    Instruction Manual Ver1.1

    STYRAX INSTRUMENTS INDIA PVT LTD

    INSTRUCTION MANUALPASS BOX CONTROLLER DYNAMIC

    MODEL NO: SPB105-S

    VERSION: 1.1

    Styrax Instruments India Pvt. Ltd.PLOT No.48-51,Dollapally,

    Phase V Extn. I.D.A.,Jeedimetla, Hyderabad 055

    Andhra Pradesh, INDIA

    Ph: +91 40 2309 3314Fax: +91 40 2309 3315

    www.styraxinstruments.com

    Email: [email protected]

  • 8/6/2019 Static Pass Box Controller Manual

    2/8

    Instruction Manual Ver1.1

    Instrument SI. No :

    STANDARD PACK

    1) Instruction Manual --- 1 No

    2) DPB105 Instrument --- 1 No

    Checked By :

    Date :

    1.Description:

    Pass box is used for material transfer from one Clean room to another clean room. It consists of a

    rectangular box with two doors. The two doors are interlocked such that two doors cannot be open at the same

    time, to prevent pressure losses and cross contamination.

    Our Pass Box controller achieve this by preventing simultaneous access of door by ensuring that when one

    door for a Pass box is open then other door is closed.

    before opening of door.

    1. When Door 1 is opened the display on the Door 2 will indicate No Go and at the Door 1 the display will

    indicate Door Open

    2. When any of the Door is kept open beyond a particular time say 10 min, a buzzer will be blown and the

    display at the doors will indicate Close Door

    3. When any door is opened the CF lamp is switched on, and the UV lamp will be switched off

    4. Both Doors are provided with EM Locks

    2. Features:

    Advanced Micro controller based.

    Supports Maximum of 2 doors

    Supports Electro magnetic locks/Electric strikers

    Door status indication with high intensity Bar LEDs

    Status indication: GO, WAIT, DOOR OPEN, CLOSE DOOR

    Buzzer output for door is kept open more than specified time from 2 to 30 min.

    UV Lamp controlling

    o UV Lamp ON for both Doors closed

    o UV Lamp OFF for both Doors Open

  • 8/6/2019 Static Pass Box Controller Manual

    3/8

  • 8/6/2019 Static Pass Box Controller Manual

    4/8

    Instruction Manual Ver1.1

    5. Status Indication on Display Panel :

    GO = Indicate that person can access the door.

    WAIT = Indicate that person cannot access the door. DOOR OPEN = Indicate Door is opened

    CLOSE DOOR = Indicates that the door is opened more than specified time.

    6. OPERATING MODES:

    MODE1: Normally Closed ( Default mode)MODE2: Normally Open

    MODE1:

    *Locking device is Electro magnetic lock or Electric striker.

    Normally Close:

    In this mode, initially both doors in locked condition. Door access by using push button. Status LEDs providedon the Display module indicates whether person can access the door or not.GO = GO LED glow indicates that person can access the door.WAIT = WAIT LED glow indicates that person cannot access the door.CLOSE DOOR = CLOSE DOOR LED indicates that the door is opened more than specified time (default time 2

    min). When the door operated from open to closed condition, both doors locked for 4 sec, with in this time person

    cannot access both doors. Door Buzzer time select from 2 - 30 min. Using jumpers (Refer Section 7.4)

    Auto close time delay i.e. the time between the door release and energizing is 4 Sec

    MODE2:

    *Locking device is Electro magnetic lock or Electric striker.

    Normally Open:

    In this mode, initially both doors in un locked condition; here push buttons are used only for indication purpose

    (no need to press the button). Status LEDs provided on the push button station indicates whether person canaccess the door or not.GO = GO LED glow indicates that person can access the door.WAIT = WAIT LED glow indicates that person cannot access the door.

    CLOSE DOOR = CLOSE DOOR LED indicates that the door is opened more than specified time (default time 2min). When the door operated from open to closed condition, both doors locked for 4 sec, with in this time person

    cannot access both doors. Door Buzzer time select from 2 - 30 min. Using jumpers (Refer Section 7.4)

    Auto close time delay i.e. the time between the door release and energizing is 4 Sec

  • 8/6/2019 Static Pass Box Controller Manual

    5/8

    Instruction Manual Ver1.1

    7. Wiring diagram:

    C UV CF LNEEML1-EML1+ FB1EML2- EML2+ FB2

    FBFB+- FBFB+-

    DOOR2 EM LOCK DOOR1 EM LOCKPower Supply

    230V AC, 50Hz

    CFLAM

    P

    UVLAM

    P

    L

    N

    1. Connect Door1 Display Module with Door1 Display Module Cable

    2. Connect Door2 Display Module with Door2 Display Module Cable

    8. MECHANICAL DRAWINGS

    Fig (8.1) Figure for Installation of SPB105C-S Controller:

    STYRAXINSTRUMENTS

    INDIA PVT LTD

    STATICPASSBOX CONT

    ROLLER

  • 8/6/2019 Static Pass Box Controller Manual

    6/8

    Instruction Manual Ver1.1

    Fig (8.2) Front View Diagram of SPB105C-S Controller:

    STYRAX INSTRUMENTS INDIA PVT LTD

    STATIC PASS BOX CONTROLLER

    230 mm

    287mm

    Fig (8.3) Figure for Installation of SPB105D Display Panel:

    Pass Box

  • 8/6/2019 Static Pass Box Controller Manual

    7/8

    Instruction Manual Ver1.1

    Fig (8.4) Front View Diagram of SPB105D:

    45mm

    185 mm

    OPEN

    SPB105

    9. Ordering Information:

    I). Pass Box Controller

    includes,

    a. Controller:

    Variants:

    SPB105 X Model. No.

    SPB105C S SPB105C-S

    SPB105C D SPB105C-D

    b. Pass Box Display Module:

    Variants:

    SPB105D X Model. No.

    SPB105D X SPB105D -X

    X

    Model No Pass Box TypeS- Static

    SPB105D- Dynamic

    X

    Model No Pass Box Type

    S- StaticSPB105C

    D- Dynamic

    SPB105D X

    Model No Future Use

    SPB105D X

  • 8/6/2019 Static Pass Box Controller Manual

    8/8

    Instruction Manual Ver1.1

    NOTE:

    For Further Information Contact:

    Styrax Instruments India Pvt. Ltd.PLOT No.48-51, Dollapally, Phase V Extn. I.D.A., Jeedimetla, Hyderabad 055

    Andhra Pradesh, INDIAPh: +91 40 2309 3314, Fax: +91 40 2309 3315

    www.styraxinstruments.com

    Email: [email protected]