Standardization and Applications Multicolor Imunophenotyping Multicolor Immunophenotyping Dr Shanaz...
-
Upload
nyah-judson -
Category
Documents
-
view
237 -
download
2
Transcript of Standardization and Applications Multicolor Imunophenotyping Multicolor Immunophenotyping Dr Shanaz...
Standardization and
Applications
Multicolor Imunophenotyping
Multicolor Immunophenotyping
Dr Shanaz KhodaijiConsultant Hematology
P.D.Hinduja National HospitalMumbai
History of FC at the PDHNH & MRC
FACSCalibur in 1995FACSCanto II in 2008•CD4 and CD8 counts•Detailed Lymphocyte Subset analysis•HLA-B27 •Immunophenotyping of Acute Leukemias and CLPDs •Platelet antigen studies
Multicolor Imunophenotyping
Calibration using 7-colour setup beads
Prior to sample acquisition, the FC is set up using 7-Colour
setup beads. Controls need to address 3 issues
Detectors, Thresholds, and Spectral Overlap.
Beads are used
• To adjust Detector Voltages: this ensures that fluorescence brightness is correct for stained cells in each detector.
• Adjust the signal for events displayed in plots by changing detector voltages. Higher voltages amplify the signal. Lower voltages decrease the signal.
• BD FACSCanto clinical software recalculates spectral overlap when you change detector voltages.
Multicolor Imunophenotyping
BD FACS 7-colour setup beads
Adjusting Thresholds
• A threshold sets a channel number below which events will
not be processed. Use threshold to filter out unwanted
events. You can set one or more thresholds at a time, and
choose whether any one or all need to be met.
Adjusting Spectral Overlap
• Set fluorescence compensation. Fluorochromes emit light
over a range of wavelengths. During cytometer setup,
fluorescence spillover is automatically determined and
corrected. If necessary, you can use the spectral overlap
controls to make manual adjustments.
Multicolor Imunophenotyping
Monitor daily instrument performance
• The LJ Chart for PMT Voltages and Blue and Red laser current are to be reviewed on a monthly basis to monitor Cytometer performance and see shifts or trends in parameters as they occur. The results are to be signed by the consultant and filed.
• For the HLA B27 and Lymphocyte subset analysis the Lyse/No wash setup is used and for leukemia/lymphoma Immunophenotyping a lyse/wash setup is used.
Multicolor Imunophenotyping
Multicolor Imunophenotyping
Cytometer setup report
Multicolor Imunophenotyping
Optimization – why?
Optimizing Cytometer Settings• When you performed cytometer QC, voltage settings
were adjusted to set each parameter at a target value. These settings might not be appropriate for the stained samples you plan to analyze. Before recording data, adjust FSC, SSC, and threshold settings; gate on the population of interest (such as lymphocytes) and adjust voltages to optimize fluorescence signal.
Multicolor Imunophenotyping
Reference range for Lymphocyte subsets
CD4 Ab/%
CD4Ab/%
CD8 Ab/%
CD8 Ab/%
CD3 Ab/%
CD3 Ab/%
CD19 Ab/%
CD16+56
CD16+56 Ab/%
Mean 790/ 34
838/ 35
821/35
642/ 27
1612/70
1590/67
353/ 15
419/ 17
Minimum
355/14
401/ 23
245/13
243/ 13
760/49
796/ 49
125 133/7
115/6
Maximum
1960/62
1451/48
1876/62
1206/41
2787 /84
2679/80
707 714/28
1009/37
Multicolor Imunophenotyping
Reference ranges in children Surg Cdr Gaurav Narula, Dr Shanaz Khodaiji, Col M.S. Bindra
Multicolor Imunophenotyping
Age GroupCD4
Range (Mean)
CD8
Range (Mean)
CD3
Range (Mean)
CD 16/56
Range (Mean)
CD 19
Range (Mean
Cord Blood
2209- 3205 (1707) 312- 1360 (836)3100- 5200
(2859)--- ---
0-6 mo 1516-4348
(2932)
970- 2118 (1544)
1767- 5495 (3631)
5.8- 35.8 (15.8)20- 77*
(405)
7-12 mo 1056- 3799
(2427)
541- 2807
(1133)
1278- 4710
(2994)
7.8- 89
(47.6)
25- 1852*
(790)
13-36 mo1113- 2946
(2029)
523- 2015
(1269)
1222- 3790
(2506)
54.5- 261.5
(103.5)
15- 2445
(563)
37-60 mo 839- 3115
(1977)
749- 1997
(1373)
1368- 3812
(2590)
53- 1160
(556)
4- 10997*
(741)
Age Our study (mean) Kotylo et al Denny T et al#
Cord blood 2209-3205 (1707) 2310-2990* 1460-5110
0-6m 1516-4348 (2932) 2800-3900 1690-4610
7-12m 1056-3799 (2427) 2600-3500 --------------
13-36m 1113-2946 (2029) 1200-2300 1020-3600
37-60m 839-3115 (1977) 560-2700 900-2860
Comparative study of CD 4 counts (x 103/ml)
Multicolor Imunophenotyping
Role of flow cytometric evaluation of lymphocyte subsets in predicting acute rejection episodes in renal transplant pts A. Dasgupta, S.Khodaiji et al
Aim of the study• To study short term results of renal transplant using low dose (1 ml/day) anti CD3 monoclonal antibody induction• to examine role of flow cytometrically determined lymphocyte counts in predicting AREConclusion• Low dose Mab significantly decreases the CD3 count & CD4:CD8 ratio•FC allows monitoring of lympho subsets even at very low counts• AREs tend to be milder and later with use of Mab induction •CD3 counts< 10% and CD4:CD8 < 0.11 of baseline decrease the risk of ARE
Multicolor Imunophenotyping
Flow cytometric analysis helps in the diagnosis of dense granule deficiency and qualitative deficiency of GPIIB/IIIA in platelets
Sheeba Abraham & Amar Das Gupta
Multicolor Imunophenotyping
To correlate aggregometry & flow cytometry
findings
To evaluate the role of FC in diagnosis PFD
Flow cytometry helped in distinguishing between quantitative (5 cases) and qualitative (1 case) defects of platelet GpIIb/IIIa in GT patients In 4 patients with bleeding diathesis but normal aggregation response, the diagnosis of SPD could be made from flow cytometric demonstration of reduced mepacrine uptake Reduced P-selectin (CD62P) expression by flow cytometry was seen as an isolated defect in 6 patients with near normal aggregation response FC analysis of platelet structure & function supplements information obtained by aggregometry and helps in further diagnosis & sub classification of platelet function disorders
Results and observations
Multicolor Imunophenotyping
Flow cytometric estimation of CD41, CD61 for GT
Multicolor Imunophenotyping
Comparison of platelet counts by Sysmex XE-2100 (I and O) and LH 750 with the International Flow Reference method in thrombocytopenic patients
Multicolor Imunophenotyping
Analysis of Sysmex reported, Sysmex Impedence, Sysmex Optical and LH750 Methods with the IRM
at Different Transfusion Thresholds
Multicolor Imunophenotyping
PNH projectStandardisation of Six colour Flowcytometry assay for the diagnosis and monitoring of Paroxysmal Nocturnal Hemoglobinuria using FLAER, as per 2010 International Guidelines
FLAER CD24PE
CD45 Per-CP
CD33 PECy7
CD15 APC
CD14 APC- Cy7
PNH marker
NeutroPNH marker
Gating marker
Mono Gating marker
Neutro Gating marker
MonoPNH marker
RBCs- CD55/CD59/CD235a
Multicolor Imunophenotyping
MRD for B ALL• MRD assay Standardisation based on the COG protocol
• Normal samples/Non B ALL post induction Bone marrows /Staging marrows shall be used for making normal templates
• Dilutional studies shall be done for linearity assessment and establishing cutoffs
FITC PE PerCP-Cy5.5
PE CY7
APC APC-H7
Tube 1 CD20 CD10 CD38 CD19 CD58 CD45
Tube 2 CD9 CD13+33 CD34 CD19 CD10 CD45
Tube 3 SYTO 16 CD33 CD3 CD19 CD71 CD45
Tube 4 CD15 CD24 CD20 CD19 CD10 CD45
Multicolor Imunophenotyping
Comparison of FC analysis on LNs with histopathologic study to diagnose NHL
Why should we do it?
FCM facilitates the analysis of cells within discrete subpopulations defined and selected (gated) based on other parameters, allowing a valid and reliable diagnosis, especially in NHLs and enabling their subclassification.
FCM is faster than the histopathologic examination, allowing for therapeutic decisions to be made quickly.
Allows a clear-cut correlation of multiple measurements (antigen expressions, DNA content, light scatter) in individual cells.
Multicolor Imunophenotyping
• Conducted workshops
and seminars
•Training programs
•Research projects
•ILCP
•CAP