description of trabeculae located beneath the articular surface of the proximal human tibia. They [38] found that the trabeculae were oriented in vertical columns perpendicular to the surface, but when viewed in the transverse plane, the plate and rod formation of the trabeculae appeared quite curved. This curvature in the horizontal plane minimizes the Euler bending [8, 33] of the trabeculae under axial compression, contributing to an increased mechanical strength [38]. The difficulties of accurately measuring three dimensional structures from a single test plane, and the lack of quantitative morphologic analyses reported with mechanical testing support the need for an accurate and efficient three-dimensional analysis technique. Stereoscopy

Two methods have been cited in the literature for generating stereomicrographs using the scanning electron microscope (SEM). These methods allow for visualization of complex three dimensional structures, and ultimately the measurement of trabecular orientation. The Shift Method involves photographing the specimen at two different positions in the field after laterally displacing the specimen [20, 27]. While this method uses simple calculations for analysis, it is limited by the amount of overlap available for each

14

micrograph [20, 27]. Consequently, at magnifications greater than 50x, sufficient overlap of the image is not obtainable (20, 27]. The Tilt Method involves tilting the specimen surface between exposures, providing complete overlap of the stereomicrographs at any magnification [3, 4, 16, 20, 21 , 22, 27]. This proven method was chosen to generate the stereomicrographs for this study.

The purpose of this study was to utilize three dimensional stereoscopic techniques to measure the angle of trabecular orientation in the proximal human tibia resected axis-perpendicular and surface-parallel. This information will help explain the load carrying capabilities of cancellous bone and how the choice of resection angle can improve the clinical results of primary total knee arthroplasty .

Materials and Methods

Specimen Preparation Five matched pair of fresh frozen proximal

human tibia were obtained. All specimens were from individuals who died from either traumatic deaths or from cardiovascular disease. As shown in Table 1, the tibias were from 3 male donors and 2 female donors. The age of the donors ranged from 53 to 64 years (58.4 ± 4.7). The posterior slope of the tibias ranged from 8.5° to 12.0° (10.2° ± 1.8°). None of the bone specimens had any evidence of gross deformities or advanced disease states . Following the osteometric methods published by Ruff et al. [30], the long-axis of the tibia was determined for each specimen. This reference axis was used to control the tibias for reproducible alignment and sectioning.

To prepare the tibias for sectioning, a transverse cut was made through the midshaft region of each tibia using a 0.625 cm (0.25 inch) wide, 0.0625 cm (0.025 inch) thick band saw blade with 6 teeth per cm (15 teeth per inch). The distal half of each tibia was discarded from this study. The distal portion of the midshaft of each proximal specimen was inserted into a 5 cm diameter cylinder fabricated of aluminum. Set screws in the aluminum device were advanced to the cortical surface of the tibia, and hand-tightened with an allen wrench so that no movement of the bone was evident. TABLE 1. Gender, age, and posterior tilt of the

proximal tibial condyles of the five human donors used in this study.

Specimen Gender Donor Age Posterior Tilt Number (years) (degrees)

I Male 60 8.5 2 Female 53 12.0 3 Male 64 8.5 4 Male 54 12.0 5 Female 61 10.0

Mean 58 10.2 Std. Dev. 5 1.8

Page 4: Stereoscopic Analysis of Trabecular Bone Orientation in ...

Measurement of Trabecular Bone Orientation

The proximal tibia was then oriented according to the previously marked reference axis such that either the articular surface of the medial condyle was parallel to the saw blade, or the long axis of the tibia was perpendicular to the saw blade, then clamped in a steel vise. The vise was firmly held to the table of the saw using an electromagnetic chuck (O.S. Walker Company, Inc., Worchester, Massachusetts). Once the specimen was aligned, a cut in the transverse plane was made through the lowest point of the ellipses defined by the articulating surfaces of the tibial condyles of each tibia using a water-cooled bone saw with a 20.3 cm (8 inch) diameter circular saw blade with 4 teeth per cm (10 teeth per inch) (Cleveland Twist Drill Co., Cleveland, OH). The proximal tibia was further resected to a level 6 mm below the initial cut, producing 5 mm thick wafers of cancellous bone.

The angles of these cuts were determined from two main surgical resection techniques currently used by orthopaedic surgeons in total joint replacement [18, 23]. One specimen from each pair was cut parallel to the natural articulating surface of the tibia [18], while the contralateral tibia was cut perpendicular to the long axis of the tibial shaft [23]. To avoid data bias, a coin toss was used to determine which specimen of each matched pair was cut surface-parallel and which was cut axis-perpendicular. This entire process resulted in five paired wafers of bone.

After cutting, the cancellous surface of each specimen was Water Piked (Teledyne, Ft. Collins, CO) to remove marrow and fatty tissue. The specimens were dehydrated using ascending grades of ethyl alcohol from 70% to 100%, and then placed in a vacuum desiccating chamber to dry for over 48 hours. Since the full wafers could not be imaged in the SEM chamber, the 5 mm wafers were sagittally sectioned into thirds and the medial third was used for this analysis (Figure 2). The medial third was chosen due to its large surface area, making it an important region for supporting the loads of the tibial component [24, 25]. Each specimen was then placed into a vacuum sputter coater (Hummer, Model VI-A, Chestnut Hill, MA), where the pressure was reduced to 40 mTorr and a thin layer of gold was sputtered onto the cancellous surface for two minutes. The specimen was removed from the coater and placed onto the specimen stage of the JEOL JSM-330A for imaging. Generating Stereomicrograph Pairs

A tilt goniometer stage on the SEM allowed the specimen to easily be tilted about the X-axis. While there are many angles of tilt used by other investigators [ 17, 22, 27], the 10° used in the current study is recommended by Boyde [3, 4], and Howell [20] . This angle provided the necessary cues to perceive relative depth, and allowed for comfortable viewing with the stereoscope based on the interocular distance of the general population [3, 4, 20].

To generate stereomicrographs using the JEOL secondary electron detector, the specimen was first tilted to -5° with respect to the horizontal. The image

15

Anterior

Posterior Figure 2. Diagram representing the sagittal cuts made in each 5 mm wafer enabling the medial third to be used in analysis.

was then photographed using Type 55 Polaroid film (Polaroid Corp., Santa Ana, CA), at a magnification of 50x, an accelerating voltage of 25kV, a load current of l~O µA, a working distance of 15 mm, and an aperture diameter of 240 µm. After three distinct features were marked on the viewing screen using a wax pencil, the spe~imen was tilted to +5° with respect to the ?onzon.tal. ~nless the area of interest happened to lie m the tilt axis of the specimen stage, translation of the specimen o~curred in both the x and y planes. Using the appropnate SEM stage controls, the specimen was moved so that the image was realigned with the marked ~osition on ~he viewing screen. As the specimen was tilted, the bnghtness and contrast of the image changed due to the slope of the specimen with respect to the electron beam and collection system. After these features were readjusted, the image was photographed a second time. To insure that the second image had the same magnification and rotation as the first image, and to reduce the tilt error, the z height stage control was used to adjust the focus [3, 4, 16, 21]. Viewing and Measuring in the Third Dimension

~ four mirrored Topcon stereoviewer (Ted Pella, Reddmg, CA) was used to view the three dimensional ~mage created from the stereomicrographs. This mstrument has two lenses which are positioned at their focal length above the stereomicrographs and separated by the interocular distance of the observer [ 4 ]. It was placed over a light box so that the negatives of the photomicrographs could be viewed via transmitted light. This provided better detail and decreased potential errors associated with viewing positive prints [4,27).

Care was taken to properly orient the stereomicrographs beneath the stereoviewer to avoid incorrect perception by the observer. Following the convention indicated by Howell, et al. [21], the photomicrograph with the lower goniometer angle (-50) was placed on the left beneath the stereoviewer, and that with the higher angle ( +5°) placed on the right. Failure to follow this convention resulted in an inverted image where elevations appeared as depressions.

To measure the parallax, the difference in vertical displacement of two points, a floating dot parallax bar

Page 5: Stereoscopic Analysis of Trabecular Bone Orientation in ...

K.N. Bachus, Ph.D., M.K. Hannan, M.S. and R.D. Bloebaum, Ph.D.

stereomicrometer (Gordon Enterprises: Ladd Research Industries, Inc., Burlington, VT) was used. This instrument consists of two glass disks, each engraved with a small dot, mounted onto a bar whose length can be varied by means of a screw micrometer. The micrometer is accurate to 0.01 mm, which corresponds to an actual distance of 0.2 µm on the magnified scale of the photomicrographs. To obtain a correct measurement of the parallax, it must be measured perpendicular to the axis of tilt [17, 27]. For this reason, the stereomicrographs were rotated 90 degrees counterclockwise beneath the stereoviewer and the parallax bar rested along the horizontal. The parallax bar was aligned so that each disk was positioned over the stereomicrographs being viewed under the stereoviewer. For each of the five paired specimens, six to nine trabecular tilt angles were calculated from the measured parallaxes and lateral displacements. As seen on the photomicrographs (Figure 3 and 4 ), the orientation of the points chosen to be measured were arranged in a right triangle. Points PO and Pl were located on the cut surface of the trabecular bone, while point P2 was located perpendicular to Pl on the sloped surface of an individual trabeculae. By matching the altitude of the floating dot with various points on the stereomicrographs, their parallax separation could be read from the micrometer. If the two dots were positioned on exactly the same surface as the corresponding points on the stereomicrographs, a point of light was seen lying in the plane of the surface [ 4].

For two points whose parallax readings are Pl and P2, their relative displacement is given by [4, 15, 16, 21, 22, 27]:

h= CPl - P2) (1) 2 * M * sin (8/2)

where M = the magnification of the stereomicrographs and 8 = the difference in tilt angles between stereomicrographs. The height displacement and the corresponding angle of tilt is determined by the location of the image points chosen to be measured. An investigation by Heidenreich, et al. [ 15] showed that the parallax of the two image points needs to be 0.4 mm or greater to produce accurate results. Due to limitations with the optical depth of field of the SEM and the resulting photomicrographs, there was difficulty in distinguishing individual features of the trabeculae at large height displacements. This limited the depth at which the parallax could be measured. Also, the trabeculae were not oriented into ideal columns. Their bases were continuous with the connecting rod and plate formation of cancellous bone [12, 28, 34, 38] creating trabeculae with wide, flared bases. For these reasons, point P2 (Figure 4) was chosen to lie on the sloped surface of the trabeculae and within the focused depth of field, not on the flared bases of the trabeculae. By meeting these conditions with all of the parallax measurements made in this study, the angle of tilt could

16

be accurately calculated. Using a digital micrometer (Mitutoyo, Model 500-

215, Japan), the lateral displacement between the two points was measured from the surface of the photomicrograph with the lower tilt angle. It was determined that the error associated with measuring from the tilted photomicrograph was small when compared to measurements made from a 0° tilted photomicrograph. The micron bar scale on the photomicrographs allowed this lateral displacement value to be converted into real dimensions. Through the use of simple trigonometry, the angle of the slope 0 between the two points could be calculated with the following equation:

0 = tan -1 (I I h) (2)

where h = the calculated height displacement and 1 = the measured lateral displacement in real dimensions. This corresponds to a two dimensional projection of the two points. (Figure 5)

The previous equations assume there is a parallel projection geometry in the SEM, which is nearly the case for pictures taken with magnifications above 500x [4, 20, 21]. While the magnification used in this study was only 50x, work completed by Kristensen, et al. [27] indicated that these equations are accurate at any magnification.

Results

Qualitative stereoscopic analysis showed the trabeculae of the surface-parallel cut specimens oriented nearly vertical (Figure 3), while the trabeculae of the axis-perpendicular cut specimens displayed an obvious tilt (Figure 4 ).

The calculated angles data are shown in Table 2. A mean angle of 4.5° ± 14.7° (range, 0° to 56.3°) was measured for the surface-parallel cut specimens. For the axis-perpendicular cut specimens, a mean angle of 36.0° ± 12.2° (range, 16.1° to 67.4°) was measured. Using an Analysis of Variance [31], it was shown that the measured trabecular orientation from the surface­parallel cutting technique was significantly different from the axis-perpendicular cutting technique (p $. 0.01).

Discussion

This analysis provided preliminary morphologic evidence that can help explain why the cancellous bone of the proximal tibia cut parallel to the anatomic posterior slope of the tibial articular surface was stronger and stiffer than the cancellous bone cut perpendicular to the long axis of the tibia [19]. This stereoscopic analysis demonstrated that the specimens cut parallel to the articular surface had trabeculae predominantly oriented normal to the resected surface. All tibias cut perpendicular to the long axis of the tibia had trabeculae predominantly oriented at an angle to

Page 6: Stereoscopic Analysis of Trabecular Bone Orientation in ...

Measurement of Trabecular Bone Orientation

Figure 3. SEM stereo­micrograph of trabecular bone cut using the surface­parallel technique.

Figure 4. SEM stereo­micrograph of trabecular bone cut using the axis­perpendicular technique.

Figures 3 and 4. The lettered points show the orientation of the points used to measure parallax and angle of trabecular tilt. Points PO and Pl are located on the cut surface of the bone, while point P2 is located on the surface of an individual trabecula.

the resected surface. From basic mechanics theory, Euler buckling will

occur on vertically oriented trabeculae loaded in compression (8, 33]. As the trabeculae are angled further from vertical, the mode of failure changes from Euler buckling to bending, and the load carrying capability decreases. A surface-parallel resection technique would result in trabeculae being oriented more normal to the tibial component in total knee replacement. This would improve the load carrying capabilities of the cancellous bone in the resected proximal tibia.

Methods for determining the angles of trabecular orientation are necessary to understand the role the trabeculae play in the mechanical strength and stiffness of bone. Investigators need to be aware of the morphology of the test specimens since it has been shown that the trabecular orientation significantly influences the stiffness and strength of cancellous bone (2, 5, 11, 28, 29, 34, 38].

The wide range of trabecular orientations measured in this study indicate that the loading pattern [9, 12, 26, 28, 29, 39] in the proximal tibia is complex. Due to the anisotropic structure of cancellous bone, the trabecular orientation may change with anatomic

17

h

Figure 5. Two dimensional projection of two points, showing the trigonometric relationship between the angle of tilt 0 and the height (h) and the length (I).

location. Further investigations are needed to quantify the trabecular orientation as a function of anatomic location across the resected tibial plateau. When revision surgery is being considered, a better understanding of the trabecular orientation distal to the articular surface of the tibia is necessary if trabecular orientation is to help improve implant stability. The application of the know ledge gained in this study can help improve the clinical results of primary total knee

Page 7: Stereoscopic Analysis of Trabecular Bone Orientation in ...

K.N. Bachus, Ph.D., M.K. Hannan, M.S. and R.D. Bloebaum, Ph.D.

arthroplasty. The 3-dimensional techniques applied in this study can further the understanding of the load carrying capabilities of cancellous bone.

Acknowled2ments

This work was supported by the Department of Veterans Affairs Medical Center in Salt Lake City, Utah.

References

1. Bartel DL, Burstein AH, Santavicca EA, Insall JN (1982) Performance of the tibial component in total knee replacement. J Bone Joint Surg. 64A, 1026-1033.

2. Behrens JC, Walker PS, Shoji H (1974) Variations in strength and structure of cancellous bone at the knee. J Biomech.1, 201-207.

3. Boyde A (1970) Practical problems and methods in the three-dimensional analysis of scanning electron microscope images. Scanning Electron Microsc. 1970, 105-112.

4. Boyde A (1973) Quantitative photogrammetric analysis and qualitative stereoscopic analysis of SEM images. J Microsc. 98, 452-471.

5. Brown TD, Ferguson AB (1980) Mechanical property distributions in the cancellous bone of the human proximal femur. Acta Orthop Scand. 21_, 429-437.

6. Carter DB, Hayes WD (1977) The compressive behavior of bone as a two-phase porous structure. J Bone Joint Surg. 59A, 954-962.

7. Cruz-Orive LM, Hunziker EB (1986) Stereology for anisotropic cells: Application to growth cartilage. J Microsc. 143, 47-80.

8. Currey J (1984) The Mechanical Adaptations of Bones. (eds), Princeton University Press, Princeton, NJ, 122-126.

9. Currey J (1984) The Mechanical Adaptations of Bones. (eds), Princeton University Press, Princeton, NJ, 133-157.

10. Elias H (1971) Three-dimensional structure identified from single sections. Science. 17 4, 993-1000.

11. Galante J, Rostoker W, Ray RD (1970) Physical properties of trabecular bone. Calcif Tissue Res.~. 236-246.

12. Gibson LJ (1985) The mechanical behavoiur of cancellous bone. J Biomech.18., 317-328.

13. Goldstein SA, Wilson DL, Sonstegard DA, Matthews LS (1983) The mechanical properties of human tibial trabecular bone as a function of metaphyseal location. J Biomech. J__Q, 965-969.

14. Gundersen HJG, Bendtsen TF, Korba L, Marcussen N, Moller A, Nielsen K, Nyengaard JR, Pakkenberg B, Sorensen FB, Vesterby A, West MJ (1988) Some new, simple, and efficient stereological methods and their use in pathological research and diagnosis. Acta Pathol Microbial Immunol Scand. 90, 379-394.

18

TABLE 2. Results of 38 measured angles from surface-parallel and axis-perpendicular cuts.

ANGLE OF TILT (d e_g_rees Specimen Parallel Perpendicular

Cut Cut 0.0 0.0 24.5 21.7

1 0.0 0.0 36.2 49.3 0.0 54.8 33.7 25.0 0.0 32.1 32.0 0.0 0.0 43.9 67.4

2 0.0 8.4 31.8 24.2 0.0 0.0 34.0 32.8 0.0 0.0 41.9

56.3 50.7 47.6 46.4 3 0.0 0.0 21.1 39.0

0.0 0.0 25.2 40.9 0.0 0.0 42.6 41.8 0.0 0.0 30.0 51.5 0.0 0.0 30.0 45.5

4 0.0 0.0 34.7 33.6 0.0 0.0 63.0 0.0 0.0 0.0 28.3 41.8

5 0.0 0.0 61.9 24.5 0.0 0.0 16.1 23.0

27.3 22.5 Mean 4.5 36.0

Std. Dev. 14.7 12.2 Range 0-56.3 16.1-67.4

n 38 38

15. Heidenreich RD, Matheson LA (1944) Electron microscope determination of surface elevations and orientations. J Appl Phys . .12, 423-435.

16. Hepworth A, Sikorski J (1973) Stereoscopy of cylindrical objects in the scanning electron microscope. J Microsc. 98, 436-451.

17. Hilliard JE (1971) Quantitative analysis of scanning electron micrographs. J Mkrosc. 95, 45-48.

18. Hofmann AA (1988) The Intermedics Natural­Knee™ System. Intermedics Orthopedics™; Austin, TX 78752. 1-27.

19. Hofmann AA, Bachus KN, Wyatt RWB (1991) Effect of the tibial cut on subsidence following total knee arthroplasty. Clin Orthop. 269, 63-69.

20. Howell P (1975) Taking, presenting and treating stereo data from the SEM. Scanning Electron Microsc. 1975, 697-706.

21. Howell PGT, Boyde A (1972) Comparison of various methods for reducing measurements from stereo-pair scanning electron micrographs to "real 3-D data". Scanning Electron Microsc. 1972, 233-240.

22. Hudson B (1973) The application of stereo­techniques to electron micrographs. J Microsc. 98, 396-401.

Page 8: Stereoscopic Analysis of Trabecular Bone Orientation in ...

Measurement of Trabecular Bone Orientation

23. Hungerford DS, Kenna RV (1983) Preliminary experience with a total knee prosthesis with porous coating used without cement. Cl in Orth op. 176, 95-107.

24. Hvid I (1988) Mechanical strength of trabecular bone at the knee. Danish Med Bull. 35, 345-365.

25 . Johnson F, Leitl S, Waugh W (1980) The distribution of load across the knee: a comparison of static and dynamic measurements. J Bone Joint Surg. 62-B, 346-349.

26. Koch JC (1917) The laws of bone architecture. Am J Anat. 2..L 177-298.

27. Kristensen S, Papadimitriou J (1981) The measurement of cell volume and surface area by SEM photogrammetry. J Microsc. 124, 155-161.

28. Pugh JW, Rose RM, Radin EL (1973) Elastic and viscoelastic properties of trabecular bone: Dependence on structure. J Biomech . .Q, 475-486.

29. Pugh JW, Rose RM, Radin EL (1973) A structural model for the mechanical behavior of trabecular bone. J Biomech . .Q, 657-670.

30. Ruff C, Jones H (1983) Cross-sectional geometry of Pecos Pueblo femora and tibiae--A biomechanical investigation: I. Method and general patterns of variation. Am J Phys Anthrop. 60, 359-381.

31. Sokal RR, Rohlf FJ (1981) Biometry. The Principles and Practice of Statistics in Biological Research. Wilson J, Cotter S (eds), W. H. Freeman & Co., New York, 179-453.

32. Takechi H (1977) Trabecular architecture of the knee joint. Acta Orthop. Scand. 48, 673-681.

33. Townsend P, Rose R, Radin E (1975) Buckling studies of single human trabeculae. J Biomech . .8., 199-201.

34. Townsenq PR, Raux P, Rose RM, Miegel RE, Radin EL (1975) The distribution and anisotropy of the stiffness of cancellous bone in the human patella. J Biomech . .8., 363-367.

35. Whitehouse W, Dyson E (1974) Scanning electron microscopy studies of trabecular bone in the proximal end of the human femur. J Anat. ill, 417-444.

36. Whitehouse WJ (1974) The quantitative morphology of anisotropic trabecular bone. J Microsc . . ill.L 153-169.

37. Whitehouse WJ (1975) Scanning electron micrographs of cancellous bone from the human sternum. J Pathol. 116, 213-224.

38. Williams JL, Lewis JL (1982) Properties and an anisotropic model of cancellous bone from the proximal tibial epiphysis. J Biomech Eng. 104, 50-56.

39. Wolff J (1892) The Law of Bone Remodelling (Das Gesetz der Transformation der Knochen, Hirschwald). Springer-Verlag, Berlin, 3-100.

19

Discussion With Reviewers

Anonymous Reviewer 5: Scanning electron microscopy was not required for this investigation and macrophotography would have sufficed. The data from this study could have been obtained from longitudinal cuts only with the two different slopes interpolated from the same specimen. Authors: Experts [10, 14, 35, 37] in the field of stereology have published that it is impossible to measure precisely three dimensional structures from a single plane -- especially with an anisotropic material such as cancellous bone. The best that one could hope for, is a good estimate. To obtain accurate, 3-dimensional data, extensive sectioning of the structure in three planes is required if proper stereological principles are to be followed from single plane images. While the structure could appear to be fairly uniform in one plane of sectioning, it could not be proven that the structure does not alter along other planes. We feel that the SEM was the best imaging technique for this study due to its superb image resolution, the ease of specimen preparation, and the accuracy of the 3D stereoscopic principles that are well documented in the literature.

S. B. Goodman: The authors imply that resection of the tibia parallel to the articular surface would be the optimal surgical technique during total knee replacement. However, this assumes that trabecular orientation for dtffe rent types of arthritis (e.g. osteoarthritis, rheumatoid arthritis, etc.) is the same as that fund for "normal" cadaveric specimens. Do the authors have any stereoscopic analyses to support this assumption? Authors: It is well established in the world literature that osteoarthritis is a focal disease within articular cartilage and subchondral bone. The surgeon commonly resects away these regions of pathology during total joint arthroplasty. We are not aware of any data which might suggest there are correlated changes in the trabecular orientation with osteoarthritis. This requires future investigations.

Rheumatoid arthritis is a systemic autoimmune disease in which the disease and pharmaceutical therapies are both known to affect the load carrying capabilities of bone. To our knowledge, there are no reported studies which demonstrate changes in trabecular orientation with this disease. This also would be worth investigating in the future using accurate quantitative techniques as described in this study.

S. B. Goodman: The authors examined the most medial portion of the medial plateau of the tibia for this study. Do the authors believe that their conclusions are applicable to other portions of the proximal tibia? Authors: We have anticipated this question and are currently completing the analysis of a 3-dimensional mapping study of the resected proximal tibia. We will

Page 9: Stereoscopic Analysis of Trabecular Bone Orientation in ...

K.N. Bachus, Ph.D., M.K. Harman, M.S. and R.D. Bloebaum, Ph.D.

be reporting the results of our studies of the entire tibial wafer as well as "maps" at incremental depths throughout the proximal tibia. We are interested in determining how the "parallel-cut versus perpendicular­cut" comparison affect trabecular orientation distal to the tibial articular surface. This would be especially important when the surgeon is dealing with revision arthroplasty.

K. Draenert: There is a hypothesis which claims that bone is a shell around an inner pressure system -- like a shell of an egg. What are the authors trying to say with respect to the slope of the shell? Authors: Assuming the trabecular bone is like a "shell" around an inner pressure system, the orientation of the resected "shells" would be angled as a unit if an axis­perpendicular resection technique is used. This places the entire "shell" in a bending moment and decreases its load carrying capability. The "inner hydraulic pressure system" would not aid in the load carrying capabilities since mechanical testing has shown that trabecular bone is not hydraulically strengthened by the presence of marrow under physiological loading conditions [6]. The goal of this study, was to develop a technique which would allow measurement of trabecular orientation or outer "shell" in three dimensions. This was accomplished.

K. Draenert: It is important to map the measurements and show the topography of the select area. Authors: We agree. We are currently completing the mapping phase of our investigations and will be reporting this data within a few months.

K. Hodde: Does the difference in measured and calculated mean tilt angles of the trabeculae to the cut surface agree with the differences in resection angles between the two main surgical techniques? Authors: The results of this study represents the trabecular orientation of the complete medial side of the resected specimens. When the data is averaged on final analysis, it represents a broad range of trabecular orientations. This is clinically pertinent since the tibial component is supported by the entire plateau represented by the average trabecular orientation.

20

T. A. W. Gruen: It is not known if the observer knew the nature of the specimen, as this may have led to the high percentage of zero readings in Table 1. Authors: This is true. Observer bias can skew the data. However, at the onset of this study, we developed measurement criteria which we hoped would minimize this problem. Using the parallax technique described, it is difficult to bias the data collection process. However, future mapping studies will use a randomized sampling method to help guarantee that observer bias does not occur.

T.A. W. Gruen: The precise region of interest within the medial third of the 5 mm thick wafer of cancellous bone was not specified in this study. This is critical as there are substantial variations within the medial condyle region. Authors: We recognize these variations across the tibial plateau. For this reason , a second study is being completed which maps the angular variations throughout the proximal tibia at specific, anatomic regions. However, the purpose of this present study, was mainly to develop and demonstrate the application of the technique.

T. A. W. Gruen: What is known about the inherent variation between the left and right sides from the same pair of normal specimens? Au th ors: There are numerous factors which could influence the orientation of the trabeculae. From the biomechanical literature, left/right comparisons are common. The literature base for trabecular orientation is, at this time, sparse. To try and avoid any bias in this study, we used a coin toss to randomly select which tibia of the pair would be cut perpendicular and which would be cut parallel.