Standard Data Analysis -Calibration

ChemStation – Level 1 - Training

Page 2: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

1. Inject standards

2. Set-up appropriate integration events for the low standard.

3. Begin a new calibration table with level one, the lowestlevel. Fill in the dialog.

4. Integrate each additional standard level and add to the calibration table.

5. Examine the calibration curve and save it to the method.

Building a Calibration Table

Page 3: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Example Calibration

1. Standard 30.0 ISTD 50.02. Standard 50.0 ISTD 50.03. Standard 75.0 ISTD 50.0

ng/L ng/L

3 Levelsmin1 2 3 4 5 6

mAU

0

20

40

60

80

100

DAD1 A, Sig=254,4 Ref=550,100 (DEMO\005-0101.D)

0.7

47

1.0

21

2.5

65

5.8

37

Standard ISTD

Page 4: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Define signals to be evaluated during a method.

Signal Details

Page 5: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Calibration Settings

Page 6: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Load the Signal(s) for the Low Level Standard and Integrate

Page 7: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Create a Calibration Table: Level One

Page 8: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Fill in:Compound NameGroupAmount

Check if:Reference PeakISTDIndicate #

Calibration Table Overview: Fill in the Dialog

Page 9: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Load the Signal - Add Level Two

Page 10: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Add Additional Levels

Page 11: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Compound, Amt, Area, Rsp Factor,Ref, ISTD#

Compound, Grp, Amt, Low LimitHigh Limit

Compound, Amt, Area, Def, CurveType, Origin

Compound, Area, Rsp%, Def, +-, Pk Usage

Calibration Table Options

Page 12: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Use Low and High Limits as qualifiers to confirm peak

identification.

Compound Details

Page 13: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Check Def to apply thedefault curve type and origin from the Calibration Settingsdialog box.

Curve TypesPiece wise ExponentLinear QuadraticLog CubicPower Average Rsp/Amt

OriginIgnoreIncludeForceConnect straight segment

Peak Details

Page 14: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Define the columns included in your Calibration Table

Edit Calibration Table Options

Page 15: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Calibration TaskTool New Calibration

Table

Recalibrate withCurrent Chromatogram

Add New Level Signal Details

Add New Peaks

Select Calibrated peak(s) fromChromatogram and delete fromcalibration table.

Select peak(s) from chromatogram and addto calibration table

Select compoundsfrom chromatogram and recalibrate

Edit CalibrationTable Options

Quantification Tools

Page 16: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Save Table to Method

To Save the Calibration Table as Part of a Method...

Page 17: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Signal Options

Customized display on screen and in reports

Page 18: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Annotations

Use it for a customized chromatogram in slides or publications

Easy placing and moving of annotations

Selectable fonts and colors

The Annotation sidebar can be activated within all toolsets

Page 19: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Signal Manipulations

Overlay, subtract and align signals

Page 20: Standard Data Analysis - Calibration ChemStation – Level 1 - Training.

Lab Time !!!! Lab 7 - Calibration