SMARTMIX CEMVAC ® Dual Vacuum Footpump + Air Hose Produces a nominal vacuum of 0.66 bar (495 mmHg) for bowl setting and 0.92 bar (690 mmHg) for syringe setting. Stepped Outlet Port The dual diameter outlet port allows connection of the vacuum system tubing. Vacuum Gauge The clear and easy to read gauge shows that a vacuum is being generated. The green and green/yellow bands around the gauge show the user that the pump is operating at the correct vacuum level. Latching Foot Switch The pump is turned on and off via a latching foot switch. This means that the user can depress the pedal once to turn on and once to turn off and it is not necessary to keep the foot on the pedal during pump operation. The metal foot plate has a textured surface for grip. Air Hose The air hose is a standard compressed air hose meeting BS 5682-3. 1 STEPS FOR USE ® Connect the hose to the footpump. 1 Connect the other end of the hose to a supply of compressed air at 5-8 bar. 2 Select either syringe or bowl setting, by turning the device selector switch and clicking into the correct position. 3 Connect filter of the mixer vacuum tubing to the outlet on the front of the pump. 4 Turn on the pump by depressing and releasing the foot pedal. Block the outlet of the pump, for example by placing finger over the end. The green and green/yellow bands indicate that the pump is functioning correctly in bowl and syringe settings respectively 5 7 When the mixing stage is complete, turn off the footpump by depressing and releasing the foot pedal. 8 At the appropriate point in the mix, turn on the footpump by depressing and then releasing the foot pedal. 6 The green and green/yellow bands indicate that the pump is functioning correctly in bowl and syringe settings respectively. DSEM-BIO-0614-0011 SmartMix Cemvac Dual Vacuum Footpump.indd 1 27/04/2015 14:12

Transcript of STEPS FOR USE VACU-MIX Low Vacuum Footpump + Air Hose 1 ...synthes.vo.llnwd.net/o16/LLNWMB8/INT...

Page 1: STEPS FOR USE VACU-MIX Low Vacuum Footpump + Air Hose 1 ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · VACU-MIX® Low Vacuum Footpump + Air Hose Produces a nominal vacuum

SMARTMIX™ CEMVAC® Dual Vacuum Footpump + Air HoseProduces a nominal vacuum of 0.66 bar (495 mmHg) for bowl setting and 0.92 bar (690 mmHg) for syringe setting.

Stepped Outlet PortThe dual diameter outlet port allows connectionof the vacuum system tubing.

Vacuum GaugeThe clear and easy to read gauge shows that a vacuum is being generated. The green and green/yellow bands around the gauge show the user that the pump is operating at the correct vacuum level.

Latching Foot SwitchThe pump is turned on and off via a latching foot switch. This means that the user can depress the pedal once to turn on and once to turn off and it is not necessary to keep the foot on the pedal during pump operation. The metal foot plate has a textured surface for grip.

Air HoseThe air hose is a standard compressed air hose meeting BS 5682-3.1

STEPS FOR USE

Stepped Outlet PortThe dual diameter outlet port allows connection of the VACU-MIX® system tubing.

VACU-MIX® Low Vacuum Footpump + Air HoseProduces a nominal vacuum of 0.40 bar (300 mmHg).

Vacuum GaugeThe clear and easy to read gauge shows that a vacuum is being generated. The green band around the gauge shows the user that the pump is operating at the correct vacuum level.

Latching Foot SwitchThe pump is turned on and off via alatching foot switch. This means that the user can depress the pedal once to turn on and once to turn off. The metal foot plate has a textured surface for grip.

Air HoseThe air hose is a standard compressed air hose meeting BS 5682-3. [1]

STEPS FOR USE

1

Connect the hose to the footpump.

Connect the other end of the hose to a supply of compressed air at 4-8 bar.

2

Connect filter of the mixer vacuum tubing to the inlet on the front of the pump.

3

At the appropriate point in the mix, turn on the footpump by depressing and then releasing the foot pedal.

4

When the mixing stage is complete, turn off the footpump by depressing and releasing the foot pedal.

5

®

Connect the hose to the footpump.

1

Connect the other end of the hose to a supply of compressed air at 5-8 bar.

2

Select either syringe or bowl setting, by turning the device selector switch and clicking into the correct position.

3

Connect fi lter of the mixer vacuum tubing to the outlet on the front of the pump.

4

Turn on the pump by depressing and releasing the foot pedal. Block the outlet of the pump, for example by placing fi nger over the end. The green and green/yellow bands indicate that the pump is functioning correctly in bowl and syringe settings respectively

5

7

When the mixing stage is complete, turn off the footpump by depressing and releasing the foot pedal.

8

At the appropriate point in the mix, turn on the footpump by depressing and then releasing the foot pedal.

6

The green and green/yellow bands indicate that the pump is functioning correctly in bowl and syringe settings respectively.

DSEM-BIO-0614-0011 SmartMix Cemvac Dual Vacuum Footpump.indd 1 27/04/2015 14:12

Page 2: STEPS FOR USE VACU-MIX Low Vacuum Footpump + Air Hose 1 ...synthes.vo.llnwd.net/o16/LLNWMB8/INT Mobile/Synthes... · VACU-MIX® Low Vacuum Footpump + Air Hose Produces a nominal vacuum

Ordering Information

SMARTMIX CEMVAC Hardware

3210026

SMARTMIX CEMVAC Dual Vacuum Footpump + Air Hose(SMARTMIX CEMVAC Dual Vacuum Footpump cannot be purchased separately)

3210035

Air Hose for Vacuum Footpump(Available as a spare part)

References

1. BS5682: 1998 Specification for probes (quick connectors) for use with medical gas pipeline systems. Certification on file with Summit Medical Ltd.

CA#DSEM/BIO/0614/0011 Issued: 10/14

This publication is not intended for distribution in the USA.

The CE mark relates to theadvertised product 3210026

NONSTERILE

0086 Summit Medical Ltd.Industrial ParkBourton on the WaterGloucestershire GL54 2HQTel: +44 (0) 1451 821311

DePuy International, Ltd.St Anthony’s RoadLeeds LS11 8DTTel: +44 (0) 113 270 0461

Manufactured by: Distributed by:

BIOMATERIALSSOLUTIONS BY DEPUY SYNTHES

Legal Manufacturer Accessory

Summit Medical LtdAirhose for Vacuum Foot Pump

Airhose for Vacuum Foot Pump (Square Connector)

©Johnson & Johnson Medical Limited. 2014. All rights reserved.

Johnson & Johnson Medical Limited PO BOX 1988, Simpson Parkway, Livingston, West Lothian, EH54 0AB, United Kingdom.Incorporated and registered in Scotland under company number SC132162.

depuysynthes.com

DePuy (Ireland)LoughbegRingaskiddyCo. CorkIrelandTel: +353 21 4914 000 Fax: +353 21 4914 199

DePuy Orthopaedics, Inc. 700 Orthopaedic DriveWarsaw, IN 46582USATel: +1 (800) 366 8143Fax: +1 (574) 267 7196

DePuy International LtdSt Anthony’s RoadLeeds LS11 8DTEnglandTel: +44 (0)113 270 0461 Fax: +44 (0)113 272 4101

DSEM-BIO-0614-0011 SmartMix Cemvac Dual Vacuum Footpump.indd 2 27/04/2015 14:12